Atec ESA-620 User Manual

ESA620
Electrical Safety Analyzer
The ESA620 Electrical Safety Analyzer, featuring smart technology to enhance productivity under any standard, represents the next generation in portable electrical safety testers. With selections of three test loads, two protective earth test currents and two insulation test voltages, this versatile device performs all primary electrical safety tests as well as several additional leakage tests for premium standards compliance worldwide. A convenient 20 A device receptacle broadens the range of equipment that can be tested using the ESA620. Standard 2-wire and optional 4-wire protective earth measurement capabilities offer first-rate time savings, while new DSP technology offers better accuracy of leakage measurements throughout specified ranges. Equipped with ten unique safety-enhanced ECG posts, the ESA620 offers simulation of ECG and performance waveforms so both electrical safety and basic tests on patient monitors can be performed with a single connection. When combined with optional Ansur computer-based software, the ESA620 allows for test procedure automation, the capture of results and comparison to standard limits, printed reports, and total digital data management.
Technical Data
Key features
Superior compliance with multiple standards:
IEC60601:2005, EN62353, VDE 751, ANSI/AAMI ES1:1993, NFPA-99, AN/NZS 3551, IEC61010
Three test loads
Expanded leakage ranges through 10,000 µA
AC only, dc only and true-rms leakage readings
100 % and 110 % mains voltage for mains on
applied part (lead isolation) test
200 mA and 25 A ac PE test current
DSP filter technology for improved accuracy in
leakage measurements
20 A equipment current
More applied parts selections
ECG and performance waveforms
Intuitive user interface
Easy-to-use applied parts (ECG) connections
Insulation posts on applied parts connections
Five different insulation tests
Varying insulation test voltage
500 V dc and 250 V dc
2- or (optional) 4-wire ground wire resistance
Optional Ansur plug-in software
USB connection
CE, C-TICK and CSA for USA and Canada
RoHS compliance
Designed, tested, and built to incomparable
Fluke quality standards
Specifications
Voltage
Range (mains voltage) 90 V to 132 V ac rms
180 V to 264 V ac rms
Range (accessible voltage) 0 V to 300 V ac rms Accuracy ± (2 % of reading +2 LSD) Voltage tests Mains, accessible, and point-to-point
Earth resistance
Two-terminal mode test current/range and accuracy
Four-terminal mode test current/range and accuracy
Resistance tests Earth resistance and point-to-point
Equipment current
Mode AC rms Range/accuracy 0 A to 20 A ± 5 % of reading ± (2 counts or .2 A,
Duty cycle 15 A to 20 A, 5 min on/5 min off
Leakage current
Modes* AC + DC (True-rms)
* Modes are available in all leakage tests with the exception of MAP leakages that
are available only in true-rms
Patient load selection (input impedance)
Crest factor 3 Ranges 0 µA to 199.9 µA
Frequency response/accuracy
Leakage tests Earth (ground wire)
> 200 mA ac 10 A to 25 A ac
> 200 mA ac 10 A to 25 A ac
10 A to 15 A, 7 min on/3 min off 0 A to 10 A continuous
AC only DC only
AAMI ES1-1993 Fig.1 IEC 60601: Fig 15
IEC 61010: Fig. A-1
200 µA to 1999 µA 2 mA to 10 mA
DC to 1 kHz 1 kHz to 100 kHz 100 kHz to 1 MHz
Chassis (enclosure) Patient (lead to ground) Patient auxiliary (lead to lead) Mains on applied part (lead isolation) Direct equipment Direct applied part Alternative equipment Alternative applied part Accessible Point to point
0.0 to 2.0 Ω ± (2 % of reading + 0.015 Ω)
0.0 to 0.2 Ω ± (2 % of reading + 0.015 Ω)
0.2 to 2.0 Ω ± (5 % of reading + 0.015 Ω)
0.0 to 2.0 Ω ± (2 % of reading + 0.005 Ω)
0.0 to 0.2 Ω ± (2 % of reading + 0.005 Ω)
0.2 to 2.0 Ω ± (5 % of reading + 0.005 Ω)
whichever is greater)
± (1 % of reading + 1 µA) ± (2 % of reading + 1 µA) ± (5 % of reading + 1 µA)
2 Fluke Biomedical ESA620 Electrical Safety Analyzer
Mains on applied part test voltage
110 % of mains at 230 V for IEC 60601 100 % of mains for AAMI at 115 V per AAMI 100 % of mains at 230 V per 62353
Differential leakage
Ranges 10 µA to 199 µA
200 µA to 1999 µA 2 mA to 20 mA
Accuracy ± 10 % of reading ± (2 counts or 20 µA, whichever is greater)
Insulation resistance
Ranges/accuracy
Source test voltage 500 V dc
Insulation resistance tests Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible
Mamimum load capacitance 1 µF
0.5 MΩ to 20 MΩ
20 MΩ to 100 MΩ
250 V dc
conductive part) and AP- NE (non-earthed accessible conductive part)
± (2 % of reading + 2 counts)
± (7.5 % of reading + 2 counts)
ECG performance waveforms
Accuracy ± 2 %
± 5 % for amplitude of 2 Hz square wave only, fixed at 1 mV Lead II configuration
Waveforms
Rates
ECG complex (BPM) 30, 60, 120, 180, and 240
Ventricular fibrillation
Square wave (50 % duty cycle) (Hz) 0.125 and 2 Sine wave (Hz) 10, 40, 50, 60, and 100
Triangle wave (Hz) 2 Pulse (63 ms pulse width) 30 and 60
Power ratings
Mains voltage outlet 120 V ac 230 V ac Mains voltage inlet power range 90 to 132 V ac rms 180 to 264 V ac rms
Maximum current 20 A 16 A Hz 50 or 60 50 or 60
Physical case
Dimensions (L x W x H) 31 cm x 23 cm x 10 cm (12.2 in x 9 in x 2.9 in) Weight 4.7 kg (10.25 lb)
Environmental specifications
Operating temperature 10 °C to 40 °C Storage temperature -20 °C to 60 °C Operating humidity 10 % to 90 % non-condensing Altitude To 2,000 meters
General
Warranty Two-year extended warranty*
*No-cost extended warranty available after first-year calibration at any Fluke Biomedical authorized service center.
ESA620 Electrical Safety Analyzer Fluke Biomedical 3
Ordering information
Models
2785725 ESA620 Electrical Safety Analyzer US, 115 V 20 A 30514 0 8 ESA620 Electrical Safety Analyzer EUR, 230 V 3051390 ESA620 Electrical Safety Analyzer FR, 230 V 3051413 ESA620 Electrical Safety Analyzer ISR, 230 V 3051424 ESA620 Electrical Safety Analyzer ITA, 230 V 30514 36 ESA620 Electrical Safety Analyzer AUS, 230 V 30514 49 ESA620 Electrical Safety Analyzer UK, 230 V 30514 51 ESA620 Electrical Safety Analyzer SWI, 230 V
Standard accessories
281496 7 Operator’s Manual CD 2814971 Multilingual Getting Started Guide 2195732 15 A to 20 A Adapter (USA only) 281498 0 Carrying Case 1626219 Data Transfer Cable Power Cord (country specific) ESA620 Accessory Kit (country specific)
Optional accessories
311 64 6 3 Ansur ESA620 Plug-In 1903307 Retractable Test Leads 22 42165 Ground Pin Adapter 2067864 Kelvin Cable Set for 4-Wire Measurement
About Fluke Biomedical
Fluke Biomedical is the world’s leading manufacturer of quality biomedical test and simulation products. In addition, Fluke Biomedical provides the latest medical imaging and oncology quality-assurance solutions for regulatory compliance. Highly credentialed and equipped with a NVLAP Lab Code 200566-6 accredited laboratory, Fluke Biomedical also offers the best in quality and customer service for all your equipment calibration needs.
Today, biomedical personnel must meet the increasing regulatory pressures, higher quality standards, and rapid technological growth, while performing their work faster and more efciently than ever. Fluke Biomedical provides a diverse range of software and hardware tools to meet today’s challenges.
Fluke Biomedical Regulatory Commitment
As a medical test device manufacturer, we recognize and follow certain quality standards and certications when developing our products. We are ISO 9001 certied and our products are:
•  CE Certied, where required
•  NIST Traceable and Calibrated
•  UL, CSA, ETL Certied, where required
•  NRC Compliant, where required
Fluke Biomedical.
Better products. More choices. One company.
Fluke Biomedical
6045 Cochran Road Cleveland, OH 44139-3303 U.S.A.
Fluke Biomedical Europe
Science Park Eindhoven 5110 5692EC Son, The Netherlands