Measure what you see.
Operation
operate
mode wheel:
• Switch on press
• Show menu press
• Select function scroll
• Activate function press
esc button:
• Change between menu
and measurement mode
std button:
• Short-cut to measure standard
Operate button:
• Switch on, select and measure
How to measure Differences?
How to Calibrate?
ENGLISH
Black/white calibration
• Select Setup – Calibrate – Black/White Calibration
• Place unit on black trap and confirm OK
• Place unit on white standard and confirm OK
Black/white calibration is requested every 3 months.
Cyan check
After a black and white calibration the instrument must be checked on the
Cyan standard.
• Place unit on Cyan standard and confirm OK
The Cyan check is requested every 30 days. Make sure to perform it in
laboratory environment (Temperature 20 - 25°C, Humidity 30 - 60%).
BYK-Gardner GmbH · Lausitzer Strasse 8 · 82538 Geretsried · Germany · Tel. +49-8171-3493-0 · Fax +49-8171-3493-140 · www.byk.com/instruments
• First page displays color
differences
• Select a stored standard, use
„AUTO standard“ or measure a standard
• Input standard name:
Change character ==> turn scroll wheel
Move to next digit ==> press scroll wheel
Move one digit back ==> press esc
• Input number of readings
• Place instrument on standard
• Press operate to take a reading
• Copy predefined settings for data display
• Continue with sample readings
Note: All color data inclusive spectral data are
saved and transferred to software.
• Scroll to second page for sparkle
and graininess results*
* Only available with BYK-mac i
269 020 584 E 0806
Please note detailed operating instructions in English on CD-ROM.
Additives & Instruments

Measure what you see.
How to measure Absolute Data?
• Select Measure - Absolute
• Place instrument on sample
• Press operate to take a reading
• Default settings: average of 3
• Input sample name:
Change character ==> turn scroll wheel
Move to next digit ==> press scroll wheel
Move one digit back ==> press esc
How to use an Organizer?
An organizer is a predefined sampling
procedure that needs to be downloaded
from the software smart-chart.
• Select style (Par1), color (Par2), paint
ENGLISH
line (Par3)
• Press operate to measure the first
measurement zone
• After a measurement zone has been
completed, press operate to move to
the next zone
• To skip a measurement zone, press the
mode wheel and select „Interrupt Statistic“.
The number of readings is set to zero.
Note: „Interrupt“ must be activated in the
organizer!
How to change Settings?
Instrument settings and measurement parameters for each individual
standard can be changed in Setup.
Overview Setup
Measurement Parameters
Statistic Define number of readings or select SMC mode
Color Select angles, color scale, illuminant/observer for display
Effect* Select effect scales
Fluorescence Limit Define acceptable max. difference between fluorescent
compensated and non-compensated reading as percentage of used color tolerance.
Sample Temperature Alarm Set minimum and maximum temperature to alert alarm
Instrument
Sample Curvature Select pin sensitivity according to sample curvature
Automatic Measurement Execute readings by pins on/off
Beeper Turn beeper on/off
Language Select language
Info Info on Version, Calibration Dates, Free Memory …
Date/Time Set date and time
Display Time Define time instrument is on
Display Brightness Adjust display brightness.
Advanced Activate/deactivate Interrupt Statistic, Input Parameter 2,
Input Comment, Input ID, Input Checkzone
* Only available with BYK-mac i