Current Probe
-10
-20
The F-16-1 is used to measure conducted emissions from 10kHz to 50MHz. Its useable frequency range is 10Hz to 70MHz. The basic F-16 probe has a transfer impedance of 4Ω from 30kHz to 50MHz. The F-16-1 with the power absorbing terminator has a constant transfer impedance of .5Ω from 20kHz to 50MHz. All configurations of the F-16 probe can handle 400 amperes of
rimary power line currents from DC to 60Hz without saturating o
impacting the transfer impedance of the probe.
0
-30
Tranfer Impedance (dB ohms)
-40
-50 100Hz 1kHz 10kHz 100kHz 1MH z 10MH z 100MH z
Frequency
F-16-1 Specifications
Frequency: 100 Hz – 70 MHz Connector: Type-N Internal diameter: 32mm DC to 60 Hz: 400 amperes External diameter: 94mm 400 MHz: 400 amperes Height: 64mm RF(CW): 50 amperes
1
Z
Ω
: 0.5 Pulse2: 300 amperes
t
1
dBΩ
: 6
2: Depends upon the pulse width and pulse rep. Rate.
1: Probe calibrated with 50Ω+j0Ω Load Impedance