The limiting factor for high
throughput in such systems is the
speed of the HPLC analysis.
Standard HPLC cycle times from
injection to injection for gradient
analysis lie between 15 and 20 min
using columns of 100 to 200 mm in
length.
To significantly increase the sample throughput, cycle times must
be shortened. This can be
achieved using fast gradients with
short columns and high flow rates.
In the following example we
demonstrate how to optimize the
Agilent 1100 Series high-pressure
gradient system to obtain rapid
gradients and high sample
throughput. Hints are given on the
influence of chromatographic
parameters on cycle times, and on
how run times of less than two
minutes can be expected to affect
performance.
Equipment
All HPLC experiments were carried out on the Agilent 1100 Series
high-pressure gradient system
comprising:
• Agilent 1100 Series high-pressure
pump for lowest delay volume.
In this design each solvent is
pumped by its own pump
assembly, and mixing takes
place on the high-pressure side.
This means gradient changes
reach the column much faster
than in low-pressure gradient
systems where mixing takes
place on the low-pressure side.
• Agilent 1100 Series vacuum
degasser for optimum baseline
stability.
• Agilent 1100 Series autosampler
for sampling from 2-ml standard
vials.
• Optional Agilent 220 micro
plate sampler for flexible sampling from deepwell and/or
microtiter plates.
• Agilent 1100 Series thermostatted column compartment for
highest stability from 10 °C
below ambient up to 80 °C.
• Agilent 1100 Series diode array
detector with standard flow cell
(10-mm pathlength, 13-microliter volume).
• Optional Agilent 1100 Series
variable wavelength detector.
• Optional Agilent 1100 Series
LC/MSD module for mass and
structural information.
• Agilent ChemStation with 3D
HPLC single instrument software
for instrument control, data handling and sample tracking.
Compounds and chromatographic conditions
For our experiments we selected
the following compounds which
differ considerably in polarity:
• caffeine
• primidone
• phenacetin
• mandelic acid benzylester
• biphenyl
The chromatographic conditions
are listed next to the figures.
Optimization of chromatographic parameters
The following parameters have to
be adapted to obtain short cycle
times, sufficient resolution and
best performance over a wide
range of polarity:
• column length
• gradient
• flow rate
• delay volume
• data rate of detector
• column temperature
The aim was to achieve cycle
times of about 2 min and baseline
separation for all compounds.