NEMP radiated susceptibility test setup
This test set-up can be used for the susceptibility testing of small pieces of equipment such as printed boards, small subsystems, etc., according to MIL-STD-461 / RS105 (NEMP: nuclear electromagnetic pulse). Associated with the generator EMP25K-2-23, the minimum field under the line RL50-50 is 50 kV/m at full output. Even higher fields can be achieved for smaller objects placed close to the beginning of the line. The line can also be used for other pulse or CW tests.
SPECIFICATIONS - Generator
Type EMP25K-2-23 Standard Peak voltage Impedance Rise time (10 – 90 %) Pulse length on 100 ohm (50 – 50 %) Output connector Power rating Dimensions Weight
MIL-STD461-E / F, paragraph RS105
typically ≤ +0.2 kV to ≥ +25 kV
100 ohm
2.3 ns ± 0.5 ns 23 ns ± 5 ns HVM50K 85 - 264 Vac/ 50 - 60 Hz / 150 VA 56 x 19 x 45 cm (L x H x W) 15 kg
SPECIFICATIONS - Line
Type RL50-50 Structure Impedance
bounded wave line / TEM mode 100 ohm
Maximum height of the EUT Input connector Dimensions Weight
montena technology - 1728 Rossens - Switzerland - phone + 41 26 411 84 84 - fax + 41 26 411 17 79
www.montena.com e-mail: products@montena.com
17 cm (according to MIL-STD461 / RS105) HVM50K 200 x 55 x 80 cm (L x H x W) 25 kg