Beurer FT 95 User Manual

ENGLISH
Non-contact Thermometer
Instruction manual ����������������� 3
ESPAÑOL
Termómetro sin contacto
Manual de instrucciones ����� 28
T
hermomètre sans contact
Mode d’emploi ����������������������54
FT95
Distributed by/Distribuido por/ Distribué par: Beurer North America LP 1 Oakwood Boulevard, Suite 255 Hollywood, FL 33020, USA www.beurer.com
Questions or Comments? Call our US-based customer service toll free at 1-800-536-0366�
¿Preguntas o comentarios? Llame a nuestro servicio al cliente en Estados Unidos gratis al teléfono 1-800-536-0366�
Questions ou commentaires? Appelez gratuitement notre service clientèle situé aux États-Unis au 1-800-536-0366�
READ THIS MANUAL COMPLETELY AND
CAREFULLY BEFORE USING THIS PRODUCT
Keep this manual in a safe location for future reference
LEA TODO ESTE MANUAL CON ATENCIÓN
ANTES DE USAR ESTE PRODUCTO
Conserve este manual en un lugar seguro para consultarlo en el futuro
LISEZ CE MODE D’EMPLOI COMPLÈTEMENT ET ATTENTIVEMENT
Conservez ce manuel en lieu sûr pour y faire référence ultérieurement
AVANT D’UTILISER CE PRODUIT
2
ENGLISH
Contents
1. Important safety notes .................................. 4
2. Getting to Know your Instrument .................. 9
3. Package Contents ....................................... 11
4. Parts and Controls ..................................... 12
5. To Set Up ..................................................... 14
6. To Use ......................................................... 15
SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE
If the instruction manual is damaged or if you no longer have the instruction manual in your possession, please contact Beurer customer service� Please see warranty for service contact�
7. Care, Maintenance and Disposal .............. 20
8. Troubleshooting Guide .............................. 21
9. Technical Specification..............................22
10. FCC Compliance Information ................... 23
11. Warranty .................................................... 24
3
Dear Valued Customer,
Thank you for choosing one of our products. Our name stands for high-quality, thoroughly tested products for application in the areas of heating, weight management, blood pressure, body temperature, pulse, gentle therapy, massage, and beauty.
Register your product today to receive further benefits:
• added protection for your purchase
• health and wellness tips
• new product information
• continual outstanding customer service and support. Visit www.registerbeurer.com or contact customer service at 1-800-536-0366. Please read these instructions for use carefully and keep them for later use. Make them accessible to other users and observe the information they contain.
With kind regards, Your Beurer Team
1� Important safety notes
Signs and symbols
Whenever used, the following signs identify safety and property damage messages and designate a level of hazard or seriousness.
READ THIS ENTIRE MANUAL INCLUDING THE SAFETY SECTION AND ALL INSTRUCTIONS AND WARNINGS COMPLETELY AND CAREFULLY BEFORE USING THIS PRODUCT� FOLLOW ALL SAFETY INSTRUCTIONS AND
4
WARNINGS TO AVOID HAZARDOUS SITUATIONS AND TO MAKE CORRECT
Operating
USE OF THIS PRODUCT�
This is the safety alert symbol. It alerts you to potential personal injury hazards. Obey all safety messages that follow this symbol to avoid possible injury or death.
Indicates a hazardous situation which, if not avoided, may result in minor or
CAUTION
NOTICE
moderate injury.
Addresses practices not related to personal injury, such as product and/or property damage.
Note
Note on important information
Type BF applied part
Follow instructions for use
Permissible operating temperature and humidity
5
Storage/Transport
Permissible storage temperature and humidity
Serial number
IP 22
w
Device protected against foreign objects ≥12.5 mm and against water dripping at an angle
Cross-reference with another section.
CAUTION - To reduce the risk of personal injury or product/property damage:
• This device is designed solely for non-commercial, household use in non-invasively measuring temperatures via an infrared sensor of the human body for persons of all ages, as well as detecting surface temperatures of objects and ambient temperatures of surrounding areas. Do not use for any other purpose.
• This thermometer is a sensitive electronic device. Treat it carefully and do not expose it to any sharp impacts.
• D
o not expose the thermometer to direct sunlight.
• D
o not disassemble this unit.
• T
he thermometer is NOT waterproof; keep away from water and other liquids.
• A
fter physical activity, a bath or spending time outdoors, wait approx. 30 minutes before you take
a measurement on your forehead.
• B
efore each use, ensure the sensor is undamaged; otherwise, contact customer service.
6
• The device needs to be in the room in which the measurement is taken for at least 30 minutes before use.
• T
his device is not intended for commercial or clinical use.
• Measure your temperature regularly to obtain your normal average temperature. If you suspect you have a fever, simply compare your normal average temperature with the temperature measured currently.
• D
o not self-diagnose or self-medicate on the basis of the measurements without consulting your doctor. In particular, do not start taking any new medication or change the type and/or dosage of any existing medication without prior approval.
• C
lean the thermometer after each use (see 7. Care and Maintenance).
• T
he manufacturer is not liable for damage resulting from improper or careless use.
• C
hildren must not be allowed to use the device. Medical devices are not toys.
• D
o not touch the lens with your fingers.
• R
ead these instructions for use carefully and keep them for later use.
• Packaging materials are a deadly hazard for children and can cause suocation. Remove all packaging materials immediately and keep them away from children at all times.
NOTICE – To reduce the risk of product/property damage:
• Taking your temperature provides information about your current state of well-being. If you are uncertain about interpreting the results or if the result is abnormal (e.g. fever), consult your doctor. This also applies in the case of slight temperature changes if there are other symptoms of illness such as agitation, severe sweating, flushed skin, fast pulse rate, tendency to collapse, etc.
• Temperatures measured with dierent thermometers cannot be compared. You should tell your doctor or keep in mind if you are measuring for yourself what type of thermometer you use and in which part of the body.
• P
ortable and mobile HF communication systems may interfere with this device.
7
Notes on electromagnetic compatibility
• The device is suitable for use in all environments listed in these instructions for use, including domestic environments.
• T
he use of the device may be limited in the presence of electromagnetic disturbances. This could
result in issues such as error messages or the failure of the display/device.
• A
void using this device directly next to other devices or stacked on top of other devices, as this could lead to faulty operation. If, however, it is necessary to use the device in the manner stated, this device as well as the other devices must be monitored to ensure they are working properly.
• T
he use of accessories other than those specified or provided by the manufacturer of this device can lead to an increase in electromagnetic emissions or a decrease in the device’s electromagnetic immunity; this can result in faulty operation.
• F
ailure to comply with the above can impair the performance of the device.
Battery Handling Safety Precautions
• U
se only the size and type of batteries specified.
• B
e sure to follow the correct polarity when installing the batteries. Reversed batteries may cause
damage to the device.
• Do not mix dierent types of batteries together (e.g., alkaline and carbon-zinc or rechargeable batteries) or old batteries with fresh ones. Always replace batteries as a simultaneous set.
• I
f the batteries in the device are depleted or the device will not be used for a long period of time,
remove the batteries to prevent damage or injury from possible battery leakage.
• Do not try to recharge batteries not intended to be recharged; they can overheat and rupture (follow battery manufacturer’s directions).
• D
o not dispose of batteries in fire, batteries may explode or leak.
• C
lean the battery contacts and also those of the device prior to battery installation.
8
• Remove discharged batteries from the product and dispose/recycle in compliance with all applicable laws.
• Should battery fluid leak and come into contact with your eyes or skin, immediately rinse with plenty of clean water. Please call 911 immediately.
• B
atteries should not be taken apart, thrown in the fire or short circuited.
• If a battery has leaked, put on protective gloves and clean the battery compartment with a dry cloth.
• Keep batteries away from children and pets. Batteries may be harmful if swallowed. Should a child or pet swallow a battery, seek medical assistance immediately.
2� Getting to Know your Instrument
This infrared thermometer measures:
• T
emperature of human body
• S
urface temperature of objects and liquids
• A
mbient temperatures
Additional features include:
• S
tores last 60 measurements for plotting changes in temperature
• S
tores date and time with each measurement
• F
ever alarm for temperatures over 99.5°F (37.5° C)
• S
witchable between °F and °C
• Bluetooth® communication via our free mobile app (requires a mobile device compatible with Bluetooth version 4.0 or later). For a list of compatible devices, scan this QR Code:
9
Dierent types of thermometers measure body temperature dierently. The resulting temperature varies depending on the part of the body where the measurement is taken. Temperatures in various parts of the body can vary up to two degrees Fahrenheit.
Measurement Location Normal Temperature Range
Forehead temperature 96.4° F to 99.7° F (35.8° C to 37.6° C) Ear temperature 96.8° F to 100.0° F (36.0° C to 37.8° C) Oral temperature 96.8° F to 99.3° F (36.0° C to 37.4° C) Rectal temperature 97.3° F to 100.0 ° F (36.3° C to 37.8° C)
NOTE:
Temperatures measured with dierent thermometers cannot be compared. You should tell your doctor or keep in mind if you are measuring for yourself what type of thermometer you use and in which part of the body.
10
Temperature is also influenced by:
• I
ndividual metabolism
• Age (body temperature is higher in babies and small children and falls with increasing age. Greater temperature fluctuations occur faster and more often in children)
• C
lothing
• O
utside temperature
• The time of day (body temperature is lower in the morning and increases slightly throughout the day).
• P
hysical and, to a lesser extent, mental activity.
NOTE:
Taking your temperature provides information about your current state of well-being. If you are uncertain about interpreting the results or if the result is abnormal (e.g. fever), consult your doctor. This also applies in the case of slight temperature changes if there are other symptoms of illness such as agitation, severe sweating, flushed skin, fast pulse rate, tendency to collapse, etc.
3� Package Contents
• N
on-Contact Thermometer FT 95
• 2
x 1.5 V AAA batteries (LR03)
• I
nstruction Manual
11
4� Parts and Controls
1 LCD Display 2 MODE/MEM button 3
button 4 LIGHT/SET button 5 STA RT button 6 Fever indicator 7 Measuring sensor 8 Battery Compartment
Buttons Functions
MODE/MEM
LIGHT/SET
START Press to measure temperature.
Fever Indicator Illuminates green/yellow/red depending on the measured temperature.
MODE Sets the measurement mode. MEM
LIGHT Turns LCD Display backlight on and o. SET Sets the basic functions.
Press to switch device on/o.
1
2
Displays the stored measurements.
12
7
3 4
5 6
8
Display description
1 Bluetooth® Indicator 2 Object Temperature mode 3 Room (Ambient) Temperature Mode 4 Forehead Temperature mode 5 Beep Indicator 6 Year 7 Memory Function 8 Date
9 Temperature/Memory Space Number Display 10 Battery Level Indicator 11 Fever Indicator (≥ 100.4° F / ≥ 38.0° C ) 12 Normal Body Temperature Indicator
(< 99.5° F /< 37.5° C )
13 Backlight Mode Indicator -
If on, backlight activates during measurement 14 Celsius Scale Indicator 15 Fahrenheit Scale Indicator 16 Year/date/time display
2 3 4 5
1
16
15
14
13 12 11 10
6 7 8
9
13
5� To Set Up
Before first use, remove the battery compartment cover and insert two “AAA” size batteries (included), observing the correct polarity as indicated inside the compartment. Slide the battery compartment cover back into place.
Setting the time and date:
1�
Press the
Press and hold the Fahrenheit scale indicator will appear on the LCD Display.
2�
Press the
or Fahrenheit (° F) as desired. Press the Button to save and continue. The time format will appear on the LCD Display. Press the select 12 or 24-hour display format and then press the
LIGHT/SET
ON/OFF
Button to switch the device on.
LIGHT/SET
MODE/MEM
button to save and continue.
Button; the Celsius/
Button to select Celsius (° C)
MODE/MEM
LIGHT/SET
button to
14
3�
Continue in this fashion to set hour, minute, year, month,
day, backlight mode, beep on/o, and Bluetooth on/o in turn.
4�
Press the
LIGHT/SET
button to save and exit; the device will shut o.
6� To Use
Forehead Temperature Measurement:
NOTES:
hysical activity, increased perspiration on the forehead, taking vasoconstrictive medication and
• P skin irritations can distort the measurement.
• T
he forehead, or the temples, must be free from perspiration and cosmetics.
• If possible, take three measurements on the forehead. If the three measurements dier, use the highest value.
• I
f the forehead area is covered in hair, sweat or dirt, clean the area and then wait for 10 minutes
before measurement.
• Keep the thermometer still during measurement. Moving the thermometer during measurement may cause incorrect results.
15
1�
Press the ON/OFF Button “ “ to switch the thermometer on; the LCD
Display will activate in forehead temperature measurement mode as indicated by the forehead icon at the top of the display.
2�
Hold the thermometer 1-2 inches from the forehead while pointing the
sensor toward the middle of the forehead. Press the measure.
3�
After about a second, the thermometer will beep once and display the
measurement. Press the
ON/OFF
Button to shut the thermometer o.
START
button to
99.5
16
4�
If the measurement is less than 99.5° F (37.5° C), the “no fever” icon will display and the Fever Indicator ring around the illuminate in green. If the measurement is 99.5° - 100.3° F (37.5° - 37.9° C), the Fever Indicator will illuminate yellow. If the measurement is over 100.4° F, the fever icon will display and the Fever Indicator will illuminate red.
5�
The device is ready to take another measurement when two short beeps sound and the forehead
symbol stops flashing. The device will shut o after about one minute; to manually shut the device o, press and hold the
ON/OFF
button.
ON/OFF
Button will
Bluetooth® and the Beurer HealthCoach app
To store measurements on your mobile device, download and install the free HealthCoach app from the Google Play Store (for Android devices) or the Apple App Store (for iOS devices). After downloading the app, start it and add FT95 under “Settings--> My devices.” If you find that measurements do not transfer to the app, try removing the protective cover from your mobile device to ensure the signal is not being blocked.
17
If Bluetooth is active on the thermometer and it is paired with your mobile device, measurements will transfer to the Beurer HealthCoach mobile app. The Bluetooth® symbol will flash on the LCD Display while the thermometer attempts to connect to the app for about 30 seconds. The Bluetooth icon will stop flashing when the thermometer connects to your device’s mobile app and stored measurements will then transfer. Afterwards, the thermometer will shut o after 60 seconds, or press the Button to shut it o manually.
ON/OFF
Viewing Stored Measurements
This device stores 60 measurements, along with the date/time and “fever”/”no fever” classification. With the device switched on, press and hold the most recent measurement will display. Press the the older measurements.
MODE/MEM
MODE/MEM
button for about three seconds; the
button repeatedly to cycle through
Surface temperatures:
To measure the surface temperature of an object, change the thermometer to Object Temperature Mode.
1�
Press the
ON/OFF
thermometer to surface temperature mode by pressing the
MEM
Button. The object temperature mode icon will display.
Button to switch the thermometer on. Change the
18
MODE/
2�
Hold the sensor 1-1 ½ inches (2-3 cm) from the object to be measured
and press the will beep once and display the measurement. Press the to shut the thermometer o.
START
Button. After about a second, the thermometer
ON/OFF
button
Ambient temperature:
To measure the ambient temperature in your location, switch the thermometer to ambient temperature mode.
1�
Press the
ON/OFF
thermometer to ambient temperature mode by pressing the
MEM
Button twice. The ambient temperature mode icon will
display.
2�
The ambient temperature around the thermometer will display. The
thermometer will shut o about one minute if no buttons are pressed; press the that ambient temperature measurements are not saved.
Button to switch the thermometer on. Change the
ON/OFF
button to shut the thermometer o manually. Note
19
MODE/
7� Care, Maintenance and Disposal
• The sensor is the most sensitive part of the thermometer. Work carefully when cleaning the sensor to avoid product damage. Use a clean cloth or cotton swab moistened with disinfectant, alcohol, or warm water.
• D
o not use any harsh cleaning products.
• B
efore prolonged storage, remove the batteries.
• Do not use or store the device in excessively high or low temperature or humidity, in direct sunlight, near an electrical current or in dusty locations.
• T
o clean the device, use a soft, damp cloth and a mild detergent. Ensure water does not get into the device. If the device does get wet, remove the battery immediately and let dry completely before reuse.
• N
ever immerse the device in water.
• W
hen storing the device for an extended period, remove the battery.
Disposal
• Observe the local regulations for material disposal. Dispose of the device in accordance with local regulations.
• I
f you have any questions, please contact the local authorities responsible for waste disposal.
20
8� Troubleshooting Guide
Error message
Problem Solution
Autotest in progress, device not ready.
Ambient temperature is not within measurement range.
Forehead temperature above 108 °F (42.2° C) or surface temperature above 176° F (80° C).
Forehead temperature below 93.2° F (34° C) or surface temperature below 32° F (0° C).
Batteries low or depleted. Replace batteries.
Wait for forehead symbol to stop flashing.
Place the thermometer in an ambient temperature between 59° F and 104° F (15° C - 40° C) for at least 30 minutes.
Only use thermometer within measurement range.
Only use thermometer within measurement range.
21
9� Technical Specification
Type Temperature Measurement Range Forehead: 93.2° - 108° F (34.0° - 42.2° C)
Measurement Accuracy Forehead: ±0.4° F (0.2° C) from 93.2° - 108° F (34° - 42.2° C);
Measurement Duration Approx. one second Operating conditions 59° - 104° F (15° - 40° C), Humidity <95% (non condensing) Storage conditions 4° - 122° F ( 20° - 50° C), Humidity <85% (non condensing) Memory Slots 60 Display resolution 0.1° Power 2 x 1.5V AAA (LR03) Dimensions 6.5 in x 1.6 in x 1.1 in (167 x 41.5 x 28 mm) Weight Approx. 2.6 oz (73 g) without batteries Automatic shuto 60 seconds
The device complies with the EU Medical Devices Directive 93/42/EEC, the German Medical Devices Act (Medizinproduktgesetz), the ASTM E 1965 - 98 and the European standard EN 60601-1-2 (in accordance with CISPR 11, IEC 61000-3-2, IEC 61000-3-3) and is subject to particular precautions with regard to electromagnetic compatibility. The device is suitable for use in all environments listed in these instructions for use, including domestic environments.
FT 95
Object temperature: 32° - 176° F (0° - 80 °C) Ambient temperature: 32° - 176° F (0° - 80° C)
Object temperature: ±4° F (2° C) Room temperature: ±4° F (2° C)
22
We hereby confirm that this product complies with the European RED Directive 2014/53/EU. The CE Declaration of Conformity for this product can be found under: https://www.beurer.com/web/ we-landingpages/de/cedeclarationofconformity.php
10� FCC Compliance Information
Non-Contact Thermometer FT 95
Responsible Party U.S. Contact Information: Beurer North America LP 1 Oakwood Boulevard, Suite 255 Hollywood, FL 33020 United States 1-800-536-0366 info@beurer.com
FCC Compliance Statement
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can
23
radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment o and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• R
eorient or relocate the receiving antenna.
• I
ncrease the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit dierent from that to which the receiver is connected.
• C
onsult the dealer or an experienced radio/TV technician for help.
The equipment complies with FCC RF exposure limits set forth for an uncontrolled environment. The equipment must not be co-located or operating in conjunction with any other antenna or transmitter.
11� Warranty
Limited Lifetime Warranty For Original Purchaser Your Beurer Non-Contact Thermometer, model FT 95, is warranted to be free from defects in
materials and workmanship for the life of the product under normal conditions of intended use and service. This warranty extends only to the original retail purchaser and does not extend to retailers or subsequent owners.
We will, at our option, repair or replace the Beurer Non-Contact Thermometer, model FT 95, without additional charge, for any part or parts covered by these written warranties. No refunds will be given. Repair or replacement is our only responsibility and your only remedy under this written
24
Loading...