ARAG DIGIWOLF User Manual

digi
BATTERY SUPPLY VERSION (12V)
4627405A 4627506A 4627707A
Software rel. 1.0X
INSTALLATION, USE AND MAINTENANCE
=
Generic danger
=
Warning
This manual is an integral part of the equipment to which it refers and must accompany the equipment in case of sale or change of ownership. Keep it for future reference; ARAG reserves the right to modify the specications and instructions regarding the product at any time and without prior notice.
CONTENTS
1 Product description ...........................................................................................................................................................4
1.1 Intended use...............................................................................................................................................................4
2 Flowmeter assembly .........................................................................................................................................................4
2.1 Monitor rotation ..........................................................................................................................................................5
2.2 Dimensions (mm) .......................................................................................................................................................6
2.3 Hydraulic connections ................................................................................................................................................6
2.3.1 Hydraulic connection for brass/fork connections .................................................................................................... 6
2.4 Electric connections ...................................................................................................................................................7
2.5 Pump Stop Module connection (for FILLING mode ONLY) ........................................................................................8
3 Controls in the menu .........................................................................................................................................................9
3.1 First switch on .............................................................................................................................................................9
Menu structure ................................................................................................................................................................. 10
4 Preliminary setup for use................................................................................................................................................ 11
4.1 Operating mode selection (FILLING / FLOWRATE) ................................................................................................. 11
5 Use in FILLING mode ...................................................................................................................................................... 11
5.1 Preliminary setup ...................................................................................................................................................... 11
5.1.1 Valve activation time .............................................................................................................................................. 11
5.2 Flowmeter use (FILLING mode) ...............................................................................................................................12
6 Use in FLOWRATE mode ................................................................................................................................................ 13
6.1 Flowmeter use (FLOWRATE mode) .........................................................................................................................13
6.2 Partial totalizer reset .................................................................................................................................................13
7 Other settings .................................................................................................................................................................. 14
7.1 Calibration ................................................................................................................................................................14
7.1.1 Automatic calibration ............................................................................................................................................14
7.1.2 Manual calibration .................................................................................................................................................15
7.2 Flowrate alarms ........................................................................................................................................................ 15
7.3 Display ......................................................................................................................................................................16
7.4 Options .....................................................................................................................................................................16
7.4.1 Language .............................................................................................................................................................. 16
7.4.2 Units of measurement ........................................................................................................................................... 17
7.4.2.1 Rate units of measurement ................................................................................................................................... 17
7.4.2.2 Volume units of measurement ............................................................................................................................... 17
7.5 Test ...........................................................................................................................................................................18
7.5.1 Display test ............................................................................................................................................................ 18
7.5.2 Keys test ................................................................................................................................................................ 18
8 Cleaning and repair ......................................................................................................................................................... 19
8.1 Troubleshooting ........................................................................................................................................................19
8.2 Paddle cleaning and replacement ............................................................................................................................20
8.3 OR replacement .......................................................................................................................................................21
9 Technical data ................................................................................................................................................................. 22
10 Disposal at the end of service ....................................................................................................................................... 22
11 Guarantee terms ..............................................................................................................................................................26
3
APPROXIMA
TE CONST
ANT
XXX
INTRODUCTION
1 PRODUCT DESCRIPTION
DigiWolf is a paddle owmeter capable of measuring the rate of a uid in a hydraulic circuit, and it can display the results of these measurements. The device can also be set to control the lling of a tank by stopping the introduction of liquid once the required quantity has been reached (this function is possible only by connecting the owmeter to the Pump Stop Module or the relevant cable - Par. 2.4-2.5).
1.1 Intended use
This device is designed to work on agricultural machinery for crop spraying applications. The machine is designed and built in compliance with EN ISO 14982 standard (Electromagnetic compatibility - Forestry and farming machines), harmonized with 2004/108/EC Directive.
The owmeter must not be used to measure the passage of hydrocarbons, ammable, explosive or toxic liquids. The owmeter is not suitable for contact with liquids for human consumption.
2 FLOWMETER ASSEMBLY
Install the owmeter at least 20 cm from the elements that could cause turbulence inside the tubes (valves, bends, constrictions, etc.). The owmeter can be installed in a vertical or horizontal position.
CAUTION:
- Do not t the owmeter with the display facing down (Fig. 1).
- The system must have a ltering element with a lter of at least 50 mesh, together with a safety valve to limit use pressure at the specied max. valve (Tab. 2 - Par. 2.3.1).
OK
Assemble the owmeter using the suitable mounting parts (Fig. 2): t the bolts (M8) in their seats, then make them slide to their stop position to prevent them from coming out.
OK OK
NO!
Fig. 1
The owmeter body must be assembled with the arrow on the label facing the ow direction.
APPROXIMATE CONSTANT
XXX
Fig. 2
4
2.1 Monitor rotation
Fig. 3
INSTALLATION
In case of vertical assembly, to simplify the use of the DigiWolf it is possible to rotate the monitor by 90° with respect to the body (Fig. 3).
1) Remove the fork from the monitor using a screwdriver.
2) Remove the monitor from the owmeter body.
Fig. 4
Fig. 5
Fig. 6
If the monitor is turned upside down, the letters A and B can be seen: these correspond to the two possible monitor positions (parallel and perpendicular to the body).
DigiWolf is supplied with the sensor in position A (parallel to the body).
With the aid of a screwdriver, loosen the screw locking the sensor without removing it.
Rotate the sensor anti-clockwise until reaching B stop position.
Fig. 7
Tighten back the screw without forcing it. The monitor is ready to be tted in perpendicu­lar position to the body.
Fit the monitor back on the owmeter body.
5
INSTALLATION
2.2 Dimensions (mm)
A
T
2.3 Hydraulic connections
E
120
B
D
C
143
CODE A B C D E T
4627405A 140 128 205 31 41 T5
4627506A 145 148 210 36 45 T6
4627707A 157 162 222 41 45 T7
Tab. 1
95
Fig. 8
2.3.1 Hydraulic connection for brass/fork connections
Avoid bends and constrictions before connections and on tubes.
Use ARAG connections with their suitable OR with MALE CONNECTIONS [T connections - General Catalogue (Tab. 2)].
Do not use the owmeter at pressures greater than those listed in Table 2. Regarding connections, use tubes and ttings properly sized for the operating pressure of the system.
CAUTION: For the implementation on already operating systems it is necessary to follow all safety rules described herein. System assembly and start-up must be carried out by expert personnel according to the safety rules so as to ensure the same safety level of the system the owmeter is going to be installed in.
After connection, check for the perfect sealing of the tubes and fork connections.
Fig. 9
CODE l/min. US GPM Connection Ø equivalent (inch)
4627405A 10-200 2.6-53 T5 F 1 1/4" 20 290
4627506A 20-400 5-106 T6 F 1 1/2" 12 174
4627707A 40-800 10-210 T7 F 2" 7 130
Max. p
(bar)
Max. p
(PSI)
Tab. 2
The diameter in inches (Ø equivalent) is given only as an indication of the typical passage of the owmeter body. Actually, it is possible to choose different sizes depending on the fork connection used.
6
INSTALLATION
2.4 Electric connections
DigiWolf is designed to operate independently, but may be connected to ARAG devices (computers, monitors, displays) or equipment from other manufacturers as long as they are set up for use with owmeters of this type.
DigiWolf can be powered by means of a generic power supply cable (cod. 4622BA50000.100 - not supplied), which does not allow the operator to control lling functions directly from the device. Connections are listed in Table 3:
Cable colour Position Connection
Grey 1 GND
Brown 2 +12 VDC
Yellow/Green 3
Black 4 Pump control
For the use as lling owmeter, the DigiWolf must be connected to ARAG devices only.
IN CASE OF ELECTRIC ARC WELDING, MAKE SURE THAT THE POWER SUPPLY TO THE DEVICE IS DISCONNECTED; IF NECES­SARY, DISCONNECT THE POWER SUPPLY CABLES.
ARAG is not liable for damages to the equipment, persons, animals or things caused by a wrong or unsuitable installation of the owmeter. Failure to observe the above instructions automatically voids the warranty.
The DigiWolf connections on devices other than ARAG are indicated in Tab. 4.
Signal
(square wave)
Tab. 3
Position Connection
4 3 2 1
Fig. 10
1 GND
2 +12 VDC
3
4 Pump control
Signal
(square wave)
Tab. 4
To connect DigiWolf to Bravo 300/400 series computers and monitors use the "Pump Stop Kit" (code 4622BA50000.200) or the adapter cable (code 4622BA50000.110 - ARAG Catalogue), which allows for connecting the DigiWolf 4-pole connector to the suitable Bravo series computer 3-pole connector.
When connecting DigiWolf with external devices (Bravo 300/400, etc.), enter IN SUCH DEVICES the approximate constant accor­ding to the table below:
CODE
4627405A 250 945
4627506A 132 500
4627707A 64 242
Approximate constant
(puls/l)
Approximate constant
(puls/gal)
Tab. 5
7
*
T
T
*
= Optional
*
INSTALLATION
2.5 Pump Stop Module connection (for FILLING mode ONLY)
For any information about the connections with the Pump Stop Module (code 4622BA50000.200 - ARAG Catalogue), consult the rele- vant use and maintenance manual (D20059), that can be downloaded at www.aragnet.com, in the Manuals section.
Fig. 11
In case the system features an ARAG 453 series 3-wire valve, we recommend to use an adapter cable (code 46220CBU01.100 - ARAG Catalogue) which will replace the Pump Stop Module.
8
3 CONTROLS IN THE MENU
0.0 l
2
1
0.0 l/min
0.0 l
2
1
0.0 l/min
Setup menu
Calibration
Flowrate alarms Display
A
B
Flowrate constant
Min value Max value
C
1
50000
01230
B
D
A
Fig. 12
Fig. 13
Fig. 14
CONTROLS
SWITCH ON
The owmeter is turned on when powered.
Press the key to see the succession of different values written in full (central part of the display).
Whenever the device is switched on, it will briey display its name and software version.
ACCESS TO SETUP MENU
Press the keys for 2 seconds at the same time to access setup menu.
SELECTION AND ACCESS TO MENU ITEMS
A Press in succession to move through items (the selected item is highlighted with
a black band)
B Press to access the selected item
Three dots under an item show the presence of another setup menu.
MODIFYING DATA
A Press to toggle from one digit to another
B Press in succession to modify the value of the digit highlighted by the cursor
C Press to conrm the change. The display goes back to the previous screen.
D Press to exit current page without conrming changes.
Fig. 15
3.1 First switch on
At rst switch on, DigiWolf will run a guided procedure to set the language, two units of measurement (owrate and volume), and the operating mode (lling or owrate).
Press key to move between items, OK to save and move to the next setting, or ESC return to the previous setting.
9
MENU
MENU STRUCTURE
Fig. 16
10
4 PRELIMINARY SETUP FOR USE
Options
Operating mode
Language Units of measurem.
Filling
2
1
2
Operating mode
Filling
Flowrate
Setup menu
Valve activ. time
Flowrate alarms Display
0.0s
Valve activ. time
Min value Max value
C
0.0
9.9
.00
B
D
A
s
For a correct display of data regarding treatment, some preliminary set ups are necessary when installing DigiWolf on farming machines.
4.1 Operating mode selection (FILLING / FLOWRATE)
DigiWolf can be used to measure the owrate of a uid in a hydraulic circuit ("Flowrate" mode), or to control the lling of a tank by stopping the intro­duction of liquid once the set quantity has been reached ("Filling" mode).
Some menu items will be available or not according to the set mode.
1) Go to the operating mode selection menu (Setup menu > Options > Operating mode).
Under the selected item, the currently set mode will be displayed.
Fig. 17
1) Select the desired operating mode with the key.
2) Press
Fig. 18
OK to save, or ESC to exit without saving.
SETUP
5 USE IN FILLING MODE
5.1 Preliminary setup
5.1.1 Valve activation time
It is possible to set the lling valve activation time, i.e. the time interval between the sending of valve closure signal and the ow actual interruption, so that DigiWolf can calculate the exact time of operation and therefore the quantity introduced. The value to be entered is indicated in the technical sheet of the lling valve installed in the system.
1) Access the valve activation time menu (Setup menu > Valve activ. time).
Under the selected item, the currently set value will be displayed.
Fig. 19
2) Use the relevant keys to set the valve activation time:
A) Press to toggle from one digit to another
B) Press in succession to modify the value of the digit highlighted by the cursor
C) Press to save changes or D) Press to exit current page without conrming
changes.
Fig. 20
CONTINUES
11
Start
¬
¬
Quantity
¬
¬
o
3000.0
o
Start
¬
¬
Quantity
¬
¬
o
3000.0
Quantity
Min Max Value
Value
C
0.1
99999.9
B
D
A
3000.00
l
Start
¬
¬
Quantity
¬
¬
o
5000.0
5000.0 l
o
6.23
21.84 l/min
5000.0 l
o
15.98
Pause
5000.0 l
o
5000.0
Completed
USE (FILLING MODE)
5.2 Flowmeter use (FILLING mode)
The main screen shows the display divided into three horizontal sectors:
Fig. 21
Fig. 22
Fig. 23
- upper sector: The Start item refers to the key on the side, which allows starting the procedure
(OK key);
- central sector: the value represents the set lling quantity (in liters). Next to it appears the symbol
;
- lower sector: The Quantity item refers to the key on the side, which allows setting the value
of tank lling (
ESC key).
1) Keep ESC key pressed for two seconds to access the lling quantity set up menu.
2) Using the suitable keys, set the lling quantity value:
A) Press to toggle from one digit to another
B) Press in succession to modify the value of the digit highlighted by the cursor
C) Press to save changes or D) Press to exit current page without conrming
changes.
Fig. 24
Fig. 25
Fig. 26
3) Keep OK key pressed for two seconds. The lling process starts.
The display shows the following data:
- upper sector: set lling quantity
;
- central sector: partial totalizer;
- lower sector: instant rate
.
To see a datum in full, press the button several times until the desired value is in the central sector of the display.
Displayed data can not be edited.
The tank lling can be interrupted at any time by pressing ESC key. The Pau­se message will appear in the lower part of the display.
Press
OK key to restart the lling process;
Press
ESC key to denitively stop the lling; the display will go back to the main
page.
The lling stops automatically once the set quantity is reached. The Comple- ted message will appear in the lower part of the display.
Fig. 27
12
6 USE IN FLOWRATE MODE
1
0.0 l
2
1
0.0 l/min
6.23 l
2
1
21.84 l/min
6.23
6.23 l
2
1
21.84 l/min
6.23
6.23 l
2
1
21.84 l/min
0.00
50.00 l
2
1
21.84 l/min
999.99
1050.0 l
2
1
21.84 l/min
1000.0
10049 l
2
1
21.84 l/min
9999.9
10050 l
2
1
21.84 l/min
10000
...
6.1 Flowmeter use (FLOWRATE mode)
The main screen shows the display divided into three horizontal sectors.
The sectors showing the symbols
The third data represents instant rate value.
Fig. 28
Fig. 29
6.2 Partial totalizer reset
and 2 represent the partial totalizers, individually resettable.
USE (FLOWRATE MODE)
Partial totalizers
Instantaneous rate
Feed some liquid to the system. The display will start to show the increasing value of the measured quantity of liquid, and the instant rate value.
To reset a totalizer, it is necessary to visualize it in full.
To do this, press the button several times until the value to zero is in the central area of the display.
Fig. 30
Press ESC key for two seconds. The totalizer will zero.
Fig. 31
- Do not place DigiWolf under pressurized water.
- Use the owmeter only within recommended rate range (Tab. 2 - Par. 2.3.1). Beyond these limits, the ow-meter may give out incorrect data.
- To avoid damaging the owmeter, do not exceed the maximum rate except for short periods.
ARAG can not be held responsible for damage caused to persons, animals or things from the incorrect or unintended use of Digiwolf or its parts.
- Do not exceed the max. operation pressure (Tab. 2 - Par. 2.3.1).
Display of the symbol ------ shows that the rate or the totalizer are over the max. displayable value.
Totalizers have oating points and display a max. of 5 gures. Up to 999.99 two decimals are shown, it then drops to 1 and 0 with the tran­sition to thousands and tens of thousands (Fig. 32).
13
Fig. 32
End reading
0.0 l/min
¬
¬
Calibration
67.58 l
¬
¬
C
D
067.580
o
1
2
B
A
1
2
SETUP
7 OTHER SETTINGS
What follows applies to both operating modes of the owmeter.
7.1 Calibration
Rate reading may not be correct due to different system congurations (tubes, valve, etc.). Therefore, we recommend to make a spray test; in case the measured value is different from the real one, perform an automatic calibration procedure or manually calculate owrate constant.
7.1.1 Automatic calibration
Make a quantity of liquid previously measured or that can be measured with another system go through the owmeter. The greater the amount of
liquid used to perform the calibration procedure, the more accurate the calibration.
2) Access Automatic calibration menu (Setup menu > Calibration > Auto calibra­tion). Once inside the menu, the device is ready to start measuring without
any further controls.
Fig. 33
3) Make the liquid go through the system. The display will start to show the incre­asing value of the measured quantity of liquid. Once liquid passage has nished, the value on the display will not change any more.
At this point, press OK. In the lower part of the display, the message Stabi- lization will come up and the device will show the screen on the side.
Fig. 34
Setting of liquid amount actually passed through the owmeter during calibration procedure.
Display of liquid amount read by the owmeter during calibration procedure.
4) Using the keys, enter the values of the liquid quantity previously measured:
A) Press to toggle from one digit to another
B) Press in succession to modify the value of the digit highlighted by the cursor
C) Press to complete calibration procedure, or D) Press for 1 sec. to cancel calibration procedure.
If, after starting calibration, the device does not sense any ow passage (and the display remains in 0), press OK to exit the calibration procedure without saving.
If the device continues to sense the liquid ow after pressing OK, after a few seconds the error message Stop flow!will be displayed. Once the ow has been interrupted, the reading will stabilize as by standard procedure.
14
7.1.2 Manual calibration
Calibration
Man. calibration
Auto calibration
600
Flowrate constant
Min value Max value
C
1
50000
B
D
A
06000
APPROXIMA
TE CONST
ANT
XXX
Flowrate alarms
Min flow alarm
Max flow alarm
6.00/min
Min flow alarm
Min value Max value
0.1
99999.9
OFF
o
Min flow alarm
Min value Max value
C
0.1
99999.9
B
D
A
0040.00
l/min
o
To set the owrate constant manually, calculate and set the correct constant using the following formula:
SETUP
[amount measured by device]
[actually delivered quantity]
Fig. 35
Fig. 36
APPROXIMATE CONSTANT
XXX
Fig. 37
x [constant indicated on flowmeter body]
1) Access Manual calibration menu (Setup menu > Calibration > Man. calibration).
In Calibration menu, selecting Manual calibration, under the item the curren­tly set constant will be displayed.
Press OK key to access value modication.
2) Using the suitable keys, set the value of the owmeter constant:
A) Press to toggle from one digit to another
B) Press in succession to modify the value of the digit highlighted by the cursor
C) Press to save changes or D) Press to exit current page without conrming
changes. Refer to the label on the body (Fig. 37).
The owmeter body must be assembled with the arrow on the label facing the ow direction.
CODE Approximate constant
4627405A 250
4627506A 132
4627707A 64
Tab. 6
7.2 Flowrate alarms
Set the minimum and maximum values beyond which the display must show an alarm message.
1) Access Flowrate alarms menu (Setup menu > Flowrate alarms).
Min. and max. owrate alarms setting is done in the same way.
Under the selected item, the currently set value will be displayed. Press OK key to access modication of the selected item.
Fig. 38
2) To activate the alarm, press
message OFF disappears and owrate alarm value is displayed.
Carry out the same procedure to deactivate owrate alarm again.
Fig. 39
3) Set owrate alarm value:
A) Press to toggle from one digit to another
B) Press in succession to modify the value of the digit highlighted by the cursor
C) Press to save changes or D) Press to exit current page without conrming
changes.
Fig. 40
and
keys at the same time until the
15
Display
Brightness
Contrast
50%
Brightness
50%
1
2
2
Options
Language
Units of measurem.
English
2
1
2
Language
Italiano
English Español Português
SETUP
7.3 Display
Adjust display brightness and contrast.
Fig. 41
Access display menu (Setup menu > Display).
Brightness and contrast adjustment is done in the same way.
Under the selected item, the currently set value will be displayed.
Press
OK key to access modication of the selected item.
Fig. 42
7.4 Options
7.4.1 Language
Set desired language and units of measurement.
Fig. 43
1) Set the value using the key. Each time it is pressed, the value will increase
by 10% until reaching 100%, and then starts back from 0.
2) Press the OK key to save, or ESC to exit without saving.
Access language setup menu (Setup menu > Options > Language).
Under the selected item, the currently set value will be displayed.
Press
OK key to access language selection.
1) Select language using the key.
2) Press the OK key to save, or ESC to exit without saving
.
Fig. 44
16
7.4.2 Units of measurement
Units of measurem.
Flowrate
Volume
l/min
2
1
2
Flowrate
l/min.
GPM m/h
3
Units of measurem.
Volume
Flowrate
l
2
1
2
Volume
l
gal m
3
Set units of measurement for owrate values and volume read by the device.
7.4.2.1 Rate units of measurement
Fig. 45
Fig. 46
7.4.2.2 Volume units of measurement
SETUP
Access instant rate units of measurement menu (Setup menu > Options > Units of measurem. > Flowrate).
Under the selected item, the currently set value will be displayed.
Press
OK key to access owrate units of measurement selection.
1) Select measurement unit using key.
2) Press OK key to save, or ESC to exit without saving.
Fig. 47
Fig. 48
Access volume units of measurement menu (Setup menu > Options > Units of measurem. > Volume).
Under the selected item, the currently set value will be displayed.
Press
OK key to access volume units of measurement selection.
1) Select measurement unit using key.
2) Press OK key to save, or ESC to exit without saving.
17
Test
Display test
Sensor frequency Battery voltage
Test
Keys test
Battery voltage Display test
1
2
1
1
SETUP
7.5 Test
In this menu is it possible to see some information and run a device operation test:
- Firmware version:
the display shows the rmware version installed on the device.
- Sensor frequency:
in the presence of ow passage, the display shows in real time the frequency of the signal from the sensor reading the owrate.
- Battery voltage:
the display shows the device supply voltage level.
7.5.1 Display test
Display test checks the correct operation of the display on the device.
Fig. 49
Access display test menu (Setup menu > Test > Display test).
Press
OK key to access the test.
The display shows all pixels on.
Press
ESC key to return to the previous screen.
Fig. 50
7.5.2 Keys test
Keys test checks the correct operation of the keys on the device.
Fig. 51
Fig. 52
Access keys test menu (Setup menu > Test > Keys test).
Press
OK key to access the test.
1) Pressing one key, the corresponding portion of the display will light up.
2) To exit, press
ESC key: after lighting up the corresponding portion of the
display, it will return to the previous screen.
18
MAINTENANCE
8 CLEANING AND REPAIR
- At the end of each treatment, run clean water through the tubes.
- If necessary, periodically clean or replace the owmeter paddle (Par. 8.2). Periodically clean the paddle anyway.
Do not use metal or abrasive objects to clean the paddle.
Do not use solvents or fuel to clean the case outer surface.
CAUTION: Before each of the following operations, use the following precautions:
1) Wear gloves, goggles and protective clothing.
2) Stop the machine and disconnect the power of the system.
3) Make sure that the system is no longer under pressure.
When replacing the sensor or the paddle, check for the correct reading of the owmeter. Proceed to a new calibration if necessary.
8.1 Troubleshooting
FAULT CAUSE REMEDY
The display is off No power supply
Rate is beyond operation limits of the owmeter
Totalizers do not progress during the passage of liquid.
Displayed owrate is not stable
The sensor is not connected correctly
The paddle-wheel is locked
Presence of turbulence or air in the circuit
Worn paddle • Replace the paddle.
• Check the presence and proper
connection of power supply.
• The model of the owmeter is
not adequate for the owrate to be measured. Replace the owmeter.
• Check the sensor connection.
• Clean or replace the paddle-
wheel group if necessary.
• Check the circuit.
Tab. 7
19
APPROXIMA
TE CONST
ANT
XXX
MAINTENANCE
8.2 Paddle cleaning and replacement
lock
unlock
Fig. 53 Fig. 54 Fig. 55
1) Remove the fork from the monitor using a screwdriver.
2) Remove the monitor from the owmeter body.
3) Unscrew the ring nut in a counter clockwise direction and remove the sensor housing block from the owmeter body.
4) With half a rotation, remove the paddle group from the sensor housing block in a counter clockwise direction.
5) Immerse the paddle group in detergent liquid for several hours.
6) Wash the paddle group thoroughly with running water and check its correct operation. If necessary, replace the complete paddle group with its suitable spare part (code 4626000.500).
7) Ret the paddle group on the electronic sensor with a clockwise rotation until it clicks in place.
2
1
Arrows direction on the body must correspond to the one on the label
Fig. 56
8) Ret the sensor housing block on the owmeter body tightening the ring nut in a clockwise direction until it stops paying attention to keep the arrow on the coupling facing the ow direction.
20
8.3 OR replacement
With the removed sensor housing block (see Par. 8.2) proceed as follows:
1) Remove the fork from the ring nut using a screwdriver.
Fig. 57
2) Remove the ring nut.
- OR replacement: Replace the ORs (Code G10051V - ARAG spare parts catalogue).
3) Ret the sensor housing block ensuring that the fork is correctly inserted in the ring nut.
Fig. 58
MAINTENANCE
21
TECHNICAL DATA
Data Min. Max. Default UoM Notes
Calibration Manual calibration 1 50,000 600 -- --
Min. owrate alarm 0.1 99999.9 OFF l/min.
Flowrate alarms
Max. owrate alarm 0.1 999999.9 OFF l/min.
Brightness 0% 100% 50% % --
Display
Contrast 0% 100% 50% % --
Valve activation time 0.0 9.9 0.0 sec.
Filling
Filling quantity 0.1 99999.9 3000.0 litres
Language - - English -
Options
Rate units of measurement
Volume units of measurement
9 TECHNICAL DATA
Description DigiWolf
Power supply 8 ÷ 32 Vdc
Pump stop enabling voltage = power supply
Max. load current 100 mA
Max. load current - pump stop enabling 200 mA
Short circuit protection
Flowrate signal square wave (0 ÷ V suppl.)
Max. frequency 1,2 KHz
Max. absorption 50 mA (100% backlight)
Working temperature
Storage temperature -30 °C ÷ 80 °C / -22 °F ÷ +176 °F
Weight 700 ÷ 770 g (depending on the type)
The alarm can be deactivated
setting the value "OFF"
The alarm can be deactivated
setting the value "OFF"
Language settings: Italiano, English, Español, Português, Français, Deutsch, Cesky, Polski,
, Русский, Magyar.
ニホン
- - l/min. -
Units of measurement settings: l/min, GPM,
3
m
/h
- - litres - Units of measurement settings: l, gal, m
0 °C ÷ 60 °C
+32 °F ÷ +140 °F
Tab. 9
3
Tab. 8
10 DISPOSAL AT THE END OF SERVICE
Dispose of the system in compliance with the established legislation in the country of use.
22
Notes
23
Notes
24
Notes
25
11 GUARANTEE TERMS
1. ARAG s.r.l. guarantees this apparatus for a period of 360 day (1 year) from the date of sale to the client user (date of the goods delivery note).
The components of the apparatus, that in the unappealable opinion of ARAG are faulty due to an original defect in the
material or production process, will be repaired or replaced free of charge at the nearest Assistance Centre operating at the moment the request for intervention is made. The following costs are excluded:
- disassembly and reassembly of the apparatus from the original system;
- transport of the apparatus to the Assistance Centre.
2. The following are not covered by the guarantee:
- damage caused by transport (scratches, dints and similar);
- damage due to incorrect installation or to faults originating from insufficient or inadequate characteristics of the electri­cal system, or to alterations resulting from environmental, climatic or other conditions;
- damage due to the use of unsuitable chemical products, for spraying, watering, weedkilling or any other crop tre­atment, that may damage the apparatus;
- malfunctioning caused by negligence, mishandling, lack of know how, repairs or modications carried out by unautho­rised personnel;
- incorrect installation and regulation;
- damage or malfunction caused by the lack of ordinary maintenance, such as cleaning of lters, nozzles, etc.;
- anything that can be considered to be normal wear and tear.
3. Repairing the apparatus will be carried out within time limits compatible with the organisational needs of the Assistan­ce Centre.
No guarantee conditions will be recognised for those units or components that have not been previously washed and
cleaned to remove residue of the products used.
4. Repairs carried out under guarantee are guaranteed for one year (360 days) from the replacement or repair date.
5. ARAG will not recognise any further expressed or intended guarantees, apart from those listed here.
No representative or retailer is authorised to take on any other responsibility relative to ARAG products. The period of the guarantees recognised by law, including the commercial guarantees and allowances for special pur-
poses are limited, in length of time, to the validities given here. In no case will ARAG recognise loss of prots, either direct, indirect, special or subsequent to any damage.
6. The parts replaced under guarantee remain the property of ARAG.
7. All safety information present in the sales documents regarding limits in use, performance and product characteristics must be transferred to the end user as a responsibility of the purchaser.
8. Any controversy must be presented to the Reggio Emilia Law Court.
Conformity Declaration
ARAG s.r.l. Via Palladio, 5/A 42048 Rubiera (RE) - Italy P.IVA 01801480359
Dichiara
che il prodotto descrizione: Flussometro a palette
modello: DigiWolf serie: 4627xxx, 4628xxx risponde ai requisiti di conformità contemplati nella seguente Direttiva Europea:
2004/108/CE e successive modificazioni (Compatibilità Elettromagnetica)
Riferimenti alle Norme Applicate:
EN ISO 14982
(Macchine agricole e forestali - Compatibilità elettromagnetica Metodi di prova e criteri di accettazione)
CEI EN 61326-1:2007
(Apparecchi elettrici di misura, controllo e laboratorio, Prescrizioni di compatibilità elettromagnetica - Parte 1 : Prescrizioni generali)
Rubiera, 08 Marzo 2011
Giovanni Montorsi
(Presidente)
Only use original ARAG accessories and spare parts, to maintain safety conditions foreseen by the constructor. Always refer to the ARAG spare parts catalogue.
42048 RUBIERA (Reggio Emilia) - ITALY Via Palladio, 5/A
Tel. +39 0522 622011 Fax +39 0522 628944
http://www.aragnet.com info@aragnet.com
D20245_GB-m01 07/2011
Loading...