Badger Meter Register ILR 750 Programming Manual

0 (0)
®
Badger Meter Europa GmbH
Register ILR 750
PROGRAMMING MANUAL
AC_ILR750_PA_02_1810
Programming Page 1/4
press for 2-3 seconds
Press „RESET“ to
Press „RESET“ to
Predefined pulse rates
3“ = 3
Programmed pulse rate on figure: 52,40 ppl
Press „RESET“ to
Selection of meter type
Programming on figure: o (RCDL und Vision)
press for 2-3 seconds
press for 2-3 seconds
Programming the electronic ILR750 register
To reach menu item 1 in the programming mode, press the following key combination: 3x then 3x (Keys must be pressed quickly one behind the other!)
Note: You can leave the programming mode any time by pressing the keys „TOTAL“ + „RESET“ for 2-3 seconds at the same time. The set data will also be saved.

Menu 1: Programming the calibration factor and the unit of measure

change the flashing digit

Menu 2: Change meter pulse rate

Menu 3: Register alignment

Press „TOTAL“ to jump to the next digit
change the flashing digit Press „TOTAL“ to
jump to the next digit
change the flashing digit
RCDL: IOG: M25 = 52,40 ½“ = 100 M35 = 33,50 ¾“ = 66 M40 = 23,70 1“ = 66 M70 = 12,30 1 HF = 43 M120 = 6,30 1 ½“ = 17 M170 = 3,80 2“ = 9
or rather the alignment:
RCDL und Vision =
IOG: Depending on alignment
I = horizontal P = vertical
o
AC_ILR750_PA_02_1810
Programming Page 2/4
Press „RESET“ to
press for 2-3 seconds
Press „RESET“ to
Flow rate that should be equal to
the maximum flow rate (menu 6).
press for 2-3 seconds
Press „RESET“ to
Flow rate that should be equal to
the minimum flow rate (menu 5).
press for 2-3 seconds
Programmed max. flow rate: 30,0 l/min
Press „RESET“ to
press for 2-3 seconds

Menu 4: Change display mode

Menu 5: Analog minimum flow rate

Menu 6: Analog maximum flow ra te

Menu 7: Set output pulse duration

change the flashing digit
change the flashing digit Press „TOTAL“ to
jump to the next digit
change the flashing digit Press „TOTAL“ to
jump to the next digit
change the flashing digit
Programming line 1 on the display:
C = Totalizer (L, Gal, Qt, Pt) F = Flow indicator (e.g. L/min)
Note: A totalizer without decimal places is always in the second line.
the 4mA Output:
Minimum: 0,0 l/min Maximum: 999,8 l/min Default: 0,0 l/min
The value of the minimum flow rate must be smaller than the value of
the 20mA output:
Minimum: 0,1 l/min Maximum: 999,9 l/min Default: 30,0 l/min
The value of the maximum flow rate must be higher than the value of
0 = disabled 2 = 2msec 10 = 10msec 20 = 20msec 40 = 40msec 100 = 100msec
AC_ILR750_PA_02_1810