Agilent Technologies Performing the EP stray User Manual

Technical Note
Performing the EP stray-light test with potassium chloride on UV-visible spectrophotometers
Introduction
Measured stray light has two com­ponents:
light coming from the light
n
• ambient light that reaches the detector either directly or by simple reflections, see figure 1
For instruments with reversed optics such as the Agilent 8453
UV-visible spectrophotometer
1
ambient light can not reach the detector, see figure 2. The equa­tion used to calculate transmit­tance and thereby absorbance is:
T = (I + Is)/(I0+ Is)
Where T is transmittance, I0is
intensity of incident light, I is intensity of transmitted light and
Isis intensity of stray light. Stray
light has an increasing influence on spectroscopic measurements at low levels of intensity, that is, high absorbances.
The result is that stray light caus­es a negative bias in instrument response and eventually becomes the limiting factor for absorbance and therefore concentration that can be measured. The effect of various levels of stray light on measured absorbance compared with actual absorbance is shown in Figure 3.
This Technical Note examines the influence of sample composition, sample temperature, bandwidth and wavelength accuracy on the result of the stray-light test using potassium chloride as described in the European Pharmacopoeia
(EP).
2
Figure 1 Terminology used when describing stray light
Ambient light
λ
n-1
λ
n
λ
n+1
Sample
Source
Ambient light
Detector
Agilent Technologies
Innovating the HP Way
Measurement of stray light
To measure stray light, a filter is required that absorbs all light of the wavelength at which the mea­surement is to be made and trans­mits higher and lower wave­lengths. Figure 4 shows this ideal stray light filter. At the measured wavelength (200 nm) the transmis­sion is 0% whereas at all other wavelengths it is 100%. Such filters do not exist in practice, so cut-off filters are used which transmit all light above a certain wavelength and block all light at lower wave­lengths.
Salt solutions, for example, potas­sium chloride (12 g/l), sodium iodide (10 g/l) and sodium nitrite (50 g/l) in water, can be used as standard stray-light filters at 200, 220 and 340 nm respectively (see figure 5).
The user should keep in mind that all 3 filters are only approaches to the ideal test filter. The slope of the absorption edge shows no infi­nite value like the ideal filter does. In addition the used filters block all light from wavelengths shorter than the measured one. The con­tribution of stray light that might result from these wavelengths is thus eliminated. This leads to smaller stray light values than would be expected from the ideal filter. The user should be aware of this systematic deviation as a con­sequence of using non-ideal stray light filters.
Source
Ambient light
Detector
Grating
Entrance slit
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 True absorbance [AU]
Measured absorbance [AU]
0% Stray light
0.01% Stray light
0.1% Stray light
1% Stray light
Figure 3 The effect of stray light on measured sample absorbance
Figure 2 Simplified schematic of a diode-array spectrophotometer
The EP stray-light test
Stray-light measurement with the Agilent 8453 UV-visible spec­trophotometer comprises the three test described above. The potassium chloride test is speci-
fied in the EP2:
“…Stray light may be detected at a given wavelength with suitable filters or solutions: for example the absorbance of a 12 g/l solu­tion of potassium chloride in a 1 cm cell should be greater than two at 200 nm when compared with water as the compensation liquid.”
An absorbance of greater than two means a transmittance of less than one percent. For a number of rea­sons the potassium chloride test is the most critical of the three stray­light tests. As shown in figure 5, this is the only test in which the measured wavelength is situated very close to the cut-off slope. As a consequence the test is extreme­ly sensitive to the wavelength accuracy of the spectrophotome­ter. Even small deviations will result in test failure. For this rea­son a wavelength recalibration should be performed before each stray-light test.
Further, the measurement time is also relevant. When measuring the transmittance with the test solu­tion in the light path the intensity at the detector is very low. The signal-to-noise ratio can be improved by increasing the inte­gration time of the spectropho­tometer. All three stray-light tests with the Agilent 8453 spectopho­tometer were performed using integration times of 5 s.
0
20
40
60
80
100
100 200 300 400 500 600 700 800 900 1000
Wavelength [nm]
Transmittance [%]
0
20
40
60
80
100
200 250 300 350 400 450 500 550 600
Wavelength [nm]
Transmittance [%T]
NaNO
2
NaIKCl
Figure 5 The spectra of potassium chloride (12 g/l), sodium iodide (10 g/l) and sodium nitrite (50 g/l) in water
Figure 4 The ideal spectrum of a stray-light filter at 200 nm
Another important issue is the quality of the selected potassium chloride. The concentration of bromide plays an important role on the position of the absorption edge. The EP allows a maximum of 0.15 % potassium bromide (0.1
% bromide)
2
. Figure 6 shows spec­tra of potassium chloride solu­tions with different bromide per­centages measured on the Agilent 8453 spectrophotometer. From left to right the bromide concentration is increasing. Spectra 1 to 3 shows the transmittance of potassium chloride samples with bromide contents of <0.005%, 0.05% and
0.10%, respectively.
Increased bromide concentrations lead to lower percentage transmit­tance values at 200 nm (see table1), because the wavelength at 50% transmittance is gradually shifted from 204.9 to 210.7 nm. Spectrum 4 shows a non-EP con­form sample with bromide content above the EP limit of 0.1% and thus resulting in lower transmit­tance value at 200 nm. Table 1 summarizes the stray light data of the different samples.
It should be emphasized that prob­lems can arise when using highly pure potassium chloride samples. At a very low bromide concentra­tions the transmittance value at 200 nm may exceed the EP limit of 1%. It would make no sense to use ultra-pure potassium chloride samples for stray light analysis, although the quality of the chemi­cals are according to the specifica­tions of the EP.
0
10
20
30
40
50
60
70
80
90
100
190 200 210 220 230 240 250
Wavelength [nm]
Transmittance [%]
0
1
2
195 200 205
[%]
Wavelength [nm]
195 205
Bromide [%] T <200nm>[%] Wavelength <T=50%> [nm]
<0.005 >0.64139 204.9
0.05 0.25893 208.3
0.10 0.18658 210.7
>0.15 <0.15534 217.6
Table 1 Measured stray light on the Agilent 8453 spectrophotometer as a function of bromide concentration (blank on water)
Figure 6 Potassium chloride samples measured on the Agilent 8453 spectrophotometer: Influence of increasing bromide concentration on the shape of the absorption edge 1: < 0.005% 2: 0.05% 3: 0.10% 4: >0.15%
1432
Another factor that has a lasting effect on the measured transmit­tance at 200 nm is the spectral bandwidth (SBW) of the spec­trophotometer. Decreasing the SBW leads to decreasing readings when the natural bandwidth (NBW) of the absorbing substance and SBW have the same order of magnitude. However, the wave­length at 50% transmittance is vir­tually independent of the selected bandwidth. These dependences are shown in table 2 for the KCl solution with 0.1% bromide. The data were aquired with a high per­formance conventional scanning spectrophotometer that possesses a better stray-light specification than the Agilent 8453 spectropho­tometer.
The SBW of the Agilent 8453 spectrophotometer is comparable to a conventional scanning instru­ment with a spectral bandwidth of about 1.5 nm. The bandwidth is fixed and cannot be changed by the operator.
The temperature also strongly affects results of stray-light mea­surements, see Figure 7. With increasing temperature the absorption band is broadend, lead­ing to smaller transmittance val­ues at 200 nm. The EP recom­mends a temperature of 20°C ±1°C for the potassium chloride stray­light test.
Figure 7 The effect of temperature variation on the transmission of potassium chloride solution at 200 nm
Slit width <nm> %T <200nm> Wavelength <T=50%> [nm]
0.5 0.0111 210.8
1.0 0.0169 210.7
1.5 0.0286 210.7
Table 2 Stray light data of a conventional scanning spectrophotometer with variable bandwidth. Sample: potassium chloride (bromide: 0.10%), Blank: water
0.30
0.40
0.50
0.60
0.70
0.80
15.0 20.0 25.0 30.0 Temperature [˚C]
Transmittance 200 nm [%]
References
1. HP 8453 Spectrophotometer and Open Sample Area, Agilent Technologies Technical Note, 1999, publication number
2. European Pharmacopoeia, third edition, 1997, pages 29 and 1361
3. ASTM E 387-84, Standart Test Method for Estimating Stray Radiant Power Ratio of Spectrophotometers by the Opaque Filter Method
ASTM stray light tests
3
The sodium nitrite and iodide tests both have cut-off wave­lenghts that are far away (50 and 60 nm) from the measured wave­lenths (see figure 5). For this rea­son the two tests are less sensitive to wavelength accuracy of the spectrophotometer. On the other hand both stray light tests vary more from the ideal test filter than the EP test.
Summary
Many factors influence the collect­ed stray-light data. When using liq­uid filters such as potassium chlo­ride users should be aware that they are applying non-ideal stray­light filters. For the EP stray-light test, the obtained results depend not only on several external fac­tors like sample temperature and wavelength accuracy of the spec­trophotometer, but also to a high degree on the bromide content of the potasssium chloride used.
The EP allows a maximum of 0.1% bromide. Independent from the spectral bandwidth of the spec­trophotometer, a bromide concen­tration of maximum 0.1% corre­sponds to a wavelength at 50% transmittance of less than 211 nm. If the potassium chloride solution used shows a wavelength at 50% transmittance that is more than 211 nm, it does not conform to the requirements of the EP.
The Agilent OQ/PV standards kit (order number 5063-6503) offers an easy-to-use, cost-effective and fully-EP-compliant solution for stray-light determination of UV­visible spectrophotometers. All three solutions (KCl, NaI and NaNO2) are part of this kit and are
provided in sealed ampules for use with standard 10 mm quartz cuvettes.
www.agilent.com/chem
The information in this publication is subject to change withou notice.
Copyright © 2000 Agilent Technologies, Inc. All Rights Reserved. Reproduction, adaptation or translation without prior written permission is prohibited, except as allowed under the copyright laws.
Printed in Germany 11/2000 Publication Number 5988-0945EN
Agilent Technologies
Innovating the HP Way
Loading...