Pre-Install Manual: 700000006
Related Publications:
700000088 Operator Manual
This manual contains proprietary information of BERCHTOLD Corporation. It shall not be reproduced in
whole or in part without the written permission of BERCHTOLD Corporation.
ETL/CETL LISTED TO IEC 60601-1. MEDICAL ELECTRICAL EQUIPMENT
CHROMOPHARE® IS A REGISTERED TRADEMARK OF BERCHTOLD CORPORATION.
PRE-INSTALLATION GUIDE
TABLE OF CONTENTS
TABLE OF CONTENTS
INTRODUCTION & GENERAL INFORMATION....................................................................................................................2
SUPERSTRUCTURE DESIGN AND SPECIFICATION INFORMATION.............................................................................3
RECOMMENDED DESIGN SPECIFICATIONS........................................................................................................................................3
MINIMUM DESIGN SPECIFICATIONS BY PRODUCT TYPE....................................................................................................................3
ELECTRICAL REQUIREMENTS...............................................................................................................................................5
ELECTRICAL REQUIREMENTS FOR STANDARD MOUNTING...............................................................................................................5
STANDARD MOUNTING – DETAILS (SEE PAGE 6 FOR LAYOUT DIAGRAM) ......................................................................................6
LOW PROFILE MOUNTING – LAYOUT DIAGRAM ...............................................................................................................................7
LOW PROFILE MOUNTING – DETAILS (SEE PAGE 7 FOR LAYOUT DIAGRAM)...........................................................................8
FLAT PANEL PRODUCTS (WITH AND WITHOUT LIGHTS)..................................................................................................................9
WALL CONTROL BOX DIMENSIONS....................................................................................................................................10
SK BOX DIMENSIONS ...............................................................................................................................................................10
PRE-INSTALLATION GUIDE PUBLICATION HISTORY...................................................................................................11
PUBLICATION HISTORY ................................................................................................................................................................11
REV. 2 700000006
Page 1
PRE-INSTALLATION GUIDE
INTRODUCTION & GENERAL INFORMATION
INTRODUCTION & GENERAL INFORMATION
This Pre-Installation Guide details and describes the structural, mechanical, and electrical requirements of
the room or site to allow proper installation of BERCHTOLD CHROMOPHARE® D-Series Surgical
Lights and Flat Panel products. (D-Series Surgical Lights include D650, D660, D530, and D540.)
It is intended for use by technical individuals involved in the specification, design, and construction
of the room or site, such as architects, structural engineers, contractors, electricians, and pipe fitters.
The Hospital and/or its contractors are responsible for the preparation of the room or site including the
superstructure design and installation and all other mechanical and electrical requirements described within
this guide and those specified by AHJ (Authorities Having Jurisdiction).
Painting and flooring need to be completed before the product can be installed.
Questions pertaining to the information contained in this guide can be addressed by contacting your local
BERCHTOLD Technical Service Representative or BERCHTOLD Technical Service at 1-800-243-
5135.This guide is applicable to the following BERCHTOLD products:
CHROMOPHARE® D650 and D660
CHROMOPHARE® D530 and D540
All multiple combinations of the CHROMOPHARE® D-Series.
Stand alone Flat Panel Products
Flat Panel combined with either one or two CHROMOPHARE® D-Series lights.
700000006 REV. 2
Page 2
PRE-INSTALLATION GUIDE
SUPERSTRUCTURE DESIGN AND SPECIFICATION INFORMATION
SUPERSTRUCTURE DESIGN AND SPECIFICATION INFORMATION
The customer/contractor has responsibility to ensure the superstructure design meets specifications.
BERCHTOLD recommends that a Structural Engineer be involved in the design of the superstructure.
Important: Weights and moments must be adjusted when designing for use in seismic zones. Check
with your local construction-engineering agency for applicable local and state codes.
RECOMMENDED DESIGN SPECIFICATIONS
BERCHTOLD Corporation STRONGLY recommends that the superstructure be designed to meet the
following specifications. (Reference drawing on next page). Designing to these specifications will allow
combinations of Light or Flat Panel product to be hung from the structure.
Weight (T)
Moment (M)
Product Type
lbs.
Light / Flat Panel Combinations
375 1538
ft-lb with less than 0.1 º
Loads to support structure
Case Fp Tmax (LBS) Vmax (LBS) Mmax (Ft-#)
1 2.83 760 1107 7713
2 1.89 6.37 739 5663
3 1.54 592 602 4893
MINIMUM DESIGN SPECIFICATIONS BY PRODUCT TYPE
If the customer decides to design the superstructure to the requirements of the specific product type,
the superstructure design specifications can be found in the table below.
Weight (T)
Moment (M)
Product Type
º
lbs.
ft-lb with less than 0.1
rotation at
Single Light 114 360
Double Light or Double Flat Panel 197 656
Triple Light 311 1016
Flat Panel only 114 360
Single Light with Flat Panel 197 656
Double Light with Flat Panel 311 1016
Single Light with Double Flat Panel 311 1016
REV. 2 700000006
Page 3