Product Leadership • Training • Service • Reliability
WARRANTY
Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in material
and manufacture and will conform substantially to Bacharach Inc.’s applicable specifications. Bacharach’s
liability and Buyer’s remedy under this warranty are limited to the repair or replacement, at Bacharach’s option,
of this Product or parts thereof returned to Seller at the factory of manufacture and shown to Bacharach Inc.’s
reasonable satisfaction to have been defective; provided that written notice of the defect shall have been given
by Buyer to Bacharach Inc. within two (2) years after the date of delivery of this Product by Bacharach, Inc.
The Oxygen (O2) sensor has a limited warranty of one (1) year. Routine service and calibration are excluded
from this warranty.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s liability and Buyer’s
remedy under this warranty of title are limited to the removal of any title defects or, at the election of Bacharach,
to the replacement of this Product or parts thereof that are defective in title.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY
AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II) ANY
OBLIGATION, LIABILITY, RIGHT, CLAIM OR REMEDY IN CONTRACT OR TORT, WHETHER OR NOT ARISING FROM BACHARACH’S NEGLIGENCE, ACTUAL OR IMPLIED. The remedies of the Buyer shall be limited
to those provided herein to the exclusion of any and all other remedies including, without limitation incidental or
consequential damages. No agreement varying or extending the foregoing warranties, remedies or this limitation will be binding upon Bacharach, Inc. unless in writing, signed by a duly authorized officer of Bacharach.
Register Your Warranty by Visiting
www.bacharach-inc.com
Notice:
Product improvements and enhancements are continuous, therefore the specifications and information
contained in this document may change without notice.
Bacharach, Inc. shall not be liable for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of this material.
No part of this document may be photocopied, reproduced, or translated to another language without the prior
written consent of Bacharach, Inc.
The PCA 2 is a commercial-grade hand-held combustion and emissions
analyzer designed for on-demand sampling of light industrial, institutional, commercial and residential furnaces, appliances, and boilers. The
basic instrument is supplied with a probe and hose assembly, instruction
manual, factory calibrated smart sensors, 4 'A A' alkaline batteries, Data
Download Software with USB cable and carrying case.
Because of the PCA 2’s ability to measure up to four gases simultaneously,
it is the perfect tool for service technicians, inspectors and boiler contractors who need to determine combustion efficiency, excess air, stack gas O2
and CO levels, stack temperature, draft, and differential pressure. The
analyzer can also directly measure and display NO, NO2 and SO2 with the
installation of the appropriate sensors. Combustion efficiency calculations
can be conducted for the following fuels: natural gas, oil #2, oil #4, oil #6,
propane, coal, wood, kerosene, bagasse, and digester gas. A large backlit
graphical display shows up to eight combustion test values simultaneously,
and includes a zoom capability that provides an extra large display.
Smart sensor technology allows a new sensor to be installed in the field
without having to calibrate the analyzer before use. New and innovative
probe and analyzer designs allow the PCA 2 to be easily serviced, thus
lowering the cost of ownership.
Advanced data storage and communication features allow the operator
to store up to 500 individual combustion test records, which can later be
recalled for viewing, printing, or downloading to a personal computer. In
its data logging mode, the analyzer can store an additional 500 data logged
records.
An optional AC power adapter allows the analyzer to run continuously for
data logging purposes.
The optional sample conditioning probe is recommended when measuring
NO2 and SO2 to ensure the highest degree of measurement accuracy.
Instruction 24-94481-1
Introduction
1.2 Sales Combo & Model Configurations
Sales Combo24-835024-835124-8352
Sales Combo (Kit)24-837024-837124-8372
Model Type225235245
PCA2 Only Part Number24-730124-730224-7303
Measurements
Oxygen (O2)
Stack Temperature
Primary / Ambient Air Temperature
Carbon Monoxide Low (CO
Low
)
Pressure / Draft
Carbon Monoxide High (CO
High
)
Nitric Oxide (NO)
Nitrogen Dioxide (NO2)
Sulfur Dioxide (SO2)
Calculations
Combustion Efciency
Excess Air
Carbon Dioxide (CO2)
NOx (NOx = NO + NO2)
NOx referenced to %O
CO referenced to %O
NO referenced to %O
NO2 referenced to %O
SO2 referenced to %O
2
2
2
2
2
333
333
333
333
333
3
3
333
333
333
333
3
Refer to Section 8.2 for a listing of standard and optional accessories.
Instruction 24-94481-2
Introduction
Sales Combo24-835324-835424-8355
Sales Combo (Kit)24-837324-837424-8375
Model Type255265275
PCA2 Only Part Number24-730424-730524-7306
Measurements
Oxygen (O2)
Stack Temperature
Primary / Ambient Air Temperature
Carbon Monoxide Low (CO
Low
)
Pressure / Draft
Carbon Monoxide High (CO
High
)
Nitric Oxide (NO)
Nitrogen Dioxide (NO2)
Sulfur Dioxide (SO2)
333
333
333
333
333
33
3
33
Calculations
Combustion Efciency
Excess Air
Carbon Dioxide (CO2)
NOx (NOx = NO + NO2)
NOx referenced to %O
CO referenced to %O
NO referenced to %O
NO2 referenced to %O
SO2 referenced to %O
2
2
2
2
2
333
333
333
3
3
333
33
3
33
Instruction 24-94481-3
Introduction
1.3 Features & Benefits
• Powered by 4 ‘AA’ alkaline batteries, or NiMH rechargeable batteries.
An optional AC power adapter provides continuous operation.
• Internal charging circuit allows rechargeable batteries to be charged
inside the analyzer with the use of the optional AC power adapter.
• O2 and CO
two additional gases: CO
measurement standard. Optional measurement of up to
Low
, NO, NO2, or SO2.
High
• With the appropriate sensors installed, the analyzer optionally displays pollution conversions for CO, NO, NO2, and SO2. Pollution conversions include ppm, #/MBTU, mg/m3, and g/GJ.
• Smart sensor technology allows pre-calibrated sensors to be installed
in the field.
• Automatic zero of all sensing channels on ambient air when the analyzer is first turned ON.
• Automatic flushing of the CO
exceeds 4,000 ppm, thus protecting the CO
sensor with fresh air if the CO level
Low
sensor from high CO
Low
levels. To measure CO levels above 4,000 ppm, the analyzer automatically switches to its CO
sensor, if installed.
High
• Automatic purging of the gas-sample system if the detected gas levels
are abnormally high when the analyzer is turned OFF.
• Displays temperatures in either °F or °C.
• Displays pressure in either inwc, mb, Pa, or hPa.
• Backlit graphic LCD with zoom capabilities.
• Low battery alarm.
• Stores 500 individual combustion records, which can later be recalled
for viewing, printing, or downloading to a personal computer. Stores an
additional 500 data logged records.
• Wireless IrDA link used for printing current and stored combustion
records, pressure records, sensor calibration data, and diagnostic data.
• USB connectivity for downloading stored data to a personal computer.
• Field replaceable sensors and thermocouple.
• Two year warranty on analyzer and all gas sensors except the O2 sensor which has a one (1) year warranty.
Instruction 24-94481-4
Introduction
•Language options including English, French, and Spanish
•Custom Display Formats
•Calibration Reminders - PCA2 can be set up to remind the user that
calibration is past due.
1.4 Operational Overview
The PCA 2 is powered by either its 4 internal batteries, or by an optional
AC power adapter that operates from any convenient source of 100–
240 VAC, 50/60 Hz power. The type of batteries used can be either disposable alkaline or rechargeable NiMH. Note that rechargeable batteries can
be charged inside the analyzer using the optional AC power adapter.
The PCA 2 is controlled by 11 front panel push buttons, while a graphical
LCD is used to display all combustion and emission test data and analyzer
parameters.
A probe and hose assembly, with an integral thermocouple and filter/water-trap connect to the bottom of the analyzer, thus providing the means of
drawing in gas samples, and for measuring stack temperature and draft.
The PCA 2 is turned ON by pressing its red I/O button. A warm-up period
of 60 seconds then begins, during which time the analyzer performs self
diagnostics. At the end of the warm-up period, if no errors were detected
the instrument will display the Combustion Test HOLD screen. If errors
were detected, the message “ERRORS DETECTED” is displayed along
with a list of the errors. These errors must be corrected before proceeding
with the combustion test.
Before starting a test be sure to select the fuel being burned. The default
fuel selected is Natural Gas. Note that the name of the fuel being burned
is indicated at the top of the display. To change the fuel: first, press the
MENU (F2) button; next, select FUEL from the menu; then use the but-
tons to highlight the fuel being burned; and finally, press the green ENT
button to select the highlighted fuel.
To assure correct combustion-efficiency calculations, the analyzer must
know the burner’s primary-air temperature. The analyzer normally uses
its internal temperature sensor for the primary-air temperature value, but
this method is only acceptable if the burner is using ambient room air. If
the burner is drawing in cold outside air, we recommend that the optional TAIR thermocouple be used. This thermocouple plugs into the bottom of the
Instruction 24-94481-5
Introduction
analyzer and is placed in the burner’s primary-air stream.
Begin the combustion test by first inserting the analyzer’s probe tube into
the stack-gas stream of the appliance under test, and then pressing the
RUN/HOLD button to display the Combustion Test RUN screen. The analyzer will begin to continuously monitor the stack temperature, %O2 and
emission levels in the stack gas and then display measured and calculated
values on its LCD. Values are listed in Section 2 Specifications.
During a test, the CO
sensor is protected from high CO levels by being
Low
automatically flushed with fresh air when the detected CO level exceeds
4,000 ppm. The analyzer will automatically start using its optional CO-
sensor, if installed, at CO levels starting at 4,001 ppm, thus providing
High
continuous CO readings up to 20,000 ppm.
A backlight enables a user to read the display in dimly-lit areas. Turn the
backlight ON and OFF by briefly pressing the I/O button.
The analyzer is turned OFF by pressing and holding down the I/O button
for at least 2 seconds. Note that there is a 5-second delay before the analyzer actually turns OFF, during which time the analyzer can be turned back
ON by pressing the RUN/HOLD button. In addition, there is a gas-purge
feature that keeps the analyzer’s pump running if the gas level inside the
sensor chambers is abnormally high at shutdown. With the probe removed
from the stack and sampling fresh air, the analyzer purges itself until the
detected gas concentrations drop below predetermined levels.
1.5 Connector Descriptions
1.5.1 Probe Connections (Gas, Pressure, T-Stack)
Attach the probe and hose assembly to the analyzer by connecting its . . .
• stack-gas thermocouple to the analyzer's T-STACK connector,
• stack-gas hose to the analyzer's GAS connector,
• draft hose to the analyzer's +∆P connector.
Observe that the probe connectors are of different sizes and shapes, which
prevent incorrect connection to their associated connectors on the analyzer.
Instruction 24-94481-6
Introduction
T-STACKT-AIR
IrDA – Wireless printer
communications port
ENT:
•Selects a highlighted
menu item, or conrms the
entry of data
RUN / HOLD:
• Starts and stops a combustion
efciency test
• Pressing this button during
the 5 second turn-off period will
keep the analyzer turned ON
ESC – Displays previous menu
or previously viewed screen
I/O – Press from between
1 and 2 seconds to turn
analyzer ON and OFF
– With analyzer turned
ON, press briey to turn
backlight ON and OFF
LCD – 160 x 160
graphic display
F1 / F2 / F3 –
Soft Menu Buttons,
whose functions are
dened by labels
appearing above them
on LCD
USB – Computer
communications
connector
Arrow Buttons:
– Moves cursor
up the display,
or increments an
alphanumerical value
– Moves cursor
down the display,
or decrements an
alphanumerical value
– Moves cursor left,
or moves to top of
menu list
– Moves cursor right,
or moves to bottom of
menu list
If thermocouple P/N 104-1797 (10 feet long) or Utility Wand P/N 104-1799
(12 inch ridged probe with handle and 5 foot coiled cable) is to be used
to measure the burner’s primary air temperature, then connect either of
these thermocouples to the analyzer’s T-AIR connector.
1.5.3 POWER (AC Adapter)
The AC power adapter P/N 24-1404 can be used as an external power supply, which will run the analyzer on a continuous basis.
When using rechargeable NiMH batteries, the AC power adapter can also
be used to charge the batteries while inside the analyzer. The analyzer’s
rapid-charger circuit, however, must first be turned ON per Section 3.16.
The rapid charger will charge a set of depleted batteries in approximately
2 - 3 hours.
When disposable alkaline batteries are used, the analyzer’s battery charger circuit must be OFF to prevent the batteries from overheating. As a
precaution, the charger circuit is automatically toggled back to its OFF
state when the analyzer is turned OFF.
1.5.4 ∆P (Differential Pressure)
Draft is measured by connecting the probe’s draft hose to the +∆P fitting,
while leaving the –∆P fitting open to the atmosphere.
In addition to measuring draft, the “+” and “–” ∆P fittings can also be used
to measure the differential pressure between two areas by first connecting
a hose P/N 24-1103 to the –∆P fitting, and then inserting the open end of
this hose into the area being used as the reference pressure. The analyzer’s
probe is then inserted into the area where differential pressure is to be
measured. Refer to Section 4.6.
1.5.5 USB (Computer Interface)
Data that has been stored in the analyzer’s memory can be downloaded to
a personal computer by connecting USB data cable P/N 104-4032 between
the USB ports of the computer and analyzer. Refer to Section 4.14.2.
1.5.6 IrDA (Printer Interface)
Data that has been stored in the analyzer’s memory can be printed on a
Instruction 24-94481-8
Introduction
compatible IrDA wireless printer by aligning their IrDA communication
ports. Refer to Section 4.16.
1.5.7 OPT (Option)
The option connector is used for optional external measurement features.
1.6 Front Panel Buttons
Descriptions of the front panel buttons are given below. Note that a control
may perform multiple functions as determined by what screen is being
displayed at the time.
The functions of these buttons are defined by labels appearing above them on the LCD. The labels that appear
depend on the functions that can be performed in the
particular screen being displayed.
PRINT (F1): Transmits the data displayed on the screen
to a printer through the IrDA communications port.
MENU (F2): Displays the Main Menu.
SAVE (F3): Saves the data currently displayed on the
LCD in memory. Up to 500 individual Combustion Test
and Pressure records can be saved. After 500 records
have been saved, the memory must be cleared to continue
saving additional data. The analyzer will not overwrite
old data.
ZERO (F2): When viewing the Pressure screen, this
button zeros the pressure sensor to current atmospheric
conditions. When viewing the Temperature screen, this
button zeroes the temperature channel difference.
PAGE– (F1): When viewing the Memory or Logging
Directory, each press of this button pages down through
the directory. Holding this button down speeds up the
paging process.
PAGE+ (F3): When viewing the Memory or Logging
Directory, each press of this button pages up through
the directory. Holding this button down speeds up the
paging process.
Instruction 24-94481-9
Introduction
The arrow buttons move the cursor on the LCD in the direction of the arrow. In screens that require the entry of
alphanumerical data, use the buttons to move cursor
across the screen and then use the buttons to increment and decrement the data. When viewing a menu, use
the buttons to quickly move to the top and bottom of
the menu.
Selects a highlighted menu. In addition, if changes were
made to one of the analyzer’s operating parameters (e.g.,
date, time, O2 reference, etc.), pressing this button confirms those changes and saves them in memory.
Starts and stops a combustion test when the Combustion Test screen is displayed. Pressing this button in any
other screen returns the analyzer to the Combustion Test
HOLD screen. Pressing this button during the 5 second
turn-off-delay period will abort the turn-off process and
also return the analyzer to the Combustion Test HOLD
screen.
Displays a previously viewed screen. In addition, if
changes were made to one of the analyzer’s operating
parameters (e.g., date, time, O2 reference, etc.), pressing
this button aborts those changes, restores the old values,
and then displays the previously viewed screen.
Turns the analyzer ON and OFF, and is also used to
turn the backlight and button LEDs ON and OFF.
Note that when the analyzer is turned OFF, there is a
5 second delay, during which time an operator can keep
the analyzer turned ON by pressing the RUN / HOLD
button. Also note that if the measured emission levels
are above predetermined limits at the time the instrument is turned OFF, the pump is automatically started
and purges the sensor compartment with fresh air until
the gas levels inside the analyzer are reduced. If desired,
the purging process can be aborted by again pressing the
I/O button.
Instruction 24-94481-10
Specications
2 Specifications
The PCA 2 Directly Measures and Displays:
The gases displayed depend on the analyzer’s model number.
Refer to Section 1.2.
System Flow Rate with Probe . . . . . . . 200 cc/min minimum
Instruction 24-94482-2
Specications
Power Requirements:
Four disposable ‘AA’ alkaline batteries provide at least 15 hours of continuous operation. NiMH rechargeable batteries can also be used, with the
operating time dependent on battery type and condition.
An optional AC power adapter, which runs from any convenient source of
100–240 VAC, 50/60 Hz power, can be used to power the analyzer on a continuous basis. If using rechargeable batteries, the AC power adapter can also
be used to charge the batteries while inside the analyzer.
Warm-Up Time:
60 seconds. Sensors are checked and auto zeroed during warm-up.
Memory:
• 500 complete combustion test records
• 500 complete logged combustion test records
Interfaces:
• Printer – Infrared (IrDA) communications
• Computer – USB
Dimensions:
9H x 3W x 2.5D inches (22.9 x 7.6 x 6.3 cm)
Weight:
• Analyzer – 1.4 lb (0.6 kg) w/ batteries
• Probe & Hose Assembly – 1 lb (0.5 kg)
Instruction 24-94482-3
Specications
Notes:
Instruction 24-94482-4
Initial Setup
Release and then
remove battery cover
by pushing latch up
3 Initial Setup
3.1 Scope
Before using the PCA 2, you MUST:
• Install batteries, or plug in the optional AC power adapter (Section 3.2)
• Connect the probe and hose assembly (Section 3.3)
• Check, and if necessary, make changes to the analyzer’s configuration
(Section 3.4)
3.2 Power
3.2.1 Installing or Replacing Batteries
Either alkaline or NiMH rechargeable batteries can used to power the analyzer. Note that if rechargeable batteries are used, they can be recharged
while installed inside the analyzer using the optional AC power adapter
(refer to Section 3.2.2).
Install or replace the batteries as described below:
1. Remove battery cover from back of unit (Figure 3-1.)
2. Remove (and properly dispose of) any old batteries.
3. Install a set of four ‘AA’ alkaline or NiMH batteries, per the “+” and
“–” markings inside the battery compartment.
4. Replace battery cover.
Figure 3-1. Installing Batteries
Instruction 24-94483-1
Initial Setup
3.2.2 Using the AC Power Adapter
The AC power adapter is capable of powering the analyzer on a continuous
basis. The adapter plugs into an appropriate 100–240 VAC, 50/60 Hz wall
outlet, and produces an output of +9 VDC. The adapter’s output connector
plugs into the analyzer’s POWER jack located on the bottom of the unit
(Figure 3-2).
If NiMH rechargeable batteries are used, the adapter can also rapid
charge these batteries in approximately 2 - 3 hours while still inside the
analyzer. For the batteries to be charged, however, the analyzer’s battery
charger circuit must be turned ON per Section 3.16.
3.3 Connecting the Probe and Hose Assembly
Do the following to attach the probe and hose assembly to the analyzer
(Figure 3-2):
1. Push the gas-sample hose connector, the larger of the two connectors (giving a slight twist), onto the analyzer’s GAS fitting.
2. Push the draft-hose connector, the smaller connector (giving a slight
twist), onto the analyzer’s +∆P fitting.
3. Push the stack-gas thermocouple connector into the T-STACK jack (connector fits in only one way).
NOTE: The analyzer has a built-in temperature sensor for
measuring ambient temperature. Perform Step 4 only if the
optional primary / ambient air thermocouple is used.
4. Push the optional primary / ambient air thermocouple into the T-AIR
jack (connector fits in only one way).
IMPORTANT: To assure the accurate calculation of combustion
efficiency, the optional primary / ambient air thermocouple
must be used when the burner’s primary-air temperature is not
the same as the room temperature.
5. Inspect all hoses for cracks. If any hose is found to be defective, replace the entire probe and hose assembly. Check that the water trap is
empty, and that the filter is not dirty or saturated with water.
Instruction 24-94483-2
Initial Setup
POWER
P
T-STACK T-AIR
GAS
Differential
Pressure Hose
(— ∆P) (Optional)
Probe
Handle
Water Trap /
Filter Assembly
Stack Gas
Thermocouple
Connector
(T-STACK)
AC Power
Adapter Jack
(POWER)
Option
Connector
(Optional External
Measurement)
Draft
Hose
(+ ∆P)
Gas Sample
Hose
(GAS)
Probe Tube
Probe Stop
Primary /
Ambient Air
Thermocouple
(T-AIR) (Optional)
Figure 3-2. Connecting the Probe and Hose Assembly to the PCA 2
Instruction 24-94483-3
Initial Setup
FUEL MENU
NATURAL GAS
OIL #2
OIL #4
OIL #6
PROPANE
COAL
WOOD
(more)
MENU
F1F2F3
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
3.4 Operating Parameters
The PCA 2 is set up at the factory for the following operating parameters:
Time . . . . . . . . . . . . . . . . . . . . Current EST HH:MM AM/PM
O2 Reference . . . . . . . . . . . . . .0%
Print Pressure . . . . . . . . . . . .No
Zoom . . . . . . . . . . . . . . . . . . . .No
Battery Charger . . . . . . . . . . .OFF
Logging . . . . . . . . . . . . . . . . . . No
Button Sound . . . . . . . . . . . . .ON
To change any of these parameters, perform the associated procedure provided in Sections 3.5 thru 3.22.
3.5 Fuel Selection
To assure the accurate calculation of combustion
efficiency, select the fuel being burned as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight FUEL, and
then press ENT to display the FUEL MENU.
3. Use the buttons to scroll through the list of
available fuels until the desired fuel is highlighted. In the example shown, PROPANE
has been selected. (If custom fuels are added,
they will be displayed at the bottom of the
list)
TIP: Use the buttons to quickly scroll to the
bottom and top of the list.
4. Press ENT to save the selection and display
the Combustion Test HOLD screen. Observe
Instruction 24-94484
that the name of the selected fuel should now appear at the
SETUP MENU
TEMP UNITS
PRESSURE UNITS
POLLUTION UNITS
DATE
TIME
O2 REFERENCE
PRINT PRESSURE
(more)
MENU
F1F2F3
TEMP UNITS MENU
Celsius
Fahrenheit
MENU
F1F2F3
05/26/06 09:25:30 AM
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
top of the screen.
Note: In addition to the standard fuels programmed into the
PCA2, Bacharach can develop custom fuel codes based on the
customers specific needs. The PCA2 can be programmed with
2 additional fuels which can be added to the instrument using
the PC based software provided. Consult factory for price and
delivery.
3.6 Temperature Units Selection
Select to display temperature in either °F or °C
as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
Initial Setup
3. Use the buttons to highlight TEMP
UNITS, and then press ENT to display the
TEMP UNITS MENU.
4. Use the buttons to highlight the desired
temperature units. In the example shown,
Fahrenheit has been selected.
5. Press ENT to save the selection and re-display
the SETUP MENU.
Instruction 24-94485
Initial Setup
PRESSURE UNITS MENU
InchesWater
milliBar
Pascals
hectoPascals
MENU
F1F2F3
05/26/06 09:25:30 AM
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
SETUP MENU
TEMP UNITS
PRESSURE UNITS
POLLUTION UNITS
DATE
TIME
O2 REFERENCE
PRINT PRESSURE
(more)
MENU
F1F2F3
3.7 Pressure Units Selection
Select to display pressure in Inches of Water Column (inwc), millibar (mb),
Pascals (Pa), or hectoPascals (hPa) as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight PRESSURE
UNITS, and then press ENT to display the
PRESSURE UNITS MENU.
4. Use the buttons to highlight the desired pressure units. In the example shown,
InchesWater has been selected.
5. Press ENT to save the selection and re-display
the SETUP MENU.
Instruction 24-94486
Initial Setup
SETUP MENU
TEMP UNITS
PRESSURE UNITS
POLLUTION UNITS
DATE
TIME
O2 REFERENCE
PRINT PRESSURE
(more)
MENU
F1F2F3
POLLUTION UNITS MENU
ppm
#/Mbtu
mg/m3
g/GJ
MENU
F1F2F3
05/26/06 09:25:30 AM
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
3.8 Pollution Units Selection
The PCA 2 is capable of converting the measured ppm levels of CO, NO,
NO2, and SO2 to various pollution units using CFR40 Part 60 emission
factors. Note that the pollution-unit conversions for NO, NO2 and NOx are
based on the molecular weight of NO2.
Select to display pollution units in parts per million (ppm), pounds of pollutant per million BTU (#/Mbtu), milligrams of pollutant per cubic meter
of gas (mg/m3), or grams of pollutant per gigajoule (g/GJ) as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight POLLUTION
UNITS, and then press ENT to display the
POLLUTION UNITS MENU.
4. Use the buttons to highlight the desired
pollution units. In the example shown, ppm
has been selected.
5. Press ENT to save the selection and re-display
the SETUP MENU.
Instruction 24-94487
Initial Setup
SETUP MENU
TEMP UNITS
PRESSURE UNITS
POLLUTION UNITS
DATE
TIME
O2 REFERENCE
PRINT PRESSURE
(more)
MENU
F1F2F3
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
DATE MENU
DATE 05/26/06
Press ENTER to Save
05/26/06 09:25:30 AM
MENU
F1F2F3
3.9 Date Setup
The date is stored in the format: MM/DD/YY. Its
value is part of the date and time stamp that is
saved along with each combustion test record.
Set the analyzer’s internal clock to the current
date as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP,
and then press ENT to display the SETUP
MENU.
3. Use the buttons to highlight DATE, and
then press ENT to display the DATE MENU.
4. First use the buttons to move the cursor
across the screen until it is over the digit to
be changed, and then press the buttons
until the desired value is displayed.
5. Repeat Step 4 until the values for month, day,
and year have been set.
6. Press ENT to save the displayed date values
and re-display the SETUP MENU, or press
ESC to abort this procedure and retain the
old date values.
Note: The Date and Time real time clock is powered by the main batteries and is maintained by a supercap on the Main PCB for approximately 1-2 days in the absence of batteries. The supercap is intended
to maintain the real time clock when the batteries are changed when
exhausted. If the batteries are removed for extended periods of time
such as when the PCA2 is not in use (off-season storage), simply reset
the time and date after fresh batteries are installed when it is placed
back in service.
Instruction 24-94488
SETUP MENU
TEMP UNITS
PRESSURE UNITS
POLLUTION UNITS
DATE
TIME
O2 REFERENCE
PRINT PRESSURE
(more)
MENU
F1F2F3
3.10 Time Setup
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
TIME MENU
TIME 09:25 AM
Press ENTER to Save
05/26/06 09:25:30 AM
MENU
F1F2F3
The time is stored in the format: hh:mm:ss AM/
PM. Its value is part of the date and time stamp
that is saved along with each combustion test record.
Set the analyzer’s internal clock to the current
time as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP,
and then press ENT to display the SETUP
MENU.
3. Use the buttons to highlight TIME, and
then press ENT to display the TIME MENU.
4. First use the buttons to move the cursor
across the screen until it is over the digit to
be changed, and then press the buttons
until the desired value is displayed.
Initial Setup
5. Repeat Step 4 until the values for hour, minute, and meridiem have been set.
NOTE:The value for seconds cannot be en-
tered, but are displayed and stored as part of
the combustion test record.
6. Press ENT to save the displayed time values
and re-display the SETUP MENU, or press
ESC to abort this procedure and retain the old
time values.
Instruction 24-94489
Initial Setup
O2 REFERENCE
CO Ref to O2
NOx Ref to O2
SO2 Ref to O2
05/26/06 09:25:30 AM
MENU
F1F2F3
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
SETUP MENU
TEMP UNITS
PRESSURE UNITS
POLLUTION UNITS
DATE
TIME
O2 REFERENCE
PRINT PRESSURE
(more)
MENU
F1F2F3
SET O2 REFERENCE
CO REF TO O2: 0
05/26/06 09:25:30 AM
MENU
F1F2F3
3.11 O2 Reference Setup
The measured values of CO, NOx, and SO2 can be
individually referenced to a specific O2 percentage
of between 0 and 15%.
Individually set up the O2 reference value for each
of the above gases as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight O2 REF, and
then press ENT to display the O2 REFERENCE screen.
4. Use the buttons to highlight the desired
measurement, and then press ENT to display
the REF TO O2 screen for that measurement.
In the example shown, CO has been selected.
5. First use the buttons to move the cursor
across the screen until it is over the digit to
be changed, and then press the buttons
until the desired value is displayed.
6. Press ENT to save the displayed value and redisplay the O2 REFERENCE screen, or press
ESC to abort this procedure and retain the
old O2 reference value.
7. If the O2 reference value for more than one gas
is being set, then repeat Steps 4, 5, and 6 for
each measurement.
Instruction 24-944810
3.12 Print Pressure Selection
PRINT PRESSURE
No
Yes
05/26/06 09:25:30 AM
MENU
F1F2F3
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
SETUP MENU
(more)
TIME
O2 REFERENCE
PRINT PRESSURE
ZOOM
BATTERY CHARGER
LOGGING
(more)
MENU
F1F2F3
Select whether to print or not print the pressure
measurement on the combustion test printout as
follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight PRINT PRESSURE, and then press ENT to display the
PRINT PRESSURE screen.
4. Use the buttons to highlight either No (do
not print pressure) or Yes (print pressure). In
the example shown, Yes has been selected.
5. Press ENT to save the selection and re-display
the SETUP MENU.
Initial Setup
Instruction 24-944811
Initial Setup
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
SETUP MENU
(more)
TIME
O2 REFERENCE
PRINT PRESSURE
ZOOM
BATTERY CHARGER
LOGGING
(more)
MENU
F1F2F3
ZOOM
Standard
2x
3x
05/26/06 09:25:30 AM
MENU
F1F2F3
O24.0 %
CO12 ppm
EFF82.6%
PRINT MENU SAVE
F1F2F3
Hold NGas
3.13 Zoom-Display Selection
Combustion test data in the Run/Hold screen can
be shown with enlarged characters to make viewing easier. The operator can set zoom levels to
Standard, 2X, or 3X. The Standard zoom setting
will display 8 lines of combustion test data at one
time, 2X which will display 4 lines of combustion test data with enlarged characters, and 3X
which will display 3 lines of combustion test data
with enlarged characters. The operator can scroll
through the complete list of measured and calculated data no matter what zoom level has been
selected.
Select desired zoom level as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight ZOOM, and
then press ENT to display the ZOOM screen.
4. Use the buttons to select the desired Zoom
level. Options include STANDARD, 2X, and
3X. Standard will display eight lines of Combustion test data, 2X will display four, and 3X
will display 3.
5. Press ENT to save the selection and re-display
the SETUP MENU.
Instruction 24-944812
3.14 Battery Charger Selection
BATTERY CHARGER
OFF
ON
05/26/06 09:25:30 AM
MENU
F1F2F3
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
SETUP MENU
(more)
TIME
O2 REFERENCE
PRINT PRESSURE
ZOOM
BATTERY CHARGER
LOGGING
(more)
MENU
F1F2F3
When using rechargeable NiMH batteries, the AC
power adapter can be used to charge the batteries
while inside the analyzer. The analyzer’s rapidcharger circuit, however, must be first turned ON.
IMPORTANT: When using disposable alkaline batteries, the analyzer’s battery charger
circuit should be OFF to prevent the batteries from overheating.
As a precaution, the charger circuit is automatically toggled to its OFF state when the analyzer is
turned OFF, thus requiring the operator to turn
the charger back ON when needed.
Note: The PCA2 must be left turned ON to
charge the batteries.
Turn the battery charger circuit ON and OFF as
follows:
Initial Setup
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP,
and then press ENT to display the SETUP
MENU.
3. Use the buttons to highlight BATTERY
CHARGER, and then press ENT to display
the BATTERY CHARGER screen.
4. Use the buttons to highlight either OFF
(turn charger OFF) or ON (turn charger ON).
In the example shown, ON has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
Instruction 24-944813
Initial Setup
BUTTON SOUND
OFF
ON
05/26/06 09:25:30 AM
MENU
F1F2F3
SETUP MENU
(more)
LOGGING
TEST ID
USERNAME
RUN/HOLD FORMAT
LANGUAGE SELECTION
BUTTON SOUND
CAL REMINDER PERIOD
MENU
F1F2F3
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
SETUP MENU
(more)
LOGGING
TEST ID
USERNAME
RUN/HOLD FORMAT
LANGUAGE SELECTION
BUTTON SOUND
CAL REMINDER PERIOD
MENU
F1F2F3
3.15 Logging Selection
–
When the logging function is activated, up to 500
combustion test records will be automatically
stored in memory at a preset interval over a predetermined length of time.
Refer to Section 4.14 for detailed information on
how to select the logging function; how to set the
interval and duration time periods; and how to
view or download the stored data.
3.16 Button Sound
The audible sound used to signal when a button is
pressed can be turned OFF and ON as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight BUTTON
SOUND, and then press ENT to display the
BUTTON SOUND screen.
4. Use the buttons to highlight either OFF
(turn sound OFF) or ON (turn sound ON). In
the example shown, OFF has been selected.
5. Press ENT to save the selection and
re-display the SETUP MENU.
Instruction 24-944814
Initial Setup
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
Test ID Menu
Select Test ID
Edit Test ID
Clear Test ID
F1F2F3
MENU
3.17 Test ID Information
Test records can be identified (e.g. customer's name, burner number, and
location) by manually entering up to three lines of text, with each line containing a maximum of 20 alphanumeric characters. When a Test ID is selected this information will be associated with all succeeding test records,
and will appear at the top of each test record when printed, and in Excel
files when records are downloaded to a PC. The chosen Test ID remains in
effect until it is deselected, a new Test ID is selected, or the instrument is
turned off. Up to 30 Test ID's can be pre-entered
for later retrieval.
To enter or edit a Test ID:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary,
press ESC until MENU appears above
F2 .
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight TEST ID, and
then press ENT to display the TEST ID menu
4. Use the buttons to highlight EDIT TEST
ID, and then press ENT to display the EDIT
TEST ID menu, which displays the first line of
each Test ID record.
5. Use the buttons to select which Test ID to
edit and then press ENT to display all three
lines of that record. Periods (.....) are used to
identify empty lines.
6. Use the buttons to choose which of the three Test ID lines to edit
and then press ENT to begin editing the chosen line.
7. Use the buttons to enter the desired character and then move to the
next character position by pressing the right arrow key. Repeat this
step until the line is complete. A maximum of 20 alphanumeric characters can be entered. Press ENT to accept the information.
8. At this time, either return to step 6 to choose another line to edit, or
Instruction 24-944815
Initial Setup
Test ID Menu
Select Test ID
Edit Test ID
Clear Test ID
F1F2F3
MENU
Test ID Menu
Select Test ID
Edit Test ID
Clear Test ID
F1F2F3
MENU
end this procedure by highlighting Edit Complete and pressing the ENT
key to return to the Edit Test ID Menu.
9. Press ESC to go to the Test ID Menu or the RUN/HOLD key to return
to the Run/Hold screen.
Select a Test ID:
1. From the Test ID menu, use the buttons to
highlight SELECT TEST ID, and then press
ENT to display the SELECT TEST ID menu,
which displays the first line of each Test ID
record.
2. Use the buttons to choose a pre-entered
Test ID or choose NO Test ID (located at
bottom of list) to deselect a previously chosen
record, then press ENT to make the selection
and return to the Test ID menu.
3. Press ESC to end the procedure and return to the Setup Menu or press
the RUN/HOLD key to return to the Run/Hold Screen.
Clear Test ID Information:
1. From the Test ID menu, use the buttons
to highlight CLEAR TEST ID, and then press
ENT to display the CLEAR TEST ID menu.
2. Do one of the following to clear Individual Records or All Records.
*Individual Records - Use the buttons to
highlight Individual Records, then press ENT
to display the Clear Individual menu. Again
use the buttons to highlight the individual
record to clear, then press ENT to clear the
record.
*All Records - Use the buttons to highlight All Records, then press
ENT to display the Clear All menu. Again use the buttons to highlight Yes, then press ENT to clear all records.
3. Press ESC to end this procedure and return to the Test ID menu or
press the RUN/HOLD key to return to the Run/Hold screen.
Instruction 24-944816
Initial Setup
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
SETUP MENU
(more)
LOGGING
TEST ID
USERNAME
RUN/HOLD FORMAT
LANGUAGE SELECTION
BUTTON SOUND
CAL REMINDER PERIOD
MENU
F1F2F3
3.18 Username
The name of the user or owner of the analyzer (e.g. company name, address, phone number) can be stored in memory by manually entering up
to three lines of text, with each line containing up to 20 alphanumeric
characters. This information will appear at the top of each printout, until
such time as new information is entered or cleared.
Username Information can be entered as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary,
press ESC until MENU appears above
F2 .
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight USERNAME,
and then press ENT to display all three lines
of the EDIT USERNAME screen. Periods(.....)
identify empty lines.
4. Use the buttons to choose which of the
three Username lines to edit, then press ENT
to begin editing the chosen line.
5. Use the buttons to enter the desired character and then move to the next character position by pressing the right arrow key. Repeat
this step until the line is complete. A maximum of 20 alphanumeric characters can be entered.
6. Press ENT to accept the entered information.
7. At this time, either return to step 4 to choose another line to edit, or
end this procedure by highlighting Edit Complete and pressing the
ENT key to return tot he Setup Menu.
8. Press ESC to go to the Main Menu or the RUN/HOLD key to return to
the Run/Hold screen.
Instruction 24-944817
Initial Setup
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
SETUP MENU
(more)
LOGGING
TEST ID
USERNAME
RUN/HOLD FORMAT
LANGUAGE SELECTION
BUTTON SOUND
CAL REMINDER PERIOD
MENU
F1F2F3
Edit Username
............
............
............
Edit Complete
MENU CLEAR
F1F2F3
To Clear a Username proceed as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary,
press ESC until MENU appears above
F2 .
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight USERNAME,
and then press ENT to display all three lines
of the EDIT USERNAME screen. Periods(.....)
identify empty lines.
4. Use the buttons to choose which of the
three Username lines to clear, then press
CLEAR (F3) to clear that line.
5. At this time, either return to step 4 to choose
another line to clear, or end this procedure by
highlighting EDIT COMPLETE and pressing
the ENT key to return tot he Setup Menu.
6. Press ESC to go to the Main Menu or the RUN/
HOLD key to return to the Run/Hold screen
Instruction 24-944818
Initial Setup
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
SETUP MENU
(more)
Logging
Test ID
Username
Run/Hold Format
Language Selection
Button Sound
CAL Reminder Period
MENU
F1F2F3
Language Selection
English
Français
Español
06/28/06 11:45:30 AM
MENU
F1F2F3
3.19 Language
Information on the display screen can be shown in English, French, or
Spanish. Select the desired language as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary,
press ESC until MENU appears above
F2 .
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight LANGUAGE
SELECTION, and then press ENT to display
The Language Selection Menu.
4. Use the buttons to highlight the desired
language, then press ENT to activate the selection and re-display the Setup Menu.
Instruction 24-944819
Initial Setup
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
SETUP MENU
(more)
Logging
Test ID
Username
Run/Hold Format
Language Selection
Button Sound
CAL Reminder Period
MENU
F1F2F3
CAL Reminder Period
6 months
8 months
10 months
12 months
15 months
MENU
F1F2F3
3.20 Cal Reminder Period
The analyzer can be set to indicate a calibration reminder during the 60
second warm-up period. Calibration reminders can be preset to occur 6, 8,
10, 12, or 15 months after the last calibration. When the preset period is
exceeded the instrument will display the reminder, and how long since the
sensors were last calibrated. The reminder will be displayed at the end of
the 60 second warm-up period. If a calibration reminder is displayed the
operator can press the RUN/HOLD key to move to the Run/Hold screen for
normal operation. Regular calibration periods of 6 months to a year for all
gas sensors (except Oxygen) are recommended.
Set the calibration reminder period as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary,
press ESC until MENU appears above
F2 .
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight CAL REMINDER PERIOD, and then press ENT to
display CAL Reminder Period Menu.
4. Use the buttons to highlight the desired
CAL reminder period, and then press the ENT
key to activate the selection and re-display the
Setup Menu.
Note: The date and time settings must be correct to
get accurate cal reminders.
Instruction 24-944820
Initial Setup
3.21 Run/Hold Screen Format
PCA2 testdata is located in the Run/Hold screen. By pressing the Run/
Hold key, you should hear the pump running and see the word RUN at
the upper-left hand corner of the display. The instrument is continuously
measuring and calculating the data that is shown in the Run/Hold screen.
Press the RUN/HOLD key again, the pump should stop running and the
word HOLD should be shown at the upper-left hand corner of the display.
The instrument will now show the last measured and calculated data
taken before the instrument was placed in HOLD. Use the up and down
arrow keys to scroll through the complete list of measured and calculated
values when the instrument is running or in the hold mode.
The order in which data appears in the Run/Hold screen is as follows:
O2 = Oxygen
CO = Carbon Monoxide
EFF = Combustion Efciency
CO2 = Carbon Dioxide
T-STK = Stack Temperature
T-AIR = Ambient/Primary Air Temperature
EA = Excess Air
CO(O) = Carbon Monoxide content referenced to an Oxygen
percentage
NO = Nitric Oxide
NO2 = Nitrogen Dioxide
NOx = Oxides of Nitrogen (NO and NO2 combined)
SO2 = Sulfur Dioxide
NO(O) = Nitric Oxide content referenced ro an Oxygen percentage
NO2(O) = Nitrogen Dioxide content referenced to an Oxygen
percentage
NOx(O) = Oxides of Nitrogen content referenced to an Oxygen
percentage
SO2(O) = Sulfur Dioxide content referenced to an Oxygen percentage
Note: (O) denotes the current O2 Reference selected.
Note: Stars (***) appear in measurement and calculation elds of sensors
that are not installed.
Instruction 24-944821
Initial Setup
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
SETUP MENU
(more)
Logging
Test ID
Username
Run/Hold Format
Language Selection
Button Sound
CAL Reminder Period
MENU
F1F2F3
Run/Hold Format
Edit Format
Reset Format
MENU
F1F2F3
Edit Run/Hold Format
O2
CO
EA
CO2
T-STK
T-AIR
NO
(more)
MENU
F1F2F3
Change the order in which data is displayed as
follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary,
press ESC until MENU appears above
F2 .
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight RUN/HOLD
FORMAT, and then press ENT to display Run/
Hold Format Menu.
4. Use the buttons to highlight Edit Format,
and then press the ENT key to show the Edit
Run/Hold Format, where the current order of
combustion data is displayed.
5. Change data shown for a particular location
by first using the buttons to highlight the
location. Then press the ENT key to select the
location; the cursor will start to flash.
6. Use the buttons to scroll through and select
the desired data to appear at that location.
Press the ENT key to make the selection.
7. Change data displayed at other locations by
repeating steps 5 & 6.
8. When finished, use the button to highlight
Edit complete at the bottom of the list, then
press ENT to save the new display format and
return to the Run/Hold Format screen.
Instruction 24-944822
Reset Display format back to factory default settings as follows:
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
SETUP MENU
(more)
Logging
Test ID
Username
Run/Hold Format
Language Selection
Button Sound
CAL Reminder Period
MENU
F1F2F3
Run/Hold Format
Edit Format
Reset Format
MENU
F1F2F3
Reset Format
No
Yes
MENU
F1F2F3
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary,
press ESC until MENU appears above
F2 .
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight RUN/HOLD
FORMAT, and then press ENT to display Run/
Hold Format Menu.
4. Use the buttons to highlight RESET FORMAT, then press ENT to display the Reset
Format screen.
5. Use the buttons to select YES, then press
ENT to reset the display and return to the
Run/Hold Format screen.
Initial Setup
Instruction 24-944823
Operation
4 Operation
4.1 Operating Tips
• When an analyzer is brought in from a cold vehicle, let it warm up slowly
to minimize condensation. Temperatures below freezing will not damage the analyzer; however, bringing a cold analyzer into a warm, humid
environment may cause condensate to form inside the case.
CAUTION: Although the analyzer itself is not damaged by an extremely
cold environment, the electrochemical sensors may be damaged. The O2
sensor's electrolyte will freeze at approximately -20 ºF and the other sensors at approximately -94 ºF. If the analyzer is exposed to an extremely
cold condition, it is strongly suggested that the sensor housings be examined for hairline cracks. Be aware that a leaking sensor can cause chemical burns to the skin and possibly damage the PCB assemblies.
• Ensure that the analyzer is sampling fresh air when turned ON. Pulling a
stack-gas sample through the analyzer during its warm-up period will not
damage the analyzer, but it will result in incorrect sensor readings, and may
result in sensor error messages appearing after the warm-up cycle completes.
• Note that flue-gas condensate is acidic and very corrosive. It is important
not to allow the analyzer’s internal components to come in contact with
condensate for long periods of time.
• Before each use, inspect the filter element of the water-trap / filter assembly. Replace the filter if it looks dirty. Refer to Section 6.2.
• When sampling flue-gas, keep the analyzer above the water-trap, and keep
the trap in a vertical position. This will maximize the effectiveness of the
trap and keep liquid condensate from being drawn directly into the analyzer.
• When liquid condensate is seen inside the water trap, empty the trap
before it becomes full. Refer to Section 4.9.
• The analyzer should be purged after performing a combustion test. After
removing probe from the stack, let the pump run for at least 10 minutes
to remove any stack gases and dry any condensate from inside the sensor
chamber and probe assembly. If the analyzer is turned OFF with high
levels of flue gas remaining in the analyzer, then the pump will continue
to run and the message “PURGING SENSORS” will appear on the display until all flue gas levels fall below predetermined levels.
• When storing the analyzer, it’s a good idea to empty the water trap and
leave it open to further dry it out.
• Calibrate the analyzer every 6 months - 1 year to assure its accuracy.
Instruction 24-94484-1
Operation
MENU
Warm Up 60
ERRORS DETECTED
Low Battery
MENU
MENUMENU
PCA 2
FIRMWARE VERSION
x.xx
MODEL xxx
SERIAL # x xxxxxx
HOLD NATURAL GAS
O220.9 %
CO 0 ppm
EFF --- %
CO2 --- %
1. Connect the probe and hose assembly, and make sure that the analyzer
is properly set up per Section 3 Initial Setup.
IMPORTANT: DO NOT insert probe into stack
before turning ON the analyzer!
2. Place the probe in an area that contains fresh
air. This ensures that the sensors will be
properly zeroed during the warm-up cycle.
3. Turn ON the analyzer by pressing the I/O
button for at least 1 second, or until a single
beep is heard. Observe that the analyzer’s
firmware version, model and serial numbers
are briefly displayed followed by the Warm Up
screen.
4. Wait for the analyzer to count down its 60 second warm-up period; after which, the instrument will display the Combustion Test HOLD
screen.
If problems were detected during warm up,
the message “ERRORS DETECTED” is displayed along with a list of those errors. As an
example, the screen to the right shows that the
battery is low. Refer to Section 7.3 for a listing
and possible remedy for the errors displayed.
TIP: If the sensors in error are not critical to
the combustion test, then press the RUN/HOLD
button to display the Combustion Test HOLD
screen and proceed with the test.
Instruction 24-94484-2
4.3 Selecting a Fuel
FUEL MENU
NATURAL GAS
OIL #2
OIL #4
OIL #6
PROPANE
COAL
WOOD
(more)
MENU
F1F2F3
MAIN MENU
FUEL
PRESSURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
HOLD NATURAL GAS
O220.9 %
CO 0 ppm
EFF --- %
CO2 --- %
The top line of the Combustion Test HOLD screen
shows the fuel currently selected. In the example
shown, the current fuel is NATURAL GAS. If necessary, change the fuel as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button.
2. Use the buttons to highlight FUEL, and
then press ENT to display the FUEL MENU.
3. Use the buttons to scroll through the list of
available fuels until the desired fuel is highlighted. In the example shown, PROPANE has
been selected.
NOTE: If custom fuels were added to the
instrument, they will be located at the bottom of
the list.
TIP: Use the buttons to quickly scroll to the
bottom and top of the list.
Operation
4. Press ENT to save the selection and return
to the Combustion Test HOLD screen. The
newly selected fuel should now appear in the
top line of the display.
Instruction 24-94484-3
Operation
O2,
CO,
Stack Temp.
Undiluted Flue
Gas Sample Taken
Under Draft
Diverter in Top
of Fire Tube
O2, CO,
Stack Temp.
O2, CO,
Stack Temp.
4.4 Sampling Point
FORCED AIR FURNACE – For atmospheric burner
or gravity vented, forced air heating equipment
with a clamshell or sectional heat exchanger design, test each of the exhaust ports at the top of the
heat exchanger. The probe should be inserted back
into each of the exhaust ports to obtain a flue-gas
sample, before any dilution air is mixed in.
HOT WATER TANK – Domestic hot water tanks
with the ‘bell’ shaped draft diverter can be accurately tested by inserting the probe tip directly
into the top of the fire tube below the diverter.
80% EFFICIENCY FAN ASSIST OR POWER VENTED
– Combustion testing of fan assist or power vented,
furnaces/boilers should be done through a hole drilled
in the vent immediately above the inducer fan.
90% EFFICIENCY CONDENSING – Condensing furnaces/boilers can be tested through a hole
drilled in the plastic vent pipe (when allowed by
the manufacturer or local authority of jurisdiction)
or taken from the exhaust termination.
Forced Air Furnace
ATMOSPHERIC OR GRAVITY VENTED BOILER
– Boilers, which have a ‘bell’ shaped draft diverter
on top, should be tested directly below the diverter
through a hole drilled in the vent connector.
O2, CO,
Stack Temp.
80% Eff. Fan Assist or
90% Eff. Condensing
Power Vented
Hot Water Tank
Combustion Air
Sampling Point
Atmospheric or Gravity
Vented Boiler
Instruction 24-94484-4
Operation
RUN NATURAL GAS
O2 4.0 %
CO 12 ppm
EFF 82.6 %
CO2 9.5 %
T-STK 374 °F
T-AIR 68.0 °F
EA 21.3 %
CO(3) 13 ppm
PRINT MENU SAVE
F1F2F3
4.5 Performing a Combustion Test
Ensure that the following has been completed, and then proceed with the
combustion test as described below:
• Turn ON analyzer and allow it to warm up (Section 4.2).
• Select fuel being burned (Section 4.3).
• Inset probe into stack (Section 4.4).
• If necessary, insert optional primary air thermocouple into combustionair stream of burners that use an outside source of combustion air.
1. Press the RUN/HOLD button to start the test.
You should hear the pump start running and
see the word RUN appear at the top of the
Combustion Test screen.
Sensor Indicators: The following indicators
appear in the sensor's data field depending on
certain conditions:
(∗ ∗ ∗) Sensor that is not calibrated or installed.
(XXX) Sensor overrange
(
- - -
) The calculated data cannot be displayed
because the measured data necessary to make the calculation is out of
range (i.e., oxygen level above 16%).
2. Use the up and down arrow keys to scroll to the T-STK reading.
Loosen the thumbscrew on probe stop and move probe in and out of
the stack until the stack’s core temperature (hot spot) is located as
indicated by the highest T-STK reading; then tighten thumbscrew to
prevent further probe movement. Locating the highest stack tempera-
ture is very important for accurate efficiency calculations.
3. You can now begin burner-service procedures. The analyzer readings
will change quickly to show changes in burner performance.
CAUTION: Position the Water Trap with its gas-flow arrow
pointing upward. Do not let water condensate go above the tip
of the riser tube. The sensors could be damaged if water would
enter the analyzer. Empty the Water Trap after every combustion test (refer to Section 4.9)
4. Pressing the RUN/HOLD button freezes all readings, stops the pump and
displays the Combustion-Test HOLD screen. Press the ENT button to view
all test values at the moment the RUN/HOLD button was pressed. Press-
ing RUN/HOLD again restarts the pump and resumes testing.
Instruction 24-94484-5
Operation
TABLE 4-1. LIST OF COMBUSTION TEST DATA
Display
Name
O
2
Description of Measurement or Calculation
% Oxygen
COCarbon Monoxide (1)
EFFCombustion Efficiency
CO
2
% Carbon Dioxide
T-STKStack Temperature
T-AIRPrimary / Ambient Air Temperature as measured either in-
ternally or by an optional external thermocouple plugged into
the analyzer’s T-AIR connector
CO(O)Carbon Monoxide ppm level referenced to a % of oxygen (2)
NONitric Oxide (1)
NO
2
Nitrogen Dioxide (1)
NOxOxides of Nitrogen (NO and NO2 combined) (1)
SO
2
Sulfur Dioxide (1)
NO(O)Nitric Oxide ppm level referenced to a % of oxygen (2)
NO2(O)Nitrogen Dioxide ppm level referenced to a % of oxygen (2)
NOx(O)Oxides of Nitrogen ppm level referenced to a % of oxygen (2)
SO2(O)Sulfur Dioxide ppm level referenced to a % of oxygen (2)
NO Temp Nitric Oxide Sensor Temperature (3)
(1) Pollution unit of measure selected per Section 3.9
(2) The letter “O” represents the oxygen reference level of between 0 and 15% as
selected per section 3.12
(3) Shown only on printout, not on LCD display.
Note: Stars (***) appear in measurement and calculation fields of sensors that are not
installed
Instruction 24-94484-6
Operation
MAIN MENU
FUEL
PRESSURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
Pressure
Measured
6.00 inwc
Type: Draft Reading
F1F2F3
Print Zero Save
4.6 Pressure Label Selection
The pressure measurement can be labeed with types including gas pressure, differential across heat exchanger, draft reading, and differential
pressure. See Section 4.6 of the PCA2 Instruction Manual 24-9448 for
pressure and draft measurement procedures.
Label data as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight PRESSURE,
and then press ENT to display the PRESSURE MENU.
3. Use the buttons to scroll through the
available pressure label types. The current
label will be used, saved, and printed with the
record
4. Press ENT to save the selection and re-display
the PRESSURE MENU or ESC to exit back to
the Main Menu Press ENT to save the selection and re-display the PRESSURE MENU or
ESC to exit back to the Main Menu
Instruction 24-94484-7
Operation
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
Temperature
Measured Delta T:
*** °F
Type: Diff Temp
Print Zero Save
4.7 Temperature Measurement Label
The difference in temperature between two areas can be measured by using the analyzer's two temperature channels and the Temperature screen.
By using the T-Air channel as a reference, the temperature applied to
the T-Stack channel will be displayed on the Temperature Measurement
screen as differential temperature between the two channels. Additionally, the temperature measurement can be labeled. Label types include
differential temperature across heat exchange or differential temperature.
Perform a differential temperature measurement as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight TEMPERATURE, and then press ENT to display the
TEMPERATURE MEASUREMENT screen.
3. Install thermocouples in both temperature
channel connectors.
4. Before taking a measurement, the temperature
channels may need to be zeroed, if not already
displaying zero with both thermocouples exposed to the same condition. Press ZERO (F2)
if needed.
5. To label the data use the buttons to scroll
through the available temperature label types.
The current label will be used, saved, and
printed with the record.
6. Press ENT to save the selection and re-display
the TEMPERATURE MENU or ESC to exit
back to the Main Menu
Instruction 24-94484-8
Operation
POWER
P
T-STACK T-AIR
GAS
— ∆P (Pressure
Reference Port)
+∆P (Pressure
Measurement Port)
MAIN MENU
FUEL
PRESSURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
PRESSURE
Measured:
0.00 inwc
PRINT ZERO SAVE
F1F2F3
4.8 Making a Draft / Pressure Measurement
The difference in pressure (∆P) between two areas can be measured by
using the analyzer’s two pressure ports and the PRESSURE screen. By
using the –∆P port as the reference, the pressure applied to the +∆P port
will be displayed on the PRESSURE screen as the differential pressure
between the two ports.
Perform a draft / pressure measurement as follows:
1. Turn ON the analyzer and allow it to complete
its warm-up cycle (Section 4.2).
TIP: The pressure units of measure is selected
per Section 3.8.
2. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
3. Use the buttons to highlight PRESSURE,
and then press ENT to display the PRESSURE
screen.
4. Before taking a measurement, the pressure
sensor may need to be re-zeroed if it is not
already displaying zero with both pressure
ports open to the atmosphere. If necessary,
zero the pressure sensor as follows:
RUN NATURAL GAS
O2 4.0 %
CO 12 ppm
EFF 82.6 %
CO2 9.5 %
T-STK 374 °F
T-AIR 68.0 °F
EA 21.3 %
CO(3) 13 ppm
PRINT MENU SAVE
F1F2F3
a. Press the ZERO (F2) button.
b. Disconnect any hoses connected to the
+∆P and –∆P ports, and then press ENT
to zero the pressure sensor.
c. Reconnect any hoses. When measuring
draft, simply leave the –∆P port open to
the atmosphere and connect the probe’s
draft hose to the +∆P port (see Figure 3-2
on Page 3-3).
5. Do one of the following to measure draft or
differential pressure:
• To measure draft, simply insert the probe
into the stack and observe the draft reading on the PRESSURE screen.
• To measure differential pressure, connect
two sampling hoses to the +∆P and –∆P
ports, and place the open end of each hose
into the areas being measured. The differential pressure between the
two areas is now displayed on the PRESSURE screen. If the pressure
at the +∆P port is higher than the –∆P port, then the pressure reading
will be positive. If it is lower, then the reading will be negative.
4.9 Saving Test Data
Up to 500 individual sets (“snap shots”) of combustion-test, pressure, or
temperature data can be saved in memory, which can later be recalled for
viewing from the Memory Directory (Section 4.13.1).
1. First display the screen that contains the
data to be saved. In the example to the right,
all data associated with the Combustion Test
RUN screen will be saved.
2. Press the SAVE (F3) button to save the test
data in the next available memory location.
NOTE: When memory is full, the next reading
will not be saved until space is made available
by clearing the data. (See 4.15.4: Clearing
Memory)
Instruction 24-94484-10
4.10 Ending a Combustion Test
To Gas Port
on Analyzer
Gas
Flow
Do not allow water
level to go above
tip of riser tube
Stack Gas
From Probe
To empty water,
pull apart using
a slight twisting
motion
WARNING! Burn Hazard. Do not touch the probe after
removing it from the stack. Allow the probe to cool before
handling (about 5 minutes).
1. Remove probe from stack.
2. Allow the pump to run until all combustion gases have been flushed
from the analyzer as indicated by the O2 reading returning to 20.9%.
4.11 Emptying the Water Trap
The Water Trap / Filter Assembly removes stack-gas condensate, and also
prevents soot from contaminating the internal components of the analyzer.
IMPORTANT: Use the Water Trap /
Filter Assembly in a vertical position
with the gas-flow arrow pointing up as
shown in the illustration to the right.
Empty the water trap chamber after each
combustion test, or stop the test and empty
the chamber if the liquid condensate level approaches the tip of the riser tube.
To empty the trap, first pull apart the two
halves of the Water Trap using a slight twisting motion; empty the water trap chamber;
and then reassemble the trap.
After each combustion test, also check the
Water Trap’s filter element. If it looks dirty,
replace the filter per Section 6.2.
Operation
Instruction 24-94484-11
Operation
MENU
SHUTDOWN IN 5 sec
MENU
PURGING SENSORS
RUN NATURAL GAS
O2 4.0 %
CO 12 ppm
EFF 82.6 %
CO2 9.5 %
T-STK 374 °F
T-AIR 68.0 °F
EA 21.3 %
CO(3) 13 ppm
PRINT MENU SAVE
F1F2F3
Low Batter y Alarm
Empty Battery Icon
4.12 Turning OFF the Analyzer & Purging
Turn OFF the analyzer by pressing the I/O button for at least 2 seconds, or until two beeps are
heard. The unit will count down 5 seconds before
shutting down, giving the operator an opportunity to keep the analyzer turned ON by pressing
the RUN/HOLD button.
If the PCA 2 was not purged with fresh air as
described in Section 4.8, then the analyzer may
remain ON with its pump running and display
the message “PURGING SENSORS” as the result of combustion gases still being present inside
the analyzer. At this time the operator should
ensure that the probe is removed from the stack,
allowing the analyzer to purge itself with fresh
air. The 5-second-shutdown sequence will not
begin until the gas levels inside the analyzer drop
below predetermined levels:
TIP: Although not recommended, the purging process can be bypassed by pressing the
I/O button a second time.
4.13 Low Battery Alarm
When the batteries are nearly depleted, an empty battery icon appears in
the upper-right corner of the display, and a short beep is sounded every
10 seconds.
After a low battery alarm occurs, the analyzer
will continue to operate for only a few minutes.
The amount of operating time that remains depends on many factors (e.g., pump and backlight
being ON or OFF, and the type and condition of
the batteries).
Instruction 24-94484-12
Operation
4.14 Data Logging
When the logging function is activated, up to 500 combustion test records
will be automatically stored in memory at a preset interval (1, 5, 10, 15,
30 seconds, 1, 2, 5, 10 minutes) over a predetermined duration (5, 10, 15,
30 minutes, 1, 2, 5, 10, 24, 48 hours).
The maximum duration that data can be collected is determined by the interval. For example, if the interval is set to 10 seconds, then the maximum
selectable length of time in which data can be collected to fill 500 memory
locations would be 1 hour (500 x 10 seconds = 5000 seconds or 83 minutes).
If the operator chooses a duration that is longer than possible for the interval chosen, then the analyzer automatically selects the highest duration
possible for the selected interval.
The stored logged data can either be viewed on the display using the analyzer’s memory function (refer to Section 4.13), or downloaded to a personal
computer using the supplied PCA 2 data recovery software and USB cable
(refer to Section 4.14). Downloaded data is stored on the computer’s hard
drive as a comma-separated-value ASCII text file with a CSV extension,
which can be opened by most spreadsheet programs for analysis.
TIP: The latest PCA 2 data recovery software can be downloaded from http://www.bacharach-inc.com/downloads.
Instruction 24-94484-13
Operation
SETUP MENU
(more)
LOGGING
TEST ID
USERNAME
RUN/HOLD FORMAT
LANGUAGE SELECTION
BUTTON SOUND
CAL REMINDER PERIOD
MENU
F1F2F3
MAIN MENU
FUEL
PRESSURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
LOGGING
NO
YES
nnn MEMORY
LOCATIONS OPEN
MENU
F1F2F3
4.14.1 Turning ON Data Logging
IMPORTANT: Before turning ON data logging and starting the
data logging process, the analyzer should already be set up to
perform a combustion test per Section 4.5.
Turn ON data logging as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight LOGGING,
and then press ENT to display the LOGGING
screen.
TIP: The amount of memory available for
storing new data is displayed in the LOGGING screen (maximum of 500 locations). If
previous logging sessions are stored, and additional memory is needed, clear the logging
memory per Section 4.13.3.
NOTE: At no time will new logging data over
write old data.
4. Use the buttons to highlight YES (turn
logging ON).
5. Press ENT to make the selection and display
the LOGGING INTERVAL screen.
Instruction 24-94484-14
Operation
LOGGING INTERVAL
1 sec
5 sec
10 sec
15 sec
30 sec
1 min
2 min
(more)
MENU
F1F2F3
LOGGING DURATION
5 min
10 min
15 min
30 min
1 hr
2 hr
5 hr
(more)
MENU
F1F2F3
4.14.2 Setting the Logging Interval and Duration
The logging interval is the length of time between measurements, while
the logging duration is the time allocated to the logging process. Set the
logging interval and duration as follows:
TIP: If the duration is set for more than 10 hours, we recommend using the optional AC power adapter to power the
analyzer.
1. As soon as logging is turned ON as described
in Section 4.12.1, the LOGGING INTERVAL
screen appears.
2. Use the buttons to highlight the desired
interval, and then press ENT to make the
selection and display the LOGGING DURATION screen. In this example, 1 minute has
been selected
3. Use the buttons to highlight the desired
duration, and then press ENT to make the
selection and display the LOGGING SUMMARY screen. In this example, 2 hours has
been selected.
NOTE: If the selected duration requires more
memory than is currently available, then the
analyzer will automatically select the longest
duration possible for the selected interval.
Instruction 24-94484-15
Operation
LOGGING SUMMARY
Interval 1 min
Duration 2 hr
Press ENTER
to Start
MENU
F1F2F3
LOG NATURAL GAS
O2 4.0 %
CO 12 ppm
EFF 82.6 %
CO2 9.5 %
T-STK 374 °F
T-AIR 68.0 °F
EA 21.3 %
CO(3) 13 ppm
Log Entry xx
F1F2F3
4.14.3 Starting the Data Logging Process
After turning ON data logging and setting the
interval and duration, the analyzer will pause
at the LOGGING SUMMARY screen, where the
currently selected interval and duration time
periods are displayed.
Press ENT to start the combustion test and logging process. At this time the Combustion Test
LOG screen will appear, indicating that the analyzer is now performing a combustion test and the
data is being stored in memory.
Note the following:
• At the bottom of the screen, the memory location where the current log entry is being saved
is displayed.
• If there were previously stored log entries,
each new logging session will start to store
data in the next available memory location.
• At no time will new data overwrite old data.
• The ENT button can be pressed while logging to view the other Combustion Test screens.
Instruction 24-94484-16
Operation
LOG NATURAL GAS
O2 4.0 %
CO 12 ppm
EFF 82.6 %
CO2 9.5 %
T-STK 374 °F
T-AIR 68.0 °F
EA 21.3 %
CO(3) 13 ppm
yy Entries Stored
F1F2F3
Where: “yy” is the number of
log entries stored during the
last logging session.
4.14.4 Ending the Data Logging Process
Data logging will stop and the pump will turn OFF after the prescribed
duration, or after all 500 memory locations are filled.
To exit the Combustion Test LOG screen, press
the RUN/HOLD button twice to display the Combustion Test HOLD screen.
Note the following:
• To end the logging process at any time, press
the ESC button.
• At the end of the logging period, or if the
ESC button was pressed, the total number
of log entries that were stored during that
period appear at the bottom of the screen.
• If the logging process was stopped by pressing ESC, it cannot be resumed without first turning logging back ON per Section 4.12.1.
4.15 Memory
There are two memory banks, each containing 500 memory locations. The
first bank is used to store combustion test data as described in Section 4.7.,
while the second bank is used to store logged combustion test data as
described in Section 4.12. Each bank is independent of each other, and cannot share data or be combined.
Individual memory locations in each memory bank can be recalled for
viewing on the display or printed (refer to Sections 4.13.1 & 4.13.2), or the
entire contents of each memory bank can be individually downloaded to a
computer and viewed in a spreadsheet program for analysis (refer to Section 4.14).
Instruction 24-94484-17
TIP: When displaying the contents of either memory bank,
the operator can quickly page through the screens by pressing the PAGE– (F1) and PAGE+ (F3) buttons. Or move to
the first or last memory location by pressing the buttons,
respectively.
Mem: 1 NATURAL GAS
O2 4.0 %
CO 12 ppm
EFF 82.6 %
CO2 9.5 %
T-STK 374 °F
T-AIR 68.0 °F
EA 21.3 %
CO(3) 13 ppm
PRINT MENU
F1F2F3
MAIN MENU
FUEL
PRESSURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
MEMORY MENU
MEMORY DIRECTORY
LOGGING DIRECTORY
CLEAR MEMORY DATA
CLEAR LOGGING DATA
CLEAR ALL MEMORY
F1F2F3
MENU
4.15.1 Recalling Combustion Test Data
Recall individual combustion test data records as
follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight MEMORY,
and then press ENT to display the MEMORY
MENU.
3. Use the buttons to highlight MEMORY
DIRECTORY, and then press ENT to display
the MEMORY DIRECTORY screen.
4. Use the buttons to highlight the desired
memory location to be recalled. Each memory
location is identified by the date and time at
which data was saved. The word EMPTY signifies that the memory location does not contain
data.
5. Press ENT to display the data contained in
the selected memory location.
Note the following:
• The top line of the recalled combustion test
data screen shows the memory location being
viewed. In this example, “MEM: 1” is being
displayed
• The ENT button can be pressed to view other
memory test data screens.
• The recalled combustion test data can be
printed by pressing the PRINT (F1) button
(refer to Section 4.16).
Log: 6 NATURAL GAS
O2 4.0 %
CO 12 ppm
EFF 82.6 %
CO2 9.5 %
T-STK 374 °F
T-AIR 68.0 °F
EA 21.3 %
CO(3) 13 ppm
PRINT MENU
F1F2F3
4.15.2 Recalling Logged Test Data
MAIN MENU
FUEL
PRESSURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
MEMORY MENU
MEMORY DIRECTORY
LOGGING DIRECTORY
CLEAR MEMORY DATA
CLEAR LOGGING DATA
CLEAR ALL MEMORY
F1F2F3
MENU
Recall individual logged combustion test data records as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight MEMORY,
and then press ENT to display the MEMORY
MENU.
3. Use the buttons to highlight LOGGING
DIRECTORY, and then press ENT to display
the LOGGING DIRECTORY screen.
4. Use the buttons to highlight the desired
data logging location to be recalled. Each
logging location is identified by the date and
time at which the data was saved. The word
EMPTY signifies that the logging location
does not contain data.
5. Press ENT to display the logging data contained in the selected memory location.
Operation
Note the following:
• The top line of the log test data screen shows
the log location being viewed. In this example,
“Log: 6” is being displayed.
• The ENT button can be pressed to view other
log test data screens.
• The recalled logged test data can be printed
by pressing the PRINT (F1) button (refer to
Section 4.16).
Instruction 24-94484-19
Operation
CLEAR MEMORY DATA
NO
YES
F1F2F3
MENU
CLEAR MEMORY DATA
WAIT
Erasing memory
F1F2F3
MENU
MAIN MENU
FUEL
PRESSURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
MEMORY MENU
MEMORY DIRECTORY
LOGGING DIRECTORY
CLEAR MEMORY DATA
CLEAR LOGGING DATA
CLEAR ALL MEMORY
F1F2F3
MENU
4.15.3 Clearing Memory
When all memory (“snap shot”) locations used to
store individual combustion test records have been
filled, the next combustion test record saved will
overwrite the oldest.
When all logging memory locations in the logging
directory are full, they must be manually cleared
in order to store new data. At no time will the logging process overwrite older data.
Each memory bank can be individually cleared, or
all memory locations in both banks can be cleared
simultaneously.
Do the following to clear memory:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight MEMORY,
and then press ENT to display the MEMORY
MENU.
3. Use the buttons to highlight one of the
following:
• CLEAR MEMORY DATA clears only the
individual saved combustion test records.
• CLEAR LOGGING DATA clears only the
combustion test records that were saved
during the logging process.
• CLEAR ALL MEMORY clears all memory
locations in both memory banks.
4. Press ENT to display the CLEAR MEMORY, LOGGING, or ALL DATA conformation screen. Highlight YES to confirm that
memory is to be cleared, and then press ENT
to actually clear memory as evidenced by
the display of the “WAIT Erasing memory”
screen.
Instruction 24-94484-20
Operation
4.16 Downloading Stored Data to a Computer
The combustion test data that was stored in either the analyzer’s Memory
Directory (Section 4.9), or Logging Directory (Section 4.14), can be downloaded to a computer using the PCA 2 Data Recovery Program and USB
cable that are supplied with the analyzer.
The following procedures assume that the operator is familiar with creating folders and navigating the file structure of the Windows operating system. If necessary, consult the Windows help files for instructions on how to
perform these procedures.
The downloaded data is stored on the computer’s hard drive – or removable
media of the operator’s choosing – as a comma-separated-value ASCII text
file with a CSV extension. This type of file can be opened by most spreadsheet programs for analysis.
Computer requirements:
• Windows 98SE or higher
• CD ROM drive
• USB 1.1 or USB 2.0 port
• 12 MB of hard drive space for the PCA 2 data recovery program, plus
up to an additional 350 KB for each download
Instruction 24-94484-21
Operation
4.16.1 PCA 2 Data Recovery Program Installation
The PCA 2 Data Recovery Program is supplied with the analyzer on a CD
(P/N 24-1448). Install this program as follows:
1. Insert the PCA 2 CD into the computer’s CD-ROM drive.
2. Locate the CD-ROM drive in Windows Explorer and open the PCA2
folder. Double-click the Setup.exe program to start the installation
process.
3. Click Next on the “Welcome to the InstallShield Wizard for PCA2”
window.
4. Click Next to accept the default destination folder, or click Change to
install to a different folder.
Default
Installation Folder
Click to change
installation folder
Instruction 24-94484-22
Operation
5. Click Install to confirm the destination folder and user information.
6. Click Finish after all files have been copied into the destination folder.
7. At this time the PCA2 Data Recovery Program icon should appear
under "Start All Programs menu.
8. This completes the installation of the software. Remove CD from drive.
Instruction 24-94484-23
Operation
4.16.2 Computer to PCA 2 Connection &
USB Device Driver Installation
Connect the USB cable (P/N 104-4032) that was supplied with the PCA 2,
and, if necessary, install the analyzer’s USB device driver as follows:
1. Insert the PCA 2 CD into the computer's CD-ROM drive.
2. With both the PCA 2 and computer turned ON, connect the PCA 2 to
the computer using the USB cable supplied with the instrument (see
Figure 4-2 in instruction manual)
3.. If this is the first time the PCA 2 is being connected to the computer,
then the “Found New Hardware Wizard” should shortly appear. Select
“No, not this time” and click Next.
NOTE: The USB
device driver
only needs to be
installed once. It
does not require
to be re-installed
each time the
PCA 2 is connected to the
computer.
Figure 4-2. Computer to PCA 2 Connection
Instruction 24-94484-24
Operation
4. Select “Install the software automatically (Recommended)” and click
Next.
Name of new USB
hardware
Select (Advanced).
5. At the Hardware Installation window, click Continue Anyway.
6. After the Found New Hardware Wizard has finished installing the
software, click Finish to close the Wizard.
Instruction 24-94484-25
Operation
4.16.3 Recovering Data
Before data can be recovered from the analyzer, install the PCA 2 Data
Recovery Program, USB cable, and USB device driver as described in Sections 4.14.1 and 4.14.2.
Recover either the logging or combustion data as follows:
1. Start the program by either double-clicking the PCA2 icon on
the Windows deskt op, or click ing the PCA 2 shortcut in
start menu. The “Bacharach PCA 2 Data Recovery
Program” window should appear.
Observe that the window is divided into two sections. The left-hand
section is used to recover logging data, while combustion data is recovered from the right-hand section.
In the following steps, the logging data will be recovered. The same
procedure can be used to recover combustion data using the righthand section of the window.
2. Select where the downloaded file containing the recovered data will
be located and give it a filename by clicking the Open Log File button.
The following example screens show that the downloaded file will be
placed into a pre-existing PCA2 Logging Data folder and given the
filename Customer XYZ.
Instruction 24-94484-26
Operation
Use this section
of window
to Recover
Logging Data
Status line
showing PCA 2
is successfully
connected to
the computer
Select a
pre-existing
folder or create
a new folder
to store the
downloaded le
Enter lename
File type preset
to *.cvs
Use this section
of window
to Recover
Combustion
Data
Start the
download
process by
clicking the
desired Open Log / Data File
button
Click here to
create a new
folder
Click Open after
selecting a folder
and entering a
lename
Location of
downloaded le
Click Start
to begin the
download
process
Total number
of records
received out of
512 total
Status line
showing les
are currently
being received
Instruction 24-94484-27
Operation
NOTE: The filename is automatically given a CSV (Comma
Separated Value) extension, allowing the file to be directly
opened by most spreadsheet programs for analysis.
3. After selecting a folder and entering the filename, click Open and then
click Start to begin the download process.
While recovering the stored data, observe that the “Records received”
line displays the total number of records currently downloaded. And
after all records have been received, the Status line will read “Log files
finished.” The PCA 2 can download 500 combustion, pressure, and
temperature records from memory, and an additional 500 records from
logged memory.
4. This completes the process of recovering logging data. Either click Exit
to close the program, or proceed to recover the combustion data by
starting over at Step 2, but this time using the right-hand section of
the “Bacharach PCA 2 Data Recovery Program” window.
NOTE: Once either the logging data or combustion data has
been recovered, that portion of the recovery program’s window
will turn gray.
Status line
showing
recovery
process has
nished
Click Exit
to end the
program
Instruction 24-94484-28
Operation
4.17 Importing Saved Data Into a Spreadsheet
Data that was recovered and saved as an ASCII text file with a “CSV”
extension, as described in Section 4.14, can easily be opened for viewing
in most spreadsheet programs by simply double-clicking the filename. For
example: double-clicking the filename Customer XYZ.csv should automatically open the spreadsheet program and display the contents of the file.
If the spreadsheet program does not recognize the “CSV” file extension,
then refer to the spreadsheet’s documentation for information on how to
manually import comma-delimited text files.
Table 4-2 contains a listing and description of the 27 data fields that are
downloaded with each data record. Note that for sensors that are not
installed, all related data fields for those sensors will be marked as “N/A”
(Not Applicable). For example: if the analyzer does not contain an SO2 sensor, then “N/A” will appear in data fields 19 and 20.
Instruction 24-94484-29
Operation
TABLE 4-2. DOWNLOADED DATA FIELDS
Field Column NameData Name or Value
1DateDate of Test in mm/dd/yy
2TimeTime of Test in 12 hour format (AM/PM)
3-5Test ID Information Input by User
6FuelFuel Name
7%O2Oxygen Level
8COCarbon Monoxide Level (see field 26)
9ppm CO(O2)Carbon Monoxide Level referenced to a
% of O2 (see field 21)
10%EfficiencyCombustion Efficiency
11%CO2Carbon Dioxide Level
12TStkStack Temperature
13TAirPrimary / Ambient Air Temperature
14TempUnitTemperature Units
15%EAExcess Air
16NONitric Oxide Level (see field 26)
17ppm NO(O2)Nitric Oxide Level referenced to a % of O
(see field 22)
18NO2Nitrogen Dioxide Level (see field 26)
19ppm NO2(O2)Nitrogen Dioxide Level referenced to a
% of O
(see field 22)
2
20NOxNitrogen Oxides (NO + NO2) (see field 26)
21ppm NOx(O2)Nitrogen Oxides Level referenced to a
% of O
(see field 22)
2
22SO2Sulfur Dioxide (see field 26)
23ppm SO2(O2)Sulfur Dioxide Level referenced to a
% of O
24CO_O2Ref%O2 Reference for CO
25NOx_O2Ref%O2 Reference for NOx
26SO2 _O2Ref%O2 Reference for SO
(see field 23)
2
(1)
(1)
(1)
2
27PressurePressure (Draft) Value
28Pressure UnitsPressure Units of Measure
29Pollution UnitsPollution Units of Measure
(2)
(3)
30PS VoltagePower Supply Voltage
31NO TempNitric Oxide Sensor Temperature (4)
2
Instruction 24-94484-30
Operation
(1) %O2 reference as selected per Section 3.12
(2) Pressure units as selected per Section 3.8
(3) Pollution units as selected per Section 3.9
(4) Shown only on printout, not on LCD display.
4.18 Printing Test Data
Combustion or pressure data that is currently being displayed can be sent
to a printer using IrDA protocol as described below.
Data that is stored in memory can also be printed by first displaying the
stored test data as described in Sections 4.13.1 & 4.13.2.
In addition to printing combustion and pressure data, the contents of any
screen that shows the label “PRINT” above the F1 button can be printed.
For example, the information shown in the DIAGNOSTIC screen can be
printed.
1. Turn ON printer. Refer to the printer’s instruction manual for detailed
operating information. If not already done, set up the printer for:
• 8 bit
• No parity
• 9600 baud
• IrDA is set to IrDA-SIR
• DTR handshaking
2. Align the printer with the top of the analyzer as shown in Figure 4-6.
3. Press the PRINT (F1) button to begin printing.
The printout shown in Figure 4-6 shows typical combustion-test results of an analyzer containing O2, CO
that since the CO
and SO2 sensors are not installed, three stars
HIGH
, NO, and NO2 sensors. Note
LOW
(∗∗∗) appear in their data fields.
Instruction 24-94484-31
Operation
BACHARACH, INC.
PCA 2
SN: xxxxxx
=====================
TIME 01:00:00 PM
DATE 10/06/06
Optional
pressure
printout as
selected per
Section 3.13.
Figure 4-3. Printer Alignment & Sample Printout
NOTES:
Three stars (∗∗∗) appear in the data
eld of sensors that are not calibrated
or installed, and in the data elds of
that sensor’s related calculated values.
Three hyphens or dashes (- - -) indicate
that the calculated data cannot be
displayed because the measured data
necessary to make the calculation is
out of range (i.e., oxygen level above
16%).
Three (XXX) indicate sensor overrange.
Instruction 24-94484-32
Operation
Instruction 24-94484-33
Calibration
5 Calibration
IMPORTANT: Before performing any calibration procedure,
ensure that fresh batteries are installed or use the optional AC
power adapter. Also ensure that the analyzer is at room temperature and will be sampling fresh air when turned ON.
5.1 Smart Sensors
The PCA 2 uses Bacharach’s new “Smart Sensor” technology, meaning
that the calibration data for each sensor is stored in nonvolatile memory
on the sensor’s printed circuit board.
Benefits of Smart Sensors:
New sensors can be installed without needing to be calibrated.
•
Sensors can be pre-calibrated and installed when needed.
•
Sensors can be moved from one analyzer to another.
•
The analyzer’s diagnostics screen shows what sensors are installed,
•
and their current end-of-life condition (Good, Low, Bad).
Each sensor stores its calibration history and operating parameters,
•
which can be downloaded to a computer and analyzed. This information is useful in determining when the sensor was last calibrated, and
predicting its end-of-life.
The smart sensors should be calibrated by an authorized Bacharach
Service Center (Section 8.3) every 6 months to 1 year to assure that the
analyzer continues to meet its published accuracy specifications. The
smart sensors, however, can be calibrated in the field if your facility has
the necessary equipment and qualified personnel to perform the procedures described in the following sections of this instruction manual.
5.2 Starting a Calibration
Start any calibration procedure by doing the following:
1. Place the probe in an area of fresh air, turn ON the analyzer, allowing
it to cycle through its 60 second warm-up period. During warm-up,
the analyzer’s operation is checked and the sensors are set to the following ambient conditions:
Oxygen sensor spanned to 20.9%
•
All gas sensors are zeroed
•
The pressure sensor is zeroed
•
Instruction 24-94485-1
Calibration
ERRORS DETECTED
Low Battery
MENU
F1F2F3
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1F2F3
MENU
Calibration Password
Enter Password
Menu
F1F2F3
Any errors detected during warm-up will
be listed on the display immediately following warm-up. For example, the screen
to the right shows that the battery is low.
Correct any errors before proceeding. Refer
to Section 7.3 for a listing of error messages
and their meaning.
2. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
3. Use the buttons to highlight
CALIBRATION, and then press ENT to display the CALIBRATION PASSWORD screen.
4. Before calibration can begin a 4-place alphanumeric password must be entered. Use the
and buttons to enter the password.
(Note that the default password is 1111)
5. Press the ENT key to accept the password.
If the correct password was entered the
CALIBRATION Menu will be displayed.
6. Use the buttons to highlight the sensor to
be calibrated, and then perform the calibration procedure for that sensor as described in
the following sections.
5.3 Pressure Sensor
Calibration
This procedure calibrates the pressure sensor to a known pressure value.
Material Required:
Bellows
•
Manometer
•
Range: ±8 in. of water column (±20 mb)
Accuracy: ±0.01 in. of water column (±0.025 mb)
-
Procedure:
NOTE: The unit-of-measure for pressure is selected per
selected, but note that any unit-of-measure can
be used for calibration purposes.
1. Assemble the pressure sensor calibration
equipment as shown in Figure 5-1, but DO
NOT connect the analyzer to the calibration
equipment at this time.
2. If not already done, turn ON the analyzer and
display the CALIBRATION LIST screen per
Section 5.2.
3. Use the buttons to highlight Pressure,
and then press ENT to display the
CALIBRATE PRESSURE screen.
“Measured” is the pressure value currently
being detected by the pressure sensor, while
“Applied” is a known value of pressure that
will be applied for calibration purposes.
4. With both the –∆P and +∆P ports open to the
atmosphere, observe that the current Measured pressure reading should
be 0 ±0.01 inwc. If necessary, zero the pressure sensor per Section 4.6, and
repeating Steps 2 thru 4.
Calibration
5. Connect the hose from the manometer to the +∆P port and apply a
negative pressure to this port by adjusting the bellows for a manometer reading of –4.00.
6. Use the and buttons to enter an Applied value that exactly
equals the manometer reading.
Instruction 24-94485-3
Figure 5-1. Pressure Sensor Calibration Equipment
Calibration
CALIBRATION MENU
Pressure
T-Stack
T-Air
CO-LO
SO2
NO
NO2
(more)
MENU
F1F2F3
The calibration range is from –6 to –2 inwc (–15 to –5 mb). An attempt
to calibrate outside this range will cause the message “Applied Value
High” (or Low) to appear at the bottom of the screen.
7. Wait until the Measured reading stabilizes, and then press ENT to calibrate the pressure sensor’s Measured value to that of the Applied value;
after which the message “Good Calibration” should briefly appear followed by the CALIBRATION LIST screen being re-displayed.
8. Remove calibration equipment.
5.4 T-Stack Calibration
This procedure first zeros and then spans the stack-temperature channel
to known temperature values.
The use of an electronic thermocouple simulator is the preferred method of
producing the desired calibration temperatures. Alternatively, containers
of ice water and boiling water can be used.
1. Set thermocouple simulator to room temperature and plug its output into the T-STACK
connector located at the bottom of the
analyzer.
Alternatively: Plug the probe’s thermocouple
into the T-STACK connector located at the
bottom of the analyzer. DO NOT attach the
probe’s gas hose to the analyzer’s GAS
port; otherwise water will be drawn into
the analyzer!
2. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen per Section 5.2.
3. Use the buttons to highlight T-Stack, and then press ENT to display the CALIBRATE TS-ZERO screen.
“Measured” is the current temperature reading, while “Applied” is a
Instruction 24-94485-4
known temperature that will be applied for
CALIBRATE TS-ZERO
Measured 34 °F
Applied 032 °F
Press ENT
to calibrate
06/27/06 11:55:30 PM
PRINT
F1F2F3
CALIBRATE TS-SPAN
Measured 574 °F
Applied 572 °F
Press ENT
to calibrate
06/27/06 11:55:30 PM
PRINT
F1F2F3
calibration purposes.
4. Set thermocouple simulator to 32 °F (0 °C),
and then use the and buttons to enter
an Applied value that exactly equals the setting of the simulator.
Alternatively: Submerge probe tip into an ice-
water bath with a thermometer, wait several
minutes, and then use the and buttons
to enter an Applied value that exactly equals
the thermometer reading.
The calibration range is from 32 to 41 °F (0
to 5 °C). An attempt to calibrate outside this
range will cause the message “Applied Value
High" (or Low) to appear at the bottom of the
screen.
5. Wait until the Measured reading stabilizes,
and then press ENT to calibrate the TS-Zero
Measured value to that of the Applied value;
after which the message “Good Calibration” should briefly appear followed by the CALIBRATE TS-SPAN screen.
Calibration
TS-Span Procedure:
6. Set thermocouple simulator to 572 °F (300 °C), and then use the
and buttons to enter an Applied value that exactly equals the
setting of the simulator.
Alternatively: Submerge probe tip into a container of boiling water with
a thermometer, wait several minutes, and then use the and
buttons to enter an Applied value that exactly equals the thermometer
reading.
The calibration range is from 175 to 625 °F (80 to 330 °C). An attempt
to calibrate outside this range will cause the message “Applied Value
High” (or Low) to appear at the bottom of the screen.
7. Wait until the Measured reading stabilizes, and then press ENT to calibrate the TS-Span Measured value to that of the Applied value; after
which the message “Good Calibration” should briefly appear followed by
the CALIBRATION LIST screen being re-displayed.
Instruction 24-94485-5
Calibration
CALIBRATION MENU
Pressure
T-Stack
T-Air
CO-LO
SO2
NO
NO2
(more)
MENU
F1F2F3
CALIBRATE TA-ZERO
Measured 33 °F
Applied 032 °F
Press ENT
to calibrate
06/27/06 11:55:30 PM
PRINT
F1F2F3
5.5 T-Air Calibration
This procedure first zeros and then spans the ambient-temperature channel to known temperature values.
The use of an electronic thermocouple simulator is the preferred method of
producing the desired calibration temperatures. Alternatively, containers
of ice water and boiling water can be used.
1. Set thermocouple simulator to room temperature and plug its output
into the T-AIR connector located at the bottom of the analyzer.
Alternatively: Plug the probe’s thermocouple into the T-AIR connector
located at the bottom of the analyzer. DO NOT attach the probe’s
gas hose to the analyzer’s GAS port; otherwise water will be
drawn into the analyzer!
2. If not already done, turn ON the analyzer and
display the CALIBRATION LIST screen per
Section 5.2.
3. Use the buttons to highlight T-Air, and
then press ENT to display the CALIBRATE
TA-ZERO screen.
“Measured” is the current temperature read-
ing, while “Applied” is a known temperature
that will be applied for calibration purposes.
4. Set thermocouple simulator to 32 °F (0 °C),
and then use the and buttons to enter
an Applied value that exactly equals the setting of the simulator.
Alternatively: Submerge probe tip into an ice-
water bath with a thermometer, wait several
minutes, and then use the and buttons
to enter an Applied value that exactly equals
the thermometer reading.
Instruction 24-94485-6
Calibration
CALIBRATE TA-SPAN
Measured 210 °F
Applied 212 °F
Press ENT
to calibrate
06/27/06 11:55:30 PM
PRINT
F1F2F3
The calibration range is from 32 to 41 °F (0 to 5 °C). An attempt to
calibrate outside this range will cause the message “Applied Value
High” (or Low) to appear at the bottom of the screen.
5. Wait until the Measured reading stabilizes, and then press ENT to
calibrate the TA-Zero Measured value to that of the Applied value;
after which the message “Good Calibration” should briefly appear followed by the CALIBRATE TA-SPAN screen.
TA-Span Procedure:
6. Set thermocouple simulator to 212 °F
(100 °C), and then use the and buttons
to enter an Applied value that exactly equals
the setting of the simulator.
Alternatively: Submerge probe tip into a con-
tainer of boiling water with a thermometer,
wait several minutes, and then use the
and buttons to enter an Applied value that
exactly equals the thermometer reading.
The calibration range is from 194 to 230 °F (90 to 110 °C). An attempt
to calibrate outside this range will cause the message “Bad Calibration
Wrong CAL Entry” to appear in the following step.
7. Wait until the Measured reading stabilizes, and then press ENT to
calibrate the TA-Span Measured value to that of the Applied value;
after which the message “Good Calibration” should briefly appear followed by the CALIBRATION LIST screen being re-displayed.
Instruction 24-94485-7
Calibration
CALIBRATION MENU
Pressure
T-Stack
T-Air
CO-LO
SO2
NO
NO2
(more)
MENU
F1F2F3
CALIBRATE CO
Measured 480 PPM
Applied 500 PPM
Press ENT
to calibrate
06/27/06 11:55:30 PM
PRINT
F1F2F3
5.6 CO-LO Sensor Calibration
Note that the CO-LO sensor also measures H2 for the purpose of compensating the CO reading for the presence of H2 in the gas sample. This
procedure first spans the CO-LO sensor and, optionally, spans the H2 part
of the sensor to known gas levels.
Material Required:
Calibration Kit, P/N 24-7059
•
Gas Cylinder: 500 ppm CO in air, P/N 24-0492
•
Gas Cylinder: 1,000 ppm CO & 1,000 ppm H2 in Nitrogen, P/N 24-0794
•
Procedure:
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen per Section 5.2.
2. Use the buttons to highlight CO-LO, and
then press ENT to display the CALIBRATE
CO screen.
“Measured” is the current CO reading, while
“Applied” is a known CO level that will be
applied for calibration purposes.
3. Attach a 500 ppm CO cylinder to the regulator of the calibration fixture shown in
Figure 5-2 on Page 5-14.
4. Use the and buttons to enter an
Applied value that exactly equals the concentration stamped on the CO cylinder.
The calibration range is from 250 to
1,000 ppm. An attempt to calibrate outside
this range will cause the message “Applied
Value High” (or Low) to appear at the bottom
of the screen.
5. Adjust the regulator for a flowmeter indication of approximately 2 SCFH. Wait until the
Measured reading stabilizes (approximately
3 minutes), and then press ENT to calibrate the CO Measured value
to that of the Applied value. The message “Good Calibration” should
briefly appear followed by the TEST GAS CO/H2 screen.
If the sensor’s output is low, but still usable, then the message “Good
Calibration WARNING Low Sensor” will appear. The sensor will now
be marked as being Low in the DIAGNOSTICS screen.
Instruction 24-94485-8
Calibration
TEST GAS CO/H2
Enter CO VALUE
1000 PPM
06/27/06 11:55:30 PM
PRINT
F1F2F3
CALIBRATE H2
Measured 1000 PPM
Applied 1000 PPM
Press ENT
to calibrate
06/27/06 11:55:30 PM
PRINT
F1F2F3
If the sensor’s output is too low to be usable, then the message “Bad
Calibration Sensor End of Life, Entry Not Saved” will appear. The sensor will now be marked as being BAD in the DIAGNOSTICS screen.
NOTE: H2 calibration can be bypassed by pressing the ESC
button, after which the CALIBRATION LIST screen is re-displayed. Skip to Step 10 if the ESC button was pressed.
6. Turn OFF the regulator of calibration fixture and remove the CO
cylinder.
7. Attach a combination 1,000 ppm CO and
1,000 ppm H2 cylinder to the regulator of the
calibration fixture, and then use the and
buttons to enter an Applied value that
exactly equals the CO concentration stamped
on the cylinder.
8. Adjust the regulator for a flowmeter indication of approximately 2 SCFH, and then press
ENT to display the CALIBRATE H
screen.
2
The calibration range is from 400 to
1,500 ppm. An attempt to calibrate outside this range will cause the
message “Bad Calibration Wrong CAL Entry” to appear in the following step.
9. Wait until the Measured reading stabilizes
(approximately 3 minutes), and then press
ENT to calibrate the H
Measured value to
2
that of the Applied value; after which the
message “Good Calibration” should briefly
appear followed by the CALIBRATION LIST
screen being re-displayed.
If the sensor’s output is low, but still
usable, then the message “Good Calibration
WARNING Low Sensor” will appear. The sensor will now be marked
as being Low in the DIAGNOSTICS screen.
If the sensor’s output is too low to be usable, then the message
“Bad Calibration Sensor End of Life” will appear followed by the
CALIBRATION LIST screen being re-displayed. The sensor will now
be marked as being BAD in the DIAGNOSTICS screen.
10. Turn OFF the regulator and remove the gas cylinder.
Instruction 24-94485-9
Calibration
CALIBRATION MENU
Pressure
T-Stack
T-Air
CO-LO
SO2
NO
NO2
(more)
MENU
F1F2F3
CALIBRATE SO2
Measured 98 PPM
Applied 100 PPM
Press ENT
to calibrate
06/28/06 11:55:30 PM
PRINT
F1F2F3
5.7 SO2 Sensor Calibration
This procedure spans the optional sulfur dioxide sensor to a known
gas level.
Material Required:
Calibration Kit, P/N 24-7059
•
Gas Cylinder: 100 ppm SO2 in Nitrogen, P/N 24-1158
•
Procedure:
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen per Section 5.2.
2. Use the buttons to highlight SO2, and then
press ENT to display the CALIBRATE SO2
screen.
“Measured” is the current SO2 reading, while
“Applied” is a known SO2 level that will be
applied for calibration purposes.
3. Attach a 100 ppm SO2 cylinder to the regulator of the calibration fixture shown in
Figure 5-2 on Page 5-14.
4. Use the and buttons to enter an
Applied value that exactly equals the concentration stamped on the SO2 cylinder.
The calibration range is from 9 to 150 ppm.
An attempt to calibrate outside this range will
cause the message “Bad Calibration Wrong
CAL Entry” to appear in the following step.
5. Adjust regulator for a flowmeter indication
of approximately 2 SCFH. Wait until the
Measured reading stabilizes (approximately
3 minutes), and then press ENT to calibrate the SO2 Measured value
to that of the Applied value. The message “Good Calibration” should
briefly appear followed by the CALIBRATION LIST screen being
re-displayed.
If the sensor’s output is low, but still usable, then the message “Good
Calibration WARNING Low Sensor” will appear. The sensor will now
be marked as being Low in the DIAGNOSTICS screen.
Instruction 24-94485-10
Calibration
CALIBRATION MENU
Pressure
T-Stack
T-Air
CO-LO
SO2
NO
NO2
(more)
MENU
F1F2F3
CALIBRATE NO
Measured 245 PPM
Applied 250 PPM
Press ENT
to calibrate
06/28/06 11:45:30 PM
PRINT
F1F2F3
If the sensor’s output is too low to be usable, then the message “Bad
Calibration Sensor End of Life, Entry Not Saved” will appear followed
by the CALIBRATION LIST screen being re-displayed. The sensor will
now be marked as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove gas cylinder.
5.8 NO Sensor Calibration
This procedure spans the optional nitric oxide sensor to a known gas level.
Material Required:
Calibration Kit, P/N 24-7059
•
Gas Cylinder: 250 ppm NO in Nitrogen, P/N 24-1156
•
Procedure:
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen per Section 5.2.
2. Use the buttons to highlight NO, and then
press ENT to display the CALIBRATE NO
screen.
“Measured” is the current NO reading, while
“Applied” is a known NO level that will be
applied for calibration purposes.
3. Attach a 250 ppm NO cylinder to the regulator of the calibration fixture shown in
Figure 5-2 on Page 5-14.
4. Use the and buttons to enter an
Applied value that exactly equals the concentration stamped on the NO cylinder.
The calibration range is from 9 to 550 ppm.
An attempt to calibrate outside this range will
cause the message “Bad Calibration Wrong
CAL Entry” to appear in the following step.
5. Adjust regulator for a flowmeter indication
of approximately 2 SCFH. Wait until the
Measured reading stabilizes (approximately 3 minutes), and then
press ENT to calibrate the NO Measured value to that of the Applied
value. The message “Good Calibration” should briefly appear followed
Instruction 24-94485-11
Calibration
CALIBRATION MENU
(more)
T-Air
CO-LO
SO2
NO
NO2
CO-HI
T-Ref
MENU
F1F2F3
CALIBRATE NO2
Measured 96 PPM
Applied 100 PPM
Press ENT
to calibrate
06/28/06 11:45:30 AM
PRINT
F1F2F3
by the CALIBRATION LIST screen being re-displayed.
If the sensor’s output is low, but still usable, then the message “Good
Calibration WARNING Low Sensor” will appear. The sensor will now
be marked as being Low in the DIAGNOSTICS screen.
If the sensor’s output is too low to be usable, then the message “Bad
Calibration Sensor End of Life, Entry Not Saved” will appear followed
by the CALIBRATION LIST screen being re-displayed. The sensor will
now be marked as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove gas cylinder.
5.9 NO2 Sensor Calibration
This procedure spans the optional nitrogen dioxide sensor to a known
gas level.
Material Required:
Calibration Kit, P/N 24-7059
•
Gas Cylinder: 100 ppm NO2 in Nitrogen, P/N 24-1157
•
Procedure:
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen per Section 5.2.
2. Use the buttons to highlight NO2, and
then press ENT to display the CALIBRATE
NO2 screen.
“Measured” is the current NO2 reading, while
“Applied” is a known NO2 level that will be
applied for calibration purposes.
3. Attach a 100 ppm NO2 cylinder to the regulator of the calibration fixture shown in
Figure 5-2 on Page 5-14.
4. Use the and buttons to enter an
Applied value that exactly equals the concentration stamped on the NO2 cylinder.
The calibration range is from 9 to 150 ppm.
An attempt to calibrate outside this range will
cause the message “Bad Calibration Wrong
CAL Entry” to appear in the following step.
Instruction 24-94485-12
Calibration
CALIBRATION MENU
(more)
T-Air
CO-LO
SO2
NO
NO2
CO-HI
T-Ref
MENU
F1F2F3
5. Adjust regulator for a flowmeter indication of approximately 2 SCFH.
Wait until the Measured reading stabilizes (approximately 3 minutes),
and then press ENT to calibrate the NO2 Measured value to that of the
Applied value. The message “Good Calibration” should briefly appear
followed by the CALIBRATION LIST screen being re-displayed.
If the sensor’s output is low, but still usable, then the message “Good
Calibration WARNING Low Sensor” will appear. The sensor will now
be marked as being Low in the DIAGNOSTICS screen.
If the sensor’s output is too low to be usable, then the message “Bad
Calibration Sensor End of Life, Entry Not Saved” will appear followed
by the CALIBRATION LIST screen being re-displayed. The sensor will
now be marked as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove gas cylinder.
5.10 CO-HI Sensor Calibration
This procedure spans the optional carbon monoxide high sensor (4,001 to
20,000 ppm) to a known gas level.
Material Required:
Calibration Kit, P/N 24-7059
•
Gas Cylinder: 4,000 ppm CO in Air, P/N 24-1155
•
Procedure:
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen per Section 5.2.
2. Use the buttons to highlight CO-HI, and then press ENT to display
the CALIBRATE CO-HI screen.
“Measured” is the current CO reading, while “Applied” is a known CO
level that will be applied for calibration purposes.
3. Attach a 4,000 ppm CO cylinder to the
regulator of the calibration fixture shown in
Figure 5-2 on Page 5-14.
4. Use the and buttons to enter an
Applied value that exactly equals the concentration stamped on the CO cylinder.
The calibration range is from 250 to
11,000 ppm. An attempt to calibrate outside
Instruction 24-94485-13
Calibration
Parts:
1. Gas Cylinder
2. Regulator *
3. 3/16" I.D. Tubing *
4. Tee *
5. Flowmeter *
6. Barbed Hose Connector *
* Part of Calibration Kit 24-7059
this range will cause the message “Bad Calibration Wrong CAL Entry”
to appear in the following step.
5. Adjust regulator for a flowmeter indication of approximately 2 SCFH. Wait until
the Measured reading stabilizes (approximately 3 minutes), and then press ENT to
calibrate the CO Measured value to that
of the Applied value. The message “Good
Calibration” should briefly appear followed
by the CALIBRATION LIST screen being
re-displayed.
If the sensor’s output is low, but still usable, then the message “Good
Calibration WARNING Low Sensor” will appear. The sensor will now
be marked as being Low in the DIAGNOSTICS screen.
If the sensor’s output is too low to be usable, then the message “Bad
Calibration Sensor End of Life, Entry Not Saved” will appear followed
by the CALIBRATION LIST screen being re-displayed. The sensor will
now be marked as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove gas cylinder.
Figure 5-2. Gas Sensor Calibration Equipment
Instruction 24-94485-14
Maintenance
6 Maintenance
Customer maintenance of the PCA 2 is limited to the following:
Battery replacement or charging using AC power adapter (Section 3.2)
•
Sensor re-calibration (Section 5)
•
Water trap / filter assembly maintenance (Section 6.2)
•
Sensor replacement (Section 6.3)
•
Probe thermocouple replacement (Section 6.6)
•
Pump (gas or purge) replacement (Section 6.7)
•
Cleaning the probe (Section 6.8)
•
All other maintenance should be performed by an authorized Bacharach
Service Center. Refer to Section 8.3.
Instruction 24-94486-1
Maintenance
Rear Case Screws
(typical 4 places)
Release and then
remove battery
cover by pushing
latch up
6.1 PCA 2 Disassembly
The following procedure describes how to disassemble the analyzer, while
Figures 6-1 thru 6-4 illustrate how the analyzer is put together.
Tools Required:
Medium Phillips Screwdriver
•
Procedure:
1. Unplug all thermocouples from
bottom of analyzer.
2. Remove battery cover and then
remove batteries.
TIP: In Step 3, if the sensors are
not being replaced, leave the tubing
connected to each sensor’s gas cap,
being careful not to put unnecessary strain on the tubing during the
disassembly process.
3. Pull off sensor retainer; and
Figure 6-1. Removing Rear Case
then unplug all sensors.
4. Lay analyzer face down on a padded work surface; and then using a
medium Phillips screwdriver, remove the unit’s four rear-case screws.
5. Lift rear case from analyzer and set aside.
6. Unplug electrical connectors J8, J9, and J14 from printed circuit
board.
CAUTION: In Step 7, note that there is tubing connected
between the bottom hose-connector plate and the pressure sensor
on the printed circuit board. Do not put unnecessary strain on
this tubing during the disassembly process.
7. Carefully lift battery-and-pump chassis, along with the bottom hose-
8. Lift printed circuit board from analyzer.
connector plate, from analyzer.
Instruction 24-94486-2
Sensor Positions
Sensor Retainer
Purge Pump
Gas Pump
GAS
Purge Air In
Pressure
+∆P
–∆P
Battery & Pump
Chassis
Hose Connector
Plate
Purge Pump
Motor Connector
Gas Pump
Motor Connector
Sensors
Type and position
determined by
model number
Battery Connector
on printed circuit board
Pressure Sensor
on printed circuit board
+∆P (Top)
–∆P (Bottom)
Model
P/N
Position
#1O
#2COLOCOLOCOLOCOLOCOLOCO
#3SO2NO2SO
#4NOCO
225
24-7301
2
235
24-7302
O
2
245
24-7303
O
255
24-7304
2
HI
265
24-7305
O
2
275
24-7306
O
O
2
NONO
Maintenance
2
LO
2
Figure 6-2. Inside View with Rear Case Removed
Instruction 24-94486-3
Maintenance
Models 235 & 245
P/Ns 24-7302 & 24-7303
Model 255
P/N 24-7304
Models 265 & 275
P/Ns 24-7305 & 24-7306
Model 225
P/N 24-7301
Elbow
PRESSURE SENSOR
on printed circuit board
–∆P
+∆P
–∆P (Bottom)
+∆P (Top)
PRESSURE
PURGE AIR
GAS
Gas IN
GAS PUMP
PURGE PUMP
Gas OUT
OUTIN
To Sensors
Elbow
Elbow
Elbow
Figure 6-3. Tubing Connections
Instruction 24-94486-4
Maintenance
Solder
(2 Places)
Red dot on pump
indicates positive
(red wire) connection
Battery
to J14
Battery
to J14
Gas
Pump
to J8
Purge
Pump
to J9
Sensor Connector
(typical 4 places)
Factory Test
Connector
LCD Connector
(opposite side of board)
Ribbon Cable Conection
between LCD Assembly and
Printed Circuit Board
Lock Tabs
(both ends)
J8
J9
J14
Wire Routing
Solder Connections
Connectors
Instruction 24-94486-5
Figure 6-4. Wiring
Maintenance
6.2 Water Trap / Filter Maintenance
6.2.1 Emptying the Water Trap Chamber
The water trap chamber should be emptied after every test, or when the
water condensate approaches the tip of the riser tube (refer to Section 4.9).
1. Remove water trap chamber per Figure 6-5.
2. Pour out liquid condensate, and then reassemble trap.
6.2.2 Replacing the Filter Element
Replace the filter element when it becomes visibly dirty or becomes saturated with water.
Material Required: • Filter Element, P/N 07-1644
• Small Flat Blade Screwdriver
1. Remove water trap chamber per Figure 6-5.
2. Pry apart filter chamber using a small flat-blade screwdriver. Remove
and discard old filter.
3. Install new filter and reassemble filter chamber, making sure that
surfaces “A” and “B” contact each other.
4. Reassemble trap.
Figure 6-5. Disassembling the Water Trap / Filter Assembly
Instruction 24-94486-6
Maintenance
Gas Cup
Sensor Retainer
Sensor
and Printed
Circuit Board
Twist gas cap
bayonet fitting
counterclockwise
to release
Sensor electrical
connector mates with
pins on analyzer’s main
board
6.3 Smart Sensor Replacement
Bacharach’s Smart sensor technology allows new sensors to be installed
without needing to be calibrated inside the analyzer.
NOTE: Sensors may be purchased with and without the Smart sensor technology
PCB. See Section 6.4 on how to change just the sensor itself. Calibration must be
performed if the sensor is purchased without the Smart technology PCB. Refer to
Section 8.1 Replacement Parts for list of sensors and part numbers of both types.
Replace the oxygen sensor when its automatic calibration fails and the analyzer displays an O2 sensor error at the end of warm up. Refer to Section 4.2.
NOTE: The O2 sensor and PCB is not Smart sensor technology. Only the sensor itself, not the PCB, needs to be purchased when the sensor needs replaced.
Replace a gas sensor when it can no longer be calibrated, as evidenced by
“Bad Calibration Sensor End of Life” being displayed during its calibration
procedure. Refer to Sections 5.6 thru 5.10.
Refer to Section 8.1 for a listing of sensor part numbers. Also, see
Figure 6-2 on page 6-3 for sensor positions.
Do the following to replace a Smart sensor:
1. Turn OFF the analyzer.
2. Remove the battery cover.
3. Remove the sensor retainer.
4. Remove the tubing from the gas
cup of sensor being replaced.
5. Unplug the sensor along with
its printed circuit board from
the analyzer’s main board.
6. Remove the sensor’s gas cup
by twisting its bayonet fitting
counterclockwise, and then
pulling the cup straight up.
7. Attach the gas cup to new sensor, making sure that when the sensor
is plugged into analyzer’s main board, the tubing connections on top
of gas cup are aligned as shown in Figure 6-3 on page 6-4.
8. Plug the sensor into analyzer; and then reattach tubing to gas cup.
9. Install the sensor retainer and battery cover.
10. Turn ON the analyzer and confirm that no sensor errors occur during
warm up.
Instruction 24-94486-7
Maintenance
Gas Cup
Sensor Retainer
Sensor
Twist gas cap
bayonet fitting
counterclockwise
to release
Sensor electrical
connector mates with
pins on analyzer’s main
board
Printed
Circuit Board
Adhesive
foam
disk
Be sure to remove any
shorting spring from the
new sensor pins before
installing! There is no
shorting spring on the
O2 or NO sensors.
6.4 Sensor Only Replacement
All sensors are replaced in a similar manner. Do the following to
1.
replace either the O2, CO-LO, CO-HI, NO, NO2, or SO2 sensor. Refer
to Section 8.1 Replacement Parts for list of sensors and part numbers.
Turn OFF the analyzer.
2.
Remove the battery cover.
3.
Remove the sensor retainer.
4.
Remove the tubing from the gas cup of sensor being replaced.
5.
Unplug the sensor along with its printed circuit board from the analyz-
6.
er’s main board.
Remove the sensor’s gas cup by twisting its bayonet fitting counter-
7.
clockwise, and then pulling the cup straight up.
Carefully remove the old sensor from the Smart sensor PCB, keeping
8.
in mind that there is an adhesive foam disk in between the PCB and
sensor. Gently remove any adhesive from from the PCB.
Important! When replacing the O2 sensor, be sure that the “+” sensor
pin plugs into the PCB socket that is also marked “O2+”.
When replacing an NO sensor, also replace the bias battery.
9. Remove the replacement sensor from its plastic canister. Note that
the CO-LO, CO-HI, SO2, and NO2 sensors have a shorting-spring
installed between two of
their pins. Remove and
discard this spring before
installing the sensor.
10. Place the adhesive disk on
the bottom of the new sensor and plug the new sensor into the sensor PCB.
11. Attach the gas cup to new
sensor, making sure that
when the sensor assembly
is plugged into the analyzer’s main board, the tubing
connections on top of gas
cup are aligned as shown
in Figure 6-3 on page 6-4.
Instruction 24-94486-8
Maintenance
12. Plug the sensor into analyzer; and then reattach tubing to gas cup.
13. Install the sensor retainer and battery cover.
14. Allow the sensor that was just installed time to stabilize in the circuit
before continuing with this procedure. Stabilization time for all sensors (except for the NO sensor) is about 1 hour. The NO sensor baseline
technically requires several days to stabilize but should be sufficiently
stabilized for use in approximately 4 hours.
15. Turn ON the analyzer and confirm that no sensor errors occur during
warm up.
Note: Discard an old sensor in accordance with local and federal hazardous waste disposal laws.
16. Calibrate the new sensor(s) per Section 5 (except the O2 sensor which
does not require calibration).
Instruction 24-94486-9
Maintenance
6.5 Nitric Oxide Sensor Battery Replacement
A single lithium battery, located on the NO Smart Sensor assembly, applies
a bias voltage to the NO sensor to prevent the sensor from destabilizing
when the analyzer is turned off. The NO bias battery is expected to last at
least the life of the NO sensor.
Note: It is recommended that the bias battery be replaced whenever the NO
sensor is replaced.
Material Required:
• Bias battery (refer to Section 8.1 Replacement Parts)
Procedure:
1. Follow the instructions in Section 6.3 Smart Sensor Replacement to
remove the Smart Sensor assembly from position 4.
2. Remove the old battery from its holder on the Nitric Oxide printed
circuit board.
3. Insert the new battery with the positive side facing away from the
sensor (the battery contact is stamped with a + symbol) .
4. Re-install the Smart Sensor assembly.
5. Before powering up and using the analyzer, allow the Nitric Oxide
sensor to stabilize as described below. Note that recalibration of the
Nitric Oxide sensor is not required after replacing its bias battery.
Depending on how long the Nitric Oxide sensor was without bias
voltage, the time required for the sensor to completely stabilize varies
from less than a minute to several days. Typical stabilization times
are shown below. Generally, however, the sensor is sufficiently stable
after 4 hours for measurement purposes.
Bias battery removed for . . .Stabilization time
Less than 15 minutesLess than 1 minute
Less than 1 hourLess than 5 minutes
Less than 2 daysLess than 4 hours
Greater than 2 daysUp to 2 days
Instruction 24-94486-10
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.