Ametek CSC200 Data sheet

s i g n a l
Specification Sheet
SS-CP-2332-US
Model CSC200
Compact Signal Calibrator
Input and Output
RTD: 14 types, TC:13 types Resistance: 0 to 4,000 5 to 4,000 mV: -10 to 75 mV
Ω (read) Ω (source)
High level of protection
Fuse-less protection for internal circuitry
- a common cause of failure in other units without this protection
CJC
Selectable feature - use with or without cold junction compensation depending on application
Setpoints
User-defined - make testing easier by storing the necessary settings for repeated test points and sequences
High accuracy
Accuracy up to ± 0.2°C / 0.36°F - including programmed curves for easy use in most any application
Easy connections
Jacks for 2, 3, or 4-wire RTDs and a dedicated TC plug - allows for connection of sensur-under-test
Complete marine program
Part of a complete program of marine approved temperature, pressure and signal calibrators; including temperature sensors See more at www.jofra.com
PRODUCT DESCRIPTION
The JOFRA CSC200 handheld calibrator offers you an easy-to­operate interface including a knob for precise and rapid adjustment. With features including a large display and an auto stepping sig­nal, the CSC200 makes trouble­shooting and calibration of your temperature sensors a snap.
Features
The CSC line of signal calibrators are dedicated units designed to make your calibration and maintenance tasks easier to perform. The JOFRA CSC200 is engineered for the calibration of temperature instruments offering the ability to measure and simulate several different types of RTDs and TCs. This instrument makes the job easy by ulitizing a thermo­couple plug for easy connection and can work with 2, 3, or 4-wire RTDs.
The CSC200 also offers features such as user-defined setpoints with a recall function, selectable cold junction compensation, and installed curves for 14 RTD and 13 TC types. These handy time savers are com­bined in a package offering an easy-to-read, two line display and an intuitive keypad with a knob for fine adjustments.
This calibrator is more than a simple tool but it is as straightforward as any device that you have in your toolbox. The CSC200 signal calibrator offers you the accuracy and features that you need in a convenient, com­pact, and easy-to-use package.
ISO 9001 Manufacturer
Specification Sheet
SS-CP-2332-US
Model CSC200 Compact Signal Calibrator
2 www.jofra.com
Clear dual line display
The CSC200 features a large two-line display. The upper line shows the measured or sourced temperature, resis­tance, or mV value. The second line indicates the type of sensor. The display also features several icons to allow for quick recognition of the status of the unit.
Auto shutdown
The CSC200 offers an automatic shutdown feature. This power saving feature turns the calibrator off after it has been idle for a specified period of time. The time delay is factory set to 30 minutes; but can be adjusted to meet specific needs.
RTD modes
The CSC200 offers the ability to source and read 14 types of RTDs. The formulas (curves) for the different types of RTDs are programmed into the firmware of this unit. This makes it possible to select the curve mach­ing the sensor and perform the test routine without the need for math­ematical conversions. The CSC200 can be used with 2, 3, or 4-wire RTDs making it flexible enough for virtually all RTD calibrations.
TC modes
The CSC200 is also programmed with the formulas for 13 TC types. Like the RTD modes, the unit will source or read TC sensors and provide a stable and reliable reading. The unit has a convenient TC plug for easy connec­tion to the TCs and is also equipped with CJC: the CJC may be turned off it desired.
CSC200 as TC input with an ITC
CSC200 as RTD input with an ITC
Type
This is used to scroll through the various types of sensors that are installed in the CSC200 firmware.
Electrical connections
Inputs for 2, 3, and 4-wire RTD probes, TC jack, and switch test connections.
I/O
Selects whether the CSC200 is reading or sourcing values.
Set and Recall operation
The CSC200 offers you a timesaving feature that is not offered amongst lesser calibrators. You can store up to 3 test points for each of the TC and RTD functions. This means that there is no need to repeatedly set each value for your test needs; just store the setpoints for your test and call them up for each sensor. The Recall button is your “hot key” to retrieve the stored values in sequential order. If you have several types of sensors in your process, the CSC200 makes it easy because you can store set val­ues for each type.
Left / Right arrows
Move the cursor left or right to allow for adjustment of the desired decade.
Adjustment Knob
Allows for easy adjustment of values. Combined with the Arrow Keys, reaching the desired value is easy.
FUNCTIONAL SPECIFICATIONS
Display
LCD .......................................................Dual line plus icons
Display resolution ..................................................... 5 digits
Display update ...................................... 4 times per second
Temperature range
Operating ..................................... -10 to 50°C / 14 to 122°F
Storage .........................................-20 to 60°C / -4 to 140°F
Power supply
Battery ......................................................... 1 x 9 V Alkaline
Battery life ..............................................Minimum 20 hours
Low battery indicator ..................................................at 7 V
Instrument dimensions
L x W x H ................. 188 x 84 x 52 mm / 7.4 x 3.3 x 2.1 in
Weight (including battery): ...........................400 g / 14.1 oz
Shipping dimensions
L x W x H ............... 203 x 101 x 64 mm / 8.0 x 4.0 x 2.5 in
Weight (including battery): ..............................567 g / 20 oz
Model CSC200 Compact Signal Calibrator
Specification Sheet
SS-CP-2332-US
Thermocouple - TC
TC types ............................... B C E J K L N R S T U BP XK
Cold Junction Compensation ON/OFF control: ............ Yes
Thermocoupe mV
TC mV read
TC mV source
Maximum current output is 1 mA with an output impedance of = 1 Ω
Thermocouple Cold junction
CJC compensation
CJC outside above
TC Type
From To From To
B
C
E
J
K
L
N
R
S
T
U
XK
BP
600.0 800.0 1112.0° 1472.0 1.40 2.52
800.0 1000.0 1472.0 1832.0 1.50 2.70
1000.0 1820.0 1832.0 3308.0 1.70 3.06
0.0 1000.0 32.0 1832.0 0.80 1.44
1000.0 2316.0 1832.0 4200.8 2.50 4.50
-250.0 -100.0 -482.0 -148.0 0.80 1.44
-100.0 1000.0 -148.0 1832.0 0.40 1.72
-210.0 0.0 -346.0 32.0 0.60 1.08
0.0 800.0 32.0 1472.0 0.40 0.72
800.0 1200.0 1472.0 2192.0 0.50 0.90
-200.0 0.0 -346.0 32.0 0.80 1.44
0.0 1000.0 32.0 1832.0 0.50 0.90
1000.0 1372.0 1832.0 2501.6 0.70 1.26
-200.0 0.0 -328.0 32.0 0.45 0.81
0.0 900.0 32.0 1652.0 0.40 0.72
-200.0 0.0 -328.0 32.0 1.00 1.80
0.0 1300.0 32.0 2372.0 0.60 1.08
0.0 1767.0 32.0 3212.6 1.40 2.52
0.0 1767.0 32.0 3212.6 1.40 2.52
-250.0 0.0 -418.0 32.0 0.80 1.44
0.0 400.0 32.0 752.0 0.40 0.72
-200.0 0.0 -328.0 32.0 0.70 1.26
0.0 600.0 32.0 1112.0 0.45 0.81
-200.0 800.0 -328.0 1472.0 0.40 0.72
0.0 800.0 32.0 1472.0 1.10 1.98
800.0 2500.0 1472.0 4532.0 2.50 4.50
-10.000 mV 75.000 mV 0.02% rdg ±10 µV
-10.000 mV 75.000 mV 0.02% rdg ±10 µV
18°C (64°F) 28°C (83°F) 0.20°C (0.36°F)
Temperature range
°C °F °C °F
Accuracy includes CJC error, does not include wire error. CJC error outside of 23°C ±5°C is 0.05°C/ °C / 73.4°F ± 9°F is
0.05°F/°F. Accuracies are stated for operation at 23°C ±5°C / 73.4°F ±9°F. For operation < 18°C or > 28°C add ± 0.005% of reading /°C / <
64.4°F or > 68.8°F add ±0.003% of reading / °F.
Range Accuracy±
min max 12 months
Range Accuracy ±
min max 12 months
0.05°C (0.09°F)
12 month accuracy
Resistance Temperature Detector - RTD
RTD Types: Pt10, Pt25, Pt50, Pt100, Pt200, Pt500, Pt1000
.................................... Cu10, Cu50, Cu100, Ni120, YSI400
Response time: ..................................................... 5 mSec
Connection ..................................................2, 3, and 4-wire
Ohm Range Accuracy±
min max 12 months
Ohm read (low)
Ohm read (high)
Ohm source (low)
0.1 to 0.5 mA
0.5 to 3 mA
Ohm source (high)
0.05 to 0.8 mA
0.05 to 0.4 mA Unit is compatible with pulsed transmitters - Frequency response is 5 m
Sec.
0.0 Ω
400.1 Ω 4000.0 Ω 0.025% rdg. ±0.5 Ω
5.0 Ω
5.0 Ω
400.0 Ω
1500.0 Ω
RTD Type Temperature range
°C °F °C °F
From To From To
-200.0 100.0 -328.0 212.0 1.40 2.52
Pt10 Alpha 385
Pt50 Alpha 385
Pt100 Alpha 385
Pt100 Alpha 3926
Pt100 Alpha 3916
Pt200 Alpha 385
Pt500 Alpha 385
Pt1000 Alpha 385
Cu10
Cu50
Cu100
YSI400
100.0 300.0 212.0 572.0 1.60 2.88
300.0 600.0 572.0 1112.0 1.80 3.24
600.0 800.0 1112.0 1472.0 2.00 3.60
-200.0 100.0 -328.0 212.0 0.40 0.72
100.0 300.0 212.0 572.0 0.50 0.90
300.0 600.0 572.0 1112.0 0.60 1.08
600.0 800.0 1112.0 1472.0 0.70 1.26
-200.0 100,0 -328.0 212.0 0.20 0.36
100.0 300.0 212.0 572.0 0.30 0.54
300.0 600.0 572.0 1112.0 0.40 0.72
600.0 800.0 1112.0 1472.0 0.50 0.90
-200.0 100.0 -328.0 212.0 0.20 0.36
100.0 300.0 212.0 572.0 0.30 0.54
300.0 630.0 572.0 1166.0 0.40 0.72
-200.0 100.0 -328.0 212.0 0.20 0.36
100.0 300.0 212.0 572.0 0.30 0.54
300.0 630.0 572.0 1166.0 0.40 0.72
-200.0 100.0 -328.0 212.0 0.80 1.44
100.0 300.0 212.0 572.0 0.90 1.62
300.0 630.0 572.0 1166.0 1.00 1.80
-200.0 100.0 -328.0 212.0 0.40 0.72
100.0 300.0 212.0 572.0 0.50 0.90
300.0 630.0 572.0 1166.0 0.60 1.08
-200.0 100.0 -328.0 212.0 0.20 0.36
100.0 300.0 212.0 572.0 0.30 0.54
300.0 630.0 572.0 1166.0 0.40 0.72
-100.0 260.0 -148.0 500.0 1.40 2.52
-180.0 200.0 -292.0 392.0 0.40 0.72
-180.0 200.0 -292.0 392.0 0.30 0.54
15.0 50.0 59.0 122.0 0.10 0.18
400.0 Ω 0.025% rdg. ±0.05 Ω
400.0 Ω
400.0 Ω
1500.0 Ω
4000.0 Ω
0.025% rdg. ±0.02/Iexc
0.025% rdg. ±0.05 Ω
0.025% rdg. ±0.05 Ω
0.025% rdg. ±0.05 Ω
12 month accuracy
Temperature coefficient
Stability: ..... ± 0.005% of reading /°C outside of 23°C ±5°C
..................± 0.003% of reading /°F outside of 73.4°F ±9°F
Read accuracy is based on a 4-wire input, for a 3-wire input add
0.005 Ω assuming all leads are matched. Accuracies are stated for operation at 23°C add ± 5°C / 73.4°F. For operation < 18°C or > 28°C add ± 0.005% of reading /°C / <
64.4°F or > 68.8°F add ±0.003% of reading / °F.
www.jofra.com 3
ORDERING INFORMATION
Order no. Description Base model
CSC200 Handheld Temperature Signal Calibrator
Certification
G NIST traceable temperature certificate (standard) H Accredited calibration certificate
CSC200G Sample order number
CSC200 calibrator with standard NIST traceable temperature certificate
STANDARD DELIVERY
JOFRA CSC200 loop calibrator
Traceable calibration certificate
Reference manual
Test leads
Soft carrying case
Shoulder strap
ACCESSORIES
120519 Thermocouple Male Plug - Type Cu-Cu - White 120518 Thermocouple Male Plug - Type R / S - Green 120517 Thermocouple Male Plug - Type K - Yellow 120516 Thermocouple Male Plug - Type J - Black 120515 Thermocouple Male Plug - Type T - Blue 120514 Thermocouple Male Plug - Type N - Orange 121983 Extension Cable for Type K - 5 m 122523 Extension Cable for Type N - 5 m 2206011 Thermocouple plug + K wire + alligator 2206012 Thermocouple plug + T wire + alligator 65-PT100-LB-CABLE Cable 2 m (6.6 ft.) with LEMO / Banana connectors 124717 1x 9 Volt rechargeable battery 124718 Charger for rechargeable batteries - 115/230 VAC
CSC200 in soft case
All JOFRA handheld calibrators are supplied in a handy soft case that allows for opera­tion while in the case. The convenient internal storage pocket offers the space for test leads and accessories.
is one of the world’s leading manufacturers and
AMETEK Calibration Instruments
developers of calibration instruments for
temperature, pressure and process signals
as well as for temperature sensors both from
a commercial and a technological point of view.
JOFRA Temperature Instruments
Portable precision thermometers. Dry-block and
liquid bath calibrators : 4 series, with more than
25 models and temperature ranges from
-90° to 1205°C / -130° to 2200°F. All featuring speed, portabilit y, accuracy and advanced documenting
functions with JOFRACAL calibration software.
JOFRA Pressure Instruments
Convenient electronic systems ranging from
-1 to 1000 bar (25 inHg to 14,500 psi) -
multiple choices of pressure ranges, pumps and
accuracies, fully temperature- compensated
for problem-free and accurate field use.
JOFRA Signal Instruments
Process signal measurement and simulation for
easy control loop calibration and measurement
tasks - from handheld field instruments to
laboratory reference level bench top instruments.
JOFRA / JF Marine Instruments
A complete range of calibration equipment
for temperature, pressure and signal,
approved for marine use.
FP Temperature Sensors
A complete range of temperature sensors
for industrial and marine use.
M&G Pressure Testers
Pneumatic floating-ball or hydraulic piston dead
weight testers with accuracies to 0.015% of reading.
M&G Pumps
Pressure generators from small pneumatic
“bicycle” style pumps to hydraulic pumps
generating up to 1,000 bar (15,000 psi).
.. .beca use calib ratio n is a ma tter of c onfid ence
Headquarters:
AMETEK Denmark A/S Gydevang 32-34 • 3450 Allerød • Denmark Tel: +45 4816 8000 • ametek@ametek.dk
Sales & Service: Europe, Asia, Africa, Middle East and South America
Information in this document is subject to change without notice. ©2008, by AMETEK, Inc., www.ametek.com. All rights reserved.
Pub code SS-C P-233 2-US Is sue 0901
4 www.jofra.com
www.ametekcalibration.com
www.jofra.com
Sales & Service Offices:
AMETEK Mansfield & Green (North America)
Tel: +1 800 527 9999 • cal.info@ametek.com
AMETEK Singapore Pte. Ltd. (Singapore) Tel: +65 6 484 2388 • aspl@ametek.com.sg
AMETEK Inc. Beijing Rep. Office (China)
Tel: +86 10 8526 2111 • jofra@ametek.com.cn
AMETEK GmbH (Germany)
Tel: +49 2159 9136 510 • info.mct-de@ametek.de
AMETEK Calibration Instruments (UK)
Tel: +44 (0) 1489 486 404 • jofra@ametek.co.uk
Loading...