Agilent Technologies Performing the EP stray User Manual

Technical Note
Performing the EP stray-light test with potassium chloride on UV-visible spectrophotometers
Introduction
Measured stray light has two com­ponents:
light coming from the light
n
• ambient light that reaches the detector either directly or by simple reflections, see figure 1
For instruments with reversed optics such as the Agilent 8453
UV-visible spectrophotometer
1
ambient light can not reach the detector, see figure 2. The equa­tion used to calculate transmit­tance and thereby absorbance is:
T = (I + Is)/(I0+ Is)
Where T is transmittance, I0is
intensity of incident light, I is intensity of transmitted light and
Isis intensity of stray light. Stray
light has an increasing influence on spectroscopic measurements at low levels of intensity, that is, high absorbances.
The result is that stray light caus­es a negative bias in instrument response and eventually becomes the limiting factor for absorbance and therefore concentration that can be measured. The effect of various levels of stray light on measured absorbance compared with actual absorbance is shown in Figure 3.
This Technical Note examines the influence of sample composition, sample temperature, bandwidth and wavelength accuracy on the result of the stray-light test using potassium chloride as described in the European Pharmacopoeia
(EP).
2
Figure 1 Terminology used when describing stray light
Ambient light
λ
n-1
λ
n
λ
n+1
Sample
Source
Ambient light
Detector
Agilent Technologies
Innovating the HP Way
Measurement of stray light
To measure stray light, a filter is required that absorbs all light of the wavelength at which the mea­surement is to be made and trans­mits higher and lower wave­lengths. Figure 4 shows this ideal stray light filter. At the measured wavelength (200 nm) the transmis­sion is 0% whereas at all other wavelengths it is 100%. Such filters do not exist in practice, so cut-off filters are used which transmit all light above a certain wavelength and block all light at lower wave­lengths.
Salt solutions, for example, potas­sium chloride (12 g/l), sodium iodide (10 g/l) and sodium nitrite (50 g/l) in water, can be used as standard stray-light filters at 200, 220 and 340 nm respectively (see figure 5).
The user should keep in mind that all 3 filters are only approaches to the ideal test filter. The slope of the absorption edge shows no infi­nite value like the ideal filter does. In addition the used filters block all light from wavelengths shorter than the measured one. The con­tribution of stray light that might result from these wavelengths is thus eliminated. This leads to smaller stray light values than would be expected from the ideal filter. The user should be aware of this systematic deviation as a con­sequence of using non-ideal stray light filters.
Source
Ambient light
Detector
Grating
Entrance slit
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 True absorbance [AU]
Measured absorbance [AU]
0% Stray light
0.01% Stray light
0.1% Stray light
1% Stray light
Figure 3 The effect of stray light on measured sample absorbance
Figure 2 Simplified schematic of a diode-array spectrophotometer
The EP stray-light test
Stray-light measurement with the Agilent 8453 UV-visible spec­trophotometer comprises the three test described above. The potassium chloride test is speci-
fied in the EP2:
“…Stray light may be detected at a given wavelength with suitable filters or solutions: for example the absorbance of a 12 g/l solu­tion of potassium chloride in a 1 cm cell should be greater than two at 200 nm when compared with water as the compensation liquid.”
An absorbance of greater than two means a transmittance of less than one percent. For a number of rea­sons the potassium chloride test is the most critical of the three stray­light tests. As shown in figure 5, this is the only test in which the measured wavelength is situated very close to the cut-off slope. As a consequence the test is extreme­ly sensitive to the wavelength accuracy of the spectrophotome­ter. Even small deviations will result in test failure. For this rea­son a wavelength recalibration should be performed before each stray-light test.
Further, the measurement time is also relevant. When measuring the transmittance with the test solu­tion in the light path the intensity at the detector is very low. The signal-to-noise ratio can be improved by increasing the inte­gration time of the spectropho­tometer. All three stray-light tests with the Agilent 8453 spectopho­tometer were performed using integration times of 5 s.
0
20
40
60
80
100
100 200 300 400 500 600 700 800 900 1000
Wavelength [nm]
Transmittance [%]
0
20
40
60
80
100
200 250 300 350 400 450 500 550 600
Wavelength [nm]
Transmittance [%T]
NaNO
2
NaIKCl
Figure 5 The spectra of potassium chloride (12 g/l), sodium iodide (10 g/l) and sodium nitrite (50 g/l) in water
Figure 4 The ideal spectrum of a stray-light filter at 200 nm
Loading...
+ 5 hidden pages