Atec EMCPro-PLUS User Manual

KeyTek EMCPro®PLUS
Advanced EMC test system for compliance testing to 6 IEC/EN standards
Technical Specifications
The newly configured KeyTek EMCPro®PLUS test system features resident capabilities for EMC CE Mark compliance testing to 6 IEC/EN standards, and fully addresses new requirements for a 100 kHz burst rate per IEC 61000-4-4, Edition 2 (EFT) and 80% dip per IEC 61000-4-11, Edition 2 (PQF™).
Portable and low cost, the KeyTek EMCPro PLUS is the answer to manufacturers’ demand for a mid-range, multi-capability EMC immunity tester. It’s ideal for companies who require flexibility, versatility, and the highest test level-to-cost ratio instrument on the market.
Portable, mid-range EMC test system
Resident capabilities for compliance testing to 6 IEC/EN standards
Addresses ANSI/IEEE, ITU, ETSI & UL standards
Surge testing to 6.6kV with the combination, telecom, & ring waves
Monitors surge voltage & current at the output terminals
Monitors output of the coupling unit & automatically switches connections according to coupling mode
Highest test levels, widest selection of tests & lowest in-use costs
Technical Specifications Model PRO-BASE
EMCPro PLUS Base Unit
System Voltage 90-240VAC, 50/60Hz INTEGRATED EUT MAINS
COUPLER/DECOUPLER AC Voltage 1 phase, 50 -
250VAC. 50/60Hz
AC Current 16A max.**
DC Voltage 100VDC max. DC Current: 10A max. Frequency 50/60Hz EUT Connectors Nema, British,
Schuko
CONTROL INTERFACE Interface RS232 Fiber-optic SAFETY FEATURES External Interlock for
users Interlock for CCL connector External stop input
ENVIRONMENTAL OPERATING CONDITIONS Temperature 15°- 40°C Humidity 10-75%, non-
condensing
Altitude 8000 ft. max. PHYSICAL Height 22.9cm (8.7 in) Width 43.4cm (17.1 in) Depth 64.8cm (25.5 in) Weight 39kg (85 lbs.) CE MARKING Safety and EMC
Directives
Upgradable as standards change
Analyze Detect Measure Control
TM
Model PRO-ESD
ESD per IEC 61000-4-2 and EN 61000-4-2
Trigger Modes One shot manual, multi-shot tripod Repetition Rate Single shot, 1pps or 20pps Air Discharge Voltage 500V - 8.8kV ±10% Contact Discharge Voltage 500V - 4.4kV ±10% Discharge Capacitor 150pF ±10% Discharge Resistance 330±10% Charging Resistance 50M- 100M Polarity Front panel or software controlled Shot Counter 1 - 999 discharges Energy Storage 5.8mJ @ 8.8kV
Model PRO-EFT
EFT per IEC 61000-4-4 Edition 2, EN 61000-4-4 and ANSI C62.41
Voltage Waveform 5/50ns ±30% Peak Voltage 250V - 4.4kV ±5% Burst Period 300ms ±10% Burst Duration 15ms ±20%, for pulse frequencies uo to 5kHz, 0.75ms
above 5kHz
Frequency 1-100kHz, in 0.5kHz steps, ±10% DC Blocking Capacitor 10nF (internal) Options Model CM-3CD-16/32: 16 or 32 Amp, 3 phase EFT &
surge coupler/decoupler
Model CM-CCL: Capacitive coupling clamp Model CM-CCLC: Coupling clamp cover Model EFT-ATTN: EFT attenuator for oscilloscope
monitoring
Model PRO-SURGE
Surge for compliant testing per IEC 61000-4-5, EN 61000-4-5, ANSI C62.41 Category B and UL 1449
Voltage Waveform 1.2/50µs Peak Voltage 250V - 6.6kV ±5%, 12mode
250V - 6.0kV ±5%, 2mode
Peak Current 125A - 3.3kA ±10% Additional 10Resistor Software selectable Repetition Rate Up to 4 per minute Open-circuit Voltage Front time: 1.2µs ±30%
Duration: 50 µs ±20% Undershoot: 30%
Short-circuit Current Front time: 8.0µs ±20%
*
Duration
Undershoot 30% Line sync accuracy ±15%, 50 - 277VAC Options Model CM-3CD-16/32: 16 or 32 Amp, 3 phase EFT
*
Durations are reduced in 12mode and when coupling multiple lines to PE
50µs ±20%
& surge coupler/decoupler Model CM-I/OCD: External 8 line coupler/decoupler for I/O signal lines Model CM-I/OCD-HS: High speed I/OCD option for testing data rates to >100kHz
1
Model PRO-RING
Ring Wave Surge per ANSI C62.41 Cat. A, B, and UL 864
Voltage Waveform 100kHz damped cosine Peak Voltage 250 - 6.6kV ±5% Repetition Rate <4/minute at 6kV, faster at lower voltages Open-circuit Voltage Rise Time: 0.5µs ±30% Short-circuit Current Vp/Ip: 12±3or 30±8software selectable Options Model CM-3CD-16/32: 16 or 32 Amp, 3 phase EFT
Model PRO-TELECOM
Surge Telecom compliant testing per IEC 61000-4-5, EN 61000-4-5, FCC Part 68, ITU K.17, K.20, K.21 and ETSI
Voltage Waveform 10/700µs (9/720µs FCC Part 68) Peak Voltage 250V - 6.6kV ±5% Peak Current 6.25 - 165A +10/-0%, 40mode Repetition Rate Up to 4 per minute Open-circuit Voltage Front time: 7.0µs to 11.7µs
Short-circuit Current Front time: 3.5µs to 6.5µs
Options Model CM-TELCD: External coupler for telecom lines
**
& surge coupler/decoupler
**
Duration: 576µs to 840µs
Duration: 256µs to 384µs
Surge Waveform Monitoring
Lines Monitored Monitors are automatically switched to match
generator coupling mode
Open-circuit Voltage 1000:1 ±10% Short-circuit 200:1 ±7%
Current Attenuation
Model PRO-HPOWER
Power Frequency Magnetic Field for compliant testing per IEC 61000-4-8 and EN 61000-4-8
Field Frequency 50Hz/60Hz Field Amplitude 0.5 - 4A/m, in 0.25A steps, ±10% (with CM-HCOIL)
up to 100A/m with optional external HPOWER-EXT
AC Source Internal Resolution 0.25A minimum Coil Factor 0.65 to 1.00 Coil Resistance 0.05maximum Options Model CM-HMON: Measurement probe for power
frequency magnetic fields
Model CM-HCOIL: 1m x1m magnetic field coil Model HPOWER - EXT: External generator for power
frequency magnetic field to 30A/m
**
PRO-TELECOM and PRO-RING can not be installed in same unit.