CTC-140 A -17 to 140°C / -1 to 284°F
CTC-320 A 33 to 320°C / 91 to 608°F
CTC-320 B 33 to 320°C / 91 to 608°F
CTC-650 A 33 to 650°C / 91 to 1202°F
CTC-650 B 33 to 650°C / 91 to 1202°F
CTC-1200 A 300 to 1205°C / 572 to 2200°F
»
Fast calibration is timesaving
The specially designed heating block profile
heats up to 320°C / 608°F in just 4 minutes
and to 650°C / 1202°F in only 10 minutes
»
High flexibility
You are not limited by fixed holes.
Inter-changeable insertion tubes are
used to match the diameter of your sensorunder-test
Compact Temperature Calibrator
CTC-series
»
Enhanced stability
MVI circuitry ensures stability despite mains
supply variations in the process environment
»
Timesaving features
Fast one-key-one-function access to the
automatic switch test and auto stepping
»
Documentation made easy
RS232 communication and JOFRACAL
calibration software are included in the
standard delivery
»
Complete marine program
Part of a complete program of marine
approved temperature, pressure and signal
calibrators; including temperature sensors
A fast, timesaving, and reliable true temperature calibrator designed for
on-site use.
The CTC series is a fast dry-block that offers both interchangeable
inserts, the MVI stability circuitry, and calibration software. Both speed
and portability are superior to liquid baths. Dry-block calibrators do not
require hazardous liquids and provide a wide temperature range.
Calibrate RTD’s, thermocouples, thermoswitches, thermistors, and other
common temperature sensing devices.
The CTC series is designed for both on-site and maintenance shop use.
The applications are generally critical process control but can vary based
on calibration and testing requirements. The user interface is easy and
intuitive. One-key-one-function gives you quick access to timesaving features such as the switch test or the auto-stepping function.
All models feature a large, backlit LCD display panel, which is easyto-read even in well-lit areas. Units feature an informative display that
provides icons and information regarding the status of the CTC and the
calibration in-progress. The CTC series consists of six different models
that differ in temperature ranges and immersion depths. All units offer
similar features. A rugged, slim-line, aluminum outer casing with die-cast
top and bottom protects the CTC series of dry-block calibrators.
ISO 9001 Manufacturer
Specification Sheet
SS-CTC
For easy documentation and automatic calibration, all units are delivered
with RS232 serial communication and JOFRACAL calibration software.
Fast heating and cooling
AUTO
STEP
ESC
MENU
SWITCH
TEST
The CTC-320 A and the CTC-650 A contain an innovative
heating block profile. This design heats up the CTC-320 A
to maximum temperature in just 4 minutes and the CTC650 A in only 10 minutes. The fast performance of the
heating block is due to the special profile that minimizes
mass and yet, still accepts an insertion tube with a 25 mm
/ 1 in. outer diameter. This design is a balanced compromise between temperature stability / homogeneity and
rapid heating / cooling.
Deep immersion depth
The CTC-320 B and CTC-650 B models offer a deeper
immersion depth of 200 mm / 7.9 in. If you have liquid-filled
sensors or other sensors that require a deeper immersion
depth, look for the B versions. While the units do not heat
and cool as quickly as their shorter counterparts, they offer
the capability to accommodate longer sensors.
Peltier effect (CTC-140 A)
The model CTC-140 A features Peltier elements. In 1834,
Jean Peltier, a French physicist found that an ’’opposite
thermocouple effect’’ could be observed when an electric
current was connected to a thermocouple. Heat would be
absorbed at one of the junctions and discharged at the
other junction. This effect is called the ’’PELTIER EFFECT’’.
MVI - Improved temperature stability
MVI stands for ’’Mains power Variance Immunity’’.
Unstable mains power supplies are a major contributor to
on-site calibration inaccuracies. Traditional temperature
calibrators often become unstable in production environments where large electrical motors, heating elements,
and other devices are periodically cycled on or off. The
cycling of supply power can cause the temperature regulator to perform inconsistently leading to both inaccurate
readings and unstable temperatures.
The CTC series employ the MVI, thus avoiding such stability problems. The MVI circuitry continuously monitors the
supply voltage and ensures a constant energy flow to the
heating elements.
Easy-to-use, intuitive operation
All instrument controls may be performed from the front
panel. The heat source is positioned away from the panel.
This design helps to protect the operator.
The main functions on the CTC series are designed with
one-key-one-function logic. This means that there are no
sub-menus or difficult to remember multiple keystrokes
necessary to access primary functions. The easy-to-read,
backlit display features dedicated icons, which help in
identifying instrument conditions and operational steps.
The practical Peltier element (electronic heating pump)
consists of many elements of semiconductor material
connected electrically in series and thermally in parallel.
These thermoelectric elements and their electrical interconnections are mounted between two ceramic plates.
The plates serve to mechanically hold the overall structure together and to electrically insulate the individual
elements from one another.
Maximum temperature
From the setup menu, the user can select the maximum
temperature limit for the calibrator. This function prevents
damage to the sensor-under-test caused by the application of excessive temperatures. The feature also aids in
reducing drift resulting from extended periods of exposures to high temperatures. This feature can be locked
with an access code.
2 www.jofra.comCompact Temperature Calibrator • CTC
Set temperature
The ’’Up’’ and ’’Down’’ arrow keys allow the user to set
the exact temperature desired with a resolution of 0.1°.
Instrument setups
The CTC series stores the complete instrument setup,
including: engineering units, stability criteria, resolution,
display contrast, slope (ramp) rate, auto-step settings, and
maximum temperature.
Stability indicator
AUTO
STEP
ESC
MENU
SWITCH
TEST
1
2
3
4
START
AUTOSTEP
ESC
MENU
SWITCH
TEST
START
Slope rate
(Ramp)
AUTO
STEP
ESC
MENU
SWITCH
TEST
A bold checkmark on the display indicates that the calibrator has reached the desired set temperature and is
stable. The operator may change the stability criteria and
establish a greater sense of security in the calibration
results. A convenient countdown timer is activated five
minutes before the unit reaches stability.
Automatic switch test
Operators can save a lot of time using the automatic thermoswitch test function to find values for the ’’Open’’ and
’’Close’’ temperatures. Additionally, this feature displays
the hysteresis (deadband) between the two points. The
feature ensures a very high repeatability when testing
thermoswitches. Simply press the ’’SWITCH TEST’’ key to
activate the function.
Auto-stepping
This feature saves manpower. The operator may stay in
the control room, or another remote location, monitoring the output from the sensor-under-test while the CTC
series calibrator is placed in the process and automatically
changes the temperature using a programmed step value
and rate. Up to 9 different temperature steps may be programmed, including the hold time for each step.
Re-calibration/adjustments
The CTC series has a very easy and straightforward procedure for re-calibration/adjustment. There is no need for
a screwdriver or PC software. The only thing you need is a
reliable reference thermometer.
Place the probe in the calibrator and follow the instructions on the display. Third-party labs and calibration facilities will be able to perform this function if a certificate
from an independent source is necessary. Of course,
AMETEK can provide you with a traceable calibration certificate from our labs when you require a higher level of
confidence.
Calibration of up to 24 sensors with JOFRA ASM
Using the CTC series together with the ASM Advanced
Signal Multi-scanner offers a great time-saving automatic
solution to calibrate multiple temperature sensors at the
same time.
The ASM series is an eight channel scanner controlled by
JOFRACAL software on a PC. Up to 3 ASM units can be
stacked to calibrate up to 24 sensors at the same time.
It can handle signals from 2-, 3- and 4 wire RTD’s, TC’s,
transmitters, thermisters, temperature switches and voltage.
Liquid filled sensors and switches
The tall B models with an immersion depth of 190 mm /
7.5 in are ideal for calibration of liquid filled sensors. The
specially designed non-linear heating elements in the
CTC-650 B and the increased block mass provide a very
homogeneous temperature throughout the block. It is
essential for the quality of the calibration/test that the full
lenght of the sensing part of the sensor is exposed to the
same temperature. Calibrate analog reading devices or
switches with very high repeatability.
www.jofra.com 3CTC • Compact Temperature Calibrator
JOFRACAL CALIBRATION SOFTWARE
JOFRACAL calibration software ensures easy calibration
of RTD´s, thermocouples, transmitters, thermoswithes,
pressure gauges and pressure switches. JOFRACAL can
be used with JOFRA DPC-500, HPC and IPI pressure
calibrators, all JOFRA temperature calibrators, as well as
AMC900, ASC300 multi signal calibrator and ASM-800
signal multi scanner.
JOFRACAL calibration software may also be used for
manual calibrations, as it can be set up to accept manual
entry of calibration data together with other liquid baths,
ice points or dry-block heat sources.
The calibration data collected may be stored on a PC for
later recall or analysis. The calibrator stores the calibration
procedure and may be taken out to the process site without using a personal computer.
Once all calibrations are completed, the data may be
uploaded to the JOFRACAL calibration software for postprocessing and printing of certificates. The calibration data
collected may be stored on the personal computer for
later recall or analysis.
The JOFRACAL temperature calibration software may be
donwloaded from our web-page www.jofra.com.
STANDARD DELIVERY
• CTC dry-block calibrator (user specified)
• Mains power cable (user specified)
• Traceable certificate - temperature performance
• CTC-1200 A only: Matching Insulation plug kit (3 pcs)
ACCESSORIES
122832 Cleaning Brushes - 4 mm - Package of 3 pcs
60F174 Cleaning Brushes - 6 mm - Package of 3 pcs
122822 Cleaning Brushes - 8 mm - Package of 3 pcs
65-F100 Insulation in Tube, 100 mm x Ø25 mm
65-F101 Insulation in Tube, 150 mm x Ø25 mm
65-F102 Insulation in Tube, 200 mm x Ø25 mm
65-F103 Insulation in Tube, 250 mm x Ø25 mm
65-F104 Insulation in Tube, 300 mm x Ø25 mm
65-F105 Insulation in Tube, 350 mm x Ø25 mm
65-F106 Insulation in Tube, 400 mm x Ø25 mm
65-F107 Insulation in Tube, 450 mm x Ø25 mm
105173 Set of Insulation Plates ( 10 pcs )
125066 Extra fixture for sensor grib
125067 Extra sensor grib
125002 Edgeport Converter with 4 pcs of RS232 ports
123408 Carrying Case CTC/MTC A models
123409 Carrying Case CTC B models and for CTC-1200 A
JOFRACAL software
Minimum hardware requirements for JOFRACAL calibration software.
TM
• INTEL
• (PENTIUM
• 32 MB RAM (64 MB recommended)
• 80 MB free disk space on hard disk prior to installation
• Standard VGA (800 x 600, 16 colors) compatible screen
• (1024 x 786, 256 colors recommended)
• CD-ROM drive for installation of the program
• 1 free RS232 serial port
486 processor
TM
800 MHz recommended)
Insulation Plugs for CTC-1200:
124414 Insulation plug (3 pcs) 12 mm - ½ in.
124415 Insulation plug (3 pcs.) 3, 4 mm and 1/8 in.
124416 Insulation plug (3 pcs) 5, 6 mm and 1/4, 3/16in.
124518 Insulation plug (3 pcs) 7, 8, 9 mm and 5/16 in.
124519 Insulation plug (3 pcs) 10,11 mm and 3/8, 7/16 in.
Carrying case (Optional)
The optional protective carrying case ensures safe transportation and storage of the instrument and all associated
equipment.
4 www.jofra.comCompact Temperature Calibrator • CTC
FUNCTIONAL SPECIFICATIONS
Mains specifications
Voltage CTC-140/320/650/1200 ................... 115V(90-127) /
230V(180-254)
Voltage CTC-650 B ............ 115V(105-127) / 230V(210-254)
Frequency, non US deliveries ............... 50 Hz ±5, 60 Hz ±5
Frequency, US deliveries .......................................60 Hz ±5
Power consumption (max.) CTC-140 A .................... 150 VA
Power consumption (max.) CTC-320 B .................... 600 VA
Power consumption (max.) CTC-1200 A .................. 650 VA
Power consumption (max.) CTC-320 A / 650 A/B .. 1150 VA
Temperature range
CTC-140 A
Maximum ........................................................140°C / 284°F
CTC-650 A ...................................................±0.9°C / 1.62°F
CTC-650 B ...................................................±0.6°C / 1.08°F
CTC-1200 A ...................................................±2.0°C / 3.6°F
Specification when using the internal reference. (Load 4 mm OD
reference probe in the center of the insert).
Immersion depth
CTC-140 A (insulation included) ...........................................
. 115 mm / 4.5 in
CTC-320 A / CTC-650 A / CTC-1200 A ..... 110 mm / 4.3 in
CTC-320 B / CTC-650 B ............................. 190 mm / 7.5 in
Well diameter
CTC-140 ................................................. . 19,2 mm / 0.76 in
CTC-320 / CTC-650 ...................................... 26 mm / 1.0 in
CTC-1200 ...................................................... 27 mm / 1.6 in
Heating time
CTC-140 A
-17 to 23°C / 1 to 73°F .........................................3 minutes
23 to 140°C / 73 to 284°F ................................... 15 minutes
CTC-320 A
23 to 320°C / 73 to 608°F ..................................... 4 minutes
CTC-650 A
23 to 650°C / 73 to 1202°F ................................. 10 minutes
CTC-320 B
23 to 320°C / 73 to 608°F ................................... 20 minutes
CTC-650 B
23 to 650°C / 73 to 1202°F ................................. 39 minutes
CTC-1200 A
23 to 1205°C / 73 to 2200°F ............................... 45 minutes
Cooling time
CTC-140 A
100 to 0°C / 212 to 32°F ..................................... 10 minutes
0 to -15°C / 32 to 5°F .........................................16 minutes
140 to 100°C / 284 to 212°F ................................. 2 minutes
CTC-320 A
320 to 100°C / 608 to 212°F ............................... 16 minutes
CTC-650 A
650 to 100°C / 1202 to 212°F ............................. 28 minutes
CTC-320 B
320 to 100°C / 608 to 212°F ............................... 22 minutes
CTC-650 B
650 to 100°C / 1202 to 212°F ............................. 65 minutes
CTC-1200 A
1205 to 300°C / 2200 to 572°F ......................... 120 minutes
Switch input (dry contact)
Test voltage .............................................. Maximum 5 VDC
Test current ............................................. Maximum 2.5 mA
Thermal protection shield
(Optional) - 104216
An external heat shield is available and may be placed on top of
the calibrator to reduce the hot air
stream around the sensor-undertest. This is especially important for
testing thermocouples having headmounted transmitters with coldjunction compensation.
Support rod set (Optional) 125068
Support rod for sensors to be
mounted on JOFRA dry-block calibrators. Holds the sensor under test
in their position, while calibrating.
Includes 2 sensors grips and 2 fixtures for sensor gribs.
www.jofra.com 5CTC • Compact Temperature Calibrator
Calibration status icons ...................................................Yes
Training mode (heating/cooling block disabled)
Simulation of all functions ...............................................Yes
Simulating heating and cooling ..... Approx. 100° per minute
Service facilities
Adjustment of the unit from the keypad .........................Yes
Self explaining guide in display .......................................Yes
Other information ............................ Displays serial number,
.................. software revision level, and last calibration date
Setup facilities
Stability criteria .......................................... Extra time before
.................................................’’stable indication’’ is shown
Display resolution ........................................... 0.1° or 1°C/°F
Temperature units ....................................................°C or °F
Slope rate .................................................0.1 to 9.9°/minute
Maximum temperature ..................... Any value within range
Miscellaneous
Serial data interface ..............................RS232 (9-pin Male)
Operating temperature .................... 0 to 40°C / 32 to 104°F
Storage temperature .................... -20 to 50°C / -4 to 122°F
Humidity ..........................................................0 to 90% RH
Protection class .......................................................... IP-10
DNV Marine Approval, Certificate no ...................... A-10384
Instrument dimensions
CTC-140 A / CTC-320 A / CTC-650 A
L x W x H: ...........241 x 139 x 325 mm / 9.5 x 5.5 x 12.8 in
CTC-320 B / CTC-650 B / CTC-1200 A
L x W x H: ...........241 x 139 x 408 mm / 9.5 x 5.5 x 16.1 in
Instrument weight
CTC-140 A / CTC-320 B ................................. 7 kg / 15.5 lb
CTC-320 A ......................................................... 5 kg / 11 lb
CTC-650 A ......................................................... 6 kg / 13 lb
CTC-650 B .................................................... 10.5 kg / 23 lb
CTC-1200 A .................................................. 12 kg / 26.5 lb
Insert dimensions
CTC-140 A outer diameter ....................... 19,1 mm / 0.75 in
CTC-140 A inner diameter ....................... 15,0 mm / 0.59 in
CTC-140 A lenght ....................................... 100 mm / 3.9 in
CTC-320 / CTC-650 A outer diameter ..... 25,7 mm / 1.01 in
CTC-320 / CTC-650 A inner diameter ..... 21,5 mm / 0.85 in
CTC-320 / CTC-650 A lenght .................... 120 mm / 4.7 in
CTC-320 / CTC-650 B outer diameter ..... 25,7 mm / 1.01 in
CTC-320 / CTC-650 B inner diameter ..... 21,5 mm / 0.85 in
CTC-320 / CTC-650 B lenght .................... 200 mm / 7.9 in
CTC-1200 A outer diameter ........................ 25 mm / 0.98 in
CTC-1200 A inner diameter .......................... 22 mm / 1.6 in
CTC-1200 A lenght ..................................... 155 mm / 6.1 in
Weight of non-drilled insert (approximate)
CTC-140 A ....................................................... 75 g / 2.6 oz
CTC-320 A ..................................................... 170 g / 5.8 oz
CTC-650 A ................................................... 510 g / 17.8 oz
CTC-320 B ..................................................... 280 g / 9.8 oz
CTC-650 B ................................................... 860 g / 30.3 oz
CTC-1200 A ................................................. 460 g / 16.3 oz
Use of other inserts may reduce performance of the calibrator. To get the best results out of the calibrator, the
insert dimensions, tolerance and material is critical. We
highly advise using JOFRA inserts, as they guarantee
trouble free operation.
Shipping (including optional carrying case)
CTC-140 A ................................................. 12.5 kg / 27.6 lb
CTC-320 A ....................................................... 11 kg / 24 lb
CTC-650 A ....................................................... 12 kg / 27 lb
CTC-320 B .................................................... 13.5 kg / 21 lb
CTC-650 B ....................................................... 17 kg / 37 lb
CTC-1200 A ..................................................... 18 kg / 39 lb
Size L x W x H: ..507 x 232 x 415 mm / 19.9 x 9.1 x 16.3 in
Shipping (without carrying case)
CTC-140 A ....................................................... 10 kg / 22 lb
CTC-320 A ...................................................... 8 kg / 17.5 lb
CTC-650 A ....................................................... 9.5 kg /21 lb
Size L x W x H: 465 x 255 x 470 mm / 18.3 x 10.0 x 18.5 in
CTC-320 B ....................................................... 11 kg / 24 lb
CTC-650 B ....................................................... 14 kg / 26 lb
CTC-1200 A ..................................................... 15 kg / 32 lb
Size L x W x H: 465 x 255 x 470 mm / 18.3 x 10.0 x 18.5 in
Shipping (carrying case only)
Weight: ........................................................... 5.0 kg / 11 lb
Size L x W x H: ..507 x 232 x 415 mm / 19.9 x 9.1 x 16.3 in
6 www.jofra.comCompact Temperature Calibrator • CTC
INSERTS FOR CTC SERIES
Inserts for CTC-140 A and CTC-320 A/B are made of aluminum. Inserts for CTC-650 A/B are made of
brass. Inserts for CTC-1200 A are made of high-temperature steel alloy.
All specifications on hole sizes are referring to the outer diameter of the sensor-under-test.
The correct clearance size is applied in all predrilled inserts.
Inserts, undrilled
Instruments
2
InsertsCTC-140 A
5-pack, undrilled
inserts
Undrilled
insulation plug
60F44810017510019460F35660F420124403
123937N/AN/AN/AN/Asee below
CTC-320 ACTC-650 ACTC-320 BCTC-650 BCTC-1200 A
Undrilled inserts
(CTC-140 A)
4
4
Spare part no. for predrilled inserts - metric (mm)
Instruments
Probe diameterInsert code
3 mm003
4 mm004
5 mm005
6 mm006
7 mm007
8 mm008
9 mm009
10 mm010
11 mm011
12 mm012
13 mm013
14 mm014
15 mm015
16 mm016
18 mm018
20 mm020
Package of the above inserts
Multi-hole type 1M01
Spare part no. for predrilled inserts - imperial (inch)
Probe diameterInsert code
1/8 in125
3/16 in187
1/4 in250
5/16 in312
3/8 in375
7/16 in437
1/2 in500
9/16 in562
5/8 in625
11/16 in688
3/4 in750
13/16 in813
7/8 in875
Package of the above inserts
Multi-hole type 2M02
1
CTC-140 A 2CTC-320 ACTC-650 ACTC-320 BCTC-650 BCTC-1200 A
123428123436123444N/AN/A124503
60F45110017710019660F35960F423124406
123429123437123445123452123460124504
60F45310017910019860F36160F425124407
123430123438122516123453123461124505
105185100182100201105190105195124506
105186100183100202105191105196124507
105187100185105188105192105197124508
123431100188100204105193105198124509
123432100186100206105194105199124510
12343360F339105189123454123462N/A
N/A100190100208123455123463N/A
N/A100191100209123456123464N/A
N/A123439123446123457123465N/A
N/A123440122517123458123466N/A
N/A123441122518123459123467N/A
1246 79124 6811246 851246831246 87124 68 9
3
123479
1
CTC-140 A 2CTC-320 ACTC-650 ACTC-320 BCTC-650 BCTC-1200 A
60F45010017610019560F35860F422124511
60F45210017810019760F36060F424124512
60F45410018010019960F36260F426124404
60F45610018110020060F36460F428124513
60F45810018410020360F36660F430124514
60F46010018710020560F36860F432124515
60F46210018910020760F37060F434124405
60F46460F34460F40860F37260F436N/A
60F46610019210021060F37460F438N/A
N/A60F34860F41260F37660F440N/A
N/A10019310021160F37860F442N/A
N/A60F35260F41610518460F444N/A
N/A60F35460F41860F37760F446N/A
124680124 68 21246861246 8412468812469 0
123480
123475123476N/AN/AN/A
Instruments
3
123477123478N/AN/AN/A
Note 1: Use the insert code, when ordered as the standard insert together with a new calibrator.
Note 2: CTC-140 A only: Remember to use matching insulation plugs (see accessories).
Note 3: CTC-140 A only: All multi-hole inserts are delivered with a matching insulation plug.
Note 4: CTC-1200 A only: Remember to order matching insulation plugs (see accessories).
Undrilled inserts
(CTC-320 / 650 A/B)
4
Undrilled inserts
(CTC-1200 A)
4
Multi-hole type 1
(CTC-140 A)
3
4
6
Multi-hole type 1
(CTC-320 A / 650 A)
3/16
1/8
Multi-hole type 2
(CTC-140 A)
1/8
Multi-hole type 2
(CTC-320 A / 650 A)
5
3/8
1/4 in
4,5
3
6
9
4
5
1/4
3/16
www.jofra.com 7CTC • Compact Temperature Calibrator
ORDERING INFORMATION
Order number Description
Base model number
CTC140A CTC-140 A, -17 to 140°C / -1 to 284°F
CTC320A CTC-320 A, 33 to 320°C / 91 to 608°F
CTC650A CTC-650 A, 33 to 650°C / 91 to 1202°F
CTC320B CTC-320 B, 33 to 320°C / 91 to 608°F
CTC650B CTC-650 B, 33 to 650°C / 91 to 1202°F
CTC1200A CTC-1200 A, 300 to 1205°C / 572 to 2200°F
Power supply (US deliveries 60 Hz only)
115 115VAC
230 230VAC
Mains power cable type
A European, 230 V,
B USA/CANADA, 115 V
C UK, 240 V
D South Africa, 220 V
E Italy, 220 V
F Australia, 240 V
G Denmark, 230 V
H Switzerland, 220 V
I Israel, 230 V
Insert type and size
XXX 1 x Insert for dry-block configuration
(please see the previous insert pages for the
right insert codes)
Calibration certificate
F NLP Traceable temperature certificate
(standard for Europe, Asia, Australia and Africa)
G NIST traceable temperature certificate
(standard for Western Hemisphere)
H Accredited certificate with 5 std. points
(except CTC-1200 A)
H Accredited certificate with 4 std. points
(for CTC-1200 A)
Options
C Carrying case
X No option used
CTC650A230AM01FX Sample order number
CTC-650 A dry-block, 230 VAC power with
European power cord and insert:
Pre-drilled multi-hole type 1 (1 x 3 mm, 1 x 4 mm, 1 x 5 mm, 1 x 6 mm, 1 x 9 mm), and
NLP traceable certificate.
A business unit of AMETEK Measurement & Calibration
AMETEK Test & Calibration Instruments
Technologies Division offering the following industry
leading brands for test and calibration instrumentation.