These instructions must be provided to users before use of the product and retained for ready reference by the user. Read
this manual carefully before using or maintaining the device. The device will perform as designed only if it is used and
maintained in accordance with the manufacturer's instructions. Otherwise, it could fail to perform as designed, and persons
who rely on this device could sustain serious injury or death.
The warranties made by MSA with respect to the product are voided if the product is not installed and used in accordance
with the instructions in this manual. Please protect yourself and your employees by following the instructions.
Please read and observe the WARNINGS and CAUTIONS inside. For additional information relative to use or repair, call
1-800-MSA-2222 during regular working hours.
For countries of Russian Federation, Republic of Kazakhstan and Republic of Belarus, the gas detector will be delivered
with a passport document that includes valid approval information. On the CD with manual instruction attached to the gas
detector the user will find the documents "Type Description" and "Test Method" - appendixes to Pattern Approval Certificate
of Measuring instrument, valid in the countries of use.
MSA is a registered trademark of MSA Technology, LLC in the US, Europe and other Countries. For all other trademarks
visit https://us.msasafety.com/Trademarks.
1000 Cranberry Woods Drive
Cranberry Township, PA 16066
USA
Phone: 1-800-MSA-2222
Fax: 1-800-967-0398
For your local MSA contacts, please go to our website www.MSAsafety.com
MSA Bacharach warrants this instrument, excluding sensors, to be free from defects in materials and workmanship for a period of two
years from the date of purchase by the original owner. The sensors have a pro-rated warranty period of 6 to 18 months, depending on
the sensor type. If the product should become defective within this warranty period, we will repair or replace it at our discretion.
The warranty status may be affected if the instrument has not been used and maintained per the instructions in this manual or has been
abused, damaged, or modified in any way. This instrument is only to be used for purposes stated herein. The manufacturer is not liable
for auxiliary interfaced equipment or consequential damage.
Due to ongoing research, development, and product testing, the manufacturer reserves the right to change specifications without notice.
The information contained herein is based on data considered accurate. However, no warranty is expressed or implied regarding the
accuracy of this data.
All goods must be shipped to the manufacturer by prepaid freight. All returned goods must be pre-authorized by obtaining a return
merchandise authorization (RMA) number. Contact the manufacturer for a number and procedures required for product transport.
SERVICE POLICY
MSA Bacharach maintains an instrument service facility at the factory. Some MSA Bacharach distributors/agents may also have repair
facilities; however, MSA Bacharach assumes no liability for service performed by anyone other than MSA Bacharach personnel.
Repairs are warranted for 90 days after date of shipment (sensors, pumps, filters and batteries have individual warranties). Should your
instrument require non-warranty repair, you may contact the distributor from whom it was purchased or you may contact MSA
Bacharach directly.
If MSA Bacharach is to do the repair work, send the instrument, prepaid, to the closest Service Center.
Service LocationService Contact InformationService Shipping Address
MSABacharach
621 Hunt Valley Circle
New Kensington, PA 15068, USA
ATTN: Service Department
MSAEdmonton Repair Service
12130 – 154th Street
Edmonton, Alberta
T5V 1J2
Always include your RMA #, address, telephone number, contact name, shipping/billing information and a description of the defect as
you perceive it. You will be contacted with a cost estimate for expected repairs prior to the performance of any service work. For liability
reasons, MSA Bacharach has a policy of performing all needed repairs to restore the instrument to full operating condition.
Prior to shipping equipment to MSA Bacharach, contact our office for an RMA # (returned merchandise authorization). All returned
goods must be accompanied with an RMA number.
Pack the equipment well (in its original packing if possible), as MSA Bacharach cannot be held responsible for any damage incurred
during shipping to our facility.
NOTICES
This manual is subject to copyright protection; all rights are reserved under international and domestic copyright laws. This manual may
not be copied or translated, in whole or in part, in any manner or format, without the written permission of MSA Bacharach, Inc.
All software utilized and/or distributed by MSA Bacharach is subject to copyright protection. All rights are reserved. No party may use or
copy such software in any manner or format, except to the extent that MSA Bacharach grants them a license to do so. If this software is
being loaded onto more than one computer, extra software licenses must be purchased.
USMGS-5503
Page 4
Contents
1Safety5
1.1General Safety Statements5
1.2Safe Connection of Electrical Devices5
2Description5
2.1Product Overview5
2.2Key Product Features6
2.3General Overview7
2.4Sensor Styles8
3Installation8
3.1General Information for Installation8
3.2Installation Restrictions9
3.3Mechanical Installation9
3.4Electrical Installation9
4Operation15
4.1Overview of Normal Operation15
4.2Menus19
4.3Functions22
4.4Parameters25
5Maintenance31
5.1Maintenance Intervals31
5.2Making Adjustments to Sensors31
5.3Troubleshooting34
5.4Sensor Maintenance39
5.5Replacing the Instrument Electronics44
5.6Replacing the Interface Board45
5.7Cleaning the Instrument45
6Factory Default Settings46
7Sensor Principle47
7.1Electrochemical Sensors47
7.2Catalytic Bead Sensors47
7.3Semiconductor Sensors48
7.4Infrared Sensors48
8Disposing of the Instrument48
8.1Disposing of Electrical and Electronic Equipment48
8.2Disposing of Electrochemical Sensors49
9Technical Data49
9.1General Specifications49
9.2Sensor Specifications50
9.3Modbus Registers51
10Ordering Information57
10.1MGS-550 Instrument Only57
10.2MGS-550 Gas Detector, IP66 with IP66 Sensor57
10.3MGS-550 5m Remote and Second Sensing Heads59
10.4MGS-550 Replacement Parts and Accessories60
4MGS-550US
Page 5
1Safety
1.1General Safety Statements
WARNING!
l
Before using this product, carefully read and strictly follow the instructions in the manual.
l
Use the product only for the purposes specified in this document and under the conditions listed.
l
Ensure that product documentation is retained, made available, and appropriately used by anyone operating the
product.
l
Comply with all local and national laws, rules, and regulations associated with this product.
l
Only trained and competent personnel may use this product.
l
Only trained and competent personnel may inspect, repair and maintain the product as detailed in this manual.
Maintenance that is not detailed in this manual must be completed by MSA Bacharach or personnel qualified by
MSA Bacharach.
l
Use only genuine MSA Bacharach spare parts and accessories. Otherwise, operation may be impaired.
l
Only operate the product within the framework of a risk-based alarm signaling concept.
Failure to follow these warnings can result in serious personal injury or death.
1 Safety
1.2Safe Connection of Electrical Devices
WARNING!
Before connecting this instrument to electrical devices not mentioned in this manual, consult the manufacturer or a
qualified professional.
Failure to follow this warning can result in serious personal injury or death.
2Description
2.1Product Overview
The MSA Bacharach MGS-550 continuously monitors indoor or outdoor ambient air for the following gases:
l
toxic and combustible gases
l
oxygen
l
refrigerants.
The instrument is housed in a rugged ABS enclosure (general purpose or “GP” housing).
The instrument can be connected to a MSA Bacharach MGS-408 controller or a Programmable Logic Controller (PLC).
With the integrated alarm relays, the instrument can be operated as a stand-alone unit (with additional local alarm
signaling). The instrument is designed to be installed in non-classified, non-hazardous, permanent locations.
USMGS-5505
Page 6
2 Description
WARNING!
l
Danger of explosions. This product is neither certified nor approved to be operated in oxygen-enriched
atmospheres.
l
The device is NOT intended to be used in areas classified as hazardous.
Failure to follow these warnings can result in serious personal injury or death.
Power options (refer to section 9.1 General Specifications):
o
24 VAC
o
19.5 to 28.5 VDC
l
Multi-function, 5-digit LED display
o
gas concentrations
o
status messages
o
menu choices
l
Diagnostic/status LEDs (3)
l
Digital output Modbus RTU signal
l
Independently configurable analog outputs (2) (based on measured gas concentration)
o
Analog output 4 to 20 mA
o
Analog output 0 to 5 V
o
Analog output 0 to 10 V
o
Analog output 1 to 5 V
o
Analog output 2 to 10 V
l
Redundant sensor mapping option (one sensor can be mapped to both analog outputs)
l
Menu navigation options:
o
Tactile switches on the front cover
o
Non-intrusive magnetic wand on the front cover
l
Non-intrusive magnetic wand can be used to configure, calibrate, and maintain the device
6MGS-550US
Page 7
2.3General Overview
2 Description
USMGS-5507
Page 8
3 Installation
2.4Sensor Styles
Local Sensor5 m (16 ft) Remote Sensor
3Installation
3.1General Information for Installation
Every detail of installation site selection is critical to ensure overall system performance and effectiveness. Strict
compliance and considerable thought must be given to every detail of the installation process, including, but not limited
to the following:
l
Regulations as well as local, state, and national codes that govern the installation of gas monitoring equipment
l
Electrical codes that govern the routing and connection of electrical power and signal cables to gas monitoring
equipment
l
The full range of environmental conditions to which the instruments will be exposed
l
The physical characteristics of the gas or vapor to be detected
l
The specifics of the application (e.g., possible leaks, air movement/draft, etc.)
l
The degree of accessibility required for maintenance purposes
l
The types of optional equipment and accessories that will be used with the system
l
Any limiting factors or regulations that would affect system performance or installations
l
Wiring details, including the following:
o
The general purpose enclosure provides six M16 entry points, which can be used for field wiring, direct
attachment of a sensor, or wiring of a remote sensor.
o
Unused openings must be closed with a suitable plug and gasket, maintaining the IP rating.
o
Secondary circuit must be supplied from an isolating source (not applicable for relay circuits).
o
The wiring for the relays must be selected and fused according to the rated voltages, currents, and
environmental conditions.
o
If stranded conductors are used, a ferrule should be used.
o
To improve RFI immunity in extreme environments, it might be necessary to ground the shield of the
communications cable at the PLC, front-end controller, or Building Management System (e.g., the chassis,
the ground bus-bar, etc.).
8MGS-550US
Page 9
3 Installation
3.2Installation Restrictions
The installation location must have appropriate supply power available for the instrument (i.e., 19.5 to 28.5 VDC or 24
VAC). Refer to Section 9 Technical Data. This ultimately determines the distance the instrument can be mounted from
the controller or power supply.
WARNING!
The MGS-550 must be powered by either:
l
a suitable UL/CSA/IEC 60950 certified power supply that is isolated from line voltage by double insulation, or
l
an appropriately rated UL listed/CSA/IEC Class 2 transformer.
Failure to follow this warning can result in serious personal injury or death.
l
The instrument accepts wire sizes of 16 AWG (1.5 mm2) to 20 AWG (0.5 mm2).
l
Depending on the configuration, use at least a shielded, multi-conductor cable.
l
The instrument must not be exposed to radiant heat that will cause the temperature to rise beyond the limits
stated in Section 9 Technical Data. The use of a reflecting shield is recommended.
l
The enclosure is weatherproof within environmental specifications and suitable for outdoor installation.
l
Each instrument must be installed and operated in an environment that conforms to the specifications listed in
Section 9 Technical Data.
3.3Mechanical Installation
l
Select a mounting location that is accessible for maintenance and adjustment.
l
Ensure that any targeted gas or vapor has unobstructed access to the sensor.
l
Consider implications of the future use of accessories and maintenance equipment.
l
Ensure that the mounting surface is flat and plumb.
l
Ensure that any installed sensor is pointing downwards.
l
MSA Bacharach recommends using M5 bolts (or smaller) with hex socket caps to mount the device.
3.4Electrical Installation
3.4.1Preparations
WARNING!
• Ensure wiring for relays and connections for sensor(s) are made before applying power.
• This product uses semiconductors which can be damaged by electrostatic discharge (ESD). When handling the
printed circuit boards (PCBs), observe proper ESD precautions so that the electronics are not damaged.
Failure to follow these warnings can result in serious personal injury or death.
NOTE: For unused 4 to 20 mA analog outputs, ensure that a jumper (pin 3 to 4 and pin 5 to 6) is installed. Otherwise, a
fault may be displayed if the wiring does not match the configuration. The analog outputs are designed as sources.
USMGS-5509
Page 10
3 Installation
• Open the enclosure lid. Loosen six (6) screws using
an M5 hex key and remove the lid from the base.
• Disconnect the ribbon cable from the processor
board on the lid. Set the lid aside and continue to
wiring procedure.
3.4.2Power and Signal Wiring
l
Using appropriate cable glands and/or conduit, connect the wires for power and signal to the appropriate terminal
as indicated in the figure and wiring table that follow.
l
Polarity must not be reversed.
l
For 24 VAC installations in a daisy-chain configuration, the neutral polarity must be maintained for all instruments.
l
Fasten terminal screws.
10MGS-550US
Page 11
3 Installation
ConnectionDescriptionPinLabelWiring Termination
24 VAC
Power
24 VDC
Analog Output 1*
Analog Output
Analog Output 2*
Digital Output
Modbus Network
Communications
* For 3-wire, single-sensor DC installations: connect pins 1 and 2 of 24 VDC, and connect pin 4 to the analog input of
the control system. For 4-wire, dual-sensor DC installations: connect pins 1 and 2 of 24 VDC, connect pin 4 to one
analog input of the control system, and connect pin 5 to another input of the control system.
If central monitoring or a PLC is used, connect the signal cable shielding at the controller only.
l
If an analog output is unused and configured as a 4 to 20 mA output, then the corresponding output connectors
must be shorted or jumpered, otherwise a fault will occur. For unused analog output 1 configured as 4 to 20 mA
output, wire pin 3 to pin 4. For unused analog output 2 configured as a 4 to 20 mA output, wire pin 5 to pin 6.
These jumper wires are installed at the factory, but should be removed for voltage outputs or if connections are
made to the analog outputs.
3.4.3Relay Wiring
WARNING!
At voltages > 30 VAC or > 42.2 VDC, the relay cables must be enclosed in protective conduit, or double-insulated
cables must be used.
Failure to follow this warning can result in serious personal injury or death.
USMGS-55011
Page 12
3 Installation
l
Using appropriate cable glands and/or conduit, connect the wires for relay 1, relay 2, and relay 3 to the terminals
(see previous wiring figure) as indicated in the following wiring table. (Note that any one of 6 alarms or 3 fault
types may be mapped to any relay.)
FunctionPinLabelWiring Termination
10RELAY 1 NCRelay 1 NC contact
Relay 1 Output
Relay 2 Output
Relay 3 Output
To change relay designations, see section 4.4.2 Relay Designation (RX-xx). For default values, see section6 Factory
Default Settings.
When configured according to the factory default settings, the relays are de-energized during normal operation (not
failsafe). Fail-safe mode can be configured and relay operation is opposite of the wiring table. See section 4.4.2 Relay
Designation (RX-xx). The terminal designators in the wiring table show factory defaults in normal operation (not
failsafe) mode with the relays de-energized.
11RELAY 1 CRelay 1 common contact
12RELAY 1 NORelay 1 NO contact
13RELAY 2 NCRelay 2 NC contact
14RELAY 2 CRelay 2 common contact
15RELAY 2 NORelay 2 NO contact
16RELAY 3 NCRelay 3 NC contact
17RELAY 3 CRelay 3 common contact
18RELAY 3 NORelay 3 NO contact
NOTE: To ensure that a fault is easily “recognized” (that is, without needing to look directly at the instrument’s display),
one relay should be designated for instrument faults and an alarm device should be connected to the fault relay.
3.4.4Install Remote Sensing Head
l
Remove a blind-plug from the enclosure (if applicable).
l
Feed the connector of the sensor (see below, left) through a cable gland or conduit (if applicable), through the
opening of the enclosure, and then into the enclosure.
l
Secure the cable gland (with attached gasket) or appropriate conduit into the housing in order to maintain IP
rating.
l
Plug the sensor connector into the socket until the lock engages. For a new instrument with no sensor connected,
it does not matter which socket is used first. However, if a sensor is already connected, then it should be left
plugged into its socket.
l
If two remote sensors are to be installed, register only one sensor at a time. Refer to section 4.3.6 Register
Sensor (F-06).
12MGS-550US
Page 13
3 Installation
NOTE: Sensors are not automatically recognized and must but individually registered using function F-06. See section
4.3.6 Register Sensor (F-06), section4.3.7 De-register One Sensor (F-07), and section 4.3.8 De-register All Sensors
and Reset Node Address (F-08).
3.4.5Connecting One or More MGS-550s to an MSA Bacharach Controller
NOTE: For wiring and configuration information, please refer to the manual which was included with the MSA
Bacharach controller (e.g., MGS, etc.).
l
At the central control system, connect the shield of the wires to the earth ground of the controller (e.g., the
chassis, the ground bus-bar, etc.).
l
For 24 VDC installations, the input is protected. If the polarity is reversed, the instrument will not power-up.
l
For 24 VAC installations in daisy-chain, the neutral polarity must be maintained for all instruments (see example
below).
USMGS-55013
Page 14
3 Installation
3.4.6Modbus RTU Interface
l
For the Modbus network use an 18 to 24 AWG (0.5 to 1 mm2) shielded twisted pair wire with a 120Ω
characteristic impedance.
l
The Modbus address, baud rate, stop bit, parity and slave termination is configured through the setup menu. No
jumpers or hardware switch settings are required.
l
Ensure that the communication parameters within the network, including the Building Management System, are
configured identically. See section 4.4.5 Modbus Configuration (MB-xx).
l
If the MGS-550 is at the end of the Modbus network, the terminating resistor must be set to “IN”. All other
instrument terminating resistor must be set to “OUT” (factory default). See section 4.4.5 Modbus Configuration
(MB-xx).
3.4.7Conclusion
After all wiring is completed, connect the ribbon cable, replace the lid, and tighten the six (6) screws with an M5 hex key.
NOTICE
DO NOT allow the lid / sensor to hang from the ribbon cable. Failure to follow this notice may result in damage to the
product.
14MGS-550US
Page 15
4 Operation
4Operation
4.1Overview of Normal Operation
WARNING!
Before leaving the instrument for normal operation, check the configuration for proper settings and check calibration.
Failure to follow this warning can result in serious personal injury or death.
4.1.1Applying Power and the Start-up Sequence
After applying power, the instrument will go through a start-up sequence (LED test, software version, and initialization)
and start the warm-up period. The power LED will blink and, by default, the display will remain blank. To enable the
display of gas type and concentration, use parameter P1-01 which will cause the display to toggle between the target
gas name and a value of “0” for the sensor in warm-up. See section 4.4.6 Display Mode (P1-01).
The instrument output will be OFFLINE (see section 4.3.1 Offline Mode (F-01)). If two sensors are installed, both target
gas names and “0”s will be toggled sequentially.
NOTE: Sensor warm-up times may be different for dual sensor configurations. The power LED will continue to blink as
long as at least one sensor is in warm-up.
USMGS-55015
Page 16
4 Operation
After the warm-up period, the instrument begins normal operation. The display toggles between the current gas
concentration and target gas name.
During normal operation, the left green LED is lit solid. Depending on the unit of measurement, the green LED for ppm
or %LEL is lit. If both of these LED are off, the unit of measurement is Vol%.
All configurations must be checked at least initially, and the calibration must be checked initially and as needed.
4.1.2Verifying Analog Signals
During normal operation the current output of the instrument is proportional to the detected gas concentration and can
be selected from the following.
l
4 to 20 mA
l
0 to 5 V
l
1 to 5 V
l
0 to 10V
l
2 to 10 V
The MGS-550 uses different current values to indicate various modes of operation. See section 9.1 General
Specifications for additional information.
4.1.3Verifying the Digital Modbus Signal
The MGS-550 provides a Modbus RTU digital interface. All status messages and most parameters which can be
accessed and/or configured through the menu can also be accessed and/or configured via a Building Management
using a Modbus network. See section 4.4.5 Modbus Configuration (MB-xx).
4.1.4The 5-Digit Display and LEDs
In normal operation, the 5-digit, 7-segment LED display toggles between the gas name (CO2 in this example, and the
measured gas concentration (291 ppm in this example).
In addition, the following special symbols and unique messages may also be displayed during operation of the
instrument.
16MGS-550US
Page 17
Symbol/MessageDescription
The measuring range of the sensor has been exceeded (“upper hockey sticks”
symbol).
The sensor drifted negative (<0) (“lower hockey sticks” symbol).
If a fault has been detected, the display toggles between gas name and "Exxx",
and the green power LED is off. "E100" indicates a critical fault and "E300"
indicates a non-critical fault. A non-critical fault does not need immediate
attention, but should be mitigated at the next scheduled maintenance (see
section 5 Maintenance). If a relay has been designated to the critical fault, it is
asserted. E400 is a configuration warning that happens after a second sensor is
added to warn the user that the sensor is not driving any relays or analog
outputs.
When the first alarm has been triggered, the display will toggle between gas
name, "A1" and current gas concentration. If a relay has been designated to the
first alarm, it is asserted.
4 Operation
When the second alarm has been triggered the display will toggle between gas
name, "A2" and current gas concentration. If a relay has been designated to the
second alarm, it is asserted.
When the third alarm has been triggered the display will toggle between gas
name, "A3" and current gas concentration. If a relay has been designated to the
third alarm, it is asserted.
The instrument is offline. See section 4.3.1 Offline Mode (F-01).
The first character represents an “M”. This is found in the Modbus parameters
(MB-xx). Refer to section 4.4.5 Modbus Configuration (MB-xx).
This is a confirmation that is displayed before a sensor is registered. Tap
[ENTER] to confirm registration of sensor 1 or registration of sensor 2. For more
information, refer to section 4.3.6 Register Sensor (F-06).
This message is shown after successful completion of a functional operation of
one of the sensors (for example, after registering a sensor, after calibrating a
sensor, etc.).
This is a confirmation display prior to de-registering all sensors and resetting
node addresses (F-08). Tap [ENTER] to confirm reset. See section 4.3.8 De-
register All Sensors and Reset Node Address (F-08) for additional information.
USMGS-55017
Page 18
4 Operation
Symbol/MessageDescription
This is a second confirmation display prior to de-registering all sensors and
resetting node addresses (F-08). Tap [ENTER] to confirm reset. section 4.3.8
De-register All Sensors and Reset Node Address (F-08) for additional
information.
The unit of measurement is indicated by two green LEDs on the right side (upper LED for ppm, lower LED for %LEL,
both LEDs off for Vol%). The green LED on the left indicates that the instrument is in normal operation.
NOTE: If two sensors are installed for the same gas, but have different measuring ranges, the unique sensor identifier
or UID (see label on the sensor) will be displayed in addition to the gas name.
NOTE: Depending on the sensor and measuring range, the gas concentration may or may not be shown with a decimal
point.
18MGS-550US
Page 19
4.2Menus
4.2.1General Navigation
The MGS-550 offers two methods of navigating through the menu.
4 Operation
• Use the non-intrusive magnetic wand (tapping above
the “magnetic” icons with dashes)
Magnetic Switch
Points
Internal Push
Button Points
Tapping/pressing and holding the [i] key for more than 3 seconds gives
access to the user menu (which displays the firmware version first). It is
also used to return to the next higher menu item without saving any
changes (“escape”). A short tap/press (< 3 seconds) of the [i] key also
brings up a sensor identification on the display, toggling between the serial
number (UID) of Sensor 1 and Sensor 2.
Tapping/pressing the magnetic wand over the [UP] / [DOWN] keys scrolls
through the menu selections. Holding the magnetic wand at the key is
interpreted as repeat tapping with an eventual acceleration. When the last
menu item is reached, the menu will roll over to the first menu item in the
list. A diagnostic scan can be entered from normal operation by tapping
and holding [UP] for more than 3 seconds.
Tapping/pressing the [ENTER] key (↵ ) confirms a selection. In normal
operation tapping/pressing and holding the [ENTER] key for more than 3
seconds will cancel the OFFLINE mode, or release any latched alarm, or
acknowledge any acknowledgeable alarm, with priority given to OFFLINE
mode.
• Use the internal push buttons (pressing the
push button points).
Function(s)
NOTE: The instrument is designed for the magnetic wand to be used with the enclosure lid in place. If the enclosure lid
is not in place, the magnetic wand may activate two or more keys at once due to cross-talk.
NOTE: After 3 minutes of inactivity within a menu, the instrument times-out and returns to normal operation. When
changing a parameter, 3 minutes of inactivity will cause the instrument to time out and return to normal operation
without accepting any changes.
4.2.2Checking Status and Changing Parameter Values
Select the menu item to be accessed by tapping [UP] / [DOWN].
USMGS-55019
Page 20
4 Operation
l
When the desired menu item is displayed, tap [ENTER]. The current value or status will flash to indicate the user
interface has switched to data entry mode.
l
[UP] / [DOWN] adjust the value of a numerical parameter or scroll through preset choices.
l
Once the display shows the intended value or choice, tap [ENTER] to validate the new parameter and return to
the previous menu.
4.2.3Exiting the Menu
To get back into measurement mode, tap and hold the [i] key for more than 3 seconds. This will step back through the
menus and eventually return to normal operation.
4.2.4Menu Overview
The menu is divided in functions and parameters. Individual parameters are grouped into logical menus.
FunctionDescription
F-0100 = instrument online, 01 = instrument OFFLINE
F-021 = zero sensor 1, 2 = zero sensor 2
F-031 = span sensor 1, 2 = span sensor 2
F-04Instrument test
Access parameters:
ParametersDescription
S1-XXSensor 1 settings
S2-XXSensor 2 settings – if connected
F-05
F-06Register sensor
F-07De-register one sensor
F-08De-register all sensors and reset node address
F-09Access diagnostics, system information, and fault parameters
F-10Reset system to factory default setting
An example of the layered structure of functions, menus, and parameters is illustrated below. Individual functions and
parameters are described in detail in the sections that follow.
In OFFLINE mode the instrument does not respond to alarm conditions, but allows the execution of functions and the
setting of parameters. OFFLINE mode is useful in eliminating false alarms while performing maintenance.
l
To enter OFFLINE mode, change F-01 to 01.
l
To cancel OFFLINE mode, change F-01 to 00 or tap and hold [ENTER] for more than 3 seconds from the top level
menu.
NOTE: OFFLINE mode automatically times out 30 minutes after returning to normal operation (i.e., after exiting the
function menu).
ItemBehavior in OFFLINE Mode
Green LEDOn
Display
Analog Output
Displays “oFFLn” or function and parameter numbers or user interaction in place of the gas
level when applicable
4 to 20 mAChanges to 3 mA
0 to 5 VStays at last valid value
0 to 10 VStays at last valid value
1 to 5 V0 V
2 to 10 V0 V
Modbus Registers
Offline flag (Modbus register 10024) is active (must be monitored).
Concentration registers stay at their last valid values.
Alarm StatesAny pre-existing alarm condition is cancelled
Faults
Faults remain active, but actions based on any pre-existing fault are de-asserted (e.g., relays,
buzzer, and Modbus flags).
4.3.2Zero Adjustment (F-02)
This function is used to adjust the zero reference point of the sensor. See section 5.2.3 Zero Adjustment for
information.
4.3.3Span Adjustment (F-03)
This function is used to adjust the sensitivity to match the known concentration of an applied calibration gas. See
section 5.2.4 Span Adjustment for information.
4.3.4Instrument Test (F-04)
This function tests the relays, display, and analog outputs by temporarily overriding them. It might be necessary to
inhibit the alarms at the central controller to avoid false alarms.
l
Use the [UP] / [DOWN] key to select the test to be performed:
o
1 = relay test
o
2 = analog output test
o
3 = display test.
After exiting this function, the instrument returns to the function menu and all test overrides are removed.
22MGS-550US
Page 23
4 Operation
#TestDescription
The relay test changes the state of each relay. Use the [UP] / [DOWN] key to toggle the relay. Tap
1Relay
Analog
2
Output
3Display The display test lights up all segments and LEDs.
4.3.5Parameter Menu (F-05)
Use this function to access the parameter menu. See section 4.2.4 Menu Overview and section 4.4 Parameters for
detailed information.
[ENTER] to test the next relay. Tap [i] to exit the function.
Note that changing the state of the relays can trigger alarms in connected equipment.
The analog output test allows you to set the output for interface test purposes (e.g., to check the
programming of a central controller).
Tapping [ENTER] will set the analog output to the level equivalent to zero gas. For a 4 to 20 mA
configuration this output will be 4 mA; for a 1 to 5 V configuration this output will be 1 V; and so on. Use
the [UP] / [DOWN] key to change the value of the zero output.
After tapping [ENTER], the analog output will be set to full scale. Use the [UP] / [DOWN] key to adjust
the value of the full-scale output.
NOTE: Changing the analog outputs can trigger alarms in connected equipment.
4.3.6Register Sensor (F-06)
Function F-06 is used to register a new sensor without having to cycle power.
In general, sensor registration is the association of parameters in the instrument to the set in the sensor. Registration
occurs at the factory for the local sensor and is based on the sensor configuration that is ordered. If the sensor
configuration must be changed (e.g., changing a sensor type or adding a second sensor), then the new sensor must be
registered.
NOTICE
Before unplugging ANY sensor, it must be de-registered using function F-07.
Failure to do this will require you to de-register all sensors using F-08 (with the sensors still connected to the main
electronics), remove the sensors, and then re-install and re-register the sensors one at a time using function F-06.
If the sensors are not connected when an F-08 is performed, then their node addresses will not be reset. Refer to the
following NOTICE for information on resetting node addresses.
New sensors all have address 100. The first sensor registered will be sensor #1 and the second sensor #2.
Resetting sensor registration through F-08 will reset them all to 100; then they need to be registered one at a time.
To register a newly added sensor, select F-06 and tap [ENTER]. There are 3 possible scenarios.
ScenarioDescription
Sensor is
already
registered
Normal
registration/
adding a
sensor
Address
conflict
If there is already one sensor connected and it has already been registered, the F-06 display is shown
after a brief delay.
If the newly added sensor was previously unregistered or de-registered (i.e., address 100), “reg 1” or
“reg 2” will be displayed (depending on the configuration). Tap [ENTER] to confirm registration of the
sensor. “PASS1” or “PASS2” will be displayed.
If a sensor was previously registered in an instrument, was not de-registered, was removed, and then
USMGS-55023
Page 24
4 Operation
ScenarioDescription
plugged in to a different instrument, then a critical fault (E100 error) will occur.
Refer to function F-08 to resolve.
4.3.7De-register One Sensor (F-07)
This function resets the registration of a sensor and changes a 2-sensor system to a 1-sensor system.
NOTE: Removing a sensor without first de-registering it will result in a fault. Always de-register a sensor before
permanently removing it.
Tap [ENTER] to display the unique sensor identifier (UID). Use the [UP] / [DOWN] keys to select the sensor to be
removed. After tapping [ENTER], the sensor can be disconnected. See section 5.4 Sensor Maintenance for
information.
NOTE: A sensor’s user-editable parameters (e.g., alarm setpoints, etc.) are not retained after it is de-registered. If a
sensor is registered again, all parameters will be set to their factory default values.
4.3.8De-register All Sensors and Reset Node Address (F-08)
Two sensors connected to the instrument that have the same node address will result in a fault. Function F-08 can be
used to recover from this situation by resetting the registration of all sensors and resetting their node addresses to 100.
l
Tap [ENTER]. The instrument will show “Reset”.
l
Tap [ENTER] to confirm reset. Instrument will show a second confirmation (“sure”).
l
Tap [ENTER] again to reset the node address. “Pass” will be displayed.
l
Unplug the sensors.
l
Plug in one sensor and register it. See section 4.3.6 Register Sensor (F-06).
l
Plug in second sensor and register it.
Be sure to plug in the sensors and re-register them one at a time to avoid address conflicts.
NOTICE
Before unplugging ANY sensor, it must be de-registered using function F-07.
Failure to do this will require you to de-register all sensors using F-08 (with the sensors still connected to the main
electronics), remove the sensors, and then re-install and re-register the sensors one at a time using function F-06.
If the sensors are not connected when an F-08 is performed, then their node addresses will not be reset.
4.3.9Diagnostics, System Information and Fault Data (F-09)
This function will display any diagnostics and fault codes.
l
Tap [ENTER] to initiate the function.
l
Use the [UP] / [DOWN] keys to select a diagnostics code.
l
After tapping [ENTER], the respective code will be displayed.
See section 5.3 Troubleshooting for additional information.
NOTE: The diagnostics attributes can be accessed directly during normal operation by tapping and holding [UP] for
more than 3 seconds.
24MGS-550US
Page 25
4 Operation
4.3.10 Reset System to Factory Default Setting (F-10)
This function will restore factory default values of all application specific parameter settings. See section 6 Factory
Default Settings for a list.
NOTE: Resetting the system to its factory default values will not reset sensor calibrations.
NOTE: Function F-10 resets all parameters (except sensor registration information) to their factory default values (see
section 6 Factory Default Settings). Before executing this function, consider recording all parameter settings in case
you wish to reset one or more of them to their former values.
l
Tap [ENTER] to initiate this function. A "SURE" confirmation prompt is given.
l
Tap [ENTER] to confirm the reset.
l
All LED segments will light up for 3 seconds.
l
Parameter settings will change to factory default values.
l
The internal buzzer will sound for 3 seconds.
l
After another 3 seconds the instrument will return to the main menu.
After an F-10 system reset, the instrument will be in a “known state”, and the parameters can be set.
4.4Parameters
4.4.1Sensor 1 Settings (S1-xx) and Sensor 2 Settings (S2-xx) if Connected
S1 & S2
Name (n=1 or 2)Description (n=1 or 2)
Param
S1-01S2-01 Sensor n Gas NameAbbreviated Gas Name
S1-02S2-02 Sensor n UIDUnique 5-digit sensor ID (serial number)
Value above which a low alarm condition occurs. Low alarm value must
be less than the medium and high alarm values (Sn-03 < Sn-04 < Sn-
S1-03S2-03
S1-04S2-04
S1-05S2-05
S1-06S2-06
S1-07S2-07
S1-08S2-08
Sensor n Low Alarm
ppb/ppm/%LEL/Vol%
Sensor n Medium Alarm
ppb/ppm/%LEL/Vol%
Sensor n High Alarm
ppb/ppm/%LEL/Vol%
Sensor n Low Alarm
Behavior
Sensor n Medium Alarm
Behavior
Sensor n High Alarm
Behavior
S1-09S2-09 Sensor n Type Code4-digit sensor code (read-only)
05). (See oxygen sensor notice below.)
This parameter has a fixed minimum limit that is sensor–specific and
not editable.
Value above which a medium alarm condition occurs. Medium alarm
value must be between low and high alarm values (Sn-03 < Sn-04 <
Sn-05). (See oxygen sensor notice below.)
Value above which a high alarm condition occurs. High alarm value
must be greater than low and medium alarm values (Sn-03 < Sn-04 <
Sn-05). (See oxygen sensor notice below.)
S1-11S2-11 Sensor n Full ScaleRead-Only. Sensor n full-scale units: ppb or ppm or Vol% or %LEL.
When set to 1, this flag indicates that a calibration is recommended.
This flag can be reset to 00 by either successfully completing a span
adjustment (see section 5.2.4 Span Adjustment) or by changing the
S1-12S2-12 Sensor n Test Flag
The following parameters are sensor dependent.
l
Low alarm limit (S1-03 and S2-03)
l
High alarm limit (S1-05 and S2-05)
l
Unit of measurement (S1-10 and S2-10)
l
Factory default values for the alarm thresholds
NOTE: A fixed hysteresis of 5% of full scale is set in order to avoid chatter at an alarm threshold.
NOTE: For all instruments except oxygen, the alarm occurs on gas concentrations that increase beyond the set-points.
value manually.
0 = Sensor OK
1 = Sensor in operation for >6 months without calibration. Recalibration
recommended.
NOTE: For all instruments monitoring oxygen, the low and medium alarms occur on concentrations that decrease
below the set-points. The high alarm occurs on gas concentrations that increase beyond the set-point. This permits the
detection of oxygen displacement and enrichment scenarios.
26MGS-550US
Page 27
4 Operation
NOTE: Alarm hierarchy exists. An A2 alarm overrides an A1 alarm on the display. However, the A1 and A2 states
operate independently. An example follows.
l
A1 is acknowledgeable
l
A2 is not acknowledgeable
l
The gas concentration is such that it triggers A1 and A2
Acknowledging will cause the A1 relay to release. However, the display will still show an A2 alarm as long as the A2
condition continues to exist. The same is true for A3 and A2.
Latching and acknowledgment settings of alarms may be combined to create unique scenarios.
l
Latching means that once the alarm level is reached, the instrument will trigger the alarm and it will remain in
alarm status even if the gas concentration subsequently does not meet the alarm condition any more. To clear a
latching alarm it has to be acknowledged by tapping and holding [ENTER] for more than 3 seconds.
l
Non-latching means that the alarm status clears if the gas concentration does not meet the alarm condition
anymore.
l
Acknowledgeable means that the alarm relay can be reset before the alarm condition clears.
l
Not acknowledgeable means that the alarm relay cannot be reset until the alarm condition clears.
USMGS-55027
Page 28
4 Operation
LatchingAcknowledgeableExplanation
NoNo
NoYes
YesNo
YesYes
Latching and acknowledgment does not just affect relay states, but also the buzzer state and Modbus status flags.
4.4.2Relay Designation (RX-xx)
The alarm state will reset automatically when the alarm condition clears. The
alarm state cannot be reset manually before the alarm condition clears.
The alarm state will reset automatically when the alarm condition clears or
can be reset manually.
The alarm state must be reset manually. The alarm state cannot be reset
before the alarm condition clears.
The alarm state must be reset manually and can be reset before the alarm
condition clears.
ParameterNameDescription
R1-01
R1-02Relay 1 Source
R2-01
R2-02Relay 2 Source
R3-01
R3-02Relay 3 Source
Relay 1 Contact
Behavior
Relay 2 Contact
Behavior
Relay 3 Contact
Behavior
Select behavior for Relay 1: 0 = NO, 1 = Failsafe.
Select alarm configuration (01 to 14) to activate Relay 1:
01: Sensor 1 Low Alarm
02: Sensor 1 Medium Alarm
03: Sensor 1 High Alarm
04: Sensor 1 Fault
05: Sensor 2 Low Alarm
06: Sensor 2 Medium Alarm
07: Sensor 2 High Alarm
08: Sensor 2 Fault
09: Instrument Electronics Critical Fault
10: Any Low Alarm
11: Any Medium Alarm
12: Any High Alarm
13: Any Sensor Fault or Instrument Electronics Critical Fault
14: Any Alarm or Sensor Fault or Instrument Electronics Critical Fault
Select behavior for Relay 2: 0 = NO, 1 = Failsafe.
Select alarm configuration (01 to 14) to activate Relay 2. See R1-02 above for
source codes.
Select behavior for Relay 3: 0 = NO, 1 = Failsafe.
Select alarm configuration (01 to 14) to activate Relay 3. See R1-02 above for
source codes.
The relays can be independently designated to any configuration below.
For example, consider a 2-sensor system with relays configured as follows:
l
R1-02=01Relay 1 configured as 01 (Sensor 1 Low Alarm).
l
R2-02=06Relay 2 configured as 06 (Sensor 2 Medium Alarm).
l
R3-02=13Relay 3 configured as 13 (Any Sensor Fault or Instrument Electronics Critical Fault).
28MGS-550US
Page 29
4 Operation
l
In this configuration, the following would occur:
l
Relay 1 will activate when the low alarm of sensor 1 is triggered.
l
Relay 2 will activate when the medium alarm of sensor 2 is triggered.
l
Relay 3 will activate when any sensor or instrument fault occurs (either sensor).
It is possible to dedicate the same alarm configuration to multiple relays. For example, sensor 1 high alarm could be
mapped to relay 1 and relay 2 (for redundancy).
NOTICE
With integrated relays, the instrument can be operated stand-alone without a central controller (with additional local
alarm signaling). It is recommended to designate one relay as fault relay (e.g., configuration 13).
4.4.3Alarm Configuration (AF-xx)
ParameterNameDescription
00 = no delay, 01 to 15 = delay in minutes.
If an alarm ON delay is set, the gas concentration has to be continuously above (below
for oxygen low and medium alarms) the alarm threshold for the delay time before the
alarm is triggered or activated.
AF-01
Alarm ON
Delay Value
NOTE: Once programmed, all alarms have the same ON delay.
00 = no delay, 01 to 15 = delay in minutes.
If an alarm OFF delay is set, the gas concentration has to be continuously below
AF-02
Alarm OFF
Delay Value
(above for oxygen low and medium alarms) the alarm threshold for the delay time
before the alarm is deactivated. If the alarm is acknowledgeable, the alarm OFF delay
will be terminated as soon as it is acknowledged.
NOTE: Once programmed, all alarms have the same OFF delay.
0 = non-latching, 1= latching
AF-03
Critical Fault
Latch Mode
Latching means that any critical fault state will remain active (even if the event creating
the fault is cleared) until the user manually clears the fault state with the "ENTER" key.
00 = no timeout, 01 to 59 = delay in minutes
NOTE: If an alarm state has been acknowledged, but the condition still exists after the
timeout period, the alarm state will be triggered again.
AF-04
Alarm
Acknowledge
Timeout
4.4.4Analog Output Configuration (AX-xx)
ParameterNameDescription
A1-01Analog Output 1 Source
A1-02Analog Output 1 Type
A1-03Analog Output 1 Scaling
USMGS-55029
1 = Sensor 1
2 = Sensor 2
0 = 0 to 5 V
1 = 1 to 5 V
2 = 0 to 10 V
3 = 2 to 10 V
4 = 4 to 20 mA
Scales the FSD (full-scale deflection), i.e., the measuring range, between
20 and 100% FSD, to the maximum analog output (e.g., 20 mA).
The selected full scale deflection determines the analog output (e.g., 20 mA
Page 30
4 Operation
ParameterNameDescription
signal) on the analog interface.
Example: Consider an application with a required range of 0 to 20,000 ppm
CO2 with a min and max of 4,000 ppm and 20,000 ppm, respectively.
To select full scale deflection as 10,000 ppm, set this parameter to 50%.
The analog output will be linear between 4 mA (= 0 ppm) and 20 mA (=
10,000 ppm). However, the resolution of the sensor stays at the value for
the max range.
NOTE: Alarm thresholds will not be changed when scaling the analog
output. These must be set separately. See section 4.4.1 Sensor 1 Settings
(S1-xx) and Sensor 2 Settings (S2-xx) if Connected.
Sets an analog offset (± 410 = ± 10% of full scale) at the zero point. This
parameter adds an offset to the analog output. The offset is constant over
the entire range of the analog signal. No re-calibration necessary. This
A1-04Analog Output 1 Offset
A2-01Analog Output 2 SourceSame as A1-01, but for output 2 (see description above).
A2-02Analog Output 2 TypeSame as A1-02, but for output 2 (see description above).
A2-03Analog Output 2 ScalingSame as A1-03, but for output 2 (see description above).
A2-04Analog Output 2 OffsetSame as A1-04, but for output 2 (see description above).
parameter is essential for an installation where the analog current at the
instrument differs from the current at the central controller.
NOTE: The offset can be positive or negative, but can’t take the output
below the zero gas level (e.g., 4 mA for 4-20 mA).
4.4.5Modbus Configuration (MB-xx)
ParameterNameDescription
MB-01Modbus Node AddressSet the RS-485 node address (001 to 255)
MB-02RS-485 Baud Rate
MB-03RS-485 Stop Bits1 or 2
MB-04RS-485 Parity
MB-05
4.4.6Display Mode (P1-01)
Modbus Slave
Termination
0 = 9600
1 = 19200
0 = None
1 = Odd
2 = Even
0 = Out
1 = In
ParameterNameDescription
Defines whether or not the gas name and concentration are displayed
during normal operation.
P1-01Display Mode
0 = OFF (gas name and concentration will not be shown)
1 = ON (gas name and concentration will be shown)
If an alarm or fault condition occurs, the display will be turned on to show
the error code(s).
30MGS-550US
Page 31
4.4.7Buzzer Designation (B1-xx)
ParameterNameDescription
5 Maintenance
B1-01
Buzzer Enable and
Source
Select alarm configuration (01 to 14) to activate Buzzer.
00 = Buzzer disabled
Each MGS-550 has a built-in buzzer which can be designated to any alarm configuration. Refer to section 4.4.2 Relay
Designation (RX-xx) for a list of alarm configurations.
5Maintenance
5.1Maintenance Intervals
IntervalFunction
Check calibration.
During Commissioning
Every 6 Months*
As RequiredReplace sensor(s). See section 5.4 Sensor Maintenance.
* Maintenance intervals must be established for each individual installation. Depending on safety considerations, application-specific conditions,
and local regulations, the 6-month maintenance intervals might need to be shortened.
Check signal transmission to the central controller.
Check LEDs for proper operation.
Check for proper triggering of alarm devices.
Inspection by trained service personnel.
Check signal transmission to the central controller.
Check LEDs for proper operation.
Check for proper triggering of alarm devices.
Calibrate the sensor or contact MSA Bacharach for sensor
exchange with factory-calibrated sensor.
5.2Making Adjustments to Sensors
5.2.1Introduction
Adjustment of the instrument must be performed at regular intervals as detailed above.
WARNING!
Breathing Hazard: Calibration gas must not be inhaled! See appropriate Material Safety Data Sheets. Calibration gas
should be vented into a fume hood or to the outside of the building.
Zero First, Then Span: For proper operation, never adjust the span before completing a zero adjustment. Performing
these operations out of order will cause faulty calibration.
MSA MSA Bacharach recommends calibrating instruments within the application-specific condition and with target gas.
This method of zeroing the instrument in the application environment and performing a target gas calibration is more
accurate. A surrogate gas calibration may only be performed as an alternative if a target gas calibration is not possible.
Failure to follow these warnings can result in serious personal injury or death.
NOTE: The sensor should be fully warmed-up (refer to section 4.1.1 Applying Power and the Start-up Sequence).
NOTE: When entering the functions F-02 (zero adjustment) or F-03 (span adjustment), the instrument will automatically
enter OFFLINE mode. The instrument will remain OFFLINE until either the OFFLINE mode is canceled using function
F-01 (see section 4.3.1 Offline Mode (F-01)), or the OFFLINE mode times out within 30 minutes after the adjustment
has ended. This is to avoid false alarms caused by residual cal gas. If the adjustment has not been completed, the
USMGS-55031
Page 32
5 Maintenance
instrument will abort the zero or span adjustment after 5 minutes of the last of interaction. No new data will be saved
and the instrument remains OFFLINE.
5.2.2General Procedure
l
Ensure the instrument has been powered on for at least 1 hour prior to beginning the sensor adjustment
procedure. For semiconductor sensors, you must wait at least 24 hours.
l
Attach the pressure regulator to the calibration gas cylinder.
l
Attach tubing to the pressure regulator.
l
Attach calibration adapter to the tubing.
l
Connect the tubing to the calibration port (the barbed fitting) of the splash guard.
l
The gas flow should be approximately 0.3 to 1.0 L/min.
l
Access the appropriate function (as needed) and begin the zero, span, or bump test.
NOTE: If the intended operation is at higher altitudes, the factory calibration will result in a reading that is lower than the
reading at sea level (a result of reduced partial pressure). A new span adjustment is recommended if the altitude or the
ambient pressure is changed. The factory calibration is set to sea level.
32MGS-550US
Page 33
5.2.3Zero Adjustment
5 Maintenance
WARNING!
• Ambient air can be used to zero the sensor instead of nitrogen or synthetic air only if the area is known to be free
of the target gas or any gas to which the sensor may be cross-sensitive (as listed on the sensor data sheet). In
this case, no cylinder or calibration adapter is needed for the zero adjustment. For semiconductor sensors, you
must use synthetic air only.
• For oxygen (O2) and carbon dioxide (CO2) sensors, use only nitrogen (N2) gas.
Failure to follow these warnings can result in serious personal injury or death.
NOTE: A sensor must be re-zeroed if it shows underflow saturation.
l
Enter the menu and select function F-02 Zero Sensor.
l
Use [UP] / [DOWN] to select the sensor to be calibrated, then tap [ENTER].
l
Apply “synthetic” air (i.e., 20.9% zero air) or nitrogen (see section 5.2.2 General Procedure) and tap [ENTER].
The display will show the current value blinking (e.g., “2”). The gas value may go negative. This is normal.
l
Wait for the current gas concentration value to stabilize.
l
Tap [ENTER] to save the calibration data.
l
The instrument will indicate success status with a “PASS” display. Otherwise a “FAIL” message is displayed.
l
Turn off gas flow and remove the calibration adapter from the sensor or disconnect the tubing.
l
For oxygen, ensure that the concentration is above the displacement alarm thresholds.
l
Tap [ENTER] to exit the function.
5.2.4Span Adjustment
l
Enter the menu and select function F-03 Span Sensor.
l
Use [UP] / [DOWN] to select the sensor to be calibrated, then tap [ENTER].
l
The last span gas concentration will be displayed.
l
Use [UP] / [DOWN] to change the value to match the concentration of the calibration gas (in the units of the
sensor – i.e., ppm, Vol %, %LEL, etc.), and tap [ENTER].
l
Apply span gas (see section 5.2.2 General Procedure) and tap [ENTER].
USMGS-55033
Page 34
5 Maintenance
l
The display will show the current gas concentration value blinking in the units of the sensor – i.e., ppm, Vol %,
%LEL, etc. (e.g., “100”).
l
Wait for the current gas concentration value to stabilize.
l
Tap [ENTER] to perform the calibration adjustment.
l
The instrument will indicate success status with a “PASS” display. Otherwise a “FAIL” message is displayed.
l
Turn off gas flow and remove the calibration adapter from the sensor or disconnect the tubing.
l
Ensure that the concentration is below the alarm thresholds.
l
For oxygen, ensure that the concentration is below the enrichment alarm thresholds.
l
Tap [ENTER] to exit the function.
5.2.5System Bump Test
A system bump test is a live test of a system to verify that the instrument responds to gas and that all connected alarm
devices, controllers, etc. are operating appropriately. In this case, be sure to inform all involved persons about the test
and that certain alarms might have to be inhibited (e.g., process shutdown, notification of authorities, etc.)
l
Apply target gas; if required a sufficiently high concentration (low concentration for oxygen displacement) to
trigger alarms.
l
The display will show the current gas concentration value.
l
Once the alarm thresholds are exceeded, verify that all designated gas alarm relays are activated and the analog
and digital outputs properly transmitted the corresponding gas concentrations.
l
Turn off gas flow and remove the calibration adapter from the sensor or disconnect the tubing.
5.3Troubleshooting
5.3.1Hexadecimal Format
All fault codes are shown in hexadecimal (hex) format. A hex digit can represent multiple codes as shown below.
0001 = Reported temperature outside operating range(Non-critical, E300)
D-01
0002 = Master RX buffer full(Non-critical, E300)
0004 = Slave RX buffer full(Non-critical, E300)
0008 = CRC error in received packet(Non-critical, E300)
0010 = CRC error in received packet(Non-critical, E300)
0020 = Packet timeout(Non-critical, E300)
34MGS-550US
Page 35
AttributeDescription
0040 = Either Current loop open(Non-critical, E300)
0080 = Stuck key or magnetic switch(Non-critical, E300)
0100 = No registered sensors(Critical, E100)
0200 = Sensor 1 reporting a critical fault(Critical, E100)
0400 = Sensor 2 reporting a critical fault(Critical, E100)
5 Maintenance
0800 =
1000 =
Sensor 1 is registered, but not responding correctly or wrong
sensor installed
Sensor 2 is registered, but not responding correctly or wrong
sensor installed
2000 = Power supply voltage(s) out of range(Critical, E100)
4000 = MPU clock fault(Critical, E100)
8000 = Can’t read EEPROM(Critical, E100)
For more information on these faults, refer to table in section 5.3.3 Instrument Electronics Critical (E100)
Faults and section 5.3.4 Non-Critical (E300) Faults. Refer to section 5.3.1 Hexadecimal Format for
combined codes.
D-02
Displays the last fault that occurred since first start-up or since the data logger was last erased. See D-01
for more information. Tap [ENTER] to clear.
D-03Supply voltage
D-04
D-05
Number of days since the last span calibration of Sensor 1. This value automatically resets to 0000 after
completing a span adjustment via F-03.
Number of days since the last span calibration of Sensor 2. This value automatically resets to 0000 after
completing a span adjustment via F-03.
Sensor 1 current fault code xxxx:
0000 = No faults
0001 = No sensor signal
0002 = Sensor board hardware fault
0004 = EEPROM checksum fault
0008 = Zero adjust out of range
0010 = Span adjustment out of range
0020 = Input voltage fault
0040 = Microprocessor fault
D-06
0080 = Software fault
0100 = RS485 receiver fault
0200 = RS485 checksum fault
0400 = Modbus fault
0800 = Temperature sensor fault
1000 = Temperature out of range fault
2000 = Analog front end communications fault
4000 = Negative gas concentration fault
8000 = Sensor configuration fault
(Critical, E100)
(Critical, E100)
For more information on these faults, Refer to table in section 5.3.5 Sensor Faults. Refer to section 5.3.1
Hexadecimal Format for combined codes.
USMGS-55035
Page 36
5 Maintenance
AttributeDescription
D-07Sensor 1 last fault code. See D-07 list. To clear, see Diagnostics D-02.
D-08Sensor 2 current fault code. Same as D-07. See D-07 list.
D-09Sensor 2 last fault code. See D-07 list. To clear, see Diagnostics D-02.
D-10Temperature of main electronics (in °C).
Critical faults are indicated by “E 100” on the display. Critical faults indicate a functional problem that results in the gas
detector no longer monitoring gas reliably. See D-01 and D-02.
CodeCritical FaultPossible CausesRemedy
0200
0400
0800Sensor 1 error
1000Sensor 2 error
2000
4000
8000
Sensor 1 critical
fault
Sensor 2 critical
fault
One or more
power supply
voltages out of
range
Micro-processor
fault
Can’t read
EEPROM
No sensor or defective
sensor.
Sensor removed and not
deregistered.
Two with same node
address.
Same as Sensor 1. See
above.
Sensor is registered, but
not responding correctly or
the wrong sensor
installed.
Input supply power to the
MGS-550 is out of range.
Internal electronics
hardware error.
Add sensor. Replace sensor.
Deregister sensor, then remove. See section 4.3.7 De-
register One Sensor (F-07).
De-register both sensors and re-register them one at a time.
See section 4.3.8 De-register All Sensors and Reset Node
Address (F-08) and section 4.3.6 Register Sensor (F-06).
Same as Sensor 1. See above.
Sensor was removed. Replace sensor. Sensor was replaced
without re-registration. Register sensor. See section 4.3.6
Register Sensor (F-06).
Correct the input supply voltage to restore normal operation.
See section 9.1 General Specifications.
Cycle power to the instrument.
A possible hardware problem may exist. Contact the factory.
5.3.4Non-Critical (E300) Faults
Non-critical faults are indicated by “E 300” on the display. Non-critical faults indicate issues such as environmental
conditions being outside of specification, an installation error has occurred (wrong wiring, for example), and so on.
The gas detector will continue to monitor, but may produce less accurate readings and may not respond correctly to gas
alarm events. A non-critical fault might not need immediate attention, but should be mitigated at the next scheduled
maintenance (at the latest). See D-01 and D-02.
CodeCritical FaultPossible CausesRemedy
Reported
0001
0002
36MGS-550US
temperature
outside operating
range
Master receiver
buffer full
Instrument temperature is
outside specification
RS-485 Message too long for
receiver buffer
Verify that ambient air conditions are within specifications
range for the instrument (see section 9.1 General
Specifications) and for the sensor (see section 9.2 Sensor
Specifications). Correct ambient air conditions.
Cycle power to the instrument.
Cycle power to the PLC or control system (if applicable).
Page 37
CodeCritical FaultPossible CausesRemedy
Verify proper shielding on network communications
wiring. Refer to section 3.1 General Information for
0004
Slave receiver
buffer full
Installation and section 9.1 General Specifications for
more information.
5 Maintenance
Checksum
0008
0010
0020Packet timeout
0040
5.3.5Sensor Faults
NOTE: If a sensor fault occurs during a gas alarm condition, then the fault overrides the alarm condition for that sensor
(i.e., the alarm state(s) for that sensor “de-assert” and the display shows “E 100”). Similarly, if a critical fault occurs, then
all active alarm conditions will “de-assert”.
Sensor faults (for diagnostic attributes D-07 to D-10) may be decoded using the following table. Note that several faults
may be reported at the same time (see section 5.3.1 Hexadecimal Format for more information). For example, fault
code “0003” is a combination of fault codes 0001 (No sensor signal) and 0002 (sensor board hardware fault). See D-07
through D-10.
(CRC) error in
received packet
Checksum
(CRC) error in
received packet
Either current
loop open
Transmission is corrupted
(computed checksum doesn’t
match transmitted checksum)
Modbus message was
truncated or timed out early
Possible wiring, connection,
and/or termination issue
exists. Analog output is set to
4 to 20 mA and loop is open.
Verify proper grounding of the communications wiring.
Refer to section 3.4.5 Connecting One or More MGS-
550s to an MSA Bacharach Controller.
Verify proper communications settings for the network.
Refer to section 4.4.5 Modbus Configuration (MB-xx) for
more information.
A possible hardware problem may exist. Contact the
factory.
Wire pin 3 to 4 or pin 5 to 6 on analog output 4 to 20 mA
terminal (see section 3.4.1 Preparations).
Check the integrity of the 4 to 20 mA connections to the
interface PCB.
NOTE: If a “last fault” attribute (D-08 or D-10) indicates that a fault has occurred at some point in time, but the
corresponding “current fault” attribute (D-07 or D-09) shows no fault, then the problem has self-healed and no service
action is required.
CodeCritical FaultPossible CausesRemedy
0000No Fault
Check all connections in the sensing
0001No sensor signalCannot detect sensor
0002
0004
0008
0010
0020Input voltage faultPower supply voltage out of rangeCall service
0040
Sensor board
hardware fault
EEPROM
checksum fault
Zero adjust out of
range
Span adjustment
out of range
Microprocessor
fault
Cannot read EEPROM or analog front end
Error reading EEPROMReplace sensor module.
Field zero adjustment exceeds limitsPerform zero adjustment
Field span adjustment exceeds limitsPerform span adjustment
MPU clock faultCall service
head.
Check connections to the interface board.
Check all connections in the sensing
head.
USMGS-55037
Page 38
5 Maintenance
CodeCritical FaultPossible CausesRemedy
0080Software faultFirmware error
0100
0200
RS485 receiver
fault
RS485 checksum
fault
0400Modbus fault
0800
1000
Temperature
sensor fault
Temperature out of
range fault
Receiver buffer full (message too long to
receive buffer)
Checksum error in received packet
Packet timeout. Modbus message
truncated or timed out early
Temperature sensor disconnected or no
valid reading
Temperature out of range
Analog front end
2000
communications
EC analog front end failureReplace sensor module
fault
4000
8000
Negative gas
concentration fault
Sensor
configuration fault
Sensor has drifted too negativePerform zero and span calibration
General error in sensor configurationReplace sensor module
Power-cycle. If it re-occurs then call
service
Power-cycle. If it re-occurs then call
service
Power-cycle. If it re-occurs then call
service
Power-cycle. If it re-occurs then call
service
Check connection to sensor PCB
Confirm that sensing head is operating
within specifications
38MGS-550US
Page 39
5 Maintenance
5.4Sensor Maintenance
5.4.1Components Overview
CAUTION!
This product uses semiconductors which can be damaged by electrostatic discharge (ESD). When handling the PCB,
care must be taken so that the electronics are not damaged.
Failure to follow this caution can result in minor or moderate injury.
NOTE: If the instrument has two sensors installed, and both sensors, modules or heads should be replaced, then
replace and register them one at a time. Failing to do so might cause a fault. Observe the sensor type (target gas,
measuring range) and replace the sensor with the same type.
If sensors should be replaced with different types, be sure to de-register the sensor(s) first, and then register them one
at a time. Refer to the following sections:
l
4.3.8 De-register All Sensors and Reset Node Address (F-08)
l
4.3.7 De-register One Sensor (F-07)
l
4.3.6 Register Sensor (F-06)
ItemDescriptionReplacement Instructions
1Sensor cableSee figure above
2Sensor connectorSee figure above
3Sensing head baseSee figure above
4O-ringSee figure above
5Sensor control board5.4.3 Replacing the Sensor Control Board
6Sensor module (sensor plus PCB)5.4.2 Replacing the Sensor Module
7Optional spacerSee figure above (used with SC combustible and SC VOC sensors)
8Foam sensor gasketSee figure above
9Sensing head capSee figure above
10Permeable membraneSee figure above
11Lock nut (M40 X 1.5)See figure above
12Splash guardSee figure above
1-12Entire Assy (Local)5.4.4 Replacing the Local Sensing Head Assembly
1-12Entire Assy (Remote)5.4.6 Replacing the Remote Sensing Head Assembly
USMGS-55039
Page 40
5 Maintenance
5.4.2Replacing the Sensor Module
See illustration in 5.4.1 Components Overview.
1. Power down the instrument.
2. Unscrew M40 locknut (item 10).
3. Pull sensor cap (item 9) down.
4. Remove sensor spacer (optional on certain sensors) if it is in place.
5. Remove foam gasket (item 8) if it is adhered to the sensor module.
6. Remove the sensor module (item 6) from sensor control board (item 5).
7. With flat edges of PCBs aligned, plug new sensor module into sensor control board.
8. Ensure that the spacer (optional item 7 on certain sensors), is in place (if applicable).
9. Ensure that the round foam gasket is in place and not deformed.
10. Slide the sensing head cap back on the sensing head base.
11. Ensure that O-Ring (item 4) is in place.
12. Screw locknut back on until it clicks three times (hand tight) while compressing the
sensor cap against the base.
13. Power-up the instrument.
14. Verify calibration.
5.4.3Replacing the Sensor Control Board
See illustration in section 5.4.1 Components Overview.
1. Power down the instrument.
2. Unscrew M40 locknut (item 10).
3. Pull sensor cap (item 8) down.
4. Remove sensor spacer (optional on certain sensors) , if it is in place (if applicable).
5. Remove foam gasket (item 7) if it is adhered to the sensor module.
6. Pull out sensor and PCB stack (items 5 and 6) from the sensing head base (item 3).
7. Remove sensor control board (item 5) from sensor module (item 6).
8. With flat edges of PCBs aligned, plug new sensor module into sensor control board.
9. Plug PCB stack with sensor back into sensing head base.
10. Ensure that the spacer (optional on certain sensors) is in place (if applicable).
11. Ensure that the round foam gasket is in place and not deformed.
12. Slide the sensing head cap back on the sensing head base.
13. Ensure that O-Ring (item 4) is in place.
14. Screw locknut back on until it clicks three times (hand tight) while compressing the
sensor cap against the base.
15. Power-up the instrument. Replacing the sensor control board requires sensor
registration. See section 4.3.6 Register Sensor (F-06).
16. Verify calibration.
40MGS-550US
Page 41
5.4.4Replacing the Local Sensing Head Assembly
See illustration in section 5.4.1 Components Overview.
1. Deregister the old sensor (see section 4.3.7 De-
register One Sensor (F-07)).
2. Power down the instrument.
3. Loosen the six screws of the lid using an M5 hex
key and remove the lid.
4. Unplug the old sensing head.
5. Unscrew the old sensing head.
6. Screw sensing head with attached gasket into the
housing and tighten (maintain IP rating).
7. Plug new sensing head connector into socket.
8. Place bezel back into the enclosure and secure
with the three screws to the standoffs.
9. Put lid back on and tighten the six screws.
10. Power-up instrument.
11. The instrument will prompt to start the sensor
registration (e.g., “rEg 1”). Acknowledge with
[ENTER]. The instrument will then indicate
successful registration (e.g., “PASS 1”).
5 Maintenance
USMGS-55041
Page 42
5 Maintenance
5.4.5Adding a Second Sensor
1. Power down the instrument.
2. Open the housing. Loosen the six screws of
the lid using an M5 hex key and remove the lid.
3. Remove an unused plug from the enclosure, if
necessary.
4. Feed the connector end of the sensor
assembly into the enclosure.
5.
Screw cable gland with attached gasket into
the housing and tighten (maintain IP rating).
6. Plug the new sensing head connector into the
open sensor connector socket on the interface
board.
7. Place the bezel back into the enclosure and
secure with the three screws to the standoffs.
8. Replace the enclosure lid as appropriate.
9. Power-up the instrument.
10.
The instrument will prompt to start the sensor
registration (e.g., “rEg 2”). Acknowledge with
[ENTER]. The instrument will then indicate
successful registration (e.g., “PASS 2”).
42MGS-550US
Page 43
5.4.6Replacing the Remote Sensing Head Assembly
See illustration in section 5.4.1 Components Overview.
1. Power down the instrument.
2. Loosen the six screws of the lid using an M5 hex key and
remove the lid.
3. Unplug the old sensing head.
4. Unscrew cable gland of old sensing head and remove.
5. Screw cable gland into the housing and tighten (maintain IP
rating).
6. Plug new sensing head connector into socket.
7. Place bezel back into the enclosure and secure with the three
screws to the standoffs.
8. Put lid back on and tighten the six screws.
9. Power-up instrument.
10. The instrument will prompt to start the sensor registration
(e.g., “rEg 1”). Acknowledge with [ENTER]. The instrument
will then indicate successful registration (e.g., “PASS 1”).
5 Maintenance
USMGS-55043
Page 44
5 Maintenance
5.5Replacing the Instrument Electronics
CAUTION!
This product uses semiconductors which can be damaged by electrostatic discharge (ESD). When handling the PCB,
care must be taken so that the electronics are not damaged.
Failure to follow this caution can result in minor or moderate injury.
1. The user should note parameters, settings such as alarm setpoints,
relay configurations, etc., as the unit will have factory settings after
replacement of the processor board.
2. Power down the instrument.
3. Remove the housing lid.
Loosen the six screws of the lid using an M5 hex key and remove the
lid.
4. Unplug ribbon cable from Interface Board.
5. Attach bezel to new Processor Board.
6. Plug ribbon cable of new Processor Board into socket of Interface
Board.
7. Place bezel (and attached Processor Board) back into the enclosure
and secure with the three screws to the standoffs.
8. Power-up instrument.
9. Register the sensor(s) (both sequentially, if applicable). See section 4.3.6 Register Sensor (F-06).
10.Place bezel back into the enclosure and secure with the three screws to the standoffs.
11.Replace the housing lid (maintain IP rating). Tighten the six screws of the lid using an M5 hex key.
12.Check settings of the parameters (see section 4.4 Parameters), perform an instrument test (see section 4.3.4
Instrument Test (F-04)), and check calibration (see section 5.2.4 Span Adjustment).
44MGS-550US
Page 45
5 Maintenance
5.6Replacing the Interface Board
CAUTION!
This product uses semiconductors which can be damaged by electrostatic discharge (ESD). When handling the PCB,
care must be taken so that the electronics are not damaged.
Failure to follow this caution can result in minor or moderate injury.
1. Power down the instrument.
2. Remove the housing lid. Loosen the six screws of the lid using an M5 hex
key and remove the lid.
3. Unplug ribbon cable from Interface Board.
4. Label all connected wires, then disconnect all interface board wiring.
5. Using a #4 Torx head screwdriver or Torx key, loosen the screws that hold
the Interface Board in place. Remove the Interface Board from the
enclosure.
6. Replace the old Interface Board with the new one.
7. Secure the Interface Board to the enclosure by reversing the previous
removal procedure.
8. With the Interface Board in place, re-attach the wiring using the labels
from the earlier step.
9. Replace the housing lid. Tighten the six screws of the lid using an M5 hex
key.
10.Power up the instrument and verify proper operation.
11.Register sensors. See section 4.3.6 Register Sensor (F-06).
5.7Cleaning the Instrument
Clean the instrument with a soft cloth using water and a mild detergent. Rinse with water.
14 = any alarm or sensor fault or
critical instrument fault
7Sensor Principle
7.1Electrochemical Sensors
Electrochemical sensors measure the partial pressure of gases under atmospheric conditions. The monitored ambient
air diffuses through a membrane into the liquid electrolyte in the sensor. The electrolyte contains a measuring
electrode, a counter-electrode and a reference electrode. An electronic “potentiostat” circuit ensures a constant
electrical voltage between measuring electrode and reference electrode. Voltage, electrolyte, and electrode material
are selected to suit the gas being monitored so that it is transformed electrochemically on the measuring electrode and
a current flows through the sensor. This current is proportional to the gas concentration. At the same time, oxygen from
the ambient air reacts at the counter electrode electrochemically. The current flowing through the sensor is amplified
electronically, digitized and corrected for several parameters (e.g., the ambient temperature).
7.2Catalytic Bead Sensors
A catalytic bead sensor measures the partial pressure of combustible gases and vapors in ambient air. It uses the heatof-combustion principle.
The monitored air diffuses through the sintered metal disc into the sensor. The mixture of combustible gases, vapors,
and air are catalytically combusted at a heated detector element (called a pellistor). The oxygen content in the air must
be greater than 12 Vol%. Due to the resulting heat-of-combustion, the temperature of the detector element rises. This
increase in temperature causes a change of resistance in the detector element, which is proportional to the
concentration of the mixture of combustible gases and vapors in the monitored air. In addition to the catalytically active
detector element, there is a compensator element. Both elements are parts of a Wheatstone bridge. Thus
environmental effects like changes in ambient temperature or humidity are almost entirely compensated.
Certain substances in the atmosphere to be monitored may impair the sensitivity of the sensors. Such
substances include, but are not limited to:
1.
Polymerizing substances such as acrylonitrile, butadiene and styrene,
2.
Corrosive compounds such as halogenated hydrocarbons (releasing halogens such as bromine,
chlorine or fluorine when oxidized) and halogen hydride acids as well as acidic gaseous compounds
such as sulfur dioxide and nitrogen oxides,
3.
Catalyst poisons such as sulfurous and phosphorous compounds, silicon compounds (especially
silicones), and metal-organic vapors.
USMGS-55047
Page 48
8 Disposing of the Instrument
It may be necessary to check the calibration if the sensor has been exposed for a long time to a high concentration of
flammable gases, vapors, or the above-mentioned contaminating substances.
7.3Semiconductor Sensors
Semiconductor or metallic oxide sensors (MOSs) are among the most versatile of all broad-range sensors. They can be
used to detect a variety of gases and vapors in low ppm or even combustible ranges. The sensor is made up of a
mixture of metallic oxides. They are heated to a temperature between 150º and 300º C depending on the gas(es) to be
detected. The temperature of operation as well as the “recipe” of mixed oxides determines the sensor selectivity to
various toxic gases, vapors, and refrigerants. Electrical conductivity greatly increases as soon as a diffusion process
allows the gas or vapor molecules to come in contact with the sensor surface. Water vapor, high ambient humidity,
temperature fluctuations, and low oxygen levels can result in higher readings.
Certain substances in the environment to be monitored may impair the sensitivity of the sensors:
1.
Materials containing silicone or silicone rubber/putty
2.
Corrosive gases such as hydrogen sulfide, sulfur oxide, chlorine, hydrogen chloride, etc.
3.
Alkaline metals, salt water spray.
7.4Infrared Sensors
The infrared (IR) gas sensor is designed to measure the concentration of combustible gases and vapors in the ambient
air. The sensor principle is based on the concentration-dependent absorption of infrared radiation in measured gases.
The monitored ambient air diffuses through a sintered metal material into the housing of an optical “bench”. The
broadband light emitted by an IR source passes through the gas in the optical bench and is reflected by the walls from
where it is directed towards a dual-element detector. One channel of the detector measures the gas-dependent light
transmission, while the other channel is used as a reference. The ratio between measurement and reference signal is
used to determine the gas concentration. Internal electronics and software calculate the concentration and produce an
output signal.
8Disposing of the Instrument
8.1Disposing of Electrical and Electronic Equipment
EU-wide regulations governing the disposal of electrical and electronic appliances which have been
defined in the EU Directive 2012/19/EU and in national laws have been effective since August 2012 and
apply to this device.
Common household appliances can be disposed of using special collecting and recycling facilities.
However, this device has not been registered for household usage. Therefore it must not be disposed of
through these channels. The device can be returned to your national MSA Bacharach Sales Organization
for disposal. Please do not hesitate to contact MSA Bacharach if you have any further questions on this
issue.
48MGS-550US
Page 49
9 Technical Data
8.2Disposing of Electrochemical Sensors
Dispose of sensors in accordance with local laws.
WARNING!
• EXPLOSION DANGER! Do not dispose of sensors in fire due to the risk of explosion and resulting chemical
burns.
• Do not force open electrochemical sensors.
• Observe the applicable local waste disposal regulations. For information, consult your local environmental
agency, local government offices or appropriate waste disposal companies.
Failure to follow these warnings can result in serious personal injury or death.
9Technical Data
9.1General Specifications
CategorySpecifications
Normal operation:4 to 20 mA
Drift below zero:3.8 mA
Measuring range exceeded: 20.5 mA
Analog Current
Analog Voltage
Signals to Central Controller
Modbus RTU over RS-485
Operating Voltage19.5 to 28.5 VDC; 24 VAC ± 20%, 50/60 Hz
Power MonitoringGreen LED
Inrush Current1.5 A
Operating current, max.330 mA at 24 VDC
Power Supply and Relays
Relay rating
Audible alarmInternal Buzzer ≥85 dB at 4” (10 cm); open enclosure
Alarm delay0 to 15 minutes (selectable)
Power and analog signal
Wiring
Modbus network
Instrument fault:≤ 1.2 mA
Fault on analog interface:> 21 mA
Offline mode /
Maintenance signal:
0 to 5 V; 1 to 5 V; 0 to 10 V; 2 to 10 V (selectable)
During fault condition, 1 to 5 V and 2 to 10 V outputs are 0
V.
Baud rate:9,600 or 19,200 (selectable)
Start bits:1
Data bits:8
Parity:
Stop bits:1 or 2 (selectable)
Retry time:
End of message:Silent 3.5 characters
3 SPDT, arbitrary designation
1 A at 24 VAC/VDC, 0.5 A at 125 VAC, resistive load
2-, 3-, 4-, or 6-core shielded cable, 16 to 20 AWG (0.5 to
1.5 mm2)
2-core twisted pair shielded cable 16 to 20 AWG (0.5 to
Enclosure protectionIP66
Display5 digit, 7 segment LED
Enclosure Size (WxHxD)
(Approx.)
Temperature- 40 to 120 ºF (-40 to 50 ºC)
Storage temperature- 5 to 100 ºF (-20 to 40 ºC)
Humidity5 to 90 %RH, non-condensing
Pressure20.7 to 38.4 in. of Hg (700 to 1300 hPa)
Elevation0 to 10,000 ft. (3050 m) altitude
SensorsSee Section 9.2 Sensor Specifications
Influences
USA Certification (for units assembled in USA): MSA
Bacharach (USA) is a TÜV-SÜD certified company.
Ireland Certification (for units assembled in Ireland):
Murco (a Bacharach Company) is an NSAI certified
company.
1.5 mm2) with 120 Ohm characteristic impedance
ABS
210x225x85 mm (8.3x8.9x3.1 in)
For influences on the measurement performance and
restrictions of a particular sensor see sensor data sheet.
9.2Sensor Specifications
ECSensorsFormulaMeasure Range(s)
AmmoniaNH
3
Carbon MonoxideCO0 to 1,000 ppm
Chlorine
Fluorine
HydrogenH
Hydrogen Chloride
1
1
1
Cl
2
F
2
2
HCl0 to 10 ppm
Hydrogen CyanideHCN0 to 30 ppm
Hydrogen SulfideH2S0 to 100 ppm
Nitrogen DioxideNO
OxygenO
Ozone
1
Suflure DioxideSO
1
No IPrating
2
2
O
3
2
0 to 100, 0 to 1,000, 0 to 5,000 ppm
0 to 10 ppm
0 to 1 ppm
0 to 10,000 ppm
0 to 20 ppm
0 to 30 Vol%
0 to 1 ppm
0 to 10 ppm
50MGS-550US
Page 51
9 Technical Data
IRSensorsFormulaMeasure Range(s)
0 to 5,000 ppm = 0 to 0.5 Vol%
0 to 10,000 ppm = 0 to 1 Vol%
0 to 20,000 ppm = 0 to 2 Vol%
Carbon DioxideCO
2
0 to 30,000 ppm = 0 to 3 Vol%
0 to 40,000 ppm = 0 to 4 Vol%
0 to 50,000 ppm = 0 to 5 Vol%
Hydrocarbons (Butane, Methane, and
Propane)
C4H
CH
C3H
10
4
8
0 to 100 %LEL
0 to 100 %LEL
0 to 100 %LEL
CATSensorsFormulaMeasuring Range
Combustible Gasses (including
Ammonia)
Various0 to 100 %LEL
SCSensorsTypical ExampleGasesMeasure Range
HFCsR134a, R404A, R407C, R410A, R507 0 to 1,000, 0 to 10,000 ppm
R220 to 1,000, 0 to 10,000 ppm
HCFCs and HFOs
R1234YF0 to 1,000 ppm
R1234ZE0 to 1,000, 0 to 10,000 ppm
R-448a, R-449a, R-452a, R-513a0 to 1,000 ppm
(SC Ethylene is 0 to 2,000 ppm)
0 to 10,000 ppm
0 to 5,000 ppm
Acetone, Chloroform, Ethanol,
VOCs
Methanol, Methylene Chloride,
0 to 1,000 ppm
Ethylene Chloride
9.3Modbus Registers
NOTE: If items span two registers (e.g., 30024 and 30025), then the registers are “long” or “float” data types. Otherwise
the registers are integer data types or ASCII.
NOTE: If a sensor is not registered, then reading from a Modbus register associated with that sensor will generate a
Modbus exception.
9.3.1Analog Input Registers
Analog input registers are read only and use Modbus function code 04 (Read Input Register).
RegDescriptionRangeUnitsParam
3000116-bit Current Fault Code Instrument electronics
3000216-bit Last Fault Code Instrument electronics-D-02
30003Software Version Instrument electronics--30004
Reserved---
See 5.3.4 Non-Critical
(E300) Faults and 5.3.3
Instrument Electronics
Critical (E100) Faults
-D-01
USMGS-55051
Page 52
9 Technical Data
RegDescriptionRangeUnitsParam
-
30010
30011Sensor 1 Type Code--S1-09
1 = ppm
30012Display units sensor 1
-S1-10
3 = Vol%
4 = %LEL
Full Scale in ppb, ppm, %LEL or Vol% Sensor 1 (Note: %
2 = ppb
30013
LEL and Vol% sensors are always displayed as “x 10”
the actual value. For example, a reading of "205" is
0 to 65535
Unit
dependent
S1-11
"20.5%".)
30014Sensor 1 Gas Type Text Char 1, 20 to 9; a to zASCII Text
S1-0130015Sensor 1 Gas Type Text Char 3, 40 to 9; a to zASCII Text
30016Sensor 1 Gas Type Text Char 5, NULL0 to 9; a to zASCII Text
30017
%LEL/Vol% Conversion Factor * 10 Sensor 1. This is the
factor to convert the integer gas output level from %LEL
or Vol% to ppm.
0 to 65535
Unit
dependent
-
30018Alarm Low Set-Point Sensor 10 to 100% Full scale S1-03
30019Alarm Medium Set-Point Sensor 10 to 100% Full scale S1-04
30020Alarm High Set-Point Sensor 10 to 100% Full scale S1-05
Concentration ppb, ppm, %LEL or Vol% Sensor 1 (Note:
30021
% LEL and Vol% sensors are always displayed as “x 10”
the actual value. For example, a reading of "205" is
0 to 65535
Unit
dependent
-
"20.5%".) Units are in register 30012.
30022Concentration Sensor 10 to 100
% Full
Scale
-
30023Sensor timer Sensor 10 to 65535Hours-
30024ppm Hours Sensor 1 (upper long integer)
-
ppm Hours
-
30025ppm Hours Sensor 1 (lower long integer)
Temperature Deg C Sensor 1
30026
-40 to +80°C-
NOTE: Sensor PCB temp, not gas temp
3002716 bit Fault Code Sensor 1See 5.3.5 Sensor Faults-D-07
30028Sensor 1 Instrument electronics UID Char 1,20 to 9; a to zASCII Text
S1-0230029Sensor 1 Instrument electronics UID Char 3,40 to 9; a to zASCII Text
30030Sensor 1 Instrument electronics UID Char 5,NULL0 to 9; a to zASCII Text
30031Sensor 2 Type Code--S2-09
1 = ppm
2 = ppb
30032Display units Sensor 2
-S2-10
3 = Vol%
4 = %LEL
30033
52MGS-550US
Full Scale in ppb, ppm, %LEL or Vol% Sensor 2. (Note:
0 to 65535
Unit
S2-11
Page 53
9 Technical Data
RegDescriptionRangeUnitsParam
% LEL and Vol% sensors are always displayed as “x 10”
the actual value. For example, a reading of "205" is
"20.5%".)
30034Sensor 2 Gas Type Text Char 1,20 to 9; a to zASCII Text
30036Sensor 2 Gas Type Text Char 5,NULL0 to 9; a to zASCII Text
%LEL/Vol% Conversion Factor * 10 Sensor 2. This is the
30037
30038Alarm Low Set-Point Sensor 2
30039Alarm Medium Set-Point Sensor 230040Alarm High Set-Point Sensor 2-
30041
30042Concentration Sensor 20 to 100
30043Sensor timer Sensor 20 to 65535Hours30044ppm Hours Sensor 2 (upper long integer)
30045ppm Hours Sensor 2 (lower long integer)
30046Temperature Deg C Sensor 2-40 to +80°C3004716-bit Fault Code Sensor 2See 5.3.5 Sensor Faults-D-09
30048Sensor 2 Instrument electronics UID Char 1,20 to 9; a to zASCII Text
30050Sensor 2 Instrument electronics UID Char 5,NULL0 to 9; a to zASCII Text
30051Sensor 1 Module SID Char 1,20 to 9; a to zASCII Text
30053Sensor 1 Module SID Char 5,NULL0 to 9; a to zASCII Text
30054Sensor 2 Module SID Char 1,20 to 9; a to zASCII Text
30055Sensor 2 Module SID Char 3,40 to 9; a to zASCII Text
30056Sensor 2 Module SID Char 5,NULL0 to 9; a to zASCII Text
factor to convert the integer gas output level from %LEL
or Vol% to ppm.
Concentration ppb, ppm, %LEL or Vol% Sensor 2. (Note:
% LEL and Vol% sensors are always displayed as “x 10”
the actual value. For example, a reading of "205" is
"20.5%".)
0 to 65535
0 to 100% Full scale
0 to 65535
-ppm Hours-
dependent
Unit
dependent
Unit
dependent
% Full
Scale
S2-0130035Sensor 2 Gas Type Text Char 3,40 to 9; a to zASCII Text
-
-
-
-
S2-0230049Sensor 2 Instrument electronics UID Char 3,40 to 9; a to zASCII Text
-30052Sensor 1 Module SID Char 3,40 to 9; a to zASCII Text
-
9.3.2Analog Output Registers
Analog output registers are readable (using function code 03) and writable (using function code 06).
RegDescriptionRangeUnitsParam
40001Display Mode0=Off, 1=On-P1-01
40002RS-485 Node Address1 to 250-MB-01
40003Baud Rate0=9600, 1=19200-MB-02
40004Stop Bits1 or 2-MB-03
40005Parity0=None, 1=Odd, 2=Even-MB-04
40006Alarm Delay ON Value00 to 15MinutesAF-01
USMGS-55053
Page 54
9 Technical Data
RegDescriptionRangeUnitsParam
40007Alarm Delay OFF Value00 to 15MinutesAF-02
40008Controller UID Char 1,20 to 9; a to zASCII Text
40009Controller UID Char 3,40 to 9; a to zASCII Text
40010Controller UID Char 5,NULL0 to 9; a to zASCII Text
Sensor 1 Low Alarm ppb, ppm, %LEL or Vol%. (Note:
40011
% LEL and Vol% sensors are always displayed as “x
10” the actual value. For example, a reading of "205" is
0 to sensor full scale
"20.5%".)
Sensor 1 Med Alarm ppb, ppm, %LEL or Vol%. (Note:
40012
% LEL and Vol% sensors are always displayed as “x
10” the actual value. For example, a reading of "205" is
0 to sensor full scale
"20.5%".)
Sensor 1 High Alarm ppb, ppm, %LEL or Vol%. (Note:
40013
% LEL and Vol% sensors are always displayed as “x
10” the actual value. For example, a reading of "205" is
0 to sensor full scale
"20.5%".)
40014Sensor 1 Low Alarm BehaviorS1-06
40015Sensor 1 Medium BehaviorS1-07
40016Sensor 1 High BehaviorS1-08
Sensor 2 Low Alarm ppb, ppm, %LEL or Vol%. (Note:
40017
% LEL and Vol% sensors are always displayed as “x
10” the actual value. For example, a reading of "205" is
0 to sensor full scale
"20.5%".)
Sensor 2 Med Alarm ppb, ppm, %LEL or Vol%. (Note:
40018
% LEL and Vol% sensors are always displayed as “x
10” the actual value. For example, a reading of "205" is
0 to sensor full scale
"20.5%".)
Sensor 2 High Alarm ppb, ppm, %LEL or Vol%. (Note:
40019
% LEL and Vol% sensors are always displayed as “x
10” the actual value. For example, a reading of "205" is
0 to sensor full scale
"20.5%".)
40020Sensor 2 Low Alarm BehaviorS2-06
40021Sensor 2 Medium BehaviorS2-07
40022Sensor 2 High BehaviorS2-08
Unit
dependent
Unit
dependent
Unit
dependent
Unit
dependent
Unit
dependent
Unit
dependent
S1-03
S1-04
S1-05
S2-03
S2-04
S2-05
40023Analog Output 1 Source
-A1-01
2 = Sensor 2
0 = 0 to 5 V
1 = 1 to 5
1 = Sensor 1
40024Analog Output 1 Type
2 = 0 to 10 V
-A1-02
3 = 2 to 10 V
4 = 4 to 20 mA
40025Analog Output 1 Scaling20% to 100% of full scale%A1-03
40026Analog Output 1 Offset
±410 (can take an output
below zero)
410 = 10%
of full scale
A1-04
40027Analog Output 1 Full Scale PPM
54MGS-550US
Page 55
9 Technical Data
RegDescriptionRangeUnitsParam
40028Analog Output 2 Source
-A2-01
2 = Sensor 2
0 = 0 to 5 V
1 = 1 to 5
1 = Sensor 1
40029Analog Output 2 Type
2 = 0 to 10 V
-A2-02
3 = 2 to 10 V
4 = 4 to 20 mA
40030Analog Output 2 Scaling20% to 100% of full scale%A2-03
40038Relay 3 Source-R3-01
40039Buzzer enable and source
40040Critical Fault LatchAF-03
4.4.7 Buzzer Designation
(B1-xx)
-
B1-01
5.3.4 Non-Critical (E300)
40041Instrument electronics fault code
Faults and 5.3.3
Instrument Electronics
-D-01
Critical (E100) Faults
5.3.4 Non-Critical (E300)
40042Instrument electronics last fault
Faults and 5.3.3
Instrument Electronics
-D-02
Critical (E100) Faults
40043Power supply voltage
15.5-28.5 VDC or 24 VAC
±20%
VoltD-04
40044Hours since last test sensor 1HoursD-05
40045Hours since last test sensor 2HoursD-06
40046Reserved
40047Instrument electronics 12V supply voltage x100
40048Instrument electronics 6V supply voltage x100
40049Instrument electronics 5.4V supply voltage x100
40050Instrument electronics 3.3V supply voltage x100
40051Instrument electronics temperature x100
40052Instrument electronics tact and magnetic switch state
USMGS-55055
Page 56
9 Technical Data
9.3.3Input Status Flags
Input status flags are readable (using function code 02).
RegDescriptionRange
10001Sensor 1 Low Alarm Flag0 or 1 = alarm
10002Sensor 1 Medium Alarm Flag0 or 1 = alarm
10003Sensor 1 High Alarm Flag0 or 1 = alarm
10004Sensor 1 Fault0 or 1 = fault
10005Sensor 1 Saturation Overflow0 or 1 = gas > full scale
10006Sensor 1 Saturation Underflow0 or 1 = gas < 0
10007Sensor 1 Start-up0 or 1 = start-up
10008 - 10010Reserved
10011Sensor 2 Low Alarm Flag0 or 1 = alarm
10012Sensor 2 Medium Alarm Flag0 or 1 = alarm
10013Sensor 2 High Alarm Flag0 or 1 = alarm
10014Sensor 2 Fault0 or 1 = fault
10015Sensor 2 Saturation Overflow0 or 1 = gas > full scale
10016Sensor 2 Saturation Underflow0 or 1 = gas < 0
10017Sensor 2 Start-up0 or 1 = start-up
10018 - 10020Reserved
10021Relay 1 State0 or 1 = energized
10022Relay 2 State0 or 1 = energized
10023Relay 3 State0 or 1 = energized
10024Instrument Electronics Offline0 or 1 = offline
10025Instrument Electronics Non-Critical Fault0 or 1 = fault
10026Instrument Electronics Critical Fault0 or 1 = fault
NOTE: A sensor must be re-zeroed if it shows underflow saturation.
9.3.4Output Status Flags
Output status flags are readable using Modbus function code 01 and writable using function code 05.
RegDescriptionRange
00001
00002
00003Alarm Flag (0 or 1 = alarm) for any alarm (not fault).
00004
00005
00006
Sensor 1 calibration expired. This flag can be cleared by performing a calibration or by
resetting.
Sensor 2 calibration expired. This flag can be cleared by performing a calibration or by
resetting.
Writing zero clears any acknowledgeable or latched alarm. This is the same behavior as
holding the ENTER key for 5 seconds.
Relay closed test. Setting this flag to zero closes all 3 relays simultaneously for 5 seconds. At
the end of the test the relays revert to their normal operation.
Relay opened test. Setting this flag to zero opens all 3 relays simultaneously for 5 seconds. At
0 or 1 =
need
calibration
0 or 1 =
need
calibration
0 or 1 =
alarm
56MGS-550US
Page 57
10 Ordering Information
RegDescriptionRange
the end of the test the relays revert to their normal operation
Analog Output Zero Test. Setting this to one drives the analog outputs to their minimum value.
This depends on the configuration of the analog output.
00007
For 4-20mA it will drive 4mA
l
For all voltage ranges it will drive 0V
0 or 1 = test
in process
l
During the test the Modbus flag will remain ON. When the test is completed the flag will turn
OFF
Display Test Mode. Setting this sets all 7-segment display segments, power LED and both units
LEDs active simultaneously for 5 seconds. At the end the display will revert to normal
00008
operation.
0 or 1 = test
in process
During the test the Modbus flag will remain ON. When the test is completed the flag will turn
OFF. This test is only operative outside the menus.
10Ordering Information
10.1MGS-550 Instrument Only
NOTE: In the descriptions below, "MGS-550 Gas Detector" includes an instrument and one sensing head mounted
directly to the instrument housing. Remote or secondary local sensors must be ordered as separate items.
P/NMGS-550 Gas Detector, IP66 with IP66 Sensor Installed at the Factory
6600-8010MGS-550 Gas Detector, IP66, IR, CO2, 0-5000 ppm
6600-8011MGS-550 Gas Detector, IP66, IR, CO2, 0-10000 ppm
6600-8012MGS-550 Gas Detector, IP66, IR, CO2, 0-20000 ppm
6600-8013MGS-550 Gas Detector, IP66, IR, CO2, 0-30000 ppm
6600-8014MGS-550 Gas Detector, IP66, IR, CO2, 0-40000 ppm
6600-8015MGS-550 Gas Detector, IP66, IR, CO2, 0-50000 ppm
6600-8016MGS-550 Gas Detector, IP66, IR, Butane, 0-100 %LEL
6600-8017MGS-550 Gas Detector, IP66, IR, Methane, 0-100 %LEL
6600-8018MGS-550 Gas Detector, IP66, IR, Propane, 0-100 %LEL
6600-8019MGS-550 Gas Detector, IP66, EC, NH3, 0-100 ppm
6600-8020MGS-550 Gas Detector, IP66, EC, NH3, 0-1000 ppm
6600-8021MGS-550 Gas Detector, IP66, EC, NH3, 0-5000 ppm
6600-8023MGS-550 Gas Detector, IP66, EC, CO, 0-1000 ppm
6600-8024MGS-550 Gas Detector, IP66, EC, NO2, 0-20 ppm
6600-8025MGS-550 Gas Detector, IP66, EC, O2, 0-30 Vol%
6600-8026MGS-550 Gas Detector, no IP rating, EC, Cl2, 0-10 ppm
6600-8027MGS-550 Gas Detector, no IP rating, EC, F2, 0-1 ppm
6600-8028MGS-550 Gas Detector, no IP rating, EC, O3, 0-1 ppm
6600-8029MGS-550 Gas Detector, IP66, EC, SO2, 0-10 ppm
USMGS-55057
Page 58
10 Ordering Information
P/NMGS-550 Gas Detector, IP66 with IP66 Sensor Installed at the Factory
6600-8030MGS-550 Gas Detector, IP66, EC, H2S, 0-200 ppm
6600-8031MGS-550 Gas Detector, no IP rating, EC, HCl, 0-10 ppm
6600-8032MGS-550 Gas Detector, IP66, EC, HCN, 0-30 ppm
6600-8033MGS-550 Gas Detector, IP66, EC, H2, 0-10000 ppm
6600-8034MGS-550 Gas Detector, IP66, CT, NH3, 0-100 %LEL
6600-8035MGS-550 Gas Detector, IP66, CT, LPG, 0-100 %LEL
6600-8036MGS-550 Gas Detector, IP66, CT, Methane, 0-100 %LEL
6600-8038MGS-550 Gas Detector, IP66, CT, Propane, 0-100 %LEL
6600-8039MGS-550 Gas Detector, IP66, CT, Butane, 0-100 %LEL
6600-8040MGS-550 Gas Detector, IP66, SC, R22, 0-1000 ppm
6600-8041MGS-550 Gas Detector, IP66, SC, R32, 0-1000 ppm
6600-8042MGS-550 Gas Detector, IP66, SC, R134a, 0-1000 ppm
6600-8043MGS-550 Gas Detector, IP66, SC, R404a, 0-1000 ppm
6600-8044MGS-550 Gas Detector, IP66, SC, R407a, 0-1000 ppm
6600-8045MGS-550 Gas Detector, IP66, SC, R407c, 0-1000 ppm
6600-8046MGS-550 Gas Detector, IP66, SC, R407f, 0-1000 ppm
6600-8047MGS-550 Gas Detector, IP66, SC, R410a, 0-1000 ppm
6600-8048MGS-550 Gas Detector, IP66, SC, R422, 0-1000 ppm
6600-8049MGS-550 Gas Detector, IP66, SC, R422d, 0-1000 ppm
6600-8050MGS-550 Gas Detector, IP66, SC, R427a, 0-1000 ppm
6600-8051MGS-550 Gas Detector, IP66, SC, R507, 0-1000 ppm
6600-8052MGS-550 Gas Detector, IP66, SC, HFO1234YF, 0-1000 ppm
6600-8053MGS-550 Gas Detector, IP66, SC, HFO1234ZE, 0-1000 ppm
6600-8054MGS-550 Gas Detector, IP66, SC, R22, 0-10000 ppm
6600-8055MGS-550 Gas Detector, IP66, SC, R32, 0-10000 ppm
6600-8056MGS-550 Gas Detector, IP66, SC, R134a, 0-10000 ppm
6600-8057MGS-550 Gas Detector, IP66, SC, R404a, 0-10000 ppm
6600-8058MGS-550 Gas Detector, IP66, SC, R407a, 0-10000 ppm
6600-8059MGS-550 Gas Detector, IP66, SC, R407c, 0-10000 ppm
6600-8060MGS-550 Gas Detector, IP66, SC, R407f, 0-10000 ppm
6600-8061MGS-550 Gas Detector, IP66, SC, R410a, 0-10000 ppm
6600-8062MGS-550 Gas Detector, IP66, SC, R422, 0-10000 ppm
6600-8063MGS-550 Gas Detector, IP66, SC, R422d, 0-10000 ppm
6600-8065MGS-550 Gas Detector, IP66, SC, R507, 0-10000 ppm
6600-8067MGS-550 Gas Detector, IP66, SC, HFO1234ZE, 0-10000 ppm
6600-8069MGS-550 Gas Detector, IP66, SC, NH3, 0-10000 ppm
6600-8070MGS-550 Gas Detector, IP66, SC, R290, 0-5000 ppm
6600-8071MGS-550 Gas Detector, IP66, SC, R600, 0-5000 ppm
6600-8072MGS-550 Gas Detector, IP66, SC, H2, 0-5000 ppm
6600-8073MGS-550 Gas Detector, IP66, SC, CH4, 0-5000 ppm
6600-8074MGS-550 Gas Detector, IP66, SC, VOC/Ethanol, 0-1000 ppm
58MGS-550US
Page 59
10 Ordering Information
P/NMGS-550 Gas Detector, IP66 with IP66 Sensor Installed at the Factory
6600-8075MGS-550 Gas Detector, IP66, SC, Ethylene, 0-2000 ppm
6600-8076MGS-550 Gas Detector, IP66, SC, R-448a, 0-1,000 ppm
6600-8077MGS-550 Gas Detector, IP66, SC, R-449a, 0-1,000 ppm
6600-8078MGS-550 Gas Detector, IP66, SC, R-452a, 0-1,000 ppm
6600-8079MGS-550 Gas Detector, IP66, SC, R-513a, 0-1,000 ppm
10.3MGS-550 5m Remote and Second Sensing Heads
NOTE: Below is the MGS-550 Sensing Head Part Number Configurator. Part number format is: 6600-8ABC.
NOTE: All 5m remote and second, directly mounted sensing heads must be ordered separately.
Code
"A"
1IP66 Sensing Head: 5 meter cable, for remote sensor applications
8IP66 Sensing Head: For mounting directly to MGS-550 instrument
IP66 Sensing Head Cable Type "A" Select the Required Cable Length, Noting the "A"
Code
Codes "B &C"Sensing Head Target Gas and Range "B & C" Select the Target Gas and Range, Noting
6600-8402Replacement display PCB Assembly, MGS-550
6600-8403Ribbon Cable, MGS-550. Connects top and bottom PCBs
6600-8404Accessory Kit, MGS-550. Includes spare cable glands and blanking plugs, gaskets.
60MGS-550US
Replacement interface I/O PCB Assembly, MGS-550. Base PCB with terminal blocks for IP66
housing.
Page 61
P/NMGS-550 Replacement Parts and Accessories
6600-8405Sensing Head Common Control PCBA, MGS-550
Sensing Head Front End Kit, MGS-550, IP66. Includes lock nut, nosepiece with membrane, O-ring,
sensor foam spacers, and splash guard with calibration port. Also included are gaskets to be used as
follows.