Atec F6150 User Manual

F6150
Over 10,000 Doble® protection test instruments are currently used by test engineers around the world,
®
to ensure the reliability of electric power systems.
Total Sources
The F6150 can provide up to twelve simultaneous user configurable AC/DC sources including: six convertible sources and six current sources.
Convertible Sources
Each 150 VA Convertible Source can be used as a voltage source or optionally as a high-power, low range current source. The F6150 has up to six Convertible Sources.
Source Configurations
Output Power
Continuous Transient for 1.5 Seconds Number of Sources 75 VA 97.5 VA 6 150 VA 195 VA 3 300 VA 390 VA 2x150 VA or 1x150 VA+2x75 VA 450 VA 585 VA 1
Each 150 VA convertible source can be split into two 75 VA sources. Two 150 VA convertible sources can be combined in parallel into one 300 VA current source. Three 150 VA convertible sources can be combined in parallel into one 450 VA current source.
Ranges and Resolution
75 VA Source F6150 Ranges (Resolution)
AC Voltage 75, 150 V rms (0.01V)
DC Voltage 106, 212 V dc (0.01V)
AC Current Transient 0.75, 1.5 A rms (0.0001
A)
Continuous 0.5, 1.0 A rms (0.0001 A) DC Current Transient 0.53, 1.06 A dc (0.0001
A)
Continuous 0.354, 0.707 A dc
(0.0001 A)
150 VA Source F6150 Ranges (Resolution)
AC Voltage 75, 150, 300 V rms
(0.01 V)
DC Voltage 106, 212 V (0.01 V), 300
VDC (0.1 V)
AC Current Transient 0.75, 1.5, 3.0 A rms
(0.0001 A)
Continuous 0.5, 1.0, 2.0 A rms
(0.0001 A)
DC Current Transient 0.53, 1.06 (0.0001 A),
2.12 A dc (0.001 A)
Continuous 0.354, 0.707 (0.0001 A),
1.41 A dc (0.001 A)
300 VA Source F6150 Ranges (Resolution)
AC Current
Transient 1.5, 3.0, 6.0 A rms
(0.001 A)
Continuous 1.0, 2.0, 4.0 A rms
(0.001 A)
DC Current Transient 1.06 A (.0.0001 A), 2.12,
4.24 A dc (0.001 A)
Continuous 0.707 A (0.0001 A), 1.41
A, 2.83 A dc (0.001 A)
450 VA Source F6150 Ranges (Resolution)
AC Current
Transient 2.25, 4.5, 9.0 A rms
(0.001 A)
Continuous 1.5, 3.0, 6.0 A rms
(0.001 A)
DC Current Transient 1.59 (0.0001 A), 3.18,
6.36 A dc (0.001 A)
Continuous 1.06, 2.12, 4.24 A dc
(0.001 A)
Australia
strstr
China
Ch
India
Norway
UK
U
United States
SS
ta
Current Sources
The F6150 has up to twelve current sources available including: six current sources and six high­power, low-range convertible sources.
Source Configurations
Output Power
Continuous Transient for 1.5 Seconds Number of Sources 75 VA 112.5 VA 6 150 VA 225 VA 3 300 VA 450 VA 2x150 VA or 1x150 VA+2x75 VA 450 VA 675 VA 1
Each 150 VA current source can be split into two 75 VA current sources. Two 150 VA current sources can be combined in series or in parallel into one 300 VA current source. Three 150 VA current sources can be combined in parallel into one 450 VA current source.
Ranges and Resolution:
75 VA Source F6150 Ranges (Resolution)
AC Current Transient 15, 30 A rms (0.001 A)
Continuous 7.5, 15 A rms (0.001 A) DC Current Transient 10, 20 A dc (0.01 A)
Continuous 5 A (0.001 A),
10 A dc (0.01 A)
150 VA Source F6150 Ranges (Resolution)
AC Current Transient 15, 30 A (0.001 A),
60 A rms (0.01 A)
Continuous 7.5, 15 A (0.001 A), 30 A
rms (0.01 A)
DC Current Transient 10, 20, 40 A dc (0.01 A)
Continuous 5 A (0.001 A), 10,
20 A dc (0.01 A)
300 VA Source F6150 Ranges (Resolution)
AC Current Transient 15, 30 A (0.001 A), 60,
120 A rms (0.01 A)
Continuous 7.5, 15 A (0.001 A), 30,
60 A rms (0.01 A)
DC Current Transient 0 A (0.001 A), 20, 40, 80
A dc (0.01 A)
Continuous 5 A (0.001 A), 10, 20, 40
A dc (0.01 A)
450 VA Source: F6150 Ranges (Resolution)
AC Current Transient 15, 30 A (0.001 A), 45,
90, 180 A rms (0.01 A)
Continuous 7.5, 15, 22.5 (0.001 A),
45 A, 90 A rms (0.01 A)
DC Current Transient 10 A (0.001 A), 20, 30,
60, 120 A dc (0.01 A)
Continuous 5 A (0.001 A), 10, 15, 30,
60 A dc (0.01 A)
Australia
strstr
China
Ch
India
Norway
UK
U
United States
SS
ta
F6150 Technical Specifications
AC Amplitude Accuracy at 50/60 Hz Voltage and Current Sources
From 20° to 30° C: <0.02% typical, 0.09% guaranteed
Typically 0.02% of reading
Convertible Source in Current Mode
From 20° to 30°C: <0.5% guaranteed
Distortion at 50/60 Hz Voltage and Current Sources
Total Harmonic Distortion (THD) < 0.02% typical < 0.1% guaranteed
Phase Angle
Range: 0 to +359.9° (Lead) / 0
to –359.9° (Lag)
Accuracy: ±0.25° at 50/60 Hz
Resolution: ±0.1° at 50/60 Hz
Frequency
Bandwidth: dc to 3 kHz at full pow-
er for transient playback
Range: dc; ac from 0.1 Hz to 2
kHz at full power continu­ous load
Resolution: 0.001 Hz Accuracy: 0.5 ppm Typical
1.5 ppm 20° to 30° C
10 ppm 0° to 50° C
Ramp/Set
Ramp: increments/decrements
voltage, current, phase angle, and frequency at user defined ramp rates. Ensures smooth, linear changes in value.
Metering Functions
DC Meter Inputs
Input Range: 0 to ±10V dc or 0 to
±20mA dc
Accuracy: <0.003% typical <+0.05%
guaranteed
AC Sources
Accuracy: <0.02% for typical meter
loads
Logic Inputs as Counters
Frequency: 10 kHz
Pulsewidth: >175 microseconds.
Timers and Triggers
Timers Number: 8
Max Recording Time: <24 Hours Accuracy: ±0.0005% of reading,
±50 microseconds
Resolution: 100 microseconds
Time can be displayed as milliseconds, seconds, or cycles
Triggers Number: 8
Boolean combination of logic inputs can be used to define triggers
Logic Inputs
Number:
logic inputs : 8 total
Isolated inputs:
Number: 2
Configurable as Voltage Sense or Contact Sense
Voltage Sense: Up to 250 V ac or dc
Open Circuit Test Voltage: 12 V dc nominal
Short Circuit Test Current: 20 mA dc nominal
Response Time: 0.1 millisecond max pickup
and dropout
Isolation: ±500 V peak
Paired Logic Inputs:
Number: 3 pairs (6 total)
Configurable as Voltage Sense or Contact Sense
Voltage Sense: Up to 250 V ac or dc
Open Circuit Test Voltage: 4 V dc nominal
Short Circuit Test Current: >50 mA dc nominal
Response Time: 0.1 millisecond max
pickup and dropout
Isolation: ±500 V peak
Logic Outputs
Number: 8
Configurable as Normally Open (NO) or Normally Closed (NC) switches.
High-Speed Electronic Switches
Number: 4
Input Voltage: 250 V dc or ac
Switching Current: 0.5 A make or break,
maximum
Response Time: 0.1 millisecond maximum
pickup and dropout
Isolation: ±500 V peak
Outputs: Relays.
Relays: 4
Breaking Capacity AC: 2000 VA with Vmax
250 V, Imax 8 A
Breaking Capacity DC: 50 W with Vmax
300 V, Imax 8 A
Response Time: <10 millisecond max
pickup and dropout
Isolation between pairs: ±500 V peak
Variable Output Battery Simulator
Range: Adjustable 6 to 300 V dc
Resolution: 0.3 V
Power: 90 W, 1.5 A max
50/60 Hz Ripple: <0.2% of range Accuracy < +/- 5%
Analog Input Measurement (F6820 Option)
Recording: 8 external analog and
digital channels
Source Recording: 12 internal sources
Ranges: 250 mV rms, 2.5 V rms,
25 V rms, 250 mV rms
Accuracy: ±0.06% typical, ±0.15%
maximum
Bandwidth: dc to 5 kHz
Input Impedance: 150 kΩ
Max. Input Voltage: 250 V rms/ dc
Isolation: ±500 V peak channel-to-
channel
General Specifications
Quality Assurance Management System
Third-party certification to ISO 9001:2000
Calibration
Certification traceable to N.I.S.T. standards
Electrostatic Discharge Immunity
IEC 801-2 I.E.C. performance level 1 @ 10 kV: normal performance within specifications. I.E.C. performance level 2 @ 20 kV: no permanent damage.
Surge Withstand Capability
ANSI/IEEE C37.90. The simulator functions as a source during surge withstand capability tests, when the ANSI/IEEE specified isolating circuit is interposed between the simulator and the test relay.
Interfaces
Ethernet or USB control to PC
Line Power Supply
105–132 V or 210–264 V, 47–63 Hz
EMC Emissions
FCC 47 CFR Part 15 Class A (USA) EN55011:1998/A1:1999/A2:2002 Group1 Class A ISM(EU) AS/NZS CISPR 11:2004 Class A ISM (Australia) ICES-001 Issue 3 ISM (Canada)
EMC Immunity
EN 61000-6-2:2005; IEC 61000-4-2/3/4/5/6/11
Safety
EN 61010-1; UL 61010-1; CSA 27.2 # 61010-1
Environmental
IEC 60068-2-2 Dry Heat (+ 85°C storage; + 50°C Operating) IEC 60068-2-1 Cold (- 50°C storage; 0°C operating) IEC 60068-2-30 Damp Heat (+55°C, 6 cycles, 95% humidity) NEMA Enclosure Rating Type 1 IEC Enclosure Rating IP20
Mechanical
IEC 60068-2-27 Shock (15g/11ms, half sine) IEC 60068-2-6 Vibration (10-150 Hz, 20 m/s2) IEC 60068-2-6 Drop test
IEC61850 Communication
Certified by KEMA as being compliant with IEC61850 protocol (IEC 61850-7-2 and 8-1)
Humidity
Up to 95% relative humidity non-condensing
GPS Accuracy
With F6895 (Antenna and Receiver): +/-50 nanoseconds With F6050: +/-10 microseconds
Enclosure
High-impact, molded, flame-retardant ABS – meets National Safe Transit Association testing specification No. 1A for immunity to severe shock and vibration
Dimensions
15 x 9.5 x 18 inches 38 x 24 x 45.7 cm
Weight
37.5 lb 17.05 kg (with front cover and carrying strap)
Additional external F6150 options described in other Doble product brochures:
F6010 Controller
F6080 Field Calibration Unit
F6050 Universal Time
Synchronizer
F6300 High-Power
Current Amplifier
Specifications are subject to change without notice.
®
Doble Engineering Company
85 Walnut Street
Watertown, MA 02472 USA
tel +1 617 926 4900
fax +1 617 926 0528
TOGETHER WE POWER THE WORLD
F6816 External
Input/Output Unit
Doble is certified ISO 9001:2000
Doble is an ESCO Technologies Company
MKT-SL-F6150TS-03/10
Loading...