The use of advanced technology and the high quality standard of our
instruments are the result of continuous development. This may result in
differences between this operating manual and your instrument. Also, we
cannot guarantee that there are absolutely no errors in this manual.
Therefore, we are sure you will understand that we cannot accept any legal
claims resulting from the data, figures or descriptions.
The BYK LC 4 hand-held meter and a suitable sensor module enable you to
measure the conductivity, specific resistance, salinity or TDS (total dissolved
solids) of a solution very easily.
Depending on the measuring range expected, the following sensor modules
are suitable for operation with the BYK LC 4 hand-held meter.
Measuring rangeSensor module
1 µS/cm ... 200 mS/cmTetraCon V
All functions and settings are operated via the touch screen.
B
Y
K
1
C
1 3 1 4
2 5 8
ESC
L
C
4
A
L
Tr
e
f
2
5
nLF
M
E
M
S/cm
C
2
1Display (touch screen)
2Sensor module
5
Page 6
OverviewBYK LC 4
6
Page 7
BYKLC4Safety
2Safety
This operating manual contains basic instructions that you must follow during
the commissioning, operation and maintenance of the meter. Consequently,
all responsible personnel must read this operating manual before working
with the meter. The operating manual must always be available within the
vicinity of the meter.
User qualificationThe meter was developed for work in the laboratory.
We assume that, as a result of their professional training and experience, the
operators will know the necessary safety precautions to take when handling
chemicals.
Safety
instructions
Further notes
The individual chapters of this operating manual use safety instructions such
as the label shown below to indicate various hazards or dangers:
Caution
indicates instructions that must be followed precisely in order to avoid
the possibility of slight injuries or damage to the instrument or the environment.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
2.1Authorized use
The authorized use of the meter consists exclusively of the measurement of
the conductivity, salinity, temperature and TDS (total dissolved solids).
The technical specifications as given in chapter 7 T
observed. Only the operation and running of the meter according to the
instructions given in this operating manual is authorized.
Any other use is considered to be unauthorized.
ECHNICALDATA must be
7
Page 8
SafetyBYK LC 4
2.2General safety instructions
This instrument is built and inspected according to the relevant guidelines
and norms for electronic meters (see chapter 7 T
It left the factory in a safe and secure technical condition.
ECHNICALDATA).
Function and
operating safety
The smooth functioning and operational safety of the meter can only be guaranteed if the generally applicable safety measures and the specific safety
instructions in this operating manual are followed during operation.
The smooth functioning and operational safety of the meter can only be guaranteed under the environmental conditions that are specified in chapter 7
T
ECHNICALDATA.
If the instrument was transported from a cold environment to a warm environment, the formation of condensate can lead to the faulty functioning of the
instrument. In this event, wait until the temperature of the instrument reaches
room temperature before putting the instrument back into operation.
Safe operationIf safe operation is no longer possible, the instrument must be taken out of
service and secured against inadvertent operation!
Safe operation is no longer possible if the meter:
! has been damaged in transport
! has been stored under adverse conditions for a lengthy period of time
! is visibly damaged
! no longer operates as described in this manual.
Obligations of the
purchaser
8
If you are in any doubt, please contact the supplier of the instrument.
The purchaser of this meter must ensure that the following laws and guidelines are observed when using dangerous substances:
! EEC directives for protective labor legislation
! National protective labor legislation
! Safety regulations
! Safety datasheets of the chemical manufacturers.
Page 9
BYKLC4Commissioning
3Commissioning
3.1Scope of delivery
! BYK LC 4 hand-held meter
! Sensor module (TetraCon V)
! Accompanying documents:
– Manual for commissioning
– Short instructions to be glued in the lid of the case
– CD-ROM with a description of all instrument functions
! Transport case with insert
3.2Initial commissioning
Perform the following activities:
! Connect the sensor module to the meter (see section 3.3)
! Switch on the meter (see section 4.2)
! Set the date and time (see section 4.10.1)
! Calibrate the meter with the sensor module (see section 4.4)
9
Page 10
CommissioningBYK LC 4
3.3Connecting the sensor module
1Unscrew the coupling ring (1) from the meter.
2Place the sensor module (2) in a straight position to the meter (turn
the wide cut-out to the front) and plug it in carefully with slight force.
2
1
3Using the coupling ring (1), screw the sensor module tight.
The instrument is ready for operation.
10
Page 11
BYKLC4Commissioning
3.4Storing the meter and sensor module
To maintain the operability of the meter and sensor module for a long time,
leave the sensor module connected to the meter as possible and avoid to
plug it on or off when it is not necessary. Thus, the watertight plug-in
connector will be well protected.
In order to store an already used sensor module separately from the meter,
protect the socket of the sensor module with the original protective cap.
11
Page 12
CommissioningBYK LC 4
12
Page 13
BYKLC4Operation
4Operation
4.1Operating elements
The entire operation is carried out via the touch sensitive surface of the
display (touch screen). A slight pressure on the surface suffices to start a
function. Any pressure on the surface is confirmed by an acoustic signal.
Tref20Tref20
Display elements
1
2
CellCALLAP
TDS
S
1 3 1 4
CellTC
1
5 5 8
LoBat
Day.Month
ESC
Tref25
nLF
Lin
h:min
CAL
STO
RCL
SAL
min:s
CLR
RESET
TC
MEM
7
O
2
mg/lmbarm
S/cm
M
cm
cm
k
C
1
cm
Sal
Lap
Ident
6
K
5
3
4
1Upper display line - example: stored conductivity value
2Lower display line - example: storing time
3Function keys
4Status display indicators on the current function of the meter:
Example: r = read out memory
5Additional information
Example: h = unit of the indication in the second line
6Sensor symbol
7Status display of stored measuring data
Example: Tref25, nLF
13
Page 14
OperationBYK LC 4
Tre
f25
n
L
F
MEMMEM
1 3 1
m
S/cm
C
2 5
ESC
Tre
f25
n
L
F
CAL
Tre
f25
n
L
F
MEMMEM
1 3 1
m
MEMMEM
S/cm
13
2 5
m
S/
C
14
ESC
2
8
Touch screenThe touch screen contains 8 touch sensitive tactile areas that start different
functions when being pressed, depending on the operating situation.
The function of the keys is connected to the display indication in this area.
CAL
Tre
f25
n
L
F
M
EM
1314
m
S/cm
258
ESC
C
14
Page 15
BYKLC4Operation
Touch screen
operation
In this operating manual, keys are indicated by angle brackets <..>.
The touch screen symbol (example <3>) shows you which area you have
to press. Function keys are indicated by the relevant name or symbol.
The keys can have different functions with a long and short keystroke.
In this operating manual, the key symbol (e. g. <E>) generally means a
short keystroke (shorter than 1 sec). A long keystroke (approx. 1 sec) is indicated by a line after the key symbol (e. g. <E_>).
Operating examples:
SymbolMeaning
<(>Press any spot on the display shortly.
(Example: switch on the meter)
<"_>Press any key in the upper display line for a
longer time (approx. 1 sec).
(Example: switch over the measured value
display between conductivity, specific resistance etc.)
<C>Press "CAL" shortly.
<U>Press the "Clock symbol" shortly.
<E_>Press "ESC" for a longer time (approx. 1 sec).
(Example: switch off the meter)
<<> or <>>Press <<> or <>> shortly.
Increase or reduce the displayed value by 1.
<<_> or <>_>Press <<> or <>> for a longer time (approx.
1sec).
Increase or reduce the displayed value
consecutively.
Note
Any keystroke is confirmed by an acoustic signal.
With a short keystroke you hear a short sound, after a long keystroke you
hear a longer sound.
15
Page 16
OperationBYK LC 4
4.2Switching on the meter
1Switch on the meter with <(>.
After a short display test, a measured value is displayed.
If no sensor module is connected, the Err4 error message appears
on the display (see chapter 6 W
Note
The meter has an energy saving feature to avoid unnecessary battery depletion. The energy saving feature switches the meter off if no entry is made
during the specified switch off-interval (see section 4.11).
HATTODOIF...).
16
Page 17
BYKLC4Operation
4.3Measuring
4.3.1General information
Depth of
immersion
AutoRead
(drift control)
Temperature
compensation
Reference
temperature Tref
For measurements in a measuring solution we recommend an immersion
depth of approx. 36 mm for optimum and reproducible measurement results.
The sensor module may only be submersed in a measuring solution up to
below the coupling ring. If it is submersed deeper, liquid can penetrate the
meter and damage it.
The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of
measured values. The AutoRead function is automatically active. The unit of
the measured value flashes on the display as long as the stability criterion is
not met. A stable measured value is achieved when the unit of the measured
parameter no longer flashes.
The meter has a linear and nonlinear temperature compensation (see section
4.5 S
ETTINGTHETEMPERATURECOMPENSATION). The temperature compensa-
tion can be switched off.
The reference temperature (Tref) can be switched over between 20 °C and
25 °C. It appears on the display as Tref20 or Tref25. To switch over the reference temperature, see section 4.6 S
ETTINGTHEREFERENCETEMPERATURE.
Caution
Special measuring media (e.g. strong acids and lyes, organic solvents) or too
high temperatures shorten the lifetime considerably or cause damage. The
warranty does not cover damage caused by such measuring media and
mechanical damage.
17
Page 18
OperationBYK LC 4
4.3.2Carrying out measurements
1Make sure a valid calibration exists for the measuring system (see
section 4.4.2), or that the cell constant is correctly adjusted (see
section 4.4.2).
2Immerse the sensor module in the test sample.
The display shows the main measured value (conductivity, specific
resistance, salinity or TDS) and the measured temperature value.
The unit of the measured value display flashes (AutoRead) as long
as the measured value is not stable.
3If necessary, switch over the display between conductivity, specific
resistance, salinity and TDS with <"_> [<Measured value>].
Setting menu,
measured
parameter
Tref20Tref20
Cell
CAL
LAP
TDS
S
1 3 1 4
CellTC
8
LoBat
Day.Month
ESC
RESET
Tref25
TC
nLF
MEM
Lin
SAL
2 5 8
h:minmin:s
CAL
STO
RCL
CLR
Lap
k
1
Sal
O
mg/lmbarm
S/cm
M
C
cm
Ident
CAL
LAP
2
cm
cm
K
<"_>
CellTC
LoBat
<"_>
Tref20Tref20
Cell
CAL
LAP
TDS
S
13
CellTC
8
LoBat
Day.Month
ESC
RESET
Tref25
TC
nLF
MEM
Lin
SAL
3 1 4
2 5 8
h:minmin:s
CAL
STO
RCL
CLR
Lap
O
mg/l
mbar
mS/cm
Mk
C
1
cm
Sal
Ident
CAL
LAP
2
cm
cm
K
<"_>
CellTC
LoBat
Tref20Tref20
Cell
Tref25
nLF
Lin
TDS
SAL
S
0 7 6 1
8
2 5 8
Day.Month
h:minmin:s
CAL
STO
ESC
RCL
CLR
Tref20Tref20
Cell
Tref25
nLF
Lin
TDS
SAL
S
1
0 2
8
2 5 8
Day.Month
h:minmin:s
CAL
STO
ESC
RCL
CLR
RESET
TC
MEM
O
2
mg/lmbar
mS/cm
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
<"_>
RESET
TC
MEM
O
2
mg/l
mbar
S/cm
m
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
18
4Wait for a stable measured value.
As soon as the measured value is stable, the unit of the measured
parameter no longer flashes.
Page 19
BYKLC4Operation
4.3.3Setting the TDS factor
1If necessary, press <"_> [<Measured value>] repeatedly until the
measured parameter TDS is displayed (see section 4.3.2).
2Using <ß> [<Temperature>], display the adjusted TDS factor.
3Using <ß> [<TDS factor>], select the TDS factor.
The TDS factor flashes.
Tref20Tref20
CAL
LAP
S
CellTC
LoBat
Cell
TDS
1
8
Day.Month
ESC
RESET
Tref25
TC
nLF
MEM
Lin
SAL
3 1 4
1 0 8
h:minmin:s
CAL
STO
RCL
CLR
Lap
O
mg/l
mbarmS/cm
M
k
C
1
Sal
Ident
2
cm
cm
cm
K
4Using <<> or <>>, set the TDS factor (0.40 ... 1.00).
The measured TDS value display is always updated while doing so.
5Confirm the TDS factor with <g>.
The TDS factor no longer flashes.
6Using <ß> [<TDS factor>], display the measured temperature value.
19
Page 20
OperationBYK LC 4
4.3.4Switching over the temperature unit
For reproducible conductivity measurements, it is essential to measure the
temperature of the test sample. A temperature sensor is integrated in the
sensor module.
You can switch the temperature display from °C to °F:
1Toggle the display between °C and °F with <ß_> [<Temperature>].
Setting menu of
the temperature
unit
Tref20Tref20
Cell
CAL
LAP
TDS
S
1 3 1 4
CellTC
8
LoBat
Day.Month
ESC
RESET
Tref25
TC
nLF
MEM
Lin
2 5 87
h:minmin:s
CAL
STO
RCL
SAL
CLR
O
2
mg/lmbarm
S/cm
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
<ß_>
Cell
CAL
LAP
TDS
S
1 3 1 4
CellTC
7 8 4
LoBat
Day.Month
ESC
Tref20Tref20
Tref25
nLF
Lin
SAL
h:minmin:s
CAL
STO
RCL
CLR
RESET
TC
MEM
O
2
mg/lmbar
m
S/cm
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
20
Page 21
BYKLC4Operation
4.4Adjusting/calibrating the cell constant
Why adjust/
calibrate the cell
constant?
The actual cell constant of the sensor module can slightly differ from the
nominal value of the cell constant.
Depending on the required measuring accuracy, you can measure with the
nominal value of the cell constant that was set in the factory (see chapter
7T
ECHNICALDATA), or adjust the value of the cell constant.
You have the following options to adjust the value of the cell constant:
! (Determining and) Adjusting the cell constant manually
The cell constant can be adjusted in the following ranges:
– TetraCon V0.450 ... 0.500 cm
-1
! Calibrating the cell constant (automatic determination)
This method can be selected for the TetraCon V sensor module (range
-1
that can be calibrated: 0.450 ... 0.500 cm
).
Calibration is carried out in the 0.01 mol/l KCl control standard.
Cell constants outside the above mentioned ranges cannot be adjusted or
calibrated.
Both the calibration data and the value of the manually adjusted cell constant
are stored in the sensor module and can be selected and used for measurements.
Note
Aging slightly changes the characteristics of the cell, e. g. by coatings. As a
result, an inexact measured value is displayed. The original characteristics of
the cell can often be restored by cleaning the cell.
When to adjust/
calibrate the cell
constant?
! If the measuring accuracy with the nominal value of the cell constant that
was set in the factory does not suffice (see chapter 7 T
ECHNICALDATA)
! After the calibration interval has expired (when the sensor symbol flashes)
Calibration intervalThe interval for the determination of the cell constant is set to 180 days in the
factory. You can select the interval in the range of 1 ... 999 days (see section
4.8).
You should calibrate at regular intervals (we recommend: every 6 months).
21
Page 22
OperationBYK LC 4
AutoRead
(drift control)
The calibration procedure automatically activates the AutoRead function.
The display of the unit flashes during the AutoRead measurement.
The current AutoRead measurement can be terminated at any time (without
accepting the current value) by pressing <E>.
Calibration dataThe calibration data is stored in the sensor module. It contains the value of
the determined cell constant. The value of the determined cell constant is
displayed after a valid calibration and is accessible any time.
Calibration statusThe meter has a simple display of the calibration status.
Display indicationCalibration status
A valid calibration is available.
flashes
Prompt to calibrate or adjust the cell constant. It is
still possible to measure.
Causes:
– The calibration interval has expired
– The cell constant that can be calibrated was
reset to the delivery condition (e. g. after a
reset)
– The last calibration was erroneous (error
message, Err3)
No sensor symbol on
the display.
The meter measures with an adjusted cell
constant
4.4.1Adjusting the cell constant manually
You can adjust the cell constant manually as follows:
1Open the setting menu for the cell constant with <C>.
The display shows the setting last selected.
2If necessary, press <"_> repeatedly until the CELL display
appears.
22
Page 23
BYKLC4Operation
Tref20Tref20
Cell
LoBat
CALLAP
S
TC
Cell
TDS
Day.Month
ESC
E LC
Tref25
nLF
Lin
SAL
7 80 4
h:minmin:s
CAL
STO
RCL
CLR
RESET
TC
MEM
L
O
2
mg/lmbarmS/cm
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
3Confirm the displayed cell constant with g .
The displayed cell constant is used immediately.
or
Select the cell constant setting with <ß_>.
The cell constant flashes.
Tref20Tref20
Cell
LoBat
CALLAP
S
TC
Cell
TDS
Day.Month
ESC
4 91
4 8 10
Tref25
nLF
Lin
SAL
h:minmin:s
CAL
STO
RCL
CLR
RESET
TC
MEM
3
O
2
mg/lmbarm
S/cm
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
4Adjust the cell constant with <<> or <>>.
Note
You can terminate the setting of the cell constant with <E>.
The data of the terminated setting is not used.
The cell constant adjusted before remains active.
5Confirm the adjusted cell constant with g .
The cell constant no longer flashes.
6Return to the measuring mode with <g>.
The adjusted cell constant is used.
23
Page 24
OperationBYK LC 4
4.4.2Determining the cell constant (calibration in the control standard)
Note
This method of automatically determining the cell constant by calibration in
the 0.01 mol/l KCL control standard solution can only be used for measuring
cells with cell constants in the range 0.450 ... 0.500 cm
-1
(sensor module
TetraCon V).
This is how you can determine the cell constant:
1Open the setting menu for the cell constant with <C>.
The display shows the setting last selected.
2If necessary, press <"_> repeatedly until the CAL display appears.
Tref20Tref20
Cell
Tref25
nLF
Lin
TDS
SAL
S
A LC
1 3 0 5
h:minmin:s
Day.Month
CAL
STO
ESC
RCL
CLR
RESET
TC
MEM
4
Cell
CAL
LAP
O
2
mg/lmbarmS/cm
M
cm
<ß_>
cm
k
C
1
cm
K
Sal
Lap
Ident
S
CellTC
LoBat
TDS
A LC
1
2 5 7
Day.Month
ESC
Cell
LoBat
CAL
LAP
S
TC
Cell
TDS
Day.Month
ESC
A LC
4 6 80
Tref20Tref20
Tref25
nLF
Lin
SAL
h:minmin:s
CAL
STO
RCL
CLR
RESET
MEM
4
TC
O
2
mg/lmbarmS/cm
M
cm
<ß_>
cm
k
C
1
cm
K
Sal
Lap
Ident
CAL
LAP
CellTC
LoBat
3If necessary, switch between the cell constant last calibrated, the
date of the calibration and the calibration temperature with <ß>.
Note
You can return to the measuring mode with <g>.
The displayed calibrated cell constant is used immediately.
4To calibrate:
Immerse the sensor module in the control standard solution.
5Start the measurement with <C>.
The calibrated cell constant is displayed.
The unit of the measured value display flashes (AutoRead) as long
as the measured value is not stable.
Tref20Tref20
Tref25
nLF
Lin
SAL
h:minmin:s
CAL
STO
RCL
CLR
RESET
TC
MEM
4
O
2
mg/lmbarmS/cm
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
24
Note
You can terminate the calibration with <E>.
The data of the terminated calibration is not used.
Page 25
BYKLC4Operation
6Wait until the measured value is stable.
The unit of the measured value display stops flashing.
7Return to the measuring mode with <g>.
The calibrated cell constant is used.
Note
If no valid calibration can be accomplished, the Err3 error message appears
on the display. Measuring is not possible in this condition. Possible causes
and actions for error elimination can be found in chapter 6 W
HATTODOIF....
25
Page 26
OperationBYK LC 4
4.5Setting the temperature compensation
The calculation of the temperature compensation is based on the preset
reference temperature, Tref20 or Tref25 (see section 4.6 S
ENCETEMPERATURE).
You can select one of the following temperature compensation methods:
! Nonlinear temperature compensation (nLF)
! Linear temperature compensation (Lin) with selectable coefficients of
0.001 ... 3.000 %
/K
! No temperature compensation (- - - -)
Note
Select the following temperature compensations given in the table according
to the respective test sample:
ETTINGTHEREFER-
Application tips
Measuring solutionTemperature compensation
(TC)
Natural water (ground
water, surface water
and drinking water)
nLF
according to ISO 7888, EN
(*)
27888
Ultrapure waternLF
according to ISO 7888, EN
(*)
27888
Other aqueous solutions
Set linear
temperature coefficient 0.001
... 3.000 %/K
Salinity
(seawater)
Automatic nLF according to
IOT
(*) Corresponds to DEV C8
Display
indicator
Sal,
26
Page 27
BYKLC4Operation
4.5.1Selecting the temperature compensation
1Open the setting menu for the cell constant with <C>.
The display shows the setting last selected.
2Open the setting menu for the temperature compensation with <T>.
The display shows the setting last selected.
3If necessary, press <ß_> repeatedly until the nLF, Lin or OFF display
appears.
– nLF = nonlinear temperature compensation
– Lin = linear temperature compensation
– OFF = temperature compensation switched off
Tref20Tref20
Setting menu for
the temperature
compensation
Cell
LoBat
CAL
LAP
TC
S
TDS
0
Day.Month
ESC
Cell
RESET
Tref25
TC
nLF
MEM
Lin
SAL
r 2t
5
n L F
h:minmin:s
CAL
STO
RCL
CLR
Tref20Tref20
Cell
CAL
Cell
LoBat
LAP
TC
S
TDS
2
Day.Month
ESC
O
2
mg/l
mbar
S/cm
m
M
cmcm
k
C
1
cm
K
Sal
Lap
Ident
<ß_>
RESET
Tref25
TC
nLF
MEM
Lin
SAL
5
r 2t
L i n
h:minmin:s
CAL
STO
RCL
CLR
O
2
mg/l
mbar
S/cm
m
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
<ß_>
Tref20Tref20
Cell
LAP
TDS
Tref25
nLF
Lin
SAL
CAL
S
Cell
TC
0
O F F
LoBat
Day.Month
h:minmin:s
CAL
STO
ESC
RCL
CLR
4– Confirm the setting with <g>.
in the nLF menu --> back to the measuring mode
–in the Lin menu --> the menu for setting the temperature coef-
ficient opens up (see section 4.5.2)
–in the OFF menu --> back to the measuring mode
The displayed setting is used immediately.
RESET
TC
MEM
02rt
<ß_>
O
2
mg/lmbar
S/cm
m
M
cmcm
k
C
1
cm
K
Sal
Lap
Ident
27
Page 28
OperationBYK LC 4
4.5.2Setting the linear temperature coefficient
1Select the linear temperature compensation (Lin menu) (see section
4.5.1).
2Open the setting of the linear temperature coefficient with <ß>.
Tref20Tref20
Cell
LoBat
CALLAP
S
TC
Cell
TDS
2
Day.Month
ESC
RESET
Tref25
TC
nLF
MEM
Lin
SAL
7
3 81
1 1 8
h:minmin:s
CAL
STO
RCL
CLR
Lap
k
1
Sal
O
mg/lmbarm
M
C
Ident
2
S/cm
cm
cm
cm
K
3Set the temperature coefficient with <<> or <>>.
Note
You can terminate the setting of the temperature coefficient with <E>.
The data of the terminated setting is not used.
The temperature coefficient set before remains active.
4Confirm the adjusted value with <g>.
5Return to the measuring mode with <g>.
The adjusted temperature coefficient is used.
28
Page 29
BYKLC4Operation
4.6Setting the reference temperature
The setting is only effective if the linear or nonlinear temperature compensation is switched on.
You can set the reference temperature as follows:
1Open the setting menu for the cell constant with <C>.
The display shows the setting last selected.
2Open the setting menu for the temperature compensation with <T>.
The display shows the setting last selected.
3If necessary, select the nLF or Lin temperature compensation with
<ß_>.The display shows the reference temperature set last.
Setting menu of
the reference
temperature
Tref20Tref20
Cell
LoBat
CALLAP
TC
S
Cell
TDS
0
Day.Month
ESC
RESET
Tref25
TC
nLF
MEM
Lin
SAL
r 2t
5
n L F
h:minmin:s
CAL
STO
RCL
CLR
O
2
mg/lmbar
S/cm
m
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
<ß_>
4Using <"_>, switch between tr20 and tr25.
– tr20 = reference temperature 20 °C
– tr25 = reference temperature 25 °C
The adjusted reference temperature is used.
5Confirm the setting with <g>.
–in the nLF menu --> back to the measuring mode
–in the Lin menu --> the menu for setting the temperature coef-
ficient opens up (see section 4.5.2)
Cell
LoBat
CAL
LAP
S
TC
0
Day.Month
ESC
Tref20Tref20
Cell
Tref25
nLF
Lin
TDS
r 2t
n L F
h:minmin:s
CAL
STO
RCL
RESET
SAL
CLR
TC
MEM
0
Lap
Sal
O
mg/lmbarm
Mk
C
1
Ident
2
S/cm
cm
cm
cm
K
29
Page 30
OperationBYK LC 4
4.7Calibration data
The calibrated cell constant is displayed after calibration (see section 4.4.2)
and is accessible any time.
1Switch on the meter with <(>.
The display test is displayed.
2During the display test:
Display the calibrated cell constant with <C>.
CellCALLAP
TDS
S
A LC
TC
Cell
0 4 6 8
Day.Month
LoBat
h:minmin:s
ESC
Tref20Tref20
Tref25
nLF
Lin
CAL
STO
RCL
RESET
SAL
CLR
TC
MEM
4
Lap
O
mg/lmbarmS/cm
M
k
C
1
Sal
Ident
Tref20Tref20
Cell
CAL
LAP
2
TDS
S
cm
<ß_>
cm
cm
K
CellTC
1 3 0 5
LoBat
Day.Month
ESC
A LC
Tref25
nLF
Lin
SAL
h:minmin:s
CAL
STO
RCL
CLR
RESET
TC
MEM
4
O
2
mg/lmbarmS/cm
M
cm
<ß_>
cm
k
C
1
cm
K
Sal
Lap
Ident
CellTC
LoBat
CALLAP
S
TDS
1
Day.Month
ESC
Tref20Tref20
Cell
Tref25
nLF
Lin
A LC
2 5 7
h:minmin:s
CAL
STO
RCL
3If necessary, switch between the cell constant last calibrated, the
date of the calibration and the calibration temperature with <ß_>.
SAL
CLR
RESET
TC
MEM
4
O
2
mg/lmbarmS/cm
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
30
Page 31
BYKLC4Operation
4.8Setting the calibration interval
The calibration interval determines how many days after the last calibration
the instrument should remind you to recalibrate.
The sensor symbol on the display flashes when the calibration interval has
expired.
1Open the Init initialization menu with <C_>.
Tref20Tref20
CellCALLAP
TDS
S
1 n i t
CellTC
8
LoBat
Day.Month
ESC
Y
Tref25
nLF
Lin
SAL
n o
h:minmin:s
CAL
STO
RCL
CLR
RESET
TC
MEM
O
2
mg/l
mbar
S/cm
m
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
2Press <g> to confirm NO for the initialization.
The setting of the calibration interval is displayed.
The display flashes.
RESET
MEM
I n t
1
08d
min:s
h:min
CLR
mV
pH
C
Lap
Ident
LoBat
CALLAP
d
8
Day.Month
ESC
3Set the calibration interval in days (d) with <<> and <>>.
4Confirm the setting with <g>.
5Switch to the measuring mode with <g>.
31
Page 32
OperationBYK LC 4
4.9Saving
The BYK LC 4 meter has a data memory. Up to 50 measurement datasets
can be stored in it.
A dataset consists of:
! Main measured value (conductivity, specific resistance, salinity or TDS)
! Measured temperature value
! Time and date
! ID number (0000 - 9999)
4.9.1Saving measured data
You can save measured data as follows:
1Open the save menu with <M>.
The memory number of the next free memory location is displayed
under the current measured value.
The X status indicator appears.
2Switch to the display of the ID number with <g>.
The ID number that was last entered appears on the display.
3If necessary, edit the four-digit ID number with <5_> and <6_>
[<ID number>].
Tref20Tref20
Cell
CALLAP
TDS
S
1 3 1 4
CellTC
LoBat
Day.Month
ESC
0 0 10
Tref25
nLF
Lin
SAL
h:minmin:s
CAL
STO
RCL
CLR
RESET
TC
MEM
O
2
mg/lmbar
m
S/cm
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
4Using <<> and <>>, enter the required ID number (0 ... 9999).
The selected display flashes.
32
5Confirm the ID number with <g>.
6Save the measurement dataset with <g>.
or
Cancel the storing procedure without storing by pressing <E>.
Page 33
BYKLC4Operation
If all storage locations are occupied, the StoFull message appears. You have
the following options:
Save the current measured value.
Press <g>
The oldest measured value (storage location 1)
will be overwritten by this
Return to the measuring mode without savingPress <E>
Erase the entire memorySee section 4.9.3
4.9.2Displaying measured data from the memory
Stored measurement datasets always have the main measured value in the
upper display line and a variable display of further elements of the measurement dataset in the lower line.
Display stored datasets as follows:
1Display the memory with <M_>.
The W status indicator appears on the display.
The measurement dataset that was saved last is displayed.
2Scroll through the stored datasets with <<> and <>>.
3If necessary, display further elements of the dataset (ID number,
date, time, temperature) with <ß_> [element of the measurement
dataset].
When scrolling further, this element is displayed in the second line
together with the measured value.
Tref20Tref20
CellCALLAP
TDS
S
1 3 1 4
CellTC
LoBat
Day.Month
ESC
Tref25
nLF
Lin
SAL
52
h:minmin:s
CAL
STO
RCL
CLR
RESET
TC
MEM
1n
O
2
mg/lmbarm
S/cm
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
4Scroll through the stored datasets with <<> <>>.
33
Page 34
OperationBYK LC 4
5If necessary, display further elements of the dataset (ID number,
date, time, temperature) with <ß_> [element of the measurement
dataset].
6Switch to the measuring mode with <E>.
4.9.3Erasing measured data from the memory
To erase all datasets proceed as follows:
1Display the memory with <M_>.
The W status indicator appears on the display.
The measured value that was saved last is displayed.
2Erase the entire memory with <D>.
A security prompt appears.
3Confirm the deletion with <g>.
After the deletion, NO DATA appears on the display.
4Switch to the measuring mode with <E>.
34
Page 35
BYKLC4Operation
4.10Clock
The meter has a clock.
When saving measured values and calibrating, the date and current time are
automatically saved as well.
The clock has the following additional functions:
! Alarm function (see section 4.10.2).
! Stopwatch (see section 4.10.3) and
! Timer (see section 4.10.4).
Note
The automatic switch-off (section 4.11) is not activated as long as the clock
is indicated on the display.
4.10.1 Displaying and setting the date and time
Setting the date
and time
The correct setting of the date and time is important for the following functions
and displays:
! Current date and time
! Calibration date
! Alarm function
! Identification of stored measured values.
Therefore, check the time regularly.
Note
The date and time are reset to 01.01.2004, 00:00 hours with a battery
change. Set the clock after each battery change.
1Switch on the meter with <(>.
After a short display test, a measured value is displayed.
2Press <U> to indicate the time and date on the display.
The v status indicator appears.
3Select the hours or minutes with <3_> or <4_> [<Hours> or
<Minutes>].
The h symbol flashes.
35
Page 36
OperationBYK LC 4
Tref20Tref20
CellCALLAP
TDS
S
1 1 0 9
CellTC
1
LoBat
Day.Month
ESC
RESET
Tref25
TC
nLF
MEM
Lin
SAL
3 0 5
h:min
CAL
STO
RCL
min:s
CLR
Lap
k
1
Sal
O
mg/lmbarm
M
C
Ident
2
S/cm
cm
cm
cm
K
4Set the time with <>> and <<>.
5Select the [<Day> or <Month>] with <5_> or <6_>.
The d symbol flashes.
6Set the date with <>> and <<>.
7Confirm the setting with <g>.
The year is displayed.
8Set the year with <>> and <<>.
9Confirm the setting with <g>.
The time and date are displayed again.
10Switch to the measured value display with <E>.
or
Set the alarm time (see section 4.10.2).
36
Page 37
BYKLC4Operation
4.10.2 Alarm function
With the alarm function you set a point of time (alarm time) at which an alarm
signal should sound. Prerequisite is the correct setting of the time (see
section 4.10.1).
The alarm signal also sounds when the meter is switched off.
Setting the alarm
time
1Switch on the meter with <(>.
After a short display test, a measured value is displayed.
2Press <U> to indicate the time and date on the display.
The v status indicator appears.
3Indicate the alarm time with <U>.
The alarm time appears on the display.
4Select the hours or minutes with <5_> or <6_> [<Hours> or
<Minutes>].
The selected display flashes.
RESET
MEM
A L -
00
00
min:s
h:min
CLR
mV
pH
C
Lap
Ident
LoBat
CALLAP
d
-
Day.Month
ESC
5Set the alarm time with <>> and <<>.
6Confirm with <g>.
The alarm time is set.
37
Page 38
OperationBYK LC 4
Switching on the
alarm function
Switching off the
alarm function
7Switch to the date and time display with <g>.
The alarm function is switched on. The v status indicator flashes.
or
Switch to the date and time display with <E>.
The alarm time is stored. The alarm function is switched off.
8When the alarm time is due and the alarm function is switched on, an
alarm signal can be heard for one minute and the measured value
display flashes.
Switch off the alarm signal with <(>.
9Switch to the date and time display with <g>.
The alarm function is switched off.
1Display the alarm time (see above).
2Switch off the alarm function with <E>.
The display shows the date and time.
The alarm function is switched off.
4.10.3 Stopwatch
With the stopwatch, you measure time periods of up to 59:59
(minutes:seconds). In the case of longer periods the stopwatch starts
counting at 00:00 again.
1Display the stopwatch with <U_>.
The v status indicator appears.
The stopwatch is on 00:00.
2Start the stopwatch with <g>.
The stopwatch starts counting the seconds.
Tref20Tref20
CellCAL
LAP
TDS
S
0 0 0 9
CellTC
8
LoBat
Day.Month
ESC
Tref25
nLF
Lin
SAL
582
h:minmin:s
CAL
STO
RCL
CLR
RESET
MEM
TC
O
2
mg/l
mbar
S/cm
m
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
38
Page 39
BYKLC4Operation
3Display the current time as an intermediate time with <L>.
The intermediate time is displayed under the running stopwatch.
4Stop the stopwatch with <g>.
5Let it continue with <g>.
or
Reset the stopwatch to 00:00 with <R>.
or
Terminate the stopwatch and switch to the measuring mode display
with <E>.
4.10.4 Timer
The timer function can count down the seconds of a specified time interval
(timer intervall) of maximum 59:59 (minutes:seconds). When the specified
timer interval has expired, an alarm signal sounds for one minute and the
display flashes.
1Display the stopwatch with <U_>.
The v status indicator appears.
The stopwatch is on 00:00.
2Select the minutes or seconds with <5_> or <6_> [<Minutes> or
<Seconds>].
The selected display flashes.
Tref20Tref20
CellCALLAP
TDS
S
0 5 0 9
CellTC
8
LoBat
Day.Month
ESC
RESET
Tref25
TC
nLF
MEM
Lin
SAL
258
h:min
CAL
STO
RCL
min:s
CLR
Lap
Sal
Mk
1
O
mg/l
mbar
S/cm
m
C
cm
Ident
2
cm
cm
K
3Set the timer interval (minutes:seconds) with <>> and <<>.
The stopwatch has been converted to a timer.
4Confirm the setting with <g>.
5Start the timer with <g>.
6If necessary, display the remaining time as an intermediate time with
<L>.
39
Page 40
OperationBYK LC 4
7If necessary, stop the timer before the time has expired with <g>
and make it continue with <g>.
8As soon as the time has expired the alarm signal sounds.
Switch off the alarm signal with <(>. The timer goes on counting
with a negative sign.
9Stop the timer with <g>.
Note
If the timer has been stopped you can
! Reset the timer to the adjusted timer interval with <R>.
! Terminate the timer and switch to the measured value display with
<E>.
4.11Switch-off interval
The automatic switch-off function switches off the meter automatically if no
entry is made for a specified time (switch-off interval). This saves the battery
of the meter.
You can specify the switch-off interval yourself. The switch-off interval is from
10 minutes minimum to 24 hours maximum.
1Switch on the meter with <(>.
The display test is displayed.
2During the display test (approx. 2 sec.):
Display the switch-off interval with <M>.
RESET
MEM
O f F
0 4
00
min:s
h:min
CLR
mV
pH
C
Lap
Ident
LoBat
CALLAP
d
t
Day.Month
ESC
40
3Select a switch-off interval with <>> and <<>.
4Confirm the setting with <g>.
The display switches to the measured value display.
The automatic switch-off function switches off the meter if no entry is
made during the switch-off interval.
Page 41
BYKLC4Operation
4.12Reset
The following parameters are stored in the sensor module. The reset function
resets them to the delivery condition.
Sensor module, TetraCon V
Parameter
Adjustable valuesDelivery
condition
Measuring mode, Spec. resistance, Sal, TDS,
with adjusted cell
constant
Cell constant, to be
0.450 ... 0.5000.475
calibrated
Cell constant, to be
0.450 ... 0.5000.475
adjusted
Temperature
nLF, Lin, OffnLF
compensation
linear TC0.001 ... 3.000 %/K2.000 %/K
Tref20 °C, 25 °C25 °C
Note
The calibration data is lost when the measuring parameters are reset. Recalibrate after performing a reset. Stored data are retained.
41
Page 42
OperationBYK LC 4
ESC
Resetting the
measuring param-
eters
Setting the
calibration interval
1Start the INIT initialization menu with <C_>.
2For initialization, select YES with <ß>.
Tref20Tref20
Cell
CALLAP
TDS
S
1 n i t
CellTC
8
LoBat
Day.Month
ESC
RESET
Tref25
TC
nLF
MEM
Lin
SAL
Y E S
h:minmin:s
CAL
STO
RCL
CLR
O
2
mg/lmbar
S/cm
m
M
cm
cm
k
C
1
cm
K
Sal
Lap
Ident
3Confirm with <g>.
The reset is performed. The measuring instrument switches to the
setting of the calibration interval (see section 4.8).
The display flashes.
RESET
MEM
I n t
1
08d
min:s
h:min
CLR
mV
pH
C
Lap
Ident
LoBat
CALLAP
d
8
Day.Month
ESC
42
4Terminate the setting with <E>.
or
Set the calibration interval in days (d) with <<> and <>>.
5Confirm the setting with <g>.
6Switch to the measuring mode with <g>.
Page 43
BYK LC 4Maintenance, cleaning, disposal
5Maintenance, cleaning, disposal
5.1Maintenance
The meter is almost maintenance-free.
The only maintenance activity is exchanging the battery when it is depleted
(LoBat display indicator).
5.2Battery exchange
The LoBat display indicator indicates that the battery should be changed.
The battery is then largely depleted.
Stored calibration data and measuring data are retained when the battery is
exchanged.
Caution
Only use leakproof alkaline manganese batteries.
Replacing the
battery
1
1Unscrew the screw (1) on the back of the meter and remove the lid
of the battery compartment by pulling at the screw.
2Remove the depleted battery from the battery compartment.
43
Page 44
Maintenance, cleaning, disposalBYK LC 4
3Place the new battery (type Mignon AA) in the battery compartment.
Make sure that the poles of the battery are the right way round. The
±signs on the battery must correspond to the ± signs in the battery
compartment.
4Close the battery compartment.
5After the battery exchange the meter switches itself on automatically.
The display to set the time is displayed (see also section 4.10.1).
5.3Cleaning
Occasionally wipe the outside of the meter with a damp, soft cloth.
Caution
The housing components are made out of synthetic materials (polyurethane, ABS and PMMA). Thus, avoid contact with acetone and similar
detergents that contain solvents. Remove any splashes immediately.
Cleaning the
touch screen
Activate the cleaning mode so the instrument will not interpret the contact
with the touch screen as entries.
1Switch the meter off with <E_>.
2Switch the meter on with <(_>.
CLR PAD is displayed. The cleaning mode is active. Short touches
do not start any functions.
3Clean the touch screen.
4Switch to the measuring mode with <(_>.
RESET
MEM
C I r
dap
min:s
h:min
CLR
mV
pH
C
Lap
Ident
LoBat
CALLAP
d
I0
Day.Month
ESC
44
Page 45
BYK LC 4Maintenance, cleaning, disposal
Cleaning the
sensor module
We recommend to clean the sensor module thoroughly, especially before
measuring low conductivity values. Before measuring, thoroughly rinse the
sensor with deionized water.
ContaminationCleaning agentReaction time at
room temperature
Water-soluble substancesDeionized waterAny
Fats and oilsWarm water and
household washing-up
liquid
Heavy contaminationMethylated spiritsMax. 5 minutes
Lime and hydroxide coatingsAcetic acid (10 %)Any
Any
5.4Transport
Activate the cleaning mode. Thus, the instrument will not interpret any
contact with the touch screen as an entry while being transported, e. g. in a
pocket (see chapter 5.3).
5.5Disposal
Dispose of the meter without the battery and the sensor module as electronic
waste at an appropriate collection point.
45
Page 46
Maintenance, cleaning, disposalBYK LC 4
46
Page 47
BYKLC4What to do if...
6What to do if...
Error message
CauseRemedy
0FL
0FL, UFL
UFL
0FL0FL
UFLUFL
Error message
Err3
Err3
Err3Err3
Error message
– The measured value lies outside
the measuring range
– Sensor module defective– Replace sensor module
CauseRemedy
– Erroneous calibration
– The data of the sensor module is
outside the specified characteris-
– Recalibrate
– Reset to default setting (see
section 4.12)
tics.
– Sensor module contaminated– Clean and replace sensor
module if necessary
– Moisture in the plug connection– Dry plug and socket
– Calibration solution not suitable– Check calibration solution
CauseRemedy
Err4
Err4
Err4Err4
Error message
Err5
Err5
Err5Err5
No sensor module– Connect sensor module
– Check the plug connection
– Switch the meter off and then on
again.
– Perform a processor reset (see
section 6.1)
– Replace sensor module
CauseRemedy
Sensor module defective
or
Wrong sensor module
– Check the plug connection
– Switch the meter off and then on
again.
– Perform a processor reset (see
section 6.1)
– Replace sensor module
47
Page 48
What to do if...BYK LC 4
No stable
CauseRemedy
measured value
– Sensor module contaminated– Clean sensor module
– Temperature not stable– Adjust temperature if necessary
– Conductivity too low– Use suitable sensor module
– Organic liquids– Use suitable measuring system
Indication, l
CauseRemedy
– Battery almost empty– Exchange the battery (see
section 5.2)
Meter does not
CauseRemedy
react to keystroke
– Operating condition undefined or
EMC load unallowed
– Perform a processor reset (see
section 6.1)
Sensor symbol
flashes
Message, StoFull
StoFull
StoFullStoFull
CauseRemedy
– Calibration interval expired– Recalibrate the measuring
system
CauseRemedy
– All memory locations are full– Clear the memory (see section
4.9.3)
6.1Processor reset:
The processor reset is a simple action that can be helpful in the case of error
messages or error modes.
1Press <1> and <8> at the same time and release them again.
Stored data and calibration values are retained.
48
Page 49
BYKLC4Technical data
7Technical data
7.1Measuring instrument
Dimensionsapprox. 140 x 80 x 33 mm (without sensor module)
Weightapprox. 115 g (without sensor module, without battery)
Mechanical
structure
Electrical safetyProtective classIII
Test certificatesCE
Ambient
conditions
Power
supply
Guidelines
and norms used
Type of protectionIP 65 (hose-proof)
Storage- 25 °C ... + 65 °C
Operation-10 °C ... + 55 °C
Climatic class2
Battery1 x 1.5 V, type Mignon/AA
Battery life timeMaximum approx. 500 operating hours
EMCEC guideline 89/336/EEC
Instrument safetyEC guideline 73/23/EEC
EN 61326-A1:1998
EN 61010-1 A2:1995
Climatic classVDI/VDE 3540
IP protectionEN 60529:1991
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with the
limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is
likely to cause harmful interference in which case the user will be
required to correct the interference at his own expense.
49
Page 50
Technical dataBYK LC 4
Measuring ranges
and
[µS/cm]
0.0 ... 199.9
0 ... 1999
resolution
[mS/cm]
0.00 ... 19.99
0.0 ... 199.9
Accuracy
(± 1 digit)
Spec. resistance
[kΩ*cm]
0.000 ... 1.999
0.00 ... 19.99
0.0 ... 199.9
0 ... 1999
Spec. resistance
[MΩ*cm]
0.00 ... 19.99
0.0 ... 199.9
0 ... 1999
SAL0.0 ... 70.0 according to the IOT table
TDS [mg/l]0 ... 1999
Factor can be set between 0.40 ... 1.00
T [°C]
T [°F]
,
Specific resistance
− 5.0 ... + 105.0
23.0 ... + 212.0
No compensation:
Accuracy ± 0.5 %
Nonlinear compensation:
Accuracy Temperature of test sample
± 0.5 %0 °C ... 35 °C
acc. to ISO 7888, EN 27888
± 0,5 % 35 °C ... 50 °C
enhanced nLF function
(*)
50
Linear compensation
AccuracyTemperature of test sample
± 0.5 % 10 °C ... 75 °C
(The accuracy percentage always refers to the
measured value.)
(*)Corresponds to DEV C8
Page 51
BYKLC4Technical data
Accuracy
(± 1 digit)
Cell constant,
calibrated
Cell constant,
adjusted
Reference
temperature
SALRange 0.0 ... 42.0
AccuracyTemperature of test sample
± 0.15 °C ... 25 °C
± 0.225 °C ... 30 °C
TDS [mg/l]1
T [°C]
T [°F]
± 0.1
± 0.54
C [cm-1]0.450 ... 0.500
C [cm-1]0.090 ... 0.110
0.450 ... 0.500
TrefCan be set to 20 °C or 25 °C
Memory locations50
51
Page 52
Technical dataBYK LC 4
7.2Sensor module TetraCon V
Electrodes
Material
1
2
3
4
Number of electrodes4
Electrode materialGraphite
ShaftEpoxy, black
Connection headPVC
1Plug head connector
2Temperature sensor
3Current electrode (ring)
4Voltage electrode
Dimensions
Shaft diameter15.3 mm
Shaft length54 mm
Diameter of plug head
connector
Total length of sensor91.5 mm
Depth of
immersion
min.: 36 mm
max.: 75.5 mm (T < 50 °C)
max.: 54.0 mm (50 °C < T < 100 °C)
Connection
Vario plug head connector with integrated data storage
technique
Cell constant0.475 cm
-1
± 1.5 %
Application range1 µS/cm .. 200 mS/cm
Temperature
Thermistor typeIntegrated NTC (30 kΩ / 25 °C)
measurement
19 mm
52
Page 53
BYKLC4Technical data
Material of thermistor
enclosure
Working temperature-5 .. +100°C
Thermistor response
behavior
Sensor accuracy± 0.2 K
Storage mediumAir
Accessories
DesignationOrder number
E/SETCalibration set300 572
Graphite
< 20 sec.
t
99
53
Page 54
Technical dataBYK LC 4
54
Page 55
BYKLC4Lists
8Lists
This chapter provides additional information and orientation aids.
AbbreviationsThe list of abbreviations explains the indicators and the abbreviations that
appear on the display and in the manual.
Specialist termsThe glossary briefly explains the meaning of the specialist terms. However,
terms that should already be familiar to the target group are not described
here.
IndexThe index will help you to find the topics that you are looking for.
55
Page 56
ListsBYK LC 4
Abbreviations
Conductivity value
°CTemperature unit, Celsius
°FTemperature unit, Fahrenheit
ARAutoRead (drift control)
CCell constant cm
-1
CalCalibration
CELL Cell constant
E3Error message
CHAPTER 6 WHATTODOIF...
see
IdentID number
InitInitialization
Resets individual basic functions to the status they had
on delivery
IntInterval
LinLinear temperature compensation
LoBatBatteries almost empty (Low Battery)
nLFNonlinear temperature compensation
OFLDisplay range exceeded (Overflow)
SALSalinity
TCTemperature coefficient
TDSTotal dissolved solids
Tref 20/T20Reference temperature 20 °C
Tref 25/T25Reference temperature 25 °C
UFLValue is below the display range (Underflow)
56
Page 57
BYKLC4Lists
Glossary
AdjustingTo manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or a value
that is regarded as correct, or that the difference remains within the
tolerance.
AutoReadName for a function to check the stability of the measured value.
CalibratingComparing the value from a measuring system (e. g. the displayed value)
to the correct value or a value that is regarded as correct. Often, this
expression is also used when the measuring system is adjusted at the same
time (see adjusting).
CalibrationThe cell constant is determined through calibration. To do so, the
conductivity measuring cell is immersed into a series of aqueous salt
solutions with an exactly known electric conductivity. The relevant
conductivity values are determined using the conductivity meter.
Cell constantLinear factor describing the geometrical dimensions of a measuring cell and
its electrodes.
! With the value of the cell constant you can roughly describe the
application range of a sensor module.
! The cell constant is determined through calibration in one or several
calibration solutions.
ConductivityThe conductivity value is a sum parameter for the ion concentration of a test
sample.
Control standard
Solution with a known conductivity to determine or check the conductivity.
solution
Drift controlSee A
Measured
parameter
The measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D. O. concentration.
UTOREAD.
Measured valueThe measured value is the special value of a measured parameter to be
determined. It is given as a combination of the numerical value and unit
(e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
Measuring mediumThe substance to be measured. It can be liquid or solid.
Measuring solution /
test sample
Designation of the sample ready to be measured. Normally, a test sample
is made by processing the original sample. The test sample and original
sample are identical if the test sample was not processed.
Measuring systemThe measuring system comprises all the devices used for measuring, e. g.
meter and sensor. In addition, there is the cable and possibly an amplifier,
terminal strip and armature.
57
Page 58
ListsBYK LC 4
Reference
temperature
In order to compare values measured at different temperatures, the values
have to be converted to a fixed temperature. This temperature is 25 °C, or,
as an exception, 20 °C.
ResetRestoring the original condition of all settings of a measuring system.
ResistanceAll substances (solids, liquids, or gases) with mobile charge carriers, like for
example electrons or ions, have a finite ohmic resistance, which means they
have an electric conductance or an electric conductivity that can be
measured.
ResolutionSmallest difference between two measured values that can be displayed by
a meter.
SalinityThe salinity is a sum parameter especially for seawater.
It quotes the salt content.
Standard solutionThe standard solution is a solution where the measured value is known by
definition. It is used to calibrate a measuring system.
Temperature
compensation (TC)
The temperature has a very strong impact on the electrical conductivity. To
be able to compare measured values, it is necessary to temper or convert
the test sample to a reference temperature.
Total dissolved
solids (TDS)
Mass that remains of the substances dissolved in an aqueous solution after
a fixed filtering and drying procedure, as far as these substances are not
volatile under the conditions of this procedure. The total dissolved solids
refer to the volume of the filtered aqueous sample used and are given in
mg/l.
58
Page 59
BYKLC4Index
9Index
A
Alarm function .........................................37
Authorized use ...........................................7