Analog Devices AD1385TD-883B, AD1385TD, AD1385KD Datasheet

a
FEATURES 16-Bit Resolution 500 kHz Sampling Rate Differential Linearity Autocalibration Specified over –558C to +1258C Range SNR 90 dB @ 100 kHz (min) THD –88 dB @ 100 kHz (min)
0.0006% FSR DNL (typ)
0.0015% FSR INL (typ) No Missing Codes 65, 610 V Bipolar Input Ranges Zero Offset Autocalibration
APPLICATIONS Medical Imaging
CAT
Magnetic Resonance Radar Vibration Analysis Parametric Measurement Unit (ATE) Digital Storage Oscilloscopes Waveform Recorders Analytical Instruments
Wide Temperature Range Sampling ADC
AD1385
FUNCTIONAL BLOCK DIAGRAM
PRODUCT DESCRIPTION
The AD1385 is a complete 500 kHz, 16-bit, sampling analog­to-digital converter contained in a single package. Its differential linearity autocalibration feature allows this high resolution, high speed converter to offer outstanding noise and distortion perfor­mance, as well as excellent INL and DNL specifications, over the full military temperature range. Autocalibration effectively eliminates DNL drift over temperature.
The AD1385 architecture includes a low noise, low distortion track/hold, a three pass digitally corrected subranging ADC, and linearity calibration circuitry. A complete linearity calibration requires only 15 ms. Precision thin-film resistors and a propri­etary DAC contribute to the part’s outstanding dynamic and static performance.
REV. 0
Information furnished by Analog Devices is believed to be accurate and reliable. However, no responsibility is assumed by Analog Devices for its use, nor for any infringements of patents or other rights of third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of Analog Devices.
The AD1385 uses four power supplies, ± 5 V and ±15 V, and an external 10 MHz clock. Power dissipation is nominally 2.76 W. Two user selectable bipolar input ranges, ± 5 V and ±10 V, are provided. Careful attention to grounding and a single package make it easy to design PCBs to achieve specified performance.
The AD1385’s pinout is nearly identical to that of the AD1382, a factory calibrated 16-bit, 500 kHz SADC. Just two additional connections, to enable and monitor autocalibration, are required. This commonality provides an easy upgrade path to extend sys­tem performance and operating temperature range.
One Technology Way, P.O. Box 9106, Norwood, MA 02062-9106, U.S.A. Tel: 617/329-4700 Fax: 617/326-8703
AD1385–SPECIFICA TIONS
(TA = +258C, VS = 615 V, VDD = +5 V, VSS = –5 V, 10 MHz External Clock, unless otherwise noted)
Parameter Min Typ Max Min Typ Max Units
AD1385KD AD1385TD
RESOLUTION 16 16 Bits ANALOG INPUT
Input Ranges ±5, ±10 ±5, ±10 V Input Impedance 2.45 2.5 2.55 2.45 2.5 2.55 k
TRANSFER CHARACTERISTICS
(Combined ADC/Track/Hold) Integral Nonlinearity Differential Nonlinearity
Drift, T
Drift, T
MIN
4
MIN
Missing Codes, T Gain Error
Bipolar Zero
Drift, T
MIN
to T
to T
4
to T
MAX
MIN
MAX
MAX
1, 2
, T
to T
1
MIN
MAX
to T
MAX
±0.0015 ±0.0015 % FSR ±0.0006 ±0.0015 ±0.0006 ±0.0015 % FSR
0.3 0.3 ppm/°C None None
±0.05 ±0.15 ±0.05 ±0.15 % FSR 8 15 8 15 ppm/°C ±0.05 ±0.10 ±0.05 ±0.10 % FSR
5 15 5 15 ppm/°C PSRR ±0.006 ±0.10 ±0.006 ±0.10 % FSR/V Noise 70 70 µV rms
MIN
2
to T
MAX
DYNAMIC CHARACTERISTICS
±5 V FSR, VIN = –0.4 dB, T
Sample Rate 500 500 kHz Signal-to-Noise Ratio
5
f = 5 kHz 90 93 90 93 dB f = 100 kHz 90 92 90 92 dB f = 200 kHz 88 91 88 91 dB
Peak Distortion
f = 5 kHz –90 –107 –90 –107 dB f = 100 kHz –88 –95 –88 –95 dB f = 200 kHz –82 –88 –82 –88 dB
Total Harmonic Distortion
6
f = 5 kHz –90 –105 –90 –105 dB f = 100 kHz –88 –95 –88 –95 dB f = 200 kHz –82 –88 –82 –88 dB
MIN
2
to T
MAX
DYNAMIC CHARACTERISTICS
±10 V FSR, VIN = –0.4 dB, T
Sample Rate 500 500 kHz Signal-to-Noise Ratio
5
f = 5 kHz 90 95 90 95 dB f = 100 kHz 90 94 90 94 dB f = 200 kHz 88 93 88 93 dB
Peak Distortion
f = 5 kHz –90 –108 –90 –108 dB f = 100 kHz –80 –87 –80 –87 dB f = 200 kHz –74 –82 –74 –82 dB
Total Harmonic Distortion
6
f = 5 kHz –90 –105 –90 –105 dB f = 100 kHz –80 –87 –80 –87 dB f = 200 kHz –74 –82 –74 –82 dB
DIGITAL INPUTS
Input Voltage
V
IL
V
IH
2.25 2.25 V
0.8 0.8 V
Input Current ±200 ±200 µA Input Capacitance 2 2 pF Clock
Frequency 2.5–10 2.5–10 MHz Duty Cycle 40-60 40-60 %
Aperture Delay
7
77ns
DIGITAL OUTPUTS
Output Voltage
@ IOL = 3.2 mA 0.2 0.4 0.2 0.4 V
V
OL
@ IOH = –3.2 mA 2.4 4.5 2.4 4.5 V
V
OH
Output Capacitance 4 4 pF Leakage, Outputs Disabled ±200 ±200 µA
3
–2–
REV. 0
AD1385
Parameter Min Typ Max Min Typ Max Units
AD1385KD AD1385TD
OUTPUT CODING Complementary Offset Binary or Complementary Twos Complement INTERNAL REFERENCE
Voltage 9.990 10.010 9.990 10.010 V Current 2 5 2 5 mA Drift 5 15 5 15 ppm/°C
TEMPERATURE RANGE, CASE
Specified 0 +70 –55 +125 °C Storage –65 +150 –65 +150 °C
POWER REQUIREMENTS
Specified Operating Range
±V +V –V
S DD
SS
14.25 15.75 14.25 15.75 V
4.75 5.25 4.75 5.25 V –5.25 –4.75 –5.25 –4.75 V
Current Drains
+V –V +V –V
S
S
DD
SS
52 80 52 80 mA 48 75 48 75 mA 104 160 104 160 mA 148 200 148 200 mA
Power Dissipation 2.76 4.125 2.76 4.125 Watts
NOTES
1
Integral linearity is inferred from FFTs. Differential linearity is derived from histograms.
2
Performance over temperature is specified at the temperature at which the last calibration was performed.
3
FSR = Full-Scale Range.
4
Adjustable to zero.
5
SNR excludes harmonics 2-9 of the fundamental.
6
THD includes harmonics 2-9 of the fundamental.
7
Aperture delay is the time from the rising edge on the Hold Command Input to the opening of the switch in the Track/Hold.
Specifications subject to change without notice.
1, 2
TIMING SPECIFICATIONS
(TA = –558C to +1258C, VS = 615 V, VDD = +5 V, VSS = –5 V)
Parameter Design Minimum Typ Unit Description
START COMMAND
t
SCS
t
SCH
10 ns Setup Time 10 ns Hold Time
AUTOZERO
t
AZS
t
AZH
DATA VALID
t
DVS
t
DVH
HOLD COMMAND
t
H
t
D
DATA STROBE
t
DS
t
DSD
10 ns Setup Time 20 ns Hold Time
1.5 CP
0.5 CP
13 CP
3 3
3
Setup Time Hold Time
Hold Time
7 ns Delay Time
2CP
16.5 CP
3 3
Pulse Width
Delay CALIBRATE PULSE WIDTH 20 ns CALIBRATION STATUS 15 ms Duration
NOTES
1
Refer to Figures 17, 18 and 24.
2
Design minimums are derived from worst case design analysis and/or simulation results. Typical values are based on characterization data. These specifications are not guaranteed or tested.
3
The time duration for this parameter varies in direct proportion to the width of the Clock Pulse (CP).
REV. 0
–3–
AD1385
WARNING!
ESD SENSITIVE DEVICE
ABSOLUTE MAXIMUM RATINGS*
+VS to AGND . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 V
–V
to AGND . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –18 V
S
V
to PGND . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 V
DD
V
to PGND . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –7 V
SS
AGND to PGND . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±0.3 V
Analog Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .±V
S
Reference Input . . . . . . . . . . . . . . . . . . . . . . . . . .0 V to +11 V
Digital Inputs . . . . . . . . . . . . . . . . . . . . –0.3 V to V
+ 0.3 V
DD
Output Short Circuit Duration
Reference Output . . . . . . . . . . . . . . . . . . . . . . . . . Indefinite
Track/Hold Output . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 sec
Digital Outputs . . . . . . . . . . . . . .1 sec for Any One Output
Case Temperature (Operating) . . . . . . . . . . –55°C to +125°C
Storage Temperature . . . . . . . . . . . . . . . . . . –65°C to +150°C
*Stresses greater than those listed under “Absolute Maximum Ratings” may cause
permanent damage to the device. This is a stress rating only and functional operation of the device at these or any other conditions above those indicated in the operational section of this specification is not implied. Exposure to absolute maximum rating conditions for extended periods may affect device reliability.
ORDERING GUIDE
Temperature
Model Range (Case) Package Option*
AD1385KD 0°C to +70°C DH-48A AD1385TD –55°C to +125°C DH-48A AD1385TD/883B –55°C to +125°C DH-48A
*DH-48A = Bottom Brazed Ceramic DIP.
AD1385 PIN CONNECTIONS
The AD1385 is housed in a 48-pin bottom-brazed ceramic bathtub package. The pinout is as follows:
Pin Function Pin Function
11 CLOCK IN 48 V
(+5 V POWER)
DD2
12 POWER GROUND 47 POWER GROUND 13 B1/B9 (MSB) 46 V
(–5 V POWER)
SS2
14 B2/B10 45 AUTOZERO 15 B3/B11 44 B1 SELECT 16 B4/B12 43 POWER GROUND 17 B5/B13 42 POWER GROUND 18 B6/B14 41
CAL
19 B7/B15 40 GAIN ADJUST
10 B8/B16 (LSB) 39 +10 V REFERENCE OUT 11 V
(+5 V SIGNAL) 38 –VS1 (–15 V)
DD1
12 POWER GROUND 37 SIGNAL GROUND 13 V
(–5 V SIGNAL) 36 +VS1 (+15 V)
SS1
14 SIGNAL GROUND 35 SIGNAL GROUND 15 DATA STROBE 34 DNC 16 HI/LO BYTE SELECT 33 DNC 17
OE DATA ENABLE 32 +10 V REFERENCE IN
18 START CONVERT 31 V
B
IN
19 HOLD COMMAND OUT 30 VIN A 20 SIGNAL GROUND 29 OFFSET ADJUST 21 +V
(+15 V) 28 CAL STATUS
S2
22 HOLD COMMAND IN 27 TRACK/HOLD OU TPUT 23 –VS2 (–15 V) 26 SIGNAL GROUND 24 POWER GROUND 25 TRACK/HOLD INPUT
DNC = DO NOT CONNECT.
CAUTION
ESD (electrostatic discharge) sensitive device. Electrostatic charges as high as 4000 V readily accumulate on the human body and test equipment and can discharge without detection. Although the AD1385 features proprietary ESD protection circuitry, permanent damage may occur on devices subjected to high energy electrostatic discharges. Therefore, proper ESD precautions are recommended to avoid performance degradation or loss of functionality.
–4–
REV. 0
AD1385
Figure 1. Spurious-Free Range vs. Input Amplitude,
±
5 V Range, 2048-Point FFT, 500 kHz Sample Rate
Figure 2. Spurious-Free Range vs. Input Amplitude,
±
10 V Range, 2048-Point FFT, 500 kHz Sample Rate
Figure 4. Full-Scale Sine Wave Power Spectral Density,
±
5 V Range, 16384-Point FFT, 500 kHz Sample Rate
Figure 5. Full-Scale Sine Wave Power Spectral Density,
±
5 V Range, 16384-Point FFT, 500 kHz Sample Rate
Figure 3. Full-Scale Sine Wave Power Spectral Density,
±
5 V Range, 16384-Point FFT, 500 kHz Sample Rate
REV. 0
Figure 6. 100 kHz Intermodulation Performance, ±5 V Range, 16384-Point FFT, 500 kHz Sample Rate
–5–
AD1385
Figure 7. 200 kHz Intermodulation Performance, ±5 V Range, 16384-Point FFT, 500 kHz Sample Rate
Figure 8. Full-Scale Sine Wave Power Spectral Density,
±
10 V Range, 16384-Point FFT, 500 kHz Sample Rate
Figure 10. Full-Scale Sine Wave Power Spectral Density,
±
10 V Range, 16384-Point FFT, 500 kHz Sample Rate
Figure 11. 100 kHz Intermodulation Performance, ±10 V Range, 16384-Point FFT, 500 kHz Sample Rate
Figure 9. Full-Scale Sine Wave Power Spectral Density,
±
10 V Range, 16384-Point FFT, 500 kHz Sample Rate
Figure 12. 200 kHz Intermodulation Performance, ±10 V Range, 16384-Point FFT, 500 kHz Sample Rate
–6–
REV. 0
Loading...