Bacharach PCA2 Operating Manual

Portable Combustion Analyzer
Instruction 24-9448
Operation & Maintenance
Rev. 4 - July 2009
Product Leadership • Training • Service • Reliability
WARRANTY
Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in material and manufacture and will conform substantially to Bacharach Inc.’s applicable specifications. Bacharach’s liability and Buyer’s remedy under this warranty are limited to the repair or replacement, at Bacharach’s option, of this Product or parts thereof returned to Seller at the factory of manufacture and shown to Bacharach Inc.’s reasonable satisfaction to have been defective; provided that written notice of the defect shall have been given by Buyer to Bacharach Inc. within two (2) years after the date of delivery of this Product by Bacharach, Inc. The Oxygen (O2) sensor has a limited warranty of one (1) year. Routine service and calibration are excluded from this warranty.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s liability and Buyer’s remedy under this warranty of title are limited to the removal of any title defects or, at the election of Bacharach, to the replacement of this Product or parts thereof that are defective in title.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE IM­PLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II) ANY OBLIGATION, LIABILITY, RIGHT, CLAIM OR REMEDY IN CONTRACT OR TORT, WHETHER OR NOT ARIS­ING FROM BACHARACH’S NEGLIGENCE, ACTUAL OR IMPLIED. The remedies of the Buyer shall be limited to those provided herein to the exclusion of any and all other remedies including, without limitation incidental or consequential damages. No agreement varying or extending the foregoing warranties, remedies or this limita­tion will be binding upon Bacharach, Inc. unless in writing, signed by a duly authorized officer of Bacharach.
Register Your Warranty by Visiting
www.bacharach-inc.com
Notice:
Product improvements and enhancements are continuous, therefore the specifications and information contained in this document may change without notice.
Bacharach, Inc. shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of Bacharach, Inc.
Copyright © 2006, Bacharach, Inc., all rights reserved. BACHARACH is a registered trademark of Bacharach, Inc. All other trademarks, trade names, service marks and logos referenced herein belong to their respective companies.
A
Instruction 24-9448
Contents
Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1 PCA 2 General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
1.2 Sales Combo & Model Configurations . . . . . . . . . . . . . . . . . . . . . . 1-2
1.3 Features & Benefits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.4 Operational Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
1.5 Connector Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
1.5.1 Probe Connections (Gas, Pressure, T-Stack) . . . . . . . . . . . .1-6
1.5.2 T-AIR (Primary Air Thermocouple) . . . . . . . . . . . . . . . . . . .1-6
1.5.3 POWER (AC Adapter) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
1.5.4 ∆P (Differential Pressure) . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
1.5.5 USB (Computer Interface) . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
1.5.6 IrDA (Printer Interface) . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
1.5.7 OPT (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
1.6 Front Panel Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
2 Specications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
3 Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1 Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
3.2 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
3.2.1 Installing or Replacing Batteries . . . . . . . . . . . . . . . . . . . . .3-1
3.2.2 Using the AC Power Adapter . . . . . . . . . . . . . . . . . . . . . . . .3-2
3.3 Connecting the Probe and Hose Assembly . . . . . . . . . . . . . . . . . . .3-2
3.4 Operating Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3.5 Fuel Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3.6 Temperature Units Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.7 Pressure Units Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3.8 Pollution Units Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
3.9 Date Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3.10 Time Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3.11 O2 Reference Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
3.12 Print Pressure Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-11
3.13 Zoom-Display Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
3.14 Battery Charger Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
3.15 Logging Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
3.16 Button Sound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
3.17 Test ID Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15
3.18 Username. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-17
3.19 Language. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
3.20 Cal Reminder Period. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-20
3.21 Run/Hold Screen Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-21
Instruction 24-9448
i
Contents
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.1 Operating Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
4.2 Turning ON the Analyzer and Warm Up . . . . . . . . . . . . . . . . . . .4-2
4.3 Selecting a Fuel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4.4 Sampling Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4.5 Performing a Combustion Test . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4.6 Pressure Label Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
4.7 Temperature Measurement Label . . . . . . . . . . . . . . . . . . . . . . . . 4-8
4.8 Making a Draft / Pressure Measurement . . . . . . . . . . . . . . . . . . 4-9
4.9 Saving Test Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
4.10 Ending a Combustion Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-11
4.11 Emptying the Water Trap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-11
4.12 Turning OFF the Analyzer & Purging . . . . . . . . . . . . . . . . . . .4-12
4.13 Low Battery Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-12
4.14 Data Logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
4.14.1 Turning ON Data Logging . . . . . . . . . . . . . . . . . . . . . . . .4-14
4.14.2 Setting the Logging Interval and Duration . . . . . . . . . . .4-15
4.14.3 Starting the Data Logging Process . . . . . . . . . . . . . . . . .4-16
4.14.4 Ending the Data Logging Process . . . . . . . . . . . . . . . . . .4-17
4.15 Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
4.15.1 Recalling Combustion Test Data . . . . . . . . . . . . . . . . . . .4-18
4.15.2 Recalling Logged Test Data . . . . . . . . . . . . . . . . . . . . . . .4-19
4.15.3 Clearing Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
4.16 Downloading Stored Data to a Computer . . . . . . . . . . . . . . . . . .4-21
4.16.1 PCA 2 Data Recovery Program Installation . . . . . . . . . 4-22
4.16.2 Computer to PCA 2 Connection & USB Device
Driver Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
4.16.3 Recovering Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
4.17 Importing Saved Data Into a Spreadsheet . . . . . . . . . . . . . . . . . .4-29
4.18 Printing Test Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-31
5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.1 Smart Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
5.2 Starting a Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
5.3 Pressure Sensor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.4 T-Stack Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5.5 T-Air Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
5.6 CO-LO Sensor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
5.7 SO2 Sensor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-10
5.8 NO Sensor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-11
5.9 NO2 Sensor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
5.10 CO-HI Sensor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-13
ii
Instruction 24-9448
Contents
6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.1 PCA 2 Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
6.2 Water Trap / Filter Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
6.2.1 Emptying the Water Trap Chamber . . . . . . . . . . . . . . . . . 6-6
6.2.2 Replacing the Filter Element . . . . . . . . . . . . . . . . . . . . . . . 6-6
6.3 Smart Sensor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
6.4 Sensor Only Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6.5 Nitric Oxide Sensor Battery Replacement . . . . . . . . . . . . . . . . . .6-10
6.6 Thermocouple Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-11
6.7 Pump Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-13
6.8 Cleaning the Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
7 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.1 Analyzer Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.2 Error Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
7.3 Error Messages Displayed After Warm-Up . . . . . . . . . . . . . . . . . .7-2
7.4 Diagnostics and Status Screens . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
8 Parts & Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8.1 Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
8.2 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
8.3 Service Centers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
Instruction 24-9448
iii
Contents
Notes:
iv
Instruction 24-9448
Introduction
1 Introduction
1.1 PCA 2 General Description
The PCA 2 is a commercial-grade hand-held combustion and emissions analyzer designed for on-demand sampling of light industrial, institu­tional, commercial and residential furnaces, appliances, and boilers. The basic instrument is supplied with a probe and hose assembly, instruction manual, factory calibrated smart sensors, 4 'A A' alkaline batteries, Data Download Software with USB cable and carrying case.
Because of the PCA 2’s ability to measure up to four gases simultaneously, it is the perfect tool for service technicians, inspectors and boiler contrac­tors who need to determine combustion efficiency, excess air, stack gas O2 and CO levels, stack temperature, draft, and differential pressure. The analyzer can also directly measure and display NO, NO2 and SO2 with the installation of the appropriate sensors. Combustion efficiency calculations can be conducted for the following fuels: natural gas, oil #2, oil #4, oil #6, propane, coal, wood, kerosene, bagasse, and digester gas. A large backlit graphical display shows up to eight combustion test values simultaneously, and includes a zoom capability that provides an extra large display.
Smart sensor technology allows a new sensor to be installed in the field without having to calibrate the analyzer before use. New and innovative probe and analyzer designs allow the PCA 2 to be easily serviced, thus lowering the cost of ownership.
Advanced data storage and communication features allow the operator to store up to 500 individual combustion test records, which can later be recalled for viewing, printing, or downloading to a personal computer. In its data logging mode, the analyzer can store an additional 500 data logged records.
An optional AC power adapter allows the analyzer to run continuously for data logging purposes.
The optional sample conditioning probe is recommended when measuring NO2 and SO2 to ensure the highest degree of measurement accuracy.
Instruction 24-9448 1-1
Introduction
1.2 Sales Combo & Model Configurations
Sales Combo 24-8350 24-8351 24-8352
Sales Combo (Kit) 24-8370 24-8371 24-8372
Model Type 225 235 245
PCA2 Only Part Number 24-7301 24-7302 24-7303
Measurements
Oxygen (O2)
Stack Temperature
Primary / Ambient Air Temperature
Carbon Monoxide Low (CO
Low
)
Pressure / Draft
Carbon Monoxide High (CO
High
)
Nitric Oxide (NO)
Nitrogen Dioxide (NO2)
Sulfur Dioxide (SO2)
Calculations
Combustion Efciency
Excess Air
Carbon Dioxide (CO2)
NOx (NOx = NO + NO2)
NOx referenced to %O
CO referenced to %O
NO referenced to %O
NO2 referenced to %O
SO2 referenced to %O
2
2
2
2
2
3 3 3
3 3 3
3 3 3
3 3 3
3 3 3
3
3
3 3 3
3 3 3
3 3 3
3 3 3
3
Refer to Section 8.2 for a listing of standard and optional accessories.
Instruction 24-94481-2
Introduction
Sales Combo 24-8353 24-8354 24-8355
Sales Combo (Kit) 24-8373 24-8374 24-8375
Model Type 255 265 275
PCA2 Only Part Number 24-7304 24-7305 24-7306
Measurements
Oxygen (O2)
Stack Temperature
Primary / Ambient Air Temperature
Carbon Monoxide Low (CO
Low
)
Pressure / Draft
Carbon Monoxide High (CO
High
)
Nitric Oxide (NO)
Nitrogen Dioxide (NO2)
Sulfur Dioxide (SO2)
3 3 3
3 3 3
3 3 3
3 3 3
3 3 3
3 3
3
3 3
Calculations
Combustion Efciency
Excess Air
Carbon Dioxide (CO2)
NOx (NOx = NO + NO2)
NOx referenced to %O
CO referenced to %O
NO referenced to %O
NO2 referenced to %O
SO2 referenced to %O
2
2
2
2
2
3 3 3
3 3 3
3 3 3
3
3
3 3 3
3 3
3
3 3
Instruction 24-9448 1-3
Introduction
1.3 Features & Benefits
• Powered by 4 ‘AA’ alkaline batteries, or NiMH rechargeable batteries. An optional AC power adapter provides continuous operation.
• Internal charging circuit allows rechargeable batteries to be charged inside the analyzer with the use of the optional AC power adapter.
• O2 and CO two additional gases: CO
measurement standard. Optional measurement of up to
Low
, NO, NO2, or SO2.
High
• With the appropriate sensors installed, the analyzer optionally dis­plays pollution conversions for CO, NO, NO2, and SO2. Pollution con­versions include ppm, #/MBTU, mg/m3, and g/GJ.
• Smart sensor technology allows pre-calibrated sensors to be installed in the field.
• Automatic zero of all sensing channels on ambient air when the ana­lyzer is first turned ON.
• Automatic flushing of the CO exceeds 4,000 ppm, thus protecting the CO
sensor with fresh air if the CO level
Low
sensor from high CO
Low
levels. To measure CO levels above 4,000 ppm, the analyzer automati­cally switches to its CO
sensor, if installed.
High
• Automatic purging of the gas-sample system if the detected gas levels are abnormally high when the analyzer is turned OFF.
• Displays temperatures in either °F or °C.
• Displays pressure in either inwc, mb, Pa, or hPa.
• Backlit graphic LCD with zoom capabilities.
• Low battery alarm.
• Stores 500 individual combustion records, which can later be recalled for viewing, printing, or downloading to a personal computer. Stores an additional 500 data logged records.
• Wireless IrDA link used for printing current and stored combustion records, pressure records, sensor calibration data, and diagnostic data.
• USB connectivity for downloading stored data to a personal computer.
• Field replaceable sensors and thermocouple.
• Two year warranty on analyzer and all gas sensors except the O2 sen­sor which has a one (1) year warranty.
Instruction 24-94481-4
Introduction
•Language options including English, French, and Spanish
•Custom Display Formats
•Calibration Reminders - PCA2 can be set up to remind the user that calibration is past due.
1.4 Operational Overview
The PCA 2 is powered by either its 4 internal batteries, or by an optional AC power adapter that operates from any convenient source of 100– 240 VAC, 50/60 Hz power. The type of batteries used can be either dispos­able alkaline or rechargeable NiMH. Note that rechargeable batteries can be charged inside the analyzer using the optional AC power adapter.
The PCA 2 is controlled by 11 front panel push buttons, while a graphical LCD is used to display all combustion and emission test data and analyzer parameters.
A probe and hose assembly, with an integral thermocouple and filter/wa­ter-trap connect to the bottom of the analyzer, thus providing the means of drawing in gas samples, and for measuring stack temperature and draft.
The PCA 2 is turned ON by pressing its red I/O button. A warm-up period of 60 seconds then begins, during which time the analyzer performs self diagnostics. At the end of the warm-up period, if no errors were detected the instrument will display the Combustion Test HOLD screen. If errors were detected, the message “ERRORS DETECTED” is displayed along with a list of the errors. These errors must be corrected before proceeding with the combustion test.
Before starting a test be sure to select the fuel being burned. The default fuel selected is Natural Gas. Note that the name of the fuel being burned is indicated at the top of the display. To change the fuel: first, press the MENU (F2) button; next, select FUEL from the menu; then use the but- tons to highlight the fuel being burned; and finally, press the green ENT button to select the highlighted fuel.
To assure correct combustion-efficiency calculations, the analyzer must know the burner’s primary-air temperature. The analyzer normally uses its internal temperature sensor for the primary-air temperature value, but this method is only acceptable if the burner is using ambient room air. If the burner is drawing in cold outside air, we recommend that the optional T­AIR thermocouple be used. This thermocouple plugs into the bottom of the
Instruction 24-9448 1-5
Introduction
analyzer and is placed in the burner’s primary-air stream.
Begin the combustion test by first inserting the analyzer’s probe tube into the stack-gas stream of the appliance under test, and then pressing the RUN/HOLD button to display the Combustion Test RUN screen. The ana­lyzer will begin to continuously monitor the stack temperature, %O2 and emission levels in the stack gas and then display measured and calculated values on its LCD. Values are listed in Section 2 Specifications.
During a test, the CO
sensor is protected from high CO levels by being
Low
automatically flushed with fresh air when the detected CO level exceeds 4,000 ppm. The analyzer will automatically start using its optional CO-
sensor, if installed, at CO levels starting at 4,001 ppm, thus providing
High
continuous CO readings up to 20,000 ppm.
A backlight enables a user to read the display in dimly-lit areas. Turn the backlight ON and OFF by briefly pressing the I/O button.
The analyzer is turned OFF by pressing and holding down the I/O button for at least 2 seconds. Note that there is a 5-second delay before the analyz­er actually turns OFF, during which time the analyzer can be turned back ON by pressing the RUN/HOLD button. In addition, there is a gas-purge feature that keeps the analyzer’s pump running if the gas level inside the sensor chambers is abnormally high at shutdown. With the probe removed from the stack and sampling fresh air, the analyzer purges itself until the detected gas concentrations drop below predetermined levels.
1.5 Connector Descriptions
1.5.1 Probe Connections (Gas, Pressure, T-Stack)
Attach the probe and hose assembly to the analyzer by connecting its . . .
• stack-gas thermocouple to the analyzer's T-STACK connector,
• stack-gas hose to the analyzer's GAS connector,
• draft hose to the analyzer's +∆P connector.
Observe that the probe connectors are of different sizes and shapes, which prevent incorrect connection to their associated connectors on the analyzer.
Instruction 24-94481-6
Introduction
T-STACK T-AIR
IrDA – Wireless printer communications port
ENT:
Selects a highlighted
menu item, or conrms the
entry of data
RUN / HOLD:
• Starts and stops a combustion
efciency test
• Pressing this button during the 5 second turn-off period will keep the analyzer turned ON
ESC – Displays previous menu or previously viewed screen
I/O – Press from between 1 and 2 seconds to turn analyzer ON and OFF
– With analyzer turned
ON, press briey to turn
backlight ON and OFF
LCD – 160 x 160 graphic display
F1 / F2 / F3 – Soft Menu Buttons, whose functions are
dened by labels
appearing above them on LCD
USB – Computer communications connector
Arrow Buttons:
– Moves cursor up the display, or increments an alphanumerical value
– Moves cursor down the display, or decrements an alphanumerical value
– Moves cursor left, or moves to top of menu list
– Moves cursor right, or moves to bottom of menu list
T-STACK – Probe’s stack-gas thermocouple connector
OPT – Option external measurement connector
T-AIR – Primary /
POWER – AC power adapter
connector
ambient air thermo­couple connector
GAS – Probe’s gas hose connector
“+P” Probe’s draft
hose tting
“–P” Reference
pressure hose tting
Figure 1-1. PCA 2 Components
Instruction 24-9448 1-7
Introduction
1.5.2 T-AIR (Primary Air Thermocouple)
If thermocouple P/N 104-1797 (10 feet long) or Utility Wand P/N 104-1799 (12 inch ridged probe with handle and 5 foot coiled cable) is to be used to measure the burner’s primary air temperature, then connect either of these thermocouples to the analyzer’s T-AIR connector.
1.5.3 POWER (AC Adapter)
The AC power adapter P/N 24-1404 can be used as an external power sup­ply, which will run the analyzer on a continuous basis.
When using rechargeable NiMH batteries, the AC power adapter can also be used to charge the batteries while inside the analyzer. The analyzer’s rapid-charger circuit, however, must first be turned ON per Section 3.16. The rapid charger will charge a set of depleted batteries in approximately 2 - 3 hours.
When disposable alkaline batteries are used, the analyzer’s battery char­ger circuit must be OFF to prevent the batteries from overheating. As a precaution, the charger circuit is automatically toggled back to its OFF state when the analyzer is turned OFF.
1.5.4 ∆P (Differential Pressure)
Draft is measured by connecting the probe’s draft hose to the +P fitting, while leaving the –P fitting open to the atmosphere.
In addition to measuring draft, the “+” and “–” P fittings can also be used to measure the differential pressure between two areas by first connecting a hose P/N 24-1103 to the –P fitting, and then inserting the open end of this hose into the area being used as the reference pressure. The analyzer’s probe is then inserted into the area where differential pressure is to be measured. Refer to Section 4.6.
1.5.5 USB (Computer Interface)
Data that has been stored in the analyzer’s memory can be downloaded to a personal computer by connecting USB data cable P/N 104-4032 between the USB ports of the computer and analyzer. Refer to Section 4.14.2.
1.5.6 IrDA (Printer Interface)
Data that has been stored in the analyzer’s memory can be printed on a
Instruction 24-94481-8
Introduction
compatible IrDA wireless printer by aligning their IrDA communication ports. Refer to Section 4.16.
1.5.7 OPT (Option)
The option connector is used for optional external measurement features.
1.6 Front Panel Buttons
Descriptions of the front panel buttons are given below. Note that a control may perform multiple functions as determined by what screen is being displayed at the time.
The functions of these buttons are defined by labels ap­pearing above them on the LCD. The labels that appear depend on the functions that can be performed in the particular screen being displayed.
PRINT (F1): Transmits the data displayed on the screen to a printer through the IrDA communications port.
MENU (F2): Displays the Main Menu.
SAVE (F3): Saves the data currently displayed on the
LCD in memory. Up to 500 individual Combustion Test and Pressure records can be saved. After 500 records have been saved, the memory must be cleared to continue saving additional data. The analyzer will not overwrite old data.
ZERO (F2): When viewing the Pressure screen, this button zeros the pressure sensor to current atmospheric conditions. When viewing the Temperature screen, this button zeroes the temperature channel difference.
PAGE– (F1): When viewing the Memory or Logging Directory, each press of this button pages down through the directory. Holding this button down speeds up the paging process.
PAGE+ (F3): When viewing the Memory or Logging Directory, each press of this button pages up through the directory. Holding this button down speeds up the paging process.
Instruction 24-9448 1-9
Introduction
The arrow buttons move the cursor on the LCD in the di­rection of the arrow. In screens that require the entry of alphanumerical data, use the buttons to move cursor across the screen and then use the buttons to incre­ment and decrement the data. When viewing a menu, use the buttons to quickly move to the top and bottom of the menu.
Selects a highlighted menu. In addition, if changes were made to one of the analyzer’s operating parameters (e.g., date, time, O2 reference, etc.), pressing this button con­firms those changes and saves them in memory.
Starts and stops a combustion test when the Combus­tion Test screen is displayed. Pressing this button in any other screen returns the analyzer to the Combustion Test HOLD screen. Pressing this button during the 5 second turn-off-delay period will abort the turn-off process and also return the analyzer to the Combustion Test HOLD screen.
Displays a previously viewed screen. In addition, if changes were made to one of the analyzer’s operating parameters (e.g., date, time, O2 reference, etc.), pressing this button aborts those changes, restores the old values, and then displays the previously viewed screen.
Turns the analyzer ON and OFF, and is also used to turn the backlight and button LEDs ON and OFF.
Note that when the analyzer is turned OFF, there is a 5 second delay, during which time an operator can keep the analyzer turned ON by pressing the RUN / HOLD button. Also note that if the measured emission levels are above predetermined limits at the time the instru­ment is turned OFF, the pump is automatically started and purges the sensor compartment with fresh air until the gas levels inside the analyzer are reduced. If desired, the purging process can be aborted by again pressing the I/O button.
Instruction 24-94481-10
Specications
2 Specifications
The PCA 2 Directly Measures and Displays:
The gases displayed depend on the analyzer’s model number. Refer to Section 1.2.
Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 20.9%
Stack Temperature . . . . . . . . . . . . . . . . . . . . –4 to 2,192 ºF (–20 to 1,200 ºC)
Primary / Ambient Air Temperature . . . . . . . . –4 to 999 ºF (–20 to 537 ºC)
Carbon Monoxide (CO) (H2 compensated) . . . . . . . . . . . . . . . .0 to 4,000 ppm
Pressure / Draft . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±72 "H2O (±180 mb)
CO High Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4,001 to 20,000 ppm
Nitric Oxide (NO) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to 3,000 ppm
Nitrogen Dioxide (NO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 500 ppm
Sulfur Dioxide (SO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to 5,000 ppm
The PCA 2 Calculates and Displays:
Calculations are performed only when the measured oxygen level is below
16.0%, and the stack temperature is below 2,000 °F (1,093 ºC).
Combustion Efficiency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.1 to 100%
Excess Air . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 to 250%
Carbon Dioxide (dry basis) . . . . . . . . . . 0.1 to fuel dependent maximum in %
NOx (NOx = NO + NO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 3,500 ppm
NOx referenced to %O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 9,999 ppm
CO referenced to %O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 9,999 ppm
NO reference to %O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 9,999 ppm
NO2 reference to %O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 9,999 ppm
SO2 reference to %O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 9,999 ppm
Fuels Available for Combustion Calculations:
• Natural Gas • Coal
• Oil #2 • Wood
• Oil #4 • Kerosene
• Oil #6 • Bagasse
• Propane • Digester Gas
Instruction 24-9448 2-1
Specications
Normal Operating Conditions:
Temperature:
Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 to 104 ºF (0 to 40 ºC)
Probe Tip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1,472 ºF (800 ºC) Max.
Humidity:
Analyzer . . . . . . . . . . . . . .15 to 90% Relative Humidity, non-condensing
Air Pressure:
Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Atmospheric
Probe . . . . . . . . . . . . . . . . . . . . . 10" H2O (25 mb) draft max. at probe tip
Performance:
Accuracy:
O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .±0.3% O2 on practical concentrations
of stack gas (mix of O2, CO2 and N2)
CO . . . . . . . . . . . . . . . . . . . . . . . . . . ±5% of reading or ±10 ppm,
whichever is greater between 0–2,000 ppm, and ±10% of reading between 2,001–20,000 ppm.
NO . . . . . . . . . . . . . . . . . . . . . . . . . . ±5% of reading or ±5 ppm,
whichever is greater between 0–2,000 ppm
NO2 . . . . . . . . . . . . . . . . . . . . . . . . . ±5% of reading or ±5 ppm,
whichever is greater between 0–500 ppm
SO2 . . . . . . . . . . . . . . . . . . . . . . . . . ±5% of reading or ±10 ppm,
whichever is greater between 0–2,000 ppm
Stack Gas Temp. . . . . . . . . . . . . . . . ±4 ºF between 32 and 255 ºF
(±2 ºC between 0 and 124 ºC) ±6 ºF between 256 and 480 ºF (±3 ºC between 125 and 249 ºC) ±8 ºF between 481 and 752 ºF (±4 °C between 250 and 400 °C)
Primary / Ambient Air Temp. . . . . . ±2 ºF between 32 and 212 ºF
(±1 °C between 0 and 100 °C)
Pressure / Draft . . . . . . . . . . . . . . . . ±2% of reading or ±0.02 "H2O
(±0.05 mb), whichever is greater
System Flow Rate with Probe . . . . . . . 200 cc/min minimum
Instruction 24-94482-2
Specications
Power Requirements:
Four disposable ‘AA’ alkaline batteries provide at least 15 hours of continu­ous operation. NiMH rechargeable batteries can also be used, with the operating time dependent on battery type and condition.
An optional AC power adapter, which runs from any convenient source of 100–240 VAC, 50/60 Hz power, can be used to power the analyzer on a con­tinuous basis. If using rechargeable batteries, the AC power adapter can also be used to charge the batteries while inside the analyzer.
Warm-Up Time:
60 seconds. Sensors are checked and auto zeroed during warm-up.
Memory:
• 500 complete combustion test records
• 500 complete logged combustion test records
Interfaces:
• Printer – Infrared (IrDA) communications
• Computer – USB
Dimensions:
9H x 3W x 2.5D inches (22.9 x 7.6 x 6.3 cm)
Weight:
• Analyzer – 1.4 lb (0.6 kg) w/ batteries
• Probe & Hose Assembly – 1 lb (0.5 kg)
Instruction 24-9448 2-3
Specications
Notes:
Instruction 24-94482-4
Initial Setup
Release and then remove battery cover by pushing latch up
3 Initial Setup
3.1 Scope
Before using the PCA 2, you MUST:
• Install batteries, or plug in the optional AC power adapter (Section 3.2)
• Connect the probe and hose assembly (Section 3.3)
• Check, and if necessary, make changes to the analyzer’s configuration (Section 3.4)
3.2 Power
3.2.1 Installing or Replacing Batteries
Either alkaline or NiMH rechargeable batteries can used to power the an­alyzer. Note that if rechargeable batteries are used, they can be recharged while installed inside the analyzer using the optional AC power adapter (refer to Section 3.2.2).
Install or replace the batteries as described below:
1. Remove battery cover from back of unit (Figure 3-1.)
2. Remove (and properly dispose of) any old batteries.
3. Install a set of four ‘AA’ alkaline or NiMH batteries, per the “+” and “–” markings inside the battery compartment.
4. Replace battery cover.
Figure 3-1. Installing Batteries
Instruction 24-9448 3-1
Initial Setup
3.2.2 Using the AC Power Adapter
The AC power adapter is capable of powering the analyzer on a continuous basis. The adapter plugs into an appropriate 100–240 VAC, 50/60 Hz wall outlet, and produces an output of +9 VDC. The adapter’s output connector plugs into the analyzer’s POWER jack located on the bottom of the unit (Figure 3-2).
If NiMH rechargeable batteries are used, the adapter can also rapid charge these batteries in approximately 2 - 3 hours while still inside the analyzer. For the batteries to be charged, however, the analyzer’s battery charger circuit must be turned ON per Section 3.16.
3.3 Connecting the Probe and Hose Assembly
Do the following to attach the probe and hose assembly to the analyzer (Figure 3-2):
1. Push the gas-sample hose connector, the larger of the two connectors (giv­ing a slight twist), onto the analyzer’s GAS fitting.
2. Push the draft-hose connector, the smaller connector (giving a slight twist), onto the analyzer’s +∆P fitting.
3. Push the stack-gas thermocouple connector into the T-STACK jack (con­nector fits in only one way).
NOTE: The analyzer has a built-in temperature sensor for measuring ambient temperature. Perform Step 4 only if the optional primary / ambient air thermocouple is used.
4. Push the optional primary / ambient air thermocouple into the T-AIR jack (connector fits in only one way).
IMPORTANT: To assure the accurate calculation of combustion efficiency, the optional primary / ambient air thermocouple must be used when the burner’s primary-air temperature is not the same as the room temperature.
5. Inspect all hoses for cracks. If any hose is found to be defective, re­place the entire probe and hose assembly. Check that the water trap is empty, and that the filter is not dirty or saturated with water.
Instruction 24-94483-2
Initial Setup
POWER
P
T-STACK T-AIR
GAS
Differential Pressure Hose
(— ∆P) (Optional)
Probe Handle
Water Trap /
Filter Assembly
Stack Gas
Thermocouple
Connector
(T-STACK)
AC Power
Adapter Jack (POWER)
Option
Connector
(Optional External Measurement)
Draft Hose
(+ ∆P)
Gas Sample
Hose
(GAS)
Probe Tube
Probe Stop
Primary /
Ambient Air
Thermocouple (T-AIR) (Optional)
Figure 3-2. Connecting the Probe and Hose Assembly to the PCA 2
Instruction 24-9448 3-3
Initial Setup
FUEL MENU NATURAL GAS OIL #2 OIL #4 OIL #6
PROPANE
COAL WOOD
(more)
MENU
F1 F2 F3
MAIN MENU
FUEL
PRESSURE TEMPERATURE MEMORY SETUP CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
3.4 Operating Parameters
The PCA 2 is set up at the factory for the following operating parameters:
Fuel . . . . . . . . . . . . . . . . . . . . .Natural Gas
Temperature Units . . . . . . . . . °F
Pressure Units . . . . . . . . . . . .Inches of Water Column (inwc)
Pollution Units . . . . . . . . . . . .ppm
Date . . . . . . . . . . . . . . . . . . . . . Current MM/DD/YY
Time . . . . . . . . . . . . . . . . . . . . Current EST HH:MM AM/PM
O2 Reference . . . . . . . . . . . . . .0%
Print Pressure . . . . . . . . . . . .No
Zoom . . . . . . . . . . . . . . . . . . . .No
Battery Charger . . . . . . . . . . .OFF
Logging . . . . . . . . . . . . . . . . . . No
Button Sound . . . . . . . . . . . . .ON
To change any of these parameters, perform the associated procedure pro­vided in Sections 3.5 thru 3.22.
3.5 Fuel Selection
To assure the accurate calculation of combustion efficiency, select the fuel being burned as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight FUEL, and then press ENT to display the FUEL MENU.
3. Use the buttons to scroll through the list of available fuels until the desired fuel is high­lighted. In the example shown, PROPANE has been selected. (If custom fuels are added, they will be displayed at the bottom of the list)
TIP: Use the buttons to quickly scroll to the bottom and top of the list.
4. Press ENT to save the selection and display the Combustion Test HOLD screen. Observe
Instruction 24-94484
that the name of the selected fuel should now appear at the
SETUP MENU
TEMP UNITS
PRESSURE UNITS POLLUTION UNITS DATE TIME O2 REFERENCE PRINT PRESSURE
(more)
MENU
F1 F2 F3
TEMP UNITS MENU Celsius
Fahrenheit
MENU
F1 F2 F3
05/26/06 09:25:30 AM
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
top of the screen.
Note: In addition to the standard fuels programmed into the
PCA2, Bacharach can develop custom fuel codes based on the customers specific needs. The PCA2 can be programmed with 2 additional fuels which can be added to the instrument using the PC based software provided. Consult factory for price and delivery.
3.6 Temperature Units Selection
Select to display temperature in either °F or °C as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
Initial Setup
3. Use the buttons to highlight TEMP UNITS, and then press ENT to display the TEMP UNITS MENU.
4. Use the buttons to highlight the desired temperature units. In the example shown, Fahrenheit has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
Instruction 24-9448 5
Initial Setup
PRESSURE UNITS MENU
InchesWater
milliBar Pascals hectoPascals
MENU
F1 F2 F3
05/26/06 09:25:30 AM
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
SETUP MENU TEMP UNITS
PRESSURE UNITS
POLLUTION UNITS DATE TIME O2 REFERENCE PRINT PRESSURE
(more)
MENU
F1 F2 F3
3.7 Pressure Units Selection
Select to display pressure in Inches of Water Column (inwc), millibar (mb), Pascals (Pa), or hectoPascals (hPa) as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight PRESSURE UNITS, and then press ENT to display the PRESSURE UNITS MENU.
4. Use the buttons to highlight the de­sired pressure units. In the example shown, InchesWater has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
Instruction 24-94486
Initial Setup
SETUP MENU TEMP UNITS PRESSURE UNITS
POLLUTION UNITS
DATE TIME O2 REFERENCE PRINT PRESSURE
(more)
MENU
F1 F2 F3
POLLUTION UNITS MENU
ppm
#/Mbtu mg/m3 g/GJ
MENU
F1 F2 F3
05/26/06 09:25:30 AM
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
3.8 Pollution Units Selection
The PCA 2 is capable of converting the measured ppm levels of CO, NO, NO2, and SO2 to various pollution units using CFR40 Part 60 emission factors. Note that the pollution-unit conversions for NO, NO2 and NOx are based on the molecular weight of NO2.
Select to display pollution units in parts per million (ppm), pounds of pol­lutant per million BTU (#/Mbtu), milligrams of pollutant per cubic meter of gas (mg/m3), or grams of pollutant per gigajoule (g/GJ) as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight POLLUTION UNITS, and then press ENT to display the POLLUTION UNITS MENU.
4. Use the buttons to highlight the desired pollution units. In the example shown, ppm has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
Instruction 24-9448 7
Initial Setup
SETUP MENU TEMP UNITS PRESSURE UNITS POLLUTION UNITS
DATE
TIME O2 REFERENCE PRINT PRESSURE
(more)
MENU
F1 F2 F3
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
DATE MENU
DATE 05/26/06
Press ENTER to Save
05/26/06 09:25:30 AM
MENU
F1 F2 F3
3.9 Date Setup
The date is stored in the format: MM/DD/YY. Its value is part of the date and time stamp that is saved along with each combustion test record.
Set the analyzer’s internal clock to the current date as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight DATE, and then press ENT to display the DATE MENU.
4. First use the buttons to move the cursor across the screen until it is over the digit to be changed, and then press the buttons until the desired value is displayed.
5. Repeat Step 4 until the values for month, day, and year have been set.
6. Press ENT to save the displayed date values and re-display the SETUP MENU, or press ESC to abort this procedure and retain the old date values.
Note: The Date and Time real time clock is powered by the main bat­teries and is maintained by a supercap on the Main PCB for approxi­mately 1-2 days in the absence of batteries. The supercap is intended to maintain the real time clock when the batteries are changed when exhausted. If the batteries are removed for extended periods of time such as when the PCA2 is not in use (off-season storage), simply reset the time and date after fresh batteries are installed when it is placed back in service.
Instruction 24-94488
SETUP MENU TEMP UNITS PRESSURE UNITS POLLUTION UNITS DATE
TIME
O2 REFERENCE PRINT PRESSURE
(more)
MENU
F1 F2 F3
3.10 Time Setup
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
TIME MENU
TIME 09:25 AM
Press ENTER to Save
05/26/06 09:25:30 AM
MENU
F1 F2 F3
The time is stored in the format: hh:mm:ss AM/ PM. Its value is part of the date and time stamp that is saved along with each combustion test re­cord.
Set the analyzer’s internal clock to the current time as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight TIME, and then press ENT to display the TIME MENU.
4. First use the buttons to move the cursor across the screen until it is over the digit to be changed, and then press the buttons until the desired value is displayed.
Initial Setup
5. Repeat Step 4 until the values for hour, min­ute, and meridiem have been set.
NOTE: The value for seconds cannot be en-
tered, but are displayed and stored as part of the combustion test record.
6. Press ENT to save the displayed time values and re-display the SETUP MENU, or press ESC to abort this procedure and retain the old time values.
Instruction 24-9448 9
Initial Setup
O2 REFERENCE
CO Ref to O2
NOx Ref to O2 SO2 Ref to O2
05/26/06 09:25:30 AM
MENU
F1 F2 F3
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
SETUP MENU TEMP UNITS PRESSURE UNITS POLLUTION UNITS DATE TIME
O2 REFERENCE
PRINT PRESSURE
(more)
MENU
F1 F2 F3
SET O2 REFERENCE
CO REF TO O2: 0
05/26/06 09:25:30 AM
MENU
F1 F2 F3
3.11 O2 Reference Setup
The measured values of CO, NOx, and SO2 can be individually referenced to a specific O2 percentage of between 0 and 15%.
Individually set up the O2 reference value for each of the above gases as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight O2 REF, and then press ENT to display the O2 REFER­ENCE screen.
4. Use the buttons to highlight the desired measurement, and then press ENT to display the REF TO O2 screen for that measurement. In the example shown, CO has been selected.
5. First use the buttons to move the cursor across the screen until it is over the digit to be changed, and then press the buttons until the desired value is displayed.
6. Press ENT to save the displayed value and re­display the O2 REFERENCE screen, or press ESC to abort this procedure and retain the old O2 reference value.
7. If the O2 reference value for more than one gas is being set, then repeat Steps 4, 5, and 6 for each measurement.
Instruction 24-944810
3.12 Print Pressure Selection
PRINT PRESSURE
No
Yes
05/26/06 09:25:30 AM
MENU
F1 F2 F3
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
SETUP MENU
(more) TIME O2 REFERENCE
PRINT PRESSURE
ZOOM BATTERY CHARGER LOGGING
(more)
MENU
F1 F2 F3
Select whether to print or not print the pressure measurement on the combustion test printout as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight PRINT PRES­SURE, and then press ENT to display the PRINT PRESSURE screen.
4. Use the buttons to highlight either No (do not print pressure) or Yes (print pressure). In the example shown, Yes has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
Initial Setup
Instruction 24-9448 11
Initial Setup
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
SETUP MENU
(more) TIME O2 REFERENCE PRINT PRESSURE
ZOOM
BATTERY CHARGER LOGGING
(more)
MENU
F1 F2 F3
ZOOM
Standard
2x 3x
05/26/06 09:25:30 AM
MENU
F1 F2 F3
O2 4.0 %
CO 12 ppm
EFF 82.6%
PRINT MENU SAVE
F1 F2 F3
Hold NGas
3.13 Zoom-Display Selection
Combustion test data in the Run/Hold screen can be shown with enlarged characters to make view­ing easier. The operator can set zoom levels to
Standard, 2X, or 3X. The Standard zoom setting will display 8 lines of combustion test data at one time, 2X which will display 4 lines of combus­tion test data with enlarged characters, and 3X which will display 3 lines of combustion test data with enlarged characters. The operator can scroll through the complete list of measured and cal­culated data no matter what zoom level has been selected.
Select desired zoom level as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight ZOOM, and then press ENT to display the ZOOM screen.
4. Use the buttons to select the desired Zoom
level. Options include STANDARD, 2X, and 3X. Standard will display eight lines of Com­bustion test data, 2X will display four, and 3X will display 3.
5. Press ENT to save the selection and re-display the SETUP MENU.
Instruction 24-944812
3.14 Battery Charger Selection
BATTERY CHARGER OFF
ON
05/26/06 09:25:30 AM
MENU
F1 F2 F3
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
SETUP MENU
(more) TIME O2 REFERENCE PRINT PRESSURE ZOOM
BATTERY CHARGER
LOGGING
(more)
MENU
F1 F2 F3
When using rechargeable NiMH batteries, the AC power adapter can be used to charge the batteries while inside the analyzer. The analyzer’s rapid­charger circuit, however, must be first turned ON.
IMPORTANT: When using disposable alka­line batteries, the analyzer’s battery charger circuit should be OFF to prevent the batter­ies from overheating.
As a precaution, the charger circuit is automati­cally toggled to its OFF state when the analyzer is turned OFF, thus requiring the operator to turn the charger back ON when needed.
Note: The PCA2 must be left turned ON to charge the batteries.
Turn the battery charger circuit ON and OFF as follows:
Initial Setup
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight BATTERY CHARGER, and then press ENT to display the BATTERY CHARGER screen.
4. Use the buttons to highlight either OFF (turn charger OFF) or ON (turn charger ON). In the example shown, ON has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
Instruction 24-9448 13
Initial Setup
BUTTON SOUND
OFF
ON
05/26/06 09:25:30 AM
MENU
F1 F2 F3
SETUP MENU
(more)
LOGGING
TEST ID USERNAME RUN/HOLD FORMAT LANGUAGE SELECTION BUTTON SOUND CAL REMINDER PERIOD
MENU
F1 F2 F3
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
SETUP MENU
(more) LOGGING TEST ID USERNAME RUN/HOLD FORMAT LANGUAGE SELECTION
BUTTON SOUND
CAL REMINDER PERIOD
MENU
F1 F2 F3
3.15 Logging Selection
When the logging function is activated, up to 500 combustion test records will be automatically stored in memory at a preset interval over a prede­termined length of time.
Refer to Section 4.14 for detailed information on how to select the logging function; how to set the interval and duration time periods; and how to view or download the stored data.
3.16 Button Sound
The audible sound used to signal when a button is pressed can be turned OFF and ON as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight BUTTON SOUND, and then press ENT to display the BUTTON SOUND screen.
4. Use the buttons to highlight either OFF (turn sound OFF) or ON (turn sound ON). In the example shown, OFF has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
Instruction 24-944814
Initial Setup
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
Test ID Menu Select Test ID
Edit Test ID
Clear Test ID
F1 F2 F3
MENU
3.17 Test ID Information
Test records can be identified (e.g. customer's name, burner number, and location) by manually entering up to three lines of text, with each line con­taining a maximum of 20 alphanumeric characters. When a Test ID is se­lected this information will be associated with all succeeding test records, and will appear at the top of each test record when printed, and in Excel files when records are downloaded to a PC. The chosen Test ID remains in effect until it is deselected, a new Test ID is selected, or the instrument is turned off. Up to 30 Test ID's can be pre-entered for later retrieval.
To enter or edit a Test ID:
1. Display the MAIN MENU by press­ing the MENU (F2) button. If necessary, press ESC until MENU appears above F2 .
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight TEST ID, and then press ENT to display the TEST ID menu
4. Use the buttons to highlight EDIT TEST ID, and then press ENT to display the EDIT TEST ID menu, which displays the first line of each Test ID record.
5. Use the buttons to select which Test ID to edit and then press ENT to display all three
lines of that record. Periods (.....) are used to
identify empty lines.
6. Use the buttons to choose which of the three Test ID lines to edit and then press ENT to begin editing the chosen line.
7. Use the buttons to enter the desired character and then move to the next character position by pressing the right arrow key. Repeat this step until the line is complete. A maximum of 20 alphanumeric char­acters can be entered. Press ENT to accept the information.
8. At this time, either return to step 6 to choose another line to edit, or
Instruction 24-9448 15
Initial Setup
Test ID Menu
Select Test ID
Edit Test ID Clear Test ID
F1 F2 F3
MENU
Test ID Menu Select Test ID Edit Test ID
Clear Test ID
F1 F2 F3
MENU
end this procedure by highlighting Edit Complete and pressing the ENT key to return to the Edit Test ID Menu.
9. Press ESC to go to the Test ID Menu or the RUN/HOLD key to return to the Run/Hold screen.
Select a Test ID:
1. From the Test ID menu, use the buttons to highlight SELECT TEST ID, and then press ENT to display the SELECT TEST ID menu, which displays the first line of each Test ID record.
2. Use the buttons to choose a pre-entered Test ID or choose NO Test ID (located at bottom of list) to deselect a previously chosen record, then press ENT to make the selection and return to the Test ID menu.
3. Press ESC to end the procedure and return to the Setup Menu or press the RUN/HOLD key to return to the Run/Hold Screen.
Clear Test ID Information:
1. From the Test ID menu, use the buttons to highlight CLEAR TEST ID, and then press ENT to display the CLEAR TEST ID menu.
2. Do one of the following to clear Individual Re­cords or All Records.
*Individual Records - Use the buttons to
highlight Individual Records, then press ENT to display the Clear Individual menu. Again use the buttons to highlight the individual record to clear, then press ENT to clear the record.
*All Records - Use the buttons to highlight All Records, then press
ENT to display the Clear All menu. Again use the buttons to high­light Yes, then press ENT to clear all records.
3. Press ESC to end this procedure and return to the Test ID menu or press the RUN/HOLD key to return to the Run/Hold screen.
Instruction 24-944816
Initial Setup
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
SETUP MENU
(more) LOGGING TEST ID
USERNAME
RUN/HOLD FORMAT LANGUAGE SELECTION BUTTON SOUND CAL REMINDER PERIOD
MENU
F1 F2 F3
3.18 Username
The name of the user or owner of the analyzer (e.g. company name, ad­dress, phone number) can be stored in memory by manually entering up to three lines of text, with each line containing up to 20 alphanumeric characters. This information will appear at the top of each printout, until such time as new information is entered or cleared.
Username Information can be entered as follows:
1. Display the MAIN MENU by press­ing the MENU (F2) button. If necessary, press ESC until MENU appears above F2 .
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight USERNAME, and then press ENT to display all three lines
of the EDIT USERNAME screen. Periods(.....)
identify empty lines.
4. Use the buttons to choose which of the three Username lines to edit, then press ENT to begin editing the chosen line.
5. Use the buttons to enter the desired charac­ter and then move to the next character posi­tion by pressing the right arrow key. Repeat this step until the line is complete. A maxi­mum of 20 alphanumeric characters can be entered.
6. Press ENT to accept the entered information.
7. At this time, either return to step 4 to choose another line to edit, or end this procedure by highlighting Edit Complete and pressing the ENT key to return tot he Setup Menu.
8. Press ESC to go to the Main Menu or the RUN/HOLD key to return to the Run/Hold screen.
Instruction 24-9448 17
Initial Setup
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
SETUP MENU
(more) LOGGING TEST ID
USERNAME
RUN/HOLD FORMAT LANGUAGE SELECTION BUTTON SOUND CAL REMINDER PERIOD
MENU
F1 F2 F3
Edit Username
............
............
............
Edit Complete
MENU CLEAR
F1 F2 F3
To Clear a Username proceed as follows:
1. Display the MAIN MENU by press­ing the MENU (F2) button. If necessary, press ESC until MENU appears above F2 .
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight USERNAME, and then press ENT to display all three lines
of the EDIT USERNAME screen. Periods(.....)
identify empty lines.
4. Use the buttons to choose which of the three Username lines to clear, then press CLEAR (F3) to clear that line.
5. At this time, either return to step 4 to choose another line to clear, or end this procedure by highlighting EDIT COMPLETE and pressing the ENT key to return tot he Setup Menu.
6. Press ESC to go to the Main Menu or the RUN/ HOLD key to return to the Run/Hold screen
Instruction 24-944818
Initial Setup
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
SETUP MENU
(more) Logging Test ID Username Run/Hold Format
Language Selection
Button Sound CAL Reminder Period
MENU
F1 F2 F3
Language Selection English Français Español
06/28/06 11:45:30 AM
MENU
F1 F2 F3
3.19 Language
Information on the display screen can be shown in English, French, or Spanish. Select the desired language as follows:
1. Display the MAIN MENU by press­ing the MENU (F2) button. If necessary, press ESC until MENU appears above F2 .
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight LANGUAGE SELECTION, and then press ENT to display The Language Selection Menu.
4. Use the buttons to highlight the desired language, then press ENT to activate the se­lection and re-display the Setup Menu.
Instruction 24-9448 19
Initial Setup
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
SETUP MENU
(more) Logging Test ID Username Run/Hold Format Language Selection Button Sound
CAL Reminder Period
MENU
F1 F2 F3
CAL Reminder Period
6 months
8 months 10 months 12 months 15 months
MENU
F1 F2 F3
3.20 Cal Reminder Period
The analyzer can be set to indicate a calibration reminder during the 60 second warm-up period. Calibration reminders can be preset to occur 6, 8, 10, 12, or 15 months after the last calibration. When the preset period is exceeded the instrument will display the reminder, and how long since the sensors were last calibrated. The reminder will be displayed at the end of the 60 second warm-up period. If a calibration reminder is displayed the operator can press the RUN/HOLD key to move to the Run/Hold screen for normal operation. Regular calibration periods of 6 months to a year for all gas sensors (except Oxygen) are recommended.
Set the calibration reminder period as follows:
1. Display the MAIN MENU by press­ing the MENU (F2) button. If necessary, press ESC until MENU appears above F2 .
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight CAL RE­MINDER PERIOD, and then press ENT to display CAL Reminder Period Menu.
4. Use the buttons to highlight the desired CAL reminder period, and then press the ENT key to activate the selection and re-display the Setup Menu.
Note: The date and time settings must be correct to
get accurate cal reminders.
Instruction 24-944820
Initial Setup
3.21 Run/Hold Screen Format
PCA2 testdata is located in the Run/Hold screen. By pressing the Run/ Hold key, you should hear the pump running and see the word RUN at the upper-left hand corner of the display. The instrument is continuously measuring and calculating the data that is shown in the Run/Hold screen. Press the RUN/HOLD key again, the pump should stop running and the word HOLD should be shown at the upper-left hand corner of the display. The instrument will now show the last measured and calculated data taken before the instrument was placed in HOLD. Use the up and down arrow keys to scroll through the complete list of measured and calculated values when the instrument is running or in the hold mode.
The order in which data appears in the Run/Hold screen is as follows:
O2 = Oxygen CO = Carbon Monoxide
EFF = Combustion Efciency
CO2 = Carbon Dioxide T-STK = Stack Temperature T-AIR = Ambient/Primary Air Temperature EA = Excess Air CO(O) = Carbon Monoxide content referenced to an Oxygen percentage NO = Nitric Oxide NO2 = Nitrogen Dioxide NOx = Oxides of Nitrogen (NO and NO2 combined) SO2 = Sulfur Dioxide NO(O) = Nitric Oxide content referenced ro an Oxygen percentage NO2(O) = Nitrogen Dioxide content referenced to an Oxygen percentage NOx(O) = Oxides of Nitrogen content referenced to an Oxygen percentage SO2(O) = Sulfur Dioxide content referenced to an Oxygen percentage
Note: (O) denotes the current O2 Reference selected.
Note: Stars (***) appear in measurement and calculation elds of sensors
that are not installed.
Instruction 24-9448 21
Initial Setup
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
SETUP MENU
(more) Logging Test ID Username
Run/Hold Format
Language Selection Button Sound CAL Reminder Period
MENU
F1 F2 F3
Run/Hold Format
Edit Format
Reset Format
MENU
F1 F2 F3
Edit Run/Hold Format
O2
CO EA
CO2 T-STK T-AIR NO
(more)
MENU
F1 F2 F3
Change the order in which data is displayed as follows:
1. Display the MAIN MENU by press­ing the MENU (F2) button. If necessary, press ESC until MENU appears above F2 .
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight RUN/HOLD FORMAT, and then press ENT to display Run/ Hold Format Menu.
4. Use the buttons to highlight Edit Format, and then press the ENT key to show the Edit Run/Hold Format, where the current order of combustion data is displayed.
5. Change data shown for a particular location by first using the buttons to highlight the location. Then press the ENT key to select the location; the cursor will start to flash.
6. Use the buttons to scroll through and select the desired data to appear at that location. Press the ENT key to make the selection.
7. Change data displayed at other locations by repeating steps 5 & 6.
8. When finished, use the button to highlight Edit complete at the bottom of the list, then press ENT to save the new display format and return to the Run/Hold Format screen.
Instruction 24-944822
Reset Display format back to factory default settings as follows:
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
SETUP MENU
(more) Logging Test ID Username
Run/Hold Format
Language Selection Button Sound CAL Reminder Period
MENU
F1 F2 F3
Run/Hold Format
Edit Format
Reset Format
MENU
F1 F2 F3
Reset Format
No
Yes
MENU
F1 F2 F3
1. Display the MAIN MENU by press­ing the MENU (F2) button. If necessary, press ESC until MENU appears above F2 .
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight RUN/HOLD FORMAT, and then press ENT to display Run/ Hold Format Menu.
4. Use the buttons to highlight RESET FOR­MAT, then press ENT to display the Reset Format screen.
5. Use the buttons to select YES, then press ENT to reset the display and return to the Run/Hold Format screen.
Initial Setup
Instruction 24-9448 23
Operation
4 Operation
4.1 Operating Tips
• When an analyzer is brought in from a cold vehicle, let it warm up slowly to minimize condensation. Temperatures below freezing will not dam­age the analyzer; however, bringing a cold analyzer into a warm, humid environment may cause condensate to form inside the case.
CAUTION: Although the analyzer itself is not damaged by an extremely cold environment, the electrochemical sensors may be damaged. The O2 sensor's electrolyte will freeze at approximately -20 ºF and the other sen­sors at approximately -94 ºF. If the analyzer is exposed to an extremely cold condition, it is strongly suggested that the sensor housings be exam­ined for hairline cracks. Be aware that a leaking sensor can cause chemi­cal burns to the skin and possibly damage the PCB assemblies.
• Ensure that the analyzer is sampling fresh air when turned ON. Pulling a stack-gas sample through the analyzer during its warm-up period will not damage the analyzer, but it will result in incorrect sensor readings, and may result in sensor error messages appearing after the warm-up cycle completes.
• Note that flue-gas condensate is acidic and very corrosive. It is important not to allow the analyzer’s internal components to come in contact with condensate for long periods of time.
• Before each use, inspect the filter element of the water-trap / filter as­sembly. Replace the filter if it looks dirty. Refer to Section 6.2.
• When sampling flue-gas, keep the analyzer above the water-trap, and keep the trap in a vertical position. This will maximize the effectiveness of the trap and keep liquid condensate from being drawn directly into the analyzer.
• When liquid condensate is seen inside the water trap, empty the trap before it becomes full. Refer to Section 4.9.
• The analyzer should be purged after performing a combustion test. After removing probe from the stack, let the pump run for at least 10 minutes to remove any stack gases and dry any condensate from inside the sensor chamber and probe assembly. If the analyzer is turned OFF with high levels of flue gas remaining in the analyzer, then the pump will continue to run and the message “PURGING SENSORS” will appear on the dis­play until all flue gas levels fall below predetermined levels.
• When storing the analyzer, it’s a good idea to empty the water trap and leave it open to further dry it out.
• Calibrate the analyzer every 6 months - 1 year to assure its accuracy.
Instruction 24-9448 4-1
Operation
MENU
Warm Up 60
ERRORS DETECTED Low Battery
MENU
MENUMENU
PCA 2
FIRMWARE VERSION
x.xx
MODEL xxx SERIAL # x xxxxxx
HOLD NATURAL GAS O2 20.9 % CO 0 ppm EFF --- % CO2 --- %
T-STK 71.5 °F T-AIR 71.5 °F EA --- % CO(0) --- ppm
PRINT MENU SAVE
F1 F2 F3
4.2 Turning ON the Analyzer and Warm Up
1. Connect the probe and hose assembly, and make sure that the analyzer is properly set up per Section 3 Initial Setup.
IMPORTANT: DO NOT insert probe into stack before turning ON the analyzer!
2. Place the probe in an area that contains fresh air. This ensures that the sensors will be properly zeroed during the warm-up cycle.
3. Turn ON the analyzer by pressing the I/O button for at least 1 second, or until a single beep is heard. Observe that the analyzer’s firmware version, model and serial numbers are briefly displayed followed by the Warm Up screen.
4. Wait for the analyzer to count down its 60 sec­ond warm-up period; after which, the instru­ment will display the Combustion Test HOLD screen.
If problems were detected during warm up, the message “ERRORS DETECTED” is dis­played along with a list of those errors. As an example, the screen to the right shows that the battery is low. Refer to Section 7.3 for a listing and possible remedy for the errors displayed.
TIP: If the sensors in error are not critical to the combustion test, then press the RUN/HOLD button to display the Combustion Test HOLD screen and proceed with the test.
Instruction 24-94484-2
4.3 Selecting a Fuel
FUEL MENU NATURAL GAS OIL #2 OIL #4 OIL #6
PROPANE
COAL WOOD
(more)
MENU
F1 F2 F3
MAIN MENU
FUEL
PRESSURE MEMORY SETUP CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
HOLD NATURAL GAS O2 20.9 % CO 0 ppm EFF --- % CO2 --- %
T-STK 71.5 °F T-AIR 71.5 °F EA --- % CO(0) --- ppm
PRINT MENU SAVE
F1 F2 F3
HOLD PROPANE O2 20.9 % CO 0 ppm EFF --- % CO2 --- %
T-STK 71.5 °F T-AIR 71.5 °F EA --- % CO(0) --- ppm
PRINT MENU SAVE
F1 F2 F3
The top line of the Combustion Test HOLD screen shows the fuel currently selected. In the example shown, the current fuel is NATURAL GAS. If nec­essary, change the fuel as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button.
2. Use the buttons to highlight FUEL, and then press ENT to display the FUEL MENU.
3. Use the buttons to scroll through the list of available fuels until the desired fuel is high­lighted. In the example shown, PROPANE has been selected.
NOTE: If custom fuels were added to the
instrument, they will be located at the bottom of the list.
TIP: Use the buttons to quickly scroll to the bottom and top of the list.
Operation
4. Press ENT to save the selection and return to the Combustion Test HOLD screen. The newly selected fuel should now appear in the
top line of the display.
Instruction 24-9448 4-3
Operation
O2, CO, Stack Temp.
Undiluted Flue Gas Sample Taken Under Draft Diverter in Top of Fire Tube
O2, CO, Stack Temp.
O2, CO, Stack Temp.
4.4 Sampling Point
FORCED AIR FURNACE – For atmospheric burner
or gravity vented, forced air heating equipment with a clamshell or sectional heat exchanger de­sign, test each of the exhaust ports at the top of the heat exchanger. The probe should be inserted back into each of the exhaust ports to obtain a flue-gas sample, before any dilution air is mixed in.
HOT WATER TANK – Domestic hot water tanks with the ‘bell’ shaped draft diverter can be ac­curately tested by inserting the probe tip directly into the top of the fire tube below the diverter.
80% EFFICIENCY FAN ASSIST OR POWER VENTED – Combustion testing of fan assist or power vented, furnaces/boilers should be done through a hole drilled in the vent immediately above the inducer fan.
90% EFFICIENCY CONDENSING – Condens­ing furnaces/boilers can be tested through a hole drilled in the plastic vent pipe (when allowed by the manufacturer or local authority of jurisdiction) or taken from the exhaust termination.
Forced Air Furnace
ATMOSPHERIC OR GRAVITY VENTED BOILER – Boilers, which have a ‘bell’ shaped draft diverter on top, should be tested directly below the diverter through a hole drilled in the vent connector.
O2, CO, Stack Temp.
80% Eff. Fan Assist or
90% Eff. Condensing
Power Vented
Hot Water Tank
Combustion Air Sampling Point
Atmospheric or Gravity
Vented Boiler
Instruction 24-94484-4
Operation
RUN NATURAL GAS O2 4.0 % CO 12 ppm EFF 82.6 % CO2 9.5 %
T-STK 374 °F T-AIR 68.0 °F EA 21.3 % CO(3) 13 ppm
PRINT MENU SAVE
F1 F2 F3
4.5 Performing a Combustion Test
Ensure that the following has been completed, and then proceed with the combustion test as described below:
• Turn ON analyzer and allow it to warm up (Section 4.2).
• Select fuel being burned (Section 4.3).
• Inset probe into stack (Section 4.4).
• If necessary, insert optional primary air thermocouple into combustion­air stream of burners that use an outside source of combustion air.
1. Press the RUN/HOLD button to start the test. You should hear the pump start running and see the word RUN appear at the top of the Combustion Test screen.
Sensor Indicators: The following indicators
appear in the sensor's data field depending on certain conditions:
(∗ ∗ ∗) Sensor that is not calibrated or installed.
(XXX) Sensor overrange (
- - -
) The calculated data cannot be displayed because the measured data necessary to make the calculation is out of range (i.e., oxygen level above 16%).
2. Use the up and down arrow keys to scroll to the T-STK reading. Loosen the thumbscrew on probe stop and move probe in and out of the stack until the stack’s core temperature (hot spot) is located as indicated by the highest T-STK reading; then tighten thumbscrew to prevent further probe movement. Locating the highest stack tempera-
ture is very important for accurate efficiency calculations.
3. You can now begin burner-service procedures. The analyzer readings will change quickly to show changes in burner performance.
CAUTION: Position the Water Trap with its gas-flow arrow pointing upward. Do not let water condensate go above the tip of the riser tube. The sensors could be damaged if water would enter the analyzer. Empty the Water Trap after every combus­tion test (refer to Section 4.9)
4. Pressing the RUN/HOLD button freezes all readings, stops the pump and displays the Combustion-Test HOLD screen. Press the ENT button to view all test values at the moment the RUN/HOLD button was pressed. Press- ing RUN/HOLD again restarts the pump and resumes testing.
Instruction 24-9448 4-5
Operation
TABLE 4-1. LIST OF COMBUSTION TEST DATA
Display Name
O
2
Description of Measurement or Calculation
% Oxygen
CO Carbon Monoxide (1)
EFF Combustion Efficiency
CO
2
% Carbon Dioxide
T-STK Stack Temperature
T-AIR Primary / Ambient Air Temperature as measured either in-
ternally or by an optional external thermocouple plugged into the analyzer’s T-AIR connector
CO(O) Carbon Monoxide ppm level referenced to a % of oxygen (2)
NO Nitric Oxide (1)
NO
2
Nitrogen Dioxide (1)
NOx Oxides of Nitrogen (NO and NO2 combined) (1)
SO
2
Sulfur Dioxide (1)
NO(O) Nitric Oxide ppm level referenced to a % of oxygen (2)
NO2(O) Nitrogen Dioxide ppm level referenced to a % of oxygen (2)
NOx(O) Oxides of Nitrogen ppm level referenced to a % of oxygen (2)
SO2(O) Sulfur Dioxide ppm level referenced to a % of oxygen (2)
NO Temp Nitric Oxide Sensor Temperature (3)
(1) Pollution unit of measure selected per Section 3.9
(2) The letter “O” represents the oxygen reference level of between 0 and 15% as
selected per section 3.12
(3) Shown only on printout, not on LCD display.
Note: Stars (***) appear in measurement and calculation fields of sensors that are not
installed
Instruction 24-94484-6
Operation
MAIN MENU FUEL
PRESSURE
MEMORY SETUP CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
Pressure
Measured
6.00 inwc
Type: Draft Reading
F1 F2 F3
Print Zero Save
4.6 Pressure Label Selection
The pressure measurement can be labeed with types including gas pres­sure, differential across heat exchanger, draft reading, and differential pressure. See Section 4.6 of the PCA2 Instruction Manual 24-9448 for pressure and draft measurement procedures.
Label data as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If nec­essary, press ESC until MENU appears above F2.
2. Use the buttons to highlight PRESSURE, and then press ENT to display the PRES­SURE MENU.
3. Use the buttons to scroll through the available pressure label types. The current label will be used, saved, and printed with the record
4. Press ENT to save the selection and re-display the PRESSURE MENU or ESC to exit back to the Main Menu Press ENT to save the selec­tion and re-display the PRESSURE MENU or ESC to exit back to the Main Menu
Instruction 24-9448 4-7
Operation
MAIN MENU FUEL PRESSURE
TEMPERATURE
MEMORY SETUP CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
Temperature
Measured Delta T:
*** °F
Type: Diff Temp
Print Zero Save
4.7 Temperature Measurement Label
The difference in temperature between two areas can be measured by us­ing the analyzer's two temperature channels and the Temperature screen. By using the T-Air channel as a reference, the temperature applied to the T-Stack channel will be displayed on the Temperature Measurement screen as differential temperature between the two channels. Addition­ally, the temperature measurement can be labeled. Label types include differential temperature across heat exchange or differential temperature. Perform a differential temperature measurement as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight TEMPERA­TURE, and then press ENT to display the TEMPERATURE MEASUREMENT screen.
3. Install thermocouples in both temperature channel connectors.
4. Before taking a measurement, the temperature channels may need to be zeroed, if not already displaying zero with both thermocouples ex­posed to the same condition. Press ZERO (F2) if needed.
5. To label the data use the buttons to scroll through the available temperature label types. The current label will be used, saved, and printed with the record.
6. Press ENT to save the selection and re-display the TEMPERATURE MENU or ESC to exit back to the Main Menu
Instruction 24-94484-8
Operation
POWER
P
T-STACK T-AIR
GAS
— ∆P (Pressure Reference Port)
+∆P (Pressure Measurement Port)
MAIN MENU FUEL
PRESSURE
MEMORY SETUP CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
PRESSURE
Measured:
0.00 inwc
PRINT ZERO SAVE
F1 F2 F3
4.8 Making a Draft / Pressure Measurement
The difference in pressure (P) between two areas can be measured by using the analyzer’s two pressure ports and the PRESSURE screen. By using the –P port as the reference, the pressure applied to the +∆P port will be displayed on the PRESSURE screen as the differential pressure between the two ports.
Perform a draft / pressure measurement as follows:
1. Turn ON the analyzer and allow it to complete its warm-up cycle (Section 4.2).
TIP: The pressure units of measure is selected per Section 3.8.
2. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
3. Use the buttons to highlight PRESSURE, and then press ENT to display the PRESSURE screen.
4. Before taking a measurement, the pressure sensor may need to be re-zeroed if it is not already displaying zero with both pressure ports open to the atmosphere. If necessary, zero the pressure sensor as follows:
Sampling Hoses with quick
connect ttings are available
as an optional accessory.
Instruction 24-9448 4-9
Figure 4-1. Draft / Pressure Measurement Hose Connections
Refer to Section 8.2.
Operation
PRESSURE ZERO
Disconnect hose then press ENTER
PRINT ZERO SAVE
F1 F2 F3
PRESSURE ZERO
Reconnect hose
PRINT ZERO SAVE
F1 F2 F3
RUN NATURAL GAS O2 4.0 % CO 12 ppm EFF 82.6 % CO2 9.5 %
T-STK 374 °F T-AIR 68.0 °F EA 21.3 % CO(3) 13 ppm
PRINT MENU SAVE
F1 F2 F3
a. Press the ZERO (F2) button.
b. Disconnect any hoses connected to the
+∆P and –∆P ports, and then press ENT to zero the pressure sensor.
c. Reconnect any hoses. When measuring
draft, simply leave the –∆P port open to
the atmosphere and connect the probe’s draft hose to the +P port (see Figure 3-2 on Page 3-3).
5. Do one of the following to measure draft or differential pressure:
• To measure draft, simply insert the probe into the stack and observe the draft read­ing on the PRESSURE screen.
• To measure differential pressure, connect two sampling hoses to the +P and –P ports, and place the open end of each hose into the areas being measured. The differential pressure between the two areas is now displayed on the PRESSURE screen. If the pressure at the +P port is higher than the –P port, then the pressure reading will be positive. If it is lower, then the reading will be negative.
4.9 Saving Test Data
Up to 500 individual sets (“snap shots”) of combustion-test, pressure, or temperature data can be saved in memory, which can later be recalled for viewing from the Memory Directory (Section 4.13.1).
1. First display the screen that contains the data to be saved. In the example to the right, all data associated with the Combustion Test RUN screen will be saved.
2. Press the SAVE (F3) button to save the test data in the next available memory location.
NOTE: When memory is full, the next reading will not be saved until space is made available by clearing the data. (See 4.15.4: Clearing Memory)
Instruction 24-94484-10
4.10 Ending a Combustion Test
To Gas Port on Analyzer
Gas Flow
Do not allow water level to go above tip of riser tube
Stack Gas From Probe
To empty water, pull apart using a slight twisting motion
WARNING! Burn Hazard. Do not touch the probe after removing it from the stack. Allow the probe to cool before handling (about 5 minutes).
1. Remove probe from stack.
2. Allow the pump to run until all combustion gases have been flushed from the analyzer as indicated by the O2 reading returning to 20.9%.
4.11 Emptying the Water Trap
The Water Trap / Filter Assembly removes stack-gas condensate, and also prevents soot from contaminating the internal components of the analyzer.
IMPORTANT: Use the Water Trap / Filter Assembly in a vertical position with the gas-flow arrow pointing up as shown in the illustration to the right.
Empty the water trap chamber after each combustion test, or stop the test and empty the chamber if the liquid condensate level ap­proaches the tip of the riser tube.
To empty the trap, first pull apart the two halves of the Water Trap using a slight twist­ing motion; empty the water trap chamber; and then reassemble the trap.
After each combustion test, also check the Water Trap’s filter element. If it looks dirty, replace the filter per Section 6.2.
Operation
Instruction 24-9448 4-11
Operation
MENU
SHUTDOWN IN 5 sec
MENU
PURGING SENSORS
RUN NATURAL GAS O2 4.0 % CO 12 ppm EFF 82.6 % CO2 9.5 %
T-STK 374 °F T-AIR 68.0 °F EA 21.3 % CO(3) 13 ppm
PRINT MENU SAVE
F1 F2 F3
Low Batter y Alarm Empty Battery Icon
4.12 Turning OFF the Analyzer & Purging
Turn OFF the analyzer by pressing the I/O but­ton for at least 2 seconds, or until two beeps are heard. The unit will count down 5 seconds before shutting down, giving the operator an opportu­nity to keep the analyzer turned ON by pressing the RUN/HOLD button.
If the PCA 2 was not purged with fresh air as described in Section 4.8, then the analyzer may remain ON with its pump running and display the message “PURGING SENSORS” as the re­sult of combustion gases still being present inside the analyzer. At this time the operator should ensure that the probe is removed from the stack, allowing the analyzer to purge itself with fresh air. The 5-second-shutdown sequence will not begin until the gas levels inside the analyzer drop below predetermined levels:
TIP: Although not recommended, the purg­ing process can be bypassed by pressing the
I/O button a second time.
4.13 Low Battery Alarm
When the batteries are nearly depleted, an empty battery icon appears in the upper-right corner of the display, and a short beep is sounded every 10 seconds.
After a low battery alarm occurs, the analyzer will continue to operate for only a few minutes. The amount of operating time that remains de­pends on many factors (e.g., pump and backlight being ON or OFF, and the type and condition of the batteries).
Instruction 24-94484-12
Operation
4.14 Data Logging
When the logging function is activated, up to 500 combustion test records will be automatically stored in memory at a preset interval (1, 5, 10, 15, 30 seconds, 1, 2, 5, 10 minutes) over a predetermined duration (5, 10, 15, 30 minutes, 1, 2, 5, 10, 24, 48 hours).
The maximum duration that data can be collected is determined by the in­terval. For example, if the interval is set to 10 seconds, then the maximum selectable length of time in which data can be collected to fill 500 memory locations would be 1 hour (500 x 10 seconds = 5000 seconds or 83 minutes). If the operator chooses a duration that is longer than possible for the inter­val chosen, then the analyzer automatically selects the highest duration possible for the selected interval.
The stored logged data can either be viewed on the display using the ana­lyzer’s memory function (refer to Section 4.13), or downloaded to a personal computer using the supplied PCA 2 data recovery software and USB cable (refer to Section 4.14). Downloaded data is stored on the computer’s hard drive as a comma-separated-value ASCII text file with a CSV extension, which can be opened by most spreadsheet programs for analysis.
TIP: The latest PCA 2 data recovery software can be download­ed from http://www.bacharach-inc.com/downloads.
Instruction 24-9448 4-13
Operation
SETUP MENU
(more)
LOGGING
TEST ID USERNAME RUN/HOLD FORMAT LANGUAGE SELECTION BUTTON SOUND CAL REMINDER PERIOD
MENU
F1 F2 F3
MAIN MENU FUEL PRESSURE MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
LOGGING
NO
YES
nnn MEMORY
LOCATIONS OPEN
MENU
F1 F2 F3
4.14.1 Turning ON Data Logging
IMPORTANT: Before turning ON data logging and starting the data logging process, the analyzer should already be set up to perform a combustion test per Section 4.5.
Turn ON data logging as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight SETUP, and then press ENT to display the SETUP MENU.
3. Use the buttons to highlight LOGGING, and then press ENT to display the LOGGING screen.
TIP: The amount of memory available for storing new data is displayed in the LOG­GING screen (maximum of 500 locations). If previous logging sessions are stored, and ad­ditional memory is needed, clear the logging memory per Section 4.13.3.
NOTE: At no time will new logging data over write old data.
4. Use the buttons to highlight YES (turn logging ON).
5. Press ENT to make the selection and display the LOGGING INTERVAL screen.
Instruction 24-94484-14
Operation
LOGGING INTERVAL 1 sec 5 sec 10 sec 15 sec 30 sec
1 min
2 min
(more)
MENU
F1 F2 F3
LOGGING DURATION 5 min 10 min 15 min 30 min 1 hr
2 hr
5 hr
(more)
MENU
F1 F2 F3
4.14.2 Setting the Logging Interval and Duration
The logging interval is the length of time between measurements, while the logging duration is the time allocated to the logging process. Set the logging interval and duration as follows:
TIP: If the duration is set for more than 10 hours, we recom­mend using the optional AC power adapter to power the analyzer.
1. As soon as logging is turned ON as described in Section 4.12.1, the LOGGING INTERVAL screen appears.
2. Use the buttons to highlight the desired interval, and then press ENT to make the selection and display the LOGGING DURA­TION screen. In this example, 1 minute has been selected
3. Use the buttons to highlight the desired duration, and then press ENT to make the selection and display the LOGGING SUM­MARY screen. In this example, 2 hours has been selected.
NOTE: If the selected duration requires more memory than is currently available, then the analyzer will automatically select the longest duration possible for the selected interval.
Instruction 24-9448 4-15
Operation
LOGGING SUMMARY
Interval 1 min Duration 2 hr
Press ENTER to Start
MENU
F1 F2 F3
LOG NATURAL GAS O2 4.0 % CO 12 ppm EFF 82.6 % CO2 9.5 %
T-STK 374 °F T-AIR 68.0 °F EA 21.3 % CO(3) 13 ppm
Log Entry xx
F1 F2 F3
4.14.3 Starting the Data Logging Process
After turning ON data logging and setting the interval and duration, the analyzer will pause at the LOGGING SUMMARY screen, where the currently selected interval and duration time periods are displayed.
Press ENT to start the combustion test and log­ging process. At this time the Combustion Test LOG screen will appear, indicating that the ana­lyzer is now performing a combustion test and the data is being stored in memory.
Note the following:
• At the bottom of the screen, the memory loca­tion where the current log entry is being saved is displayed.
• If there were previously stored log entries, each new logging session will start to store data in the next available memory location.
• At no time will new data overwrite old data.
• The ENT button can be pressed while logging to view the other Com­bustion Test screens.
Instruction 24-94484-16
Operation
LOG NATURAL GAS O2 4.0 % CO 12 ppm EFF 82.6 % CO2 9.5 %
T-STK 374 °F T-AIR 68.0 °F EA 21.3 % CO(3) 13 ppm
yy Entries Stored
F1 F2 F3
Where: “yy” is the number of log entries stored during the last logging session.
4.14.4 Ending the Data Logging Process
Data logging will stop and the pump will turn OFF after the prescribed duration, or after all 500 memory locations are filled.
To exit the Combustion Test LOG screen, press the RUN/HOLD button twice to display the Com­bustion Test HOLD screen.
Note the following:
• To end the logging process at any time, press the ESC button.
• At the end of the logging period, or if the ESC button was pressed, the total number of log entries that were stored during that period appear at the bottom of the screen.
• If the logging process was stopped by pressing ESC, it cannot be re­sumed without first turning logging back ON per Section 4.12.1.
4.15 Memory
There are two memory banks, each containing 500 memory locations. The first bank is used to store combustion test data as described in Section 4.7., while the second bank is used to store logged combustion test data as described in Section 4.12. Each bank is independent of each other, and can­not share data or be combined.
Individual memory locations in each memory bank can be recalled for viewing on the display or printed (refer to Sections 4.13.1 & 4.13.2), or the entire contents of each memory bank can be individually downloaded to a computer and viewed in a spreadsheet program for analysis (refer to Sec­tion 4.14).
Instruction 24-9448 4-17
TIP: When displaying the contents of either memory bank, the operator can quickly page through the screens by press­ing the PAGE– (F1) and PAGE+ (F3) buttons. Or move to the first or last memory location by pressing the buttons, respectively.
Operation
MEMORY DIRECTORY
1 05/23/06 10:00:00
2 05/23/06 10:30:14 3 05/23/06 11:45:38 4 05/23/06 01:50:10 5 05/23/06 03:35:25 6 EMPTY 7 EMPTY
(more)
F1 F2 F3
PAGE- MENU PAGE+
Mem: 1 NATURAL GAS O2 4.0 % CO 12 ppm EFF 82.6 % CO2 9.5 %
T-STK 374 °F T-AIR 68.0 °F EA 21.3 % CO(3) 13 ppm
PRINT MENU
F1 F2 F3
MAIN MENU FUEL PRESSURE
MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
MEMORY MENU
MEMORY DIRECTORY
LOGGING DIRECTORY
CLEAR MEMORY DATA CLEAR LOGGING DATA
CLEAR ALL MEMORY
F1 F2 F3
MENU
4.15.1 Recalling Combustion Test Data
Recall individual combustion test data records as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight MEMORY, and then press ENT to display the MEMORY MENU.
3. Use the buttons to highlight MEMORY DIRECTORY, and then press ENT to display the MEMORY DIRECTORY screen.
4. Use the buttons to highlight the desired memory location to be recalled. Each memory location is identified by the date and time at which data was saved. The word EMPTY signi­fies that the memory location does not contain data.
5. Press ENT to display the data contained in the selected memory location.
Note the following:
• The top line of the recalled combustion test data screen shows the memory location being viewed. In this example, “MEM: 1” is being displayed
• The ENT button can be pressed to view other memory test data screens.
• The recalled combustion test data can be printed by pressing the PRINT (F1) button (refer to Section 4.16).
Instruction 24-94484-18
LOGGING DIRECTORY
1 05/23/06 10:00:00 2 05/23/06 10:30:00 3 05/23/06 11:00:00 4 05/23/06 11:30:00 5 05/23/06 12:00:00
6 05/23/06 12:30:00
7 05/23/06 01:00:00
(more)
F1 F2 F3
PAGE- MENU PAGE+
Log: 6 NATURAL GAS O2 4.0 % CO 12 ppm EFF 82.6 % CO2 9.5 %
T-STK 374 °F T-AIR 68.0 °F EA 21.3 % CO(3) 13 ppm
PRINT MENU
F1 F2 F3
4.15.2 Recalling Logged Test Data
MAIN MENU FUEL PRESSURE
MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
MEMORY MENU
MEMORY DIRECTORY
LOGGING DIRECTORY
CLEAR MEMORY DATA CLEAR LOGGING DATA
CLEAR ALL MEMORY
F1 F2 F3
MENU
Recall individual logged combustion test data records as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight MEMORY, and then press ENT to display the MEMORY MENU.
3. Use the buttons to highlight LOGGING DIRECTORY, and then press ENT to display the LOGGING DIRECTORY screen.
4. Use the buttons to highlight the desired data logging location to be recalled. Each logging location is identified by the date and time at which the data was saved. The word EMPTY signifies that the logging location does not contain data.
5. Press ENT to display the logging data con­tained in the selected memory location.
Operation
Note the following:
• The top line of the log test data screen shows the log location being viewed. In this example, “Log: 6” is being displayed.
• The ENT button can be pressed to view other log test data screens.
• The recalled logged test data can be printed by pressing the PRINT (F1) button (refer to Section 4.16).
Instruction 24-9448 4-19
Operation
CLEAR MEMORY DATA
NO
YES
F1 F2 F3
MENU
CLEAR MEMORY DATA
WAIT
Erasing memory
F1 F2 F3
MENU
MAIN MENU FUEL PRESSURE
MEMORY
SETUP
CALIBRATION DIAGNOSTICS STATUS
F1 F2 F3
MENU
MEMORY MENU
MEMORY DIRECTORY LOGGING DIRECTORY
CLEAR MEMORY DATA
CLEAR LOGGING DATA
CLEAR ALL MEMORY
F1 F2 F3
MENU
4.15.3 Clearing Memory
When all memory (“snap shot”) locations used to store individual combustion test records have been filled, the next combustion test record saved will overwrite the oldest.
When all logging memory locations in the logging directory are full, they must be manually cleared in order to store new data. At no time will the log­ging process overwrite older data.
Each memory bank can be individually cleared, or all memory locations in both banks can be cleared simultaneously.
Do the following to clear memory:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the buttons to highlight MEMORY, and then press ENT to display the MEMORY MENU.
3. Use the buttons to highlight one of the following:
• CLEAR MEMORY DATA clears only the individual saved combustion test records.
• CLEAR LOGGING DATA clears only the combustion test records that were saved during the logging process.
• CLEAR ALL MEMORY clears all memory locations in both memory banks.
4. Press ENT to display the CLEAR MEMO­RY, LOGGING, or ALL DATA conforma­tion screen. Highlight YES to confirm that memory is to be cleared, and then press ENT to actually clear memory as evidenced by the display of the “WAIT Erasing memory” screen.
Instruction 24-94484-20
Operation
4.16 Downloading Stored Data to a Computer
The combustion test data that was stored in either the analyzer’s Memory Directory (Section 4.9), or Logging Directory (Section 4.14), can be down­loaded to a computer using the PCA 2 Data Recovery Program and USB cable that are supplied with the analyzer.
The following procedures assume that the operator is familiar with creat­ing folders and navigating the file structure of the Windows operating sys­tem. If necessary, consult the Windows help files for instructions on how to perform these procedures.
The downloaded data is stored on the computer’s hard drive – or removable media of the operator’s choosing – as a comma-separated-value ASCII text file with a CSV extension. This type of file can be opened by most spread­sheet programs for analysis.
Computer requirements:
• Windows 98SE or higher
• CD ROM drive
• USB 1.1 or USB 2.0 port
• 12 MB of hard drive space for the PCA 2 data recovery program, plus up to an additional 350 KB for each download
Instruction 24-9448 4-21
Operation
4.16.1 PCA 2 Data Recovery Program Installation
The PCA 2 Data Recovery Program is supplied with the analyzer on a CD (P/N 24-1448). Install this program as follows:
1. Insert the PCA 2 CD into the computer’s CD-ROM drive.
2. Locate the CD-ROM drive in Windows Explorer and open the PCA2 folder. Double-click the Setup.exe program to start the installation process.
3. Click Next on the “Welcome to the InstallShield Wizard for PCA2” window.
4. Click Next to accept the default destination folder, or click Change to install to a different folder.
Default Installation Folder
Click to change installation folder
Instruction 24-94484-22
Operation
5. Click Install to confirm the destination folder and user information.
6. Click Finish after all files have been copied into the destination folder.
7. At this time the PCA2 Data Recovery Program icon should appear under "Start All Programs menu.
8. This completes the installation of the software. Remove CD from drive.
Instruction 24-9448 4-23
Operation
4.16.2 Computer to PCA 2 Connection &
USB Device Driver Installation
Connect the USB cable (P/N 104-4032) that was supplied with the PCA 2, and, if necessary, install the analyzer’s USB device driver as follows:
1. Insert the PCA 2 CD into the computer's CD-ROM drive.
2. With both the PCA 2 and computer turned ON, connect the PCA 2 to the computer using the USB cable supplied with the instrument (see Figure 4-2 in instruction manual)
3.. If this is the first time the PCA 2 is being connected to the computer, then the “Found New Hardware Wizard” should shortly appear. Select “No, not this time” and click Next.
NOTE: The USB device driver only needs to be installed once. It does not require to be re-installed each time the PCA 2 is con­nected to the computer.
Figure 4-2. Computer to PCA 2 Connection
Instruction 24-94484-24
Operation
4. Select “Install the software automatically (Recommended)” and click Next.
Name of new USB hardware
Select (Advanced).
5. At the Hardware Installation window, click Continue Anyway.
6. After the Found New Hardware Wizard has finished installing the software, click Finish to close the Wizard.
Instruction 24-9448 4-25
Operation
4.16.3 Recovering Data
Before data can be recovered from the analyzer, install the PCA 2 Data Recovery Program, USB cable, and USB device driver as described in Sec­tions 4.14.1 and 4.14.2.
Recover either the logging or combustion data as follows:
1. Start the program by either double-clicking the PCA2 icon on the Windows deskt op, or click ing the PCA 2 shortcut in start menu. The “Bacharach PCA 2 Data Recovery Program” window should appear.
Observe that the window is divided into two sections. The left-hand
section is used to recover logging data, while combustion data is recov­ered from the right-hand section.
In the following steps, the logging data will be recovered. The same
procedure can be used to recover combustion data using the right­hand section of the window.
2. Select where the downloaded file containing the recovered data will be located and give it a filename by clicking the Open Log File button. The following example screens show that the downloaded file will be placed into a pre-existing PCA2 Logging Data folder and given the filename Customer XYZ.
Instruction 24-94484-26
Operation
Use this section
of window
to Recover
Logging Data
Status line
showing PCA 2
is successfully
connected to
the computer
Select a
pre-existing
folder or create
a new folder
to store the
downloaded le
Enter lename
File type preset
to *.cvs
Use this section of window to Recover Combustion Data
Start the download process by clicking the desired Open Log / Data File button
Click here to create a new folder
Click Open after selecting a folder and entering a
lename
Location of
downloaded le
Click Start
to begin the
download
process
Total number
of records
received out of
512 total
Status line
showing les
are currently
being received
Instruction 24-9448 4-27
Operation
NOTE: The filename is automatically given a CSV (Comma Separated Value) extension, allowing the file to be directly opened by most spreadsheet programs for analysis.
3. After selecting a folder and entering the filename, click Open and then click Start to begin the download process.
While recovering the stored data, observe that the “Records received”
line displays the total number of records currently downloaded. And after all records have been received, the Status line will read “Log files finished.” The PCA 2 can download 500 combustion, pressure, and temperature records from memory, and an additional 500 records from logged memory.
4. This completes the process of recovering logging data. Either click Exit to close the program, or proceed to recover the combustion data by starting over at Step 2, but this time using the right-hand section of the “Bacharach PCA 2 Data Recovery Program” window.
NOTE: Once either the logging data or combustion data has been recovered, that portion of the recovery program’s window will turn gray.
Status line
showing
recovery
process has
nished
Click Exit to end the program
Instruction 24-94484-28
Operation
4.17 Importing Saved Data Into a Spreadsheet
Data that was recovered and saved as an ASCII text file with a “CSV” extension, as described in Section 4.14, can easily be opened for viewing in most spreadsheet programs by simply double-clicking the filename. For example: double-clicking the filename Customer XYZ.csv should automati­cally open the spreadsheet program and display the contents of the file.
If the spreadsheet program does not recognize the “CSV” file extension, then refer to the spreadsheet’s documentation for information on how to manually import comma-delimited text files.
Table 4-2 contains a listing and description of the 27 data fields that are downloaded with each data record. Note that for sensors that are not installed, all related data fields for those sensors will be marked as “N/A” (Not Applicable). For example: if the analyzer does not contain an SO2 sen­sor, then “N/A” will appear in data fields 19 and 20.
Instruction 24-9448 4-29
Operation
TABLE 4-2. DOWNLOADED DATA FIELDS
Field Column Name Data Name or Value
1 Date Date of Test in mm/dd/yy 2 Time Time of Test in 12 hour format (AM/PM)
3-5 Test ID Information Input by User
6 Fuel Fuel Name 7 %O2 Oxygen Level 8 CO Carbon Monoxide Level (see field 26) 9 ppm CO(O2) Carbon Monoxide Level referenced to a
% of O2 (see field 21) 10 %Efficiency Combustion Efficiency 11 %CO2 Carbon Dioxide Level
12 TStk Stack Temperature 13 TAir Primary / Ambient Air Temperature 14 TempUnit Temperature Units 15 %EA Excess Air
16 NO Nitric Oxide Level (see field 26) 17 ppm NO(O2) Nitric Oxide Level referenced to a % of O
(see field 22) 18 NO2 Nitrogen Dioxide Level (see field 26) 19 ppm NO2(O2) Nitrogen Dioxide Level referenced to a
% of O
(see field 22)
2
20 NOx Nitrogen Oxides (NO + NO2) (see field 26)
21 ppm NOx(O2) Nitrogen Oxides Level referenced to a
% of O
(see field 22)
2
22 SO2 Sulfur Dioxide (see field 26) 23 ppm SO2(O2) Sulfur Dioxide Level referenced to a
% of O 24 CO_O2Ref %O2 Reference for CO 25 NOx_O2Ref %O2 Reference for NOx 26 SO2 _O2Ref %O2 Reference for SO
(see field 23)
2
(1)
(1)
(1)
2
27 Pressure Pressure (Draft) Value 28 Pressure Units Pressure Units of Measure 29 Pollution Units Pollution Units of Measure
(2)
(3)
30 PS Voltage Power Supply Voltage 31 NO Temp Nitric Oxide Sensor Temperature (4)
2
Instruction 24-94484-30
Operation
(1) %O2 reference as selected per Section 3.12 (2) Pressure units as selected per Section 3.8 (3) Pollution units as selected per Section 3.9
(4) Shown only on printout, not on LCD display.
4.18 Printing Test Data
Combustion or pressure data that is currently being displayed can be sent to a printer using IrDA protocol as described below.
Data that is stored in memory can also be printed by first displaying the stored test data as described in Sections 4.13.1 & 4.13.2.
In addition to printing combustion and pressure data, the contents of any screen that shows the label “PRINT” above the F1 button can be printed. For example, the information shown in the DIAGNOSTIC screen can be printed.
1. Turn ON printer. Refer to the printer’s instruction manual for detailed operating information. If not already done, set up the printer for:
• 8 bit
• No parity
• 9600 baud
• IrDA is set to IrDA-SIR
• DTR handshaking
2. Align the printer with the top of the analyzer as shown in Figure 4-6.
3. Press the PRINT (F1) button to begin printing.
The printout shown in Figure 4-6 shows typical combustion-test re­sults of an analyzer containing O2, CO that since the CO
and SO2 sensors are not installed, three stars
HIGH
, NO, and NO2 sensors. Note
LOW
(∗∗∗) appear in their data fields.
Instruction 24-9448 4-31
Operation
BACHARACH, INC.
PCA 2
SN: xxxxxx ===================== TIME 01:00:00 PM DATE 10/06/06
FUEL NATURAL GAS
O2 4.0 % CO 12 ppm EFF 82.6 % CO2 9.5 % T-STACK 374 °F T-AIR 68.0 °F EA 21 % NO 18 ppm NO Temp 70.5 °F NO2 6 ppm NOX 24 ppm SO2 ∗∗∗ ppm CO(3) 13 ppm NO(3) 19 ppm NO2(3) 6 ppm NOX(3) 25 ppm SO2(3) ∗∗∗ ppm
PRESSURE – 0.25 inwc
COMMENTS:
PRINT MENU SAVE
3 to 18 in.
(7.62 to 45.72 cm)
Optional pressure printout as selected per Section 3.13.
Figure 4-3. Printer Alignment & Sample Printout
NOTES: Three stars (∗∗∗) appear in the data
eld of sensors that are not calibrated or installed, and in the data elds of
that sensor’s related calculated values.
Three hyphens or dashes (- - -) indicate that the calculated data cannot be displayed because the measured data necessary to make the calculation is out of range (i.e., oxygen level above 16%).
Three (XXX) indicate sensor over­range.
Instruction 24-94484-32
Operation
Instruction 24-9448 4-33
Calibration
5 Calibration
IMPORTANT: Before performing any calibration procedure,
ensure that fresh batteries are installed or use the optional AC power adapter. Also ensure that the analyzer is at room temper­ature and will be sampling fresh air when turned ON.
5.1 Smart Sensors
The PCA 2 uses Bacharach’s new “Smart Sensor” technology, meaning that the calibration data for each sensor is stored in nonvolatile memory on the sensor’s printed circuit board.
Benefits of Smart Sensors:
New sensors can be installed without needing to be calibrated.
• Sensors can be pre-calibrated and installed when needed.
• Sensors can be moved from one analyzer to another.
• The analyzer’s diagnostics screen shows what sensors are installed,
• and their current end-of-life condition (Good, Low, Bad). Each sensor stores its calibration history and operating parameters,
• which can be downloaded to a computer and analyzed. This informa­tion is useful in determining when the sensor was last calibrated, and predicting its end-of-life.
The smart sensors should be calibrated by an authorized Bacharach Service Center (Section 8.3) every 6 months to 1 year to assure that the analyzer continues to meet its published accuracy specifications. The smart sensors, however, can be calibrated in the field if your facility has the necessary equipment and qualified personnel to perform the proce­dures described in the following sections of this instruction manual.
5.2 Starting a Calibration
Start any calibration procedure by doing the following:
1. Place the probe in an area of fresh air, turn ON the analyzer, allowing it to cycle through its 60 second warm-up period. During warm-up, the analyzer’s operation is checked and the sensors are set to the fol­lowing ambient conditions:
Oxygen sensor spanned to 20.9%
• All gas sensors are zeroed
• The pressure sensor is zeroed
Instruction 24-9448 5-1
Calibration
ERRORS DETECTED Low Battery
MENU
F1 F2 F3
MAIN MENU FUEL PRESSURE TEMPERATURE MEMORY SETUP
CALIBRATION
DIAGNOSTICS STATUS
F1 F2 F3
MENU
Calibration Password
Enter Password
Menu
F1 F2 F3
Any errors detected during warm-up will
be listed on the display immediately fol­lowing warm-up. For example, the screen to the right shows that the battery is low. Correct any errors before proceeding. Refer to Section 7.3 for a listing of error messages and their meaning.
2. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
3. Use the buttons to highlight CALIBRATION, and then press ENT to dis­play the CALIBRATION PASSWORD screen.
4. Before calibration can begin a 4-place alpha­numeric password must be entered. Use the
and buttons to enter the password.
(Note that the default password is 1111)
5. Press the ENT key to accept the password. If the correct password was entered the CALIBRATION Menu will be displayed.
6. Use the buttons to highlight the sensor to be calibrated, and then perform the calibra­tion procedure for that sensor as described in the following sections.
5.3 Pressure Sensor
Calibration
This procedure calibrates the pressure sensor to a known pressure value.
Material Required:
Bellows
• Manometer
Range: ±8 in. of water column (±20 mb)
­Accuracy: ±0.01 in. of water column (±0.025 mb)
-
Procedure:
NOTE: The unit-of-measure for pressure is selected per
Instruction 24-94485-2
Section 3.8. In the following procedure inwc is
POWER
P
T-STACK T-AIR
GAS
4 3 2 1 0
-1
-2
-3
-4
Parts:
1. 3/16" I.D. Tubing
2. Tee
3. Barbed Hose Connector
4. Bellows
5. Manometer
1
2
3
4
5
CALIBRATION MENU
Pressure
T-Stack T-Air CO-LO SO2 NO NO2
(more)
MENU
F1 F2 F3
CALIBRATE PRESSURE Measured 0.00 inwc Applied -4.00 inwc
Press ENT to calibrate
06/27/06 11:55:30 PM
PRINT
F1 F2 F3
selected, but note that any unit-of-measure can be used for calibration purposes.
1. Assemble the pressure sensor calibration equipment as shown in Figure 5-1, but DO
NOT connect the analyzer to the calibration equipment at this time.
2. If not already done, turn ON the analyzer and display the CALIBRATION LIST screen per Section 5.2.
3. Use the buttons to highlight Pressure, and then press ENT to display the CALIBRATE PRESSURE screen.
“Measured” is the pressure value currently
being detected by the pressure sensor, while “Applied” is a known value of pressure that will be applied for calibration purposes.
4. With both the –∆P and +∆P ports open to the
atmosphere, observe that the current Measured pressure reading should be 0 ±0.01 inwc. If necessary, zero the pressure sensor per Section 4.6, and repeating Steps 2 thru 4.
Calibration
5. Connect the hose from the manometer to the +∆P port and apply a
negative pressure to this port by adjusting the bellows for a manom­eter reading of –4.00.
6. Use the and buttons to enter an Applied value that exactly equals the manometer reading.
Instruction 24-9448 5-3
Figure 5-1. Pressure Sensor Calibration Equipment
Calibration
CALIBRATION MENU Pressure
T-Stack
T-Air CO-LO SO2 NO NO2
(more)
MENU
F1 F2 F3
The calibration range is from –6 to –2 inwc (–15 to –5 mb). An attempt
to calibrate outside this range will cause the message “Applied Value High” (or Low) to appear at the bottom of the screen.
7. Wait until the Measured reading stabilizes, and then press ENT to cali­brate the pressure sensor’s Measured value to that of the Applied value; after which the message “Good Calibration” should briefly appear fol­lowed by the CALIBRATION LIST screen being re-displayed.
8. Remove calibration equipment.
5.4 T-Stack Calibration
This procedure first zeros and then spans the stack-temperature channel to known temperature values.
The use of an electronic thermocouple simulator is the preferred method of producing the desired calibration temperatures. Alternatively, containers of ice water and boiling water can be used.
Material Required:
Thermocouple Simulator (K-type)
• Range: 0 to 600 °F
­Accuracy: ±0.5 °F
-
(Alternatively) Ice-Water, Boiling Water, Thermometer
TS-Zero Procedure:
1. Set thermocouple simulator to room tempera­ture and plug its output into the T-STACK connector located at the bottom of the analyzer.
Alternatively: Plug the probe’s thermocouple
into the T-STACK connector located at the bottom of the analyzer. DO NOT attach the
probe’s gas hose to the analyzer’s GAS port; otherwise water will be drawn into the analyzer!
2. If not already done, turn ON the analyzer and display the CALIBRATION LIST screen per Section 5.2.
3. Use the buttons to highlight T-Stack, and then press ENT to dis­play the CALIBRATE TS-ZERO screen.
“Measured” is the current temperature reading, while “Applied” is a
Instruction 24-94485-4
known temperature that will be applied for
CALIBRATE TS-ZERO Measured 34 °F Applied 032 °F
Press ENT to calibrate
06/27/06 11:55:30 PM
PRINT
F1 F2 F3
CALIBRATE TS-SPAN Measured 574 °F Applied 572 °F
Press ENT to calibrate
06/27/06 11:55:30 PM
PRINT
F1 F2 F3
calibration purposes.
4. Set thermocouple simulator to 32 °F (0 °C), and then use the and buttons to enter an Applied value that exactly equals the set­ting of the simulator.
Alternatively: Submerge probe tip into an ice-
water bath with a thermometer, wait several minutes, and then use the and buttons to enter an Applied value that exactly equals the thermometer reading.
The calibration range is from 32 to 41 °F (0
to 5 °C). An attempt to calibrate outside this range will cause the message “Applied Value High" (or Low) to appear at the bottom of the screen.
5. Wait until the Measured reading stabilizes, and then press ENT to calibrate the TS-Zero Measured value to that of the Applied value; after which the message “Good Calibration” should briefly appear fol­lowed by the CALIBRATE TS-SPAN screen.
Calibration
TS-Span Procedure:
6. Set thermocouple simulator to 572 °F (300 °C), and then use the
and buttons to enter an Applied value that exactly equals the
setting of the simulator.
Alternatively: Submerge probe tip into a container of boiling water with
a thermometer, wait several minutes, and then use the and buttons to enter an Applied value that exactly equals the thermometer reading.
The calibration range is from 175 to 625 °F (80 to 330 °C). An attempt
to calibrate outside this range will cause the message “Applied Value High” (or Low) to appear at the bottom of the screen.
7. Wait until the Measured reading stabilizes, and then press ENT to cali­brate the TS-Span Measured value to that of the Applied value; after which the message “Good Calibration” should briefly appear followed by the CALIBRATION LIST screen being re-displayed.
Instruction 24-9448 5-5
Calibration
CALIBRATION MENU Pressure T-Stack
T-Air
CO-LO SO2 NO NO2
(more)
MENU
F1 F2 F3
CALIBRATE TA-ZERO Measured 33 °F Applied 032 °F
Press ENT to calibrate
06/27/06 11:55:30 PM
PRINT
F1 F2 F3
5.5 T-Air Calibration
This procedure first zeros and then spans the ambient-temperature chan­nel to known temperature values.
The use of an electronic thermocouple simulator is the preferred method of producing the desired calibration temperatures. Alternatively, containers of ice water and boiling water can be used.
Material Required:
Thermocouple Simulator (K-type)
• Range: 0 to 600 °F
­Accuracy: ±0.5 °F
-
(Alternatively) Ice-Water, Boiling Water, Thermometer
TA-Zero Procedure:
1. Set thermocouple simulator to room temperature and plug its output into the T-AIR connector located at the bottom of the analyzer.
Alternatively: Plug the probe’s thermocouple into the T-AIR connector
located at the bottom of the analyzer. DO NOT attach the probe’s
gas hose to the analyzer’s GAS port; otherwise water will be drawn into the analyzer!
2. If not already done, turn ON the analyzer and display the CALIBRATION LIST screen per Section 5.2.
3. Use the buttons to highlight T-Air, and then press ENT to display the CALIBRATE TA-ZERO screen.
“Measured” is the current temperature read-
ing, while “Applied” is a known temperature that will be applied for calibration purposes.
4. Set thermocouple simulator to 32 °F (0 °C), and then use the and buttons to enter an Applied value that exactly equals the set­ting of the simulator.
Alternatively: Submerge probe tip into an ice-
water bath with a thermometer, wait several minutes, and then use the and buttons to enter an Applied value that exactly equals the thermometer reading.
Instruction 24-94485-6
Calibration
CALIBRATE TA-SPAN Measured 210 °F Applied 212 °F
Press ENT to calibrate
06/27/06 11:55:30 PM
PRINT
F1 F2 F3
The calibration range is from 32 to 41 °F (0 to 5 °C). An attempt to
calibrate outside this range will cause the message “Applied Value High” (or Low) to appear at the bottom of the screen.
5. Wait until the Measured reading stabilizes, and then press ENT to calibrate the TA-Zero Measured value to that of the Applied value; after which the message “Good Calibration” should briefly appear fol­lowed by the CALIBRATE TA-SPAN screen.
TA-Span Procedure:
6. Set thermocouple simulator to 212 °F (100 °C), and then use the and buttons to enter an Applied value that exactly equals the setting of the simulator.
Alternatively: Submerge probe tip into a con-
tainer of boiling water with a thermometer, wait several minutes, and then use the and buttons to enter an Applied value that exactly equals the thermometer reading.
The calibration range is from 194 to 230 °F (90 to 110 °C). An attempt
to calibrate outside this range will cause the message “Bad Calibration Wrong CAL Entry” to appear in the following step.
7. Wait until the Measured reading stabilizes, and then press ENT to calibrate the TA-Span Measured value to that of the Applied value; after which the message “Good Calibration” should briefly appear fol­lowed by the CALIBRATION LIST screen being re-displayed.
Instruction 24-9448 5-7
Calibration
CALIBRATION MENU Pressure T-Stack T-Air
CO-LO
SO2 NO NO2
(more)
MENU
F1 F2 F3
CALIBRATE CO Measured 480 PPM Applied 500 PPM
Press ENT to calibrate
06/27/06 11:55:30 PM
PRINT
F1 F2 F3
5.6 CO-LO Sensor Calibration
Note that the CO-LO sensor also measures H2 for the purpose of com­pensating the CO reading for the presence of H2 in the gas sample. This procedure first spans the CO-LO sensor and, optionally, spans the H2 part of the sensor to known gas levels.
Material Required:
Calibration Kit, P/N 24-7059
• Gas Cylinder: 500 ppm CO in air, P/N 24-0492
• Gas Cylinder: 1,000 ppm CO & 1,000 ppm H2 in Nitrogen, P/N 24-0794
Procedure:
1. If not already done, turn ON the analyzer and display the CALIBRATION LIST screen per Section 5.2.
2. Use the buttons to highlight CO-LO, and then press ENT to display the CALIBRATE CO screen.
“Measured” is the current CO reading, while
“Applied” is a known CO level that will be applied for calibration purposes.
3. Attach a 500 ppm CO cylinder to the regu­lator of the calibration fixture shown in Figure 5-2 on Page 5-14.
4. Use the and buttons to enter an Applied value that exactly equals the concen­tration stamped on the CO cylinder.
The calibration range is from 250 to
1,000 ppm. An attempt to calibrate outside this range will cause the message “Applied Value High” (or Low) to appear at the bottom of the screen.
5. Adjust the regulator for a flowmeter indica­tion of approximately 2 SCFH. Wait until the Measured reading stabilizes (approximately 3 minutes), and then press ENT to calibrate the CO Measured value to that of the Applied value. The message “Good Calibration” should briefly appear followed by the TEST GAS CO/H2 screen.
If the sensor’s output is low, but still usable, then the message “Good
Calibration WARNING Low Sensor” will appear. The sensor will now be marked as being Low in the DIAGNOSTICS screen.
Instruction 24-94485-8
Calibration
TEST GAS CO/H2 Enter CO VALUE 1000 PPM
06/27/06 11:55:30 PM
PRINT
F1 F2 F3
CALIBRATE H2 Measured 1000 PPM Applied 1000 PPM
Press ENT to calibrate
06/27/06 11:55:30 PM
PRINT
F1 F2 F3
If the sensor’s output is too low to be usable, then the message “Bad
Calibration Sensor End of Life, Entry Not Saved” will appear. The sen­sor will now be marked as being BAD in the DIAGNOSTICS screen.
NOTE: H2 calibration can be bypassed by pressing the ESC button, after which the CALIBRATION LIST screen is re-dis­played. Skip to Step 10 if the ESC button was pressed.
6. Turn OFF the regulator of calibration fixture and remove the CO cylinder.
7. Attach a combination 1,000 ppm CO and 1,000 ppm H2 cylinder to the regulator of the calibration fixture, and then use the and
buttons to enter an Applied value that exactly equals the CO concentration stamped on the cylinder.
8. Adjust the regulator for a flowmeter indica­tion of approximately 2 SCFH, and then press
ENT to display the CALIBRATE H
screen.
2
The calibration range is from 400 to
1,500 ppm. An attempt to calibrate outside this range will cause the message “Bad Calibration Wrong CAL Entry” to appear in the follow­ing step.
9. Wait until the Measured reading stabilizes (approximately 3 minutes), and then press
ENT to calibrate the H
Measured value to
2
that of the Applied value; after which the message “Good Calibration” should briefly appear followed by the CALIBRATION LIST screen being re-displayed.
If the sensor’s output is low, but still
usable, then the message “Good Calibration WARNING Low Sensor” will appear. The sensor will now be marked as being Low in the DIAGNOSTICS screen.
If the sensor’s output is too low to be usable, then the message
“Bad Calibration Sensor End of Life” will appear followed by the CALIBRATION LIST screen being re-displayed. The sensor will now be marked as being BAD in the DIAGNOSTICS screen.
10. Turn OFF the regulator and remove the gas cylinder.
Instruction 24-9448 5-9
Calibration
CALIBRATION MENU Pressure T-Stack T-Air CO-LO
SO2
NO NO2
(more)
MENU
F1 F2 F3
CALIBRATE SO2 Measured 98 PPM Applied 100 PPM
Press ENT to calibrate
06/28/06 11:55:30 PM
PRINT
F1 F2 F3
5.7 SO2 Sensor Calibration
This procedure spans the optional sulfur dioxide sensor to a known gas level.
Material Required:
Calibration Kit, P/N 24-7059
• Gas Cylinder: 100 ppm SO2 in Nitrogen, P/N 24-1158
Procedure:
1. If not already done, turn ON the analyzer and display the CALIBRATION LIST screen per Section 5.2.
2. Use the buttons to highlight SO2, and then press ENT to display the CALIBRATE SO2 screen.
“Measured” is the current SO2 reading, while
“Applied” is a known SO2 level that will be applied for calibration purposes.
3. Attach a 100 ppm SO2 cylinder to the regu­lator of the calibration fixture shown in Figure 5-2 on Page 5-14.
4. Use the and buttons to enter an Applied value that exactly equals the concen­tration stamped on the SO2 cylinder.
The calibration range is from 9 to 150 ppm.
An attempt to calibrate outside this range will cause the message “Bad Calibration Wrong CAL Entry” to appear in the following step.
5. Adjust regulator for a flowmeter indication of approximately 2 SCFH. Wait until the Measured reading stabilizes (approximately 3 minutes), and then press ENT to calibrate the SO2 Measured value to that of the Applied value. The message “Good Calibration” should briefly appear followed by the CALIBRATION LIST screen being re-displayed.
If the sensor’s output is low, but still usable, then the message “Good
Calibration WARNING Low Sensor” will appear. The sensor will now be marked as being Low in the DIAGNOSTICS screen.
Instruction 24-94485-10
Calibration
CALIBRATION MENU Pressure T-Stack T-Air CO-LO SO2
NO
NO2
(more)
MENU
F1 F2 F3
CALIBRATE NO Measured 245 PPM Applied 250 PPM
Press ENT to calibrate
06/28/06 11:45:30 PM
PRINT
F1 F2 F3
If the sensor’s output is too low to be usable, then the message “Bad
Calibration Sensor End of Life, Entry Not Saved” will appear followed by the CALIBRATION LIST screen being re-displayed. The sensor will now be marked as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove gas cylinder.
5.8 NO Sensor Calibration
This procedure spans the optional nitric oxide sensor to a known gas level.
Material Required:
Calibration Kit, P/N 24-7059
• Gas Cylinder: 250 ppm NO in Nitrogen, P/N 24-1156
Procedure:
1. If not already done, turn ON the analyzer and display the CALIBRATION LIST screen per Section 5.2.
2. Use the buttons to highlight NO, and then press ENT to display the CALIBRATE NO screen.
“Measured” is the current NO reading, while
“Applied” is a known NO level that will be applied for calibration purposes.
3. Attach a 250 ppm NO cylinder to the regu­lator of the calibration fixture shown in Figure 5-2 on Page 5-14.
4. Use the and buttons to enter an Applied value that exactly equals the concen­tration stamped on the NO cylinder.
The calibration range is from 9 to 550 ppm.
An attempt to calibrate outside this range will cause the message “Bad Calibration Wrong CAL Entry” to appear in the following step.
5. Adjust regulator for a flowmeter indication of approximately 2 SCFH. Wait until the Measured reading stabilizes (approximately 3 minutes), and then press ENT to calibrate the NO Measured value to that of the Applied value. The message “Good Calibration” should briefly appear followed
Instruction 24-9448 5-11
Calibration
CALIBRATION MENU
(more) T-Air CO-LO SO2 NO
NO2
CO-HI T-Ref
MENU
F1 F2 F3
CALIBRATE NO2 Measured 96 PPM Applied 100 PPM
Press ENT to calibrate
06/28/06 11:45:30 AM
PRINT
F1 F2 F3
by the CALIBRATION LIST screen being re-displayed.
If the sensor’s output is low, but still usable, then the message “Good
Calibration WARNING Low Sensor” will appear. The sensor will now be marked as being Low in the DIAGNOSTICS screen.
If the sensor’s output is too low to be usable, then the message “Bad
Calibration Sensor End of Life, Entry Not Saved” will appear followed by the CALIBRATION LIST screen being re-displayed. The sensor will now be marked as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove gas cylinder.
5.9 NO2 Sensor Calibration
This procedure spans the optional nitrogen dioxide sensor to a known gas level.
Material Required:
Calibration Kit, P/N 24-7059
• Gas Cylinder: 100 ppm NO2 in Nitrogen, P/N 24-1157
Procedure:
1. If not already done, turn ON the analyzer and display the CALIBRATION LIST screen per Section 5.2.
2. Use the buttons to highlight NO2, and then press ENT to display the CALIBRATE NO2 screen.
“Measured” is the current NO2 reading, while
“Applied” is a known NO2 level that will be applied for calibration purposes.
3. Attach a 100 ppm NO2 cylinder to the regu­lator of the calibration fixture shown in Figure 5-2 on Page 5-14.
4. Use the and buttons to enter an Applied value that exactly equals the concen­tration stamped on the NO2 cylinder.
The calibration range is from 9 to 150 ppm.
An attempt to calibrate outside this range will cause the message “Bad Calibration Wrong CAL Entry” to appear in the following step.
Instruction 24-94485-12
Calibration
CALIBRATION MENU
(more) T-Air CO-LO SO2 NO NO2
CO-HI
T-Ref
MENU
F1 F2 F3
5. Adjust regulator for a flowmeter indication of approximately 2 SCFH. Wait until the Measured reading stabilizes (approximately 3 minutes), and then press ENT to calibrate the NO2 Measured value to that of the Applied value. The message “Good Calibration” should briefly appear followed by the CALIBRATION LIST screen being re-displayed.
If the sensor’s output is low, but still usable, then the message “Good
Calibration WARNING Low Sensor” will appear. The sensor will now be marked as being Low in the DIAGNOSTICS screen.
If the sensor’s output is too low to be usable, then the message “Bad
Calibration Sensor End of Life, Entry Not Saved” will appear followed by the CALIBRATION LIST screen being re-displayed. The sensor will now be marked as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove gas cylinder.
5.10 CO-HI Sensor Calibration
This procedure spans the optional carbon monoxide high sensor (4,001 to 20,000 ppm) to a known gas level.
Material Required:
Calibration Kit, P/N 24-7059
• Gas Cylinder: 4,000 ppm CO in Air, P/N 24-1155
Procedure:
1. If not already done, turn ON the analyzer and display the CALIBRATION LIST screen per Section 5.2.
2. Use the buttons to highlight CO-HI, and then press ENT to display the CALIBRATE CO-HI screen.
“Measured” is the current CO reading, while “Applied” is a known CO
level that will be applied for calibration purposes.
3. Attach a 4,000 ppm CO cylinder to the regulator of the calibration fixture shown in Figure 5-2 on Page 5-14.
4. Use the and buttons to enter an Applied value that exactly equals the concen­tration stamped on the CO cylinder.
The calibration range is from 250 to
11,000 ppm. An attempt to calibrate outside
Instruction 24-9448 5-13
Calibration
Parts:
1. Gas Cylinder
2. Regulator *
3. 3/16" I.D. Tubing *
4. Tee *
5. Flowmeter *
6. Barbed Hose Connector * * Part of Calibration Kit 24-7059
POWER
P
T-STACK T-AIR
GAS
Leave Top Port Open
GAS Port
CALIBRATE CO-HI Measured 3895 PPM Applied 4000 PPM
Press ENT to calibrate
06/28/06 09:45:30 AM
PRINT
F1 F2 F3
this range will cause the message “Bad Calibration Wrong CAL Entry” to appear in the following step.
5. Adjust regulator for a flowmeter indica­tion of approximately 2 SCFH. Wait until the Measured reading stabilizes (approxi­mately 3 minutes), and then press ENT to calibrate the CO Measured value to that of the Applied value. The message “Good Calibration” should briefly appear followed by the CALIBRATION LIST screen being re-displayed.
If the sensor’s output is low, but still usable, then the message “Good
Calibration WARNING Low Sensor” will appear. The sensor will now be marked as being Low in the DIAGNOSTICS screen.
If the sensor’s output is too low to be usable, then the message “Bad
Calibration Sensor End of Life, Entry Not Saved” will appear followed by the CALIBRATION LIST screen being re-displayed. The sensor will now be marked as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove gas cylinder.
Figure 5-2. Gas Sensor Calibration Equipment
Instruction 24-94485-14
Maintenance
6 Maintenance
Customer maintenance of the PCA 2 is limited to the following:
Battery replacement or charging using AC power adapter (Section 3.2)
• Sensor re-calibration (Section 5)
• Water trap / filter assembly maintenance (Section 6.2)
• Sensor replacement (Section 6.3)
• Probe thermocouple replacement (Section 6.6)
• Pump (gas or purge) replacement (Section 6.7)
• Cleaning the probe (Section 6.8)
All other maintenance should be performed by an authorized Bacharach Service Center. Refer to Section 8.3.
Instruction 24-9448 6-1
Maintenance
Rear Case Screws (typical 4 places)
Release and then remove battery cover by pushing latch up
6.1 PCA 2 Disassembly
The following procedure describes how to disassemble the analyzer, while Figures 6-1 thru 6-4 illustrate how the analyzer is put together.
Tools Required:
Medium Phillips Screwdriver
Procedure:
1. Unplug all thermocouples from bottom of analyzer.
2. Remove battery cover and then remove batteries.
TIP: In Step 3, if the sensors are not being replaced, leave the tubing connected to each sensor’s gas cap, being careful not to put unneces­sary strain on the tubing during the disassembly process.
3. Pull off sensor retainer; and
Figure 6-1. Removing Rear Case
then unplug all sensors.
4. Lay analyzer face down on a padded work surface; and then using a medium Phillips screwdriver, remove the unit’s four rear-case screws.
5. Lift rear case from analyzer and set aside.
6. Unplug electrical connectors J8, J9, and J14 from printed circuit board.
CAUTION: In Step 7, note that there is tubing connected between the bottom hose-connector plate and the pressure sensor on the printed circuit board. Do not put unnecessary strain on this tubing during the disassembly process.
7. Carefully lift battery-and-pump chassis, along with the bottom hose-
8. Lift printed circuit board from analyzer.
connector plate, from analyzer.
Instruction 24-94486-2
Sensor Positions
Sensor Retainer
Purge Pump
Gas Pump
GAS
Purge Air In
Pressure
+∆P
–∆P
Battery & Pump Chassis
Hose Connector Plate
Purge Pump Motor Connector
Gas Pump Motor Connector
Sensors
Type and position determined by model number
Battery Connector on printed circuit board
Pressure Sensor on printed circuit board
+∆P (Top) –∆P (Bottom)
Model P/N
Position
#1 O
#2 COLOCOLOCOLOCOLOCOLOCO
#3 SO2NO2SO
#4 NO CO
225
24-7301
2
235
24-7302
O
2
245
24-7303
O
255
24-7304
2
HI
265
24-7305
O
2
275
24-7306
O
O
2
NO NO
Maintenance
2
LO
2
Figure 6-2. Inside View with Rear Case Removed
Instruction 24-9448 6-3
Maintenance
Models 235 & 245
P/Ns 24-7302 & 24-7303
Model 255
P/N 24-7304
Models 265 & 275
P/Ns 24-7305 & 24-7306
Model 225
P/N 24-7301
Elbow
PRESSURE SENSOR
on printed circuit board
–∆P
+∆P
–∆P (Bottom)
+∆P (Top)
PRESSURE
PURGE AIR
GAS
Gas IN
GAS PUMP
PURGE PUMP
Gas OUT
OUTIN
To Sensors
Elbow
Elbow
Elbow
Figure 6-3. Tubing Connections
Instruction 24-94486-4
Maintenance
Solder
(2 Places)
Red dot on pump indicates positive (red wire) connection
Battery to J14
Battery to J14
Gas Pump to J8
Purge Pump to J9
Sensor Connector (typical 4 places)
Factory Test Connector
LCD Connector (opposite side of board)
Ribbon Cable Conection
between LCD Assembly and
Printed Circuit Board
Lock Tabs (both ends)
J8
J9
J14
Wire Routing
Solder Connections
Connectors
Instruction 24-9448 6-5
Figure 6-4. Wiring
Maintenance
6.2 Water Trap / Filter Maintenance
6.2.1 Emptying the Water Trap Chamber
The water trap chamber should be emptied after every test, or when the water condensate approaches the tip of the riser tube (refer to Section 4.9).
1. Remove water trap chamber per Figure 6-5.
2. Pour out liquid condensate, and then reassemble trap.
6.2.2 Replacing the Filter Element
Replace the filter element when it becomes visibly dirty or becomes satu­rated with water.
Material Required: • Filter Element, P/N 07-1644
• Small Flat Blade Screwdriver
1. Remove water trap chamber per Figure 6-5.
2. Pry apart filter chamber using a small flat-blade screwdriver. Remove and discard old filter.
3. Install new filter and reassemble filter chamber, making sure that surfaces “A” and “B” contact each other.
4. Reassemble trap.
Figure 6-5. Disassembling the Water Trap / Filter Assembly
Instruction 24-94486-6
Maintenance
Gas Cup
Sensor Retainer
Sensor and Printed Circuit Board
Twist gas cap bayonet fitting counterclockwise to release
Sensor electrical connector mates with pins on analyzer’s main board
6.3 Smart Sensor Replacement
Bacharach’s Smart sensor technology allows new sensors to be installed without needing to be calibrated inside the analyzer.
NOTE: Sensors may be purchased with and without the Smart sensor technology PCB. See Section 6.4 on how to change just the sensor itself. Calibration must be performed if the sensor is purchased without the Smart technology PCB. Refer to Section 8.1 Replacement Parts for list of sensors and part numbers of both types.
Replace the oxygen sensor when its automatic calibration fails and the ana­lyzer displays an O2 sensor error at the end of warm up. Refer to Section 4.2.
NOTE: The O2 sensor and PCB is not Smart sensor technology. Only the sen­sor itself, not the PCB, needs to be purchased when the sensor needs replaced.
Replace a gas sensor when it can no longer be calibrated, as evidenced by “Bad Calibration Sensor End of Life” being displayed during its calibration procedure. Refer to Sections 5.6 thru 5.10.
Refer to Section 8.1 for a listing of sensor part numbers. Also, see Figure 6-2 on page 6-3 for sensor positions.
Do the following to replace a Smart sensor:
1. Turn OFF the analyzer.
2. Remove the battery cover.
3. Remove the sensor retainer.
4. Remove the tubing from the gas cup of sensor being replaced.
5. Unplug the sensor along with its printed circuit board from the analyzer’s main board.
6. Remove the sensor’s gas cup by twisting its bayonet fitting counterclockwise, and then pulling the cup straight up.
7. Attach the gas cup to new sensor, making sure that when the sensor is plugged into analyzer’s main board, the tubing connections on top of gas cup are aligned as shown in Figure 6-3 on page 6-4.
8. Plug the sensor into analyzer; and then reattach tubing to gas cup.
9. Install the sensor retainer and battery cover.
10. Turn ON the analyzer and confirm that no sensor errors occur during warm up.
Instruction 24-9448 6-7
Maintenance
Gas Cup
Sensor Retainer
Sensor
Twist gas cap bayonet fitting counterclockwise to release
Sensor electrical connector mates with pins on analyzer’s main board
Printed Circuit Board
Adhesive foam disk
Be sure to remove any shorting spring from the new sensor pins before installing! There is no shorting spring on the O2 or NO sensors.
6.4 Sensor Only Replacement
All sensors are replaced in a similar manner. Do the following to
1. replace either the O2, CO-LO, CO-HI, NO, NO2, or SO2 sensor. Refer
to Section 8.1 Replacement Parts for list of sensors and part numbers.
Turn OFF the analyzer.
2.
Remove the battery cover.
3.
Remove the sensor retainer.
4.
Remove the tubing from the gas cup of sensor being replaced.
5.
Unplug the sensor along with its printed circuit board from the analyz-
6. er’s main board.
Remove the sensor’s gas cup by twisting its bayonet fitting counter-
7. clockwise, and then pulling the cup straight up.
Carefully remove the old sensor from the Smart sensor PCB, keeping
8. in mind that there is an adhesive foam disk in between the PCB and sensor. Gently remove any adhesive from from the PCB.
Important! When replacing the O2 sensor, be sure that the “+” sensor pin plugs into the PCB socket that is also marked “O2+”.
When replacing an NO sensor, also replace the bias battery.
9. Remove the replacement sensor from its plastic canister. Note that the CO-LO, CO-HI, SO2, and NO2 sensors have a shorting-spring installed between two of their pins. Remove and discard this spring before installing the sensor.
10. Place the adhesive disk on the bottom of the new sen­sor and plug the new sen­sor into the sensor PCB.
11. Attach the gas cup to new sensor, making sure that when the sensor assembly is plugged into the analyz­er’s main board, the tubing connections on top of gas cup are aligned as shown in Figure 6-3 on page 6-4.
Instruction 24-94486-8
Maintenance
12. Plug the sensor into analyzer; and then reattach tubing to gas cup.
13. Install the sensor retainer and battery cover.
14. Allow the sensor that was just installed time to stabilize in the circuit before continuing with this procedure. Stabilization time for all sen­sors (except for the NO sensor) is about 1 hour. The NO sensor baseline technically requires several days to stabilize but should be sufficiently stabilized for use in approximately 4 hours.
15. Turn ON the analyzer and confirm that no sensor errors occur during warm up.
Note: Discard an old sensor in accordance with local and federal haz­ardous waste disposal laws.
16. Calibrate the new sensor(s) per Section 5 (except the O2 sensor which does not require calibration).
Instruction 24-9448 6-9
Maintenance
6.5 Nitric Oxide Sensor Battery Replacement
A single lithium battery, located on the NO Smart Sensor assembly, applies a bias voltage to the NO sensor to prevent the sensor from destabilizing when the analyzer is turned off. The NO bias battery is expected to last at least the life of the NO sensor.
Note: It is recommended that the bias battery be replaced whenever the NO sensor is replaced.
Material Required:
• Bias battery (refer to Section 8.1 Replacement Parts)
Procedure:
1. Follow the instructions in Section 6.3 Smart Sensor Replacement to remove the Smart Sensor assembly from position 4.
2. Remove the old battery from its holder on the Nitric Oxide printed circuit board.
3. Insert the new battery with the positive side facing away from the sensor (the battery contact is stamped with a + symbol) .
4. Re-install the Smart Sensor assembly.
5. Before powering up and using the analyzer, allow the Nitric Oxide sensor to stabilize as described below. Note that recalibration of the Nitric Oxide sensor is not required after replacing its bias battery. Depending on how long the Nitric Oxide sensor was without bias voltage, the time required for the sensor to completely stabilize varies from less than a minute to several days. Typical stabilization times are shown below. Generally, however, the sensor is sufficiently stable after 4 hours for measurement purposes.
Bias battery removed for . . . Stabilization time
Less than 15 minutes Less than 1 minute
Less than 1 hour Less than 5 minutes
Less than 2 days Less than 4 hours
Greater than 2 days Up to 2 days
Instruction 24-94486-10
Loading...