The EPOCH LTC digital ultrasonic flaw
detector offers state-of-the-art technology in
a compact and lightweight package. Based
on the class-leading EPOCH XT, the
EPOCH LTC is full featured, complies with
the EN 12668-1 standard, and has various
specialized options to meet your inspection
needs.
Ultrasonic flaw detection is widely used in
locating and sizing hidden cracks, voids,
disbonds, and similar discontinuities in
welds, forgings, billets, axles, shafts, tanks
and pressure vessels, turbines, and
structural components.
Part of the EPOCH
flaw detector family
The EPOCH LTC features a bright, multicolor transflective, full-VGA (640 pixels x
480 pixels) display offering high-resolution
A-scans and superior visibility in all lighting
conditions, including direct sunlight. The
EPOCH LTC is also designed for IP67
rating, ensuring that the instrument can
perform in the harsh environments common in the inspection field.
Features
■ Designed for IP67 rating to ensure field
ruggedness
■ Complies with the EN12668-1 standard
■ Only 960 g (2.1 lb)
■ Full-color VGA display with transflective
technology
■ USB On-The-Go port for computer
communication and direct printing
■ Onboard and miniSD Card storage
capabilities for the data logger
■ Standard dynamic DAC/TVG
■ Standard onboard DGS/AVG
■ Digital high dynamic range receiver with
digital filters
■ High-performance square wave pulser
with tuning option
920-121G-EN
Powerful Performance in a Compact Flaw Detector
The EPOCH LTC offers many standard features to meet your inspection needs. These include a fixed width square wave pulser, dynamic
DAC/TVG, onboard DGS/AVG, selectable digital filters, gain range from 0 dB to 110 dB, peak memory, 0.01 mm (0.001 in.) measurement resolution, and one gate with programmable alarms. In addition, the EPOCH LTC offers many optional, application-specific software features such as AWS D1.1/D1.5 weld rating, manual PRF control, pulse width tuning, and the GageView Pro interface program.
• Standard fixed widthc square wave
pulser with PerfectSquare™ technology
• Digital receiver filtering: standard,
broadband, and high-pass settings
• Four, fully customizable measurement
displays to meet inspection needs.
Select any Gate measurement for each
display box
• Amplitude measurement resolution of
0.25% full screen height
• Auto-calibration for thickness,
soundpath, or angular depth
• Amplitude measurement from 0% to
110% full screen height
• PRF rate automatically controlled from
10 Hz to 500 Hz. Manual control is
optional
• Grid display modes
– Standard 1 to 10
– Soundpath
– Leg mode for angle beam inspection
– Selectable 100% or 110% vertical
display
Built for Tough Environments
The EPOCH LTC can withstand the rigors of field operation in
harsh environments, from offshore to deserts.
Environmental Ratings
IP rating
Explosive atmosphere
Shock tested
Vibration tested
Drop tested
Operating temperature Lithium-ion: –10 °C to 50 °C (14 °F to 122 °F)
Battery storage
temperature
Recharge temperature
range
Battery compartment
I/O door
Simplified Keypad Design
The EPOCH LTC features a simple keypad
design that allows full control with the left
or right hand for direct access to important
functions. The keypad design also makes
for rapid and easy parameter adjustments
and recall of instrument configuration
(CAL) files.
The keypad is available in English, international symbols, Japanese, and Chinese.
Four custom function keys allow operators to select preset values for instrument
configuration parameters.
Designed to meet the standards of the Ingress
Protection (IP) rating number IP67
Safe operation as defined by Class I,
Division 2, Group D, as defined in the National Fire Protection Association Code (NFPA
70), Article 500, and tested using MIL-STD810F, Method 511.4, Procedure I.
IEC 60068-2-27; 60 g, 6 ms H.S., 3 axes,
18 shocks total
Sine vibration as per IEC 60068-2-6; 50 Hz
to 150 Hz at 0.03 in. DA or 2 g, 20-sweep
cycles
MIL-STD-810F, 4.5.5 Procedure IV–Transit
drop. 48 in. drop onto 2 in. plywood surface
backed by concrete. Drop on each face, edge,
and corner for a total of 26 drops
0 °C to 50 °C (32 °F to 122 °F)
0 °C to 40 °C (32 °F to 104 °F)
Sealed – no tools necesssary to replace
battery
Sealed I/O door for USB On-The-Go
connection, miniSD Card slot, and
VGA/RS-232 port
Hand strap can be moved for use
with left or right hand
2
www.olympus-ims.com
Advanced Software Features and Data Management
Standard dynamic DAC/TVG: Calculates
signal amplitude as a percentage or decibel
level compared to a DAC curve or a reference echo amplitude fixed at a time-varied
gain. DAC versions include ASME, ASME
3, JIS, and Custom. Also includes several
key features: dynamically adjustable DAC
curves, switchable DAC and TVG views,
and custom DAC warning curves.
Standard onboard DGS/AVG: Flaw sizing
technique allowing echo signals to be
evaluated with a DGS/AVG diagram associated to a particular type of probe and
material. The DGS/AVG diagram illustrates
the relationships between echo height, flaw
size, and distance from the transducer
Corrects sound path information when using
an angle beam transducer to circumferentially inspect a curved surface.
Software Options
Manual PRF control: Allows operators to
manually adjust the EPOCH LTC’s pulse
repetition frequency (PRF) from 10 Hz to
500 Hz in 10-Hz increments
Extended range: Extends the standard range
of the EPOCH LTC to 3.36 mm to 13,404
mm (0.132 in. to 527 in.) at 5900 m/s
(0.2320 in./ms) Includes Low Pass filter
(0.5 MHz to 4.0 MHz).
Tunable square wave pulser: Allows
operators to tune the pulse width of the
square wave pulser to optimize transducer
performance. This maximizes signal-tonoise ratio and penetration in difficult
materials.
AWS D1.1 and D1.5: Provides a dynamic
reflector indication rating for various AWS
weld inspection applications. This allows
more efficient inspections by eliminating
manual calculations .
API 5UE: Allows defect sizing according to
API Recommended Practice 5UE. Uses the
Amplitude Distance Differential Technique
(ADDT) to measure the size of potential
defects during the prove-up process of
OCTG pipe. The measurement process
is simple and repeatable since all ADDT
variables are captured from a Peak Memory
envelope.
Instrument Inputs/Outputs
USBUSB On-The-Go port for PC communication, printing, and data storage.
VGA ouput port Standard
miniSDminiSD Card slot for data storage.
RS-232 portYes
Capable of storing up to 50,000 IDs with waveforms, measurements, and
Data Storage
configuration parameters in the onboard memory. The miniSD Card allows
virtually unlimited data storage.
GageView Pro
The optional GageView
program helps manage and format stored
inspection data. Data can be printed or
easily copied to word processing files and
spreadsheets for further reporting needs.
GageView Pro also allows creating customized databases of identifier (ID) strings that
can be uploaded to the EPOCH LTC. The
program is fully compatible with the
Software option API 5UE allows defect sizing according to API
Recommended Practice 5UE
EPOCH LT, XT, 600 and the 1000 Series.
With GageView Pro operators can:
• Export thickness or amplitude data
saved on the EPOCH to Microsoft Excel,
Word, or similar programs
• Create, format, and manage test
databases
• Import and export configurations
between the EPOCH and a computer
• Create custom DGS probe libraries
• Upgrade the EPOCH operating software
TM
Pro interface
3
EPOCH LTC Specifications*
General
Dimensions (W x H x D) 128.9 mm x 223.3 mm x 55.1 mm (5.07 in. x 8.79 in. x 2.17 in.)
Weight960 g (2.12 lb) with Li-ion battery
KeypadEnglish, international symbols, Japanese, or Chinese
Display languages
Transducer connectionLEMO 00
BatteryRechargeable Lithium-ion. Standard AA battery tray
Battery operating time
Power compatibility100/120 VAC, 200/240 VAC, 50-60 Hz
Display type
Display dimensions (W x H)640 pixels x 480 pixels, 75 mm x 56 mm (2.95 in. x 2.21 in.), 94 mm (3.7 in.) diagonal
English, Spanish, French, German, Italian, Japanese, Chinese, Russian, Norwegian,
Swedish
Lithium-ion rechargeable battery: 8 hours; AA Li-FeS2: 6 hours
AA NiMH cells: 4 hours; AA alkaline cells: 3 hours
Color transflective LCD, refresh rate of 60 Hz, user-selectable color schemes and
brightness, two split-screen modes and one full-screen mode
Pulser
PulserFixed width square wave pulser optimized for general flaw detection applications
Pulse width
Pulse repetition frequency
(PRF)
Energy settings100 V to 400 V in 100 V increments
Damping50 Ω, 100 Ω, 200 Ω, 400 Ω
Fixed width in base instrument. Adjustable from 25 ns to 5,000 ns (0.1 MHz) with
EPLTC-SWP option
Automatically software controlled over the range of 10 Hz to 500 Hz.
Manual PRF control option available
Receiver
Gain 0 dB to 110 dB — Two user-defined gain step adjustments and presets
Total instrument bandwidth 0.2 MHz to 26.5 MHz at –3 dB