The ESA620 Electrical Safety Analyzer, featuring
smart technology to enhance productivity under
any standard, represents the next generation in
portable electrical safety testers. With selections of
three test loads, two protective earth test currents and
two insulation test voltages, this versatile device
performs all primary electrical safety tests as well
as several additional leakage tests for premium
standards compliance worldwide.
A convenient 20 A device receptacle broadens
the range of equipment that can be tested using
the ESA620. Standard 2-wire and optional 4-wire
protective earth measurement capabilities offer
first-rate time savings, while new DSP technology
offers better accuracy of leakage measurements
throughout specified ranges.
Equipped with ten unique safety-enhanced
ECG posts, the ESA620 offers simulation of ECG
and performance waveforms so both electrical
safety and basic tests on patient monitors can
be performed with a single connection. When
combined with optional Ansur computer-based
software, the ESA620 allows for test procedure
automation, the capture of results and comparison
to standard limits, printed reports, and total digital
data management.
Range (accessible voltage)0 V to 300 V ac rms
Accuracy± (2 % of reading +2 LSD)
Voltage testsMains, accessible, and point-to-point
Earth resistance
Two-terminal mode test
current/range and accuracy
Four-terminal mode test
current/range and accuracy
Resistance testsEarth resistance and point-to-point
Equipment current
Mode AC rms
Range/accuracy0 A to 20 A± 5 % of reading ± (2 counts or .2 A,
Duty cycle15 A to 20 A, 5 min on/5 min off
Leakage current
Modes* AC + DC (True-rms)
* Modes are available in all leakage tests with the exception of MAP leakages that
are available only in true-rms
Patient load selection
(input impedance)
Crest factor≤ 3
Ranges0 µA to 199.9 µA
Frequency response/accuracy
Leakage tests Earth (ground wire)
> 200 mA ac
10 A to 25 A ac
> 200 mA ac
10 A to 25 A ac
10 A to 15 A, 7 min on/3 min off
0 A to 10 A continuous
AC only
DC only
AAMI ES1-1993 Fig.1
IEC 60601: Fig 15
IEC 61010: Fig. A-1
200 µA to 1999 µA
2 mA to 10 mA
DC to 1 kHz
1 kHz to 100 kHz
100 kHz to 1 MHz
Chassis (enclosure)
Patient (lead to ground)
Patient auxiliary (lead to lead)
Mains on applied part (lead isolation)
Direct equipment
Direct applied part
Alternative equipment
Alternative applied part
Accessible
Point to point
0.0 to 2.0 Ω ± (2 % of reading + 0.015 Ω)
0.0 to 0.2 Ω ± (2 % of reading + 0.015 Ω)
0.2 to 2.0 Ω ± (5 % of reading + 0.015 Ω)
0.0 to 2.0 Ω ± (2 % of reading + 0.005 Ω)
0.0 to 0.2 Ω ± (2 % of reading + 0.005 Ω)
0.2 to 2.0 Ω ± (5 % of reading + 0.005 Ω)
whichever is greater)
± (1 % of reading + 1 µA)
± (2 % of reading + 1 µA)
± (5 % of reading + 1 µA)
2785725 ESA620 Electrical Safety Analyzer US, 115 V 20 A
30514 0 8 ESA620 Electrical Safety Analyzer EUR, 230 V
3051390 ESA620 Electrical Safety Analyzer FR, 230 V
3051413 ESA620 Electrical Safety Analyzer ISR, 230 V
3051424 ESA620 Electrical Safety Analyzer ITA, 230 V
30514 36 ESA620 Electrical Safety Analyzer AUS, 230 V
30514 49 ESA620 Electrical Safety Analyzer UK, 230 V
30514 51 ESA620 Electrical Safety Analyzer SWI, 230 V
Standard accessories
281496 7 Operator’s Manual CD
2814971 Multilingual Getting Started Guide
2195732 15 A to 20 A Adapter (USA only)
281498 0 Carrying Case
1626219 Data Transfer Cable
Power Cord (country specific)
ESA620 Accessory Kit (country specific)
Optional accessories
311 64 6 3 Ansur ESA620 Plug-In
1903307 Retractable Test Leads
22 42165 Ground Pin Adapter
2067864 Kelvin Cable Set for 4-Wire Measurement
About Fluke Biomedical
Fluke Biomedical is the world’s leading manufacturer of quality biomedical test
and simulation products. In addition, Fluke Biomedical provides the latest medical
imaging and oncology quality-assurance solutions for regulatory compliance.
Highly credentialed and equipped with a NVLAP Lab Code 200566-6 accredited
laboratory, Fluke Biomedical also offers the best in quality and customer service
for all your equipment calibration needs.
Today, biomedical personnel must meet the increasing regulatory pressures,
higher quality standards, and rapid technological growth, while performing their
work faster and more efciently than ever. Fluke Biomedical provides a diverse
range of software and hardware tools to meet today’s challenges.
Fluke Biomedical Regulatory Commitment
As a medical test device manufacturer, we recognize and follow certain quality
standards and certications when developing our products. We are ISO 9001
certied and our products are:
• CE Certied, where required
• NIST Traceable and Calibrated
• UL, CSA, ETL Certied, where required
• NRC Compliant, where required
Fluke Biomedical.
Better products. More choices. One company.
Fluke Biomedical
6045 Cochran Road
Cleveland, OH 44139-3303 U.S.A.
Fluke Biomedical Europe
Science Park Eindhoven 5110
5692EC Son, The Netherlands
For more information, contact us:
In the U.S.A. (800) 850-4608 or
Fax (440) 349-2307
In Europe/M-East/Africa +31 40 267 5435 or
Fax +31 40 267 5436
From other countries +1 (440) 248-9300 or
Fax +1 (440) 349-2307
Email: sales@flukebiomedical.com
Web access: www.flukebiomedical.com