Berchtold Elektrotom 630 Service manual

4.5 (6)

ELEKTROTOM® 630

Service Manual (E)

Valid for version 1170, 1171

CONTENS

 

PAGE

1.

General

4

- 6

 

1.1

Introduction

4

 

 

1.2

Manufacturer’s notes

4

 

 

1.3

General information

4

- 5

 

1.4

Notes on product responsibility

5

 

 

1.5

Routine checks following delivery

5

 

 

1.6

EC certification

5

 

 

1.7

Repairs

5

- 6

 

1.7.1

Replacement of fuses

5

- 6

 

1.8

Technical safety controls

6

 

 

1.9

Type plate identification

7

 

2.

Commissioning

7

- 9

 

2.1

Installation

7

 

 

2.2

Important notes for fafe usage

7

- 8

 

2.3

First usage

8

 

 

2.4

Visual and functional checks before each use

8

 

 

2.5

Cleaning, disinfecting and sterilisation

8

- 9

 

2.5.1

Cleaning and disinfection of the unit

8

 

 

2.5.2

Cleaning, disinfection and sterilisation of accessories

8

- 9

3.

Operating the ELEKTROTOM® 630

10

- 15

 

3.1

Puhsbuttons, signal lights and symbols

10

- 15

 

3.1.1

The front of the ELEKTROTOM® 630 version 1170

10

 

 

3.1.2

The front of the ELEKTROTOM® 630 version 1171

11

 

 

3.1.3

The rear of theELEKTROTOM® 630

12

 

 

3.1.4

Description of the puhsbuttons and symbols

12

- 15

 

3.1.5

Symbol markings on the rear of the unit

15

 

4.

Technical description

15

- 17

 

4.1

Technical data

15

- 16

 

4.1.1

Mains connection

15

 

 

4.1.2

HF current output and current characteristics

15

- 16

 

4.2

Safety relevant data

17

 

 

4.3

Equipment for user support

17

 

 

4.4

Dimensions and weight

17

 

 

4.5

Certification

17

 

5.

Power output diagram

18

- 24

 

5.1

Power output in relation to resistance (output characteristic)

18

- 21

 

5.1.1

Current type CUT I

18

 

 

5.1.2

Current type CUT II

18

 

 

5.1.3

Current type CUT III

19

 

 

5.1.4

Current type CONTACT COAGULATION

19

 

 

5.1.5

Current type FORCED COAGULATION

20

 

 

5.1.6

Current type SPRAY COAGULATION

20

 

 

5.1.7

Current type BIPOLAR COAGULATION

21

 

2

ELEKTROTOM® 630 valid for version 1170, 1171 Service Manual (E)

CONTENS

 

PAGE

 

5.2

Peak voltage in relation to power regulator

21

- 24

 

5.2.1

Current type CUT I

21

 

 

5.2.2

Current type CUT II

22

 

 

5.2.3

Current type CUT III

22

 

 

5.2.4

Current type CONTACT COAGULATION

23

 

 

5.2.5

Current type FORCED COAGULATION

23

 

 

5.2.6

Current type SPRAY COAGULATION

24

 

 

5.2.7

Current type BIPOLAR COAGULATION

24

 

6.

Positiones of spare parts

25

- 26

 

6.1

Valid for version 1170

25

 

 

6.2

Valid for version 1171

26

 

7.

Spare part list

27

 

8.

Adjustment neutral electrode control

27

 

9.

Trouble shooting and repair

28

 

10.

Check of the HF power

28

 

11.

Function and connection diagram

29

 

12.

Power board

30

 

13.

Front board

31

 

14.

Tonegenerator board

32

 

ELEKTROTOM® 630 valid for version 1170, 1171 Service Manual (E)

3

1. GENERAL

1.1 Introduction

Electrosurgery belongs to the most important energy applications in surgery. In all surgical disciplines, it remains the most effective means of combining tissue cutting and haemostasis. The electrosurgical unit, the ELEKTROTOM® 630 electrosurgical unit is a modern instrument combining economy with a considerable degree of operative flexibility. The unit has a capable output of 300 Watts.

Particularly important features of the ELEKTROTOM® 630 include:

Enhanced user simplicity and safety via integrated microprocessor technology

Three current types offering excellent cutting qualities combined with simultaneous haemostasis

Bipolar coagulation with up to 95 Watts output

Contact coagulation up to 120 watts, effective forced coagulation (standard) up to 150 watts, and spray coagulation up to 100 watts.

User friendly control panel with foil covered pressure pads and digital read-out display

The highest degree of safety using modern, negative electrode monitoring technology, for both single and split negative electrodes

A malfunction, error code display system, showing user information and offering assistance during service and repair procedures

All the advantages of the ELEKTROTOM® 630 can be put to optimum use by the user if these instructions are observed. These will assist you in competent operation, correct selection and handling of your HF instruments.

1.2 Manufacturer´s notes

The manufacturer of the products specified in the service manual is

BERCHTOLD Holding GmbH Ludwigstaler Straße 25 Postfach 4052

D-78505 Tuttlingen

Internet: http://www.BERCHTOLD.de

e-mail: BERCHTOLD.Medizinelektronik@BERCHTOLD.de Tel. (+49) 7461 / 181-0

Fax (+49) 7461 / 181-200

1.3 General information

Precise observance of the user’s manual is a prerequisite for the proper use and correct operation of the equipment, which is essential for the safety of patients and operators alike.

Only accessories which are specified in this user’s manual and which have been tested together with the equipment may be used. If accessories are used which are not specified in the user’s manual, their ability to be used in accordance with safety regulations must be proved.

4

ELEKTROTOM® 630 valid for version 1170, 1171 Service Manual (E)

All literature relates to the equipment model and the prevailing basic safety regulations when printed. All rights are reserved for equipment, switches, procedures, software programs and names.

1.4 Notes on product responsibility

The BERCHTOLD company can only consider themselves responsible for the safety, reliability and function of the product under the following conditons:

a)installation, modifications or repairs have been performed only by BERCHTOLD or by an agent expressly authorized by BERCHTOLD to do so,

b)the electrical installation of the room complies with regulations IEC 60364-710,

c)the product is used in accordance with the operating instructions.

1.5 Routine checks following delivery

The product and accessories should be inspected for possible transport damage or other defects immediately on arrival.

Reclamation regarding damage or defects can only be entertained by the selling organisation (BERCHTOLD GmbH & Co.) or the delivering agency when they are immediately reported. In case of complaint, the forwarding agent or the BERCHTOLD sub-agency must immediately be informed, prior to the submission of a damage / deficiency report to the BERCHTOLD main offices in Germany for further processing by our insurance agents. When returning a unit or one of its components to BERCHTOLD or to a BERCHTOLD service centre, every effort should be made to use the original packaging material. The following information/documentation must also accompany the returned items:Name and address of the owners, product identification number (see plate affixed to unit), Detailed description of the defect.

1.6 EC certification

The equipment complies with the requirements of the EC guideline regarding medical products, 93/42/EEC.

1.7 Repairs

By obvious defect, either of the unit or its connecting cable, it must be repaired or its cable renewed before being used again.

The ELEKTROTOM® 630 may only be repaired by BERCHTOLD or their officially appointed agent.

Should the unit be repaired by an officially appointed agent, the user is required to obtain written confirmation of the work carried out.

This signed confirmation should bare the date of the repair and the details of the officially appointed agent. When repairs are not carried out by BERCHTOLD direct, the repairing organisation must append their details to the unit or, that part of the unit which has been repaired.

1.7.1 Replacement of fuses

The mains fuses is to be found the fuse housing (35) below the mains cable socket (36) at the rear of the unit.

ELEKTROTOM® 630 valid for version 1170, 1171 Service Manual (E)

5

Changing the fuse is performed as follows:

Remove the mains cable from its socket (36),

Using a small screw driver, prise open the flap covering the fuse housing (35) and fold it upwards,

Remove the red fuse holder and replace the defective fuse. Take care to use a fuse of the correct value which is written on the fuse holder. T 6,0 A,

Re-insert the red fuse holder and close the flap before re-inserting the mains cable in the socket (36).

1.8 Technical safety controls

The following controls must be carried out at least on a yearly basis:

Visual checking for any mechanical or functional defect

Safety relevant markings on the unit must be readable

Checking of the mains fuses against nominal electrical value

Checking the calibration of the HF current output against the setting of the UP/DOWN button of the control panel

The actual output measurement for the current modes and the coagulatory modes should be checked to the values the laid down in the specifications for the unit.

Checking of optical and acoustic signalisation

Compare protected resistance according to EN 60601-1

with mains connection:

Limit 0.2 Ω

• Measure case leakage according to EN 60601-1

Limit 0.50 mA (N.C.) *

 

Limit 1.00 mA (S.F.C.) **

• Gehäuseableitstrom lt. EN 60601-1 messen:

Limit 0.10 mA (N.C.) *

 

Limit 0.50 mA (S.F.C.) **

• Measure patient leakage according to EN 60601-1

Limit 0.01 mA (N.C.) *

 

Limit 0.05 mA (S.F.C.) **

• Measure patient leakage according to EN 60601-1

Limit 0.05 mA (S.F.C.) **

(Mains voltage at used instrument)

 

* N.C. = Normal condition

 

** S.F.C. = Single fault condition

 

The results of the technical safety checks should be documented.

Should the unit prove to be defective or otherwise unsafe it must not be used until repaired.

6

ELEKTROTOM® 630 valid for version 1170, 1171 Service Manual (E)

1.9 Type plate identification

1Version number

2Variant

3Manufacturing year

4 Serial number

2. COMMISSIONING

2.1 Installation

The unit is intended for use only in a medical environment and connection to the mains must be in accordance with the IEC 60364-710 regulations. Further, connection to the mains should be via a suitably protected socket using the mains lead and plug provided by the manufacturer or one of an equivalent quality. For safety reasons, extension leads or multi-socket connections should not be used. The mains socket must be protected by a fuse rated at not less than 10 Amperes.

The ELEKTROTOM® 630 can be placed on any flat surface with a tilt angle not in excess of 10°. The surface itself should be equivalent in size to that of the unit. Care must be taken not to block the air vents on the underside of the generator and ensure a free flow of air around the unit. The ELEKTROTOM® 630 should be protected from the danger of fluids entering the unit.

For intracardiac surgery this equipment must be connected to the main power stabilizer in the operating room or location where it is installed by means of the (yellow/green) power-stabilization cable supplied.

2.2 Important notes for safe usage

Misuse of the generator and a disregard of these instruction can lead to serious injury!

Take care to study these instructions supplied with your ELEKTROTOM® 630.

Warning!

The unit is not intended for use in explosion endangered areas.

Caution must be exercised when anaesthetic gas mixtures such as Oxygen (O2) and nitrous Oxide (N2O) are used during surgery in the thoracic or head regions. The use of anaesthetic gas, exhaust management systems is to be recommended. Inflammable substances used for cleaning or disinfection or, particularly, solvents used to remove adhesives, must be removed or fully evaporated before the using an electrosurgical unit. The danger of pocketing or pooling of inflammable liquids or vapours in body cavities such as the navel or vagina as well as in the depths of surgical wounds which must also be considered and not underestimated. Liquids must not be allowed to gather or pool under the patient. The presence of endogenic gases which may be ignited, must also be taken into account when using electrosurgical equipment on the gut and a system of inert gas flushing is recommended. Material such as cotton wool or gauze can, in certain circumstances, also be ignited via HF current induced sparking - particularly in the presence of oxygen.

The use of electrosurgery requires caution and the following rules should be considered:

The high frequency current output of the unit should be minimal and not more than is required for the task to be performed.

ELEKTROTOM® 630 valid for version 1170, 1171 Service Manual (E)

7

Note:

A reduced or lack of function after setting the unit output controls at 'normal' power can be caused by a number of factors such as neutral electrode problems, bad connections, damaged cables or a crusted active electrode. Theses point should be considered before selecting what might be a much higher unit output than necessary.

Do not attempt to test the unit by directly discharging against a metal object or the negative electrode.

The function of other electromedical equipment can be interfered with by the use of high frequency current.

The switching mechanisms of an electrode handle which is not completely water tight, may be penetrated by blood, saline or other rinsing liquids or amniotic fluid producing an unpredictable response from the generator.

In order to prevent accidental HF current burns, the electrode handle should be placed on the instrument trolley when not in use and not on the patient.

Placing a finger switched, electrode handle onto very damp drapes or, into pooled liquid on the drapes, may cause patient burns directly below the electrode handle.

2.3 First usage

Before the unit is first used surgically, the Manufacturer or their official agent shall:

b)have fully tested the unit in the position in which it is to be used;

c)have given full operational instructions for the unit to a responsible person.

2.4 Visual and functional checks before each use

Before each use the user must be sure that the unit and its accessories are in good working order.

The following visual checks should be made:

check for external damage to the unit, insulation and plugs,

check that the appropriate accessories a present and that they fit,

Very carefully check the insulation on endoscopic instruments.

Damaged or doubtful equipment must not be used.

Warning!

Should the flow of HF current be indicated by the unit without the attachment of a foot-switch or electrode handle with a double, finger switch then the unit is faulty and must be examined before use. An indicated malfunction following the attachment of a foot-switch or electrode handle with double, finger switch shows a defective accessory which must be checked and eventually replaced.

2.5 Cleaning, disinfecting and sterilisation

2.5.1 Cleaning and disinfecting of the ELEKTROTOM® 630

The entire exterior of the unit, including the foil covered operating panel, can be cleaned with normal, alcohol free cleaning fluids. (Spray or wipe disinfecting)

Please take note of the manufacturers instructions for disinfectant solutions.

8

ELEKTROTOM® 630 valid for version 1170, 1171 Service Manual (E)

2.5.2 Cleaning, disinfecting and sterilisation of accessories

After use, accessories may be soaked in standard disinfecting solutions following the instructions of the manufacturers, without exceeding soaking times. The life expectancy of some plastics may be shortened by certain chemicals and a thorough rinsing of all accessories is important. Phenol and chlorine solutions are not suitable. Alternatively, a mechanised washing and thermal disinfecting process is acceptable, provided temperatures do not exceed 93° C/10 min..

Good operative results can only be expected when the active and negative electrode are perfectly clean and free from any dried protein.

Connecting cables and the insulation of active electrodes must be constantly checked and maintained in perfect condition. Articles with damaged insulation must not be used.

The following sterilisation temperatures are acceptable:

 

Gas sterilisation

Steam sterilisation

Hot air sterilisation

 

up to 70 °C

up to 134 ° C

at 200 ° C

Connecting cables for

 

 

 

electrode handle

yes

yes

no

Electrode handle

yes

yes

no

Active electrodes

yes

yes

yes

Bipolar coagulation

 

 

 

forceps

yes

yes

no

ELEKTROTOM® 630 valid for version 1170, 1171 Service Manual (E)

9

3. OPERATING THE ELEKTROTOM® 630

3.1 Puhsbuttons, signal lights and symbols

3.1.1 The front of the ELEKTROTOM® 630 version 1170

1Symbol indicating „split, neutral electrode applied“

2Symbol indicating „single, neutral electrode applied“

3Signal light red for negative electrode alarm

4Button to confirm the correct application of a negative electrode (SET button)

5Service indication: when this lights up, a service technician should inspect the unit within one week

6Button for Cut I “cutting with minimal tissue charring”

7Button for Cut II “cutting with medium tissue charring, also for “transurethral resection”

8Button for Cut III “cutting with wire loop electrodes for polypectomy and papillotomy”

9Digital output indicator in watts for cutting power

10Signal light indicating activation - cutting current

11Button to increase cutting output

12Button to decrease cutting output

13Button for contact coagulation

14Button for forced coagulation

15Button for spray coagulation (ray of sparks coagulation)

16Digital output indicator in watts for coagulatory power

17Signal light indicating activation - coagulatory current

18Button to increase coagulating output

19Button to decrease coagulating output

20Button for double foot-switch use in either monopolar or bipolar coagulation

21Digital output indicator in watts for bipolar coagulatory power

22Button to decrease bipolar coagulating output

23Signal light indicating activation - bipolar coagulatory current

24Button to increase bipolar coagulating output

25Button for bipolar coagulating via forceps activation (auto-start mode)

26Mains, on/off switch

27Connecting socket for neutral electrode (split or single)

28Connecting socket for electrode handle, HF output 1 (with additional socket for disposable handles)

29Connecting socket for electrode handle, HF output 2 (with additional socket for disposable handles)

30Connecting socket for bipolare coagulations instruments

31Connecting socket for foot-switch (for HF output 1 and bipolar coagulation)

10

ELEKTROTOM® 630 valid for version 1170, 1171 Service Manual (E)

Berchtold Elektrotom 630 Service manual

3.1.2 The front of the ELEKTROTOM® 630 version 1171

1Symbol indicating „split, neutral electrode applied“

2Symbol indicating „single, neutral electrode applied“

3Signal light red for negative electrode alarm

4Button to confirm the correct application of a negative electrode (SET button)

5Service indication: when this lights up, a service technician should inspect the unit within one week

6Button for Cut I “cutting with minimal tissue charring”

7Button for Cut II “cutting with medium tissue charring, also for “transurethral resection”

8Button for Cut III “cutting with wire loop electrodes for polypectomy and papillotomy”

9Digital output indicator in watts for cutting power

10Signal light indicating activation - cutting current

11Button to increase cutting output

12Button to decrease cutting output

13Button for contact coagulation

14Button for forced coagulation

15Button for spray coagulation (ray of sparks coagulation)

16Digital output indicator in watts for coagulatory power

17Signal light indicating activation - coagulatory current

18Button to increase coagulating output

19Button to decrease coagulating output

20Button for double foot-switch use in either monopolar or bipolar coagulation

21Digital output indicator in watts for bipolar coagulatory power

22Button to decrease bipolar coagulating output

23Signal light indicating activation - bipolar coagulatory current

24Button to increase bipolar coagulating output

25Button for bipolar coagulating via forceps activation (auto-start mode)

26Mains, on/off switch

27Connecting socket for neutral electrode (split or single)

28Connecting socket for electrode handle, HF output 1, activation with foot-switch

29Connecting socket for electrode handle with 3-pin plug, HF output 2

30Connecting socket for bipolare coagulations instruments

31Connecting socket for foot-switch (for HF output 1 and bipolar coagulation)

38Slide for release of socket 28 or socket 39

39Connecting socket for electrode handle with 3-pin plug, HF output 1

ELEKTROTOM® 630 valid for version 1170, 1171 Service Manual (E)

11

Loading...
+ 23 hidden pages