Current Probe 
-10
-20
The F-16-1 is used to measure conducted emissions from 10kHz to 
50MHz. Its useable frequency range is 10Hz to 70MHz. The basic 
F-16 probe has a transfer impedance of 4Ω from 30kHz to 50MHz. 
The F-16-1 with the power absorbing terminator has a constant 
transfer impedance of .5Ω from 20kHz to 50MHz. All 
configurations of the F-16 probe can handle 400 amperes of 
rimary power line currents from DC to 60Hz without saturating o
impacting the transfer impedance of the probe. 
0
-30
Tranfer Impedance (dB ohms)
-40
-50 
100Hz 1kHz 10kHz 100kHz 1MH z 10MH z 100MH z
Frequency
F-16-1 Specifications
Frequency:       100 Hz – 70 MHz     Connector: Type-N 
Internal diameter:  32mm  DC to 60 Hz: 400 amperes   
External diameter: 94mm        400 MHz: 400 amperes 
Height:  64mm  RF(CW): 50 amperes 
1
Z
Ω
:   0.5   Pulse2: 300 amperes 
t
1
dBΩ
:   −6    
    2: Depends upon the pulse width and pulse rep. Rate. 
1: Probe calibrated with 50Ω+j0Ω Load Impedance 
F-16-1
Fischer Custom Communications, Inc. 
20603 Earl St., Torrance, CA 90503 · Phone: 310-303-3300 Fax 310-371-6268 · email: Sales@fischerc c.com 
© Fischer Custom Communications, Inc. 
 072803DS-F-16-1 