Specifications
1
(Preliminary)
External Power Meter FOT-932 FOT-932X FOT-933
Detector type Ge GeX InGaAs 
Measurement range (dBm) 10 to -70 26 to -55 6 to –73 
Range displayed (dBm) Down to -90 Down to -75 Down to -90 
Uncertainty
2, 3
± 5 % (± 0.1 nW) ± 5 % (± 3 nW) ± 5 % (± 0.05 nW) 
Wavelength range (nm) 800 to 1650 800 to 1650 800 to 1650 
Display resolution2(dB) 0.01 0.01 0.01 
Calibrated wavelengths 40 40 40
Recommended recalibration 
period (years) 3 3 3
Automatic offset nulling
4
Yes Yes Yes
Sources Standard -4 -5 -12C (second port) -12D (second port)
Wavelengths (nm) 1310 ± 20 1310 ± 20 1310 ± 20 850 ± 25 850 ± 25
1550 ± 20 1550 ± 20 1490 ± 10 1300 +50/–10  1300 +50/–10 
1625 ± 15 1550 ± 20 
Emitter type Laser  Laser Laser LED LED 
Minimum output power5(dBm) -3.5/-3.5 -3.5/-7/-7 -3.5/-7/-7 -26.5/-26.5 (50/125 µm) -24.5/-24.5 (62.5/125 µm) 
Spectral width6(nm) ≤ 5/≤ 5  ≤ 5/≤ 5/≤ 5 ≤ 5/≤ 5/≤ 5 50/135 50/135 
Stability7(8 hours) (dB) ± 0.05 ± 0.05 ± 0.05 ± 0.05 ± 0.05
FasTest Standard -4 -5 -12C (second port) -12D (second port)
Wavelengths (nm) 1310 ± 20 1310 ± 20 1310 ± 20 850 ± 25 850 ± 25
1550 ± 20 1550 ± 20 1490 ± 10 1300 +50/–10  1300 +50/–10 
1625 ± 15 1550 ± 20 
Loss range8(dB) 60 60 60 43 45 
Loss precision9(repeatability) (dB)
Side-by-side 0.15 0.15 0.15 0.25 0.25
Loopback 0.25 0.25 0.25 0.45 0.45 
Length measurement range (km) 200 200 200 5 5 
Length measurement uncertainty
10
± (10 m + 1 % x distance)
Dedicated ORL  All SM wavelengths
ORL range9(APC / UPC) (dB) 65 / 55          
ORL uncertainty11(dB) ± 0.5    
Resolution2(dB) 0.01
General Specifications
Size (H x W x D)  25.0 cm x 12.5 cm x 7.5 cm  (
97/8
in x 4 15/16 in x 3 in) 
Weight 1 kg (2.2 lb) 
Temperature  operating  –10 °C to 50 °C  (14 °F to 122 °F)
storage  –40 °C to 70 °C  (–40 °F to 158 °F) 
Storage  capacity 1024 complete tests 
Relative humidity  0 % to 95 % non-condensing 
Power
9
Li-ion battery 9 hours 
3 hours to fully recharge when unit is off
Warranty (years) 3
FOT-930 MaxTester
Notes:
1. At 23 °C ± 1 °C and 1550 nm with FC/PC connector unless otherwise specified. 
2. Resolution, uncertainty and linearity are functions of input power; uncertainty is 
valid at calibration conditions.
3. Traceable to NIST; up to 20 dBm for GeX.
4. Power of > –45 dBm for Ge, > –30 dBm for GeX and > –57 dBm for InGaAs.
5. In high-power mode.
6. As defined by Telcordia TR-TSY-000887, rms for lasers and at -3 dB for LEDs; 
typical values for LEDs.
7. After a warm-up time of 6 minutes.
8. Typical value, at 1550 nm for SM and 850 nm for MM.
9. Typical value.
10. For fiber length ≤ 120 km.
11. Typical value, for APC connectors.
12. In graded-index MM fibers.
Talk set
Emitter type Laser 
Wavelength (nm) 1550 ± 20  
Dynamic range at 1550 nm (dB) 45 
Dynamic range MM12(dB) 40
VFL
9
Emitter type Laser 
Wavelength (nm) 650 
Output power (dBm) 3
Standard Accessories
User guide, AC adapter/charger, 2 Li-ion batteries, shoulder strap, Certificate of Calibration, connector adapter (FOA) according to chosen connector, mandrel, 
RS-232 cable and software, soft carrying case and alcohol cleaning pads.