Atec EMR-20C, EMR-21C User Manual

RF Safety
Test Solutions
50 Years of
Excellence
Electric and Magnetic Field Measurement
EMR-20C and EMR-21C
E-Field Measurement Systems
100 kHz to 3 GHz
0.6 to 800 V/m (-20C)
0.2 to 320 V/m (-21C)
Completely Automatic Zeroing
Easy-to-Use Design
Lightweight and Rugged
Cost Effective
Isotropic Detection
24 Month Calibration Interval
Description
The EMR-20C and –21C are complete systems that con­sist of a meter and probe, along with a charging system for the NiCad batteries.
The EMR-20C and EMR-21C are supplied with a cali­brated Type 8 probe that provides coverage for many in­dustrial systems operating at 915 and 2450 MHz, as well as measuring extremely low field levels. Users appreciate the wide measurement range and extreme ruggedness of this multi-purpose survey system. The EMR-20C and -21C have even been designed to withstand drops from two meters directly on the sensor head!
Operation of this series of survey systems could not be any easier. All you need to do is connect the probe and turn on the meter. The system automatically detects the probe and auto-zero’s the system – even while immersed in an RF field environment. Electric field readings are dis­played in units of V/m, W/m of measurement range, without touching a button!
Both the EMR-20C and -21C feature a bi-directional fiber-optic link. With the optional software and cables, real-time readings can be displayed on a personal computer.
2
and mW/cm2, with over 62 dB
USA TEL: (1) 631 231-1700 • FAX: (1) 631 231-1711 • E-MAIL: NardaSTS@L-3COM.com • www.narda-sts.com
28
GERMANY TEL: 49-7121-9732-777 • FAX: 49-7121-9732-790 • E-MAIL: support@narda-sts.de • www.narda-sts.de
narda
Safety Test Solutions
an
RF Safety
Test Solutions
Electric and Magnetic Field Measurement
Specifications
EMR-20C EMR-21C
Frequency Range 100 kHz to 3 GHz
Field Measured E-Field Only, Isotropic Detection, Diode Sensor
0.6 to 800 V/m
Measurement Range
(CW signals > 300 kHz)
0.6 to 20 V/m (True RMS)
Measurement Resolution 0.01 V/m
Overload Levels
700 mW/cm
70 W/cm
2
2
(Peak)
(CW)
Absolute Error ±1 dB (27.5 V/m and 27.12 MHz)
0.6 to 1.25 V/m = ±3dB
Linearity Error
a
1.25 to 2.5 V/m = ±1 dB
2.5 to 400 V/m = ±0.5 dB
400 to 800 V/m = ±0.7 dB
Frequency Response Error
b
Isotropic Deviation (Typical)
100 kHz to 100 MHz = ±0.45 dB
100 MHz to 3 GHz = ±1.4 dB
Field probe only = ±0.5 dB for Freq.>1MHz
Complete system = ±1.0 dB for Freq.>1MHz
Temperature Error + 0.2 / - 1.0 dB (0 to +50°C)
100 kHz to 5 MHz = >35 dB – 20 dB x log f
H-field Immunity
(f in MHz)
Above 5 MHz=>20dB(Typical)
Settling Time 1 second (0 to 90% of measured value) Typical
Display Refresh Rate 400 msec. Typical
Units Displayed V/m, A/m, mW/cm2, W/m2and “% of limit value”
Levels Displayed Current reading, maximum reading or average reading
Audible Alarm Variable setting, piezoelectric
Interfaces Fiber Optic, serial interface for results transfer, remote operation and calibration
Power Supply Rechargeable AA batteries (supplied)
Operating Temperature 0 to +50°C
Size (HxWxD,approx.) 18.3" x 3.78" x 2.52" (465 x 96 x 64 mm)
Weight (approx.) 450 grams
0.2 to 320 V/m (CW signals > 300 kHz)
0.2 to 10 V/m (True RMS)
175 mW/cm
17.5 W/cm
2
(CW)
2
(Peak)
0.6 to 1.2 V/m = ±1dB
1.2 to 200 V/m = ±0.5 dB
200 to 320 V/m = ±0.7 dB
>20 dB Typical
50 Years of
Excellence
Notes:
a
Referenced to 27.5 V/m and 27.12 MHz
b
Error after using supplied correction factors, includes calibration accuracy
narda
Safety Test Solutions
an
USA TEL: (1) 631 231-1700 • FAX: (1) 631 231-1711 • E-MAIL: NardaSTS@L-3COM.com • www.narda-sts.com
GERMANY TEL: 49-7121-9732-777 • FAX: 49-7121-9732-790 • E-MAIL: support@narda-sts.de • www.narda-sts.de