This manual is an integral part of the equipment to which it refers and must accompany the equipment in case of sale or
change of ownership. Keep it for future reference; ARAG reserves the right to modify the specications and instructions
regarding the product at any time and without prior notice.
2.3.1 Hydraulic connection for brass/fork connections .................................................................................................... 6
2.4 Electric connections ...................................................................................................................................................7
3 Controls in the menu .........................................................................................................................................................9
3.1 First switch on .............................................................................................................................................................9
Menu structure ................................................................................................................................................................. 10
4 Preliminary setup for use................................................................................................................................................ 11
5 Use in FILLING mode ...................................................................................................................................................... 11
5.1.1 Valve activation time .............................................................................................................................................. 11
5.2 Flowmeter use (FILLING mode) ...............................................................................................................................12
6 Use in FLOWRATE mode ................................................................................................................................................ 13
6.1 Flowmeter use (FLOWRATE mode) .........................................................................................................................13
7 Other settings .................................................................................................................................................................. 14
7.4.1 Language .............................................................................................................................................................. 16
7.4.2 Units of measurement ........................................................................................................................................... 17
7.4.2.1 Rate units of measurement ................................................................................................................................... 17
7.4.2.2 Volume units of measurement ............................................................................................................................... 17
7.5 Test ...........................................................................................................................................................................18
7.5.1 Display test ............................................................................................................................................................ 18
7.5.2 Keys test ................................................................................................................................................................ 18
8 Cleaning and repair ......................................................................................................................................................... 19
8.2 Paddle cleaning and replacement ............................................................................................................................20
8.3 OR replacement .......................................................................................................................................................21
9 Technical data ................................................................................................................................................................. 22
10 Disposal at the end of service ....................................................................................................................................... 22
DigiWolf is a paddle owmeter capable of measuring the rate of a uid in a hydraulic circuit, and it can display the results of these measurements.
The device can also be set to control the lling of a tank by stopping the introduction of liquid once the required quantity has been reached (this function
is possible only by connecting the owmeter to the Pump Stop Module or the relevant cable - Par. 2.4-2.5).
1.1 Intended use
This device is designed to work on agricultural machinery for crop spraying applications.
The machine is designed and built in compliance with EN ISO 14982 standard (Electromagnetic compatibility - Forestry and
farming machines), harmonized with 2004/108/EC Directive.
The owmeter must not be used to measure the passage of hydrocarbons, ammable, explosive or toxic liquids. The owmeter
is not suitable for contact with liquids for human consumption.
2 FLOWMETER ASSEMBLY
Install the owmeter at least 20 cm from the elements that could cause turbulence inside the tubes (valves, bends, constrictions, etc.).
The owmeter can be installed in a vertical or horizontal position.
CAUTION:
- Do not t the owmeter with the display facing down (Fig. 1).
- The system must have a ltering element with a lter of at least 50 mesh, together with a safety valve to limit use pressure at
the specied max. valve (Tab. 2 - Par. 2.3.1).
OK
Assemble the owmeter using the suitable mounting parts (Fig. 2): t the bolts (M8) in their seats, then make them slide to their stop position to prevent
them from coming out.
OKOK
NO!
Fig. 1
The owmeter body must be assembled with the arrow on the label facing the ow direction.
APPROXIMATE CONSTANT
XXX
Fig. 2
4
2.1 Monitor rotation
Fig. 3
INSTALLATION
In case of vertical assembly, to simplify the use of the DigiWolf it is possible to rotate the
monitor by 90° with respect to the body (Fig. 3).
1) Remove the fork from the monitor using a screwdriver.
2) Remove the monitor from the owmeter body.
Fig. 4
Fig. 5
Fig. 6
If the monitor is turned upside down, the letters A and B can be seen: these correspond to
the two possible monitor positions (parallel and perpendicular to the body).
DigiWolf is supplied with the sensor in position A (parallel to the body).
With the aid of a screwdriver, loosen the screw locking the sensor without removing it.
Rotate the sensor anti-clockwise until reaching B stop position.
Fig. 7
Tighten back the screw without forcing it. The monitor is ready to be tted in perpendicular position to the body.
Fit the monitor back on the owmeter body.
5
INSTALLATION
2.2 Dimensions (mm)
A
T
2.3 Hydraulic connections
E
120
B
D
C
143
CODEABCDET
4627405A1401282053141T5
4627506A145 1482103645T6
4627707A157 162 2224145T7
Tab. 1
95
Fig. 8
2.3.1 Hydraulic connection for brass/fork connections
Avoid bends and constrictions before connections and on tubes.
Use ARAG connections with their suitable OR with MALE CONNECTIONS [T connections - General Catalogue (Tab. 2)].
Do not use the owmeter at pressures greater than those listed in Table 2.
Regarding connections, use tubes and ttings properly sized for the operating pressure of the system.
CAUTION: For the implementation on already operating systems it is necessary to follow all safety rules described herein.
System assembly and start-up must be carried out by expert personnel according to the safety rules so as to ensure the same
safety level of the system the owmeter is going to be installed in.
After connection, check for the perfect sealing of the tubes and fork connections.
Fig. 9
CODEl/min.US GPM Connection Ø equivalent (inch)
4627405A 10-2002.6-53T5 F1 1/4"20290
4627506A 20-4005-106T6 F1 1/2"12174
4627707A 40-80010-210T7 F2"7130
Max. p
(bar)
Max. p
(PSI)
Tab. 2
The diameter in inches (Ø equivalent) is given only as an indication of the typical passage of the owmeter body.
Actually, it is possible to choose different sizes depending on the fork connection used.
6
INSTALLATION
2.4 Electric connections
DigiWolf is designed to operate independently, but may be connected to ARAG devices (computers, monitors, displays) or equipment from
other manufacturers as long as they are set up for use with owmeters of this type.
DigiWolf can be powered by means of a generic power supply cable (cod. 4622BA50000.100 - not supplied), which does not allow the
operator to control lling functions directly from the device. Connections are listed in Table 3:
Cable colourPositionConnection
Grey1GND
Brown2+12 VDC
Yellow/Green 3
Black4Pump control
For the use as lling owmeter, the DigiWolf must be connected to ARAG devices only.
IN CASE OF ELECTRIC ARC WELDING, MAKE SURE THAT THE POWER SUPPLY TO THE DEVICE IS DISCONNECTED; IF NECESSARY, DISCONNECT THE POWER SUPPLY CABLES.
ARAG is not liable for damages to the equipment, persons, animals or things caused by a wrong or unsuitable installation of the
owmeter. Failure to observe the above instructions automatically voids the warranty.
The DigiWolf connections on devices other than ARAG are indicated in Tab. 4.
Signal
(square wave)
Tab. 3
PositionConnection
4
3
2
1
Fig. 10
1GND
2+12 VDC
3
4Pump control
Signal
(square wave)
Tab. 4
To connect DigiWolf to Bravo 300/400 series computers and monitors use the "Pump Stop Kit" (code 4622BA50000.200) or the adapter
cable (code 4622BA50000.110 - ARAG Catalogue), which allows for connecting the DigiWolf 4-pole connector to the suitable Bravo
series computer 3-pole connector.
When connecting DigiWolf with external devices (Bravo 300/400, etc.), enter IN SUCH DEVICES the approximate constant according to the table below:
For any information about the connections with the Pump Stop Module (code 4622BA50000.200 - ARAG Catalogue), consult the rele-
vant use and maintenance manual (D20059), that can be downloaded at www.aragnet.com, in the Manuals section.
Fig. 11
In case the system features an ARAG 453 series 3-wire valve, we recommend to use an adapter cable (code 46220CBU01.100 - ARAG
Catalogue) which will replace the Pump Stop Module.
8
3 CONTROLS IN THE MENU
0.0 l
2
1
0.0 l/min
0.0 l
2
1
0.0 l/min
Setup menu
Calibration
Flowrate alarms
Display
A
B
Flowrate constant
Min value
Max value
C
1
50000
01230
B
D
A
Fig. 12
Fig. 13
Fig. 14
CONTROLS
SWITCH ON
The owmeter is turned on when powered.
Press the key to see the succession of different values written in full (central part
of the display).
Whenever the device is switched on, it will briey display its name and
software version.
ACCESS TO SETUP MENU
Press the keys for 2 seconds at the same time to access setup menu.
SELECTION AND ACCESS TO MENU ITEMS
A Press in succession to move through items (the selected item is highlighted with
a black band)
B Press to access the selected item
Three dots under an item show the presence of another setup menu.
MODIFYING DATA
A Press to toggle from one digit to another
B Press in succession to modify the value of the digit highlighted by the cursor
C Press to conrm the change. The display goes back to the previous screen.
D Press to exit current page without conrming changes.
Fig. 15
3.1 First switch on
At rst switch on, DigiWolf will run a guided procedure to set the language, two units of measurement (owrate and volume), and
the operating mode (lling or owrate).
Press key to move between items, OK to save and move to the next setting, or ESC return to the previous setting.
9
MENU
MENU STRUCTURE
Fig. 16
10
4 PRELIMINARY SETUP FOR USE
Options
Operating mode
Language
Units of measurem.
Filling
2
1
2
Operating mode
Filling
Flowrate
Setup menu
Valve activ. time
Flowrate alarms
Display
0.0s
Valve activ. time
Min value
Max value
C
0.0
9.9
.00
B
D
A
s
For a correct display of data regarding treatment, some preliminary set ups are necessary when installing DigiWolf on farming machines.
4.1 Operating mode selection (FILLING / FLOWRATE)
DigiWolf can be used to measure the owrate of a uid in a hydraulic circuit ("Flowrate" mode), or to control the lling of a tank by stopping the introduction of liquid once the set quantity has been reached ("Filling" mode).
Some menu items will be available or not according to the set mode.
1) Go to the operating mode selection menu (Setup menu > Options > Operating
mode).
Under the selected item, the currently set mode will be displayed.
Fig. 17
1) Select the desired operating mode with the key.
2) Press
Fig. 18
OK to save, or ESC to exit without saving.
SETUP
5 USE IN FILLING MODE
5.1 Preliminary setup
5.1.1 Valve activation time
It is possible to set the lling valve activation time, i.e. the time interval between the sending of valve closure signal and the ow actual interruption,
so that DigiWolf can calculate the exact time of operation and therefore the quantity introduced.
The value to be entered is indicated in the technical sheet of the lling valve installed in the system.
1) Access the valve activation time menu (Setup menu > Valve activ. time).
Under the selected item, the currently set value will be displayed.
Fig. 19
2) Use the relevant keys to set the valve activation time:
A) Press to toggle from one digit to another
B) Press in succession to modify the value of the digit highlighted by the cursor
C) Press to save changes or D) Press to exit current page without conrming
changes.
Fig. 20
CONTINUES
11
Start
¬
¬
Quantity
¬
¬
o
3000.0
o
Start
¬
¬
Quantity
¬
¬
o
3000.0
Quantity
Min
Max Value
Value
C
0.1
99999.9
B
D
A
3000.00
l
Start
¬
¬
Quantity
¬
¬
o
5000.0
5000.0 l
o
6.23
21.84 l/min
5000.0 l
o
15.98
Pause
5000.0 l
o
5000.0
Completed
USE (FILLING MODE)
5.2 Flowmeter use (FILLING mode)
The main screen shows the display divided into three horizontal sectors:
Fig. 21
Fig. 22
Fig. 23
- upper sector:
The Start item refers to the key on the side, which allows starting the procedure
(OK key);
- central sector:
the value represents the set lling quantity (in liters).
Next to it appears the symbol
;
- lower sector:
The Quantity item refers to the key on the side, which allows setting the value
of tank lling (
ESC key).
1) Keep ESC key pressed for two seconds to access the lling quantity set up
menu.
2) Using the suitable keys, set the lling quantity value:
A) Press to toggle from one digit to another
B) Press in succession to modify the value of the digit highlighted by the cursor
C) Press to save changes or D) Press to exit current page without conrming
changes.
Fig. 24
Fig. 25
Fig. 26
3) Keep OK key pressed for two seconds. The lling process starts.
The display shows the following data:
- upper sector: set lling quantity
;
- central sector: partial totalizer;
- lower sector: instant rate
.
To see a datum in full, press the button several times until the desired value is in
the central sector of the display.
Displayed data can not be edited.
The tank lling can be interrupted at any time by pressing ESC key. The Pause message will appear in the lower part of the display.
Press
OK key to restart the lling process;
Press
ESC key to denitively stop the lling; the display will go back to the main
page.
The lling stops automatically once the set quantity is reached. The Comple-ted message will appear in the lower part of the display.
Fig. 27
12
6 USE IN FLOWRATE MODE
1
0.0 l
2
1
0.0 l/min
6.23 l
2
1
21.84 l/min
6.23
6.23 l
2
1
21.84 l/min
6.23
6.23 l
2
1
21.84 l/min
0.00
50.00 l
2
1
21.84 l/min
999.99
1050.0 l
2
1
21.84 l/min
1000.0
10049 l
2
1
21.84 l/min
9999.9
10050 l
2
1
21.84 l/min
10000
...
6.1 Flowmeter use (FLOWRATE mode)
The main screen shows the display divided into three horizontal sectors.
The sectors showing the symbols
The third data represents instant rate value.
Fig. 28
Fig. 29
6.2 Partial totalizer reset
and 2 represent the partial totalizers, individually resettable.
USE (FLOWRATE MODE)
Partial totalizers
Instantaneous rate
Feed some liquid to the system. The display will start to show the increasing value
of the measured quantity of liquid, and the instant rate value.
To reset a totalizer, it is necessary to visualize it in full.
To do this, press the button several times until the value to zero is in the central
area of the display.
Fig. 30
Press ESC key for two seconds. The totalizer will zero.
Fig. 31
- Do not place DigiWolf under pressurized water.
- Use the owmeter only within recommended rate range (Tab. 2 - Par. 2.3.1). Beyond these limits, the ow-meter may give out incorrect data.
- To avoid damaging the owmeter, do not exceed the maximum rate except for short periods.
ARAG can not be held responsible for damage caused to persons, animals or things from the incorrect or unintended use of Digiwolf or
its parts.
- Do not exceed the max. operation pressure (Tab. 2 - Par. 2.3.1).
Display of the symbol ------ shows that the rate or the totalizer are over the max. displayable value.
Totalizers have oating points and display a max. of 5 gures. Up to 999.99 two decimals are shown, it then drops to 1 and 0 with the transition to thousands and tens of thousands (Fig. 32).
13
Fig. 32
End reading
0.0 l/min
¬
¬
Calibration
67.58 l
¬
¬
C
D
067.580
o
1
2
B
A
1
2
SETUP
7 OTHER SETTINGS
What follows applies to both operating modes of the owmeter.
7.1 Calibration
Rate reading may not be correct due to different system congurations (tubes, valve, etc.). Therefore, we recommend to make
a spray test; in case the measured value is different from the real one, perform an automatic calibration procedure or manually
calculate owrate constant.
7.1.1 Automatic calibration
Make a quantity of liquid previously measured or that can be measured with another system go through the owmeter. The greater the amount of
liquid used to perform the calibration procedure, the more accurate the calibration.
2) Access Automatic calibration menu (Setup menu > Calibration > Auto calibration). Once inside the menu, the device is ready to start measuring without
any further controls.
Fig. 33
3) Make the liquid go through the system. The display will start to show the increasing value of the measured quantity of liquid. Once liquid passage has nished,
the value on the display will not change any more.
At this point, press OK. In the lower part of the display, the message Stabi-lization will come up and the device will show the screen on the side.
Fig. 34
Setting of liquid amount actually passed through the owmeter during calibration procedure.
Display of liquid amount read by the owmeter during calibration procedure.
4) Using the keys, enter the values of the liquid quantity previously measured:
A) Press to toggle from one digit to another
B) Press in succession to modify the value of the digit highlighted by the cursor
C) Press to complete calibration procedure, or D) Press for 1 sec. to cancel calibration procedure.
If, after starting calibration, the device does not sense any ow passage (and the display remains in 0), press OK to exit the calibration
procedure without saving.
If the device continues to sense the liquid ow after pressing OK, after a few seconds the error message Stop flow!will be displayed.
Once the ow has been interrupted, the reading will stabilize as by standard procedure.
14
7.1.2 Manual calibration
Calibration
Man. calibration
Auto calibration
600
Flowrate constant
Min value
Max value
C
1
50000
B
D
A
06000
APPROXIMA
TE CONST
ANT
XXX
Flowrate alarms
Min flow alarm
Max flow alarm
6.00/min
Min flow alarm
Min value
Max value
0.1
99999.9
OFF
o
Min flow alarm
Min value
Max value
C
0.1
99999.9
B
D
A
0040.00
l/min
o
To set the owrate constant manually, calculate and set the correct constant using the following formula:
SETUP
[amount measured by device]
[actually delivered quantity]
Fig. 35
Fig. 36
APPROXIMATE CONSTANT
XXX
Fig. 37
x [constant indicated on flowmeter body]
1) Access Manual calibration menu (Setup menu > Calibration > Man. calibration).
In Calibration menu, selecting Manual calibration, under the item the currently set constant will be displayed.
Press OK key to access value modication.
2) Using the suitable keys, set the value of the owmeter constant:
A) Press to toggle from one digit to another
B) Press in succession to modify the value of the digit highlighted by the cursor
C) Press to save changes or D) Press to exit current page without conrming
changes. Refer to the label on the body (Fig. 37).
The owmeter body must be assembled with the arrow on the label
facing the ow direction.
CODEApproximate constant
4627405A250
4627506A132
4627707A64
Tab. 6
7.2 Flowrate alarms
Set the minimum and maximum values beyond which the display must show an alarm message.
1) Access Flowrate alarms menu (Setup menu > Flowrate alarms).
Min. and max. owrate alarms setting is done in the same way.
Under the selected item, the currently set value will be displayed.
PressOKkey to access modication of the selected item.
Fig. 38
2) To activate the alarm, press
message OFF disappears and owrate alarm value is displayed.
Carry out the same procedure to deactivate owrate alarm again.
Fig. 39
3) Set owrate alarm value:
A) Press to toggle from one digit to another
B) Press in succession to modify the value of the digit highlighted by the cursor
C) Press to save changes or D) Press to exit current page without conrming
changes.
Fig. 40
and
keys at the same time until the
15
Display
Brightness
Contrast
50%
Brightness
50%
1
2
2
Options
Language
Units of measurem.
English
2
1
2
Language
Italiano
English
Español
Português
SETUP
7.3 Display
Adjust display brightness and contrast.
Fig. 41
Access display menu (Setup menu > Display).
Brightness and contrast adjustment is done in the same way.
Under the selected item, the currently set value will be displayed.
Press
OK key to access modication of the selected item.
Fig. 42
7.4 Options
7.4.1 Language
Set desired language and units of measurement.
Fig. 43
1) Set the value using the key. Each time it is pressed, the value will increase
by 10% until reaching 100%, and then starts back from 0.
2) Press the OK key to save, or ESC to exit without saving.
Access language setup menu (Setup menu > Options > Language).
Under the selected item, the currently set value will be displayed.
Press
OK key to access language selection.
1) Select language using the key.
2) Press the OK key to save, or ESC to exit without saving
.
Fig. 44
16
7.4.2 Units of measurement
Units of measurem.
Flowrate
Volume
l/min
2
1
2
Flowrate
l/min.
GPM
m/h
3
Units of measurem.
Volume
Flowrate
l
2
1
2
Volume
l
gal
m
3
Set units of measurement for owrate values and volume read by the device.
7.4.2.1 Rate units of measurement
Fig. 45
Fig. 46
7.4.2.2 Volume units of measurement
SETUP
Access instant rate units of measurement menu (Setup menu > Options > Units
of measurem. > Flowrate).
Under the selected item, the currently set value will be displayed.
Press
OK key to access owrate units of measurement selection.
1) Select measurement unit using key.
2) Press OK key to save, or ESC to exit without saving.
Fig. 47
Fig. 48
Access volume units of measurement menu (Setup menu > Options > Units of
measurem. > Volume).
Under the selected item, the currently set value will be displayed.
Press
OK key to access volume units of measurement selection.
1) Select measurement unit using key.
2) Press OK key to save, or ESC to exit without saving.
17
Test
Display test
Sensor frequency
Battery voltage
Test
Keys test
Battery voltage
Display test
1
2
1
1
SETUP
7.5 Test
In this menu is it possible to see some information and run a device operation test:
- Firmware version:
the display shows the rmware version installed on the device.
- Sensor frequency:
in the presence of ow passage, the display shows in real time the frequency of the signal from the sensor reading the owrate.
- Battery voltage:
the display shows the device supply voltage level.
7.5.1 Display test
Display test checks the correct operation of the display on the device.
Fig. 49
Access display test menu (Setup menu > Test > Display test).
Press
OK key to access the test.
The display shows all pixels on.
Press
ESC key to return to the previous screen.
Fig. 50
7.5.2 Keys test
Keys test checks the correct operation of the keys on the device.
Fig. 51
Fig. 52
Access keys test menu (Setup menu > Test > Keys test).
Press
OK key to access the test.
1) Pressing one key, the corresponding portion of the display will light up.
2) To exit, press
ESC key: after lighting up the corresponding portion of the
display, it will return to the previous screen.
18
MAINTENANCE
8 CLEANING AND REPAIR
- At the end of each treatment, run clean water through the tubes.
- If necessary, periodically clean or replace the owmeter paddle (Par. 8.2). Periodically clean the paddle anyway.
Do not use metal or abrasive objects to clean the paddle.
Do not use solvents or fuel to clean the case outer surface.
CAUTION:
Before each of the following operations, use the following precautions:
1) Wear gloves, goggles and protective clothing.
2) Stop the machine and disconnect the power of the system.
3) Make sure that the system is no longer under pressure.
When replacing the sensor or the paddle, check for the correct reading of the owmeter. Proceed to a new calibration if necessary.
8.1 Troubleshooting
FAULTCAUSEREMEDY
The display is offNo power supply
Rate is beyond operation limits
of the owmeter
Totalizers do not progress
during the passage of liquid.
Displayed owrate is not
stable
The sensor is not connected
correctly
The paddle-wheel is locked
Presence of turbulence or air in
the circuit
Worn paddle• Replace the paddle.
• Check the presence and proper
connection of power supply.
• The model of the owmeter is
not adequate for the owrate to be
measured.
Replace the owmeter.
• Check the sensor connection.
• Clean or replace the paddle-
wheel group if necessary.
• Check the circuit.
Tab. 7
19
APPROXIMA
TE CONST
ANT
XXX
MAINTENANCE
8.2 Paddle cleaning and replacement
lock
unlock
Fig. 53Fig. 54Fig. 55
1) Remove the fork from the monitor using a screwdriver.
2) Remove the monitor from the owmeter body.
3) Unscrew the ring nut in a counter clockwise direction and remove the sensor housing block from the owmeter body.
4) With half a rotation, remove the paddle group from the sensor housing block in a counter clockwise direction.
5) Immerse the paddle group in detergent liquid for several hours.
6) Wash the paddle group thoroughly with running water and check its correct operation. If necessary, replace the complete paddle group with its suitable spare part (code 4626000.500).
7) Ret the paddle group on the electronic sensor with a clockwise rotation until it clicks in place.
2
1
Arrows direction on the body must
correspond to the one on the label
Fig. 56
8) Ret the sensor housing block on the owmeter body tightening the ring nut in a clockwise direction until it stops paying attention to keep the
arrow on the coupling facing the ow direction.
20
8.3 OR replacement
With the removed sensor housing block (see Par. 8.2) proceed as follows:
1) Remove the fork from the ring nut using a screwdriver.
Fig. 57
2) Remove the ring nut.
- OR replacement:
Replace the ORs (Code G10051V - ARAG spare parts catalogue).
3) Ret the sensor housing block ensuring that the fork is correctly inserted in the ring nut.
Fig. 58
MAINTENANCE
21
TECHNICAL DATA
DataMin.Max.DefaultUoMNotes
CalibrationManual calibration150,000600----
Min. owrate alarm0.199999.9OFFl/min.
Flowrate alarms
Max. owrate alarm0.1999999.9OFFl/min.
Brightness0%100%50%%--
Display
Contrast0%100%50%%--
Valve activation time0.09.90.0sec.
Filling
Filling quantity0.199999.93000.0litres
Language--English-
Options
Rate units of
measurement
Volume units of
measurement
9 TECHNICAL DATA
DescriptionDigiWolf
Power supply8 ÷ 32 Vdc
Pump stop enabling voltage= power supply
Max. load current100 mA
Max. load current - pump stop enabling200 mA
Short circuit protection•
Flowrate signalsquare wave (0 ÷ V suppl.)
Max. frequency1,2 KHz
Max. absorption50 mA (100% backlight)
Working temperature
Storage temperature-30 °C ÷ 80 °C / -22 °F ÷ +176 °F
Weight700 ÷ 770 g (depending on the type)
The alarm can be deactivated
setting the value "OFF"
The alarm can be deactivated
setting the value "OFF"
Language settings: Italiano, English, Español,
Português, Français, Deutsch, Cesky, Polski,
, Русский, Magyar.
ニホン
--l/min.-
Units of measurement settings: l/min, GPM,
3
m
/h
--litres-Units of measurement settings: l, gal, m
0 °C ÷ 60 °C
+32 °F ÷ +140 °F
Tab. 9
3
Tab. 8
10 DISPOSAL AT THE END OF SERVICE
Dispose of the system in compliance with the established legislation in the country of use.
22
Notes
23
Notes
24
Notes
25
11 GUARANTEE TERMS
1. ARAG s.r.l. guarantees this apparatus for a period of 360 day (1 year) from the date of sale to the client user (date of the
goods delivery note).
The components of the apparatus, that in the unappealable opinion of ARAG are faulty due to an original defect in the
material or production process, will be repaired or replaced free of charge at the nearest Assistance Centre operating
at the moment the request for intervention is made. The following costs are excluded:
- disassembly and reassembly of the apparatus from the original system;
- transport of the apparatus to the Assistance Centre.
2. The following are not covered by the guarantee:
- damage caused by transport (scratches, dints and similar);
- damage due to incorrect installation or to faults originating from insufficient or inadequate characteristics of the electrical system, or to alterations resulting from environmental, climatic or other conditions;
- damage due to the use of unsuitable chemical products, for spraying, watering, weedkilling or any other crop treatment, that may damage the apparatus;
- malfunctioning caused by negligence, mishandling, lack of know how, repairs or modications carried out by unauthorised personnel;
- incorrect installation and regulation;
- damage or malfunction caused by the lack of ordinary maintenance, such as cleaning of lters, nozzles, etc.;
- anything that can be considered to be normal wear and tear.
3. Repairing the apparatus will be carried out within time limits compatible with the organisational needs of the Assistance Centre.
No guarantee conditions will be recognised for those units or components that have not been previously washed and
cleaned to remove residue of the products used.
4. Repairs carried out under guarantee are guaranteed for one year (360 days) from the replacement or repair date.
5. ARAG will not recognise any further expressed or intended guarantees, apart from those listed here.
No representative or retailer is authorised to take on any other responsibility relative to ARAG products.
The period of the guarantees recognised by law, including the commercial guarantees and allowances for special pur-
poses are limited, in length of time, to the validities given here. In no case will ARAG recognise loss of prots, either
direct, indirect, special or subsequent to any damage.
6. The parts replaced under guarantee remain the property of ARAG.
7. All safety information present in the sales documents regarding limits in use, performance and product characteristics
must be transferred to the end user as a responsibility of the purchaser.
8. Any controversy must be presented to the Reggio Emilia Law Court.
che il prodotto
descrizione: Flussometro a palette
modello: DigiWolf
serie: 4627xxx, 4628xxx
risponde ai requisiti di conformità contemplati nella seguente Direttiva Europea:
2004/108/CE e successive modificazioni
(Compatibilità Elettromagnetica)
Riferimenti alle Norme Applicate:
EN ISO 14982
(Macchine agricole e forestali - Compatibilità elettromagnetica
Metodi di prova e criteri di accettazione)
CEI EN 61326-1:2007
(Apparecchi elettrici di misura, controllo e laboratorio, Prescrizioni di
compatibilità elettromagnetica - Parte 1 : Prescrizioni generali)
Rubiera, 08 Marzo 2011
Giovanni Montorsi
(Presidente)
Only use original ARAG accessories and spare parts, to maintain safety conditions foreseen by the constructor. Always refer
to the ARAG spare parts catalogue.
42048 RUBIERA (Reggio Emilia) - ITALY
Via Palladio, 5/A
Tel. +39 0522 622011
Fax +39 0522 628944
http://www.aragnet.com
info@aragnet.com
D20245_GB-m01 07/2011
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.