Berchtold Operon D850 Service manual

OPERON® D 850 SERVICE & PARTS MANUAL
SERVICE AND PARTS MANUAL
Related Publications:
OPERON D 850 Installation and Operator Manual 700000121
OPERON D 850 Uncrating Instructions 700000122
The service and parts manual is meant to be used as a reference in performing preventative and corrective maintenance on the OPERON D 850 Surgical Table. The successful completion of a factory training program is required to properly and effectively troubleshoot, repair, and maintain the OPERON D 850 Surgical Table.
BERCHTOLD Corporation assumes no liability for table performance when used with accessories designed by others for use on the OPERON D 850 Surgical Table or damage to the OPERON D 850 Surgical Table resulting from the use of accessories designed by others.
This manual contains proprietary information of BERCHTOLD Corp. It shall not be reproduced in whole or in part without the written permission of BERCHTOLD Corp.
UL/CETL Listed to IEC 60601-1. Medical Electrical Equipment
®
OPERON
is a registered trademark of BERCHTOLD CORP.
OPERON D 850 SURGICAL TABLE TABLE OF CONTENTS
TABLE OF CONTENTS
SPECIAL SAFETY INSTRUCTIONS S-7
SAFETY SYMBOLS ....................................................................................................................S-7
USE CONDITIONS ....................................................................................................................S-7
POWER SUPPLY........................................................................................................................S-7
LOAD CAPACITY ......................................................................................................................S-7
PINCH POINTS..........................................................................................................................S-7
PATIENT POSITIONING...........................................................................................................S-9
OPERATION...............................................................................................................................S-9
TABLE MOVEMENT WITH PATIENT ......................................................................................S-9
PRIMARY HAND PENDANT FUNCTIONS ..............................................................................S-9
AUXILIARY HAND PENDANT ..................................................................................................S-9
HEADREST.................................................................................................................................S-9
REMOVABLE LEG SECTION .................................................................................................S-10
REMOVABLE Back SECTION .................................................................................................S-10
ACCESSORIES DESIGNED BY OTHERS ...............................................................................S-10
RADIOLUCENT TABLE TOP AND PADS ..............................................................................S-10
FOOTSWITCH .........................................................................................................................S-10
HYDRAULIC SYSTEM .............................................................................................................S-10
HYDRAULIC HOLLOW PLUGS..............................................................................................S-10
HYDRAULIC FLUID................................................................................................................S-10
ELECTROMAGNETIC INTERFERENCE ...............................................................................S-11
ELECTRICAL SYSTEM ............................................................................................................S-11
CLEANING AND DISINFECTION ..........................................................................................S-11
WARNING LABELS..................................................................................................................S-11
GENERAL
INFORMATION 1-1
GENERAL DESCRIPTION ........................................................................................................1-1
RADIOLUCENT TABLE TOP AND PADS ........................................................................1-2
INTERCHANGEABLE HEADREST ...................................................................................1-2
POWER SUPPLY ..................................................................................................................1-2
FLOOR LOCK SYSTEM ......................................................................................................1-3
FUSES....................................................................................................................................1-3
PRIMARY HAND PENDANT..............................................................................................1-3
AUXILIARY HAND PENDANT..........................................................................................1-3
PRIMARY HAND PENDANT INDICATORS ....................................................................1-4
PRIMARY HAND PENDANT FUNCTIONS ......................................................................1-5
FOOTSWITCH (Optional).....................................................................................................1-6
MANUAL FOOT PUMP .......................................................................................................1-6
EMERGENCY UNLOCK .....................................................................................................1-6
AUXILIARY HAND PENDANT..........................................................................................1-7
OPERATING
INSTRUCTIONS 2-1
PINCH POINTS..........................................................................................................................2-1
ELECTROMAGNETIC INTERFERENCE .................................................................................2-2
LOAD CAPACITY ......................................................................................................................2-2
PATIENT ORIENTATION ..........................................................................................................2-2
PATIENT POSITIONING...........................................................................................................2-2
OPERATION...............................................................................................................................2-3
HAND PENDANT OPERATION .........................................................................................2-3
Function Priority.....................................................................................................................2-4
Table Movement Limits .........................................................................................................2-4
HEADREST ...........................................................................................................................2-5
REMOVABLE BACK AND LEG SECTIONS.....................................................................2-6
Rev. 1 700000120 S-3
OPERON D 850 SURGICAL TABLE
TABLE OF CONTENTS
TABLE OF CONTENTS CONTINUED
FOOTSWITCH (OPTIONAL) .................................................................................................... 2-7
PRINCIPLES
OF OPERATION 3-1
HYDRAULIC SYSTEM............................................................................................................... 3-1
Hydraulic Power Unit ............................................................................................................3-1
Hydraulic Manifolds .............................................................................................................. 3-1
Column Manifold................................................................................................................... 3-1
Floor Lock Manifold.............................................................................................................. 3-2
Floor Lock System...................................................................................................................... 3-2
MANIFOLD VALVE DESCRIPTION ........................................................................................ 3-3
Manifold Valve Operation ..................................................................................................... 3-4
COLUMN HYDRAULICS .......................................................................................................... 3-5
General Information............................................................................................................... 3-5
DETAILS
OF OPERATION 4-1
Height Up and Down.................................................................................................................. 4-1
Tilt Left and Right ......................................................................................................................4-1
Trendelenburg and Reverse Trendelenburg............................................................................... 4-1
Table Top Hydraulics................................................................................................................. 4-1
HOLLOW PLUGS AND SEALS................................................................................................. 4-3
Using Hollow Plugs and Washers to Secure Hydraulic Lines............................................... 4-4
Hollow Plug Torque Specifications ....................................................................................... 4-4
SERVICE 5-1
HYDRAULIC FLUID .................................................................................................................5-1
ELECTRONIC CONTROL SYSTEM.......................................................................................... 5-1
CPU........................................................................................................................................ 5-1
Communication Interface (RS232) Board .............................................................................5-1
Position Sensors and Switches............................................................................................... 5-1
Slide & Height Sensor ...........................................................................................................5-2
Auxiliary Pendant Switch ...................................................................................................... 5-2
Trend/Tilt Sensor ................................................................................................................... 5-3
Leg and Back Angle Sensors ................................................................................................. 5-3
Power Supply System ............................................................................................................ 5-3
Batteries ................................................................................................................................. 5-4
Power Input System ............................................................................................................... 5-5
CONTROL PENDANTS ............................................................................................................. 5-6
Primary Hand Pendant ........................................................................................................... 5-6
Auxiliary Hand Pendant......................................................................................................... 5-7
WIRING...................................................................................................................................... 5-7
MECHANICAL SLIDE SYSTEM................................................................................................ 5-7
SING THE SERVICE PROGRAM 6-2
U
Connect to the Table .................................................................................................................. 6-2
Launch the Service Program......................................................................................................6-2
The Hand Pendant Graphic.................................................................................................... 6-2
The Service Program Menus ...................................................................................................... 6-3
Buttons................................................................................................................................... 6-8
Device Settings Screen ..........................................................................................................6-8
Software Commands with Primary Pendant ....................................................................... 6-12
CPU Error Codes ................................................................................................................. 6-13
Table of Figures
Figure S-3 Serial Number and Regulatory Approval Labels.................................................... S-12
Figure 1-1 OPERON D 850 Surgical Table ............................................................................... 1-1
Rev. 1 700000120 S-4
OPERON D 850 SURGICAL TABLE TABLE OF CONTENTS
Table of Figures continued
Figure 1-2 Power Center Components ........................................................................................1-3
Figure 1-3 Primary Hand Pendant Indicators..............................................................................1-4
Figure 1-4 Primary Hand Pendant Functions ..............................................................................1-5
Figure 1-5 Footswitch .................................................................................................................1-6
Figure 1-6 Deploying Foot Pump Pedal......................................................................................1-6
Figure 1-7 Emergency Switch.....................................................................................................1-7
Figure 1-8 Auxiliary Hand Pendant ............................................................................................1-7
Figure 2-1 Pinch Points...............................................................................................................2-1
Figure 2-2 Table Orientation.......................................................................................................2-2
Figure 2-4 Headrest.....................................................................................................................2-6
Figure 2-5 Leg Section Attachment ............................................................................................2-6
Figure 2-6 Assembled Leg Section .............................................................................................2-6
Figure 2-7 Footswitch .................................................................................................................2-7
Figure 3-1 Manifold Distribution Blocks....................................................................................3-1
Figure 3-2 Distribution Blocks ...................................................................................................3-1
Figure 3-3 Floor Lock Manifold .................................................................................................3-2
Figure 3-4 Location of Floor Lock Manifold In Base.................................................................3-2
Figure 3-5 Floor Lock System Operation ...................................................................................3-3
Figure 3-6 Manifold Valve .........................................................................................................3-4
Figure 3-7 Column Hydraulics ...................................................................................................3-6
Figure 4-1 Lift Column/Cylinders ..............................................................................................4-1
Figure 4-2 Tilt Cylinder ..............................................................................................................4-1
Figure 4-3 Hose Routing To Upper Frame .................................................................................4-2
Figure 4-4 Right Side Frame Cylinders ......................................................................................4-2
Figure 4-5 Back and Leg Cylinders ............................................................................................4-3
Figure 4-6 Left and Right Kidney Cylinder ................................................................................4-3
Figure 4-7 Hollow Plugs and Washers........................................................................................4-4
Figure 4-8 Securing Hydraulic Line ...........................................................................................4-4
Figure 5-1 D 850 Reservoir ........................................................................................................5-1
Figure 5-2 Height Sensor ............................................................................................................5-2
Figure 5-3 Slide Sensor ...............................................................................................................5-2
Figure 5-4 Auxiliary Pendant Switch..........................................................................................5-2
Figure 5-5 Kidney Switch ...........................................................................................................5-2
Figure 5-6 Trend/Tilt Sensor (viewed from under table) ............................................................5-3
Figure 5-7 Leg & Back Section Proximity Sensors ....................................................................5-3
Figure 5-8 Power Supply System................................................................................................5-3
Figure 5-9 Charging circuit output diagram................................................................................5-4
Figure 5-10 Batteries.....................................................................................................................5-4
Figure 5-11 Terminal ....................................................................................................................5-5
Figure 5-12 Transformer...............................................................................................................5-5
Figure 5-13 Power Supply ............................................................................................................5-5
Figure 5-14 Power Input Center....................................................................................................5-6
Figure 5-15 Primary Hand Pendant...............................................................................................5-6
Figure 5-16 Slide Screw Drive Belt And Potentiometer Cable Routing.......................................5-7
Figure 6-1 RS232 Connection.....................................................................................................6-2
Figure 6-2 Login Form................................................................................................................6-2
Figure 6-3 Service Program Main Menu Screen.........................................................................6-1
Figure 6-4 Service Program Main Menu Screen Menu Options.................................................6-2
Figure 6-5 File Menu Options.....................................................................................................6-3
Figure 6-6 Windows File open Selection Screen ........................................................................6-4
Figure 6-7 Save Settings to file ...................................................................................................6-4
Figure 6-8 Device Settings Report..............................................................................................6-4
700000120 Rev. 1
S-5
OPERON D 850 SURGICAL TABLE
TABLE OF CONTENTS
Table of Figures continued
Figure 6-9 Print Report Screen ................................................................................................... 6-4
Figure 6-10 Export Report Screen................................................................................................6-5
Figure 6-11 Zoom Menu............................................................................................................... 6-5
Figure 6-12 Device Settings Screen ............................................................................................. 6-5
Figure 6-13 System Information Screen....................................................................................... 6-5
Figure 6-14 Errors Information Screen......................................................................................... 6-6
Figure 6-15 LEDs State Information Screen ................................................................................ 6-6
Figure 6-16 Sensors Information Screen ...................................................................................... 6-6
Figure 6-17 Keypad/Pedal Information Screen ............................................................................ 6-6
Figure 6-18 Valves Information Screen ....................................................................................... 6-7
Figure 6-19 Valves Test Screen.................................................................................................... 6-7
Figure 6-20 COM Options Screen................................................................................................ 6-7
Figure 6-21 COM Unlock Screen................................................................................................. 6-8
Figure 6-22 High Power Settings Screen...................................................................................... 6-9
Figure 6-23 Sensor Setting Screen ............................................................................................. 6-10
Figure 6-24 Comfort Movements Screen ................................................................................... 6-11
Figure 6-25 Shutdown Screen ....................................................................................................6-12
Figure 6-26 About Information Screen....................................................................................... 6-12
Index of Tables
Table 1-1 Technical Data................................................................................................................ 1-10
Table 2-1 Allowed Simultaneous Movements................................................................................... 2-4
Table 2-2 Software Controlled Limits ............................................................................................... 2-5
Table 3-1 Manifold Valve Operating Specifications..................................................................... 3-4
Table 3-2 Column Solenoids ......................................................................................................... 3-5
Table 4-1 Hollow Plug Torque Specifications ..............................................................................4-4
Table 6-1 Sensor Parameters ....................................................................................................... 6-10
Table 6-2 Battery setting parameters........................................................................................... 6-12
Table 6-3 OPERON B850 Error Codes....................................................................................... 6-13
Rev. 1 700000120 S-6
OPERON D 850 SURGICAL TABLE SPECIAL SAFETY INSTRUCTIONS
IMPORTANT -
THE SAFETY INSTRUCTIONS THAT FOLLOW APPEAR WITHIN THE MANUAL.
READ THEM CAREFULLY BEFORE OPERATING THE UNIT AND FOLLOW
INSTRUCTIONS.
SPECIAL SAFETY INSTRUCTIONS
SAFETY SYMBOLS
This manual uses special symbols to help alert you to important safety information:
This symbol identifies a WARNING.
WARNING
CAUTION
NOTE
A warning indicates that the procedure in the following text involves actions that could cause harm or death to an individual if certain precautions are not taken. The warning text always provides specific information on what you must do to avoid the risk.
This symbol identifies a CAUTION. A caution indicates that the procedures in the following text involve actions that could cause damage to the equipment, including the failure to operate. The caution text always provides specific information on what you must do to avoid the risk.
This symbol identifies a NOTE. A note contains information that can help you perform the procedure in the following text more effectively.
USE CONDITIONS
WARNING — POSSIBLE
WARNING
EXPLOSION HAZARD: Do not operate the table in the presence of flammable anesthetics. The OPERON D 850 table is designed
for use in non-hazardous anesthetizing locations.
POWER SUPPLY
WARNING — ELECTRICAL
WARNING
HAZARD: Repeated breaker tripping or fuse replacement may indicate a ground fault or overload condition. The table will not operate reliably under such a condition and may be electrically unsafe. Do not use the table until this condition is corrected.
LOAD CAPACITY
In normal orientation, this table has a patient lifting and articulation capacity of 1000 lb (453 kg). In reverse orientation, this table has a patient lifting and articulation capacity of 500 lbs (227 kg). Positioning and constraints are discussed on page S-9. Do not exceed this maximum load in service. See Figure 1-1.
WARNING — Overloading may
WARNING
NOTE
result in tipping of the table or failure of the table sections to move. Do not exceed the maximum load in service.
NOTE — The use of some positioning capabilities or table accessories will put additional restrictions on patient weight. Accessories provided by Berchtold Corp., when used as intended, are designed to perform safely with the patient loads for which this table was designed. A Berchtold Corporation accessory that puts additional restrictions on patient weight, when used as intended, will be clearly labeled to indicate the maximum allowed loads.
The table load capacity is 400 lb (181 kg) when the beach chair accessory is installed on the table.
PINCH POINTS
WARNING — Cutting or crushing
WARNING
CAUTION
injury may occur if any part of a patient or staff member is caught in a pinch point when the table is articulated.
CAUTION — Damage to the table or object may occur if any foreign object is caught in a table pinch point when the table is articulated.
Due to the variety of table positions attainable with the OPERON D 850 table, pinch or crushing points exist (Ref. Figure S-1). The OR staff must be trained to ensure that body parts or surgical apparatus are not caught in pinch points when the table is operated.
Rev. 1 700000120 S-7
OPERON D 850 SURGICAL TABLE SPECIAL SAFETY INSTRUCTIONS
This list is not all-inclusive, because pinch points can be created by the attachment of accessories. Always ensure that the patient and staff are clear of all pinch points before moving the table or while the table is in motion.
Pinch points include but are not limited to:
1. Kidney elevator to seat top
2. Kidney elevator to seat channels
3. Kidney elevator to back section
4. Headrest to back section
5. Back / Leg section link to seat channel
6. Side shields to column shields (in tilt)
7. Column shields to base cover
8. Base to floor
9. Floor feet to base
10. Floor feet to floor
11. Leg section to column shield
12. Foot pump handle to floor
13. Back Section to seat top.
1 2 3
13
4
5
11
12
5
6
7
10
9
8
Figure S-1 Pinch Points
Rev. 1 700000120 S-8
OPERON D 850 SURGICAL TABLE SPECIAL SAFETY INSTRUCTIONS
PATIENT POSITIONING
WARNING — POSSIBILITY OF TIPPING: Table stability may be
WARNING
compromised if patients are positioned beyond the following limits:
The patient's pelvis should not be extended beyond the end of the seat section. The patient’s legs may be placed on a leg extension attached to the end of the leg section.
WARNING — POSSIBILITY OF TIPPING: If FULL SLIDE
WARNING
FORWARD IS REQUIRED, the patient’s pelvis must always be located over the seat section when in normal orientation. Utilizing full forward slide while the patient’s pelvis is on the leg section can impair table stability and result in patient injury or equipment damage.
WARNING — The leg section may be removed from the table. If it is
WARNING
left on the table, use caution when raising the leg section while the patient is in the normal perineal position.
WARNING — The back section may be removed from the table.
WARNING
Full articulation and full slide in either direction can be safely achieved with a 1000lb (453g) patient if these steps are followed.
OPERATION
WARNING— The table should
WARNING
WARNING
always be locked to the floor whenever a patient is on the table or being transferred to or from the table. This is accomplished by extending the table’s feet by activating the FLOOR LOCK function. This table is not intended for patient transport.
WARNINGWeight Limit in reverse, while not recommended, it is possible to place a patient in reverse orientation. Patient weight must not exceed 500 lbs. when placed in reverse.
CAUTION — Storage of clamps,
accessories, etc., on the base cover
CAUTION
may result in damage to lift column shields and prevent proper elevation or descent of the table.
TABLE MOVEMENT WITH PATIENT
WARNING — If the table needs to be moved with a patient on the
WARNING
table, it is very important to follow the following instructions to prevent patient injury or equipment damage.
WARNING — The table should not be used to transport a patient over a
WARNING
long distance. This procedure is only meant for repositioning the table in the surgical suite.
PRIMARY HAND PENDANT FUNCTIONS
AUXILIARY HAND PENDANT
WARNING — The auxiliary hand pendant functions will operate
WARNING
regardless of whether the floor feet are locked or unlocked. If the floor feet are unlocked, the table may roll when the manual pump is operated.
CAUTION — The auxiliary hand pendant does not have the following
CAUTION
functions: FLEX, REFLEX, CHAIR, RETURN TO LEVEL, UROLOGY, or SLIDE. This pendant is intended to be used for completing procedures, followed by charging the batteries or servicing the table.
WARNING — When using the auxiliary pendant, the software
WARNING
controlled limits are not active. DAMAGE MAY OCCUR. Review the user controlled limits section of this manual.
HEADREST
WARNING — Do not mount knee
WARNING
crutches or leg holders to the headrest side rail. The headrest should only be used for supporting the head or feet. Do not lean on the headrest. The headrest has a weight limit of 50 lbs.
Rev. 1 700000120 S-9
OPERON D 850 SURGICAL TABLE SPECIAL SAFETY INSTRUCTIONS
WARNING — When articulating the
WARNING
WARNING
headrest, be sure that the locking mechanism engages completely, and does not hang in between positions. Until the locking mechanism engages, the headrest may move when a load is applied.
WARNING — When releasing the locking mechanism, support the headrest, as it will fall freely if the locking mechanism is not engaged.
REMOVABLE LEG SECTION
WARNING — Failure to securely
WARNING
lock the leg section into place may result in patient injury. Always check the retention of the leg section upon installation.
REMOVABLE BACK SECTION
WARNING — Failure to securely
WARNING
lock the back section into place may result in patient injury. Always check the retention of the leg section upon installation.
ACCESSORIES DESIGNED BY OTHERS
Berchtold Corporation assumes no liability for table performance when used with accessories designed by others for use on Berchtold equipment, or damage resulting to Berchtold equipment resulting from the use of accessories designed by others.
RADIOLUCENT TABLE TOP AND PADS
CAUTION — The anti-static
CAUTION
CAUTION
properties of the table depend on the use of the recommended mattress. Compromise of the anti-static properties could expose the patient or OR personnel to an electrostatic shock. It could also deliver a shock to the table’s electronics that could damage or destroy critical components.
CAUTION — Protect the top and bottom surfaces from nicks and scratches. Scratches more than
0.005 inches (127 microns) deep may appear on x-rays. Damaged or scratched panels should be replaced if x-ray quality is affected.
FOOTSWITCH
CAUTION — The Footswitch’s
CAUTION
electrical connections are resistant to fluid intrusion, but if fluid drainage is anticipated, the footswitch should be protected by wrapping it with a clear disposable footswitch bag. This will reduce the need to disinfect the footswitch.
HYDRAULIC SYSTEM
WARNING – POSSIBILITY OF
WARNING
CAUTION
INJURY: The hydraulic system must be kept free of contamination. Contamination can cause the hydraulic system to fail, resulting in personal injury or equipment damage.
CAUTION – The hydraulic system on the OPERON D 850 surgical table contains micro-hydraulic valves, which are very sensitive to contamination. Only a qualified, trained service technician should open the hydraulic system, and only after all other repair procedures have been tried. Always use clean tools, and cover or wrap any loose hydraulic lines with clean plastic bags.
HYDRAULIC HOLLOW PLUGS
CAUTION – Do not over tighten the
CAUTION
CAUTION
hollow plugs. Over tightening can cause the seals to deform and leak.
CAUTION — Hollow plugs used in the hydraulic system have different torque values than the mechanical fasteners. See plug torque requirements.
Table 4-1 for hollow
HYDRAULIC FLUID
CAUTION – Do not mix hydraulic
CAUTION
fluid types. Other hydraulic fluids can be incompatible with the seals in the OPERON D 850 table and can cause equipment failure or reduce the service life. (See drawing 306762 for correct fluid designation.)
Rev. 1 700000120 S-10
OPERON D 850 SURGICAL TABLE SPECIAL SAFETY INSTRUCTIONS
ELECTROMAGNETIC INTERFERENCE
WARNING — When using the table
WARNING
WARNING
in the vicinity of high-frequency surgical equipment, such as cardiac defibrillators and cardiac defibrillator monitors, refer to the equipment manufacturers’ instructions to ensure compatibility issues are considered.
WARNING — When using the table in the vicinity of neuro-diagnostic monitoring equipment, interference may appear on the monitoring device. To eliminate this, set the BACKLIGHT TIMER to 1 sec using the service software
ELECTRICAL SYSTEM
WARNING – When servicing the
WARNING
WARNING
WARNING
table, turn power off or remove the appropriate fuses to prevent personal injury. If power is required, use care to avoid contact with live parts.
WARNING – Even a “dead” battery can contain enough power to cause a serious shock or explosive spark. Always disconnect the battery charger plug at the power center before disconnecting cables.
WARNING – The batteries contain acid, which can cause serious burns. While the batteries are sealed, a dead battery may be damaged and may leak. Always handle batteries in a way that will prevent acid from spilling on your skin or clothes.
CLEANING AND DISINFECTION
WARNING — Unplug the power
WARNING
CAUTION
WARNING
cord before cleaning the table if cleaning liquids will be used near the power cord receptacle. Under no circumstances should steam or extremely hot water (over 150°F/66°C) be used to disinfect or clean the table. Elevated temperatures and high humidity will weaken the hydraulic lines and result in failure of the hydraulic system and increased leakage currents. Do not clean this table in automatic cart washing equipment. Use of high­pressure water guns could lead to an electrical malfunction and must be avoided.
CAUTION — Failure to thoroughly dry the surface after cleaning may result in rust.
WARNING — Always replace fuses with fuses of the same voltage and current rating. Failure to use the correct fuse could cause a fire and injury to the patient or hospital personnel.
WARNING LABELS
Figure S-2 shows the patient safety warning label attached to the table. The label is located on the base at the foot end.
700000120 Rev. 1
S-11
SPECIAL SAFETY INSTRUCTIONS
OPERON D 850 SURGICAL TABLE
Figure S-2 Patient Safety Warning Labels
Figure S-3 Serial Number and Regulatory Approval Labels
Rev. 1 700000120 S-12
OPERON D 850 SURGICAL TABLE GENERAL INFORMATION
GENERAL INFORMATION
GENERAL DESCRIPTION
The Berchtold OPERON D 850 surgical table is a remote controlled, full-function design offering a high level of patient positioning and imaging flexibility to the surgical team. The table can be activated by AC power through the power cord (input range of 115V ±10%) or by battery power, which requires no power cord except when charging the batteries.
An electric motor with a belt driven screw is used to slide the table top head to foot.
Hydraulic cylinders are used to articulate the table's sections. Hydraulic pressure is provided by a motor driven pump during
Patient’s pelvis must not extend beyond the end of the seat section
Seat section Back section
normal use, and by a manual foot pump in manual back-up mode.
A hand-held control pendant is used to control table positions.
The tabletop is divided into the following sections:
Headrest
Back Section
Kidney Elevator
Seat Section
Leg Section
A leg extension can be added to extend the tabletop surface.
Head section
Leg section
Manual foot pump pedal
Main hand control pendant
Auxiliary Pendant Storage
Floor locking feet
Figure 1-1 OPERON D 850 Surgical Table
700000120 Rev. 1
1-1
OPERON D 850 SURGICAL TABLE GENERAL INFORMATION
RADIOLUCENT TABLE TOP AND PADS
CAUTION — The anti-static
CAUTION
properties of the table depend on the use of the recommended mattress. Compromise of the anti-static properties could expose the patient or OR personnel to an electrostatic shock. It could also deliver a shock to the table’s electronics that could damage or destroy critical components.
The tabletop surface is made from electrically conductive radiolucent panels. The table is fitted with a matching set of conductive radiolucent pads. Velcro tape holds the pads in place.
CAUTION — Protect the top and
CAUTION
Each table section has a cassette channel that can accommodate x-ray cassettes up to 14'' x 17'' (356 x 432 mm).
INTERCHANGEABLE HEADREST
The headrest can be attached to the seat section to allow for positioning the patient toward the leg end of the table.
POWER SUPPLY
WARNING
bottom surfaces from nicks and scratches. Scratches more than 0.005 inches (127 microns) deep may appear on x-rays. Damaged or scratched panels should be replaced if x-ray quality is affected.
WARNING — Repeated breaker tripping or fuse replacement may indicate a ground fault or overload condition. The table will not operate reliably under these conditions, and may be electrically unsafe. Do not use the table until these conditions are corrected.
NOTE — Recharge the batteries as
NOTE
indicated by the battery indicator on the control pendant. For longevity, it is good practice to charge the batteries at least once a week
The locations of the primary Power Supply components are shown in Figure 1-2.
When the table is plugged into a 120VAC receptacle, it is powered by AC (mains). When the power cord is not plugged in, the batteries provide power to the table. In either case, the ON/OFF switch on the base must be ON to articulate any table sections. When the ON/OFF switch on the base is OFF, the table's electronics and pump motor are isolated from the power supply and the batteries, regardless of the pendant used.
When the charge indicator lamp near the power cord receptacle is lit, the table senses AC power and activates the battery charging system. If AC is present, the battery charging system is activated whether the ON/OFF switch on the base is in the ON or OFF position. The table can remain plugged into AC indefinitely without damaging the batteries. The battery indicator on the primary hand pendant displays the battery charge status. This display has three sections: green (full charge), yellow (low charge), and red (recharge). Only one color will be activated at a time.
NOTE — In the unlikely event of a
NOTE
control pendant failure, the table movements can be stopped by switching the breaker on the base to the off (down) position.
Rev. 1 700000120 1-2
OPERON D 850 SURGICAL TABLE
R
GENERAL INFORMATION
COVE
FUSES
GROUND
ON/OFF CIRCUIT BREAKER
CHARGE INDICATOR
Figure 1-2 Power Center Components
FLOOR LOCK SYSTEM
Four self-stabilizing floor feet compensate up to
0.25'' (6 mm) for uneven floors. The floor feet can be lowered (locked) or raised (unlocked) using the control pendant. When the floor feet are raised, the table rests on four non-conductive casters.
FUSES
Fuses are accessible without removing the base covers.
They are located within the receptacle for the power cord, contains two fuses. These fuses protect against current surges in the AC (mains) line.
In addition to the fuses, the ON/OFF switch serves as a circuit breaker with a 20 Amp rating.
See Figure 1-2 for the fuse locations.
PRIMARY HAND PENDANT
A primary hand pendant controls all table functions. The floor feet must be locked by activating the FLOOR LOCK function to operate any other pendant functions. This ensures that the table base will not move while the tabletop is articulating.
NOTE — The ON/OFF function on the
NOTE
control pendant can be used to stop the table while in motion.
AUXILIARY HAND PENDANT
The Auxiliary Hand Pendant is identical to Primary Hand Pendant and can be used as a replacement for a failed Primary Hand Pendant. To replace the Primary Hand Pendant, unplug it from the Primary Pendant receptacle and plug in the Auxiliary Pendant. The Auxiliary Hand Pendant will now have all of the functions and indicators of the Primary Hand pendant.
The primary hand pendant plugs into the PATIENT LEFT side of the remote control connector box, which is located on the head end of the column beneath the table top. Insert the pendant so the red dot on the connector aligns with a red dot on the socket in the Primary Hand Pendant Connector Box.
700000120 Rev. 1
1-3
OPERON D 850 SURGICAL TABLE OPERATING INSTRUCTIONS
PRIMARY HAND PENDANT INDICATORS
See Figure 1-3 for the location of the indicators
1. AUX. PENDANT (amber flashing) —
Auxiliary hand pendant is active. The primary hand pendant will be disabled if the auxiliary hand pendant is active.
2. SERVICE (red) — Table malfunction. It is
likely that most of the functions still work properly and can still be controlled from the primary hand pendant. If the control pendant is not working, use the auxiliary hand pendant and the manual foot pump (if necessary) to articulate the table until the malfunction can be repaired.
NOTE — If the SERVICE indicator lights,
NOTE
call your local Berchtold Corporation Service Representative.
3. TABLE LIMIT (amber flashing)
Indicates that the movement commanded has reached a software controlled limit to avoid damage to the table components.
4. POWER (green) – hand pendant is active.
Backlight may or may not be lit.
5. FLOOR UNLOCK ² (amber) — Floor feet
are raised so table is mobile.
6. FLOOR LOCK ± (amber) — Floor feet are
lowered so table is locked in place.
7. SPEED ADJUST HI (green) – articulation
speed of all functions is set to normal.
8. SPEED ADJUST LO (green) – articulation
speed of all functions is set to low.
9. BATTERY — LEDs indicate the charge
status of the batteries:
Green — Full to medium charge. Normal
Operation.
Yellow — Low charge. Table still
operates normally, but batteries should be charged at the first opportunity.
Red (steady) — Recharge battery. Pump
motor is shut off and manual foot pump must be used.
Red (flashing) — Auxiliary pendant must
be used. Recharge battery immediately!
NOTE — In the event of a primary hand
NOTE
pendant failure, use the auxiliary hand pendant and foot pump to complete the procedure and contact your local Berchtold Corporation Service Representative.
10. BACKLIGHT – Pendant backlighting
(electroluminescent panel) combined with the green Power LED indicates that the pendant is active. The Backlight will time out after a preset time period, which can be set in the service software. The backlight timer can be set to a shorter time than the Controls Wakened state. This will reduce electrical current drain on the batteries, slightly extending the battery charge cycle. The lowest backlight timer setting is 1 second.
Figure 1-3 Primary Hand Pendant
Indicators
Rev. 1 700000120 1-4
OPERON D 850 SURGICAL TABLE GENERAL INFORMATION
PRIMARY HAND PENDANT FUNCTIONS
1. POWER - Turns the primary hand
pendant on and off. The pendant will be backlit when activated (unless backlight timer is exceeded. See hand pendant operation section)
NOTE: - The power switch on the base of
NOTE
NOTE
2. FLOOR LOCK - Raises (unlocks) or lowers
3. SPEED ADJUST (HI/LO) – changes the
4. AUTO DRIVE REV/FWD - (Optional)
5. BACK v - Raises and lowers the back
6. TABLE v - Raises and lowers the entire
7. LEG v - Raises and lowers the leg section.
8. TREND/REV TREND
9. SLIDE FOOT/SLIDE HEAD - Slides the
10. TILT LEFT/TILT RIGHT - Rotates the
11. UP KIDNEY/DOWN KIDNEY - Raises
12. FLEX - Combines back down with reverse
13. REFLEX - Combines back up with
14. CHAIR - Combines back up, Trendelenburg,
the table must be in the ON position before the hand pendant will operate.
NOTE - If AC power is not present, the auto-shutdown feature will turn the control pendant off after a preset period of non­use. To reactivate the pendant, press any function button. If AC is present, the auto­shutdown feature is disabled, and the hand pendant remains on until it is turned off.
(locks) the floor feet.
rate of speed of all movements between normal and slow. This is useful when gradual motion is required.
drives the table toward the head (forward) or foot (reverse). There is a slow acceleration, followed by a higher speed (~0.75m/s) if held.
section.
tabletop.
(Trendelenburg) - Tilts
the entire tabletop about the transverse axis.
table top toward and away from the head.
entire tabletop about the head to foot axis
.
and lowers the kidney elevator.
Trendelenburg to form an inverted V (jackknife position).
Trendelenburg to return from FLEX position.
and leg down functions to form a chair-like support. The three functions operate in a repeating sequence until the back section
reaches 90° relative to the seat section and the leg sections reach 105° degrees relative to the seat section.
Figure 1-4 Primary Hand Pendant
Functions
15. UROLOGY – Levels the table; then
lowers the leg section to 90° or until a software limit is reached. (To bring the table to its lowest position and have LEG at 90°, slide must be fully toward the leg end)
16. LEVEL - Moves the table to a horizontal
position that is within 1 degree of level. This function should be used prior to the Flex and Chair Functions.
700000120 Rev. 1
1-5
OPERON D 850 SURGICAL TABLE GENERAL INFORMATION
TO ITS FULL OUTERMOST POSITION
FOOTSWITCH (Optional)
A footswitch control is available for procedures where the surgeon is normally seated at the pelvic end of the patient (Figure 1-5).
ENABLES THE AUXILIARY PENDANT AND DISABLES THE MAIN HAND PENDANT.
To use the manual foot pump:
Footswitch functions include:
1. Pull the foot pedal straight out as far as it
Trendelenburg/Reverse Trendelenburg
Tilt Left/Tilt Right
Table Up/Down
will go. It will rotate counter clockwise as it is pulled outward.
The footswitch is connected into the remote control connector box by aligning the red dot on the footswitch connector with the red dot on the socket in the remote control connector box. See Figure 1-5 for location of connector.
Table Up/
Table Down
Tilt Le ft/
Tilt Right
Reverse Trendelenburg
Figure 1-5 Footswitch
Pressing one of the footswitch controls turns the table on and places it in “controls active” status. However, none of the functions will operate unless the floor locks are set. The floor locks must be operated through the hand pedant.
MANUAL FOOT PUMP
The manual foot pump is used to provide hydraulic pressure when there is no electric power to operate the hydraulic pump. This could occur due to any of the following conditions:
Battery power is low as indicated by a
steady or flashing red battery light on the hand pendant.
A combined failure of the battery and AC
mains power supplies.
A failure of the table’s hydraulic pump or
the electrical system that powers the hydraulic pump.
A failure of table controls
The manual pump pedal is located at the patient left side near the head end of the base. DEPLOYING THE FOOT PUMP HANDLE
Figure 1-6 Deploying Foot Pump Pedal
2. Extend the pedal until it clicks into
position.
3. To articulate the table, press and hold the
desired function button on the auxiliary control pendant (see next section) and operate the manual pump with your foot. For best results, depress the pump pedal completely, making full strokes, at the rate of about one stroke per second.
To re-insert the pump pedal:
1. Push the pedal back into the table until
the end is inserted into the base cover.
2. Continue to push the pedal into the table
until it snaps into the retracted position.
EMERGENCY UNLOCK
In the unlikely situation where the control system is completely disabled AND the auxiliary system is disabled, there is a back-up emergency switch that will allow the table to be unlocked and moved. It is wired directly to the batteries. There should always be enough power in the batteries to open the valve solenoids. This switch is located inside the base near the foot pump handle storage location.
Rev. 1 700000120 1-6
OPERON D 850 SURGICAL TABLE GENERAL INFORMATION
Figure 1-7 Emergency Switch
Do this to access this switch:
1. Pull the foot pedal straight out as far as it
will go. It will rotate counter clockwise as it is pulled outward.
2. Put one finger through the handle storage
hole as shown and press the button.
3. The floor locks should release and the table
should be resting on its casters.
4. Press the button again to switch off the
power to the valve solenoid.
AUXILIARY HAND PENDANT
NOTE — Activating the auxiliary
NOTE
hand pendant disables the primary hand pendant.
.The auxiliary hand pendant is identical to the main hand pendant. It has limited function when plugged into the auxiliary control connector, located in the auxiliary hand pendant compartment on the head end of the column. The active functions are Leg Up/Down, Table Up/Down, Back Up/Down, Floor Lock, Trend, Tilt Left/Right, and Kidney Up/Down. Only these buttons will be backlit when in auxiliary mode.
Figure 1-8 Auxiliary Hand Pendant
NOTE — The Auxiliary hand
NOTE
pendant is identical to the main hand pendant and may be used as a replacement for the main hand pendant. See Primary Hand Pendant for details.
Use the auxiliary control pendant when:
The primary hand pendant will not
function.
WARNING — The auxiliary hand
WARNING
CAUTION
pendant functions will operate regardless of whether the floor feet are locked or unlocked. If the floor feet are unlocked, the table may roll when the manual pump is operated.
CAUTION — The auxiliary hand pendant does not have the following functions: AUTO DRIVE, HI/LO, FLEX, REFLEX, UROLOGY, CHAIR, RETURN TO LEVEL, and SLIDE. This pendant is intended to be used for completing procedures, followed by charging the batteries or servicing the table.
700000120 Rev. 1
1-7
OPERON D 850 SURGICAL TABLE GENERAL INFORMATION
Rev. 1 700000120 1-8
OPERON D 850 SURGICAL TABLE GENERAL INFORMATION
700000120 Rev. 1
1-9
OPERON D 850 SURGICAL TABLE GENERAL INFORMATION
TABLE 1-1 TECHNICAL DATA
Electrical Power Requirements:
115 V AC +/-10% 50/60 Hz, 6A, Single-phase.
Environmental and Storage Requirements:
Operation
Ambient temperature range of 50 to 100° F (10 to 38° C), relative humidity range of 10% to 95%, at altitudes from sea level to 8000 ft.
Transport and Storage
Ambient temperature range of –30 to +150° F (-34 to 66° C) relative humidity range of 0% to 95% at altitudes from sea level to 50,000 ft.
Table Top Radiolucent and conductive carbon fiber panels conform to 21 CFR
Subchapter J, Radiation Control for Health and Safety Act of 1968.
Dimensions:
Length 82.6 inches (1930 mm)
Width 21 inches (533 mm); excluding the side rails.
Weight Approximately 750 lbs. (340 kg)
Height Range 22.6” to 46.4” (573 mm to 1179 mm)
Head Section 12.2” (310 mm) X 21” (533 mm)
Back Section 22.6” (574 mm) X 21” (533 mm)
Seat Section (w/kidney
23.1” (587 mm) X 21” (533 mm)
elevator)
Kidney Elevator 4.3” (109 mm) X 21” (533 mm)
Leg Section 24.6” (625 mm) X 21”(533 mm)
Range of Motion:
Trendelenburg
Reverse Trendelenburg
Lateral Tilt
Back Section
Up to 30°
Up to 31°
Up to 20° right or left
From 45° below horizontal to 90° above horizontal
Kidney Elevator Extendable up to 3.7” (76 mm)
Leg Section
Headrest
From 30° above horizontal to 105° below horizontal
Adjustable from 90° below horizontal to 90° above horizontal in 15° increments
Longitudinal Slide 16.8” (427mm) total
Flex
Chair
226°
Leg Section up to 105°, Back Section up to 90° relative to the Seat Section, which may be up to 30° in the Trendelenburg position.
Load Capacity:
Lifting and Articulation 1000 lbs. (453 kg) all articulations (500 lbs. if patient is reversed)
If you need assistance, call Berchtold Technical Service at 800-243-5135. .
Rev. 1 700000120 1-10
OPERON D 850 SURGICAL TABLE OPERATING INSTRUCTIONS
OPERATING INSTRUCTIONS
PINCH POINTS
WARNING — Cutting or crushing
WARNING
CAUTION
injury may occur if any part of a patient or staff member is caught in a pinch point when the table is articulated.
CAUTION — Damage to the table or object may occur if any foreign object is caught in a table pinch point when the table is articulated.
Due to the variety of table positions attainable with the OPERON D 850 table, pinch or crushing points exist (Figure 2-1). The OR staff must be trained to ensure that body parts or surgical apparatus are not caught in pinch points when the table is operated.
This list is not all-inclusive, because pinch points can be created by the attachment of accessories. Always ensure that the patient and staff are clear of all pinch points before moving the table or while the table is in motion.
Pinch points include but are not limited to:
1. Kidney elevator to seat top
2. Kidney elevator to seat channels
3. Kidney elevator to back section
4. Headrest to back section
5. Back / Leg section link to seat channel
6. Side shields to column shields (in tilt)
7. Column shields to base cover
8. Base to floor
9. Floor feet to base
10. Floor feet to floor
11. Leg section to column shield
12. Foot pump handle to floor
13. Back Section to seat top.
13
5
11
12
1 2 3
4
5
6
7
10
8
9
Figure 2-1 Pinch Points
Rev. 1 700000120 2-1
OPERON D 850 SURGICAL TABLE
OPERATING INSTRUCTIONS
ELECTROMAGNETIC INTERFERENCE
WARNING — When using the table in
WARNING
the vicinity of high-frequency surgical equipment, such as cardiac defibrillators and cardiac defibrillator monitors, refer to the equipment manufacturers’ instructions to ensure compatibility issues are considered.
This surgical table has been tested to EN 60601­1-2 to ensure proper electromagnetic compatibility. Other products used in the vicinity of the surgical table should also comply with this standard. If they do not comply, interference between products resulting in unintended responses may occur. Please contact the appropriate manufacturer(s) if any problems arise.
If the integrity of the external protective earth conductor arrangement is in doubt, the surgical table should be operated from its internal electrical power source (batteries) until the situation is corrected. The batteries require charging at least once a week. If the integrity of the external protective earth conductor arrangement is in doubt, and the batteries are not charged, operate the surgical table with the manual foot pump and the auxiliary hand pendant until the situation is corrected.
LOAD CAPACITY
WARNING — Overloading may result
WARNING
in tipping of the table or failure of the table sections to move. Do not exceed the maximum load.
Left and right, refer to the patients left
and right.
Figure 2-2 Table Orientation
PATIENT POSITIONING
WARNING — POSSIBILITY OF
WARNING
TIPPING: If FULL SLIDE FORWARD IS REQUIRED, the patient’s pelvis must always be located over the seat section. Utilizing full forward slide while the patient’s pelvis is closer to the back section can impair table stability and result in patient injury or equipment damage.
When the patient is placed in the Lithotomy position on the seat section, the following precautions must be followed to ensure safety and stability:
The table must be securely locked to the
floor and not mobilized while the patient is in this position.
WARNING — The leg section may be
WARNING
removed from the table. If it is left on the table, use caution when raising the leg section while the patient is in the normal perineal position.
The patient lifting and articulation capacity is 1000 lbs (453 kg). This means that a 1000 lb patient can be moved into any of the standard procedural positions.
WARNINGWeight Limit in
WARNING
reverse, while not recommended, it is possible to place a patient in reverse orientation. Patient weight must not exceed 500 lbs. when placed in reverse.
PATIENT ORIENTATION
Orientation is defined as follows:
The head end is the end where the power
cord is connected.
The Perineal cutout in the seat section is
exposed
Rev. 1 700000120 2-2
Figure 2-3 Normal Perineal Position
Certain ENT, Neuro, and Ortho procedures require that the table be moved after
OPERON D 850 SURGICAL TABLE OPERATING INSTRUCTIONS
anesthetizing the patient. If this is required, the following steps must be followed:
Be sure that the patient is properly
secured to the table with straps.
Center (slide) the patient over the main
lift column.
Lower the table to its lowest point.
Check the floor for and remove objects
that may interfere with moving the table, such as power cords or table accessories.
Have at least one person at each end of
the table.
Unlock the table.
When moving the table, the people at
each end must use a lateral force in the direction of motion; do not lift up on a section or push down on a section.
Immediately lock the table when the new
location is reached.
OPERATION
CAUTION — Storage of clamps,
CAUTION
NOTE
If desired, plug the table’s power cord into a mains power source. When the table is plugged in via the power cord, it is powered by mains AC. When the power cord is not plugged in, the batteries provide power to the table. In either case, the ON/OFF switch on the base must be ON to articulate any table sections. When the ON/OFF switch on the base is OFF, the table’s electronics and pump motor are isolated from the power supply and the batteries.
When the charge indicator lamp near the power cord receptacle is lit, the table senses mains AC and activates the battery charging system. Battery charging will occur whether the ON/OFF switch on the base is in the ON or OFF position. The table can remain plugged into AC indefinitely without damaging the batteries.
accessories, etc., on the base cover may result in damage to lift column shields and prevent proper elevation or descent of the table.
Note — The table can be operated with or without the power cord connected to the mains inlet receptacle. Plug in the power cord if the batteries need recharging. The table may be operated while the batteries are charging.
HAND PENDANT OPERATION
The table may be activated by pressing the hand pendant ON button or by pressing any of the hand pendant function buttons.
If you activate the table by pressing the ON button, the hand pendant LEDs for all Battery levels will flash briefly. This is a self-test function. During this time, no function is active. If there is a problem with the hand pendant, the connection to the table, or any of the table functions, the SERVICE LED will remain on.
If there are no problems, all LEDs except the Power, current floor lock status, and battery status displays will turn off, and the function buttons will become back lit to indicate that the table is in a “controls active” status.
To conserve battery power, the hand pendant will automatically go into SLEEP MODE after a preset period of non-use (default value). This time can be changed through the table’s Service Software Program, which is described in the “Service Program” section of this manual. Additionally, hand pendant backlighting will turn off after a preset period of non-use (default value). This time can also be changed through the table’s Service Software Program.
If you activate the hand pendant from the SLEEP MODE by pressing any of the function buttons (other than the ON button), the self-test will be performed after the button is released. If the floor locks are NOT locked, the back lighting will come on, and the hand pendant will go into “controls active” status, the feet will automatically descend and lock.
1. Press the ON button on the control
pendant to activate the control pendant
2. Press FLOOR LOCK to lower the floor
feet. The green LOCK indicator on the control pendant will light when the floor feet are locked.
3. Select the desired function by pressing
and holding the appropriate function switch.
4. When the desired position has been
reached, release the function button to stop movement.
5. After the patient has been transferred
from the table, press FLOOR LOCK to raise the floor lock feet and allow table transport. The yellow UNLOCK
700000120 Rev. 1
2-3
Loading...
+ 57 hidden pages