BIO RAD PROTEAN II xi CELL Setup Manual

This kit allows 18.5 cm wide x 20 cm long gels to be run on Bio-Rad’s PROTEAN II xi Cell thus accommodating commercially available immobilized pH gradient gel strips for 2-D second dimension protein separations. It contains:
• Wide sealing gaskets for the upper buffer chamber (2)
• 20 cm glass plates (2 sets)
• 20 cm notched clamps (2 sets)
• Foam gaskets for the PROTEAN II xi Casting Stand (2)
• Alignment card
• Narrow spacers (4) *
* 1 mm or 2 mm thick depending upon which kit you ordered.
Casting Gels
Gels are cast the same way as with the standard PROTEAN II xi Cell (refer to your user manual for details). For best results, replace the solid rubber gaskets on the PROTEAN II xi Casting Stand with the foam rubber gaskets included with this kit. They provide a more reliable seal when using the narrow spacers. When assembling the gel “sandwich”, use the alignment card to push the spac­ers to the extreme outer edges of the glass plates to ensure maximum gel width. Since the card is narrower than the gel cavity, you will need to slide it side-to-side to properly align the spacers.
After pouring the monomer solution, insert the comb into the gel. Using combs ensures gel-to-gel dimensional repeatability. This is especially true if using the PROTEAN II xi Multi-Gel Casting Chamber. Overlaying gels with t-amyl alcohol can be done, but unless it is applied simultaneously to all gels, the gel heights will not be the same because the weight of the overlay applied to the first gel (and each subsequent gel) will displace the solutions in the other gels.
The combs provided in this kit have two sides—one with a reference well for standards and one without. The side without the stan­dard lane allows for commercially available 18 cm IPG strips to be used without excess trimming. The side with the end reference well allows standards to be run along side your sample. However, 18 cm IPG’s will need to be trimmed more to fit into the small­er prep well. Deciding which comb side to use depends upon what you know about your sample. When running without the refer­ence well, an internal standard can still be included by adding a known protein to your sample. Providing a marker between gels helps to monitor the reproducibility of the 2-D patterns and makes post run gel comparisons easier. To determine which standard mark­er to use, run a 1-D SDS gel of your unknown sample.
Assembling the Upper Buffer Chamber
Replace the standard U-shaped gaskets on the central cooling core with wider U-shaped sealing gaskets provided with this kit. (Refer to section 5.1 in the PROTEAN II xi Cell manual.)
Once the gels have polymerized, remove the gel sandwich assemblies from the casting stand, remove combs, and mount the gel sand­wich assemblies onto the central cooling core to create the upper buffer chamber.
If running one gel, a buffer dam can be mounted on the opposite side of the core. If you do not have a buffer dam, clamp an inner and outer glass plate together without spacers and mount them on the central cooling core. Note that either U-shaped gasket size
will work with the buffer dam or glass plate set.
PR OTEAN®II xi Cell IPG Con v ersion Kit
Setup Guide
Loading the IPG strip onto the Slab Gel
Equilibrate the IPG strip in second dimension equilibration solution as you normally would before placing them onto the second dimension vertical gel. Using a pair of forceps, manually slide the IPG strip onto the slab gel such that one edge of the IPG strip abuts against the top of the slab gel.Secure the IPG in place by overlaying it with molten agarose containing 1% electrode buffer (refer to Section 14.1, solution C, in the PROTEAN II xi user manual for the recipe). It is important that no bubbles be trapped between the two gels.
Running the 2nd Dimension Gel
Electrophorese the gels until the bromophenol blue marker dye runs to the bottom of the gel. Upon completion of the run, remove and fix the gels as soon as possible as described in section 9 of the PROTEAN II xi Cell user manual.
4106107 Rev B
Ordering Information
Catalog Number Product Description
165-3183 PROTEAN II xi Cell IPG Conversion Kit, 1 mm Gel* 165-3184 PROTEAN II xi Cell IPG Conversion Kit, 2 mm Gel* 165-1834 PROTEAN II xi Basic Unit with casting stand** 165-1835 IPG Clamp Set (1R and 1L) 165-1836 IPG Spacers, 1 mm, 4 165-1837 IPG Spacers, 2 mm, 4 165-1838 IPG 2-D Prep Comb, 1 mm, 1 165-1839 IPG 2-D Prep Comb, 2 mm, 1 165-3182 IPG Sealing Gasket, 2 165-1823 Inner Glass Plate, 20 cm x 20 cm, 2 165-1824 Outer Glass Plate, 22.3 cm x 20 cm, 2 165-3181 IPG Spacers, 1.5 mm, 4 165-3186 PROTEAN II xi Cell IPG Conversion Kit, 1.5 mm Gel* 165-3187 IPG 2-D Prep Comb, 1.5 mm, 1
*
Kit includes 2 sets of IPG clamps, 2 sets of glass plates, 4 IPG spacers, 2 IPG 2-D/prep combs, 2 IPG gaskets, one alignment card, and two casting stand gaskets
**
The PROTEAN II xi Basic Unit does not include spacers, combs, gaskets, or plates. Combine it with an IPG Conversion kit for a complete system.
See Bio-Rad catalog for standard PROTEAN II xi Cell & Multi-Gel Cell accessories.
Bio-Rad Laboratories
Life Science Group
Website www.bio-rad.com Bio-Rad Laboratories Main Office 2000 Alfred Nobel Drive, Hercules, CA 94547, Ph. (510) 741-1000, Fx. (510)741-5800 Also in: Australia Ph. 02-9914-2800, Fx. 02-9914-2889 Austria Ph. (1)-877 89 01, Fx. (1) 876 56 29 Belgium Ph. 09-385 55 11, Fx. 09-385 65 54
Canada Ph. (905) 712-2771, Fx. (905) 712-2990 China Ph. (86-10) 2046622, Fx. (86-10) 2051876 Denmark Ph. 39 17 9947, Fx. 39 27 1698 Finland Ph. 90 804 2200, Fx. 90 804 1100 France Ph. (1) 43 90 46 90, Fx. (1) 46 71 24 67 Germany Ph. 089 31884-0, Fx. 089 31884-100 Hong Kong Ph. 7893300, Fx. 7891257 India Ph. 91-11-461-0103, Fx. 91-11-461-0765 Israel Ph. 03 951 4127, Fx. 03 951 4129 Italy Ph. 02-21609.1, Fx. 02-21609.399 Japan Ph. 03-5811-6270, Fx. 03-5811-6272 The Netherlands Ph. 0313 18-540666, Fx. 0313 18-542216 New Zealand Ph. 09-443 3099, Fx. 09-443 3097 Singapore Ph. (65) 272-9877, Fx. (65) 273-4838 Spain Ph. (91) 661 70 85, Fx. (91) 661 96 98 Sweden Ph. 46 (0) 8 627 50 00, Fx. 46 (0) 8 627 54 00 Switzerland Ph. 01-809 55 55, Fx. 01-809 55 00 United Kingdom Ph. 0800 181134, Fx. 01442 259118