Ardo Master 45, Senator 30 Service manual

Suction Pumps
MASTER 45
SENATOR 30
Ardo medical AG, Gewerbestrasse 19, CH-6314 Unterägeri/Switzerland
Tel. +41-(0)41 754 70 70 - Fax +41-(0)41 754 70 71 - info@ardo.ch - www.ardo.ch
11/2005
TABLE OF CONTENTS
8. INTRODUCTION Safety
Responsibility Guarantee Service
9. TECHNICAL DESCRIPTION AND DIAGRAMS
Mechanical Layout Electrical Layout Wiring Diagram/Block Diagram
10. MAINTENANCE
General Maintenance Maintenance Works for Technical Personnel
11. SERVICING AND REPAIRS
Troubleshooting Accessories/Spare Parts Replacing Spare Parts Cylinder Belt Drive Wearing Parts
12. TECHNICAL INFORMATION
Specifications Transport and Storage Disposal after Service Life
Please also follow the enclosed
OPERATING INSTRUCTIONS (chapter 1-7)!
COPYRIGHT © 2005 Ardo medical AG
3 4 4 5
5 7 8
9 9
10 11 17 17 18 18
18 19 19
Ardo medical AG 2 Service Manual MASTER/SENATOR
Subject to change
8. INTRODUCTION
Safety
The suction pumps are designed and constructed according to the safety regulations of IEC 601-1, the EMC standard IEC 60601-1-2 and EN 10079-1. For the user's and the patient’s safety as well as to prevent equipment damage, please observe these safety rules:
Safety Rules
The suction pumps MASTER/SENATOR may only be installed and operated by competent personnel.
Use the suction pumps with the ARDO hydrophobic bacterial filter only.
To ensure the proper venting of the pumps, the feet must be installed on the bottom
of the pumps if used without trolley.
Never start operating the suction pump when the housing is open (rotating parts – hazard of injury and electric shock).
Capacitors may be charged even if device is without power. Discharge capacitors with discharge set!
Check contact points of the housing for tears. If there are any tears, exchange housing.
See that MASTER/SENATOR are set into operation according to Chapter 4 - Installation -
in the operating instructions.
Store the service manual in a safe place, so that it is available to the trained technical personnel at all times.
General Safety Regulations
Never use MASTER/SENATOR in the presence of ignitable gases as anesthetic agents. Explosion hazard!
The suction pumps may only be opened by qualified technical personnel. Electric shock hazard!
Always switch the suction pump off and disconnect from mains power before cleaning and opening. Electric shock hazard!
Do not use solvents for cleaning. Otherwise the surface of the housing or the synthetic material parts may be damaged.
Ardo medical AG 3 Service Manual MASTER/SENATOR
Subject to change
Responsibility
ARDO is only responsible for effects on safety, reliability and performance of the device, if:
Installation, new adjustments, alterations or repairs and maintenance works are carried out by authorized persons.
For repairs and replacements of spare parts use original parts by ARDO only.
The electric installations of the respective room meet the requirements of the standards
IEC 601 and MPG (German Medicinal Products Act), and if the devices are maintained and repaired according to this standard and to EN 10079-1.
The pump is used in accordance with the operating instructions.
Guarantee
The suction pumps MASTER/SENATOR have a guarantee period of 2 years.
General Conditions
Maintenance
Ardo medical AG warrants for material and production faults of the manufactured products. The guarantee period can be taken from the respective operating instructions. Faulty material is replaced free of charge within the guarantee period, assumed the device was not used improperly. Wearing parts are excluded. For ensuring guarantee and faultless functioning of the device, the notes of the operating instructions have to be observed. Furthermore, only original spare parts by Ardo medical AG may be installed and/or used.
The right to guarantee is waived, when non-authorized persons interfere with the device or alterations not conforming to the standard IEC 601 and EN 10079-1 are carried out. There is no right to guarantee exceeding the described scope of guarantee, as liability to consequential damages etc.
Read operating instructions Chapter 5 - Maintenance and Care, Cleaning -.
Ardo medical AG 4 Service Manual MASTER/SENATOR
Subject to change
Service
Please contact to the following addresses for service, maintenance or any questions about this product or any other product of Ardo medical AG:
Switzerland
Germany
International
Ardo medical AG Gewerbestrasse 19 CH-6314 Unterägeri Switzerland Tel. +41-(0)41 754 70 70 Fax +41-(0)41 754 70 71 info@ardo.ch www.ardo.ch
Ardo medical GmbH Argelsrieder Feld 10 D-82234 Oberpfaffenhofen Germany Tel. +49-(0)8153 / 40 66 00 Fax +49-(0)8153 / 40 66 01 info@ardomedical.de www.ardomedical.de
Authorized Importer
9. TECHNICAL DESCRIPTION AND DIAGRAMS
Mechanical Layout
The ARDO suction pumps MASTER/SENATOR work according to the principle of a piston/ cylinder system. Both power units are based on a modular driving concept, equipped with 2 cylinder units.
MASTER 45 2 cylinder power unit (approximately 300 rpm) SENATOR 30 2 cylinder power unit (approximately 220 rpm)
A two-stage belt transmission was developed for driving the cylinder units that may be used for both power units (MASTER and SENATOR).
Transmission reduction for the two power units can be adjusted with the two belt pulleys of the first reduction level (motor pinion + belt wheel = 1. level).
Ardo medical AG 5 Service Manual MASTER/SENATOR
Subject to change
For both power units a bipolar capacitor motor with built-in temperature sensor is used. This protects the motor coil from heating to more than 125°C.
The piston is driven over a linear unit (slide). By using this system, the piston system is exposed to only minimal thrust forces, which ensures low wear and high service life. In MASTER/SENATOR the pistons work in angles of 90° to each other. This ensures a well-balanced run of the pump with low vibration levels.
1 Cylinder unit 6 Vacuum-meter 2 Discharge valve 7 Vacuum regulator 3 Inlet valve 8 Vacuum connector 4 AC motor 9 Exhaust 5 Linear drive 10 2-stage belt transmission
Ardo medical AG 6 Service Manual MASTER/SENATOR
Subject to change
Electrical Layout
The driving motor is a bipolar capacitor motor and may be used for both power units (MASTER and SENATOR).
MASTER is equipped with a electronically controlled start-up circuit (Pushstart), so that the power unit starts also under difficult conditions (high vacuum with simultaneous
mains undervoltage). This start-up circuit is also used in the SENATOR with 115V motor. This circuit increases the motor torque momentarily by parallel switching in a second start-up capacitor. After approximately 3 seconds this second capacitor is switched off again, so that the motor is not overloaded during continuous operation. This start-up circuit is activated each time the motor is powered up or mains power is interrupted.
Ardo medical AG 7 Service Manual MASTER/SENATOR
Subject to change
Wiring Diagram/Block Diagram
grounded model
Circuit board "Pushstart"
MASTER 230V / 115 V SENATOR 115V
Ardo medical AG 8 Service Manual MASTER/SENATOR
Subject to change
10. MAINTENANCE
General Maintenance
It is the user’s obligation to perform a maintenance check at least once per year. If the manufacturer performs the maintenance check, a service confirmation is placed on the suction pump. If you do not observe the procedures and cleaning instructions there is a danger of contamination!
Maintenance Works for Technical Personnel
General Maintenance
Checking of
Leak Test
Check of Tubing
Safety Check
Maintenance works include the following tests:
Checking of vacuum level, liter performance, function of vacuum regulator
Leak test
Check of tubing
Safety check (electrical and mechanical parts)
Vacuum level, liter performance, function of vacuum regulator:
There are no specifications for these checks. The permitted tolerances are:
- vacuum level ± 5%
- liter performance ± 15%
Connect 1 liter or 2 liter suction jar
Pinch off the patient tubing
Switch pump on and turn vacuum regulator to maximum
vacuum level
When maximum vacuum level of approximately 85kPa has
been reached, switch pump off and observe vacuum meter. After 1 minute, vacuum level should not have decreased more than 10kPa, otherwise complete sealing is not guaranteed.
Check tubings on cleanness* and damages.
*If an overflow has occurred, pack suction pump in a synthetic material bag and dispose of, after clarification with the customer.
Electrical and mechanical parts
Check power connection on damages. Check moving power
unit parts on abrasion or damage (cylinder system, ventilator).
According to IEC 601 and EN 10079-1, the measuring of the
leakage current (using the safety tester) has to be carried out after maintenance or service works.
Ardo medical AG 9 Service Manual MASTER/SENATOR
Subject to change
Access to power unit
The device must be disconnected from mains power before opening the housing!
- Dismount fitting rail
- On each side, unscrew two Phillips screws
- Unscrew another two Phillips screws on the bottom side
of the device
- Pull back housing backplane
- The power unit can now be accessed
Caution! Capacitors may be charged even if device is without power. Discharge capacitors with discharge set!
11. SERVICING AND REPAIRS
For repairs and replacements of spare parts use original parts by ARDO only. The item numbers can be found in the Accessories/Spare Parts list. Technical documentations from ARDO including diagrams, descriptions and adjustment instructions are available to authorized personnel.
Troubleshooting
Problem Cause Remedy
Green signal lamp on the ON/OFF switch is not illuminated after activation
Signal lamp is illuminated, but device is not working or does not start up
Poor or no suction capacity or not enough vacuum
- Power plug does not sit properly in the socket
- No power supply
- Power unit overheated
- Faulty electronics
- Leak in vacuum system
- Defective cylinder system
- Missing or defect o-ring on
plug nipple or filter housing
- Check connection in wall socket
- Check power supply (main fuse)
- Check device fuse
- Check ventilation inlets on bottom and
handle side for free airflow
- Overheat protection may respond, if the device is poorly ventilated
- Replace Pushstart
- Motor defect
- Replace Capacitor
- Check internal tubing connections and
replace if necessary
- Replace retention valves (mind correct placing of blocking side)
- Replace complete cylinder unit if necessary
- Replace o-ring
Ardo medical AG 10 Service Manual MASTER/SENATOR
Subject to change
Accessories/Spare Parts
Product Item No.
- Hydrophobic bacterial filter 50.00.05
- 1 liter suction jar, TPX, graduated 50.00.50
- 2 liter suction jar, TPX, graduated 50.00.51
- 3 liter suction jar, polysulfone, graduated 50.00.52
- 5 liter suction jar, polysulfone, graduated 50.00.53
- Lid with handle and mechanical overflow safety device, 50.00.08 for suction tubing ND 8 mm
- Lid with handle and mechanical overflow safety device, 50.00.10 for suction tubing ND NW 12.7 mm
- Connection tubing with 2 angled coupling connectors, 0.5 m 50.00.01
- Seal for lid 99.00.544
- Coupling connector, angled, ND 8 mm, green 50.00.85
- Coupling connector, angled, ND 8 mm, transparent 50.00.86
- Coupling connector, angled, ND 12.7 mm, transparent 50.00.87
- O-Ring in silicone Ø 8.1 mm, white 99.00.303
- O-Ring in silicone Ø 14 mm, white 51.00.17
- Silicone tubing, transparent, Ø 7/13 mm, sterilisable, per meter 50.00.47
- Silicone tubing, transparent, Ø 12.7/18 mm, sterilisable, per meter 50.00.46
- Retainer for float 50.00.83
- Float 50.00.84
Ardo medical AG 11 Service Manual MASTER/SENATOR
Subject to change
Overview on Spare Parts shown on pages 13-16
MASTER 45 (230V/115V)
Pos. Item No. Product
1 32.00.02 10 32.00.12 10 32.00.24 15 32.00.16 15a 32.00.17
17 32.00.06
Housing MASTER Power unit MASTER 230V Power unit MASTER 115V Capacitor set 230V MASTER (3.5µF /400V and 6µF /400V) Capacitor set 115V MASTER/SENATOR (14µF /250V and 18µF /250V)
Circuit board "Pushstart" (SENATOR only for 115V)
SENATOR 30 (230V/115V)
Pos. Item No. Product
1 32.00.18 10 32.00.19 10 32.00.23
Housing SENATOR Power unit SENATOR 230V Power unit SENATOR 115V
14 32.00.20 Capacitor 230V / 3.5µF/400V 15a 32.00.17
Capacitor set 115V MASTER/SENATOR (14µF /250V and 18µF /250V)
MASTER 45 and SENATOR 30 (230V/115V)
Pos. Item No. Product
# 32.00.03 Main socket complete
# 32.00.04 Fuse T630mA (230V) 10 pcs
# 32.00.25 Fuse T1.6A (115V) 10 pcs
4 99.00.693 Power switch
5 32.00.07 Vacuum-meter
6 32.00.08 Pneumatics set
7 32.00.09 Rubber shock-absorber set
8 32.00.10 Vacuum regulator complete
# 32.00.11 Valve set 8 pcs 11 32.00.13 Slide block 1 12 32.00.14 Slide block 2 13 32.00.15 Cylinder 17 32.00.06
# 32.00.21 Turn knob with pointer
# 32.00.01 Unit bases
# 99.00.328 Shipping packaging complete
# 32.00.22 Ground-compensate set
# = not shown in the pictures
Circuit board "Pushstart" (SENATOR only for 115V)
Ardo medical AG 12 Service Manual MASTER/SENATOR
Subject to change
MASTER 45 (230V/115V)
Pos. Item No. Product
Pos. 10 Pos. 15 (15a) Pos. 17 Pos. 1
1 32.00.02
10 32.00.12 10 32.00.24
15 32.00.16
15a 32.00.17 17 32.00.06
Housing MASTER
Power unit MASTER 230V Power unit MASTER 115V
Capacitor set 230V MASTER (3.5µF /400V and 6µF /400V)
Capacitor set 115V MASTER/SENATOR (14µF /250V and 18µF /250V) Circuit board "Pushstart" (SENATOR only for 115V)
Ardo medical AG 13 Service Manual MASTER/SENATOR
Subject to change
SENATOR 30 (230V/115V)
Pos. Item No. Product
Pos. 14 (15a) Pos. 10 Pos. 1
1 32.00.18
10 32.00.19 10 32.00.23
Housing SENATOR
Power unit SENATOR 230V Power unit SENATOR 115V
14 32.00.20 Capacitor 230V / 3.5µF/400V
15a 32.00.17
Capacitor set 115V MASTER/SENATOR (14µF /250V and 18µF /250V)
Ardo medical AG 14 Service Manual MASTER/SENATOR
Subject to change
Tubing MASTER 45 / SENATOR 30
Pos. 17 Pos. 4 Pos. 5 Pos. 8 Pos. 6
Pos. Item No. Product
4 99.00.693 Power switch
5 32.00.07 Vacuum-meter 6 32.00.08 Pneumatics set
8 32.00.10 Vacuum regulator complete
17 32.00.06
Circuit board "Pushstart" (SENATOR only for 115V)
Ardo medical AG 15 Service Manual MASTER/SENATOR
Subject to change
Power unit MASTER 45 / SENATOR 30
Pos. 13 Pos. 11 Pos. 7 Pos. 12
Pos. Item No. Product
7 32.00.09 Rubber shock-absorber set 11 32.00.13 Slide block 1, complete
12 32.00.14 Slide block 2
13 32.00.15 Cylinder, complete
Ardo medical AG 16 Service Manual MASTER/SENATOR
Subject to change
Replacing Spare Parts
Always disconnect the suction pumps MASTER/SENATOR from mains power before opening.
Only technical personnel may replace components or repair the suction pumps MASTER/ SENATOR. Due to reasons of functional safety and testing, replacement is limited to mechanical parts and individual electrical components. When using foreign parts or carrying out non-authorized works on the suction pump, we refer to operating instructions Chapter 6 - Guarantee -. Replacing components shall be carried out according to the specification, the processes and/or the photographic illustrations. If difficulties arise, please contact the manufacturer Ardo medical AG.
Cylinder
Dismantling
Assembly
Initial position; power unit tray is dismantled from the housing
Loosen the two fixing screws on the cylinder base
Loosen mounting of the piston rods on the slide
Slide complete cylinder with piston rod up and out
Place cylinder from above on the cylinder holder
Insert a 0.3 mm feeler gauge between cylinder holder
and cylinder
Set slide to the lower dead center
Pull piston rod all the way down (do not damage
mechanically!)
Fix on slide with Allen screw
Pull out feeler gauge and adjust cylinder
Tighten both Allen screws on the cylinder foot
Assembly of the second cylinder is done analogously.
CAUTION! Glass cylinders and graphite pistons must never be exchanged or combined with another cylinder. These two components are specially designed for alignment.
Valves
If device is contaminated or vacuum performance is poor, check valves on cleanness and damage.
The valves are fixed in place by a two-component epoxy adhesive (araldite)
Ardo medical AG 17 Service Manual MASTER/SENATOR
Subject to change
Belt Drive
The belts must be checked annually for extensive wear.
Wearing Parts
Ardo medical AG recommends to check the following wearing parts annually and replace them if necessary:
- Inlet and outlet valves on the cylinder
- Mechanical drive
- All tubing connections
- Rubber cushions of the power unit
12. TECHNICAL INFORMATION
Specifications
Suction Pump
MASTER 45
Suction Pump
SENATOR 30
MASTER 45 suction pump 230 Volt, 50/60 Hz MASTER 45 suction pump 115 Volt, 60Hz
SENATOR 30 suction pump 230 Volt, 50/60Hz SENATOR 30 suction pump 115 Volt, 60Hz
SENATOR 30 double protection on request
- Dimensions without trolley: 345 x 245 x 282 mm
- Dimensions with trolley: 1000 x 470 x 680 mm
- Weight without trolley: 7,4 kg
- Weight with trolley: 21 kg
- Protection class grounded: Class l (with ground safety wire)
- Protection level: BF
Drip-proof IPX1
- Double protection SENATOR 230V: Class II
- Protection level: BF
Drip-proof IPX1
- Classification acc. to MDD93/42: Class 2a
- Rated voltage: 230/115 Volt AC +/- 15%
- Rated current: 50/60Hz
- Power consumption 230V Æ 100W/0.45A
115V Æ 100W/1.00A
- Fuse: 230V Æ T630mA
115V Æ T1.6A
- Suction capacity: MASTER 45 Æ 45 l/min.
SENATOR 30 Æ 30 l/min.
- Vacuum range: 10 to 85kPa
75 to 640mmHg
Ardo medical AG 18 Service Manual MASTER/SENATOR
Subject to change
- Conformity: 0123
to the standard MDD 93/42 EEC of 14.7.1993 concerning medical devices EN 60601-1, EN60601-1-2, EN 10079-1
- UL 2601-1 (115V)
Transport and Storage
MASTER/SENATOR shall be transported and stored in the original packaging only. We recommend a separate storage, protected from UV, heat and dust.
Transport and storage up to 15 weeks
For transport and storage periods up to 15 weeks, the following storage conditions are valid:
- Temperature: -25°C to +70°C
- Relative humidity: 10% to 75%
- Air pressure: 500hPa to 1060hPa
Transport and storage for more than 15 weeks
After this period, the operating condition values are valid:
- Temperature: +10°C to +40°C
- Relative humidity: 30% to 75%
- Air pressure: 700hPa to 1060hPa
Disposal after Service Life
For disposal of a disused MASTER/SENATOR, the decontaminated suction pump may be sent back to the manufacturer or, after disinfection, separated into materials (metals, synthetics) and disposed of environment-friendly by an electronics recycling company. After decontamination, there are no special or increased risks.
Manufacturer:
Ardo medical AG Gewerbestrasse 19 CH-6314 Unterägeri Switzerland
Ardo medical AG 19 Service Manual MASTER/SENATOR
Subject to change
Loading...