Bacharach MVR-300 Maintenance Notification

MVR-300 VRF REFRIGERANT LEAK DETECTOR
MAINTENANCE NOTIFICATION
The MVR-300 VRF Refrigerant Leak Detector provides continuous monitoring of refrigerant leaks from variable refrigerant
ow (VRF) systems in occupied spaces. Bacharach recommends that equipment owners test or recalibrate the sensors in
deployed devices every 6 months to ensure optimal performance. The sensors have a 2-year operational life and should be replaced every 2 years.
There are two methods for maintainig calibration in your facility’s leak detectors:
1. Perform a sensor calibration using a calibration adapter, regulator and calibration gas.
2. Replace the sensor with a pre-calibrated, plug-and-play sensor from the manufacturer.
OPTION 1: CALIBR ATION / BUMP TEST
P/N DESCRIPTION
1100-2018 Calibration Adapter Kit (Includes extraction tool, magnetic
wand & tubing)
0051-2358 Pressure Regulator (0.5 L/Min)
0051-4208* Calibration Gas - R-410A (2,000 ppm)
0051-4209* Calibration Gas - R-407c (2,000 ppm)
0051-4210* Calibration Gas - R-404A (2,000 ppm)
0051-4211* Calibration Gas - R-404A (5,000 ppm)
0051-4206* Calibration Gas - R-32 (2,000 ppm)
*Use the calibration gas concentration listed on the label beneath the detector’s cover plate.
Please see the MVR-300 User Manual (P/N 6203-9000) for more information about performing calibration or a bump test.
OPTION 2: SENSOR REPLACEMENT
Pre-calibrated sensor modules are available for quick and easy, a plug-and-play sensor replacement.
Target Gas 0 - 2,500 PPM 0 - 5,000 PPM 0 - 10,000 PPM
R-410A 6203-0101 6203-0102 6203-0103
R-407c 6203-0111 6203-0112 6203-0113
R-404A 6203-0121 6203-0122 6203-0123
R-32 6203-0141 6203-0142 6203-0143
Please see the MVR-300 User Manual (P/N 6203-9000) for more information about sensor replacement.
Bacharach recommends the use of a maintenance log to track the maintenance service performed on the MVR-300 refrigerant monitors.
For more information or to discuss which option may be best to maintain your device, please email help@mybacharach.com or speak with a representative by calling 800-736-4666 × 5007 (US), +353 1-284-3888 × 101 (EU) or +1 724-334-5000 (Global).