Please note: All specications are given with an ambient temperature 23°C / 73.4°F ± 3°C / 5.9°F and specied at 115 V/230 V.
) In lower 30 mm incl. Stability, Uniformity, 12-month drift (typical), Hysteresis, Resolution, Load (Max 6 mm) and calibration laboratory uncertainty.
) Measured after the stability indicator has been on for 10 minutes. Measuring time is 30 minutes.
) Stability when exposed to maximum temperature for 100 hours. Stability will depend on actual use of the probe.
Mains Power
Voltage .....................115 V (90-127) / 230 V (180-254)
Max Power Consumption .............................. 130 VA
Frequency, US deliveries ............................. 60 Hz ±3
Frequency, non US deliveries ...............50 Hz ±3, 60 Hz ±3
Physical Specifications
Dimension L x W x H .
Weight
CTC-155 .......................................5.5 kg / 12.1 lb
Immersion Depth incl. insulation plug
CTC-155 .....................................120 mm / 4.53 in
Well Diameter
CTC-155 ......................................26 mm / 1.02 in
Insert Dimensions (diameter x length)
CTC-155 ................... 25.8 mm x 100 mm / 1.01 x 3.9 in
248x148x305 mm / 9.761x5.83x12.01 in
Electrical
Switch Input (dry contact)
Test Voltage ................................ Maximum 14 VDC
Test Current ...................................Maximum 1 mA
Digital Interface
USB 2.0
Environmental
Operating Temperature
0 to 50°C / 32 to 122°F
Storage Temperature
-20 to 50°C / -4 to 122°F
Humidity
5 to 90% Rh, non-condensing
Protection Class
IP-10
External Reference Sensor
STS-120-A-915 .....................-25 to 155°C /-13 to 311°F
Type .....................................................Pt100
Response Time
STS-120-A: t
STS-120-A: t
Dimensions
Diameter ......................................4 mm / 0.157 in
Length .......................................135 mm / 5.32 in
Max. height over calibrator top ................10 mm / 0.39 in
(50%) ......................................7 sec.
0.5
(90%) .....................................18 sec.
0.9
Please note:All specications are given with an ambient temperature 23°C / 73.4°F ± 3°C / 5.9°F and specied at 115 V/230 V.
) In lower 30 mm incl. Stability, Uniformity, 12-month drift (typical), Hysteresis, Resolution, Load (Max 6 mm) and calibration laboratory uncertainty.
) Measured after the stability indicator has been on for 10 minutes. Measuring time is 30 minutes.
) Stability when exposed to maximum temperature for 100 hours. Stability will depend on actual use of the probe.
Specifications CTC-652
Temperature
Temperature Range
@ Ambient 23°C/73°F ..............28 to 650°C / 82 to 1202°F
@ Ambient 0°C/32°F ................5 to 650°C / 41 to 1202°F
Voltage .............115 V (103-127) / 206 V (180-254)
Max Power Consumption .............................1150 VA
Frequency, US deliveries ............................. 60 Hz ±3
Frequency, non US deliveries ...............50 Hz ±3, 60 Hz ±3
Physical Specifications
Dimension L x W x H ..
Weight
CTC-652 .......................................9.0 kg / 19.8 lb
Immersion Depth
CTC-652 .....................................190 mm / 7.48 in
Well Diameter
CTC-652 ......................................26 mm / 1.02 in
Insert Dimensions (diameter x length)
CTC-652 .................25.7 mm x 200 mm / 1.01 x 7.874 in
248x148x390 mm / 9.76x5.83x15.35 in
Electrical
Switch Input (dry contact)
Test Voltage ................................ Maximum 14 VDC
Test Current ...................................Maximum 1 mA
Digital Interface
USB 2.0
Environmental
Operating Temperature
0 to 50°C / 32 to 122°F
Storage Temperature
-20 to 50°C / -4 to 122°F
Humidity
5 to 90% Rh, non-condensing
Protection Class
IP-10
Please note:All specications are given with an ambient temperature 23°C / 73.4°F ± 3°C / 5.9°F and specied at 115 V/230 V.
) In lower 40 mm incl. Stability, Uniformity, 12-month drift (typical), Hysteresis, Resolution, Load (Max 6 mm) and calibration laboratory uncertainty.
) Measured after the stability indicator has been on for 10 minutes. Measuring time is 30 minutes.
) Stability when exposed to maximum temperature for 100 hours. Stability will depend on actual use of the probe.
) 103V-127VAC/206-254VAC 47-63 Hz for full temperature range. At mains voltage below 103VAC/206VAC maximum temperature will be reduced.
External Reference Sensor
STS-200-A-970 ......................0 to 700°C / 32 to 1292°F
Type .....................................................Pt100
Response Time
STS-120-A: t
STS-120-A: t
Dimensions
Diameter ......................................4 mm / 0.157 in
Length .......................................151 mm / 5.95 in
Max. height over calibrator top ................25 mm / 0.94 in
(50%) ......................................8 sec.
0.5
(90%) .....................................26 sec.
0.9
Please note:All specications are given with an ambient temperature 23°C / 73.4°F ± 3°C / 5.9°F and specied at 115 V/230 V.
) In lower 30 mm incl. Stability, Uniformity, 12-month drift (typical), Hysteresis, Resolution, Load (Max 6 mm) and calibration laboratory uncertainty.
) Measured after the stability indicator has been on for 10 minutes. Measuring time is 30 minutes.
) Stability when exposed to maximum temperature for 100 hours. Stability will depend on actual use of the probe.
Inserts
Inserts for CTC-155 and CTC-350 are made of aluminum. Inserts for CTC-652 and CTC-660 are made of brass. All specications on hole sizes refer to the outer diameter of the
sensor-under-test. The correct clearance size is applied in all predrilled inserts. All CTC-155 inserts include an insulation plug.
Predrilled Inserts–metric (mm)
Part Numbers
Probe Dia.
3 mm003129 407129429130156129459
4 mm004129 4081294301301571294 60
5 mm00512 940 912 94 3113015 81294 61
6 mm00612 9410129 432130159129 462
7 mm007129411129 43313 016012 94 63
8 mm008129 412129 434130161129 46 4
9 mm009129 413129 4351301621294 65
10 mm010129 414129 436130163129466
11 mm011129 415129 43713 016412 9467
12 m m012129416129 4381301651294 68
13 m m013129 417129439130166129 469
14 mm014N/A129 44013016712 9470
15 m m015N/A129441130168129 471
16 mm016N/A129 442 *130169*129 472*
18 mm018N/A129 443*130170 *129 473*
20 mm020N/A129444*130171*129 474*
Package of the
above inserts
Multi-holeM01129489129 49113015212 94 93
Insert
CTC-155CTC-350CTC-652CTC-660
Code
—12950 2129 50413018412 950 6
Predrilled Inserts–imperial (in)
Part Numbers
Probe Dia.
1/8 in1251294201294 4713017212947 7
3/16 in187129 421129 44813017 312 9478
1/4 in25012942 21294 4913 017412 9479
5/16 in31212942 3129450130175129 48 0
3/8 in375129424129451130176129 481
7/16 in437129 425129452130177129 482
1/2 in500129 426129 453130178129 48 3
9/16 in562129 427129 45413017912 9484
5/8 in625129 42812945513 01801294 85
11/16 in688N/A1294 56*130181*12 9486*
13/16 in750N/A129457*130183*1294 87*
3/4 in813N /A129458*13 0182*12 94 88*
Package of the
above inserts
Multi-holeM0212949 0129 492130153129494
* All inserts are suppied with hole for 4mm reference sensor, execpt if marked
with *.
Insert
CTC-155CTC-350CTC-652CTC-660
Code
—12950 312950513018512 9507
Undrilled Inserts
Part Numbers
Inserts
5-Pack
No ref. holeUN1129 418129 445130154129 475
w/ref. holeUN2129419129 44613 015512 9476
CTC-155/350/652/660 Inserts
Use of other inserts may reduce performance of the calibrator. To get
the best results out of the calibrator, the insert dimensions, tolerance,
and material is critical. We highly advise using JOFRA inserts, as they
guarantee trouble free operation.
Insert
CTC-155CTC-350CTC-652CTC-660
Code
Typical Weight
CTC-155: 2.6 oz / 75 g
CTC-350: 5.8 oz/170 g
CTC-652: 28.6 oz/850 g
CTC-660: 17.8 oz/510 g
Do you need a customized insert?
Please contact us for more information.
9
4
3
5
4
6
M01 Multi-hole
3/8
4
1/8
1/4
3/16
M02 Multi-hole
UN1 Undrilled
4
UN2 Undrilled w/ref. hole
Ordering Information
Base Model Number
CTC155
CTC350
CTC652CTC-652, 28 to 650 °C (82 to 1202°F)
CTC660
Model Version
ABasic mod el, without inpu t
CFull model, includes reference sensor input
Power Supp ly (US deliverie s 60 Hz only)
115115 V AC
230230 VA C
Mains Powe r Cable Type
A
BUSA/Cana da 115 V
CUK 240 V
DSouth Africa 2 20 V
EItaly 2 20 V
FAustra lia 240 V
GDenmark 230 V
HSwitzerlan d 220 V
IIsrael 23 0 V
Inser t Type an d Size
NONWithout insert (standard delivery)
UNX1 x Undrill ed insert (Pleas e see insert sec tion for corre ct code)
XXX1 x Si ngle hole Inser t (Please see inse rt section fo r correct code)
MXX1 x Mul ti hole insert (Pl ease see inser t section for cor rect code)
SIMComplet e set of Imperial ins erts - inch
SMMComplete set o f Metric inser ts - mm
STS Reference Sensor
R1STS-102 Ref. Sens or - temperature r ange up to 155°C/311°F
R21STS-120-A- 915 Ref. Senso r for CTC-155
R22STS-120-A-935 Ref. Sens or for CTC-350
R4 STS-200 -A-970 Ref. Senso r for CTC-652
R23STS-102-A-966 Ref. Sens or for CTC-660
Calibration Certicate
FTraceable Certicate to International standards (standard)
HAccredit ed Certicate - IS O17025
HS
CTC-155, -25 to 155°C (-13 to 311°F)
CTC-350, 28 to 350°C (82 to 6 62°F)
CTC-660 , 28 to 660°C (82 to 1220°F)
European 2 30 V
System Cali bration - Accredi ted Certicate - I SO17025 (C-Model On ly)
Accessories
CXProtective Carrying Case
CRProte ctive Carry ing Case with Supp ort Rod Set
SRSupport Rod Se t
StandardDelivery
CTC dry-block calibrator (user specied)
Mains power cable (user specied)
Traceable certicate - temperature performance
Tool for insertion tubes
User manual
Test cables (1 x red, 1 x black)
USB cable
JofraCal calibration software
CTC-660 and CTC-652 includes a Heat Shield
CTC350 C 230 A M01 R21 F CR CTC-350 with standard accessories, Full model, incl. Reference sensor input, 230 VAC, European power cord,
Multi-hole insert type M01, STS-120-A-915 Ref. sensor, Traceable certicate, and Carrying case with support rod
*ISO 17025 accredited calibration lab.
Accessories
STS120A915EH ..Reference sensor for CTC-155
STS120A935EH ..Reference sensor for CTC-350
STS200A970EH ..Reference sensor for CTC-652
STS120A966EH ..Reference sensor for CTC-660
STS102A030EH ..Reference sensor STS-102
129540 .......... Carrying Case for CTC-155-350-660
130029 .......... Carrying Case for CTC-652
129539 ..........Support Rod set
125067 ..........Sensor grip
125066 ..........Fixture for sensor grip
129264 ..........Heat Shield
122832 .......... Cleaning Brushes - 4 mm - Package of 3 pcs
60F174 .......... Cleaning Brushes - 6 mm - Package of 3 pcs
122822 .......... Cleaning Brushes - 8 mm - Package of 3 pcs
Carrying Case
129539 Support Rod Set:
1 Rod
1 Fixing screw
1 Sensor Grip
1 Fixture for sensor grip
EN ISO/IEC 17025
Laboratory accreditation
AMETEK Sensors, Test & Calibration has two
EN ISO/IEC 17025 accredited laboratories that issues
ac credited certificates in accordance with international standards.
Laboratory accreditation is a reliable indicator of technical
competence assuring customers the most accurate
documentation. We believe in being clear about our
capabilities, our accuracy, and about what you
can expect from us.
Because calibration is a matter of condence!
USA, Florida
Tel +1 (800) 527 9999
cal.info@ametek.com
USA, California *
Tel +1 (800) 444 1850
crystal@ametek.com
India
Tel +91 22 2836 4750
jofra@ametek.com
129264 Heat ShieldSTS-Type Reference Sensor
Singapore
Tel +65 6484 2388
jofra@ametek.com
China, Shanghai
Tel +86 21 5868 5111
stc.china@ametek.com
China, Beijing
Tel +86 10 8526 2111 ext.19
stc.china@ametek.com
125067 Sensor Grip125066 Fixture for Grip
United Kingdom
Tel +44 (0) 1243 833 302
stc.uk@ametek.com
France
Tel +33 (0)1 30 68 89 40
general.lloyd-instruments@ametek.fr