The suction pumps are designed and constructed according to the safety regulations of
IEC 601-1, the EMC standard IEC 60601-1-2 and EN 10079-1. For the user's and the
patient’s safety as well as to prevent equipment damage, please observe these safety
rules:
Safety Rules
• The suction pumps MASTER/SENATOR may only be installed and operated by
competent personnel.
• Use the suction pumps with the ARDO hydrophobic bacterial filter only.
• To ensure the proper venting of the pumps, the feet must be installed on the bottom
of the pumps if used without trolley.
• Never start operating the suction pump when the housing is open (rotating parts –
hazard of injury and electric shock).
• Capacitors may be charged even if device is without power. Discharge capacitors with
discharge set!
• Check contact points of the housing for tears. If there are any tears, exchange housing.
• See that MASTER/SENATOR are set into operation according to Chapter 4 - Installation -
in the operating instructions.
• Store the service manual in a safe place, so that it is available to the trained technical
personnel at all times.
General Safety Regulations
• Never use MASTER/SENATOR in the presence of ignitable gases as anesthetic
agents. Explosion hazard!
• The suction pumps may only be opened by qualified technical personnel.
Electric shock hazard!
• Always switch the suction pump off and disconnect from mains power before
cleaning and opening. Electric shock hazard!
• Do not use solvents for cleaning. Otherwise the surface of the housing or the
synthetic material parts may be damaged.
Ardo medical AG 3 Service Manual
MASTER/SENATOR
Subject to change
Responsibility
ARDO is only responsible for effects on safety, reliability and performance of the device, if:
• Installation, new adjustments, alterations or repairs and maintenance works are carried
out by authorized persons.
• For repairs and replacements of spare parts use original parts by ARDO only.
• The electric installations of the respective room meet the requirements of the standards
IEC 601 and MPG (German Medicinal Products Act), and if the devices are maintained
and repaired according to this standard and to EN 10079-1.
• The pump is used in accordance with the operating instructions.
Guarantee
The suction pumps MASTER/SENATOR have a guarantee period of 2 years.
General Conditions
Maintenance
Ardo medical AG warrants for material and production faults of
the manufactured products. The guarantee period can be taken
from the respective operating instructions. Faulty material is
replaced free of charge within the guarantee period, assumed
the device was not used improperly. Wearing parts are
excluded. For ensuring guarantee and faultless functioning of
the device, the notes of the operating instructions have to be
observed. Furthermore, only original spare parts by
Ardo medical AG may be installed and/or used.
The right to guarantee is waived, when non-authorized persons
interfere with the device or alterations not conforming to the
standard IEC 601 and EN 10079-1 are carried out. There is no
right to guarantee exceeding the described scope of guarantee,
as liability to consequential damages etc.
Read operating instructions Chapter 5 - Maintenance and Care, Cleaning -.
Ardo medical AG 4 Service Manual
MASTER/SENATOR
Subject to change
Service
Please contact to the following addresses for service, maintenance or any questions
about this product or any other product of Ardo medical AG:
Switzerland
Germany
International
Ardo medical AG
Gewerbestrasse 19
CH-6314 Unterägeri
Switzerland
Tel. +41-(0)41 754 70 70
Fax +41-(0)41 754 70 71
info@ardo.ch
www.ardo.ch
The ARDO suction pumps MASTER/SENATOR work according to the principle of a piston/
cylinder system.
Both power units are based on a modular driving concept, equipped with 2 cylinder units.
MASTER 45 2 cylinder power unit (approximately 300 rpm)
SENATOR 30 2 cylinder power unit (approximately 220 rpm)
A two-stage belt transmission was developed for driving the cylinder units that may be
used for both power units (MASTER and SENATOR).
Transmission reduction for the two power units can be adjusted with the two belt pulleys
of the first reduction level (motor pinion + belt wheel = 1. level).
Ardo medical AG 5 Service Manual
MASTER/SENATOR
Subject to change
For both power units a bipolar capacitor motor with built-in temperature sensor is used.
This protects the motor coil from heating to more than 125°C.
The piston is driven over a linear unit (slide). By using this system, the piston system is
exposed to only minimal thrust forces, which ensures low wear and high service life.
In MASTER/SENATOR the pistons work in angles of 90° to each other. This ensures a
well-balanced run of the pump with low vibration levels.
1 Cylinder unit 6 Vacuum-meter
2 Discharge valve 7 Vacuum regulator
3 Inlet valve 8 Vacuum connector
4 AC motor 9 Exhaust
5 Linear drive 10 2-stage belt transmission
Ardo medical AG 6 Service Manual
MASTER/SENATOR
Subject to change
Electrical Layout
The driving motor is a bipolar capacitor motor and may be used for both power units
(MASTER and SENATOR).
MASTER is equipped with a electronically controlled start-up circuit (Pushstart), so that
the power unit starts also under difficult conditions (high vacuum with simultaneous
mains undervoltage). This start-up circuit is also used in the SENATOR with 115V motor.
This circuit increases the motor torque momentarily by parallel switching in a second
start-up capacitor.
After approximately 3 seconds this second capacitor is switched off again, so that the
motor is not overloaded during continuous operation.
This start-up circuit is activated each time the motor is powered up or mains power is
interrupted.
Ardo medical AG 7 Service Manual
MASTER/SENATOR
Subject to change
Wiring Diagram/Block Diagram
grounded model
Circuit board "Pushstart"
MASTER 230V / 115 V
SENATOR 115V
Ardo medical AG 8 Service Manual
MASTER/SENATOR
Subject to change
10. MAINTENANCE
General Maintenance
It is the user’s obligation to perform a maintenance check at least once per year. If the
manufacturer performs the maintenance check, a service confirmation is placed on the
suction pump.
If you do not observe the procedures and cleaning instructions there is a danger of
contamination!
Maintenance Works for Technical Personnel
General Maintenance
Checking of
Leak Test
Check of Tubing
Safety Check
Maintenance works include the following tests:
• Checking of vacuum level, liter performance, function of
vacuum regulator
• Leak test
• Check of tubing
• Safety check (electrical and mechanical parts)
Vacuum level, liter performance, function of vacuum regulator:
There are no specifications for these checks.
The permitted tolerances are:
- vacuum level ± 5%
- liter performance ± 15%
• Connect 1 liter or 2 liter suction jar
• Pinch off the patient tubing
• Switch pump on and turn vacuum regulator to maximum
vacuum level
• When maximum vacuum level of approximately 85kPa has
been reached, switch pump off and observe vacuum meter.
After 1 minute, vacuum level should not have decreased more
than 10kPa, otherwise complete sealing is not guaranteed.
• Check tubings on cleanness* and damages.
*If an overflow has occurred, pack suction pump in a synthetic
material bag and dispose of, after clarification with the
customer.
Electrical and mechanical parts
• Check power connection on damages. Check moving power
unit parts on abrasion or damage (cylinder system, ventilator).
• According to IEC 601 and EN 10079-1, the measuring of the
leakage current (using the safety tester) has to be carried out
after maintenance or service works.
Ardo medical AG 9 Service Manual
MASTER/SENATOR
Subject to change
Access to power unit
The device must be disconnected from mains power
before opening the housing!
- Dismount fitting rail
- On each side, unscrew two Phillips screws
- Unscrew another two Phillips screws on the bottom side
of the device
- Pull back housing backplane
- The power unit can now be accessed
Caution! Capacitors may be charged even if device is
without power. Discharge capacitors with discharge set!
11. SERVICING AND REPAIRS
For repairs and replacements of spare parts use original parts by ARDO only. The item
numbers can be found in the Accessories/Spare Parts list. Technical documentations
from ARDO including diagrams, descriptions and adjustment instructions are available to
authorized personnel.
Troubleshooting
Problem Cause Remedy
Green signal lamp on the
ON/OFF switch is not illuminated
after activation
Signal lamp is illuminated, but
device is not working or does not
start up
Poor or no suction capacity or
not enough vacuum
- Power plug does not sit
properly in the socket
- No power supply
- Power unit overheated
- Faulty electronics
- Leak in vacuum system
- Defective cylinder system
- Missing or defect o-ring on
plug nipple or filter housing
- Check connection in wall socket
- Check power supply (main fuse)
- Check device fuse
- Check ventilation inlets on bottom and
handle side for free airflow
- Overheat protection may respond, if
the device is poorly ventilated
- Replace Pushstart
- Motor defect
- Replace Capacitor
- Check internal tubing connections and
replace if necessary
- Replace retention valves (mind correct
placing of blocking side)
Housing MASTER
Power unit MASTER 230V
Power unit MASTER 115V
Capacitor set 230V MASTER (3.5µF /400V and 6µF /400V)
Capacitor set 115V MASTER/SENATOR
(14µF /250V and 18µF /250V)
Circuit board "Pushstart" (SENATOR only for 115V)
SENATOR 30 (230V/115V)
Pos. Item No. Product
1 32.00.18
10 32.00.19
10 32.00.23
Housing SENATOR
Power unit SENATOR 230V
Power unit SENATOR 115V
Always disconnect the suction pumps MASTER/SENATOR from mains power
before opening.
Only technical personnel may replace components or repair the suction pumps MASTER/ SENATOR. Due to reasons of functional safety and testing, replacement is limited to
mechanical parts and individual electrical components. When using foreign parts or
carrying out non-authorized works on the suction pump, we refer to operating instructions
Chapter 6 - Guarantee -. Replacing components shall be carried out according to the
specification, the processes and/or the photographic illustrations. If difficulties arise,
please contact the manufacturer Ardo medical AG.
Cylinder
Dismantling
Assembly
Initial position; power unit tray is dismantled from the
housing
• Loosen the two fixing screws on the cylinder base
• Loosen mounting of the piston rods on the slide
• Slide complete cylinder with piston rod up and out
• Place cylinder from above on the cylinder holder
• Insert a 0.3 mm feeler gauge between cylinder holder
and cylinder
• Set slide to the lower dead center
• Pull piston rod all the way down (do not damage
mechanically!)
• Fix on slide with Allen screw
• Pull out feeler gauge and adjust cylinder
• Tighten both Allen screws on the cylinder foot
Assembly of the second cylinder is done analogously.
CAUTION! Glass cylinders and graphite pistons must never
be exchanged or combined with another cylinder. These two
components are specially designed for alignment.
Valves
If device is contaminated or vacuum performance is poor,
check valves on cleanness and damage.
• The valves are fixed in place by a two-component epoxy
adhesive (araldite)
Ardo medical AG 17 Service Manual
MASTER/SENATOR
Subject to change
Belt Drive
The belts must be checked annually for extensive wear.
Wearing Parts
Ardo medical AG recommends to check the following wearing parts annually and replace
them if necessary:
- Inlet and outlet valves on the cylinder
- Mechanical drive
- All tubing connections
- Rubber cushions of the power unit
12. TECHNICAL INFORMATION
Specifications
Suction Pump
MASTER 45
Suction Pump
SENATOR 30
MASTER 45 suction pump 230 Volt, 50/60 Hz
MASTER 45 suction pump 115 Volt, 60Hz
SENATOR 30 suction pump 230 Volt, 50/60Hz
SENATOR 30 suction pump 115 Volt, 60Hz
SENATOR 30 double protection on request
- Dimensions without trolley: 345 x 245 x 282 mm
- Dimensions with trolley: 1000 x 470 x 680 mm
- Weight without trolley: 7,4 kg
- Weight with trolley: 21 kg
- Protection class grounded: Class l (with ground safety wire)
- Protection level: BF
Drip-proof IPX1
- Double protection SENATOR 230V: Class II
- Protection level: BF
Drip-proof IPX1
- Classification acc. to MDD93/42: Class 2a
- Rated voltage: 230/115 Volt AC +/- 15%
- Rated current: 50/60Hz
- Power consumption 230V Æ 100W/0.45A
115V Æ 100W/1.00A
- Fuse: 230V Æ T630mA
115V Æ T1.6A
- Suction capacity: MASTER 45Æ 45 l/min.
SENATOR 30Æ 30 l/min.
- Vacuum range: 10 to 85kPa
75 to 640mmHg
Ardo medical AG 18 Service Manual
MASTER/SENATOR
Subject to change
- Conformity: 0123
to the standard MDD 93/42 EEC
of 14.7.1993 concerning medical devices
EN 60601-1, EN60601-1-2, EN 10079-1
- UL 2601-1 (115V)
Transport and Storage
MASTER/SENATOR shall be transported and stored in the original packaging only.
We recommend a separate storage, protected from UV, heat and dust.
Transport and
storage up to
15 weeks
For transport and storage periods up to 15 weeks, the following
storage conditions are valid:
- Temperature: -25°C to +70°C
- Relative humidity: 10% to 75%
- Air pressure: 500hPa to 1060hPa
Transport and
storage for more
than 15 weeks
After this period, the operating condition values are valid:
- Temperature: +10°C to +40°C
- Relative humidity: 30% to 75%
- Air pressure: 700hPa to 1060hPa
Disposal after Service Life
For disposal of a disused MASTER/SENATOR, the decontaminated suction pump may be
sent back to the manufacturer or, after disinfection, separated into materials (metals,
synthetics) and disposed of environment-friendly by an electronics recycling company.
After decontamination, there are no special or increased risks.
Manufacturer:
Ardo medical AG
Gewerbestrasse 19
CH-6314 Unterägeri
Switzerland
Ardo medical AG 19 Service Manual
MASTER/SENATOR
Subject to change
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.