Electric and Magnetic Field Measurement
EMR-20C and EMR-21C
E-Field Measurement Systems
100 kHz to 3 GHz
◆
0.6 to 800 V/m (-20C)
◆
0.2 to 320 V/m (-21C)
◆
Completely Automatic Zeroing
◆
Easy-to-Use Design
◆
Lightweight and Rugged
◆
Cost Effective
◆
Isotropic Detection
◆
24 Month Calibration Interval
◆
Description
The EMR-20C and –21C are complete systems that consist of a meter and probe, along with a charging system for
the NiCad batteries.
The EMR-20C and EMR-21C are supplied with a calibrated Type 8 probe that provides coverage for many industrial systems operating at 915 and 2450 MHz, as well
as measuring extremely low field levels. Users appreciate
the wide measurement range and extreme ruggedness of
this multi-purpose survey system. The EMR-20C and -21C
have even been designed to withstand drops from two
meters directly on the sensor head!
Operation of this series of survey systems could not be
any easier. All you need to do is connect the probe and
turn on the meter. The system automatically detects the
probe and auto-zero’s the system – even while immersed
in an RF field environment. Electric field readings are displayed in units of V/m, W/m
of measurement range, without touching a button!
Both the EMR-20C and -21C feature a bi-directional
fiber-optic link. With the optional software and cables,
real-time readings can be displayed on a personal
computer.
2
and mW/cm2, with over 62 dB
USA TEL: (1) 631 231-1700 • FAX: (1) 631 231-1711 • E-MAIL: NardaSTS@L-3COM.com • www.narda-sts.com
28
GERMANY TEL: 49-7121-9732-777 • FAX: 49-7121-9732-790 • E-MAIL: support@narda-sts.de • www.narda-sts.de
narda
Safety Test Solutions
communications company
an
RF Safety
Test Solutions
Electric and Magnetic Field Measurement
Specifications
EMR-20C EMR-21C
Frequency Range 100 kHz to 3 GHz
Field Measured E-Field Only, Isotropic Detection, Diode Sensor
0.6 to 800 V/m
Measurement Range
(CW signals > 300 kHz)
0.6 to 20 V/m (True RMS)
Measurement Resolution 0.01 V/m
Overload Levels
700 mW/cm
70 W/cm
2
2
(Peak)
(CW)
Absolute Error ±1 dB (27.5 V/m and 27.12 MHz)
0.6 to 1.25 V/m = ±3dB
Linearity Error
a
1.25 to 2.5 V/m = ±1 dB
2.5 to 400 V/m = ±0.5 dB
400 to 800 V/m = ±0.7 dB
Frequency
Response Error
b
Isotropic Deviation
(Typical)
100 kHz to 100 MHz = ±0.45 dB
100 MHz to 3 GHz = ±1.4 dB
Field probe only = ±0.5 dB for Freq.>1MHz
Complete system = ±1.0 dB for Freq.>1MHz
Temperature Error + 0.2 / - 1.0 dB (0 to +50°C)
100 kHz to 5 MHz = >35 dB – 20 dB x log f
H-field Immunity
(f in MHz)
Above 5 MHz=>20dB(Typical)
Settling Time 1 second (0 to 90% of measured value) Typical
Display Refresh Rate 400 msec. Typical
Units Displayed V/m, A/m, mW/cm2, W/m2and “% of limit value”
Levels Displayed Current reading, maximum reading or average reading
Audible Alarm Variable setting, piezoelectric
Interfaces Fiber Optic, serial interface for results transfer, remote operation and calibration
Power Supply Rechargeable AA batteries (supplied)
Operating Temperature 0 to +50°C
Size (HxWxD,approx.) 18.3" x 3.78" x 2.52" (465 x 96 x 64 mm)
Weight (approx.) 450 grams
0.2 to 320 V/m (CW signals > 300 kHz)
0.2 to 10 V/m (True RMS)
175 mW/cm
17.5 W/cm
2
(CW)
2
(Peak)
0.6 to 1.2 V/m = ±1dB
1.2 to 200 V/m = ±0.5 dB
200 to 320 V/m = ±0.7 dB
>20 dB Typical
50 Years of
Excellence
Notes:
a
Referenced to 27.5 V/m and 27.12 MHz
b
Error after using supplied correction factors, includes calibration accuracy