ALTANA BYK-mac i COLOR Short Instructions

A member of
Measure what you see.
operate
mode wheel:
Switch on press
Show menu press
Select function scroll
Activate function press
esc button:
Change between menu
and measurement mode
std button:
Short-cut to measure standard
Operate button:
Switch on, select and measure
How to measure Differences?
How to Calibrate?
ENGLISH
Black/white calibration
Select Setup – Calibrate – Black/White Calibration
Place unit on black trap and confirm OK
Place unit on white standard and confirm OK
Black/white calibration is requested every 3 months.
Cyan check
After a black and white calibration the instrument must be checked on the Cyan standard.
Place unit on Cyan standard and confirm OK
The Cyan check is requested every 30 days. Make sure to perform it in laboratory environment (Temperature 20 - 25°C, Humidity 30 - 60%).
BYK-Gardner GmbH · Lausitzer Strasse 8 · 82538 Geretsried · Germany · Tel. +49-8171-3493-0 · Fax +49-8171-3493-140 · www.byk.com/instruments
First page displays color
differences
Select a stored standard, use
„AUTO standard“ or measure a standard
Input standard name:
Change character ==> turn scroll wheel Move to next digit ==> press scroll wheel Move one digit back ==> press esc
Input number of readings
Place instrument on standard
Press operate to take a reading
Copy predefined settings for data display
Continue with sample readings
Note: All color data inclusive spectral data are saved and transferred to software.
Scroll to second page for sparkle
and graininess results*
* Only available with BYK-mac i
269 020 584 E 0806
Please note detailed operating instructions in English on CD-ROM.
Additives & Instruments
Measure what you see.
A member of
How to measure Absolute Data?
Select Measure - Absolute
Place instrument on sample
Press operate to take a reading
Default settings: average of 3
Input sample name:
Change character ==> turn scroll wheel Move to next digit ==> press scroll wheel Move one digit back ==> press esc
How to use an Organizer?
An organizer is a predefined sampling procedure that needs to be downloaded from the software smart-chart.
Select style (Par1), color (Par2), paint
ENGLISH
line (Par3)
Press operate to measure the first
measurement zone
After a measurement zone has been
completed, press operate to move to the next zone
To skip a measurement zone, press the
mode wheel and select „Interrupt Statistic“. The number of readings is set to zero.
Note: „Interrupt“ must be activated in the
organizer!
How to change Settings?
Instrument settings and measurement parameters for each individual standard can be changed in Setup.
Overview Setup
Measurement Parameters
Statistic Define number of readings or select SMC mode
Color Select angles, color scale, illuminant/observer for display
Effect* Select effect scales
Fluorescence Limit Define acceptable max. difference between fluorescent
compensated and non-compensated reading as percen­tage of used color tolerance.
Sample Temperature Alarm Set minimum and maximum temperature to alert alarm
Instrument
Sample Curvature Select pin sensitivity according to sample curvature
Automatic Measurement Execute readings by pins on/off
Beeper Turn beeper on/off
Language Select language
Info Info on Version, Calibration Dates, Free Memory …
Date/Time Set date and time
Display Time Define time instrument is on
Display Brightness Adjust display brightness.
Advanced Activate/deactivate Interrupt Statistic, Input Parameter 2,
Input Comment, Input ID, Input Checkzone
* Only available with BYK-mac i
300 000 893 E 1306