Bayer DCA 2000+ User manual

Analyzer
Operating
Manual
®
50241739 ii Revised 6/03
© 2003 Bayer HealthCare LLC
All Rights Reserved
Unless otherwise noted, all
®
Trademarks are the property
of Bayer HealthCare LLC
SECTION 1. Introduction Page
System and Intended Use ......................................................................... 1.1
Features .......................................................................................... 1.1
Specifications..................................................................................... 1.1
Hazards........................................................................................... 1.1
Safety Standards...................................................................................1.1
Symbols Used .....................................................................................1.2
SECTION 2. Unpacking, Getting Acquainted and Set Up
Unpacking........................................................................................ 2.1
Carton Contents................................................................................... 2.1
Getting Acquainted
DCA 2000
®
Analyzer (Front Panel) ........................................................... 2.2
DCA 2000
®
Analyzer (Back Panel)............................................................2.3
Screen Saver....................................................................................2.3
Audible Tones.................................................................................. 2.3
Blinking Colon in Displayed Time.............................................................. 2.3
When To Turn the Power Off ................................................................... 2.3
Set Up
Placing the Instrument/Connecting the Power Cord/Inserting Program Card..................... 2.4
Checking to Verify Instrument Functions Properly .............................................. 2.5
Viewing Factory Settings ....................................................................... 2.6
Options ........................................................................................ 2.8
Accepting Factory Settings.................................................................... 2.10
Changing Factory Settings..................................................................... 2.10
Setting Date and Time......................................................................... 2.14
Setting Creatinine Concentration Units (Microalbumin/Creatinine Assay ONLY)............... 2.16
Running the Optical Test Cartridge (Standard 1) —
Prior To Analyzing Samples for the First Time .............................................. 2.17
SECTION 3. Menu
Menu ............................................................................................. 3.1
RECALL PREVIOUS TESTS?................................................................. 3.2
SET SEQUENCE NUMBER?.................................................................. 3.4
RECALL CONTROL RESULTS? .............................................................. 3.5
VIEW CALIBRATION STATUS?.............................................................. 3.8
SET DATE/TIME?............................................................................ 3.10
INSTRUMENT SETUP?...................................................................... 3.13
SET CREATININE UNITS? .................................................................. 3.15
INSTRUMENT TEST? ....................................................................... 3.16
RUN CONTROL?............................................................................. 3.17
TABLE OF CONTENTS
Revised 6/03 iii
SECTION 4. Operating Instructions — Hemoglobin A
1c
Page
Step 1: Turning the Power On ..................................................................... 4.1
Step 2: Calibration ................................................................................ 4.2
Step 3: Preparing Patient Samples and Controls.................................................... 4.3
Step 4: Analyzing the Patient Sample.............................................................. 4.8
Analyzing DCA 2000
®
Hemoglobin A1cControls, ONLY......................................... 4.12
Cancelling a Test ................................................................................ 4.15
SECTION 5. Operating Instructions — Microalbumin/Creatinine Page
Step 1: Turning the Power On ..................................................................... 5.1
Step 2: Calibration ................................................................................ 5.2
Step 3: Preparing Patient Samples and Controls.................................................... 5.4
Step 4: Analyzing the Patient Sample.............................................................. 5.8
Analyzing DCA 2000
®
Microalbumin/Creatinine Controls, ONLY................................ 5.12
Cancelling a Test ................................................................................ 5.15
SECTION 6. Error and Warning Messages, Error Codes and Troubleshooting
Chart of Contents................................................................................. 6.1
Error and Warning Messages ...................................................................... 6.2
Error Codes....................................................................................... 6.4
Troubleshooting ................................................................................. 6.10
SECTION 7. Instrument Care and Routine Maintenance
Instrument Care................................................................................... 7.1
Routine Maintenance Chart ....................................................................... 7.1
Exterior of Instrument and Bar Code Window (includes disinfection) .............................. 7.2
Changing Air Filter ............................................................................... 7.3
Cartridge Compartment ........................................................................... 7.4
Running Optical Test Cartridge.................................................................... 7.6
SECTION 8. Minor Repair
How to Replace the Fuse.......................................................................... 8.1
SECTION 9. Service Information
How to Report the Problem ....................................................................... 9.1
Accessory Items and Replacement Parts........................................................... 9.3
APPENDIX.................................................................................... App.1.1
iv Revised 6/03
INTR ODUCTION
Revised 6/03 1.1
The DCA 2000®System consists of:
DCA 2000® Analyzer
DCA 2000®Reagent Cartridges, Capillary Holders and Calibration Card
DCA 2000®Controls
Optical T est Cartridge
INTENDED USE
The DCA 2000® Analyzer quantitatively measures:
the percent concentration of Hemoglobin A
1c
in blood
the concentrations of Microalbumin and Creatinine in urine
FEATURES
easily calibrated using the bar code card provided with each reagent kit
stores calibrations for up to two lots of reagent
stores up to 16 test results
stores up to 16 control results
can be connected to a computer and/or printer
SPECIFICATIONS
Power Required:
Instrument Model No. 5031C: 100–240 VAC,
@ 0.4 amps 50/60 Hz
Dimensions/Weight:
Depth 27.2 cm (10.7 in) without power line cord Width 24.1 cm (9.5 in) Height 23.9 cm (9.4 in) Weight 5 kg (11 lbs)
Ambient Operating Temperature:
(15–32°C)—Hemoglobin A
1c
(18–30°C)—Microalbumin/Creatinine
Relative Humidity Range:
10% – 90% RH (non-condensing)
NOTE: If an instrument I/O port is utilized, the cable used should be 100% shielded to guard against EMI and RFI. There should be continuity between the cable shield and connector shell that mates with the instrument.
Instrument Safety Design
The instument is for professional, in vitro diagnostic use and must be used in the manner specified in the Operating Manual in order to provide the safety and standard performance standards specified. The instrument will operate safely in the conditions
listed below; however, results will only
be correct for the system specifications listed above:
indoors
5°– 40°C (41°– 104°F)
installation category II (IEC 1010.1)
safety tested to comply to IEC 1010.1
HAZARDS
To alert you to potential electrical or operational hazards, warning and caution statements are provided where applicable. To ensure your safety, comply with all warning and caution statements.
SAFETY STANDARDS
Underwriters’Laboratories, Inc. (UL) and the Canadian StandardsAssociation (CSA) as certified and complies with the safety standards specified in UL 3101 and CSA-C22.2, No. 1010.1.
The instrument meets the provisions of the IVD Directive 98/79/EC (Oct./1998), which includes the EMC Directive 89/336 Amendment 92/31/EEC, and the Low Voltage Safety Directive 73/23/EEC.
WARRANTY INFORMATION
Contact your Authorized Bayer Representative for complete warranty information.
1
INTRODUCTION
32
15
18
30
1
INTRODUCTION
1.2 Revised 6/03
Catalog number In vitro diagnostic device (LOT) Batch code Consult operating instructions (EXP) Use by Warning/Attention: consult
instructions for use
Temperature limitations
Do not use spray
Normal control
Abnormal control
High control
Low control
Reconstitution fluid
High temperature part
Fuse
Quantity
Reagent cartridge Menu/Next Increase Decrease Enter Escape Manufactured by
Authorized representative
Calibration card
Optical test cartridge
Do not freeze
SYMBOLS USED
The following symbols are used throughout the product labeling for the DCA 2000®System (Instrument, Instrument Manual, Quick Reference Guide, DCA Reagent Labeling and Instructional inserts and the DCA Control packaging and instructional inserts).
U.S. PATENTS
6,043,043, 5,822,071, 5,610,073, 5,385,847, 5,372,948, 5,305,093, 5,272,093, 5,258,311, 5,220,161, 5,162,237, 5,151,369, 5,084,397, 4,990,075, 4,970,171, 4,968,472, 4,898,824, 4,847,209, 4,727,036, 4,658,022, 4,647,654, 4,629,692, D400,673.
2
8
EC REP
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
Revised 6/03 2.1
UNP ACKING
The DCA 2000®System arrives in one shipping carton.
Inspect the carton for shipping damage. Unpack the carton.
Use extreme care when unpacking and handling the instrument. The instrument contains sensitive electronic and optical parts.
Check each item for shipping damage.
•Report shipping damage to the representative of the carrier or to your Bayer HealthCare Sales Representative.
CAR T ON CONTENTS
*Some instruments will require a separately packed Installation Kit
containing language specific instructional materials (Program Card, Operating Manual and Quick Reference Guide), Cleaning Kit, Air Filter Replacement Kit and the appropriate Power Cord.
DCA 2000® Analyzer
Cleaning Kit*
Optical T est Cartridge (in instrument cartridge compartment)
Program Card*
Power Line Cord*
Other (not shown)
• Operating Manual*
•Replacement Fuse (stored in fuse holder located inside the instrument)
• Air Filter Replacement Kit*
• HbA
1c
Quick Reference Guide*
6
5
4
3
2
1
UNPACKING, GETTING ACQUAINTED AND SET UP
1
2
5
4
3
Missing Items?...Contact the nearest Bayer HealthCare office or authorized distributor.
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
2.2 Revised 6/03
UNPACKING, GETTING ACQUAINTED AND SET UP
Unlock and then remove the optical test cartridge
according to the following instructions.
1. Open the cartridge compartment door.
2. Locate the button on the right side of the cartridge compartment. Push and hold it down with your right hand.
3. With your left hand, gently push the plastic tab on the cartridge to the right; this action releases (unlocks) cartridge.
4. Pull test cartridge out of compartment.
5. Put test cartridge aside for later use (page 2.17).
6. Make sure the cartridge return spring inside the cartridge compartment is intact (refer to Section 7, page 7.4, for information regarding this spring).
GETTING ACQUAINTED
DCA 2000® Analyzer FRONT PANEL
Display—indicates date, time, error
messages, test results, procedural prompts, etc.
Keys—allow you to provide input for system set-up, menu items, etc.
Reagent Cartridge Compartment Access Door—covers and protects the
reagent cartridge; closing door starts test timing (after a 5 second delay)
Reagent Cartridge Compartment— holds one reagent cartridge during sample analysis
Bar Code Reader Window—covers and protects the bar code reader
Bar Code Track—area where reagent cartridge, calibration card or control card is placed prior to scanning bar code
6
5
4
3
2
1
1
6
5
2
3
4
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
Revised 6/03 2.3
BA CK PANEL
Power Switch—turns the power to the
instrument ON and OFF
Filter Holder—contains replaceable air filter that prevents dust contamination
Program Card Connector—accepts the program card
EIA-232-D Output—accepts the plug for the computer interface cable
Printer Output—accepts the plug for the printer cable
Power Cord Connector—accepts the plug for connecting the power cord to the instrument
Fuse Holder Compartment—holds two fuses (one fuse is the replacement fuse)
AUDIBLE TONES
Beep—a short audible tone; indicates successful
completion of an activity (such as scanning the bar code)
Buzz—a long audible tone; indicates an error
condition or reminds you to perform an activity such as removing a reagent cartridge
SCREEN SAVER
When the instrument is not in use for more than 5 minutes, the display will change to a block shaped moving cursor. If the screen saver is on, press any key to return to a normal display before performing any other steps.
BLINKING COLON IN DISPLAYED TIME
The current time is displayed using a “blinking” colon. When the colon does not blink, the time displayed is the time the assay began.
WHEN TO TURN THE POWER OFF
When the instrument is not in use, the power may be turned OFF without loss of stored results. However, when the po wer is subsequently restored, a warm up period of one to eight minutes is required.
IMPORTANT: If power is turned OFF or interrupted while a test is in progress, the test must be discarded.
Turn the power off when inserting or removing the program card and whenever instructed to do so by the particular procedure (mainte­nance, etc.) in use.
7
6
5
4
3
2
1
1
2
3
5
4
6
7
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
2.4 Revised 6/03
UNPACKING, GETTING ACQUAINTED AND SET UP
SET UP
PLA CING THE INSTRUMENT/ CONNECTING THE POWER CORD/ INSERTING PROGRAM CARD
IMPORTANT: Do not place the instrument where it would be subjected to extreme temperature variations, direct sunlight, excessive humidity or air current, or excessive particulate matter.
1. Place the instrument on a firm, level* surface near a properly grounded electrical outlet.
0. *If the surface is not level, the instrument will
not function properly.
2. Set power switch to OFF (O).
3. Plug in program card (contacts facing instrument; label side up).
0. IMPORTANT: The program card can be
damaged if inserted when the power is ON (I).
4. Connect the power cord to the power cord
connector on the instrument.
5. Plug in power cord to a properly grounded
outlet.
6. Move the instrument into place on the
designated work space.
0.
Allow at least 2 inches of air space between the wall (or other surface) and the back and right sides of the instrument (ventilation panels).
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
Revised 6/03 2.5
CHECKING TO VERIFY INSTRUMENT FUNCTIONS PROPERL Y
Set the power switch to ON (I).
After about 8 seconds, the software version is displayed.
(displayed for about 8 seconds)
Copyright information is displayed for 3 seconds.
Then:
*
The instrument is checking internal optics and proper operation of mechanical features.
Then:
Instrument does not function? Above display(s) fail(s) to appear?...Contact the nearest Bayer HealthCare office or
authorized distributor. Otherwise, continue with “Viewing Factory Settings” (next).
INSTRUMENT SETUP
PRESS [<--] TO CONTINUE
INITIALIZING
KEEP DOOR CLOSED
COPYRIGHT 1991–2003
BY BAYER CORPORATION
SOFTWARE VERSION
E3.11/01.04
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
2.6 Revised 6/03
VIEWING FACTORY SETTINGS— Upon Receipt of a New or Factory-Serviced Instrument
Only upon receipt of a new or factory-serviced instrument, the following display appears (just after “INITIALIZING / KEEP DOOR CLOSED”).
Before the instrument can analyze samples for the first time, it is necessary for you to either accept or change factory settings.
To view factory settings, press .
Line 1 shows the factory setting numbers. Line 2 shows the factory setting options.
The options in Line 2 correspond directly to the number above them in Line 1.
The following chart defines the factory setting now active for each available option.
Before deciding to accept factory settings, review the chart under OPTIONS, on the following pages.
FACTORY SETTING OPTION NUMBER DEFINITION T (Time) 1 AM/PM D (Date) 1 Month/Day/Year L (Labels 4 Time Assay Began;
displayed — (0) Sequence Number (reset with results) —(0) daily, automatically
—(0) at midnight)
C (Controls) 0 Use of DCA 2000
—(0) Controls, only
P (Port) 0 Computer Port is turned
—(0) Off (O)
1 1 4 0 00000000000000 TDLCP0000000000000
INSTRUMENT SETUP
PRESS [<--] TO CONTINUE
UNPACKING, GETTING ACQUAINTED AND SET UP
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
Revised 6/03 2.7
(This page left blank on purpose.
Intended for future use)
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
2.8 Revised 6/03
UNPACKING, GETTING ACQUAINTED AND SET UP
OPTIONS
All available settings are shown in the following chart.
Review this chart before you decide to accept or change the factory setting for each option. An asterisk (*) marks factory settings.
OPTION SETTING NUMBER DEFINITION OF SETTING NUMBER T (Time 1* AM/PM*
Format)
2 24 hour format
D (Date 1* Month/Day/Year*
Format)
2Day
Month•Year
3Year–Month–Day
L (Labels displayed 0 No labels
with results)
1 Sequence number
(reset daily, automatically at midnight)
2 Sequence number
(continuous, 001 – 999)
3Time assay began
4* Time assay began; sequence number
(reset daily, automatically at midnight)*
5Time assay began; sequence number
(continuous, 001 – 999)
C (Controls) 0* DCA 2000 Controls, only*
The control bar code card enables the instrument to label the control result. The control result is then stored in the control memory (separate from patient results).
1 ANY CONTROL
If using controls other than DCA 2000 Controls, use the MENU (just prior to each control assay) to label the control result. The control result, once labeled, is stored in the control memory.
IMPORTANT: When the MENU is used
to label control results, only the next sample analyzed is labeled (and stored in the control memory).
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
Revised 6/03 2.9
OPTION SETTING NUMBER DEFINITION OF SETTING NUMBER P (Computer Port 0* OFF*Computer port is OFF.
Configuration)
2ON Computer port is ON.
All subsequent results will be transferred to the computer.
NOTE: The options below (baud rate through modem control) are selectable only if the computer port is turned ON.
B (Baud Rate) 9* 9600 bps*
4 4800 bps 2 2400 bps 1 1200 bps 3 300 bps
U (Data Bits) 8* 8 bits*
77 bits
V (Parity) 0* None*
1 Odd 2Even
W (Stop Bits) 1* 1*
22
X (Xon—Xoff 0* Off*
Protocol)
1On
Y (Block 0 Off
Transfer)
1* On*
Z (Modem 0* Off*
Control) 1 On
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
2.10 Revised 6/03
ACCEPTING FACT ORY SETTINGS
If the factory setting for each option is acceptable, press . If not, see below. If acceptable, you are
now ready to set the current date and time. Refer to instructions on page 2.14.
CHANGING FACTORY SETTINGS
Before attempting to change factory settings, read the following information on “Keys” and “Cursor.”
Keys:
moves the cursor* (underline) under
setting you desire to change
cycles through each setting choice,
such as AM/PM or 24 HR for time format
accepts all displayed settings (for
time, date, labels, controls and computer port) and immediately advances to “SET DATE/TIME” display.
Important: If you (prematurely) press before you are finished changing factory settings, refer to Section 3, MENU for instructions on how to access “INSTRUMENT SETUP?”
HINT: Don’t press until all settings on Line 1 reflect your choices.
*Cursor
The cursor is the underline in the display. The cursor shows you which setting is ready for change.
Reminder: If the underlined number is acceptable, press . The cursor then moves horizontally (right) to the next number on line 1 while line 2 defines the corresponding option and setting choice. Only when all numbers shown in Line 1 are acceptable, press .
*The cursor does not move in reverse (left).
To change a number to the left of the cursor, repeatedly press until the cursor returns to the desired location.
1 1 4 0 0 nnnnnnnnnnnn TIME FORMAT AM/PM
DISPLA Y WHA T Y OU DO
1. Press (places a cursor under factory
setting for first option, TIME FORMAT).
2. Press and to cycle through choices
for TIME FORMAT.
0. When the desired choice is displayed, press
0. (moves cursor under setting for next
0. option).
1 1 4 0 0 mmmmmmmmmmmm TIME FORMAT AM/PM
1 1 4 0 0 mmmmmmmmmmmm T D L C P mmmmmmmmmmmm
UNPACKING, GETTING ACQUAINTED AND SET UP
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
Revised 6/03 2.11
DISPLA Y WHA T Y OU DO
3. Press and to cycle through choices for
DA TE FORMAT.
0. When the desired choice is displayed, press
0. (moves cursor under setting for next
0. option).
4. Press and to cycle through choices for
LABELS.
0. When the desired choice is displayed, press
0. (moves cursor under setting for next
0. option).
5. Press and to cycle through choices for
CONTROLS.
0. When the desired choice is displayed, press
0. (moves cursor under setting for next
0. option).
6. Press and to cycle through choices for
COMPUTER PORT.
0. If you have selected “0” for COMPUTER
PORT OFF:
0.
Check to make sure all numbers on Line 1 reflect your choices.
0.
Reminder: Options Chart is found on pages
2.8–2.9.
0.
Press (accepts all settings and
0.
advances display to SET DATE/TIME?)
0.
Continue with instructions on page 2.14, “Setting the Date and Time.”
0. If you have selected “2” for
COMPUTER PORT ON:
0.
Continue with instructions on page 2.12, “Changing Factory Settings For: Computer Port ‘ON’Option.”
1 1 4 0 0mm mmmmmmmmmm COMPUTER PORT OFFmmmmm
1 1 4 0 0 mmmmmmmmmmmm CONTROLS DCA 2000
1 1 4 0 0 mmmmmmmmmmmm LABELS USED TIME/DAILY
1 1 4 0 0 mmmmmmmmmmmm DATE FORMAT MM/DD/YY
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
2.12 Revised 6/03
UNPACKING, GETTING ACQUAINTED AND SET UP
CHANGING FA CTOR Y SETTINGS FOR: COMPUTER PORT “ON” OPTION
Refer to the chart on page 2.9 for Computer Port Configurations.
NOTE: The first four numbers in each display (below) are factory settings. The display on your instrument may be different for the first four numbers (depending on whether the first four factory settings were accepted or changed).
DISPLA Y WHA T YOU DO
1. Press (places a cursor under factory
0. setting for first computer port option).
2. Press and to cycle through choices for BAUD RATE.
0. When the desired choice is displayed, press
0. (moves cursor under setting for next
0. option).
3. Press and to cycle through choices for DATA BITS.
0. When the desired choice is displayed, press
0. (moves cursor under setting for next
0. option).
4. Press and to cycle through choices for PARITY.
0. When the desired choice is displayed, press
0. (moves cursor under setting for next
0. option).
1 1 4 0 2 9801010 PARITY NONE
1 1 4 0 2 9801110 DATA BITS EIGHT
1 1 4 0 2 9801010 BAUD RATE 9600
1 1 4 0 2 9801110 COMPUTER PORT ON
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
Revised 6/03 2.13
DISPLA Y WHA T Y OU DO
5. Press and to cycle through choices for
STOP BITS.
0. When the desired choice is displayed, press
0. (moves cursor under setting for next
0. option).
6. Press and to cycle through choices for
XON/XOFF.
0. When the desired choice is displayed, press
0. (moves cursor under setting for next
0. option).
7. Press and to cycle through choices for
BLOCK TRANSFER.
0. When the desired choice is displayed, press
0. (moves cursor under setting for next
0. option).
8. Press and to cycle through choices for
MODEM.
0.
Check to make sure all numbers on Line 1 represent desired settings.
0.
Press (accepts all settings and
0.
advances display to SET DATE/TIME).
1 1 4 0 2 9801110 MODEM OFF
1 1 4 0 2 9821000 BLOCK XFER OFF
1 1 4 0 2 9821010 XON/XOFF OFF
1 1 4 0 2 9821010 STOP BITS ONE
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
2.14 Revised 6/03
UNPACKING, GETTING ACQUAINTED AND SET UP
DISPLA Y WHA T Y OU DO
The display on your instrument is displaying the date and time in your chosen format.
1. Press or until the correct two digits are
displayed.
0. Press .
2. Press or until the correct two digits are
displayed.
0. Press .
3. Press or until the correct two digits are
displayed.
0. Press .
SET DATE/TIME
02/24/03 6:20AM
SET DATE/TIME
02/24/03 6:20AM
SET DATE/TIME
02/24/03 6:20AM
SETTING THE D ATE AND TIME
Use the following keys:
and To cycle through each available setting (e.g.,
digits for day/month/year and time, or AM/PM)
To move cursor under setting you desire to change (e.g., “2” in display below is marked by cursor and indicates “26” is ready for change)
To set the displayed date and time in the instrument
SET DATE/TIME
02/24/03 6:20AM
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
Revised 6/03 2.15
DISPLAY
AM or PM if AM/PM
format is selected
AM or PM if AM/PM
format is selected
WHA T Y OU DO
4. Press or until the correct two digits are
displayed.
0. Press .
5. Press or until the correct two digits are
displayed.
0.
If 24 HR format is selected, press .
0.
If AM/PM format is selected, press .
0.
Then press or until the correct
choice is displayed.
0.
Press .
SET DATE/TIME
02/24/03 6:20AM
SET DATE/TIME
02/24/03 6:20AM
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
2.16 Revised 6/03
UNPACKING, GETTING ACQUAINTED AND SET UP
SETTING CREATININE CONCENTRA TIONUNITS (Microalbumin/Creatinine Assay ONL Y)
The concentration units reported for creatinine are selectable between “mg/dL” and “mmol/L”. The factory setting is “mg/dL”.
DISPLA Y WHA T YOU DO
1. Press .
0.
2. Repeatedly press , until
“SET CREATININE UNITS?” is displayed.
3. Press . This places a cursor below the “m” in
“mg/dL”, and the question mark disappears.
4. Press or to display “mmol/L” or
“mg/dL”.
5. Press to accept desired units.
6. Press to exit the MENU.
SET CREATININE UNITS?
mmol/L
SET CREATININE UNITS
mmol/L
SET CREATININE UNITS
mg/dL
SET CREATININE UNITS?
mg/dL
READY: SCAN BAR CODE
02/24/03 2:09PM
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
Revised 6/03 2.17
The provided optical test cartridge allows you to monitor the performance of the optical system over time. (The optical test cartridge holds the cartridge return spring in place and also simulates the mass of a reagent test cartridge.)
Before samples are analyzed for the first time, run the optical test cartridge.
IMPORTANT! Keep a permanent record of the results obtained (i.e., Mean Transmittance, Standard Deviation and Drift).* These initial values will
be used for comparison, as in control charting (and also to isolate the cause of an instrument malfunction in conjunction with instructions provided by our Bayer HealthCare Customer Service Department Representative).
*It is recommended that you record the results on
the page provided in the appendix of this manual.
After the optical test cartridge is run initially,
it is recommended that the optical test cartridge be run:
quarterly
after cleaning the cartridge compartment
after changing the air filter
when instructed to do so by our Customer Service Representative
NOTE: Refer to Section 7, page 7.6, Instrument Care and Routine Maintenance, for information regarding the comparison of initial values obtained for your instrument with values obtained thereafter.
RUNNING THE OPTICAL TEST CARTRIDGE (Standard 1) —Prior to Analyzing Samples f or the First Time
DISPLA Y WHA T Y OU DO
1. Locate the bar code on the optical test
cartridge.
2. Hold the cartridge so that the bar code faces
right.
3. Insert the cartridge (above dot on instrument)
into the bar code track.
4. Quickly (within 1 second) and smoothly, slide
the cartridge down past the dot.
0. A beep sounds to signal a successful scan.
0.
If no beep sounds, repeat the procedure. If a beep repeatedly fails to sound, refer to Troubleshooting, Section 6.
5. Press .
RUN STANDARD 1?
READY: SCAN BAR CODE
02/24/03 2:09PM
2
UNPACKING, GETTING
ACQUAINTED AND SET UP
2.18 Revised 6/03
DISPLAY
6 MIN = total test time
After 1 minute:
Upon completion of test:
WHA T Y OU DO
6. Open the cartridge compartment door.
7. Hold the optical test cartridge so that the bar
code faces right.
0. Insert the cartridge into the compartment
until a subtle snap is heard/felt.
0. HINT: The cartridge is designed to fit only
one way into the instrument.
8. Close door.
09. Record the displayed results in the blanks
provided on the last page of this manual (appendix).
10. Remove cartridge.
00. a) Open cartridge compartment door.*
00. b) Locate the button on the right side of the
cartridge compartment. Push and hold it down with your right hand.
00. c) With your left hand, gently push the plastic
tab on the cartridge to the right; this action releases (unlocks) cartridge.
00. d) Pull cartridge out of compartment.
*If the door is opened (within 15 minutes after assay
completion), the test result is displayed for only 30 seconds.
*If the door is not opened, the test result will remain
displayed for 15 minutes.
*At 15 minutes, an audible tone (error buzz) sounds and the
display changes to “READY: REMOVE TEST.”
1.0001 T 0.00012 SDnnn S1 0.00387 DRIFT
1.0001 T 0.00012 SD S1 5 MIN 10:03AM
PROCESSING STANDARD
S1 6 MIN 10:02AM
STANDARD 1
LOAD, CLOSE DOOR
UNPACKING, GETTING ACQUAINTED AND SET UP
Revised 6/03 3.1
3
MENU
MENU
The MENU consists of up to nine (9) items listed below. Note that item number two and item number nine are optional (may or may not be active—depends upon choices made during “Instrument Setup”).
RECALL PREVIOUS TESTS?
SET SEQUENCE NUMBER? (optional)
RECALL CONTROL RESULTS?
VIEW CALIBRATION STATUS?
SET DATE/TIME?
INSTRUMENT SETUP?
SET CREATININE UNITS?
INSTRUMENT TEST?
RUN CONTROL? (optional)
To access the MENU (and display the first item):
Press
To display each additional item:
Repeatedly press
To select the item displayed:
Press
To exit the MENU:
Press
Information and instructions for each MENU item are provided in this section.
9
8
7
6
5
4
3
2
1
3.2 Revised 6/03
3
MENU
MENU
RECALL PREVIOUS TESTS?
Up to 16 test results can be recalled (and printed if a printer is in use).
Recalled Test Result Format
Assay Result Reagent Lot No.
Date of Assay Sequence No. Time Assay Began
Date of Assay (optional) Time A. (optional)
6.0 % HbA1c LOT:0325 02/24/03 #011 11:24PM
1
DISPLAY
OR
OR
WHA T Y OU DO
1. Press .
(Test in progress is not aborted when MENU is pressed.)
TEST IN PROGRESS
#001 6 MIN 10:15AM
READY: SCAN BAR CODE
02/24/03 2:09PM
WAIT: WARMING UP
02/24/03 2:09PM
Date of Assay
Test
Type
Sequence No.
(optional)
Time Assay Began
Revised 6/03 3.3
3
MENU
DISPLAY
(example)
OR
(example)
OR
(example)
OR
(example)
(after all results are displayed)
WHA T Y OU DO
2. Press .
0.
The test result for the sample most recently assayed is displayed first.
3. Press to recall up to 16 test results.
NOTES:
To print the displayed test result, press (if a printer is in use).
–If is pressed for more than 3 seconds, all stored
test results are printed.
To return display to “RECALL PREVIOUS TESTS?,” press once anytime during recall of test results.
To exit the MENU, press twice.
Albumin (A), creatinine (C), and the albumin/ creatinine ratio (A/C) results for the same specimen are shown sequentially on separate displays, and are labeled with the same sequence number (optional) and date/time.
This display is retained until:
a calibration card bar code is read
or is pressed
is pressed, the display returns to the first test result
previously recalled (this is the result for the most recent sample assayed).
–If is pressed once, the display returns to “RECALL
PREVIOUS TESTS?”
–If is pressed twice, you exit the MENU.
NO MORE RESULTS STORED
IN INSTRUMENT
A/C=76 LOT:0325 02/24/03 #010 10:24PM
C=72.1mg/dL LOT:0325 09/13/02 #010 10:24PM
A=55.0mg/L LOT:0325 09/13/02 #010 10:24PM
6.0 % HbA1c LOT:0325 02/24/03 #011 11:24PM
RECALL PREVIOUS TESTS?
02/17/90 #001 11:47AM
3.4 Revised 6/03
3
MENU
MENU
SET SEQUENCE NUMBER? (optional)
2
DISPLAY
OR
OR
WHA T Y OU DO
1. Press .
2. Press .
NOTE: If a sequence number is selected while a test is in progress, the sequence number for the test in progress is changed to the newly selected sequence number.
3. Press .
4. Press or to select the digit above the
cursor.
0. Press to move cursor (right) to next
0. digit.
0.
Repeat procedure to select second and third digits.
0. Press .
5. To exit the MENU, press .
0. To display the next MENU item, press .
SET SEQUENCE NUMBER?
SET SEQUENCE NUMBER
#009
SET SEQUENCE NUMBER?
TEST IN PROGRESS
#001 6 MIN 10:15AM
READY: SCAN BAR CODE
02/24/03 2:09PM
WAIT: WARMING UP
02/24/03 2:09PM
Revised 6/03 3.5
3
MENU
RECALL CONTROL RESULTS?
Format for Recalled Results (examples)
DCA 2000 CONTROLS, ONLY
Control Result Reagent Lot Number
mmmmmmmmmmmmmmmmmmmmmmmmm
**Cl—is replaced with “C2” for the DCA 2000 Abnormal or High Control and
“C1out” or “C2out” for an out-of-range control
ALL CONTROLS EXCEPT DCA 2000 CONTROLS*
Control Result Reagent Lot Number
*unless a DCA 2000 Control was run without scanning the DCA 2000
Control bar code card provided with the control kitmmmmmmmmn
9.8% HbA1c LOT:0325 02/24/03 c1 LOTC:0223
9.8% HbA1c LOT:0325 02/24/03 C1 LOTC:0223
3
• Up to 16 control results can be stored and recalled (and also printed if a printer is in use).
• Upper case “C” in display indicates a result for a DCA 2000
®
Control.
•Lower case “c” in display indicates a result for a control other than a DCA 2000 Control*.
•A control lot number is displayed only for DCA 2000 Controls.
Date of Assay C1— indicates
a result for the DCA 2000 Normal or Low Control**
DCA 2000 Control Lot Number
Date of Assay c1—indicates control
level “1” of 9 possible levels
Test
Type
Test
Type
—continued on next page
3.6 Revised 6/03
3
MENU
MENU
Recall Control Results
DISPLAY
OR
OR
WHA T Y OU DO
1. Press .
2. Repeatedly press , until “RECALL
CONTROL RESULTS?” is displayed.
(Test in progress is not aborted.)
3. Press .
0.
The control result for the control most recently assayed is displayed first.
RECALL CONTROL RESULTS?
TEST IN PROGRESS
#001 6 MIN 10:15AM
READY: SCAN BAR CODE
02/24/03 2:09PM
WAIT: WARMING UP
02/24/03 2:09PM
Revised 6/03 3.7
3
MENU
DISPLAY
(example)
OR
(example)
OR
(example)
Reminder: Upper case “C” indicates
result for a DCA 2000
®
Control
(after all control
results are displayed)
WHA T Y OU DO
4. Press to recall up to 16 control
0. results.
NOTES:
To print the control result, press (if a printer is in use).
– If is pressed for more than 3 seconds, all stored
control results are printed.
To return display to “RECALL CONTROL RESULTS,” press once anytime during recall of results.
To exit the MENU, press twice.
Albumin (A) and creatinine (C) for the same control are shown sequentially on separate displays, and are labeled with the same date.
This display is retained until:
a bar code is scanned
or is pressed
– If is pressed, the display returns to the first control
result previously recalled (this is the result for the most recent control assayed).
– If is pressed once, the display returns to “RECALL
CONTROL RESULTS?”
– If is pressed twice, you exit the MENU.
NO MORE CONTROLS STORED
IN INSTRUMENT
C=54.1mg/dL LOT:0105 02/24/03 C2 LOTC:0345
A=24.6mg/L LOT:0105 02/24/03 C2 LOTC:0345
9.8% HbA1c LOT:0325 02/24/03 C1 LOTC:0223
3.8 Revised 6/03
3
MENU
MENU
DISPLAY
OR
OR
WHA T Y OU DO
1. Press .
0. (A test in progress is not aborted.)
2. Repeatedly press , until
0. “VIEW CALIBRATION STATUS?” is
displayed.
3. Press .
0.
The calibration status for the most recent calibration is displayed first.
VIEW CALIBRATION STATUS?
TEST IN PROGRESS
#001 6 MIN 10:15AM
READY: SCAN BAR CODE
02/24/03 2:09PM
WAIT: WARMING UP
02/24/03 2:09PM
VIEW CALIBRATION STATUS?
The instrument stores up to two calibrations for each DCA 2000 Reagent Test. The two calibrations must be for two different lot numbers.
4
Revised 6/03 3.9
3
MENU
DISPLAY
(example)
OR
(example)
(after all stored
calibrations are displayed)
WHA T Y OU DO
4. Press to recall status of each
0. calibration stored.
NOTES:
To return display to “VIEW CALIBRATION STATUS?,” press once anytime during calibration status recall.
To exit the MENU, press twice.
This display is retained until:
a bar code is scanned
or is pressed
– If is pressed, the display returns to the first
calibration status previously recalled (this is the status for the most recent calibration).
– If is pressed once, the display returns to “VIEW
CALIBRATION STATUS?”
– If is pressed twice, you exit the MENU.
NO MORE CALIBRATIONS STORED IN INSTRUMENT
LOT:0222 Malb/C 02/24/03 11:30PM
LOT:0183 HbA1c 02/24/03 11:23PM
3.10 Revised 6/03
3
MENU
MENU
SET DATE/TIME?
Use the following keys:
and To cycle through each available setting (e.g.,
digits for day/month/year and time, or AM/PM)
To move cursor under setting you desire to change (e.g., “2” in display below is marked by cursor and indicates “26” is ready for change)
To set the displayed date and time in the instrument
SET DATE/TIME...
02/24/03 6:20AM
5
DISPLAY
OR
WHA T Y OU DO
1. Press .
2. Repeatedly press until
0. “SET DATE/TIME?” is displayed.
3. Press .
SET DATE/TIME?
02/24/03 6:19AM
READY: SCAN BAR CODE
02/24/03 2:09PM
WAIT:WARMING UP
02/24/03 2:09PM
Revised 6/03 3.11
3
MENU
DISPLAY
AM or PM if AM/PM
format is selected
AM or PM if AM/PM
format is selected
WHA T Y OU DO
The display on your instrument is displaying the date and time in your chosen format.
4. Press or until the correct two digits are
displayed.
0. Press .
5. Press or until the correct two digits are
displayed.
0. Press .
6. Press or until the correct two digits are
displayed.
0. Press . The cursor will move back to the first
digit of the date
7. Press to advance the cursor to the time
setting.
8. Press or until the correct two digits are
displayed.
0. Press .
9. Press or until the correct two digits are
displayed.
0. Press .
10.
If 24 HR format is selected, press .
10.
If AM/PM format is selected, press to advance cursor to the AM/PM setting.
0. Press or until the correct choice is
displayed.
0. Press .
0. Press .
SET DATE/TIME
02/24/03 6:20AM
SET DATE/TIME
02/24/03 6:20AM
SET DATE/TIME
02/24/03 6:20AM
SET DATE/TIME
02/24/03 6:20AM
SET DATE/TIME
02/24/03 6:20AM
3.12 Revised 6/03
3
MENU
(This page left blank on purpose.
Intended for future use.)
Revised 6/03 3.13
3
MENU
DISPLAY
OR
OR
If computer port is Off.
OR
If computer port is On.
WHA T Y OU DO
1. Press .
2. Repeatedly press , until
0. “INSTRUMENT SETUP?” is displayed.
(Test in progress is not aborted.)
3. Press .
4. Refer to the chart (pages 2.8 and 2.9) to
determine the definition of each number and abbreviated option appearing on your instrument’s display.
0. Then continue with step 5 on page 3.14.
1 1 4 0 2 9801010mmmmm T D L C P BUVWXYZmmmmm
1 1 4 0 0 MMMMMMMMMMM T D L C P MMMMMMMMMMM
INSTRUMENT SETUP?
TEST IN PROGRESS
#001 6 MIN 10:15AM
READY: SCAN BAR CODE
02/24/03 2:09PM
WAIT: WARMING UP
02/24/03 2:09PM
INSTRUMENT SETUP?
6
3.14 Revised 6/03
3
MENU
MENU
DISPLAY
If computer port is Off.
OR
If computer port is On.
If computer port is Off.
OR
If computer port is On.
If computer port is Off.
OR
If computer port is On.
WHA T Y OU DO
5. Press (places a cursor under the
0. setting for the first option, TIME FORMAT).
6. Press and to cycle through choices for
TIME FORMAT.
0. When the desired choice is displayed, press
0. (moves cursor under setting for
0. next option).
7.
Repeat step 6 to change settings for each option as desired.
1 1 4 0 2 9801010mmmmm T D L C P BUVWXYZmmmmm
1 1 4 0 0 mmmmmmmmmmmm T D L C P Mmmmmmmmmmmm
1 1 4 0 2 9801010mmmmm T D L C P BUVWXYZmmmmm
1 1 4 0 0 mmmmmmmmmmmm TIME FORMAT AM/PM
1 1 4 0 2 9801010mmmmm T D L C P BUVWXYZmmmmm
1 1 4 0 0 9801010mmmmm T D L C Pmmmmmmmmmmmmm
INSTRUMENT SETUP DISPLAY OPTIONS/SETTINGS continued
Revised 6/03 3.15
3
MENU
SET CREATININE UNITS
(Microalbumin/Creatinine Assay ONL Y)
The concentration units reported for creatinine are selectable between “mg/dL” and “mmol/L”. The factory setting is “mg/dL”.
7
DISPLA Y WHA T YOU DO
1. Press .
0.
2. Repeatedly press , until
“SET CREATININE UNITS?” is displayed.
3. Press . This places a cursor below the “m”
in “mg/dL”, and the question mark disappears.
4. Press or to display “mmol/L” or
“mg/dL”.
5. Press to accept desired units.
6. Press to exit the MENU.
SET CREATININE UNITS?
mmol/L
SET CREATININE UNITS
mmol/L
SET CREATININE UNITS
mg/dL
SET CREATININE UNITS?
mg/dL
READY: SCAN BAR CODE
02/24/03 2:09PM
3.16 Revised 6/03
3
MENU
MENU
INSTRUMENT TEST?
This menu item is to be used only under the guidance of our Bayer HealthCare Customer Service Department Representative.
Using this menu item allows you and our Customer Service Representative to determine existing problems with the keyboard, display, bar code reader, printer port, memory, and computer port functions.
8
Revised 6/03 3.17
3
MENU
DISPLA Y WHA T Y OU DO
1. Press .
2. Repeatedly press until
0. “RUN CONTROL?” is displayed.
3. Press .
4. Press or until the desired
control number (choices 1–9) is displayed.
0. Press .
5. Follow instructions in Section 4, page 4.3, or
Section 5, page 5.4, under “Preparing Patient Samples and Controls.”
SCAN BAR CODE
02/24/03 c1 11:10AM
SET CONTROL NUMBER
c1
RUN CONTROL?
02/17/90 #001 11:47AM
READY: SCAN BAR CODE
02/24/03 2:09PM
RUN CONTROL? (optional)
Available only if selected via “INSTRUMENT SETUP” procedure
Marks the control result with a lower case “c”
• lower case “c” indicates controls other than DCA 2000 Controls (unless a DCA 2000 Control was analyzed without scanning the control bar code card provided with the control kit)
9
Revised 6/03 4.1
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
OPERATING INSTRUCTIONS — HEMOGLOBIN A
1c
NOTE: The instructions in this section are for use
after “INSTRUMENT SETUP” has been performed (a one time requirement upon receipt of a new or factory-serviced instrument).
STEP 1: Turning the P ower ON
DISPLAY
After about 8 seconds:
(displayed for 8 seconds)
(displayed for 3 seconds)
(a beep is heard)
WHA T Y OU DO
1. Set the power switch to ON.
0. IMPORTANT:The program card must be
inserted or removed only when the power switch is set to OFF. If the card is inserted when the power is ON, the card can be permanently damaged.
2. OPTIONAL: While the instrument is warming up (usually 1 – 2 minutes but can take up to 8 minutes), you may access certain MENU items. Refer to Section 3 for instructions.
0. Reminder: Current time is denoted by
“blinking colon.” Time assay began is denoted by stationary colon.
READY: SCAN BAR CODE
02/24/03 2:09PM
WAIT: WARMING UP
02/24/03 2:09PM
INITIALIZING
KEEP DOOR CLOSED
COPYRIGHT 1991–2003
BY BAYER CORPORATION
SOFTWARE VERSION
E3.11/01.04
9.8% HbA1c LOT:0183 02/17/90 #001 11:47AM
4.2 Revised 6/03
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
OPERATING INSTRUCTIONS — HEMOGLOBIN A
1c
STEP 2: Calibration
When to Calibrate:
Calibrate the System for each new lot number of reagent cartridges.
Materials Required:
Calibration Card (provided in DCA 2000
®
Hemoglobin A1cReagent Kit)
DISPLAY
OR
Any result or menu display
as long as testing is not in progress.
(After 5 seconds, the display
returns to the display in effect
prior to calibration.)
WHA T Y OU DO
1. Locate the dot (on the instrument) next to the bar code track.
2. Locate the bar code on the calibration card.
3. Hold the card so that the bar code faces right.
4. Insert the card into the bar code track (above dot). Hold card gently against the right side of track.
0.
5. Quickly (within 1 sec.) and smoothly, slide the card down past the dot.
0. A beep sounds to signal a successful scan.
0.
If no beep sounds, repeat procedure. If a beep repeatedly fails to sound, refer to Troubleshooting, Section 6.
CALIBRATION DATA SAVED LOT:0132 HbA1c
READY: SCAN BAR CODE
02/24/03 2:09PM
NOTES:
The instrument stores up to two calibrations for the DCA 2000 Hemoglobin A
1c
Reagent
Test. The two calibrations must be for two
different lot numbers.
The calibration stored in the instrument first is deleted when a calibration card for a third lot number is scanned.
Revised 6/03 4.3
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
STEP 3: Preparing Patient Samples and Controls
Materials Required:
DCA 2000 Hemoglobin A1cReagent Kit
Patient Sample, DCA 2000®Hemoglobin A
1c
Control Kit or other control
Lint-free tissue
Clock or timer
OPENING THE FOIL PACKAGE (Containing Reagent Cartridge)
IMPORTANT: Do not use scissors to cut open
foil package. Scissors can damage the reagent cartridge, the flexible plastic pull-tab on the cartridge or the sack containing desiccant.
1. Remove one foil package (containing a reagent cartridge) from storage.
2. Refer to “Recommended Procedures
for
Handling Reagent Cartridges” in the
DCA 2000
Hemoglobin A1cReagent Kit package insert for instructions on how and when to open foil package.
INSPECTING THE CONTENTS OF THE FOIL PACKAGE
The foil package contains:
1. reagent cartridge with flexible pull-tab and bar code label
2. small sack (filled with desiccant)
0.
When handling the reagent cartridge, do not touch or otherwise contaminate the optical window or erroneous test results may occur.
0.
Discard the reagent cartridge if:
the cartridge is damaged
the flexible pull-tab is loose or missing
the small sack (desiccant) is missing or open
loose desiccant particles are found inside the foil package
(refer to Service Information, Section 9)
OPTICAL WINDOW
4.4 Revised 6/03
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
OPERATING INSTRUCTIONS — HEMOGLOBIN A
1c
CAPILLARY HOLDER
Unused capillary holders may be saved and used with any lot of Hemoglobin A
1c
reagent cartridges.
1. Open plastic wrap by tearing wrap at serrated
1. edge.
2. Inspect the capillary holder for the presence of:
0. absorbent pad
0. glass capillary
0. latching mechanism
0.
If the capillary holder is missing any of the above parts, discard the capillary holder (refer to Service Information, Section 9).
3
2
1
1
2
3
Revised 6/03 4.5
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
FILLING THE CAPILLARY WITH WHOLE BLOOD
(Instructions for filling capillary with control sample are found in the
DCA 2000 Hemoglobin A
1c
Control package
insert.)
IMPORTANT PLEASE READ CAREFULLY:
Once the capillary is filled with sample, analysis must begin within 5 minutes. There is no need to rush. Five minutes allows enough time for proper completion of procedures.
Within 5 minutes after filling glass capillary (step 1, next page), complete steps 2-5 (page 4.7) and one of the following, whichever applies.
EITHER
For a Patient Sample or any control other than DCA 2000 Control — Steps 1 through 9 under “WHAT
YOU DO” on pages 4.8-4.9
OR
For DCA 2000 Control — Steps 1 through 14 under “WHAT
YOU DO” on pages 4.12-4.13
IMPORTANT: If sample analysis does not begin within 5 minutes after filling glass capillary, discard capillary. If capillary is in the reagent cartridge, discard both capillary and reagent cartridge.
WARNING! POTENTIAL BIOHAZARD
All products or objects which come into contact with human blood, even after cleaning, should be handled as if capable of transmitting viral diseases.
The user should follow the recommendations for prevention of blood-borne transmissible diseases in healthcare settings, as recommended for potentially infectious human blood specimens in National Committee for Clinical Laboratory Standards, Protection of Laboratory Workers from Instrument Biohazards and Infectious Disease Transmitted by Blood, Body Fluids and Tissues: Approved Guideline. NCCLS Document M29-A [ISBN 1-56238­339-6] NCCLS, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898, USA, 1997. This document has complete information on the topic of user protection and can be used as background material for instruction.
VAROITUS! MAHDOLLINEN TARTUNTAVAARA
Terveydenhoitohenkilökunnan, joka käyttää tätä laitetta useiden ihmisten tutkimiseen, tulee ottaa huomioon, että kaikki tuotteet, jotka joutuvat kosketukseen ihmisveren kanssa, ovat myös puhdistuksen jälkeen mahdollisia viirustartuntalähteitä.
Tartuntojen välttämiseksi suosittelemme laboratorion omien turvallisuusohjeiden ehdotonta noudattamista.
4.6 Revised 6/03
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
OPERATING INSTRUCTIONS — HEMOGLOBIN A
1c
FILLING THE CAPILLARY WITH WHOLE BLOOD—continued
1. Complete step A) OR step B) depending upon the blood sample source. Continue with step 2 on the next page.
0. IMPORTANT: Do not allow blood to contact
the plastic part of the capillary holder. Any blood touching the plastic will be transferred into the reaction buffer, along with the blood in the glass capillary. This can cause an invalid HbA
1c
result or possibly an error message.
If blood contacts the plastic part of the capillary holder, discard capillary holder.
0. A) If filling glass capillary with blood from
finger prick:
0. A) 1. Hold the capillary holder at an angle.
0. A) 2. Touch only the tip of the capillary to a
small drop of blood on the finger until the capillary is filled.
A) 0.
0. B) If filling glass capillary with blood
obtained by venipuncture:
0. B) 1. Mix sample well (by inversion or use of
an aliquot mixer) to prevent separation of red blood cells and plasma.
0. B) 2. Remove stopper from blood collection
tube in such a way that a small sample of blood remains on stopper.
0. B) 3. Hold the capillary holder at an angle.
0. B) 4. Touch only the tip of the capillary to
blood sample on stopper.
0. B) 0.
Do not attempt to fill capillary by touching glass capillary to blood in a blood collection tube. Attempting to fill capillary in this manner most often results in blood touching the capillary holder. If blood touches the capillary holder, discard capillary holder.
B) 0.
Revised 6/03 4.7
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
2. Using a lint-free tissue, carefully wipe the
outside of the glass capillary.
0. Do not allow the tissue to touch the open end
of the glass capillary. Contact with the open
end of the capillary could result in loss of sample (by wicking into tissue). If sample loss is obvious, discard capillary holder; then repeat procedure using a new capillary holder.
3. Inspect the glass capillary for the presence of bubble(s). If bubble(s) are obvious, discard capillary holder; then repeat procedure using a new capillary holder.
4. Position the capillary holder in the correct orientation for insertion into the reagent cartridge.
0.
5. Carefully insert the capillary holder into the
reagent cartridge until the holder gently snaps into place.
0. IMPORTANT: Avoid harsh insertion of
capillary holder. It is important not to dislodge sample from glass capillary or erroneous results may occur.
4.8 Revised 6/03
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
OPERATING INSTRUCTIONS — HEMOGLOBIN A
1c
DISPLA Y WHA T Y OU DO
1. Locate the dot (on the instrument) next to the
bar code track.
2. Locate the bar code on the reagent cartridge.
3. Hold the reagent cartridge so that the bar code
faces right.
4. Insert the reagent cartridge (above dot) into bar
code track.
0.
5. Quickly (within 1 sec.) and smoothly, slide
the reagent cartridge down past the dot.
0. A beep sounds to signal a successful scan.
0.
If no beep sounds, repeat procedure. If a beep repeatedly fails to sound, refer to Troubleshooting, Section 5.
READY: SCAN BAR CODE
02/24/03 2:09PM
STEP 4:Analyzing the Patient Sample
Revised 6/03 4.9
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
DISPLA Y WHA T YOU DO
6. Open the cartridge compartment door.
7. Hold the reagent cartridge so that the bar code
faces right.
0. Insert the reagent cartridge into the cartridge
compartment until a subtle snap is heard/felt.
High temperature part Huom. Kuumenee käytettäessä
0.
0.
HINT: The cartridge is designed to fit only one
way into the instrument. Do not force cartridge into instrument.
8. Using a smooth, slow, continuous motion, pull flexible pull-tab completely out of reagent cartridge.
9. Close door. Dispose of flexible pull-tab.
0.
Five (5) seconds after the door is closed, a beep sounds and the assay begins.
0.
NOTE: If you accidentally close the door before you
pull the flexible plastic tab, you have 5 seconds to re-open the door; the display returns to “LOAD CARTRIDGE.” You may now pull the tab or correct existing problem(s).
LOAD CARTRIDGE
PULL TAB, CLOSE DOOR
4.10 Revised 6/03
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
OPERATING INSTRUCTIONS — HEMOGLOBIN A
1c
DISPLAY
(colon does not blink)
After test is completed:
(colon does not blink)
WHA T Y OU DO
10. Record the displayed result before removing
the reagent cartridge.
11. Remove the reagent cartridge.
00. a) Open cartridge compartment door.*
00. b) Locate the button on the right side of the
cartridge compartment. Push and hold it down with your right hand.
00. c) With your left hand, gently push the tab on
the cartridge to the right; this action releases (unlocks) cartridge.
00. d) Pull the reagent cartridge out of the
compartment.
00. e) Discard cartridge in proper container,
according to your standard laboratory procedures.
0.
*If the door is opened (within 15 minutes after assay
completion), the test result is displayed for only 30 seconds.
00. If the door is not opened, the test result will remain
displayed for 15 minutes.
00. At 15 minutes, an audible tone (error buzz) sounds
and the display changes to “READY: REMOVE TEST.”
00. HINT: If the displayed test result was not recorded,
use the MENU to recall up to 16 test results (refer to SECTION 3).
%HbA1c mmm #001 10:21
TEST IN PROGRESS
#001 6 MIN 10:15AM
Revised 6/03 4.11
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
RESULTS:
The displayed test result requires no further calculation. Hemoglobin A
1c
concentrations in the
following range are reported:
2.5% to 14.0%mm
The test is linear throughout this range.
Result preceded by a less than sign (<):
A less than sign in the display indicates a concentration below the lower limit of the test (under range). Report the result as being less than
2.5% Hemoglobin A
1c
. This method does not provide for re-assay using a larger sample aliquot. Results less than 2.5% Hemoglobin A
1c
are rare and may indicate that the sample contains substantial amounts of fetal hemoglobin (does not react in the immunoassay); or that the patient may be suffering from hemolytic anemia or polycythemia (conditions which often result in a significant decrease in the life span of red blood cells).
Result preceded by a greater than sign (>):
A greater than sign in the display indicates a concentration above the upper limit of the test (over range). Report the result as being more than
14.0% Hemoglobin A
1c
. This method does not provide for re-assay using a diluted sample. To obtain a more quantitative test value at levels greater than 14%, use another test method.
All laboratory tests are subject to random error. If the test result is questionable, or if clinical signs and symptoms appear inconsistent with test results, re-assay the sample or confirm the result using another method.
4.12 Revised 6/03
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
OPERATING INSTRUCTIONS — HEMOGLOBIN A
1c
Analyzing DCA 2000
®
Hemoglobin A1cControls, ONLY
NOTE: Follow instructions on page 4.8, if
analyzing a control other than a DCA 2000 Hemoglobin A
1c
Control.
Controls are analyzed in the same manner as the patient sample.
A specially designed control bar code (that enters the control lot number, etc.) is provided with DCA 2000 HbA
1c
Controls.
DISPLAY
Therefore, when analyzing DCA 2000 Hemoglobin A
1c
Controls, use instructions that contain steps for scanning the control bar code card found on pages 4.12–4.14.
If using controls other than DCA 2000 Hemoglobin A
1c
Controls, refer to Section 3, page 3.17, for information on labeling and storing the control result.
WHA T Y OU DO
1. Locate the dot (on the instrument) next to the
bar code track.
2. Locate the bar code on the control card.
0.
NOTE: The control card is double-sided; one side for normal, the other for abnormal. Make sure you are using the correct side of the control card for the particular DCA 2000 Control level in use.
C1 = = Normal C2 = = Abnormal
3. Hold the control card so that the bar code
faces right.
4. Insert the control card into the bar code track (above dot). Hold card gently against the right side of track.
5. Quickly (within 1 sec.) and smoothly, slide the card down past the dot.
0. A beep sounds to signal a successful scan.
0.
If no beep sounds, repeat procedure. If a beep repeatedly fails to sound, refer to Troubleshooting, Section 6.
6. Press .
RUN CONTROL C1?
LOTC:1112 HbA1c
READY: SCAN BAR CODE
02/24/03 2:09PM
Revised 6/03 4.13
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
DISPLA Y WHA T Y OU DO
07. Locate the bar code on the reagent cartridge.
08. Hold the reagent cartridge so that the bar code
faces right.
09. Insert the reagent cartridge (above dot) into
bar code track.
10. Quickly (within 1 sec.) and smoothly, slide
the reagent cartridge down past the dot.
00. A beep sounds to signal a successful scan.
00.
If no beep sounds, repeat procedure. If a beep repeatedly fails to sound, refer to Troubleshooting, Section 6.
11. Open the cartridge compartment door.
12. Hold the reagent cartridge so that the bar code
faces right.
00. Insert the reagent cartridge into the cartridge
compartment until a subtle snap is heard/felt.
00. HINT: The reagent cartridge is designed to fit
only one way into the instrument. Do not force cartridge into instrument.
13. Using a smooth, slow continuous motion,
pull flexible pull-tab completely out of reagent cartridge.
14.
Close door. Dispose of flexible pull-tab.
00. NOTE: If you accidentally close the door before you
pull the tab, you have 5 seconds to re-open the door; the display returns to “LOAD CARTRIDGE.” You may now pull the tab or correct existing problem(s).
LOAD CONTROL
PULL TAB, CLOSE DOOR
SCAN BAR CODE: HbA1c 02/24/03 C1 11:10AM
4.14 Revised 6/03
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
OPERATING INSTRUCTIONS — HEMOGLOBIN A
1c
DISPLAY
(colon does not blink)
Then:mmmmmmmmnmmmmmm.
After test is completed:
For a result within the acceptable control range printed in the control package insert
OR
For a result outside the acceptable control range printed in the control package insert
WHA T Y OU DO
15. NOTE: If “CONTROL OUT OF RANGE” is
displayed, press to display value of out-of­range control.
15. Record the displayed result before removing
the reagent cartridge.
16. Remove the reagent cartridge.
00. a) Open cartridge compartment door.*
00. b) Locate the button on the right side of the
cartridge compartment. Push and hold it down with your right hand.
00. c) With your left hand, gently push the tab on
the cartridge to the right; this action releases (unlocks) cartridge.
00. d) Pull the reagent cartridge out of the
compartment.
00. e) Discard cartridge in proper container,
according to your standard laboratory procedures.
00.
*If the door is opened (within 15 minutes after assay
completion), the test result is displayed for only 30 seconds.
If the door is not opened, the test result will remain displayed for 15 minutes.
At 15 minutes, an audible tone (error buzz) sounds and the display changes to “READY: REMOVE TEST.”
15.
CONTROL OUT OF RANGE
PRESS [ESC] TO PROCEED
%HbA1cmmmmm
0 C1 10:21AM
PROCESSING CONTROL
C1-HbA1c 6 MIN 11:11AM
Revised 6/03 4.15
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
CANCELLING A TEST
You may cancel a test anytime. Important: If a test in progress is cancelled, the
test must be discarded.
DISPLAY
OR
(colon does not blink)
(an error buzz sounds)
WHA T Y OU DO
1. Press .
2. Press within 15 seconds.
0.
NOTE: After pressing , “PLEASE WAIT” is displayed until the cartridge returns to original loading position.
0. If is not pressed within 15 seconds:
0.
the display returns to the original display (i.e., either display shown in Step 1).
0.
the test in progress continues without interruption
CANCEL TEST?
PRESS [<--] TO CONFIRM
TEST IN PROGRESS
#001 6 MIN 10:15AM
LOAD CARTRIDGE
PULL TAB, CLOSE DOOR
4.16 Revised 6/03
4
OPERATING INSTRUCTIONS
HEMOGLOBIN A
1C
OPERATING INSTRUCTIONS — HEMOGLOBIN A
1c
DISPLAY
displayed if the test is cancelled after the bar code is scanned but before the cartridge compartment door is opened
a beep is heard
OR
–displayed if the test is cancelled during
sample analysis (cartridge is in instrument)
–an error buzz is heard
WHA T Y OU DO
3. The test is cancelled.
0.
Scan bar code
ORmm
0.
Open cartridge compartment door and remove cartridge.
CANCELLED: DISCARD TEST
02/24/03 10:18AM
READY: SCAN BAR CODE
02/24/03 2:09PM
Revised 6/03 5.1
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
OPERATING INSTRUCTIONS —
MICROALBUMIN/CREATININE
NOTE: The instructions in this section are for use
after “INSTRUMENT SETUP” has been performed (a one time requirement upon receipt of a new or factory-serviced instrument).
STEP 1: Turning the P ower ON
DISPLAY
After about 8 seconds:
(displayed for 8 seconds)
(displayed for 3 seconds)
(a beep is heard)
WHA T Y OU DO
1. Turn the power switch to ON.
0. IMPORTANT:The program card must be
inserted or removed only when the power switch is set to OFF. If the card is inserted when the power is ON, the card can be permanently damaged.
2. OPTIONAL: While the instrument is warming up (usually 1 – 2 minutes but can take up to 8 minutes), you may access certain MENU items. Refer to Section 3 for instructions.
0. Reminder: Current time is denoted by
“blinking colon.” Time assay began is denoted by stationary colon.
READY: SCAN BAR CODE
02/24/03 2:09PM
WAIT: WARMING UP
02/24/03 2:09PM
INITIALIZING
KEEP DOOR CLOSED
COPYRIGHT 1991-2003
BY BAYER CORPORATION
SOFTWARE VERSION
E3.11/01.04
9.8% HbA1c LOT:0183 02/17/90 #001 11:47AM
5.2 Revised 6/03
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
OPERATING INSTRUCTIONS — MICRO ALB UMIN / CREATININE
STEP 2: Calibration
When To Calibrate:
Calibrate the System for each new lot number of reagent cartridges.
Materials Required:
Calibration Card (provided in DCA 2000
®
Microalbumin/Creatinine Reagent Kit)
NOTES:
The calibration card has two bar codes, one on each side of the card. Side 1 and Side 2 are identified on the card. Either side may be scanned first. After the first scan, the display will indicate the next side to be scanned. Side 1 is Malb/C; Side 2 is Malb-2.
It is necessary to scan both bar codes to calibrate the system for each new lot of reagent cartridges.
The instrument stores up to two calibrations for the DCA 2000 Microalbumin/Creatinine Reagent Test. The two calibrations must be for
two different lot numbers.
The calibration stored in the instrument first is deleted when a calibration card for a third lot number is scanned.
DISPLAY
OR
Any result or menu display
as long as testing is not in progress.
WHA T Y OU DO
1. Locate the dot (on the instrument) next to the bar code track.
2. Locate the bar code on the calibration card.
3. Hold the card so that the bar code faces right.
4. Insert the card into the bar code track (above dot). Hold card gently against the right side of track.
0.
READY: SCAN BAR CODE
02/24/03 2:09PM
Revised 6/03 5.3
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
DISPLAY
After 5 seconds, the instrument then sounds a
2 second beep and displays:
WHA T Y OU DO
5. Quickly (within 1 sec.) and smoothly, slide the card down past the dot.
0. A beep sounds to signal a successful scan.
0.
If no beep sounds, repeat procedure. If a beep repeatedly fails to sound, refer to Troubleshooting, Section 6.
6. Repeat Steps 1–5 above using the other side of the card.
0. If the second bar code is not scanned within 5
minutes:
0.
the display returns to the original display and the calibration process must start over. (See previous page.)0.
CALIBRATION DATA SAVED LOT:0432 Malb-2
SCAN CALIBRATION CARD LOT:0432 Malb-2
CALIBRATION DATA SAVED LOT:0432 Malb/C
5.4 Revised 6/03
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
OPERATING INSTRUCTIONS — MICRO ALB UMIN / CREATININE
STEP 3: Preparing Patient Samples and Controls
Materials Required:
DCA 2000 Microalbumin/Creatinine Reagent Kit
Patient Sample, DCA 2000®Microalbumin/ Creatinine Control Kit or other control
Lint-free tissue
OPENING THE FOIL PACKAGE (Containing Reagent Cartridge)
IMPORTANT: Do not use scissors to cut open
foil package. Scissors can damage the reagent cartridge, the flexible pull-tab on the cartridge or the sack containing desiccant.
1. Remove one foil package (containing a reagent cartridge) from storage.
2. Refer to “Recommended Procedures
for
Handling Reagent Cartridges” in the
DCA 2000
Microalbumin/Creatinine Reagent Kit package insert for instructions on how and when to open foil package.
INSPECTING THE CONTENTS OF THE FOIL PACKAGE
The foil package contains:
1. reagent cartridge with flexible pull-tab and bar code label
2. small sack (filled with desiccant)
0.
When handling the reagent cartridge, do not touch or otherwise contaminate the optical window or erroneous test results may occur.
0.
Discard the reagent cartridge if:
the cartridge is damaged
the flexible pull-tab is loose or missing
the small sack (desiccant) is missing or open
loose desiccant particles are found inside the foil package
(refer to Service Information, Section 9)
OPTICAL WINDOW
Revised 6/03 5.5
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
CAPILLARY HOLDER AND PLUNGER
Unused capillary holders may be saved and used with any lot of microalbumin/creatinine reagent cartridges.
1. Remove a capillary holder and plunger from plastic bag.
2. Inspect the capillary holder for the presence of:
0. absorbent pad
0. glass capillary
0. latching mechanism
0. starch plug
0.
3. If the capillary holder is missing any of the above parts, discard the capillary holder; also discard if the starch plug is at the bottom of the capillary tube (refer to Service Information, Section 9).
4
3
2
1
plunger
1
2
2
3
4
5.6 Revised 6/03
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
OPERATING INSTRUCTIONS — MICRO ALB UMIN / CREATININE
FILLING THE CAPILLARY WITH URINE
1. Complete step A) OR step B) depending upon
the volume of urine specimen available. Continue with Step 2 on the next page.
1. IMPORTANT: Do not allow urine to contact
either the plastic part of the capillary holder or the adsorbent material in the capillary holder. Any urine touching the plastic will be transferred into the reaction buffer, along with the urine in the glass capillary. This can cause an invalid microalbumin/creatinine result or possibly an error message. If urine contacts
the plastic part of the capillary holder, discard capillary holder.
A) If filling capillary tube with urine from a
container with a large sample volume:
A) 1. Immerse the tip of the capillary tube in
the urine specimen to a level just above the plug in the capillary.
A) 2. Allow enough time for the urine
specimen to flow into the capillary tube and come in contact with the starch plug, approximately 5 seconds. Wetting the starch plug seals the capillary tube and keeps the urine within the tube.
A) 3. Remove the capillary tube from the
urine specimen. If the urine flows back down the tube, re-immerse the capillary tube in the urine specimen again, allowing enough time to ensure that the starch plug becomes saturated.
(Instructions for filling capillary with control sample are found in the DCA 2000 Microalbumin/Creatinine Control package insert.)
B) If filling capillary tube with urine from a
container with a small sample volume:
B) 1. Immerse the tip of the capillary tube
in the urine specimen. Tilt sample container and capillary holder to a more horizontal position to increase the rate of flow into the capillary. Take care not to spill the urine specimen.
B) 2. Allow enough time for the urine
specimen to flow into the capillary tube and come in contact with the starch plug. Wetting the starch plug seals the capillary tube and keeps the urine within the tube.
B) 3. Remove the capillary tube from the
urine specimen. If the urine flows back down the tube, re-immerse the capillary tube in the urine specimen again, allowing enough time to ensure that the starch plug becomes saturated.
Revised 6/03 5.7
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
2. Using a lint-free tissue, carefully wipe the outside of the glass capillary tube. Do not allow the tissue to touch the open end of the glass capillary. Contact with the open end of
the capillary tube could result in loss of sample (by wicking into tissue). If sample loss is obvious, discard the capillary holder; then repeat procedure using a new capillary holder.
3. Inspect the glass capillary tube for the presence of bubble(s). If bubbles are obvious, discard the capillary holder; then repeat procedure using a new capillary holder.
4. Position the capillary holder in the correct orientation for insertion into the reagent cartridge.
5. Carefully insert the capillary holder into the reagent cartridge until the holder gently snaps into place.
IMPORTANT: Avoid harsh insertion of the capillary holder. It is important not to dislodge sample from the capillary tube, or erroneous results may occur.
5.8 Revised 6/03
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
OPERATING INSTRUCTIONS — MICRO ALB UMIN / CREATININE
DISPLA Y WHA T Y OU DO
1. Locate the dot (on the instrument) next to the
bar code track.
2. Locate the bar code on the reagent cartridge.
3. Hold the reagent cartridge so that the bar code
faces right.
4. Insert the reagent cartridge (above dot) into bar
code track.
0.
5. Quickly (within 1 sec.) and smoothly, slide
the reagent cartridge down past the dot.
0. A beep sounds to signal a successful scan.
0.
If no beep sounds, repeat procedure. If a beep repeatedly fails to sound, refer to Troubleshooting, Section 6.
READY: SCAN BAR CODE
02/24/03 2:09PM
STEP 4:Analyzing the Patient Sample
Revised 6/03 5.9
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
DISPLA Y WHA T YOU DO
6. Open the reagent cartridge compartment door.
7. Hold the reagent cartridge so that the bar code
faces right.
0. Insert the reagent cartridge into the cartridge
compartment until a subtle snap is heard/felt.
0. HINT: The cartridge is designed to fit only
one way into the instrument. Do not force cartridge into instrument.
8. Insert plunger into hole in top of capillary
holder.
0. Depress plunger into capillary holder fully.
Plunger will lock into the capillary holder.
0. Press .
High temperature part Huom. Kuumenee käytettäessä
00.
9. Using a smooth, slow, continuous motion,
pull flexible pull-tab completely out of reagent cartridge.
10. Close door.
Dispose of flexible pull-tab.
0.
Five (5) seconds after the door is closed, a beep sounds and the assay begins.
0.
NOTE: If you accidentally close the door before
you pull the flexible tab, you have 5 seconds to re-open the door; the display returns to “LOAD CARTRIDGE.” You may now pull the tab or correct existing problem(s).
—continued on next page
PULL TAB, CLOSE DOOR
LOAD CARTRIDGE, INSERT
PLUNGER, PRESS [<--]
5.10 Revised 6/03
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
OPERATING INSTRUCTIONS — MICRO ALB UMIN / CREATININE
DISPLAY
(colon does not blink)
After test is completed:
(colon does not blink)
WHA T Y OU DO
11. Record the displayed result before removing
cartridge.
11. NOTE: If creatinine units are mg/dL, then the
Albumin/Creatinine (A/C) ratio is reported as mg/g. If the creatinine units are mmol/L, then the Albumin/Creatinine (A/C) ratio is reported as mg/mmol.
12. Remove cartridge.
00. a) Open cartridge compartment door.*
00. b) Locate the button on the right side of the
cartridge compartment. Push and hold it down with your right hand.
00. c) With your left hand, gently push the tab on
the cartridge to the right; this action releases (unlocks) cartridge.
00. d) Pull cartridge out of compartment.
00. e) Discard cartridge in proper container,
according to your standard laboratory procedures.
00. *If the door is opened (within 15 minutes
after assay completion), the test result is displayed for only 30 seconds.
00. If the door is not opened, the test result will
remain displayed for 15 minutes.
00. At 15 minutes, an audible tone (error buzz)
sounds and the display changes to “READY: REMOVE TEST.”
00. HINT: If the displayed test result was not
recorded, use the MENU to recall up to 16 test results (refer to SECTION 3).
A=102mg/L C=141mg/dL A/C=72 #001 10:17AM
TEST IN PROGRESS
#001 6 MIN 10:15AM
Revised 6/03 5.11
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
RESULTS: Albumin: The displayed test result requires no
further calculation. Albumin concentrations in the following range are reported: 5 to 300 mg/L. The test is linear throughout this range.
Creatinine: The displayed test result requires no further calculation. Creatinine concentrations in the following range are reported: 15 to 500 mg/dL or 1.3 to 44.2 mmol/L. The test is linear throughout this range.
Albumin/Creatinine Ratio: The displayed test result requires no further calculation. Albumin/ Creatinine ratios can be reported in the following range: 1 to 2,000 mg/g or 0.11 to 226 mg/mmol.
Albumin or creatinine result preceded by a less than sign (<): A less than sign in the display
indicates a concentration below the lower limit of the test (under range). This method does not provide for re-assay using a larger sample.
Albumin or creatinine result preceded by a greater than sign (>): A greater than sign in the
display indicates a concentration above the upper limit of the test (over range). This method does not provide for re-assay using a diluted sample. To obtain a more quantitative test value, use another test method.
Ratio result preceded by a less than (<) sign or greater than (>) sign or (---): If the albumin
and/or creatinine result is under or over range, the ratio will also be reported as under or over range. In certain cases, no ratio will be reported (---).
Example 1: If the albumin result is >300 mg/L and the creatinine result is 100 mg/dL (8.84 mmol/L), then the ratio will be reported as >300 mg/g (>26.5 mg/mmol).
Example 2: If the albumin result is 75 mg/L and the creatinine result is <15 mg/dL (<1.33 mmol/L), then the ratio will be reported as >500 mg/g (>56.4 mg/mmol).
Example 3: If the albumin result is >300 mg/L and the creatinine result is >500 mg/dL (>44.2 mmol/L), then no ratio will be reported (---).
Example 4: If the albumin result is <5 mg/L and the creatinine result is <15 mg/dL (<1.33 mmol/L), then no ratio will be reported (---).
All laboratory tests are subject to random error. If the test result is questionable, or if clinical signs and symptoms appear inconsistent with the test results, re-assay the sample or confirm the result using another method.
5.12 Revised 6/03
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
OPERATING INSTRUCTIONS — MICRO ALB UMIN / CREATININE
Analyzing DCA 2000
®
Controls, ONLY
NOTE: Follow instructions on page 5.8, if
analyzing a recommended control other than a DCA 2000 Microalbumin/Creatinine Control.
Controls are analyzed in the same manner as the patient sample.
A specially designed control bar code (that enters the control lot number, etc.) is provided with DCA 2000 Microalbumin/Creatinine Controls.
DISPLAY
Therefore, when analyzing DCA 2000 Microalbumin/Creatinine Controls, use instructions that contain steps for scanning the control bar code card found on pages 5.12–5.14.
If using controls other than DCA 2000 Microalbumin/Creatinine Controls, refer to Section 3, page 3.17, for information on labeling and storing the control result.
WHA T Y OU DO
1. Locate the dot (on the instrument) next to the
bar code track.
2. Locate the bar code on the control card.
0.
NOTE: The control card is double-sided; one side for Low, the other for High. Make sure you are using the correct side of the control card for the particular DCA 2000 Control level in use.
C1 = = Low C2 = = High
3. Hold the control card so that the bar code
faces right.
4. Insert the control card into the bar code track
(above dot). Hold card gently against the right side of track.
5. Quickly (within 1 sec.) and smoothly, slide
the card down past the dot.
0. A beep sounds to signal a successful scan.
0.
If no beep sounds, repeat procedure. If a beep repeatedly fails to sound, refer to Troubleshooting, Section 6.
6. Press .
RUN CONTROL C1?
LOTC:1112 Malb/C
READY: SCAN BAR CODE
02/24/03 2:09PM
Microalb umin / Creatinine
Revised 6/03 5.13
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
DISPLA Y WHA T Y OU DO
07. Locate the bar code on the reagent cartridge.
08. Hold the cartridge so that the bar code faces
right.
09. Insert the cartridge (above dot) into bar code
track.
10. Quickly (within 1 sec.) and smoothly, slide the cartridge down past the dot.
00. A beep sounds to signal a successful scan.
00.
If no beep sounds, repeat procedure. If a beep repeatedly fails to sound, refer to Troubleshooting, Section 6.
11. Open the cartridge compartment door.
12. Hold the reagent cartridge so that the bar code faces right.
00. Insert the cartridge into the cartridge
compartment until a subtle snap is heard/felt.
00. HINT: The cartridge is designed to fit only
one way into the instrument. Do not force cartridge into instrument.
13. Insert plunger into hole in top of capillary holder.
00. Depress plunger into capillary holder fully.
Plunger will lock into the capillary holder.
00. Press .
14. Using smooth, slow continuous motion, pull flexible pull-tab completely out of reagent cartridge.
15.
Close door. Dispose of flexible pull-tab.
00. NOTE: If you accidentally close the door before you
pull the tab, you have 5 seconds to re-open the door; the display returns to “LOAD CARTRIDGE.” You may now pull the tab or correct existing problem(s).
PULL TAB, CLOSE DOOR
LOAD CARTRIDGE, INSERT
PLUNGER, PRESS [<--]
SCAN BAR CODE Malb/C 02/24/03 C1 11:10AM
— continued on next page
5.14 Revised 6/03
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
OPERATING INSTRUCTIONS — MICRO ALB UMIN / CREATININE
DISPLAY
(colon does not blink)
After test is completed:
For a result within the acceptable control range printed in the control package insert
NOTE: The albumin/creatinine ratio is not
calculated for controls.
OR
For a result outside the acceptable control range printed in the control package insert
WHA T Y OU DO
16. NOTE: If “CONTROL OUT OF RANGE” is displayed, press to display value of out-of­range control.
15. Record the displayed result before removing
cartridge.
17. Remove cartridge.
00. a) Open cartridge compartment door.*
00. b) Locate the button on the right side of the
cartridge compartment. Push and hold it down with your right hand.
00. c) With your left hand, gently push the
flexible tab on the cartridge to the right; this action releases (unlocks) cartridge.
00. d) Pull cartridge out of compartment.
00. e) Discard cartridge in proper container,
according to your standard laboratory procedures.
00.
*If the door is opened (within 15 minutes after assay
completion), the test result is displayed for only 30 seconds.
If the door is not opened, the test result will remain displayed for 15 minutes.
At 15 minutes, an audible tone (error buzz) sounds and the display changes to “READY: REMOVE TEST.”
CONTROL OUT OF RANGE
PRESS [ESC] TO PROCEED
A=102mg/L C=141 mg/dL
0 C1 11:11AM
PROCESSING CONTROL
C1-Malb/C 7 MIN 11:11AM
Revised 6/03 5.15
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
CANCELLING A TEST
You may cancel a test anytime. Important: If a test in progress is cancelled, the
test must be discarded.
DISPLAY
OR
OR
(colon does not blink)
(an error buzz sounds)
WHA T Y OU DO
1. Press .
2. Press within 15 seconds.
0.
NOTE: After pressing , “PLEASE WAIT” is displayed until the cartridge returns to original loading position.
0. If is not pressed within 15 seconds:
0.
the display returns to original display (i.e., either display shown in Step 1).
0.
the test in progress continues without interruption
CANCEL TEST?
PRESS [<--] TO CONFIRM
TEST IN PROGRESS
#001 6 MIN 10:15AM
PULL TAB, CLOSE DOOR
LOAD CARTRIDGE, INSERT
PLUNGER, PRESS [<--]
— continued on next page
5.16 Revised 6/03
5
OPERATING INSTRUCTIONS
MICROALBUMIN/CREATININE
OPERATING INSTRUCTIONS — MICRO ALB UMIN / CREATININE
DISPLAY
displayed if the test is cancelled after the bar code is scanned but before the cartridge compartment door is opened
a beep is heard
OR
–displayed if the test is cancelled during
sample analysis (cartridge is in instrument)
–an error buzz is heard
WHA T Y OU DO
3. The test is cancelled.
0.
Scan bar code
ORmm
0.
Open cartridge compartment door and remove cartridge.
CANCELLED: DISCARD TEST
02/24/03 10:18AM
READY: SCAN BAR CODE
02/24/03 2:09PM
Revised 6/03 6.1
6
ERROR AND WARNING MESSAGES,
ERROR CODES & TROUBLESHOOTING
ERROR AND WARNING MESSAGES, ERR OR CODES AND TROUBLESHOOTING
Use the chart below to quickly find the correct page for the desired error and warning message, error code, or troubleshooting information.
ERROR AND WARNING MESSAGES
Error and warning messages are provided in alphabetical order.
page
CANCELLED: DISCARD TEST/(date)(time) ............................. 6.2
DOOR OPEN ERROR .................................................................. 6.2
OUT OF DATE CONTROL.......................................................... 6.2
OUT OF DATE REAGENT.......................................................... 6.2
PLEASE WAIT ............................................................................. 6.2
SCAN CALIBRATION CARD..................................................... 6.2
SEE OPERATING MANUAL (COMPUTER
PORT ERROR-51)........................................................................ 6.2
SEE OPERATING MANUAL ERROR 90—RAM MEMORY .. 6.3
SEE OPERATING MANUAL/TEST ERROR XXX................... 6.3
SEE OPERATING MANUAL (ERROR Xx)............................... 6.3
SEE OPERATING MANUAL
(TEST UNKNOWN)
................... 6.3
WRONG CARD............................................................................ 6.3
ERROR CODES
E-1 ................................................................................................. 6.4
E-2.................................................................................................. 6.4
E-3.................................................................................................. 6.4
E-4.................................................................................................. 6.5
E-5................................................................................................. 6.5
E-6.................................................................................................. 6.5
TROUBLESHOOTING
..................................................................................................... 6.10
Hemoglobin A
1c
ERROR 101 ........................ 6.6
ERROR 102 ........................ 6.6
ERROR 103 ........................ 6.6
ERROR 104 ........................ 6.6
ERROR 105 ........................ 6.6
ERROR 106 ........................ 6.6
ERROR 107 ........................ 6.6
ERROR 108 ........................ 6.7
ERROR 109 ........................ 6.7
ERROR 110 ........................ 6.7
ERROR 111 ........................ 6.7
ERROR 112 ........................ 6.7
ERROR 113 ........................ 6.7
ERROR 114 ........................ 6.7
ERROR 115 ........................ 6.7
ERROR 116 ........................ 6.7
Microalbumin / Creatinine
ERROR 301 ........................ 6.8
ERROR 302 ........................ 6.8
ERROR 303 ........................ 6.8
ERROR 304 ........................ 6.8
ERROR 305 ........................ 6.8
ERROR 306 ........................ 6.8
ERROR 307 ........................ 6.8
ERROR 308 ........................ 6.8
ERROR 309 ........................ 6.9
ERROR 310 ........................ 6.9
ERROR 311 ........................ 6.9
ERROR 313 ........................ 6.9
ERROR 315 ........................ 6.9
ERROR 316 ........................ 6.9
6.2 Revised 6/03
6
ERROR AND WARNING MESSAGES,
ERROR CODES & TROUBLESHOOTING
ERROR AND WARNING MESSAGES, ERR OR CODES AND TROUBLESHOO TING
ERROR AND WARNING MESSAGES
ERROR OR WARNING CAUSE SOLUTION
The cartridge compartment door was Open the cartridge compartment door. opened while testing was in progress. Remove and discard the reagent The door was then closed and “PLEASE cartridge. WAIT” was displayed prior to
Repeat the test.
“CANCELLED: DISCARD TEST.”
If the error message “SEE OPERATING
OR
MANUAL/(any system error)” is
was pressed in response to
displayed again, refer to Section 9, for
“SEE OPERATING MANUAL/
instructions on “How to Report the
(any system error) ” Problem.”
The cartridge compartment door was Close door. opened while testing was in progress.
Wait for next display “PLEASE WAIT.”
The bar code for an out-of-date control Discard out-of-date control(s). (past expiration date) has been read.
(displayed for 15 seconds)
Control lot number and test
name above are examples, only.
The bar code for an out-of-date (past Discard out-of-date reagent cartridge(s). expiration date) reagent cartridge has been read.
(displayed for 15 seconds)
Reagent lot number and test
name above are examples, only.
The cartridge compartment door was Wait until cartridge is in correct position for opened while testing was in progress. removal (until “CANCELLED: DISCARD Then the door was closed. TEST” is displayed).
OR
A System Error was detected and
was pressed.
The calibration card for the reagent Scan the correct calibration card. cartridge in use has not been scanned.
Check name and lot number of cartridge
in use.
Lot number and test name
above are examples, only
The computer port is ON and/or: 1. Check to make sure cable is securely
the instrument buffer is full of
0. connected to both instrument and
untransmitted results
0. computer.
the computer will not accept results
2. If the problem persists, refer to Section 9,
0. for instructions on “How to Report the
0. Problem.”
0. HINT: To facilitate testing (without
0. transmitting results to a computer) turn
0. off the computer port (refer to Section 3,
0. MENU).
SEE OPERATING MANUAL
COMPUTER PORT ERROR-E51
SCAN CALIBRATION CARD
LOT:1154 HbA1c
PLEASE WAIT
02/17/90 #001 11:47AM
OUT OF DATE REAGENT
LOT:1154 HbA1c
OUT OF DATE CONTROL
LOTC:1154 HbA1c
DOOR OPEN ERROR
CLOSE DOOR
CANCELLED: DISCARD TEST 02/24/03 10:18AM
Revised 6/03 6.3
6
ERROR AND WARNING MESSAGES,
ERROR CODES & TROUBLESHOOTING
ERROR OR WARNING CAUSE SOLUTION
When the power was turned ON (I), a Press key. failure was detected in the non-volatile
“INSTRUMENT SETUP/PRESS
memory.
[<--]
TO CONTINUE” is displayed.
All factory settings (defaults) are now active.
Turn to Section 2 and follow the instruc-
tions beginning with “Viewing Factory
Settings.” You must now, once again,
either accept or change factory settings.
An error relating to a test measurement 1.Discard the test. parameter has been detected (e.g.,
2.See pages 6.6, 6.7, 6.8, and 6.9.
abnormal hemoglobin levels, high C.V.’s,
3.Repeat the test using a new reagent
out of range hemoglobin transmission, etc.).
0.cartridge and sample.
4. If the problem persists, record the test error identification number; then refer to
Section 8, for instructions on “How to
Report the Problem.”
* The instrument has detected an instrument
If testing is in progress, press .
error (motor or optical failure, temperature,
Follow instructions under error message
etc.).
“CANCELLED: DISCARD TEST.”
*OR:
If testing is not in progress, turn the
ERROR 2x—OPTICAL
power OFF, then ON.
ERROR 3x—TEMPERATURE
If the problem persists, record the error
ERROR 4x—BAR CODE
message. Then refer to Section 9, for
ERROR 9x—INTERNAL
instructions on “How to Report the Problem.”
The instrument does not recognize the Contact Bayer HealthCare Customer Service reagent test in use. Department. See Section 9, Service
Information.
In response to “SCAN CALIBRATION Scan correct calibration card. CARD,” the wrong calibration card is
Check name and lot number of cartridge
scanned.
in use.
(displayed for 5 seconds—then “SCAN Both sides of the calibration card must
Repeat calibration process making certain
CALIBRATION CARD” appears) be scanned to enter the calibration for
to scan both sides of the calibration card.
Microalbumin / Creatinine.
WRONG CARD
SEE OPERATING MANUAL
LOT:2234 TEST: UNKNOWN
SEE OPERATING MANUAL
ERROR 11-MOTOR
SEE OPERATING MANUAL
TEST ERROR 115
SEE OPERATING MANUAL
ERROR 90-RAM MEMORY
ERROR—101 TO 116 for HbA
1c ERROR—301 to 316 for Microalbumin/Creatining
6.4 Revised 6/03
6
ERROR AND WARNING MESSAGES,
ERROR CODES & TROUBLESHOOTING
ERROR AND WARNING MESSAGES, ERR OR CODES AND TROUBLESHOO TING
ERROR CODES
ERROR CODE CAUSE SOLUTION
An internal system error has been Contact Bayer HealthCare Customer detected. Service Department. See Section 9, Service
Information.
When the power switch was set to ON (I), Set the power switch to OFF (O). the instrument detected an error in the
Contact Bayer HealthCare Customer
system’s non-volatile memory.
Service Department. See Section 9, Service Information.
1. The wrong program card is 1. Set the power switch to OFF (O).
0. plugged into the instrument.
0. Remove the wrong program card.
0. Plug the correct program card completely
0. into the instrument.
0. Set the power switch to ON (I).
2. No program card is plugged into 2. Set the power switch to OFF (O).
0. the instrument.
0. Plug the correct program card completely
0. into the instrument.
0. Set the power switch to ON (I).
3. The correct program card is correctly 3. Set the power switch to OFF (O).
0. plugged into the instrument but the
0. Use a pencil eraser to gently clean the
0. contacts on the program card need
0. contacts on the program card.
0. cleaning or are defective.
0. Use a lint-free tissue to wipe particles
0. from the contacts.
0. Correctly insert the program card.
0. Set the power switch to ON (I).
0.
If the problem persists, contact
0.
Bayer HealthCare Customer Service
0.
Department. See Section 9, Service
0.•Information.
E3
E2
E1
Revised 6/03 6.5
6
ERROR AND WARNING MESSAGES,
ERROR CODES & TROUBLESHOOTING
ERROR CODE CAUSE SOLUTION
1. The wrong program card is plugged 1. Set the power switch to OFF (O). Plug
0. into the instrument. 0. the correct program card completely into
0. the instrument. Set the power switch to
0. ON (I).
2. The correct program card is not 2. Refer to step 1.
0. completely plugged into the
0. instrument.
3. The contacts on the correct program 3. Set the power switch to OFF (O).
0. card need to be cleaned or are 0. Remove the program card.
0. defective.
0. Use a pencil eraser to gently clean the
0. contacts on the program card.
0. Use a lint-free tissue to wipe particles
0. from contact.
0. Correctly insert the program card.
0. Set the power switch to ON (I).
0. If the problem persists, contact Bayer
0. HealthCare Customer Service
0. Department. See Section 9, Service
0. Information.
When the power switch was set to ON (I), 1. Set the power switch to OFF (O). the instrument detected use of a defective
2. Remove the program card.
program card.
3. Plug a replacement program card into the
0.
instrument. (Refer to Section 9, SERVICE
0. INFORMATION, for instructions on
0. where to order replacement program card.)
4. Set the power switch to ON (I).
0. If the problem persists, contact Bayer
0. HealthCare Customer Service
0. Department. See Section 9, Service
0. Information.
When the power switch was set to ON (I), Set the power switch to OFF (O). the instrument detected an error in
Contact Bayer HealthCare Customer
system timing.
Service Department. See Section 9, Service Information.
E6
E5
E4
6.6 Revised 6/03
6
ERROR AND WARNING MESSAGES,
ERROR CODES & TROUBLESHOOTING
ERROR AND WARNING MESSAGES, ERR OR CODES AND TROUBLESHOO TING
ERROR CODE CAUSE SOLUTION
101
Buffer Reading Out of Limits — High
102
Buffer Reading Out of Limits — Low
103
High Variation in Readings for Buffer
104
106
Reading for Hemoglobin Value — Low
105
107
Reading for Hemoglobin Value — High
No cartridge present. Optical alignment problem.
Condensation on cartridge — cartridge not allowed to warm up
Cartridge not located properly in instrument.
Buffer tab not pulled. Buffer tab removed before cartridge inserted.
Cartridge defect in optical window. Cartridge optical window blocked or dirty.
Condensation on cartridge — cartridge not allowed to warm up
Rare. Particulate contamination.
No or low blood in reaction — capillary underfill or air bubble
— blood dried in capillary
— no capillary holder inserted — improper reconstitution of controls
or use of non-DCA 2000 controls Hemoglobin < 7 g/dL — anemic patient, abnormally low
hemoglobin Buffer tab not pulled or buffer not
released from tray.
Excess blood on capillary holder.
Blood not lysing — cartridge not allowed to warm up
— irregularity in patient red blood cells (rare)
Hemoglobin > 24 g/dL — patient has abnormally high hemoglobin
— very high triglycerides
Repeat test with cartridge in place. If problem persists, contact Bayer
HealthCare Customer Service Department. See Section 9, Service Information.
Allow at least 10 minutes after cartridge is removed from refrigerator before starting test.
Ensure that cartridge is inserted completely into instrument.
Pull buffer tab after inserting cartridge into holder.
If problem persists, contact Bayer HealthCare Customer Service Department. See Section 9, Service Information.
Allow at least 10 minutes after cartridge is removed from refrigerator before starting test.
If problem persists, contact Bayer HealthCare Customer Service Department. See Section 9, Service Information.
Ensure that capillary is completely filled with no air bubbles.
Wait no more than 5 minutes after filling capillary before starting test.
Repeat test with capillary holder in place. Review procedure for reconstitution of
DCA 2000 controls.
Perform test by another method.
Pull buffer tab before closing door.
Take care in sampling the blood to avoid excess blood on capillary holder.
Ensure that capillary is wiped before inserting into cartridge.
Allow at least 10 minutes after cartridge is removed from refrigerator before starting test.
Freeze/thaw specimen before use, or perform test by another method.
Perform test by another method. See reagent kit insert under Limitations.
Revised 6/03 6.7
6
ERROR AND WARNING MESSAGES,
ERROR CODES & TROUBLESHOOTING
ERROR CODE CAUSE SOLUTION
108
High Variation in Readings for Hemoglobin
109
110
111
112
Readings for Glycated Hemoglobin Out of Limits
113
114
116
Irregular Reaction Kinetics for Glycated Hemoglobin
115
Final Hemoglobin Reading Greater Than Reading at Earlier Checkpoint
Rare. Particulate contamination.
Cartridge exposed to excessive temperature and/or humidity.
Cartridge exposed to excessive temperature and/or humidity.
Particulate contamination.
Blood left too long in capillary.
If problem persists, contact Bayer HealthCare Customer Service Department. See Section 9, Service Information.
Ensure proper storage of reagent kits. Open foil package just prior to use
of cartridge.
Ensure proper storage of reagent kits. Open foil package just prior to use
of cartridge. If problem persists, contact Bayer
HealthCare Customer Service Department. See Section 9, Service Information.
Wait no more than 5 minutes after filling capillary before starting test.
6.8 Revised 6/03
6
ERROR AND WARNING MESSAGES,
ERROR CODES & TROUBLESHOOTING
ERROR AND WARNING MESSAGES, ERR OR CODES AND TROUBLESHOO TING
ERROR CODE CAUSE SOLUTION
301
Buffer Mean Out of Limits — Low Absorbance
302
Buffer Mean Out of Limits — High Absorbance
303
High Variation in Readings for Buffer
304
Sample Blank Mean Out of Limits — Low
305
Sample Blank Mean Out of Limits — High
306
High Variation in Readings for Sample Blank
307
308
Readings for Albumin Out of Limits
No cartridge present. Optical alignment problem.
Condensation on cartridge — cartridge not allowed to warm up
Cartridge not located properly in instrument.
Buffer tab not pulled. Buffer tab removed before cartridge inserted.
Cartridge defect in optical window. Cartridge optical window blocked or dirty.
Rare. Particulate contamination.
Particulate contamination.
Urine sample is turbid or highly pigmented.
Rare. Particulate contamination.
Rare. Cartridge exposed to excessive
temperature and/or humidity.
Repeat test with cartridge in place. If problem persists, contact Bayer
HealthCare Customer Service Department. See Section 9, Service Information.
Allow at least 15 minutes after cartridge is removed from refrigerator before starting test.
Ensure that cartridge is inserted completely into instrument.
Pull buffer tab after inserting cartridge into holder.
If problem persists, contact Bayer HealthCare Customer Service Department. See Section 9, Service Information.
If problem persists, contact Bayer HealthCare Customer Service Department. See Section 9, Service Information.
Repeat test. If problem persists, contact Bayer HealthCare
Customer Service Department. See Section 9, Service Information.
Centrifuge sample before assaying if sample is very turbid or cloudy.
If sample contains visible amounts of blood, or is highly pigmented, obtain a fresh sample.
If problem persists, contact Bayer HealthCare Customer Service Department. See Section 9, Service Information.
Repeat test. If problem persists, contact Bayer HealthCare
Customer Service Department. See Section 9, Service Information.
Ensure proper storage of reagent kits. Open foil package just prior to use
of cartridge.
Revised 6/03 6.9
6
ERROR AND WARNING MESSAGES,
ERROR CODES & TROUBLESHOOTING
ERROR CODE CAUSE SOLUTION
309
High Variation in Albumin Readings
310
311
Readings for Creatinine Out of Limits
313
Irregular Reaction Kinetics for Creatinine
315
Normalization Factor has not been set
316
Normalization Factor outside allowable range
Rare. Particulate contamination.
Rare. Cartridge exposed to excessive
temperature and/or humidity.
Cartridge exposed to excessive temperature and/or humidity.
Particulate contamination.
Instrument has not been adjusted to run the Microalbumin/Creatinine assay.
Normalization Cartridge is not located properly in instrument.
Normalization Cartridge is dirty or defective.
Repeat test. If problem persists, contact Bayer
HealthCare Customer Service Department. See Section 9, Service Information.
Ensure proper storage of reagent kits. Open foil package just prior to use
of cartridge.
Ensure proper storage of reagent kits. Open foil package just prior to use
of cartridge. If problem persists, contact Bayer
HealthCare Customer Service Department. See Section 9, Service Information.
Contact Bayer HealthCare Customer Service Department. See Section 9, Service Information.
Ensure that cartridge is inserted completely into instrument.
If problem persists, contact Bayer HealthCare Customer Service Department. See Section 9, Service Information.
6.10 Revised 6/03
6
ERROR AND WARNING MESSAGES,
ERROR CODES & TROUBLESHOOTING
ERROR AND WARNING MESSAGES, ERR OR CODES AND TROUBLESHOO TING
TROUBLESHOOTING
PROBLEM SOLUTION
Beep repeatedly fails to sound after scanning a bar code. 1. Insert the cartridge or expanded bottom of card (whichever
0. applies) into the bar code track below the dot.
2. Quickly (within 1 second) and smoothly, slide the card or
0. cartridge up past the dot.
0. If a beep continually fails to sound, refer to Section 8 for
0. instructions on “How to Report the Problem.”
Capillary will not fill completely with sample. Discard capillary holder.
Flexible tab tears off before it is pulled completely out of Discard the reagent cartridge. reagent cartridge.
Refer to Section 9 for instructions on “How to Report the
Problem.”
Revised 6/03 7.1
7
INSTRUMENT CARE
AND ROUTINE MAINTENANCE
INSTR UMENT CARE AND R OUTINE MAINTEN ANCE
INSTRUMENT CARE
The DCA 2000® Analyzer contains sensitive electronics and optics.
IMPORTANT! Do not use sprays. Sprays will permanently damage the optical system.
Handle the instrument with extreme care. Severe mechanical shocks can damage and/or dislodge internal parts and connections.
Do not block the ventilation panels found on the back and right sides of the instrument. Allow at least two inches of air space between the wall (or other surface) and the back and right sides of the instrument.
Do not operate the instrument beyond the
recommended maximum ambient operating temperature or relative humidity ranges listed below.
Maximum Ambient Operating T emperature:
(15–32°C)—Hemoglobin A
1c
(18–30°C)—Microalbumin/Creatinine
Relative Humidity: 10% – 90% RH
Do not place the instrument where it would be exposed to direct sunlight, extreme temperature variations, particulate matter, excessive humidity or air currents.
Do not smoke in the room where you have placed the instrument. Smoke may cause a film to form on the internal optical surfaces affecting the optical transmission qualities of the instrument.
ROUTINE MAINTENANCE CHART
NOTE: Maintenance requirements for each laboratory must be assessed individually. Use the above chart as a guide. It is good laboratory practice to properly maintain your instrument.
As Required (by spillage, troubleshooting or
Weekly Quarterly contamination)
Exterior Cartridge Compartment Exterior Bar Code Windo w Change Air Filter Change Air Filter
Run Optical Test Cartridge Cartridge Compartment
Bar Code Window Run Optical Test Cartridge
32
15
18
30
7.2 Revised 6/03
7
INSTRUMENT CARE
AND ROUTINE MAINTENANCE
INSTRUMENT CARE AND ROUTINE MAINTENANCE
EXTERIOR OF THE INSTRUMENT AND BAR CODE WINDOW
WARNING: Turn off the power and unplug the power cord before cleaning
exterior of instrument.
VAROITUS: Kytke virta pois ja irrota virtajohto ennen kuin alat puhdistaa
laitteen ulkopintoja.
Do not allow water or other cleaning fluid to drip inside instrument (bar code window area, program card area and key pad are especially vulnerable).
1. Clean the exterior of the instrument, including
the display panel and bar code window with a lint-free cloth dampened with water. A cloth dampened with ethanol may also be used.
0. If you desire to disinfect the exterior of the
instrument, expose the surface to either of the following for 10 minutes.
1
Remove liquid blood
(as much as possible) prior to disinfection.
0. a. 0.5% sodium hypochlorite (see household
bleach, below)
0. b. 2% glutaraldehyde
0.a.
1
This is as referenced in the National Committee for Clinical Laboratory Standards, Protection of Laboratory Workers from Instrument Biohazards and Infectious Disease Transmitted by Blood, Body Fluids and Tissues:Approved Guideline. NCCLS Document M29-A [ISBN 1-56238-339-6].
2. After the exterior is clean and dry, attach and
plug-in power cord.
Most household bleach is (approx.) a 5% solution of sodium hypochlorite (read the label).
Dilute household bleach containing (approx.) 5% sodium hypochlorite as follows to obtain (approx.) a 0.5% solution of sodium hypochlorite.
10 mL of household bleach + 90 mL of water or alternatively:
1 part household bleach + 9 parts water
Revised 6/03 7.3
7
INSTRUMENT CARE
AND ROUTINE MAINTENANCE
CHANGING THE AIR FIL TER
1. Remove the filter holder from the right side of the instrument by rotating the holder counter­clockwise (until it stops) and pulling it off.
2. Dispose of the old air filter.
3. Place a new air filter into the filter holder.
4. Place the filter holder back on the instrument and rotate the holder clockwise (until it stops).
Filter
Holder
Filter
Holder
Air
Filter
7.4 Revised 6/03
7
INSTRUMENT CARE
AND ROUTINE MAINTENANCE
INSTRUMENT CARE AND ROUTINE MAINTENANCE
CAR TRIDGE COMPARTMENT
WARNING: Turn off the power and unplug the power cord before cleaning
the cartridge compartment.
VAROITUS: Kytke virta pois ja irrota virtajohto, ennen kuin puhdistat
kasettipesän.
Do not allow liquid to drip into instrument. If liquid drips into instrument, optics can be destroyed.
1. Open the cartridge compartment door as far as possible.
2. Using a lint-free cloth dampened with water or ethanol, wipe the inside surface of the compartment door and surfaces on both sides of the cartridge holder.
0.
Dry surfaces using a clean, dry, lint-free cloth.
3. Locate the cartridge return spring inside the cartridge holder. Note top and bottom holes.
0.
4. Insert the tip of a straightened paper clip
(or other like device) into the top hole on the spring.
0.
Gently, pull metal end toward center of cartridge compartment to release one side of the spring from the cartridge holder.
0.
Repeat procedure to release other side of spring from cartridge holder.
5. Pull cartridge return spring completely out of
instrument.
6. Clean cartridge return spring using any of the
following:
0.
warm solution of mild detergent and water (you may immerse cartridge return spring)
0.
lint-free cloth dampened in water or ethanol
0. IMPORTANT: Make sure the leaf springs are
not bent or damaged while cleaning. Damaged leaf springs will not function properly.
0.
7. Dry cartridge return spring with a clean,
lint-free cloth. Set spring aside.
TOP HOLE
BOTTOM HOLE
CARTRIDGE RETURN SPRING
LEAF SPRINGS
Revised 6/03 7.5
7
INSTRUMENT CARE
AND ROUTINE MAINTENANCE
08. Using a clean, dry, sponge swab (provided in
“Cleaning Kit,” Part No. 95001901) remove spilled liquid from the cartridge holder.
00. IMPORTANT: Do not use a cotton swab.
Cotton fibers may be left on surface and interfere with instrument optical system.
0.0 With compartment door partially closed,
rotate the cartridge holder as shown to locate and remove any additional liquid.
00.
00.
00.
00.
NOTE: Cartridge holder cannot be rotated if compartment door is completely open.
09. Dampen (do not soak) a sponge swab with
water or ethanol. Clean cartridge holder (rotating cartridge holder as necessary).
00. IMPORTANT: Do not allow liquid to drip
off
sponge swab into instrument. If liquid
drips
into instrument, optics can be
destroyed.
10. Look inside the cartridge compartment and locate the vertical grooves (tracks). Next, locate the front and back slots (found near the
top lip of the compartment).
11. Locate the leaf spring on one side of the cartridge return spring.
12. With the leaf spring orientated toward the back of the instrument, lower the leaf spring into the instrument as follows:
00. a. Hold onto both sides of the cartridge
return spring (near the holes).
00. b. Pinch the sides together (push toward each
other) and lower the spring into instrument by sliding the sides of the spring between the vertical grooves (tracks) in the compartment. Release (let go of) spring.
00. b. NOTE: When “step b,” above, has been
properly completed, the return spring should be free to slide up and down within the cartridge compartment. If not, repeat “step b” above.
00. b. IMPORTANT: Never force the spring into
the cartridge compartment! Forcing the spring into the compartment will damage the spring.
00. c. Gently and carefully, push down on edge
of cartridge return spring (with hole) and insert edge into slot.
00. c. Repeat “step c” to attach opposite side
of cartridge return spring to cartridge compartment.
00. c. IMPORTANT:When the cartridge return
spring is placed back into the instrument, the leaf spring must be orientated toward the back of the instrument and cartridge return spring must be securely attached via the grooves (tracks) and slots.
00. c. IMPORTANT: Failure to properly
replace return spring can result in erroneous test results.
13. Run optical test cartridge according to instructions that follow.
00. (Allows you to verify that the optical system
was not damaged or contaminated by dust, etc., during this maintenance procedure.)
7.6 Revised 6/03
7
INSTRUMENT CARE
AND ROUTINE MAINTENANCE
INSTRUMENT CARE AND ROUTINE MAINTENANCE
RUNNING THE OPTICAL TEST CAR TRIDGE (Standard 1)
The provided optical test cartridge allows you to monitor the performance of the optical system over time.
IMPORTANT! Keep a permanent record of the results obtained (i.e., Mean Transmittance, Standard Deviation and Drift).*
*It is recommended that you record the results on
the page provided in the appendix of this manual.
Comparing New Values With Initial Values
Compare the new values obtained with the results recorded initially for your instrument (see appendix). The mean Transmittance should be within the range of 0.9500 to 1.0500, and should not have varied by more than ±0.0100. The Standard Deviation should be less than 0.00150 and the Drift should be less than 0.01400.
DISPLA Y WHA T Y OU DO
1. Locate the bar code on the optical test
cartridge.
2. Hold the cartridge so that the bar code faces
right.
3. Insert the cartridge (above dot on instrument)
into the bar code track.
4. Quickly (within 1 second) and smoothly, slide
the cartridge down past the dot.
0. A beep sounds to signal a successful scan.
0.
If no beep sounds, repeat the procedure. If a beep repeatedly fails to sound, refer to Troubleshooting, Section 6.
5. Press .
RUN STANDARD 1?
READY: SCAN BAR CODE
02/24/03 2:09PM
Revised 6/03 7.7
7
INSTRUMENT CARE
AND ROUTINE MAINTENANCE
DISPLAY
6 MIN = total test time
After 1 minute:
WHA T Y OU DO
6. Open the cartridge compartment door.
7. Hold the optical test cartridge so that the bar
code faces right.
0. Insert the cartridge into the compartment until a
subtle snap is heard/felt.
0. HINT: The cartridge is designed to fit only one
way into the instrument.
8. Close door.
1.0001 T 0.00012 SD0 S1 5 MIN 10:03AM
PROCESSING STANDARD
S1 6 MIN 10:02AM
STANDARD 1
LOAD, CLOSE DOOR
7.8 Revised 6/03
7
INSTRUMENT CARE
AND ROUTINE MAINTENANCE
INSTRUMENT CARE AND ROUTINE MAINTENANCE
DISPLAY
Upon completion of test:
WHA T Y OU DO
09. Record the displayed results in the blanks
provided on the last page of this manual (appendix).
10. Remove cartridge.
00. a) Open cartridge compartment door.*
00. b) Locate the button on the right side of the
cartridge compartment. Push and hold it down with your right hand.
00. c) With your left hand, gently push the plastic
tab on the cartridge to the right; this action releases (unlocks) cartridge.
00. d) Pull cartridge out of compartment.
11. Compare the results obtained with results obtained initially.
00. If the results are not within the limits found
on page 7.6 and in the appendix, refer to Section 9 for information on “How to Report the Problem.”
00.
*If the door is opened (within 15 minutes after assay
completion), the test result is displayed for only 30 seconds.
00.
If the door is not opened, the test result will remain displayed for 15 minutes.
00. At 15 minutes, an audible tone (error buzz) sounds and
the display changes to “READY: REMOVE TEST.”
1.0001 T 0.00012 SDmm S1 0.00387 DRIFT
Revised 6/03 8.1
8
MINOR REPAIR
MINOR REPAIR
You can:
replace the fuse
If your instrument requires any other type of repair, refer to Section 9, Service Information.
How to Replace the Fuse
Tool Required: Screwdriver Fuse Required:
250 V, T-1.25A (Instrument Model No. 5031C) One replacement fuse is stored in the fuse holder.
The fuse holder is located on the back panel of the instrument between ON/OFF switch and power cord connector.
1.
WARNING: Set the power switch to OFF. Unplug power cord from wall
outlet. Unplug power cord from instrument.
VAROITUS: Kytke virta pois (off
-asento). Irrota virtajohto
seinäpistokkesta. Irrota virtajohto laitteesta.
2. Locate the groove on the side of the fuse holder
that faces power cord connector.
3. Insert the tip of a screwdriver into the groove.
Then, exert pressure to unsnap fuse holder from instrument.
4. Remove the fuse holder from instrument.
replacement fuse blown fuse fuse block
5. Remove and dispose of blown fuse.
6. Push the replacement fuse out of the black box
using the tip of a screwdriver or similar device.
7. Insert replacement fuse onto fuse block.
0.
WARNING: For continued protection against fire hazard, replace only with
the indicated type and rating of fuse.
0. VAROITUS: Paloturvallisuussyistä,
käytä vain suositeltuja sulakkeita.
8. Re-insert fuse holder into instrument (snap
gently into place.)
3
2
1
1
2
3
Groove
Revised 6/03 9.1
9
SERVICE INFORMATION
SERVICE INFORMATION
When you ha v e a problem with the System
Refer to Troubleshooting, Section 6, of this Manual
If Section 6 cannot assist you in solving the problem, answer all of the following questions before contacting us for help. The requested information will help identify the cause of the particular problem.
How to Report the Problem
First
Record the following information. Instrument Serial Number ___________________
Installation Date___________________________ Test Type ________________________________ Reagent Cartridge Lot Number _______________ Control Lot Number _______________________ Control Results ___________________________
Optical Test Cartridge Results:
Initial Current Mean Transmittance ________ _______ Standard Deviation ________ _______ Drift ________ _______
Second
Complete the following “Preservice Questionnaire.”
1. Has Section 6, Trouble- YES NO
0. shooting been reviewed?
2. Have you performed the
YES NO
0. required maintenance
0. procedures (Section 7)?
3. Set the power switch to OFF.
YES NO
0. Make sure the program card
0. is securely plugged in. Set
0. the power switch to ON. Are
0. the following displayed?
(if instrument is not within
operating temperature range)
4. If displays do not appear:
0.
is the instrument plugged YES NO
0.
into a live,AC electrical
0.
outlet?
0.
Is the line fuse defective? YES NO
0.
Refer to Section 7, Minor
0.
Repair.
5. Do any characters on the dis-
YES NO
0. play appear partially defective?
READY: SCAN BAR CODE
02/24/03 2:09PM
xWAIT: WARMING UP
02/24/03 2:09PM
INITIALIZING
KEEP DOOR CLOSED
COPYRIGHT 1991-2003
BY BAYER CORPORATION
SOFTWARE VERSION
E3.11/01.04
9.2 Revised 6/03
9
SERVICE INFORMATION
SERVICE INFORMATION
6.
If your instrument is connected
YES NO
0. to a printer or computer,
are all cables securely attached?
7. List error messages that have appeared. ______
0. ______________________________________
0. ______________________________________
0. ______________________________________
0. ______________________________________
0. ______________________________________
8. Record the exact sequence of events that took
place when the failure occurred and any results obtained.
9. Be prepared to perform a complete test
procedure when you call for assistance.
Third
Contact the nearest Bayer Diagnostics office or authorized distributor.
Revised 6/03 9.3
9
SERVICE INFORMATION
ACCESSORY ITEMS
Part Number Description Hemoglobin A
1c
5035C DCA 2000®Reagent Kit 5068A DCA 2000
®
Normal &
Abnormal Control Kit
Microalbumin/Creatinine
6011A DCA 2000
®
Reagent Kit
6012A DCA 2000
®
Low & High
Control Kit
NOTE: Part numbers are subject to change without notice.
TO ORDER:
Contact the nearest Bayer HealthCare office or authorized distributor.
Distributed by:
Bayer HealthCare LLC Elkhart, IN 46515 USA
REPLA CEMENT PAR TS
Part Number Description
SR002810 Program Card for Instrument
English 40330028 N.A. Power Cord for Instrument 40330046 Euro Power Cord for Instrument 40330050 U.K. Power Cord for Instrument 50214149 Filter Holder 50546216 Cartridge Return Spring 625-0127-01 Fuse: T-1.25A, Slow Blow; 250V 95001901 Cleaning Kit 95002911 Optical T est Cartridge 95002117 Air Filter (8) Replacement Kit 95002836 Operating Manual
NOTE: Part numbers are subject to change without notice.
Bayer House Strawberry Hill Newbury, Berkshire RG14 1JA UK
Ascensia™ Diabetes Support 0845 600 6030–UK 1 890 920 111–Republic of Ireland
diabetes@bayer.co.uk
Bayer plc MERA (Middle East, Eastern Europe, Russia, Africa) Bayer House Strawberry Hill Newbury Berkshire RG14 1JA UK +44 1635 563000
Bayer Inc. Diagnostics Division Toronto, Ontario M9W 1G6
Bayer (South East Asia), Pte Ltd No.9 Benoi Sector Singapore 629844 Tel: (65) 6261 3389 Fax: (65) 6266 3376
Bayer Australia Ltd. ABN 22 000 138 714 Diagnostics Division 875 Pacific Hwy Pymble NSW 2073 AUSTRALIA 1 800 028 251 (toll free)
Revised 6/03 App. 1.1
APPENDIX
APPENDIX
OPTICAL TEST CARTRIDGE RESULTS
Initial V alues
Date Mean Transmittance Standard Deviation Drift
DA TE MEAN TRANSMITTANCE STANDARD DEVIATION DRIFT
Comparing New Values With Initial Values
Compare the new values obtained with the initial values. The Mean Transmittance should be within the range of 0.9500 to 1.0500, and should not have varied by more than 0.0100. The Standard Deviation should be less than 0.00150 and Drift should be less than 0.01400.
Revised 6/03 Index 1.1
APPENDIX
A
aborted: 3.8 air: 2.4 analysis: 2.2 analyze: 2.6 assay: 2.3 authorized: 1.1
B
bar code: 2.2 beep: 4.8 biohazard: 4.5 blood: 1.1 bubbles: 4.7 buzz: 4.10
C
calibration: 5.2 capillary: 4.5, 5.5 card: 3.3 caution: 1.1, 4.1 cleaning: 2.1 compartment: 2.2 computer: 1.1, 3.16 computer port: 2.9 concentration: 2.16 copyright: 2.5 cursor: 2.10
D
definition: 3.13 displayed: 2.3
E
electrical: 1.1, 2.4 electronic: 2.1 error: 6.1
F
finger: 4.6 foil: 4.3 format: 3.2
G
glutaraldehyde: 7.2 grounded: 2.4
H
humidity: 1.1, 2.4
I
important: 2.3 insert: 5.4 installation: 1.1 IVD: 1.1
L
laboratories: 1.1
M
mechanical: 2.5 memory: 2.8 menu: 2.8 mg/dL: 2.16 mmol/L: 2.16
O
operational: 1.1 optical: 2.1
P
power: 2.1 printer: 1.1, 3.16 procedures: 4.3 progress: 3.8 prompts: 2.2
R
ratio: 3.3 reagent: 1.1 recall: 3.7 reconstitution: 1.2
INDEX
Index 1.2 Revised 6/03
APPENDIX
S
safety: 1.1 sequence: 2.8, 3.4 sodium hypochlorite: 7.2 software version: 2.5 specifications: 1.1 spring: 2.2 standards: 1.1 sunlight: 2.4
T
tab: 2.2 temperature: 1.1, 2.4
U
unpacking: 2.8 urine: 1.1, 5.6 use by: 1.2
V
venipuncture: 4.6
W
warning: 1.1, 4.5 warranty: 1.1
INDEX
Loading...