Bayer Clinitek Service manual

Page 1
1 INTRODUCTION AND
1 OF 8
GENERAL DESCRIPTION
Introduction ......................................................................................... 2
General Specifications ......................................................................... 5
NEXT
Page 2
INTRODUCTION AND GENERAL DESCRIPTION

1.1 Introduction

2 OF 8
This manual is to be used in conjunction with the Clinitek Sta in performing routine troubleshooting and service on an instrument. Each section of the service manual is dedicated to specific areas of interest: General features, methods overview, troubleshooting, error code descriptions, repair, functional diagrams, and parts breakdown. Any future service information or instrument revisions would be covered through service bulletins and/or revisions of this manual.
The CLINITEK capable of reading a range of Bayer Urine Reagent Strips along with the Bayer Clinitest hCG Immunoassay Cassette. It was developed as a next generation instrument combining features of the CLINITEK® 50 along with the CLINITEK® 100 plus the ability to read and interpret immunoassay cassette tests.
®
Status instrument is a small, portable urine chemistry analyzer
tus Operator’s Manual
NEXT
Page 3
INTRODUCTION AND GENERAL DESCRIPTION
3 OF 8

1.2 Features

The major feature is the lower cost and size, portability, centered around small office placement, converting from visual user to instrument. Thus, reducing the subjectivity in visual interpretation and combined with better patient to sample tracking.
Touch screen display for easier and faster operation,
Internal automatic calibration to improve reliability of readings,
Memory recall which allows customers to display the test results from specific samples,
Entry of patient information along with operator for each test result for improved
traceability,
Allows for entry of color and clarity of sample which can be printed along with results,
User Interface available in English, German, French, Spanish, Italian, and Swedish.
NEXT
Page 4
INTRODUCTION AND GENERAL DESCRIPTION
Figure 1-1: Clinitek Status
4 OF 8
NEXT
Page 5
INTRODUCTION AND GENERAL DESCRIPTION
1.3 General Specifications
SIZE:
Depth – 27.2 cm (10.7 in.) Width – 17.1 cm (6.7 in.) H
eight – 15.8 cm (6.2 in.)
WEIGHT:
1.66 Kg (3.65 lb)
INPUT VOLTAGE:
110 VAC ± 20%
45-65 Hz (US only)
220 VAC ± 20%
45-65 Hz (European)
9 VDC
5 OF 8
(6 AA Batteries)
MAXIMUM POWER INPUT:
Maximum power input to battery eliminator is 18.4 VA
Maximum power output from battery eliminator / input to instrument is 7.2 VA
NEXT
Page 6
INTRODUCTION AND GENERAL DESCRIPTION
THERMAL OUTPUT:
Battery eliminator 62.7 BTU/hr
Instrument 24.6 BTU/hr
Line Leakage Current:
170 microamperes in normal condition
Power supply is double insulated type, no leakage in fault condition.
(Testing protocol and allowable limits as specified by the safety standards for laboratory equipment outlined in UL 3101-1; CSA 22.2 No. 1010.1; IEC 1000-4 Sections 2, 3, 4, 5, 6, 8, and 11; CISPR 11; EN 50082-1 Jan. 1992; EN60950 5.2.2 and 5.2.3; and EN 50081-1 Jan. 1992.)
AMBIENT OPERATING TEMPERATURE RANGE:
10˚C to 40˚C (50˚F to 104˚F)
AMBIENT OPERATING HUMIDITY RANGE:
10% to 90% relative humidity
6 OF 8
OPTIMUM OPERATING CONDITIONS:
18˚C to 30˚C (64˚F to 86˚F)
10% to 85% relative humidity
This device complies to FCC Part 15 as a Class A device.
COMPUTER INTERFACE:
9-Pin RS-232 protocol with comma deliminated output. Null modem cable required.
NEXT
Page 7
INTRODUCTION AND GENERAL DESCRIPTION
PRINTER PAPER:
(Ludlow C-696)
Width — 5.8 cm (2.28 in.)
Length — 2500 cm (82 ft.)
Shelf Life — 5 years
Storage — <25˚C; <65% RH
DEFAULT SETTINGS:
Default settings are defined in the Operator’s Manual. New instruments are shipped from the manufacture set to English-Conventional.
7 OF 8
NEXT
Page 8
INTRODUCTION AND GENERAL DESCRIPTION

Instrument Programmed Reagent Strip Multiples and Cassettes:

8 OF 8
Multistix® 10 SG (N-Multistix® SG-L)
Multistix® 8 SG (Multistix® GP)
Multistix® 7
Multistix® 5
Multistix® SG
Multistix PRO® 10LS
Multistix PRO® 7PH
Multistix PRO® 7G
Multistix PRO® 6B
Multistix PRO® 6K
N-Multistix® SG
Uro-Hema-Combistix® SG-L
Uro-Hema-Combistix
Hema-Combistix
®
®
Uristix® 4
Uristix
®
Labstix® SG
N-Neostix
Uro-Uristix
®
®
Combistix® SG
Clinitek® Microalbumin
Clinitest® hCG
TO TOP
BACK
Page 9

2 METHODS OVERVIEW

1 OF 2
Introduction ........................................................................................ 2
NEXT
Page 10
METHODS OVERVIEW
2 OF 2

2.1 Introduction

The CLINITEK® Status instrument is a reflectance photometer, which reads the changes in color of the Bayer urine reagent strips and immunoassay cassettes. A series of six light emitting diodes (LED) each at a different wavelength illuminates the entire strip or cassette length. Reflected light from the strip is then passed through a lens and focused on a photodiode detector array (CCD) which then converts the light energy into an electronic signal. This signal is then passed through an A/D converter and then onto a microprocessor. Once there, the signal is processed into percent reflectances and then decodes. These decode values are compared to lookup tables containing specific decode ranges for each reagent and their respective clinical values. Results are reported to the customer via the touch screen, onboard printer, and/or through a computer interface.
For methods used on the urine reagent strips and immunoassay cassettes, refer to their specific product inserts.
TO TOP
BACK
Page 11

3 INSTALLATION

Introduction ........................................................................................ 2
1 OF 2
NEXT
Page 12
INSTALLATION

3.1 Introduction

2 OF 2
Refer to Section 1 “Introduction”
Manual for the proper installation and set up of the CLINITEK
1
Unpack the instrument from the shipping packaging.
2
Remove package tape from the front of the instrument, table shutter.
3
Insert test table with table insert approximately one half to two thirds of the way into the instrument.
Plug the power supply into a mains outlet.
4
5
Connect the power supply plug into the back of the instrument using the power jack on the instrument.
6
Push the power button momentarily to turn instrument on, instrument should beep once.
7
For initial, first time power up for the instrument, follow the “Start-Up Wizard.”
8
Subsequent power ups after the initial, the instrument should return to the “Select Ready” screen after instrument hardware and software checks.
and Section 3 “Start-Up Wizard” of the Operators
®
Status Analyzer.
TO TOP
BACK
Page 13
4 OPERATIONS /
1 OF 18
PROCEDURES
Introduction.......................................................................................... 1
Urinalysis Strip Test ............................................................................ 3
Cassette Method ................................................................................ 12
NEXT
Page 14
OPERATIONS / PROCEDURES
2 OF 18

4.1 Introduction

Refer to Section 4 “Instrument Operations” of the Operators Manual for the operation of the CLINITEK® Status Analyzer. This section describes both strip and cassette test procedures.
NEXT
Page 15
OPERATIONS / PROCEDURES
3 OF 18

4.2 Urinalysis Strip Test

A Full Strip Test allows you the option to enter an Operator ID, Patient Name and/or Patient ID prior to inserting a strip. The procedures to enter the Operator and Patient data are presented in this section. The strip testing process is identical to a Quick Strip Test.
Testing is started from the main Select screen.
Touch Strip Test to conduct a Bayer urinalysis strip test.
The next screen that appears is Operator ID.
NEXT
Page 16
OPERATIONS / PROCEDURES
There are two options under Operator ID: Last Operator or Enter New Operator ID.
OPTION 1: LAST OPERATOR
If this option is enabled, the last operator that entered his/her ID will be displayed on the screen in the lower right side of the box. If you are this operator, then touch the Last Operator button to proceed.
OPTION 2: ENTER NEW OPERATOR ID
In order to enter the information for a new operator, touch Enter New Operator ID button. The next screen that is displayed is Enter Operator ID.
4 OF 18
NEXT
Page 17
OPERATIONS / PROCEDURES
Use the keyboards to enter Operator ID using a maximum of 13 characters. Touch Enter when you have finished entering the ID and to move to the next screen.
The next screen displayed is Patient Information.
5 OF 18
There are two options under Patient Information: Recall Patient or Enter New Patient.
NEXT
Page 18
OPERATIONS / PROCEDURES
6 OF 18
OPTION 1: RECALL PATIENT
In order to look up previous patients, touch Recall Patient.
If previous patient identification has been entered, a list of up to 200 patient results will appear on the screen. Use the up and down arrow buttons to scroll through the list of patients. The most recently performed test will be shown at the top. Once the patient is highlighted, touch Select button.
The next screen will be Prepare Test.
NOTE: A total of 200 patient tests can be stored in the analyzer. The tests are listed in
chronological order. When the limit of 200 has been reached, the oldest test will be deleted from the analyzer. Deleted information cannot be retrieved from the analyzer.
NEXT
Page 19
OPERATIONS / PROCEDURES
OPTION 2: ENTER NEW PATIENT
In order to enter the information for a new patient, touch Enter New Patient button. The next screen displayed is Enter Patient Name.
Use the keyboards to enter Patient Name using a maximum of 20 characters. Touch Enter when you have finished entering the patient’s name and to proceed to the next screen.
7 OF 18
The next screen that appears is Prepare Test.
While the strip is being analyzed, a Select Appearance screen will be displayed. The urine sample must be visually observed and then the appropriate color and clarity must be selected.
NEXT
Page 20
OPERATIONS / PROCEDURES
If the urine sample is yellow and clear, touch the Yellow and Clear button.
If the urine sample is not yellow and clear, touch the Other button for more choices.
If you touched the Other button, then select the appropriate color by touching the circle button that corresponds to the correct description.
8 OF 18
NEXT
Page 21
OPERATIONS / PROCEDURES
Select the clarity by touching the circle that corresponds to the correct description. Then touch Next.
There is a time indicator on the Select Appearance screen that is counting down the time remaining in the analysis of the strip.
After color and clarity have been entered the next screen displayed will either be:
9 OF 18
Analyzing if the strip is still being analyzed
Results if analyzing the strip has been completed
NEXT
Page 22
OPERATIONS / PROCEDURES
The test results displayed on the screen and the printout will include the following information:
10 OF 18
Patient Name or ID
Type of strip used
Test date
Time
Operator
Test Number
Color
Clarity
Results
NEXT
Page 23
OPERATIONS / PROCEDURES
NOTE: If the results are positive, an asterisk* will appear next to the results (if “mark
positive results” was selected in Instrument Set Up).
11 OF 18
NEXT
Page 24
OPERATIONS / PROCEDURES

4.3 Cassette Method

A Full Cassette Test allows you the option to enter an Operator ID,
Patient Name and/or Patient ID prior to inserting a cassette.
The procedures to enter the Operator and Patient data are
presented in this section.
The cassette testing process is identical to a Quick Cassette Test.
Testing is started from the main Select screen.
12 OF 18
Touch Cassette Test to conduct an hCG pregnancy test.
NEXT
Page 25
OPERATIONS / PROCEDURES
The next screen that appears is Operator ID.
There are two options under Operator ID: Last Operator or Enter New Operator ID.
OPTION 1: LAST OPERATOR
If this option is enabled, the last operator that entered his/her ID will be displayed on the screen in the lower right side of the box. If you are this operator, then touch the Last Operator button to proceed.
OPTION 2: ENTER NEW OPERATOR ID
In order to enter the information for a new operator, touch Enter New Operator ID button. The next screen that is displayed is Enter Operator ID.
13 OF 18
NEXT
Page 26
OPERATIONS / PROCEDURES
Use the keyboards to enter Operator ID using a maximum of 13 characters. Touch Enter when you have finished entering the ID and to move to the next screen.
The next screen displayed is Patient Information.
There are two options under Patient Information: Recall Patient or Enter New Patient.
OPTION 1: RECALL PATIENT
In order to look up previous patients, touch Recall Patient.
14 OF 18
NEXT
Page 27
OPERATIONS / PROCEDURES
15 OF 18
If the previous patient identification has been entered, a list of up to 200 patients will appear on the screen. Use the up and down arrow buttons to scroll through the list of patients. The most recently performed test will be shown at the top. Once the patient is highlighted, touch the Select button.
The next screen will be Test Type.
NOTE: A total of 200 patient tests can be stored in the analyzer. The tests are listed in
chronological order. When the limit of 200 has beenreached, the oldest test will be deleted from the analyzer. Deleted information cannot be retrieved from the analyzer.
NEXT
Page 28
OPERATIONS / PROCEDURES
OPTION 2: ENTER NEW PATIENT
In order to enter the information for a new patient, touch the Enter New Patient button.
The next screen displayed is Enter Patient Name.
16 OF 18
Use the keyboards to enter Patient Name using a maximum of 20 characters. Touch Enter when you have finished entering the patient’s name and to move to the next screen.
NEXT
Page 29
OPERATIONS / PROCEDURES
The next screen displayed is Enter Patient ID. Use the keyboards to enter Patient Identification using a maximum of 13 characters. Touch Enter when you have finished entering the patient’s ID and are ready to proceed to the next screen.
The next screen that appears is Test Type.
17 OF 18
The test results displayed on the screen and the printout will include the following information:
Patient Name or ID
Type of Clinitest cassette test
Test date
NEXT
Page 30
OPERATIONS / PROCEDURES
18 OF 18
Time
Operator
Tes t N u m b e r
Result
NOTE: If the result is positive, an asterisk* will appear next to the result (if “mark positive
results” was selected in Instrument Set Up).
TO TOP
BACK
Page 31
5 PREVENTIVE
1 OF 7
MAINTENANCE
Introduction ........................................................................................ 2
Cleaning.............................................................................................. 3
NEXT
Page 32
PREVENTIVE MAINTENANCE
2 OF 7

5.1 Introduction

This section describes cleaning and preventative maintenance test procedures that should be performed as a matter of routine anytime that a CLINITEK® Status instrument is in Bayer Diagnostics possession. Following these procedures will help identify possible system weaknesses prior to them manifesting themselves as field failures.
Wear disposable surgical gloves when working with reagents, fluids, or cleaning the instrument.
Place all cleaning or disposable items in a biohazard bag.
NEXT
Page 33
PREVENTIVE MAINTENANCE
3 OF 7

5.2 Cleaning

Refer to Section 8 “Maintenance” of the CLINITEK® Status Operators Manual for specific cleaning and disinfecting instructions for the instrument, test table insert and test table.
CAUTION:
Care should be taken when handling the test table to insure that the calibration bar is not damaged, scratched or marred in any way. If any damage does occur, the test table must be replaced.

Daily

1
With the instrument power on, gently pull the test table carriage from the instru­ment.
Remove the test table insert and drain the drip tray if required.
2
3
Examine the calibration strip for any dust, dirt, scratches, or mars.
Do not touch the calibration strip.
4
Using a cotton tipped applicator and distilled water, thoroughly scrub the strip channel and surrounding area of the insert and also scrub the drip tray. Rinse the entire test table insert under running water.
Dry insert with a soft lint-free cloth or tissue.
5
If the calibration strip appears dirty, gently clean with a new cotton tipped
6
applicator and distilled water. Allow to air dry and examine again. Obtain a new test table if the calibration strip can not be cleaned, is scratched or marred.
NEXT
Page 34
PREVENTIVE MAINTENANCE
Refer to “Disinfecting Test Table and Test Table Insert” for disinfecting the test
7
table and test table insert.
8
To reinstall test table and insert, gently push table into instrument until halfway in.
4 OF 7

Disinfecting Test Table and Test Table Insert

If the test table or insert needs to be disinfected the following solutions may be used:
CAUTION:
CAUTION:
Precept
Cidex™ (3.2% glutaraldehyde solution)
Amphyl™ liquid
Theracide™ liquid
Preparation and use of the above products or their equivalents should be in accordance with their product directions.
Do not allow the disinfection solution to come in contact with the white calibration strip.
Any solutions other than these listed may damage the test table and insert.
NEXT
Page 35
PREVENTIVE MAINTENANCE
Household Bleach (5% sodium hypochlorite) Can be used either full strength or diluted to as much as a 20:1 dilution (5 mL of 5% sodium hypochlorite to 95 mL of water). Full strength 10% bleach should not be used since any residue not thoroughly rinsed off can affect several of the reagent chemistries.
Isopropyl alcohol (70% - 85% or undiluted)
Remove the test table and insert and clean according to “Disinfecting Test Table
1
and Test Table Insert” .
2
Fill a tall, narrow container to a depth of about 4 inches (10 cm) with the prepared disinfecting solution. An empty Multistix bottle makes an ideal container.
Place the test table or insert in the solution making sure the calibration strip does
3
not come into contact with solution.
4
Soak the table or insert for no longer than 10 minutes then rinse thoroughly with water.
5 OF 7
5
Dry with a soft cloth or tissue and replace test table and table insert in the instrument.
NEXT
Page 36
PREVENTIVE MAINTENANCE
6 OF 7

Removing Heavy Buildup

If the test table and table insert are cleaned regularly, heavy buildup should not occur. However, if urine has been allowed to dry on the table and insert a strong solution may be required to clean.
CAUTION:
1
Remove the test table and insert from instrument. Wet a cotton-tipped applicator with 0.1N sodium hydroxide (NaOH).
Clean the strip holding channel, surrounding areas, drip tray and table insert
2
until all traces of urine has been removed.
Rinse thoroughly with water.
3
Dry with a soft cloth or tissue and replace into instrument.
4
Do not allow the solution to come in contact with the white calibration strip.
NEXT
Page 37
PREVENTIVE MAINTENANCE
7 OF 7
General Instrument Cleaning
T
he outside of the CLINITEK Status instrument should be kept free of dust, dirt, and urine. If the instrument requires cleaning, turn the instrument off by pressing the power button for 2 seconds. The instrument can be wiped off with a damp cloth and any of the disinfection solutions listed or a mild detergent.
CAUTION:
Care should be taken to insure that no liquid is allowed to enter the printer compartment.
TO TOP
BACK
Page 38

6 THEORY OF OPERATION

1 OF 11
General Description ............................................................................ 2
LCD display and Touch Screen assembly .......................................... 3
Main PCB ............................................................................................ 4
Printer ................................................................................................. 9
Interconnection ....................................................................... 10
Block Diagram ........................................................ 11
NEXT
Page 39
THEORY OF OPERATION
2 OF 11

6.1 General Description

The Clintiek Status urine analyzer is a reflectance photometer designed to provide semiquantative clinical results when used with Bayer Urinalysis test strips and imuno assay cassettes. The instrument can be broken down into the following subsystems:
Optical system The Processor User interface
The optical system provides both llimation of the test cassette of strip using light from six LEDs. This light is directed on to the test surface by a light plastic light guide that provides unforim ilumation across the length of the read area. This light is then reflected off the read area to a mirer which directs the light to a lens that focuses the reflected image onto a CCD array.
The Clinitek Status optical system reads the entire length of the read area in one reading. The image of the read area is refected and focused on the CCD array, each element of the CCD array is then clocked out, amplified ,sent to a 10 bit A/D convertor, and the digital value to each ellemnt is then stored in an array in the instruments memory. This entire process is refered to a “Scan”. When a test is processed the system will scan the read area ten times and averaged, thus the average of the ten scans are used in the instrument aliogrothms for generating the clinical results. At the start of eact test the instrument scans the white calibration bar on the table to calibrate the instrument so that the sample refletance readings can be determined. Test results are displayed on the screen, they can also be printed on the internal printer or sent out the serial communcations port.
NEXT
Page 40
THEORY OF OPERATION
3 OF 11

6.2 LCD display and Touch Screen assembly

The LCD display and Touch Screen assembly is the primary user interface for the instrument. The LCD and integrated touch panel are supplied as a semi– customised assembly with a housing that directly attaches to the upper cover of the instrument. This provides a fluid resistant and electrically resistant seal to the housing. The assembly contains electronics that convert a serial stream of 4 bit wide data along with some necessary control signals to the electrical signals necessary to excite the LCD panel itself. The LCD then provides a 320 by 240 monochrome, single level (Black or background) display for viewing by the user from a single 5V supply and TTL/CMOS signals. An analog voltage input from the main PCB can then vary the contrast of the whole display. The LCD is interfaced directly to the main PCB by a single 14 way Flexible cable with printed connector. The 8 by 6 touch screen overlay is completely passive and contains no electronics. It provides an 8 by 6 crosspoint matrix of normally open resistive switches on a second 14 pin flexible cable with printed connector for direct interface to the main PCB. These crosspoints are scanned sequentially by the main board electronics for any closures. Should multiple closures be detected (which may be ambiguous), the software in the main board decides on the outcome of these scenariosThis unit contains embedded electronics to format a 4-bit wide data stream into a 320 by 240 pixel display and control the contrast of the display. The resistive touch screen provides an 8 by 6 passive overlay on the LCD.
NEXT
Page 41
THEORY OF OPERATION
4 OF 11

6.3 Main PCB

A single PCB that contains all of the necessary control circuits is mounted on the This “Main PCB “contains the following:
a.
The Microcontroller used to control most of the functions within the instrument is a Hitachi H8/3006. This component incorporates an internal 32 bit CPU archi­tecture with a 16 bit register set along with embedded RAM, timers, serial com­munications, memory interface, DMA, I/O ports, ADC and DAC functions. In this design it is powered from the 5V supply and receives a system clock from an external oscillator at a frequency of 18.432MHz. External RAM and ROM (Flash) support the Microcontroller. A large number of the pins / ports on the device are used for this high-speed interface.
The on chip serial communications controllers are used to interface with the external RS232 interface, an internal serial interface to a future expansion card and to the SIP bus that communicates with the NVM, MMC card and RTC.
In general, the I/O ports on the processor are used as inputs and the output functions are segregated to the U26 CPLD, as there are insufficient on-chip ports to meet all of the system requirements. The digital input ports come from the touch screen sensing, power state sensing, drawer position sensing & printer status.
The timer / counters are used to provide:
A general system timing function.
The sounder drive.
The PWM modulation for controlling the speed of the drawer mechanism.
NEXT
Page 42
THEORY OF OPERATION
The LCD interface which stimulates the DMA to transfer the display image repeatedly independent of the CPU operation.
The CCD interface which stimulates the ADC and DMA to transfer data from the CCD into memory independent of the CPU operation.
The 8 bit DAC’s are used to set the LED illumination level when scanning the sample, the contrast level on the LCD and the offset adjustment to the CCD variable gain amplifier.
The 10 bit ADC is used in conjunction with its input multiplexer to sample analog voltages from the CCD, the raw supply voltage, the printer head current and head temperature and the drawer motor current.
The Microcontroller is reset by an external device (U11, via U26). U11 also provides a watchdog function so that the instrument is stopped from operating in the event of a timeout.
b.
The External Memory provided consists of a single 512K byte Static RAM device for transient data and a single 1 M byte Flash ROM device for the program memory and static data. The ROM is initially assembled onto the PCB with a boot loader pre-loaded into it. Subsequently, the ROM has the ‘Status’ application loaded into it during board test using the serial port at high speed. In service, the ROM can be updated either from an MMC card being inserted into the MMC connector from under the printer cover of the instrument or from the serial connection should it be required.
5 OF 11
The Programmable logic device (CPLD - U26) is generally used to provide
c.
output latches and hardware processing functions that were not available on the Microcontroller. The device is a flash memory programmable device and can be reprogrammed in-situ by attaching a programmer to the connector position
NEXT
Page 43
THEORY OF OPERATION
provided on the PBA. The CPLD complexity is quite low (72 storage elements) and so only provides simple functions. In particular, it converts and re-times the 8 bit data into 4bit data for interface to the LCD, drives the touch screen inputs, the printer head and paper feed functions, the sounder, the sample drawer driver, the CCD amplifier gain and the power switching functions.
The LED Driver circuit has a digital part that switches on the appropriate LED
d.
when required (under CPU control) and an analog part that converts the voltage output of one of the Microcontroller DAC’s to a current output. The current is then used to vary the drive level of the LED’s used for illuminating the sample. Additionally, two of the digital outputs from this circuit control the power to the drawer position sensing opto-couplers.
The Variable Gain Amplifier is used to amplify the signal from the CCD on the
e.
Emitter / Detector PCB before application to the ADC in the Microcontroller. The gain of this amplifier is by default set to +2, but may be set to gains of +1, +3 and +4 as well when required by the Microcontroller via output ports in the CPLD. To accommodate the voltage offset from ground of the CCD signal compensation is applied to the amplifier by a DAC on the Microcontroller.
6 OF 11
f.
The Drawer Motor Driver takes a PWM signal from the Microcontroller, a direction signal and control signals from the CPLD and produces a high current ‘H Bridge’ drive to the brushed dc. motor. This allows the drawer motor speed and direction to be altered by the Microcontroller as well as providing braking of the motor when required. A circuit to allow sampling of the drawer motor current for future use is also provided that feeds into an ADC channel on the Microcontroller.
NEXT
Page 44
THEORY OF OPERATION
The Drawer position sensors mounted on the flexible cable assembly are driven
g.
by two outputs from the LED driver circuit and the signals returned from the sensors interfaced to digital inputs on the Microcontroller via the CPLD. Additionally two signals are generated by the CPLD to drive interrupt inputs on the Microcontroller in order to provision for fast servicing of the drawer position state when required.
h.
The non-volatile memory (NVM) device is a serial EEPROM device that can store 256k bits of data as 32K bytes of storage that is retained without power on the instrument. The device is used amongst other things for the storage of instrument set-up data and calibration data.
The Real Time Clock (RTC) device is a serial device that stores the current date
i.
and time and is powered from the Instrument 5V supply when it is present or from a permanently fitted primary lithium cell when the instrument supply is not present. The device communicates over the SPI bus with the Microcontroller, using its own protocol. The device has its own 32.786KHz crystal timing source.
j.
The Multi-Media Card (MMC) interface is comprised of a connector and some protection components, so that an MMC card can be inserted into the connector and read over the SPI bus, using the SPI protocol. The data recorded on the card is for the purposes of software updates in the field. A specially formatted card is recognised at power up and read into memory so that reprogramming of the software application can take place.
7 OF 11
The Sounder is a simple ceramic capsule fitted to the main PCB. The sounder is
k.
driven in a bridge mode from two outputs of the CPLD. The frequency of the sound is derived from one of the timers on the Microcontroller.
NEXT
Page 45
THEORY OF OPERATION
The Printer Paper Feed Driver is a stepping motor driver that is stepped using
l.
two signals from output latches in the CPLD using a gray code sequence. The device boosts the current capabilities of the CPLD and provides a bipolar drive for each winding of the motor. An analog circuit monitors the current that the motor windings are consuming and chops the waveform to keep the current at or below one of two predetermined levels that can be set by the Microcontroller.
m.
The Printer Head monitoring circuits buffer and amplify the head current and head temperature signals before sending them to the ADC on the Microcontroller. The head temperature should not normally exceed a pre-set temperature, in the event that it does the software turns off the instrument. The printer head current is monitored by amplifying the voltage across a low value resistor in series with the head elements. The signal is then sent to the ADC on the Microcontroller.
n.
The Printer “head up” and paper out opto sensors on the printer are powered by main board through resistor feeds and are interfaced back to the Microcontroller for digital sensing of the states of these signal in order to inhibit printing under paper out or head up conditions.
8 OF 11
The RS232 Interface acts as a level changer between the 5V logic levels
o.
required at the Microcontroller pins and the +/- 6V (min) signals required to meet the RS232 specification at the connector on the end of the flexible cable assembly. The device used is a +/- 15KV ESD compliant part with on board charge pumps to generate the required voltages for the interface.
NEXT
Page 46
THEORY OF OPERATION
9 OF 11

6.4 Printer

The printer is a thermal printer that prints rows of 384 dots as required across the special paper supplied from a roll. The paper is then moved so that the next row of dots can be printed to build up a pixel by pixel graphical image on the paper. The electronics in the printer takes a serial stream of 8 parallel inputs and determines which of the thermal heads should be energized. The printing head electronics also contains a temperature sensor that monitors the temperature of the printhead. This electronics is connected directly to the main PCB via a high current, 20 way flexible cable with printed connector.
The paper feed to expose the paper and print the next row of dots is driven by a simple stepping motor. There are no electronics to drive this motor in the printer. The two motor windings are presented directly to the main PCB electronics via a 9 way flexible cable with printed connector. On the same flexible cable, the connections to two opto-couplers that sense the printer head position and paper present conditions are provided for action by the main PCB electronics.
A 5V logic supply and a 4.2 – 8.5V power source from the main PCB power the printer. The current taken by the printer has to be actively managed by the main board electronics to provide an acceptably dense and even image and regulate how much power is consumed at each printing operation.
NEXT
Page 47
THEORY OF OPERATION
10 OF 11

6.5 Interconnection

A flexible printed circuit assembly provides the interconnections between the main PCB to the power supply, Drawer mechanism, battery pack and RS232 connector. At one end it has a 30 way FPC male connector printed within it for connection to the main PCB. The connections from the main board connector are then distributed to the other connectors and assemblies directly attached to exposed pads on the flexible printed circuit.
Connections are made to:
The power inlet connector (with EMC ferrite in line), the RS232 connector with
a.
EMC ferrite on connector pins.
b.
The battery pack via soldered interconnect leads.
The drawer opto-coupler positional feedback devices mounted on a small rigid
c.
section of this circuit.
d.
The drawer motor and noise suppression component(s).
NEXT
Page 48
THEORY OF OPERATION

6.6 Block Diagram

Figure 6-1: Block Diagram
11 OF 11
TO TOP
BACK
Page 49
7 DIAGNOSTICS
1 OF 60
TROUBLESHOOTING
Introduction ........................................................................................ 2
Troubleshooting Table ....................................................................... 3
Instrument Diagnostic Screens ........................................................ 40
NEXT
Page 50
DIAGNOSTICS TROUBLESHOOTING
2 OF 60

7.1 Introduction

This chapter is divided into two different sections. The first section is devoted to troubleshooting tables and error code tables giving a brief description and steps the customer and service individual would need to take. The second section goes into a description of the diagnostic screens available on the instrument.
NEXT
Page 51
DIAGNOSTICS TROUBLESHOOTING
3 OF 60

7.2 Troubleshooting Table

The following table is provided to assist in troubleshooting the CLINITEK Status system. The first four columns are similar to those found in the Operators Manual under troubleshooting with the exception that “Contact Bayer Service” has been removed from Customer Remedy column and replaced with “Replace instrument.”
The fifth column, “Service Remedy,” gives a more advanced troubleshooting guide that should be utilized by Bayer Service Personnel or under the guidance of trained personnel. The items are listed in increasing order of technical difficulty.
In some isolated cases, the power plug must be unplugged from the back of the instrument in order to turn the instrument off or batteries removed. Once the plug or batteries has been replaced, the power button on front of the instrument must be pressed to turn the instrument back on.
NEXT
Page 52
DIAGNOSTICS TROUBLESHOOTING
4 OF 60
Symptom or Error
Touch screen does not respond
Meaning Possible
Cause
Electronic lock-upHave the customer
Faulty touch screen
Loose cable Replace instrument Same as above
Customer Remedy Service Remedy
disconnect from power and reconnect, and turn instrument on
Replace instrument
Reseat cables to the
1
main PCB
2
Replace display / bezel assemble
3
Replace main PCB.
NEXT
Page 53
DIAGNOSTICS TROUBLESHOOTING
5 OF 60
Symptom or Error
Blank Display
Meaning Possible
Cause
No power
Defective Display Unit
Battery contacts corroded
Customer Remedy Service Remedy
1
Check that the power cord is firmly plugged into the instrument and the line cord is plugged into a live AC electrical outlet and firmly plugged into the Power Supply.
Replace batteries.
2
Replace instrument
Replace instrument Replace optical chassis
1
Test instrument using a second power supply and line cord.
Inspect battery holder
2
and replace batteries as needed.
1
Remove Bezel and check that all cables are properly seated.
Troubleshoot the Main
2
PCB.
NEXT
Page 54
DIAGNOSTICS TROUBLESHOOTING
6 OF 60
Symptom or Error
Faint Display
Missing Character s or Strange Character s
Meaning Possible
Cause
Possible low contrast setting
Possible low contrast setting
Faulty display Replace instrument
Electronics lock-up
Loose Cable Replace instrument Run Display diagnostics
Customer Remedy Service Remedy
Refer to operators manual for adjusting contrast
1
Replace display / bezel assembly
Replace main PCB
2
Disconnect and reconnect from power source, and turn instrument “on”.
Reseat cables
1 2
Replace Display/Bezel Assembly
3
Replace Main PCB
Faulty display Replace Instrument Replace Main PCB
Main PCB Failure
Replace instrument Replace Main PCB
NEXT
Page 55
DIAGNOSTICS TROUBLESHOOTING
7 OF 60
Symptom or Error
Test Table cannot be installed
Meaning Possible
Cause
Broken Table Guides
Customer Remedy Service Remedy
1
Inspect the Test Ta ble opening on the instrument for any obstructions
nspect bottom of
2
I
Test Table for any obstructions blocking the teeth.
3
Tr y inserting a different Test Table.
Inspect the Test Table
4
opening on the instrument for any obstructions such as bent guide pins.
NEXT
Page 56
DIAGNOSTICS TROUBLESHOOTING
8 OF 60
Symptom or Error
Instrument does not turn “ON” or turn “OFF
Meaning Possible
Cause
Front panel switch is not functioni ng correctly
No Power
Dirty switch connect
Main PCB Failure
Customer Remedy Service Remedy
1
Check that the power cord is firmly plugged into the instrument and the line cord is plugged into a live AC electrical outlet and firmly plugged into the Power Supply.
Replace batteries.
2
Replace instrument
Replace instrument Replace Main PCB
1
Test system using a second power supply and line cord.
Inspect battery holder
2
and replace batteries as needed.
1
Remove Bezel and clean power switch and PCB switch con­tacts Troubleshoot the Main PCB.
Replace Display/Bezel
2
assembly
NEXT
Page 57
DIAGNOSTICS TROUBLESHOOTING
9 OF 60
Symptom or Error
Forgot password
Meaning Possible
Cause
Password set, and operator does not know it
Customer Remedy Service Remedy
1
Have customer check to see writ­ten operators manual.
From the main
2
screen press instrument set-up, then enter the password ”3654”. Then select choose setting option. Then select remove password. Then if they want to reset the manual instruct refer them to the operators manual.
From the main screen press instrument set-up, then enter the password” select choose setting option. Then select remove password. Then if they want to reset the manual instruct refer them to the operators manual.
3654
”. Then
NEXT
Page 58
DIAGNOSTICS TROUBLESHOOTING
10 OF 60
Symptom or Error
Test Table does not move into instrument when a test is started
Test Table does not move into instrument when instrument is powered off
Meaning Possible
Cause
Instrument “Lock-Up”
Broken teeth on test table
Defective Test Ta ble Motor Drive Unit
Instrument “Lock-Up”
Broken teeth on test table
Defective Test Ta ble Motor Drive Unit
Customer Remedy Service Remedy
Tu rn the instrument off and on.
Replace table Replace test table
Replace instrument Replace the Optical
1
Tu rn the instru­ment off and on.
Perform test table
2
diagnostics
Replace instrument Replace Test Table
Replace instrument Replace optical chassis
Perform Table Movement test through the Diagnostics menu.
Chassis that contains the Motor Drive Unit.
NEXT
Page 59
DIAGNOSTICS TROUBLESHOOTING
11 OF 60
Symptom or Error
Results are not printing
Meaning Possible
Cause
Printer option disabled
Paper in backwards
Incorrect paper being used
Batteries are low
Printer defective
Printer cable loose
Customer Remedy Service Remedy
Refer to operators manual to select printer option
Refer to operators manual for how install paper
Recommend that they use Bayer thermal paper
Replace batteries or use power supply
Run printer diagnostics; refer to operators manual for instructions. If it fails replace instrument
Run printer diagnostics; refer to operators manual for instructions. If it fails replace instrument
Run printer diagnostics
Check cable
1
connections.
2
Replace printer
Replace main PCB
3
Run printer diagnostics
1
Check cable connections.
2
Replace printer
3
Replace main PCB
NEXT
Page 60
DIAGNOSTICS TROUBLESHOOTING
12 OF 60
Symptom
Meaning Possible
or Error
E00 Not used
E01
E02
Low battery power
Battery voltage drops to
5.8 Vdc
Failure of Calibrati on data checksu m
Cause
Batteries have low voltage
Battery contacts corroded
Power supply is faulty
Battery detector circuit is damaged
Main PCB failure
Customer Remedy Service Remedy
Replace with fresh batteries.
Clean if possible, replace instrument
Replace power supply
Replace instrument
Return for service Replace Main PCB
Replace with fresh batteries.
Replace optical chassis
1
Replace power supply
Power input connector
2
damaged, replace optical chassis
1
Reload instrument software
Replace main PCB
2
NEXT
Page 61
DIAGNOSTICS TROUBLESHOOTING
13 OF 60
Symptom or Error
E03
E04
E05
E06
E07
E08
E09
E10
Meaning Possible
Cause
Failure of Flash (ROM) checksu m
Failure of RAM checksu m
Not used
Loss of test results
Memory checksum error
Main PCB failure
(Non Volatile Memory failure during storage of test results)
Customer Remedy Service Remedy
1
If customer has a program card, have them attempt to reload software
Return for service
2
Return for service Replace Main PCB
Tu rn instrument off and on and repeat the test.
1
Reload software
Replace Main PCB
2
Replace Main PCB
NEXT
Page 62
DIAGNOSTICS TROUBLESHOOTING
14 OF 60
Symptom or Error
E11
Meaning Possible
Cause
Failure of test table
Ta ble fails to move 2 steps in <4 seconds.
Insert or Cassette not fully into table
Ta ble bumped Disconnect from
Customer Remedy Service Remedy
1
Reseat the table insert.
Remove the table
2
inset and inspect that there is no debris in the table preventing the insert from seating.
3
Clean and inspect the table insert, if damaged replace the insert. Confirm that the test table is in place.
power source to reset and clear the error (
Tur ning instrument “off” and “on” will not reset and clear the error
Run table diagnostics to troubleshoot.
Replace test table.
Replace lower optical chassis.
).
NEXT
Page 63
DIAGNOSTICS TROUBLESHOOTING
15 OF 60
Symptom
Meaning Possible
or Error
E12
E13 Not used
E20
Failure of LED
During the calibratio n step the instrume nt could not set the current gain on one or more of the Led
Real time clock failure
Customer Remedy Service Remedy
Cause
Calibration strip dirty
Failure of LED Replace the
Main PCB failure
Electronics lock-up
Main PCB failure
1
Remove the table and clean the table as directed in the operators manual
Replace the table
2
Replace the instrument
instrument
Replace the instrument
Tu rn instrument off and on.
Run electronics diagnostics.
Run light source diagnostics.
Replace the table
Replace lower optical chassis
Replace the Main PCB
Run electronics diagnostics.
Replace Main PCB
E21 Not used
NEXT
Page 64
DIAGNOSTICS TROUBLESHOOTING
16 OF 60
Symptom or Error
E23
Meaning Possible
Cause
Low battery power approach ing (one segment on battery icon on display
Battery voltage is below
6.6Vdc
Batteries have low voltage
Battery contacts corroded.
Power supply is faulty
)
Battery detector circuit is damaged
Customer Remedy Service Remedy
Replace with fresh batteries.
Clean if possible, replace instrument
Replace power supply
Replace instrument
Replace with fresh batteries.
Replace optical chassis
1
Replace power supply
2
Power input connector damaged, replace optical chassis
1
Reload instrument software
2
Replace main PCB
NEXT
Page 65
DIAGNOSTICS TROUBLESHOOTING
17 OF 60
Symptom
Meaning Possible
or Error
E24 No
printer paper
Cause
No printer paper
Printer defective
Main PCB failure
Customer Remedy Service Remedy
Replace the printer paper
If paper is in installed:
1 Cycle power off
and on
2 Replace
instrument.
If paper is in installed:
1 Cycle power off
and on
2 Replace
instrument
Replace the printer paper and check that flex cables from printer to main PCB are seated in their connectors
Replace printer
Replace Main PCB
E25 Auto
Calibrati on failure
Same as E12
Same as E12 Same as E12 Same as E12
NEXT
Page 66
DIAGNOSTICS TROUBLESHOOTING
18 OF 60
Symptom
Meaning Possible
or Error
E27 Set up
failure
(loss of test results or instrume nt set up due to checksu m failure)
E28 Printer
error
Cause
Instrument lock-up
Main PCB Failure
Printer paper release lever arm is not down.
Printer cable not seated fully in connector
Printer defective
Customer Remedy Service Remedy
Tu rn instrument off and on.
Replace instrument Replace Main PCB
Lift the printer cover and push the paper holding arm back into position.
Replace instrument Check that printer cable
Replace instrument Replace printer
Cycle instrument power off and on.
Lift the printer cover and push the paper holding arm back into position
in fully seated and locked in place
Main PCB Failure
Replace instrument Replace Main PCB
NEXT
Page 67
DIAGNOSTICS TROUBLESHOOTING
19 OF 60
Symptom
Meaning Possible
or Error
E48 Loss of
test results.
The same as E10
Cause
(Non Volatile Memory failure during storage of test results)
Customer Remedy Service Remedy
Tu rn instrument off and on and repeat the test.
Replace Main PCB
NEXT
Page 68
DIAGNOSTICS TROUBLESHOOTING
20 OF 60
Symptom
Meaning Possible
or Error
E50 Incorrect
strip type
(Failure to identify strip)
Cause
Strip type used does not match setting selected.
Instrument failed to read correctly match strip ID bands to pad configuration.
Customer Remedy Service Remedy
Ensure that the strip type selected in Instrument set up is being used (see Section 6, Select Instrument Tests).
1 If the correct type
of strip is being used, check the instrument opera­tion by running another test using.
2 Excess sample is
on the color ID bands, instruct the customer on proper dipping technique (refer to Operators manual).
3 Have the
customer run a Control such as Chek-Stix.
NEXT
Page 69
DIAGNOSTICS TROUBLESHOOTING
21 OF 60
Symptom or Error
Meaning Possible
Cause
Strip was slightly tilted on table
Instrument problem Test Ta ble, Main PCB, or Lower Chassis
Customer Remedy Service Remedy
1 Have customer
rerun the sample.
2 Have the
customer run a Control such as Chek-Stix.
Replace instrument
1 Replace Test Table 2 Replace Main PCB 3 Replace lower optical
assembly.
NEXT
Page 70
DIAGNOSTICS TROUBLESHOOTING
22 OF 60
Symptom
Meaning Possible
or Error
E52 Invalid
cassette barcode
Area of barcode on cassette does not match supporte d cassette tests
Cause
Cassette barcode label damaged
Cassette not seated in insert.
Insert not installed in to table correctly
Insert damaged
Ta ble damaged
Instrument failure
Customer Remedy Service Remedy
Repeat the test using a Bayer cassette.
Clean insert and retest
Clean table and retest
Replace insert
Replace table.
Replace instrument
1 Replace test table 2 Replace lower Optical
assembly.
3 Replace Main PCB
NEXT
Page 71
DIAGNOSTICS TROUBLESHOOTING
23 OF 60
Symptom
Meaning Possible
or Error
E53 Strip Test
selected but cassette detected
Cassette barcode detected
Cause
Incorrect test was selected
Instrument failure
Customer Remedy Service Remedy
Repeat the test using the appropriate test routine.
Have the customer use the CSCC test to confirm instrument failure.
If confirmed replace instrument.
If CSCC tests are successfully, have customer run controls.
Replace lower Optical assembly.
Replace Main PCB
NEXT
Page 72
DIAGNOSTICS TROUBLESHOOTING
24 OF 60
Symptom
Meaning Possible
or Error
E54 Cassette
Test selected but strip detected
No barcode detected but a test is present
Cause
Incorrect test was selected
Instrument failure
Customer Remedy Service Remedy
Repeat the test using the appropriate test routine.
Have the customer use the CSCC test to confirm instrument failure.
If confirmed replace instrument.
If CSCC tests are successfully, have customer run controls.
Replace lower Optical assembly.
Replace Main PCB
NEXT
Page 73
DIAGNOSTICS TROUBLESHOOTING
25 OF 60
Symptom
Meaning Possible
or Error
E55 Strip
being used on cassette side of table insert
%R in all areas is > black and no cue mark is detected
Cause
Incorrect side of insert being used
Instrument optical error
Customer Remedy Service Remedy
1 Tu rn the test table
insert over so that the strip holder face of the insert is upwards.
2 Repeat the test.
Have the customer use the CSCC test to confirm instrument failure.
If confirmed, replace instrument.
If CSCC tests are successfully, have customer run controls
1 Replace lower Optical
assembly.
2 Replace Main PCB
NEXT
Page 74
DIAGNOSTICS TROUBLESHOOTING
26 OF 60
Symptom
Meaning Possible
or Error
E56 Incorrect
size test table
4.25” strip on a
3.25” test table
Cause
Long strip being used on a short strip insert
Instrument failure
Customer Remedy Service Remedy
Verify that the customer is using the correct insert with the strip that they are using.
Repeat the test using the correct test table.
Have the customer use the CSCC test to confirm instrument failure.
If confirmed, replace instrument.
If CSCC tests are successfully, have customer run controls
Replace lower Optical assembly.
Replace Main PCB
NEXT
Page 75
DIAGNOSTICS TROUBLESHOOTING
27 OF 60
Symptom
Meaning Possible
or Error
E57 Missing
strip or cassette
%R in all areas is equal to black
Cause
Test started with out either a strip in the insert
Test started without an insert in table
Instrument failure
Customer Remedy Service Remedy
Repeat the test ensuring that the strip or cassette is correctly positioned when the test is started
Repeat testing using the correct insert installed on the table
Have the customer use the CSCC test to confirm instrument failure.
If confirmed, replace instrument.
If CSCC tests are successfully, have customer run controls Replace Instrument.
1 Replace lower Optical
assembly.
2 Replace Main PCB
NEXT
Page 76
DIAGNOSTICS TROUBLESHOOTING
28 OF 60
Symptom
Meaning Possible
or Error
E58 Misplace
d strip
845nm% R <20% from cue mark + 50 pixels and 565nm% R >17%
Cause
Test strip is not correctly positioned in the insert; the tip is not to the end of the channel.
Customer Remedy Service Remedy
1 Repeat the test
ensuring that the strip is correctly positioned on the test table.
2 Have customer
clean the insert, checking to see that there is not any build up near edges of the channel Repeat the test ensuring that the strip is correctly positioned on the test table.
3 Have the
customer remove the table and clean the gears.
4 Replace the Insert 5 Replace the table
NEXT
Page 77
DIAGNOSTICS TROUBLESHOOTING
29 OF 60
Symptom or Error
Meaning Possible
Cause
Instrument failure
Customer Remedy Service Remedy
Have the customer use the CSCC test to confirm instrument failure.
If confirmed, replace instrument.
If CSCC tests are successful, have customer run controls
Replace the instrument
1 Replace lower Optical
assembly.
2 Replace Main PCB
NEXT
Page 78
DIAGNOSTICS TROUBLESHOOTING
30 OF 60
Symptom
Meaning Possible
or Error
E59 Inverted
strip
Cause
Strip placed upon the table upside down.
Instrument failure
Customer Remedy Service Remedy
Repeat the test ensuring that the strip is correctly positioned on the test table.
Have the customer use the CSCC test to confirm instrument failure.
If confirmed, replace instrument.
If CSCC tests are successfully, have customer run controls.
Replace the instrument
1 Replace lower Optical
assembly.
2 Replace Main PCB
NEXT
Page 79
DIAGNOSTICS TROUBLESHOOTING
31 OF 60
Symptom
Meaning Possible
or Error
E60 Tilted
strip
Cause
Strip not laying flat in the insert channel
Insert not flat in the table
Customer Remedy Service Remedy
1 Repeat the test
ensuring that the strip is correctly positioned on the test table.
2 Have customer
clean the insert, checking to see that there is not any build up near edges of the channel. Repeat the test ensuring that the strip is correctly positioned on the test table.
Have the customer clean and inspect the table for anything that might prevent the insert from sitting flat in the table.
NEXT
Page 80
DIAGNOSTICS TROUBLESHOOTING
32 OF 60
Symptom or Error
Meaning Possible
Cause
Instrument failure
Customer Remedy Service Remedy
Have the customer use the CSCC test to confirm instrument failure.
If confirmed, replace instrument.
If CSCC tests are successfully, have customer run controls.
Replace the instrument
1 Replace lower Optical
assembly.
2 Replace Main PCB
NEXT
Page 81
DIAGNOSTICS TROUBLESHOOTING
33 OF 60
Symptom or Error
E61 Dry strip Strip was not
Meaning Possible
Cause
fully dipped in to the sample.
Instrument failure
Customer Remedy Service Remedy
Refer the customer to the operators manual section 4 on dipping the strips. Repeat the test.
Have the customer use the CSCC test to confirm instrument failure.
If confirmed replace instrument.
If CSCC tests are successfully, have customer run controls. Replace the instrument.
1 Replace lower Optical
assembly.
2 Replace Main PCB
NEXT
Page 82
DIAGNOSTICS TROUBLESHOOTING
34 OF 60
Symptom
Meaning Possible
or Error
E62 Light
ingress
Cause
Light is leaking into the instrument.
Instrument failure
Customer Remedy Service Remedy
If instrument is in direct sunlight or strong side lighting, move out of bright light.
Have the customer use the CSCC test to confirm instrument failure.
If confirmed replace instrument.
If CSCC tests are successfully, have customer run controls
Replace the instrument
1 Check shutter door for
proper operation. Clean and repair if pos­sible.
2 Replace lower optical
chassis.
3 Replace Main PCB
NEXT
Page 83
DIAGNOSTICS TROUBLESHOOTING
35 OF 60
Symptom
Meaning Possible
or Error
E63 Failure to
find end of strip
(Fails to locate cue mark on table insert, strip side)
Cause
Insert dirty or damaged
Ta ble damaged
Instrument failure
Customer Remedy Service Remedy
1 Have the cus-
tomer clean and inspect the insert to see the white line at the end of the channel is present.
2 Replace the insert 3 Repeat the test
ensuring that the strip is correctly positioned on the test table.
Have the customer make sure table is clean and not damaged.
Have the customer use the CSCC test to confirm instrument failure.
1 Replace lower Optical
assembly.
2 Replace Main PCB
If confirmed replace instrument.
If CSCC tests are successfully, have customer run controls
Replace the instrument
NEXT
Page 84
DIAGNOSTICS TROUBLESHOOTING
36 OF 60
Symptom
Meaning Possible
or Error
E64 Over-
range CAL strip
Cause
Calibration bar is dirty.
Instrument failure
Customer Remedy Service Remedy
1 Clean and inspect
the calibration bar.
2 Replace Test
Ta ble
Have the customer use the CSCC test to confirm instrument failure.
If confirmed replace instrument.
If CSCC tests are successfully, have customer run controls
Replace the instrument
1 Replace Optical chas-
sis
2 Replace Main PCB
NEXT
Page 85
DIAGNOSTICS TROUBLESHOOTING
37 OF 60
Symptom
Meaning Possible
or Error
E65 Dirty
CAL strip
Cause
Calibration bar is dirty
Instrument failure.
Customer Remedy Service Remedy
1 Clean and inspect
the calibration bar.
2 Replace Test
Ta ble
Have the customer use the CSCC test to confirm instrument failure.
If confirmed, replace instrument.
If CSCC tests are successfully, have customer run controls
Replace the instrument
1 Replace Optical chas-
sis
2 Replace Main PCB
NEXT
Page 86
DIAGNOSTICS TROUBLESHOOTING
38 OF 60
Symptom
Meaning Possible
or Error
E66 Dry
cassette
Cause
Insufficient sample
Instrument failure
Customer Remedy Service Remedy
1 Review the direc-
tions for applying the sample to the cassette with the customer.
2 Repeat the test 3 Have customer
run a control.
Have the customer use the CSCC test to confirm instrument failure.
If confirmed replace instrument.
If CSCC tests are successfully, have customer run controls Replace the instrument
1 Replace Optical chas-
sis
2 Replace Main PCB
NEXT
Page 87
DIAGNOSTICS TROUBLESHOOTING
39 OF 60
Symptom
Meaning Possible
or Error
E67 Insufficie
nt sample
Cause
Poor immune flow
Instrument failure
Customer Remedy Service Remedy
1 Review the direc-
tions for applying the sample to the cassette.
2 Repeat the test
ensuring that the sample is correctly dispensed into the cassette and/or there is enough sample in the pipette before dispensing into the cassette.
Have the customer use the CSCC test to confirm instrument failure.
If confirmed, replace instrument.
1 Replace Optical chas-
sis
2 Replace Main PCB
If CSCC tests are successfully, have customer run controls
Replace the instrument
NEXT
Page 88
DIAGNOSTICS TROUBLESHOOTING
40 OF 60
Symptom
Meaning Possible
or Error
E68 Insufficie
nt sample
Cause
Poor immune flow
Instrument failure
Customer Remedy Service Remedy
1 Review the direc-
tions for applying the sample to the cassette.
2 Repeat the test
ensuring that the sample is correctly dispensed into the cassette and/or there is enough sample in the pipette before dispensing into the cassette.
Have the customer use the CSCC test to confirm instrument failure.
If confirmed replace instrument.
1 Replace Optical chas-
sis
2 Replace Main PCB
If CSCC tests are successfully, have customer run controls
Replace the instrument
NEXT
Page 89
DIAGNOSTICS TROUBLESHOOTING
41 OF 60
Symptom or Error
E90
E91
E92
E93
Meaning Possible
Cause
Not Used
Customer Remedy Service Remedy
NEXT
Page 90
DIAGNOSTICS TROUBLESHOOTING
42 OF 60

7.3 Instrument Diagnostic Screens

The following section is provided to assist in troubleshooting the CLINITEK Status system. This section explains the different diagnostic screens available on the instrument.

Diagnostics Screens: Initial Screens

The following is done in order to access the instruments diagnostic features.
Figure 7-1: Ready for Test Screen
1 From the ready screen, select “Instrument Set Up” by pressing the screen
button.
2 If the customer has set a Password to access the instrument set up, then you
must use the default password, 3654, to gain access to the Settings Menu.
NEXT
Page 91
DIAGNOSTICS TROUBLESHOOTING
Press the “123” button on the “Enter Password” screen to get the number pad and enter 3654 then “Enter.”
43 OF 60
3 You then get the following menu options. Press the down arrow button to scroll
down until “Diagnostics” is highlighted.
4 Once highlighted, press the “Select” button to proceed to the next menu screen.
5 If you want to exit any screen without proceeding to any of the listed options,
simply press the “Done” button to return to the previous menu screen or the ready screen.
NEXT
Page 92
DIAGNOSTICS TROUBLESHOOTING
44 OF 60

Diagnostics Screens: Display

The display test is the first test on the menu. This test is designed to highlight all 48 areas of the touch screen/display through an alternating checker board pattern. If any area is not properly displayed then there may be a problem with the display cable, display, or main PCB.
1 Scroll down using the down arrow to highlight the desired menu option.
The above screen is displayed once “Display” was selected.
2 Select “Run Test.”
3 These two screens will alternate back and forth several times in 20 seconds.
Make sure all areas are displayed in order to pass the test. If any area is not
NEXT
Page 93
DIAGNOSTICS TROUBLESHOOTING
properly displayed then follow the troubleshooting for missing characters or strange characters in
When the instrument has finished the test, the following screen is displayed.
Troubleshooting Table” .
45 OF 60
4 Select “Done” to return to the diagnostics menu screen or “Repeat Test” if
needed.
NEXT
Page 94
DIAGNOSTICS TROUBLESHOOTING
46 OF 60

Diagnostics Screens: Touch Screen

The touch screen test is designed to test the touch screen portion of the display. All 48 quadrants are highlighted and numbered. You are instructed to physically touch all 48 areas. The touch screen will respond to the pressure of the touch by changing the number into a forward arrow. An area not responding can indicate a problem with either the touch screen cable, touch screen, or main PCB.
1 Scroll down using the down arrow to select “Touch Screen” and press “Select.”
NEXT
Page 95
DIAGNOSTICS TROUBLESHOOTING
47 OF 60
Once “Touch Screen” was selected this screen is displayed.
2 Press “Run Test.”
Each of the 48 touch areas are displayed with a number except the last one which is the enter/done key.
NEXT
Page 96
DIAGNOSTICS TROUBLESHOOTING
48 OF 60
3 Press each numbered area to change the number into an arrow icon by starting
at 1 and ending at 47.
4 Press the last button to proceed to final screen.
Either “Test Passed” will be displayed (if all areas successfully changed to arrows) or the “Test was not successful” screen will be shown. If the latter screen is displayed then follow the troubleshooting guide for “Touch screen does not respond” in
Troubleshooting Table” .
NEXT
Page 97
DIAGNOSTICS TROUBLESHOOTING
49 OF 60

Diagnostics Screens: Printer

The printer test is to be used to test the printer by generating a test print. This test print is designed to utilize all areas of the thermal print head. If there are problems with the print head, voids in the printing would be seen or no printing at all. If the printer fails to respond to the test then there may be other problems like printer cables bad, printer bad, or main PCB bad.
1 Highlight “Printer” by scrolling
down using the down arrow and then pressing “Select.”
Once “Select” was press, this screen will be displayed.
2 Press “Run Test” to continue.
NEXT
Page 98
DIAGNOSTICS TROUBLESHOOTING
The “Please Wait” screen is displayed while the printer is printing.
The Final screen is displayed.
50 OF 60
3 Select “Done” to return to the diagnostic menu screen or repeat as needed.
Follow the troubleshooting guide for “Results are not printing” in
Table” if the printer fails to work properly.
Troubleshooting
NEXT
Page 99
DIAGNOSTICS TROUBLESHOOTING
51 OF 60

Diagnostics Screens: Test Table

The “Test Table” diagnostics feature allows for the user to test the table drive components along with the table position sensor. In this test the table gets drawn into the instrument table checks are performed along with a calibration check and then the table is moved back into the ready for test position. The test fails if there are any problems encountered during the checks or table movement. This test can also be used to help clear an E11 error (test table failure error) by running the diagnostics and retesting the sample.
1 Scroll down using the down
arrow to highlight “Test Table” and press “Select.”
This screen is displayed after Test Table” was selected.
2 Select “Run Test.”
NEXT
Page 100
DIAGNOSTICS TROUBLESHOOTING
The “Please wait” screen is displayed while the test table is being cycled.
Screen that is displayed when there is a successful test completed. If there were any problems, a failed test screen would be displayed.
52 OF 60
NEXT
Loading...