ARAG DIGIWOLF User Manual

digi
CELL POWERED VERSION
4628405 4628506 4628707
Software rel. 1.0X
INSTALLATION, USE AND MAINTENANCE
LEGEND SYMBOLS
=
Generic danger
=
Warning
CONTENTS
1 Product description ........................................................................................................................................................... 4
1.1 Intended use............................................................................................................................................................... 4
2 Flowmeter assembly ......................................................................................................................................................... 4
2.1 Monitor rotation .......................................................................................................................................................... 5
2.2 Dimensions (mm) ....................................................................................................................................................... 6
2.3 Hydraulic connections ................................................................................................................................................ 6
2.3.1 Hydraulic connection for brass/fork connections ....................................................................................................... 6
2.4 Power supply .............................................................................................................................................................. 7
2.4.1 Battery replacement ................................................................................................................................................... 7
3 Controls in the menu ......................................................................................................................................................... 8
Menu structure ................................................................................................................................................................... 9
4 Preliminary setup for use................................................................................................................................................ 10
4.1 Calibration ................................................................................................................................................................ 10
4.1.1 Automatic calibration ................................................................................................................................................10
4.1.2 Manual calibration .....................................................................................................................................................11
4.2 Flowrate alarms ........................................................................................................................................................ 12
4.3 Display ...................................................................................................................................................................... 12
4.4 Energy saving ........................................................................................................................................................... 13
4.5 Options ..................................................................................................................................................................... 13
4.5.1 Language ................................................................................................................................................................. 13
4.5.2 Units of measurement .............................................................................................................................................. 14
4.5.2.1 Rate units of measurement ................................................................................................................................... 14
4.5.2.2 Volume units of measurement ............................................................................................................................... 14
4.6 Test ........................................................................................................................................................................... 15
4.6.1 Display test ............................................................................................................................................................... 15
4.6.2 Keys test ................................................................................................................................................................... 15
5 Use .............................................................................................................................................................................. 16
5.1 Partial totalizer reset ................................................................................................................................................. 16
6 Cleaning and repair ......................................................................................................................................................... 17
6.1 Troubleshooting ........................................................................................................................................................ 17
6.2 Paddle cleaning and replacement ............................................................................................................................ 18
6.3 OR replacement ....................................................................................................................................................... 19
7 Technical data ................................................................................................................................................................. 20
8 Disposal at the end of service ....................................................................................................................................... 20
9 Guarantee terms .............................................................................................................................................................. 22
3
APPROXIMA
TE CONST
ANT
XXX
INTRODUCTION
1 PRODUCT DESCRIPTION
DigiWolf is a battery powered paddle owmeter capable of measuring the rate of a uid in a hydraulic circuit, and it can display the results of these measurements.
1.1 Intended use
This device is designed to work on agricultural machinery for crop spraying applications. The machine is designed and built in compliance with EN ISO 14982 standard (Electromagnetic compatibility - Forestry and farming machines), harmonized with 2004/108/EC Directive.
The owmeter must not be used to measure the passage of hydrocarbons, ammable, explosive or toxic liquids. The owmeter is not suitable for contact with liquids for human consumption.
2 FLOWMETER ASSEMBLY
Install the owmeter at least 20 cm from the elements that could cause turbulence inside the tubes (valves, bends, constrictions, etc.). The owmeter can be installed in a vertical or horizontal position.
CAUTION:
- Do not install the owmeter with the connector facing downwards (Fig. 1).
- The system must have a ltering element with a lter of at least 50 mesh, together with a safety valve to limit use pressure at the specied max. valve (Tab. 2 - Par. 2.3.1).
OK
Assemble the owmeter using the suitable mounting parts (Fig. 2): t the bolts (M8) in their seats, then make them slide to their stop position to prevent them from coming out.
OK OK
NO!
Fig. 1
The owmeter body must be assembled with the arrow on the label facing the ow direction.
APPROXIMATE CONSTANT
XXX
Fig. 2
4
2.1 Monitor rotation
Fig. 3
INSTALLATION
In case of vertical assembly, to simplify the use of the DigiWolf it is possible to rotate the monitor by 90° with respect to the body (Fig. 3).
1) Remove the fork from the monitor using a screwdriver.
2) Remove the monitor from the owmeter body.
Fig. 4
Fig. 5
Fig. 6
If the monitor is turned upside down, the letters A and B can be seen: these correspond to the two possible monitor positions (parallel and perpendicular to the body).
DigiWolf is supplied with the sensor in position A (parallel to the body).
With the aid of a screwdriver, loosen the screw locking the sensor without removing it.
Rotate the sensor anti-clockwise until reaching B stop position.
Fig. 7
Tighten back the screw without forcing it. The monitor is ready to be tted in perpendicu­lar position to the body.
Fit the monitor back on the owmeter body.
5
INSTALLATION
2.2 Dimensions (mm)
A
T
2.3 Hydraulic connections
E
120
B
D
C
143
CODE A B C D E T
4628405 140 128 205 31 41 T5
4628506 145 148 210 36 45 T6
4628707 157 162 222 41 45 T7
Tab. 1
95
Fig. 8
2.3.1 Hydraulic connection for brass/fork connections
Avoid bends and constrictions before connections and on tubes.
Use ARAG connections with their suitable OR with MALE CONNECTIONS [T connections - General Catalogue (Tab. 2)].
The tubes must be able to stand a pressure of at least twice the max. operation pressure of the owmeter (Tab. 2) CONSIDERING THE OPERATION PRESSURES WITHIN THE SYSTEM.
Hose tail tightening must be done using the suitable metal clamps to ensure perfect mechanical sealing, even at high pressures. The connection with threaded connectors must be done paying attention to operation pressure.
CAUTION: For the implementation on already operating systems it is necessary to follow all safety rules described herein. System assembly and start-up must be carried out by expert personnel according to the safety rules so as to ensure the same safety level of the system the owmeter is going to be installed in.
After connection, check for the perfect sealing of the tubes and fork connections.
Fig. 9
CODE l/min. US GPM Connection Ø equivalent (inch)
4628405 10-200 2.6-53 T5 F 1 1/4" 20 290
4628506 20-400 5-106 T6 F 1 1/2" 12 174
4628707 40-800 10-210 T7 F 2" 7 130
The diameter in inches (Ø equivalent) is given only as an indication of the typical passage of the owmeter body. Actually, it is possible to choose different sizes depending on the fork connection used.
Max. p
(bar)
6
Max. p
(PSI)
Tab. 2
0.0 l/min
1
2
0.0 l
INSTALLATION
2.4 Power supply
DigiWolf is powered by 2 AA batteries type LR6 (alkaline) or FR6 (Li-Fe S2).
2.4.1 Battery replacement
When the batteries are low, the symbol will blink on the display.
Fig. 10
To insert or replace the batteries, proceed as follows:
1) Remove the lid of the battery compartment (A) loosening the 4 screws (B);
2) Remove the battery support (C) and replace the batteries, respecting their polarity (as indicated).
3) Fit the support back in its housing and ret the lid by tightening the 4 screws without forcing them, taking care not to crush or break the wires in the battery case and that the lid seal remains in its seat.
C
B
Fig. 11
A
To avoid damaging the device, replace low batteries as soon as this symbol is displayed .
Remember to remove the batteries when the device is not going to be in use for a long period.
Use only the recommended type of batteries. Do not use combinations of different types of batteries (old and new, carbon and alkaline, etc.). Do not try to recharge the batteries.
ARAG is not liable for damage to the equipment, persons, animals or things caused by failure to observe the above instructions.
Do not dispose of exhausted batteries in the environment. Dispose of in a suitable container.
7
0.0 l
2
1
0.0 l/min
0.0 l
2
1
0.0 l/min
Setup menu
Calibration
Flowrate alarms Display
A
B
Flowrate constant
Min value Max value
C
1
50000
01230
B
D
A
SETUP
3 CONTROLS IN THE MENU
Fig. 12
Fig. 13
Fig. 14
SWITCH ON
The display turns on automatically at liquid passage. If you need to switch on the device to check or modify data, press OK.
DigiWolf switches on and displays the gure shown.
Press the key to see the succession of different values written in full (central part of the display).
At rst device switch on, or after battery replacement, the device will briey display its name and software version.
ACCESS TO SETUP MENU
Press the keys for 2 seconds at the same time to access setup menu.
SELECTION AND ACCESS TO MENU ITEMS
A Press in succession to move through items (the selected item is highlighted with
a black band)
B Press to access the selected item
Three dots under an item show the presence of another setup menu.
Fig. 15
MODIFYING DATA
A Press to toggle from one digit to another
B Press in succession to modify the value of the digit highlighted by the cursor
C Press to conrm the change. The display goes back to the previous screen.
D Press to exit current page without conrming changes.
8
Loading...
+ 16 hidden pages