1. Overview
These instructions describe how to install the NOx Filter into the probe and hose assembly of various Bacharach
combustion gas analyzers. The NOx filter removes cross-interfering acid gases, including NO, NO
and SO2, that
2
could affect the CO measurement.
2. Items Required
Replacement NO
•
Probe and Hose Assembly
•
Scissors
•
Filter (P/N 0024-1505)
x
3. Procedure
Use the following procedure and Figures 1, 2, and 3 to insert the NOx Filter in-line with the sampling system.
Step Description
1. Locate Locate the gas sample hose. Refer to Figure 1 for examples, or check the documentation
associated with your gas monitoring instrument.
2. Separate If the gas sample hose is connected to an additional parallel hose (e.g., draft hose or T-Stack
hose), separate a portion of the gas sample hose (approximately 5” or 12.7 cm) from that
parallel hose using scissors to carefully cut along the connective webbing. Be sure to cut ONLY
the connective webbing between the hoses and not the hoses themselves. Refer to Figure 1 and
Figure 2.
3. Remove Using scissors, carefully cut and remove a piece of the gas sample tubing (approximately 1.5” or
3.8 cm in length) where the NOx filter will be placed. Make the first cut approximately 2” (5.1
cm) from the end of the gas sample tubing. Cut the hose and place the 2” end piece (with metal
connector if applicable) aside. Make another cut approximately 1.5” (3.8 cm) from the previous
cut and discard the 1.5” (3.8 cm) section of hose. Refer to Figure 1, Figure 2, and Figure 3.
4. Insert Insert the NOx filter in-line where the piece of gas sample hose was removed by pushing each
section of tubing about halfway onto the NOx filter barbs. Note that the NOx filter is not
direction dependent.
5. Trim In applications that have a draft hose, trim excess tubing from the end piece of the gas sample
hose so it and the draft hose are the same length.
NOTE: Filter life is dependent on the application and has a capacity of approximately 3000 ppm•HR.
When new, the filter material is purple, but will change to dark brown as the filter loses its capacity to
remove acid gas.
NOTE: A barbed hose connector is included in the NOx Filter Kit to reconnect the gas sample hose
when the NOx filter is not in use.
IMPORTANT: The housing of the NO
glycol, as these substances will cause the NO
0024-9469 Rev 2 1
filter is made of nylon. Do not expose it to alcohol, solvents, or
x
filter to deteriorate prematurely and fail.
x
NOx Filter Installation
Figure 1. NOx Filter Placement Example 1 – Gas Sample Hose with Integrated Draft and T-Stack Hoses
Figure 2. NOx Filter Placement Example 2 – Gas Sample Hose with Integrated T-Stack Hose
2 0024-9469 Rev 2