signal
»
Current loop calibrator
With its robust custom housing the CSC101
is ideal for both field and workshop use. It
is an excellent calibrator and a convenient
tool for a process engineer
»
Highest accuracy
The CSC101 is designed to meet the
highest accuracy demands from modern
sensors and transmitters: resolution 1 µA
»
Intuitive one knob operation
The CSC101 is an advanced, but easyto-use loop calibrator, due to its one knob
operation
Compact Signal Calibrator
CSC101
»
Advanced step and ramp functions
User defined time and step time settings up
to 300 seconds and an intuitive “fast” key
step system
»
Small and handy
The size and shape of the CSC101 is
designed to fit into a technician’s hand
»
40 hour battery life
The CSC101 is powered by 6 AAA
batteries, typically giving the unit over 40
hours of battery life
»
High resolution display
Clear and bright display gives you optimal
readout in sunlight or low light conditions
The JOFRA CSC101 is a compact mA loop signal calibrator. You can use
it to calibrate or measure mA as well as loop voltage. The unit is easy-touse, with an intuitive user interface and comprehensive ramp and step
functions. All functions are displayed on a high resolution display. The
high accuracy of the CSC101 has not been achieved on account of
fragile measurement or source circuitry, the unit has fuseless protection
– which might save you a calibrator or lower your cost of ownership.
ISO 9001 Manufacturer
Specification Sheet
SS-CSC101
SPECIFICATIONS
Ambient specifications
Operating temperature ................ -10 to 55°C / 14 to 131°F
Storage temperature ....................-20 to 60°C / -4 to 140°F
Humidity range, non condensing ................... 10 to 95%RH
All specifications specified
at ambient temperature ................... 23°C ±5°C / 73°F ±9°F
Outside ambient 23°C ±5°C .......................±20 ppm FS/°C
Outside ambient 73°F ±9°F ......................... ±12 ppm FS/°F
Display
Resolution ....................................................128 x 64 pixels
Digit size (primary measure/source) ............. 8.6 mm / 0.34”
Back light .......................................................................LED
Source mA
Current measure, source and sink
Range ................................................................. 0 to 24 mA
Span ...................................................0 to 24 or 4 to 20 mA
Accuracy .............................................±0.01% RDG + 2 µA
mA resolution ................................................................1 µA
Over current protection ..............................................28 mA
Hart resistor, software on/off ...................................... 250 Ω
Loop Drive Capacity, no HART resistor .................... 1200 Ω
Loop Drive Capacity, with HART resistor ...................950 Ω
Loop Supply .................................................... 24 VDC ±5%
Voltage measure
Range ............................................................... 0 to 30 VDC
Accuracy ............................................±0.01% RDG + 2 mV
Voltage resolution.........................................................1 mV
Over voltage protection .......................................... 240 VAC
Auto ramp & step
Auto ramp time ................................................ 5 to 300 sec.
Auto step time ................................................. 5 to 300 sec.
Step size ........................................................ 25% / 5 steps
Low battery warning ....................................................... Yes
Valve test
Auto step time ................................................. 5 to 300 sec.
Number of steps ...................................................... 9 steps
Range .......................................................-1.25 to 101.25%
Simulate mA
Measure mA, CSC 101 supplying 24 VDC
Power specifications
Batteries ................................................... 6 x AAA batteries
Battery type ................................. Alkaline, Lithium or NiMH
Battery life ................................................ Approx. 40 hours
Auto power off ................... 1 to 30 min., or permanently on
Low battery warning ....................................................... Yes
Physical specifications (LxHxW)
Measure mA
Instrument ................... 150 x 90 x 30 mm / 6 x 3.6 x 1.3 in
Weight ...........................................................0.3 kg / 9.5 oz
Shipping size ...... 285 x 110 x 160 mm / 11.2 x 4.3 x 6.3 in
Shipping weight.........................................1.23 kg / 43.4 oz
Miscellaneous
CE - EMC ....... EN 61326, ANNEX A (Portable Instruments)
Accuracy includes hysteresis, nonlinearity, repeatability, reference
standard uncertainty and 1 year typical long-term stability;
operated inside the rated temperature span and range.
2 www.jofra.com Compact Signal Calibrator • CSC101