Bacharach Snifit + Owner's Manual

Instruction Manual
P/N: 0019-9371
Revision 0
Electrochemical
C)
2000 ppm)
(9.67 cm x 5.87 cm x 2 cm)
3.6 oz (105 g)
2+ years
resistant
December 2, 2013
Carbon Monoxide
Personal Safety
Monitor Kit
(P/N: 0019-7600)
Snifit+ CO Gas Monitor ................................. 0019-6108
Instruction Manual ........................................ 0019-9371
Calibration Adapter ....................................... 0019-6106
Calibration Kit (Regulator and 4’ Hose) ......... 0051-2978
100 ppm Calibration Gas Cylinder ................. 0051-1994
WARNING: The Snifit+ contains a lithium battery which my leak or explode if improperly used. Do not place the Snifit
IMPORTANT: Snifit+ has no user-serviceable parts inside. Do not disassemble the instrument.
IMPORTANT: Verify operation each day by per­forming a functional (bump) test.
IMPORTANT: Keep all openings free of dirt, debris, and foreign objects.
IMPORTANT: Do not use if damaged or improp­erly operating. Contact a service representative immediately.
IMPORTANT: Zero the Snifit+ only in a CO-free environment.
IMPORTANT: Dispose of the Snifit+ properly.
KIT CONTENTS
OPTIONAL ACCESSORIES
SAFETY INFORMATION
+ in fire or incinerate.
DESCRIPTION
The Snifit+ is a rugged personal gas monitor for detecting carbon monoxide gas. Preset alarms provide visual, audi­ble, and vibration alerts for low and high alarm conditions. Built-in time weighted average (TWA) alarms and short­term exposure limit (STEL) alarms provide an even greater margin of safety for the user. A custom display provides easy-to-view information including:
gas concentration
time remaining (months) until replacement is needed
battery low alert
calibration activation.
Sensor: Carbon Monoxide (CO);
Operating Temperature:
Humidity: 15 to 90% RH (Non-condensing)
Duty Cycle: Continuous
Display Resolution (Range):
Response: T90 < 30 seconds
Visual Alarm: LEDs and display (low battery,
Sound Alarm: 90dB @ 12” (30 cm)
Touch Alarm: Vibration
Power: Lithium battery
Size: 3.81” x 2.31” x 0.80”
Weight:
Battery Life:
Housing: IP65 environmental resistance;
Mounting: Durable belt clip and lanyard loop
SPECIFICATIONS
-4 to 122° F (-20 to 50°
1 ppm (0-999 ppm);
0.01% Volume (1000-
sensor life, calibration status)
Rubberized and impact-
FEATURES
Construction: The Snifit+ is constructed of durable ABS
plastic. The top housing is further protected with a shock absorbing material to protect the Snifit field use. The unit is sealed for an IP65 environmental resistance rating.
Operation: Two buttons operate the Snifit
button ( ) is for power. The right ( ) is for zeroing and calibrating. The Snifit
+ supports “continuous on” opera­tion. The CO sensor and battery are designed for 2 years of continuous normal use.
+ from the rigors of
+ . The left
The audible (sound), visual (LEDs), and vibrating alarms are sequenced during alarms. The alarm settings are defined during the set-up of the instrument. The instrument’s default alarm limits are factory-preset to the values shown below:
Extended Range Operation: All readings from 1000 ppm to 2000 ppm are displayed as “_10%” to “_19%”. The underscore represents a decimal point. Levels detected above “_19%” display “OL”.
FEATURES (CONTINUED)
LOW: 35 ppm TWA: 50 ppm HIGH: 200 ppm STEL: 75 ppm
Sensor: The sensor opening is on the front of the Snifit+ .
STARTUP
Press to start the Snifit+ in an area known to be CO gas free and with normal oxygen content. The following will display as a normal start up sequence.
1. All segments and icons will activate.
2. Gas type to be sensed will be displayed.
Mounting: The instrument may be attached using the metal lanyard loop or the sturdy suspender clip on the back of the instrument.
Alarm Components: The front of the instrument has an opening for the audible alarm (which is sealed to maintain the integrity of the IP65 rating). Multiple LEDs beneath the frosted area create a visual warning that is bright even in sunlight.
NOTE: The Snifit+ cannot be powered off or zeroed when in alarm mode.
Alarm Limits: The Snifit+ has two basic styles of alarm:
Low Alarms: Indicators activate at 2 sec intervals
High Alarms: Indicators activate at 1 sec intervals.
3. Software version is shown.
4. Temperature in °C is shown.
5. Alarms will sound, illuminate and vibrate.
6. “LO” Low alarm setpoint displays.
7. “HI” High alarm setpoint displays.
8. “tA” TWA alarm setpoint displays (if applicable).
9. “StL” STEL alarm setpoint displays (if applicable).
10. “Sn” Serial Number displays.
11. Displays remaining life, in months.
12. All segments flash for 6 seconds indicating zeroing.
13. The working display is shown.
If the Snifit+ is in an area that is not gas free, “BAD” will illuminate after the automatic zeroing followed by a read­ing based on factory zero set points.
BUMP TEST
To verify the operation of the instrument, a bump test should be done prior to each day’s use. Bump test failure may indicate a product failure or the need for calibration.
To perform a bump test, turn on the Snifit+ and wait for the working display to be shown. Apply test gas using the calibration adapter. Flow should be 200-400 cc/min. As­suming an alarm setpoint is less than the applied gas con­centration, an alarm should occur in less than 20 seconds.
trained personnel
Only factory trained personnel
“Pin”. Press either key to display 900.
The display will now show “LO” for the low alarm.
Press to adjust the alarm (roll-over after 300).
now display “HI”.
Press to adjust the alarm (roll-over after 300).
now display “tA” for TWA.
Press to adjust the alarm (roll-over after 125).
now display “StL” for STEL.
Press to adjust the alarm (roll-over after 300).
the menu and perform an auto zero automatically.
should be 200-400 cc/min.
2.
Press and hold until the display reads “GAS”.
3.
Attach the calibration adapter.
5.
Press . The CAL icon will flash.
over occurs after 50).
Press to accept the desired “MO”.
shows the current date.
“dd”.
AM/PM selection).
Adjust to desired “Hr”.
shows the current minute.
over after 59).
Adjust to desired “nn”.
OPERATION OVERVIEW
WARNING: Follow federal, state, local and com-
pany regulations as they relate to the use of this product.
PLACEMENT: Attach the Snifit+ to the outermost garment as close to your head/face as practical.
ALARMS: Alarms will indicate unsafe levels. The display will show concentration data in combination with audible, visual and vibration indications. TWA alarms can be acknowledged for 10-minute intervals by pressing either button.
MANUAL ZERO: Manual zeroing should only be done in an atmosphere known to be CO gas free and containing nor­mal oxygen levels. Press for 2-3 seconds and release. All segments of the display will flash followed by the working display.
OVER RANGE: Exposure to CO gas in excess of 2000 ppm will result in “OL” being displayed with the high alarm activated.
OPERATION TIME: To view the remaining operation time, press for 2-3 seconds and release. The Snifit
+ will dis-
play the time remaining in months, then perform a zero.
BATTERY DIAGNOSTICS: BAT indicates low battery output and service will be needed soon.
WARNING: Only factory should perform any internal servicing.
BAT + LO indicates that the Snifit+ should be immediately taken out of service.
SYSTEM DIAGNOSTICS: I2C indicates a communication error in the EEPROM. Factory service is required.
WARNING: should perform any internal servicing.
SHUTDOWN: To turn off, press and hold the POWER but­ton ( ) for approximately 6 seconds until “OFF” is dis­played, then release the button. The Snifit
+ cannot be
powered off (or zeroed) when in alarm mode.
P/N: 0019-9371 Rev 0
Copyright 2013 © Bacharach, Inc.
All Rights Reserved
ALARM ADJUSTMENT
1. When the Snifit+ starts, press both buttons together ( + ) after the Gas Type is displayed, then re­lease… OR From the working display, push both buttons simulta­neously ( + ) and release. The unit will display
2.
Press to scroll up to 980, then push . Roll-over occurs after 998 then restarts at 900. The display will show “Con” for Continuous On operation.
Press to adjust mode.
Press to toggle mode (Yes or No). Note: Con-
tinuous On mode (Yes) disables the Power But­ton ( ) and the TWA alarm.
Press to save the setting.
3.
Press to adjust the “LO” alarm. The current setting will be displayed.
4.
5.
Press to save the alarm setting. The Snifit+ will
6.
Press to adjust the “HI” alarm. The current setting will be displayed.
7.
8.
Press to save the alarm settings. The Snifit+ will
9.
Press to adjust the TWA alarm. The current setting will be displayed.
10.
11.
Press to save the alarm settings. The Snifit+ will
12.
Press to adjust the STEL alarm. The current setting will be displayed.
13.
14.
Press to save the alarm settings. Snifit+ will exit
CALIBRATION
Turn on the Snifit+ . The working display is shown. Allow the unit to remain on for 2 minutes prior to beginning the calibration process.
Prepare the 100 ppm CO gas cylinder and regulator
1.
with the calibration adapter attached. The flow
4. Start the gas flow.
CALIBRATION (CONTINUED)
6. Successful calibration is indicated when only the gas concentration is displayed. The CAL icon will no longer be illuminated. Bad calibration will be indi­cated by the word “BAD” and the CAL icon on the display. Press to acknowledge a bad calibration, then retry.
WARRANTY AND PRODUCT SUPPORT
Bacharach warrants to the buyer that the Snifit+ will be free from material defects, free from manufacturing de­fects, and will conform substantially to Bacharach’s appli­cable specifications at the time of delivery.
Bacharach’s liability and the buyer’s remedy under this warranty are limited to the repair or replacement, at Bacharach’s option, of this product or parts thereof re­turned to Bacharach and shown to Bacharach’s reasonable satisfaction to have been defective; provided that written notice of the defect is given by buyer to Bacharach within two (2) years after the date of delivery of this Product by Bacharach. Internal repair must be completed by a Bacha­rach authorized technician. Violation will void warranty.
Register your warranty by visiting
www.MyBacharach.com
For product support or additional information contact Bacharach at 724-334-5000 or 1-800-736-4666.
SETTING DATE AND TIME
To adjust the date/time manually, enter the menu by sim­ultaneously pressing both buttons ( + ) then releasing them. (The Snifit+ will force the user to update the time if it is invalid during startup.)
The Snifit+ shows “Pin”. Press and release either button to show “900”. Press to increase the number to 925. The display rolls over to 900 when it exceeds 998. Press and release . Next:
1.
The Snifit+ will beep continuously if invalid at startup and show “rtc”. Press either button.
2.
The display shows “Yr”. Press either button. Snifit+ shows the last two digits of the current year.
3. Press to increase year (roll­Adjust to desired “Yr”.
4.
Press to accept the desired “Yr”.
5. The month icon and current month are displayed.
6. Press to increase month (roll-over after 12). Adjust to desired “MO”.
7.
8. The display shows “dd”. Press either button. Display
9. Press to increase the day (roll-over after max num­ber of days for selected month). Adjust to desired
10.
Press to accept the desired “dd”.
11.
The display shows “Hr”. Press either button. Snifit+ shows the current hour in the 24-Hour format (no
12 Press to increase the hour (roll-over after 23).
13
Press to accept the desired “Hr”.
14.
The display shows “nn”. Press either button. Snifit+
15. Press to increase the minute (roll-
16.
Press to accept the desired “nn”.