Badger Meter RVL Operating Manual

Vortex Flow Meters
Shedding Flow Meter
RVL Series
VRX-UM-00371-EN-03 (April 2017)
User Manual
Vortex Flow Meters, Shedding Flow Meter
Page ii March 2017VRX-UM-00371-EN-03
User Manual
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Operating Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Fluids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
General Installation Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Flow Rate and Range Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Piping Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Back Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
K-Factors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Three-Pin Connection Option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Mechanical Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
RVL Inline Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
RVL Wafer Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
RVL Tube Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Specications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
RVL Inline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
RVL Wafer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
RVL Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
RVL Inline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
RVL Wafer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
RVL Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Current Loop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Over-Stressed Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Calibration Certicate Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Page iii March 2017 VRX-UM-00371-EN-03
Vortex Flow Meters, Shedding Flow Meter
Page iv March 2017VRX-UM-00371-EN-03

Description

DESCRIPTION
The RVL series meter uses vortex-shedding technology for repeatable flow measurement accurate to ±1 percent of full scale. The meter has no moving parts, and any potential for fluid contamination is eliminated by the corrosion-resistant all plastic construction. The meter includes a compact two-wire (4…20 mA) or three-wire (pulse) transmitter, contained within a conveniently replaceable plug-in electronics module. All electronics are housed in a corrosion-resistant enclosure.
Unlike meters containing metal or moving parts, the RVL is perfect for aggressive or easily contaminated fluids. Applications range from ultra-pure water to highly corrosive chemicals and slurries. Units can be recalibrated and the meter output span can be reprogrammed in the field.

OPERATING PRINCIPLE

Operation of the RVL vortex flow meter is based on the vortex shedding principle. As fluid moves around a body, vortices (eddies) are formed and move downstream. They form alternately, from one side to the other, causing pressure fluctuations. The pressure fluctuations are sensed by a piezoelectric crystal in the sensor tube, and are converted to a 4…20 mA or pulse signal. The frequency of the vortices is directly proportional to the flow rate. The results are extremely accurate and repeatable measurements using no moving parts.
Counter
Blu Body
Detector
Detector
Figure 1: Operating principle

FLUIDS

Use any clean liquid compatible with the plastic material of construction that does not contain significant amounts of fibers or abrasive materials.
DO NOT USE WITH EXPLOSIVE OR FLAMMABLE MATERIALS, FOOD OR BEVERAGES, OR GASEOUS FLUIDS.
Viscosities above 1 cSt raise the minimum usable flow rate and reduce the flow range. This effect is linear to viscosity. No adjustments are required for viscosities up to 2.0 cSt. Liquids with higher viscosities adversely affect the permissible amount and duration of over range flow. See Table 1.
Viscosity Minimum Maximum Flow Range
1 cSt 1 12 12:1
2 cSt 2 12 6:1
3 cSt 3 12 4:1
4 cSt 4 12 3:1
5 cSt 5 12 2.4:1
6 cSt 6 12 2:1
Table 1: Viscosity and flow range
Page 5 April 2017 VRX-UM-00371-EN-03

General Installation Information

GENERAL INSTALLATION INFORMATION
Before installing the meter:
• Find an area for installation away from large electrical motors, transformers or other devices that can produce high electromagnetic or electrostatic fields. The vortex transmitter contains electric circuitry that can be affected by these interferences.
• Proper grounding is required to eliminate electrical noise which may be present within the fluid and piping system or in the near vicinity of the vortex transmitter. Use exterior grounding strap for non-conductive piping systems to provide a path to earth ground. Properly ground pipes in conductive piping systems.

Flow Rate and Range Requirements

Most manufacturers state flow range capabilities by publishing the maximum allowed flow rates. Then they provide a turndown ratio to determine minimum flow rate. To use the turndown ratio, simply divide the maximum rate by the ratio to determine the minimum rate. Vortex flow meters have a 12:1 turndown ratio at a viscosity of 1 cSt. Higher viscosities will reduce the turndown.
OTE:N The 1/4 in. NPT and 1/2 in. flare end meters have a standard turndown ratio of 8:1.

Piping Requirements

Turbulence in the pipeline can affect the accuracy of flow meters. Typical sources of turbulence are pumps, valves, change in pipe diameter or changes-in-direction in the line. Install the meter away from the turbulence source to avoid turbulence issues. These distances are indicated in Pipe Diameters (PD). For example, 10 PD is ten times the inside pipe diameter away from the source of turbulence. Follow upstream and downstream distances for all sources of turbulence. See Figure 2 on page
7, Figure 3 on page 8, Figure 4 on page 9 and Figure 5 on page 9 for proper piping distance requirements.
OTE:N Pulsating flow affects accuracy. Pressure pulses affect accuracy.
Page 6 April 2017VRX-UM-00371-EN-03
General Installation Information
Piping Requirements
Configuration
(pipe diameters)
Inlet Outlet
1 plane change
1 plane change w/outlet valve 10
2 plane changes
2 plane changes w/outlet valve 10
Two Plane
Changes
20
27
20 Dia
Minimum
Flow
20 Dia
Minimum
Flow
27 Dia
Minimum
Flow
5
5
5 Dia
Minimum
10 Dia
Minimum
5 Dia
Minimum
Accuracy
(full scale)
Repeatability
(of point)
±1.00% 0.25%
Two Plane
Changes
27 Dia
Minimum
Flow
Figure 2: Horizontal flow with sensing element in vertical orientation
10 Dia
Minimum
Page 7 April 2017 VRX-UM-00371-EN-03
General Installation Information
Configuration
1 plane change
1 plane change w/outlet valve 10 PD
2 plane changes
2 plane changes w/outlet valve 10 PD
Two Plane
Changes
Piping Requirements
Inlet Outlet
20 PD
27 PD
20 Dia
Minimum
Flow
20 Dia
Minimum
Flow
27 Dia
Minimum
Flow
5 PD
5 PD
5 Dia
Minimum
10 Dia
Minimum
5 Dia
Minimum
Accuracy
(full scale)
Repeatability
(of point)
±1.50% 0.25%
Two Plane
Changes
27 Dia
Minimum
Flow
Figure 3: Horizontal flow with sensing element in horizontal position
10 Dia
Minimum
Page 8 April 2017VRX-UM-00371-EN-03
General Installation Information
Configuration
1 plane change
1 plane change w/outlet valve 10 PD
2 plane changes
2 plane changes w/outlet valve 10 PD
Two Plane
Changes
Flow
27 Dia
Minimum
10 Dia
Minimum
Piping Requirements
Inlet Outlet
20 PD
27 PD
Two Plane
Changes
Flow
Minimum
Minimum
27 Dia
5 Dia
5 PD
5 PD
Flow
20 Dia
Minimum
10 Dia
Minimum
Accuracy
(full scale)
Repeatability
(of point)
±1.00% 0.25%
Flow
20 Dia
Minimum
5 Dia
Minimum
Figure 4: Vertical flow with a change in direction or valve
25 Dia
Minimum
Minimum
Flow
5 Dia
Minimum
20 Dia
Minimum
25 Dia
Minimum
Minimum
5 Dia
Minimum
Flow
Changes
20 Dia
Minimum
30 Dia
Minimum
Minimum
5 Dia
Minimum
5 Dia
Minimum
Two Plane
Flow
5 Dia
Minimum
25 Dia
Minimum
Minimum
Figure 5: Horizontal flow with a change in pipe diameter
10 Dia
2 Dia
2 Dia
2 Dia
Page 9 April 2017 VRX-UM-00371-EN-03
General Installation Information
1

Back Pressure

Back pressure, the pressure immediately downstream of the meter, must be maintained above a minimum level to avoid cavitation. For most applications this may be ignored if the flow rate is less than 75% of maximum. For other applications, use the following formula to calculate the minimum back pressure.
Back Pressure = 2.75P + 1.25 PV - 14.7
Where:
P = Pressure drop in psi at max flow
PV = Vapor pressure in psia of the liquid at operating temp.
(For example, the PV of water at 100° F is 0.42.)
BP = Back pressure (downstream of meter) in psig.
Example
For water, at 100° F (37° C) in a 1/2 in. (12.7 mm) meter, where the maximum pressure drop is 8 psi minimum back pressure is
7.8 psig.
BP = (2.75 × 8) + (1.25 × 0.42) - 14.7
BP = 22 + 0.525 - 14.7
BP = 7.825

Outputs

The RVL series meters can be ordered with a current output or a rate frequency output. The current output can be re-scaled in the field using a PC communications cable and programming software, which are both available as PN RVS220-954.
OTE:N The two outputs use unique circuit boards and cannot be changed in the field.
The rate frequency output produces pulses whose frequency is proportional to the flow going through the meter. Each meter has a slightly different output frequency which is listed on the calibration sheet that accompanies the meter. See Table 2 for the long term average full scale output frequency for standard size meters.
Meter Size Average Full Scale Frequency Pulse Width
1/4 in. (6.35 mm) 1055 Hz 0.47 msec
1/2 in. (12.7 mm) 820 Hz 0.61 msec
1/2 in. (12.7 mm) 570 Hz 0.88 msec
3/4 in. (19.05 mm) 284 Hz 1.76 msec
1 in. (25.4 mm) 292 Hz 1.71 msec
1-1/2 in. (38.1 mm) 144 Hz 3.47 msec
2 in. (50.8 mm) 148 Hz 3.38 msec
3 in. (76.2 mm) 61 Hz 8.20 msec
Table 2: Full scale output frequency
The frequency output option generates a square wave with an amplitude that matches the input power level. The pulse width varies with frequency and is found by using the following formula.
PW in sec. =
2 x Ma ximum Frequency (Hz)

K-Factors

The K-factor is the number of pulses that must be accumulated to equal a particular volume of fluid. Think of each pulse as representing a small fraction of the totalizing unit.
Calibration reports that accompany RVL series meters include a nominal K-factor in both gallons and liters. See “Calibration
Certificate Sample” on page 23.
Page 10 April 2017VRX-UM-00371-EN-03

Electrical Installation

ELECTRICAL INSTALLATION

Power

Use the following guidelines when selecting a power source:
• Use an 8…28V DC power supply. The specific connection depends on which output is option is used.
• Use clean electrical line power.
• Do not operate this unit on circuits with noisy components such as fluorescent lights, relays, compressors or variable frequency drives.
• Use linear power supplies.
OTE:N The power and output connections share a common ground.

Wiring

4…20 mA Loop
Connect a twisted pair wire (not provided) to the terminals of the transmitter marked 8…28V DC and Output. Do not connect the shield to the transmitter if the twisted pair wire is shielded. The shield should be grounded at the receiver only. See Figure 6. The transmitter is reverse-polarity protected.
+8-28 VDC
Programming
+8-28 VDC
Output
Output
Gnd
Gnd
4…20 mA
RECEIVER
POWER SUPPLY
8…28V DC
8…28V DC
Output
Load
Figure 6: Loop connection with single load
The receiving equipment must accept industry standard true two-wire or loop powered 4…20 mA process transmitter inputs. The power can either be supplied by the receiving equipment or an external power supply that supplies 24V DC an 30 mA. See Figure 6 for the wiring setup using an external power source and Figure 7 using the receiver as the power source. Several receivers may be connected in a series as shown in Figure 7, but only one should provide power, and all should have isolated inputs.
+8-28 VDC
Programming
+8-28 VDC
Output
Output
Gnd
Gnd
4...20 mA
RECEIVERRECEIVER
Additional Loads
RECEIVER/POWER SUPPLY
8…28V DC
Output
Figure 7: Loop connection with multiple loads
Page 11 April 2017 VRX-UM-00371-EN-03
Electrical Installation
Loop Load (Ohm's)
Loop Load (Ohm's)
The voltage provided by the receiver must be within the limits shown in Figure 8.
1100
1000
900
800
700
600
500
400
300
200
100
Supply Voltage – 8V DC
0.02
= Maximum Loop Resistance
Operate in the
Shaded Region
10
12814 16 18 20 22 24 26 28
Supply Voltage (V DC)
Figure 8: Supply voltage chart
To use this figure:
1. Add the resistance of all the receivers, indicators and the wire in the loop. If the wire resistance is unknown, use a value of 50 ohm for a twisted wire of 1000 feet or less with a gauge of #22 awg or heavier.
2. Find the total load (in ohms) on the left side of the chart in Figure 8 and follow that value horizontally until it intersects with the shaded area.
3. From the intersection point look straight down to where a vertical line would intersect the voltage scale. This is the minimum voltage needed for the transmitter to operate properly under the specic load conditions.
Example
After checking the specification for all the loads in an application the total amounted to 800 ohms. Following the 800 ohm line to the right, the intersection point is about 3/4 of the way across the chart in Figure 9.
A vertical line through the intersection point crosses the voltage axis at about 24V DC, so with a load of 800 ohms a standard 24 volt power supply would be used.
1100
1000
900
800
700
600
500
400
300
200
100
Supply Voltage - 8V DC
0.02
= Maximum Loop Resistance
Operate in the
Shaded Region
10
12814 16 18 20 22 24 26 28
Supply Voltage (V DC)
Figure 9: Supply voltage example
Pulse Output
Counter
Pulse
Output
8…28V DC
POWER
SUPPLY
Programming
+8-28 VDC
Output
Gnd
8…28V DC Output Ground
Figure 10: Pulse output wiring
Page 12 April 2017VRX-UM-00371-EN-03
Electrical Installation
White

Three-Pin Connection Option

An optional three-pin connection is available for when the transmitter/meter combination is mounted remotely from the power source/receiver. The mating connector is PN RF8687000.
4…20 mA Input
8…28V DC
Output
4…20 mA Input
– 4…20 mA Input
+8-28 VDC
Output
Gnd
Meter
Electronics
P.N. RF8687000 Connector
Figure 11: Remote connection loop power
ON
1 2 3 4 5 6 7 8
2
CW
1 3
SPAN ZERO
– 4…20 mA Input
- +
8…28V DC
Power Supply
mA
Ammeter
White 4…20 mA Input
Black – 4…20 mA Input
Black
- +
Figure 12: Integral configuration for rate indicator
Page 13 April 2017 VRX-UM-00371-EN-03

Mechanical Installation

MECHANICAL INSTALLATION

RVL Inline Installation

For proper installation, follow these guidelines:
• Install the meter where pipe vibration is minimal.
• Use the upstream and downstream piping requirements shown in “Piping Requirements” on page 6.
• Do not use upstream valves to control flow rate. Always keep upstream valves fully open.
• Connect good quality ball valves with integral unions directly to the flow meter if the valves are fully open during operation for easy isolation and removal of the flow meter. Cavitation and flow rate pulsation adversely affects the flow meter performance.
• Do not use diaphragm or piston pumps.
• Do not use Teflon tape or any kind of pipe dope when piping.
• Handle the meter with care.
• Do not use excessive force. Screw mating fittings (FNPT) and flanges into the meter hand-tight; then tighten an additional 1/2…3/4 turn.
• Always use two wrenches when turning the flow meter into a fitting; one across the flats on the flow meter end, close to the fitting, and one on the fitting.
• Do not use tools inside the flow meter, as this may damage the vortex sensor, and void the warranty.
The flow meter may be mounted in any orientation. Three holes, tapped 1/4-20 UNC-2B, 0.375 in.-deep, on 3/4 in. centers are provided on the 3/4 in. and smaller flow meters. Use these holes to provide support for the flow meter if pipe supports are not practical.
Page 14 April 2017VRX-UM-00371-EN-03
Mechanical Installation

RVL Wafer Installation

The RVL Wafer series transmitters are designed with wafer style flow bodies, that mount easily between standard ANSI style pipe flanges.
For proper installation, follow these guidelines:
• Install the meter where pipe vibration is minimal.
• Use the upstream and downstream piping requirements shown in “Piping Requirements” on page 6.
• Do not use upstream valves to control flow rate. Always keep upstream valves fully open.
• Connect good quality ball valves with integral unions directly to the flow meter if the valves are fully open during operation for easy isolation and removal of the flow meter. Cavitation and flow rate pulsation adversely affects the flow meter performance.
• Do not use diaphragm or piston pumps.
• Do not use Teflon tape or any kind of pipe dope when piping.
• Do not allow gaskets to protrude into the flow stream on flanged meters.
Flange Size Recommended Torque
1/2…1-1/2 in. 10…15 ft Ibs
2…3 in. 20…30 ft Ibs
Table 3: Torque rating
Follow these steps for proper installation and operation:
1. Space anges to accommodate the width of the ow body. See “RVL Wafer” on page 21 for dimensions.
2. Align the ow body centered with respect to anges and gaskets, insert threaded rods, retaining nuts and lock washers.
3. Install all retaining nuts hand-tight, and then uniformly tighten the nuts in an alternating sequence, diametrically opposed to each other. Uniform stress across the ange prevents leakage at the gasket. Torque ratings are listed in Table 3.
4. Use grounding rings when metal pipes are used in conjunction with this meter. See Figure 13.
Grounding
Rings
Figure 13: Grounding ring installation
Page 15 April 2017 VRX-UM-00371-EN-03

Maintenance

CAUTION

RVL Tube Installation

For proper installation, follow these guidelines:
• Install the meter where pipe vibration is minimal.
• Use the upstream and downstream piping requirements shown in “Piping Requirements” on page 6.
• Do not use upstream valves to control flow rate. Always keep upstream valves fully open.
• Connect good quality ball valves with integral unions directly to the flow meter if the valves are fully open during operation for easy isolation and removal of the flow meter. Cavitation and flow rate pulsation adversely affects the flow meter performance.
• Do not use diaphragm or piston pumps.
• Do not use Teflon tape or any kind of pipe dope when piping.
• Handle the meter with care.
To install the meter:
1. Remove any burrs from the pipe ends.
2. Slide the are nut onto the pipe.
3. Push the are nut back far enough so that it will be out of the way when you use the aring tool.
4. Clip the pipe in the aring tool, keeping the end ush with the face of the tool.
5. Slowly turn the handle on the tool until it bottoms out.
6. Unscrew the handle and remove the tool to check the quality of the are.
a. If the flare is not smooth or even the first time, cut off the end with your pipe cutter, and repeat steps 4…6.
7. Line up and tighten the nut and ared pipe to the tting body. Make the connection tight, but not so tight that the ow meter body is distorted.
• Always use two wrenches when turning a fitting onto the flow meter; one across the flats on the flow meter end close to the fitting, and one on the fitting.
• Do not use tools inside the flow meter, as this may damage the vortex sensor, and invalidate the warranty.
MAINTENANCE
RVL flow meters do not require maintenance in normal service if they are properly installed. Remove the meter from service for cleaning if the flow tube becomes clogged with debris. Significant clogging often results in high (up to 20%) and/or erratic output. Do not stick tools into the tube, as this may permanently damage the vortex sensor. The vortex sensor cannot be repaired in the field. To clean the flow tube, run hot, up to 160° F (71.1° C), soapy water into the downstream end of the flow tube. Dislodge large objects jammed against the bluff body by lightly tapping the upstream end of the flow tube against a firm surface.
DO NOT REMOVE VORTEX METER DURING OPERATION. ALWAYS DISCONNECT THE PRIMARY POWER SOURCE BEFORE INSPECTION OF SERVICE. DO NOT TAP THE FLOW TUBE SO HARD THAT THE THREADS, ON THREADED UNITS, BECOME DAMAGED.
A schedule of maintenance checks should be determined based upon environmental conditions and frequency of use. Inspect the meter at least once a year.
• Visually check for evidence of overheating by noting discoloration of wires or other components.
• Check for damaged or worn parts, especially the bluff body, or indications of corrosion.
• Check for tight, clean electrical connections and that the device is operating properly.
Page 16 April 2017VRX-UM-00371-EN-03

SPECIFICATIONS

PRESSURE DROP (PSID)
PRESSURE DROP (MILLIBAR)

RVL Inline

Fluid Liquids
Connection NPT Female or Butt (PVDF only)
Turndown Ratio
Accuracy
Repeatability ±0.25% of actual flow
Materials
Output Signals
Power Supply 8…28V DC
Response Time 2 seconds minimum, step-change-in flow
Enclosure Type 4X (IP 66)
12:1 for 1/2…2 in. (12.7…50.8 mm) meters
8:1 for 1/4 in. (6.35 mm) meter
±1% of full scale (4…20 mA)
±2% of full scale, frequency pulse
PVC standard
CPVC, PVDF optional
4…20 mA standard
Frequency pulse optional push-pull driver 150 mA sink or source
Specications
Nominal Flow Rates
Tube Size Minimum Flow Maximum Flow Full Scale Frequency Weight
1/4 in. (6.35 mm) 0.6 gpm (2.3 lpm) 5 gpm (18.9 lpm) 1052 Hz 1.5 lbs (0.68 kg)
1/2 in. (12.7 mm) 1.3 gpm (4.7 lpm) 15 gpm (56.8 lpm) 570 Hz 1.6 lbs (0.72 kg)
3/4 in. (19.05 mm) 2.1 gpm (7.9 lpm) 25 gpm (94.6 lpm) 284 Hz 1.7 lbs (0.77 kg)
1 in. (25.4 mm) 4.2 gpm (15.8 lpm) 50 gpm (189.3 lpm) 292 Hz 1.8 lbs (0.80 kg)
1-1/2 in. (38.1 mm) 8.3 gpm (31.5 lpm) 100 gpm (378.5 lpm) 144 Hz 3.1 lbs (1.40 kg)
2 in. (50.8 mm) 16.7 gpm (63.1 lpm) 200 gpm (757.1 lpm) 142 Hz 2.7 lbs (1.22 kg)
Maximum Fluid
Temperature
PVC CPVC PVDF
Maximum Operating Pressure psig (KPa)
203° F (95° C) Not recommended Consult factory Consult factory
150° F (66° C) Not recommended 63 psig (434 KPa) 130 psig (896 KPa)
100° F (38° C) 93 psig (641 KPa) 120 psig (827 KPa) 150 psig (1034 KPa)
70° F (21° C) 150 psig (1034 KPa) 150 psig (1034 KPa) 150 psig (1034 KPa)
20
12
10
8 5
2
1
.5
.2
.1
.05
¼ in.
.5 2 10 25 100.3 1 5 15 50 200
FLOW (GPM)
½ in.
¾ in.
1 in.
1½ in.
2 in.
Figure 14: RVL inline pressure drop
1000
750 500 350
200
100
1 in.
¼ in.
½ in.
1½ in.
¾ in.
50 35
20
10
5
3.5
30 100 300 800
2 in.
6002 5 10 20 50 200
FLOW (LPM)
Page 17 April 2017 VRX-UM-00371-EN-03
Specications
PRESSURE DROP (PSID)
PRESSURE DROP (MILLIBAR)

RVL Wafer

Fluid Liquids
Connection Wafer
Turndown Ratio 12:1
Accuracy
Repeatability ±0.25% of actual flow
Materials
Output Signals
Power Supply 8…28V DC
Response Time 2 seconds minimum, step-change-in flow
Enclosure Type 4X (IP 66)
Tube Size Minimum Flow Maximum Flow Full Scale Frequency Weight
1/2 in. (12.7 mm) 1.3 gpm (4.7 lpm) 15 gpm (56.8 lpm) 570 Hz 0.8 lbs (0.36 kg)
3/4 in.(19.05 mm) 2.1 gpm (7.9 lpm) 25 gpm (94.6 lpm) 284 Hz 0.9 lbs (0.41 kg)
1 in. (25.4 mm) 4.2 gpm (15.8 lpm) 50 gpm (189.3 lpm) 292 Hz 1.1 lbs (0.50 kg)
1-1/2 in. (38.1 mm) 8.3 gpm (31.5 lpm) 100 gpm (378.5 lpm) 144 Hz 1.7 lbs (0.77 kg)
2 in. (50.8 mm) 16.7 gpm (63.1 lpm) 200 gpm (757.1 lpm) 148 Hz 2.6 lbs (1.17 kg) 3 in. (76.2 mm) 25.0 gpm (94.6 lpm) 300 gpm (1136 lpm) 61 Hz 4.8 lbs (2.16 kg)
±1% of full scale (4…20 mA)
±2% of full scale, frequency pulse
PVC standard
CPVC, Polypropylene, PVDF optional
4…20 mA standard
Frequency pulse optional push-pull driver 150 mA sink or source
Nominal Flow Rates
Maximum Fluid
Temperature
PVC CPVC Polypropylene PVDF
Maximum Operating Pressure, Standard
203° F (95° C) Not recommended Consult factory Not recommended Consult factory 150° F (66° C) Not recommended 63 psig (434 KPa) 90 psig (621 KPa) 130 psig (896 KPa) 100° F (38° C) 100 psig (690 KPa) 120 psig (827 KPa) 130 psig (896 KPa) 150 psig (1034 KPa)
70° F (21° C) 150 psig (1034 KPa) 150 psig (1034 KPa) 150 psig (1034 KPa) 150 psig (1034 KPa)
Maximum Fluid
Temperature
PVC CPVC Polypropylene PVDF
Maximum Operating Pressure, High Pressure
203° F (95° C) Not recommended Not recommended Not recommended Consult factory 150° F (66° C) Consult factory Consult factory 90 psig (621 KPa) 300 psig (2068 KPa) 100° F (38° C) Consult factory Consult factory 130 psig (896 KPa) 400 psig (2750 KPa)
70° F (21° C) Consult factory Consult factory 150 psig (1034 KPa) 400 psig (2750 KPa)
20
12
10
8 5
2
1
.5
.2
.1
.05
.5 2 10 25 100 300.3 1 5 15 50 200
FLOW (GPM)
½ in.
¾ in.
1 in.
2 in.
1½ in.
3 in.
Figure 15: RVL wafer pressure drop
1000
750 500 350
200
100
3/4 in.
2 in.
1 in.
1½ in.
1/2 in.
50 35
20
10
5
3.5
30 100 300 800
3 in.
6002 5 10 20 50 200
1200
FLOW (LPM)
Page 18 April 2017VRX-UM-00371-EN-03

RVL Tube

PRESSURE DROP (PSID)
PRESSURE DROP (MILLIBAR)
Fluid Liquids
Connection Tube (Flare end)
Turndown Ratio
Accuracy
Repeatability ±0.25% of actual flow
Materials
Output Signals
Power Supply 8…28V DC
Response Time 2 seconds minimum, step-change-in flow.
Enclosure Type 4X (IP 66)
Tube Size Minimum Flow Maximum Flow Weight
1/2 in. (12.7 mm) 0.6 gpm (2.3 lpm) 5 gpm (18.9 lpm) 1.5 lbs (0.68 kg)
3/4 in. (19.05 mm) 1.3 gpm (4.7 lpm) 15 gpm (56.8 lpm) 1.6 lbs (0.72 kg)
1 in. (25.4 mm) 2.1 gpm (7.9 lpm) 25 gpm (94.6 lpm) 1.7 lbs (0.77 kg)
12:1 for 3/4 in. (19.05 mm) and 1 in. (25.4 mm) meters
8:1 for 1/2 in. (12.7 mm) meter
±1% of full scale (4…20 mA)
±2% of full scale, frequency pulse
PVC standard
CPVC, Polypropylene, PVDF optional
4…20 mA standard
Frequency pulse optional push-pull driver 150 mA sink or source
Nominal Flow Rates
Specications
Maximum Fluid Temperature
150° F(66° C) 130 psig (896 KPa)
100° F (38° C) 150 psig (1034 KPa)
70° F (21° C) 150 psig (1034 KPa)
20
12
10
8 5
2
1
.5
.2
.1
.05
½ in.
1 in.
¾ in.
.5 2 10 25 100.3 1 5 15 50 200
FLOW (GPM)
Figure 16: RVL tube pressure drop
Maximum Operating Pressure
PVDF
FLOW (LPM)
1000
750 500 350
200
100
50 35
20
10
5
3.5
½ in.
¾ in.
30 100 300 800
FLOW (LPM)
1 in.
6002 5 10 20 50 200
Page 19 April 2017 VRX-UM-00371-EN-03

Dimensions

DIMENSIONS

RVL Inline

Cord Grip
F
I
NPT/BUTT
END
C
A
B
E D
Figure 17: RVL inline dimensions
Cover
Cord Grip
Conduit Adapter
Terminal Strip
Electronics Module
Three-Pin Connector
Flow Sensor Body
PVC/CPVC
Size
A
in. (mm)
B
in. (mm)
C
in. (mm)
D
in. (mm)
E
in. (mm)
F
in. (mm)
I
in. (mm)
1/4 in. (6.35 mm) 3.81 (97) 1.75 (45) 5.25 (133) 2.50 (64) 0.30 (8) 2.88 (73) 3.00 (76)
1/2 in. (12.7 mm) 3.81 (97) 1.75 (45) 7.13 (181) 2.50 (64) 0.55 (14) 2.88 (73) 3.00 (76)
3/4 in. (19.05 mm) 3.81 (97) 1.75 (45) 7.63 (194) 2.50 (64) 0.74 (19) 2.88 (73) 3.00 (76)
1 in. (25.4 mm) 3.92 (100) 1.75 (45) 8.03 (204) 2.50 (64) 0.96 (24) 2.88 (73) 3.00 (76)
1-1/2 in. (38.1 mm) 3.90 (99) 2.00 (51) 8.37 (213) 2.50 (64) 1.50 (38) 2.88 (73) 3.38 (86)
2 in. (50.8 mm) 4.31 (109) 2.00 (51) 8.37 (213) 2.50 (64) 1.94 (49) 2.88 (73) 3.38 (86)
PVDF (BUTT Fusion Only)
Size
A
in. (mm)
B
in. (mm)
C
in. (mm)
D
in. (mm)
E
in. (mm)
F
in. (mm)
I
in. (mm)
1/4 in. (6.35 mm) 5.90 (150) 0.63 (16) 4.87 (124) 1.31 (33) 0.30 (8) 2.88 (73) 3.00 (76)
1/2 in. (12.7 mm) 5.75 (146) 0.78 (20) 4.87 (124) 1.31 (33) 0.55 (14) 2.88 (73) 3.00 (76)
3/4 in. (19.05 mm) 5.75 (146) 0.94 (24) 4.87 (124) 1.44 (37) 0.74 (19) 2.88 (73) 3.00 (76)
1 in. (25.4 mm) 5.88 (149) 1.19 (30) 5.09 (129) 2.00 (51) 0.96 (24) 2.88 (73) 3.00 (76)
1-1/2 in. (38.1 mm) 6.21 (158) 1.50 (38) 6.24 (158) 2.50 (64) 1.50 (38) 2.88 (73) 3.38 (86)
2 in. (50.8 mm) 6.60 (168) 1.88 (48) 6.77 (172) 3.00 (76) 1.94 (49) 2.88 (73) 3.38 (86)
Page 20 April 2017VRX-UM-00371-EN-03

RVL Wafer

Cord Grip
Dimensions
Cord Grip
E
Cover
Conduit Adapter
Terminal Strip
Electronics Module
A
B
Three-Pin Connector
Flow Sensor Body
C
D
Figure 18: RVL wafer dimensions
RVL (Wafer) Dimensions PP/PVC/CPVC/PVDF
Size
A
in. (mm)
B
in. (mm)
C
in. (mm)
D
in. (mm)
E
in. (mm)
1/2 in. (12.7 mm) 5.85 (149) 0.78 (20) 2.03 (52) 1.75 (45) 2.88 (73)
3/4 in. (19.05 mm) 5.90(150) 0.94 (24) 2.03 (52) 1.75 (45) 2.88 (73)
1 in. (25.4 mm) 5.69 (145) 1.19 (30) 2.25 (57) 1.75 (45) 2.88 (73)
1-1/2 in. (38.1 mm) 6.00 (152) 1.50 (38) 2.63 (67) 1.75 (45) 2.88 (73)
2 in. (50.8 mm) 6.37 (162) 1.88 (48) 3.22 (82) 1.75 (45) 2.88 (73)
3 in. (76.2 mm) 6.88 (175) 2.50 (64) 4.25 (108) 1.75 (45) 2.88 (73)

RVL Tube

Cord Grip
Cover
B
Conduit Adapter
Terminal Strip Electronics Module Three-Pin Connector
Flow Sensor Body
C
Tube Size
A
A
in. (mm)
Figure 19: RVL tube dimensions
B)
in. (mm)
1/2 in. (12.7 mm) 1.31 (33.3) 6.25 (158.8) 4.87 (123.7)
3/4 in. (19.05 mm) 1.31 (33.3) 6.25 (158.8) 4.66 (118.4)
1 in. (25.4 mm) 1.44 (36.6) 6.59 (167.4) 5.42 (137.7)
C
in. (mm)
Page 21 April 2017 VRX-UM-00371-EN-03

Troubleshooting

TROUBLESHOOTING
If difficulty is encountered, locate the symptom most likely present and follow the appropriate instructions.

Current Loop

No Current Output
• Place a DC voltmeter across the two terminal block screws. With the electronics module powered there must be at least 8V DC present. If there is less than 8V DC, but more than 0V DC, check the power source for sufficient voltage to drive the loop, as shown in Figure 8 on page 12.
◊ If there is 0V DC present, check for a broken wire or connector in the loop.
• Check for the proper polarity of the current loop connections.
• Make sure the receiving device is configured to provide source current to the electronics module.
Zero Flow Indication (4 mA in Loop)
• Check that the flow is greater than the minimum specified for the particular size flow meter in use.
◊ If the flow rate is too low, replace the flow meter with the proper size flow meter.
◊ If the flow rate is sufficient, partially remove the electronic module. Check that the three pin connector that connects
the electronics module to the flow transducers is positively connected. See Figure 20. Align and insert the connector on to the bottom of the electronics module if it is disconnected.
Terminal Strip
Electronics Module
Three-Pin Connector
Flow Sensor Body
Figure 20: Electrical connection
Erratic Flow Indication
• Check that there is at least 8V DC present across the two terminal block screws.
• Check for material clogging the flow meter and in the upstream piping.
• Check for erosion of the bluff body by sighting down the meters bore. Erosion or damage to the bluff body causes erratic readings and compromise accuracy. If the erosion continues, the flow meter will need to be periodically replaced.
• Check upstream piping distance. See “Piping Requirements” on page 6.
• Check for excessive pipe vibration. Normal amounts of pipe vibration are easily tolerated. The transmitter module contains a highly effective active filter that rejects false signals caused by pipe vibration. This filter is most effective under flowing conditions. If vibration is causing the meter to indicate flow when the flow is stopped it will most likely not cause error under flowing conditions. The false flow indication may be ignored, or the pipe may be restrained by firm clamps.
• Check for electrical noise. Under some conditions there can be high common mode AC noise present between the fluid and the power supply ground. The flow meter is designed to reject up to 50 volts of AC common mode noise without loss of accuracy. If noise adjustment is used, accuracy is effected at low flow rates. Place a ground strap on the pipe on both sides of the flow meter (the flow meter is made of non-conductive plastic) and connect them both to the one point where the loop is grounded if metal piping is used. See “Wiring” on page 11. Use a grounding orifice if plastic piping is used. The transmitter module contains a highly effective active filter that will reject false signals due to high common mode voltage. This filter is most effective under flowing conditions. If a false indication of flow is encountered at zero flow, it will probably not cause error under flowing conditions.

Over-Stressed Sensor

The sensor can be over-stressed if the maximum permitted flow rate of 125% of recommended capacity (100% of HT meters) is exceeded.
Page 22 April 2017VRX-UM-00371-EN-03

CALIBRATION CERTIFICATE SAMPLE

Calibration Report
Unit Under Test (UUT) Information: Master Meter:
Description: ¾ in. In-Line NPT End Flow Meter Std uncertainty: ±0.25% Model Number: RVL075-N 1 VNN Traceability No: 30400/31801 Serial Number: 99999 Model No: FT8-8N EXW-LEG-5/FT-16 NEXW-LEG-1 Sensor Type: Vortex Shedding Serial No: 806890/16011903 Output type: 0-5V Minimum Flow: 2.1 GPM 7.9 LPM Customer Information:
Maximum Flow: 25 GPM 94.6 LPM Customer Name: Calibration Date: October 24, 2007 Customer No.: Calibration Interval: 12 Months Order No.: Cal. Liquid: Water Ambient Temperature: 71.74 °F Ambient Humidity: 31.39 %RH Linear Points: 5
UUT Calibration Data Table In GPM:
Flow
Standard
Actual
GPM
1 25.00 100.000 72.00 0.949 105.406 240.00 24.57 1.0174 1.71 5.000 5.000 0.00
1 18.00 75.000 72.00 0.949 79.055 250.00 18.43 0.9767 -1.71 3.600 3.680 0.40
1 12.00 50.000 72.00 0.949 52.703 250.00 12.29 0.9767 -1.14 2.400 2.420 0.10
1 6.00 25.000 72.00 0.949 26.352 250.00 6.14 0.9767 -0.57 1.200 1.200 0.00
1 2.10 10.000 72.00 0.949 10.541 285.71 2.46 0.8547 -1.43 0.420 0.420 0.00
UUT
Hz
UUT
Temp °F
Visc.
cSt
UUT F/V
Hz/cSt
UUT K
CYC/GAL
(Hz*60)/NK
GPM
Linear
COEFF.
Raw Err
% FS
Calibration Certicate Sample
Calc. 0-5V
Meas.
0-5V
Output
Err % FS
Nominal K (NK) 244.186
UUT Calibration Data Table In LPM:
Flow
Standard
Status: PASS
Meter Accuracy (of FS): ± 0.4 %
Average Calib. Temperature : 72 F
Average Calib. Specic Gravity : 1 Calibrated By: Ramon Benedict
Average Calib. Viscosity : 0.95 cSt
Flow Direction : Forward Certied By: Larry Perez
Racine calibrations are performed using standards traceable to National Institute of Standards and Technology.
The equipment and calibration procedures comply with ISO 9001.
Actual
GPM
1 94.64 100.000 72.00 0.949 105.406 63.40 93.01 1.0174 1.71 5.000 5.000 0.00
1 68.14 75.000 72.00 0.949 79.055 66.04 69.76 0.9767 -1.71 3.600 3.680 0.40
1 45.42 50.000 72.00 0.949 52.703 66.04 46.51 0.9767 -1.14 2.400 2.420 0.10
1 22.71 25.000 72.00 0.949 26.352 66.04 23.25 0.9767 -0.57 1.200 1.200 0.00
1 7.95 10.000 72.00 0.949 10.541 75.48 9.30 0.8547 -1.43 0.420 0.420 0.00
UUT
Hz
UUT
Temp °F
Visc.
cSt
UUT F/V
Hz/cSt
UUT K
CYC/GAL
(Hz*60)/NK
GPM
Linear
COEFF.
Raw Err
% FS
Calc. 0-5V
Meas.
0-5V
Nominal K (NK) 64.507
Output
Err % FS
Page 23 April 2017 VRX-UM-00371-EN-03
Vortex Flow Meters, Shedding Flow Meter
Control. Manage. Optimize.
Trademarks appearing in this document are the property of their respective entities. Due to continuous research, product improvements and enhancements, Badger Meter reserves the right to change product or system specications without notice, except to the extent an outstanding contractual obligation exists. © 2017 Badger Meter, Inc. All rights reserved.
www.badgermeter.com
The Americas | Badger Meter | 4545 West Brown Deer Rd | PO Box 245036 | Milwaukee, WI 53224-9536 | 800-876-3837 | 414-355-0400 México | Badger Meter de las Americas, S.A. de C.V. | Pedro Luis Ogazón N°32 | Esq. Angelina N°24 | Colonia Guadalupe Inn | CP 01050 | México, DF | México | +52-55-5662-0882 Europe, Eastern Europe Branch Oce (for Poland, Latvia, Lithuania, Estonia, Ukraine, Belarus) | Badger Meter Europe | ul. Korfantego 6 | 44-193 Knurów | Poland | +48-32-236-8787 Europe, Middle East and Africa | Badger Meter Europa GmbH | Nurtinger Str 76 | 72639 Neuen | Germany | +49-7025-9208-0 Europe, Middle East Branch Oce | Badger Meter Europe | PO Box 341442 | Dubai Silicon Oasis, Head Quarter Building, Wing C, Oce #C209 | Dubai / UAE | +971-4-371 2503 Slovakia | Badger Meter Slovakia s.r.o. | Racianska 109/B | 831 02 Bratislava, Slovakia | +421-2-44 63 83 01 Asia Pacic | Badger Meter | 80 Marine Parade Rd | 21-06 Parkway Parade | Singapore 449269 | +65-63464836 China | Badger Meter | 7-1202 | 99 Hangzhong Road | Minhang District | Shanghai | China 201101 | +86-21-5763 5412 Switzerland | Badger Meter Swiss AG | Mittelholzerstrasse 8 | 3006 Bern | Switzerland | +41-31-932 01 11
Legacy Document: 09-VRX-UM-00362
Loading...