This reference manual is aimed at users who are familiar with AMETEK calibrators, as well
as those who are not. The manual is divided into 11 sections which describe how to set up,
operate, service and maintain the calibrator. The technical specifications are described and
accessories may be ordered from the list of accessories.
Safety symbols
This manual contains a number of safety symbols designed to draw your attention to
instructions which must be followed when using the instrument, as well as any risks involved.
Warning
Conditions and actions that may compromise the safe use of the
instrument and result in considerable personal or material damage.
Caution…
Conditions and actions that may compromise the safe use of the
instrument and result in slight personal or material damage.
Note…
Special situations which demand the user’s attention.
5.5 Selecting the set-temperature ............................................................................................................. 25
5.5.1 Editing the preset set-temperature ......................................................................................... 26
5.6 Auto Step function ............................................................................................................................... 28
5.6.1 Running an Auto Step test ..................................................................................................... 28
5.6.2 The calibrator's Auto Step procedure ..................................................................................... 29
5.7 Switch Test function ............................................................................................................................ 31
5.7.1 Running a Switch Test ........................................................................................................... 31
5.7.2 The calibrator's Switch Test procedure .................................................................................. 33
6.0 Storing and transporting the calibrator .................................................................................. 34
7.0 Error messages (List of alarms) .............................................................................................. 37
8.0 Returning the calibrator for service ........................................................................................ 40
11.0 List of accessories .................................................................................................................... 56
129375 00 2016-05-18 3
1.0 Introduction
A
Congratulations on your new AMETEK JOFRA CTC Calibrator!
With the AMETEK JOFRA Compact Temperature Calibrator, you have chosen an extremely
effective instrument which we hope will live up to all your expectations.
This CTC calibrator is a fast, timesaving, and reliable true industrial temperature calibrator
designed for on-site use.
During the past several years, we have acquired extensive knowledge of industrial temperature
calibration. This expertise is reflected in our products which are all designed for daily use in an
industrial environment. Please note that we would be very interested in hearing from you if you
have any ideas or suggestions for changes to our products.
This reference manual applies to the following instruments:
JOFRA CTC-155 A – Temperature calibrator
JOFRA CTC-155 C – Temperature calibrator with reference sensor input
JOFRA CTC-350 A – Temperature calibrator
JOFRA CTC-350 C – Temperature calibrator with reference sensor input
JOFRA CTC-660 A – Temperature calibrator
JOFRA CTC-660 C – Temperature calibrator with reference sensor input
The calibrator has the following features and functions:
Wide temperature range
Fast heating and cooling time as well as a short stabilization time
Signal input for external reference sensor, which makes it possible to improve accuracy even
more.
External sensor control for running in two modes.
Multi-Information colour Display and Function keys
Useful features such as Set function, Preset mode, Auto Switch Test and Auto Stepping
IRI – Intelligent Recalibration Information
Plug and play STS reference sensors with memory chip
Broad range of inserts
Reference sensor protection
Silent mode operation
JOFRACAL calibration software
Protective carrying case with compartments for inserts, cables, manuals, plugs etc.
ISO-9001 certified
METEK Denmark A/S was ISO-9001 certified in September 1994 by Bureau Veritas
Certification Denmark.
129375 00 2016-05-18 4
CE-label
Your new temperature calibrator bears the CE label and conforms to the
Electromagnetic Compatibility (EMC) Directive and the Low Voltage Directive.
Technical assistance
Please contact the dealer from whom you acquired the instrument if you require technical
assistance.
1.1 Warranty
This instrument is warranted against defects in workmanship, material and design for two (2)
years from date of delivery to the extent that AMETEK will, at its sole option, repair or replace
the instrument or any part thereof which is defective, provided, however, that this warranty shall
not apply to instruments subjected to tampering or, abuse, or exposed to highly corrosive
conditions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES WHETHER EXPRESS OR
IMPLIED AND AMETEK HEREBY DISCLAIMS ALL OTHER WARRANTIES, INCLUDING,
WITHOUT LIMITATION, ANY WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE OR
MERCHANTABILITY. AMETEK SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR
CONSEQUENTIAL DAMAGES, INCLUDING, BUT NOT LIMITED TO, ANY ANTICIPATED OR
LOST PROFITS.
This warranty is voidable if the purchaser fails to follow any and all instructions, warnings or
cautions in the instrument’s User Manual.
If a manufacturing defect is found, AMETEK will replace or repair the instrument or replace any
defective part thereof without charge; however, AMETEK’s obligation hereunder does not
include the cost of transportation, which must be borne by the customer. AMETEK assumes
no responsibility for damage in transit, and any claims for such damage should be presented
to the carrier by the purchaser.
129375 00 2016-05-18 5
1.2 Receiving the Compact Temperature Calibrator
When you receive the instrument…
1) Unpack and check the calibrator and the accessories carefully.
2) Check the parts according to the list shown below.
If any of the parts are missing or damaged, please contact the dealer who sold
you the calibrator.
You should receive:
1 CTC Calibrator
1 USB memory stick containing electronic Reference manual and software
package JOFRACAL
1 mains cable
1 sets of test leads and test clips (black and red)
Thermal protection shield (CTC-660 only)
1 tool for insertion tube
1 USB cable
1 Calibration certificate (International traceable)
When reordering, please specify the part numbers according to the list of accessories,
section 11.0
129375 00 2016-05-18 6
1.3 Dimensioning drawing
129375 00 2016-05-18 7
2.0 Safety instructions
Read this manual carefully before using the instrument!
Please follow the instructions and procedures described in this manual. They are
designed to allow you to get the most out of your calibrator and avoid any personal
injuries and/or damage to the instrument.
Disposal – WEEE Directive
These calibrators contain Electrical and Electronic circuits and must be recycled or
disposed of properly (in accordance with the WEEE Directive 2002/96/EC).
Warning
About the use:
The calibrator must not be used for any purposes other than those described in
this manual, as it might cause a hazard.
The calibrator has been designed for indoor use only and is not to be used in
wet locations.
The calibrator is not to be used in hazardous areas, where vapour or gas leaks,
etc. may constitute a danger of explosion.
The calibrator is not designed for operation in altitudes above 2000 meters.
The calibrator is a CLASS I product and must be connected to a mains outlet with
a protective earth connection. Ensure the ground connection of the calibrator is
properly connected to the protective earth before switching on the calibrator.
Always use a mains power cable with a mains plug that connects to the protective
earth.
To ensure the connection to protective earth any extension cord used must also
have a protective earth conductor.
Only use a mains power cord with a current rating as specified by the calibrator
and which is approved for the voltage and plug configuration in your area.
Before switching on the calibrator make sure that it is set to the voltage of the
mains electricity supply.
Always position the calibrator to enable easy and quick disconnection of the
power source (mains inlet socket).
The calibrator must be kept clear within an area of 20 cm on all sides and 1 metre
above the calibrator due to fire hazard.
Never use heat transfer fluids such as silicone, oil, paste, etc. in the dry-block
calibrators. These fluids may penetrate the calibrator and cause electrical hazard,
damage or create poisonous fumes.
The calibrator must be switched off before any attempt to service the instrument
is made. There are no user serviceable parts inside the calibrator.
When cleaning the well or the insertion tube, REMEMBER to wear goggles when
using compressed air!
About the frontpanel:
The connectors, on the front panel of the calibrator, must NEVER be connected to
a voltage source.
Thermostats connected to the switch test input must not be connected to any
other voltage source during a test.
129375 00 2016-05-18 8
About insertion tubes and insulation plugs:
Never leave hot insertion tubes, which have been removed from the calibrator,
unsupervised – they may constitute a fire hazard or personal injury.
If you intend to store the calibrator in the aluminium carrying case after use, you
must ensure that the instrument has cooled down to a temperature below
50°C/122°F before placing it in the carrying case.
Never place a hot insertion tube in the optional carrying case.
Use only insulation plugs supplied by AMETEK Denmark A/S.
About the fuses:
The fuse box must not be removed from the power control switch until the mains
cable has been disconnected.
The two main fuses must have the specified current and voltage rating and be of
the specified type. The use of makeshift fuses and the short-circuiting of
fuse holders are prohibited and may cause a hazard.
Caution – Hot surface
This symbol is visible on the grid plate.
Do not touch the grid plate, the well or the insertion tube as the calibrator is
heating up – they may be very hot and cause burns.
Do not touch the tip of the sensor when it is removed from the insertion tube/well
– it may be very hot and cause burns.
Do not touch the handle of the calibrator during use – it may be very hot and
cause burns.
Over 50°C/122°F
If the calibrator has been heated up to temperatures above 50°C/122°F, you must
wait until the instrument reaches a temperature below 50°C/122°F before you
switch it off.
Do not remove the insert from the calibrator before the insert has cooled down to
less than 50°C/122°F.
Caution – Cold surface
Below 0°C/32°F (applies only to the CTC-155 C models)
Do not touch the well or insertion tube when these are below 0°C/32°F - they
might create frostbite.
If the calibrator has reached a temperature below 0°C/32°F, ice crystals may form
on the insertion tube and the well. This, in turn, may cause the material surfaces
to oxidize
To prevent this from happening the insertion tube and the well must be dried. This
is done by heating up the calibrator to 100°C/212°F until all water left has
evaporated.
Remove the insulation plug while heating up.
It is very important that humidity in the well and insertion tube is removed to
prevent corrosion and frost expansion damages.
129375 00 2016-05-18 9
Caution…
A
About the use:
Do not use the instrument if the internal fan is out of order.
Before cleaning the calibrator, you must switch it off, allow it to cool down and
remove all cables.
bout the well, insertion tube and grid plate:The well and the insertion tube must be clean and dry before use.
Do not pour any form of liquids into the well. It might damage the well or cause
a hazard.
Scratches and other damage to the insertion tubes should be avoided by storing
the insertion tubes carefully when not in use.
The insertion tube must never be forced into the well. The well could be
damaged as a result, and the insertion tube may get stuck.
Before using new insertion tubes for the calibration, the insertion tubes must be
heated up to maximum temperature 350°C(662°F) / 660°C (1220°F) for a period
of minimum 30 minutes (CTC-350/660 only).
The insertion tube must always be removed from the calibrator after use.
The humidity in the air may cause corrosion oxidation on the insertion tube
inside the instrument. There is a risk that the insertion tube may get stuck if this
is allowed to happen.
If the calibrator is to be transported, the insertion tube must be removed from
the well to avoid damage to the instrument.
Note…
The product liability only applies if the instrument is subject to a manufacturing
defect. This liability becomes void if the user fails to follow the instructions set out
in this manual or uses unauthorised spare parts.
129375 00 2016-05-18 10
3.0 Setting up the calibrator for use
3.1 Preparing the calibrator
Warning
The calibrator has been designed for indoor use only and is not to be used in
wet locations.
The calibrator is not to be used in hazardous areas, where vapour or gas leaks,
etc. may constitute a danger of explosion.
The calibrator is not designed for operation in altitudes above 2000 meters.
The calibrator is a CLASS I product and must be connected to a mains outlet with
a protective earth connection. Ensure the ground connection of the calibrator is
properly connected to the protective earth before switching on the calibrator.
Always use a mains power cable with a mains plug that connects to the protective
earth.
To ensure the connection to protective earth any extension cord used must also
have a protective earth conductor.
Only use a mains power cord with a current rating as specified by the calibrator
and which is approved for the voltage and plug configuration in your area.
Before switching on the calibrator make sure that it is set to the voltage of the
mains electricity supply.
Always position the calibrator to enable easy and quick disconnection of the
power source (mains inlet socket).
The calibrator must be kept clear within an area of 20 cm on all sides and 1 metre
above the calibrator due to fire hazard.
Note…
The instrument must not be exposed to draughts.
When setting up the calibrator, you must…
Place the calibrator on an even horizontal surface where you intend to use it.
Caution…
Do not use the instrument if the internal fan is out of order.
Ensure a free supply of air to the internal fan located at the bottom
of the instrument (pos. 2). The area around the calibrator should be
free of draught, dirt, flammable substances etc.
Check that the fuse size corresponds to the applied voltage on (pos.
3). The fuse is contained in the power control switch (on/off switch
(230V/115V)). To check; do as follows (see Fig. 1):
129375 00 2016-05-18 11
Warning
The two main fuses must have the specified current and voltage rating and be of the
specified type. The use of makeshift fuses and the short-circuiting of fuse holders are
prohibited and may cause a hazard.
A. Open the fuse box lid using a screwdriver.
B. Take out the fuse box.
C. Remove both fuses replacing them with two new fuses. These must be identical
D. Slide the fuse box back into place.
Fig. 1
Changing fuses
and should correspond to the line voltage. See section 11.0.
The calibrator is now ready for use.
Check that the earth connection for the instrument is present and attach the cable.
Select an insertion tube with the correct bore diameter. See section 3.2 for information
on how to select insertion tubes.
3.2 Choosing an insertion tube
Caution…
To get the best results out of your calibrator, the insertion tube dimensions, tolerance
129375 00 2016-05-18 12
and material are critical. We highly advise using the JOFRA insertion tubes, as they
guarantee trouble free operation. Use of other insertion tubes may reduce
performance of the calibrator and cause the insertion tube to get stuck.
Caution…
Before using new insertion tubes for the calibration in the CTC-350/660 instruments
the insertion tubes must be heated up to maximum temperature 350°C(662°F) /
660°C (1220°F) for a period of minimum 30 minutes.
Insertion tubes are selected on the basis of the diameter of the sensor to be calibrated.
Use the table for insertion tubes in section 3.2.1 to find the correct parts number.
Alternatively, you may order an undrilled insertion tube and drill the required hole yourself. The
finished dimension should be as follows:
Sensor diameter ød+0.2 +0.05 / -0.00 mm.
Reference sensor hole ø4.2mm +0.05 / -0.00 mm
3.2.1 Standard insertion tubes
Caution…
To get the best results out of your calibrator, the insertion tube dimensions, tolerance
and material are critical. We highly advise using the JOFRA insertion tubes, as they
guarantee trouble free operation. Use of other insertion tubes may reduce
performance of the calibrator and cause the insertion tube to get stuck.
PARTS NO. FOR STANDARD INSERTION TUBES – SINGLE HOLES
Sensor
size
3 mm
4 mm
5 mm
6 mm
7 mm
8 mm
9 mm
10 mm
11 mm
12 mm
13 mm
14 mm
15 mm
16 mm*
18 mm*
20 mm*
CTC-155 CTC-350 CTC-660
129407
129408
129409
129410
129411
129412
129413
129414
129415
129416
129417
-
-
-
-
-
129429
129430
129431
129432
129433
129434
129435
129436
129437
129438
129439
129440
129441
129442
129443
129444
129459
129460
129461
129462
129463
129464
129465
129466
129467
129468
129469
129470
129471
129472
129473
129474
Sensor
size
Undrilled
Undrilled/ref. hole
1/8”
3/16”
1/4”
5/16”
3/8”
7/16”
1/2”
9/16”
5/8”
11/16”*
13/16”*
3/4”*
CTC-155 CTC-350 CTC-660
129418
129419
129420
129421
129422
129423
129424
129425
129426
129427
129428
-
-
-
129445
129446
129447
129448
129449
129450
129451
129452
129453
129454
129455
129456
129457
129458
129475
129476
129477
129478
129479
129480
129481
129482
129483
129484
129485
129486
129487
129488
The CTC-155 single-hole insertion tubes are delivered with a matching insulation plug.
*Note: Insertion tubes without reference holes.
129375 00 2016-05-18 13
PART NO. FOR STANDARD INSERTION TUBES – MULTI-HOLE
Description
Metric Type 1
Sensor size
3, 4, 5, 6, 9 mm and
4mm REF
Inch Type 2
Sensor size
1/8”, 3/16”, 1/4”, 3/8” and
4 mm REF
CTC-155
129489 129491 129493
129490
CTC-350
129492
CTC-660
129494
The CTC-155 multi-hole insertion tubes are delivered with a matching insulation plug.
3.3 Inserting the sensor
Before inserting the sensor and switching on the calibrator, please note the following important
warning:
Warning
Never use heat transfer fluids such as silicone, oil, paste, etc. in the dry-block
calibrators.
These fluids may penetrate the calibrator and cause electrical hazard, damage or
create poisonous fumes.
Never try to modify the insulation plugs to make them fit the sensor. Use only
insulation plugs supplied by AMETEK Denmark A/S.
Insert the sensor as shown below in fig. 2.
Fig. 2
Inserting sensor and
insertion tube
Sensor under test
Insertion tube
Thermal protection shield
129375 00 2016-05-18 14
Check that the insulation plug fits the diameter of the sensor. Otherwise replace it (CTC-155
only).
In order to spare the sensor and its connections it is recommended to use a thermal
protection shield (129264) at high temperatures (CTC-350/660 only).
Caution…
The well and the insertion tube must be clean before use.
Do not pour any form of liquids in the well. It might damage the well.
Scratches and other damage to the insertion tubes should be avoided by storing
the insertion tubes carefully when not in use.
The insertion tube must never be forced into the well. The well could be damaged
as a result, and the insertion tube may get stuck.
Caution – Hot surface
Do not touch the grid plate, the well or the insertion tube as the calibrator is
heating up – they may be very hot and cause burns.
Do not touch the tip of the sensor when it is removed from the insertion tube/well
– it may be very hot and cause burns.
Do not touch the handle of the calibrator during use – it may be very hot and
cause burns.
Do not remove the insertion tube from the calibrator before the insertion tube has
cooled down to less than 50°C/122°F.
Caution – Cold surface
Below 0°C/32°F (applies only to the CTC-155 models)
Do not touch the well or insertion tube when these are below 0°C/32°F - they might
create frostbite.
If the calibrator has reached a temperature below 0°C/32°F, ice crystals may form
on the insertion tube and the well. This, in turn, may cause the material surfaces to
oxidize
To prevent this from happening the insertion tube and the well must be dried. This is
done by heating up the calibrator to 100°C/212°F until all water left has evaporated.
Remove the insulation plug while heating up.
It is very important that humidity in the well and insertion tube is removed to prevent
corrosion and frost expansion damages.
129375 00 2016-05-18 15
3.4 Programming intelligent STS sensors
Use the configuration software CON050 supplied with CTC to program and to update
calibration information in intelligent STS sensors.
For instructions read the software manual for CON050 installed on the USB key.
129375 00 2016-05-18 16
play
y
4.0 Calibrator Interface
4.1 Keypad - Functions
Function keys
F1, F2, F3, F4
To operate the horizontal
menu bar use the
Functions keys.
System menu key
Press the key to enter the
System menu.
Back key
Press the key to cancel a
selection or return to
previous menu.
Arrow Keys
Serve different functions
depending on the mode of
operation.
Navigation mode: Use the
four keys to move the
cursor in the desired
direction.
Edit mode: The Up and
Down Arrow keys scroll
through the lists of options.
If entering a number, the
Left and Right Arrow keys
move the cursor one
character in the desired
direction.
Action key / Enter key
Action function: Open and
close edit fields or a menu
button. The action key also
accepts the selected
option or entered value.
Enter function: Accept
selected options or
entered values. When a
value is entered with the
Enter Key the cursor
selects the next
configurable field in the
list.
4.2 Display - Functions
Reference sensor info
Shows the reference sensor
selected. The serial number
of the external reference
sensor is read from the
intelligent reference sensor
and displayed in this field.
“Time to stable”
Selection
Shows the selected
specified stability criteria
and states a time when the
stable situation can be
achieved.
Process Indicator
Indicates the status of the
current process.
Memory reading
Shows the current
memory selected
from the System
menu.
Warning/Error
s
mbols. See 4.2.2
Real Time Clock and
date dis
Horizontal menu bar
Provides you with the relevant menu
options that can be selected at the
present point.
selected and activated by pressing the
function keys (F1, F2, F3 and F4).
Each option can be
.
True temperature
reading
Shows the numeric value
of the temperature being
measured. Can be either
the internal reference
sensor or an external
reference sensor.
Units
Shows the unit of the
current measurement.
Internal Reference
temperature reading
Shows the temperature of
the well.
Set temperature reading
Shows the numeric value
of the current set
temperature selected.
129375 00 2016-05-18 17
Loading...