Optimizing Performance
of an Agilent 1100
Series HPLC System
Is sensitivity an issue?
þ Select a cell with long optical
path length, for example,
10 mm.
þ When using a DAD, set the slit
width to high values, for
example, 8 nm.
þ When using a DAD, select a
broad bandwidth for the detection wavelength, for example,
30 nm.
þ Select the reference wave-
length such that the reference
bandwidth starts close to the
end of the sample spectra. For
example, if the sample spectra
has its maximum at 252 nm and
is down to zero mAUs at
300 nm, the optimum reference
wavelength here would be
360 nm with a bandwidth of
100 nm.
þ When using a DAD, select high
response times, for example,
0.1 min. For broad peaks,
select even higher response
times.
Are retention time and area
precision an issue?
þ Select correct compressibility
and stroke value for the pump.
þ Flush instrument thoroughly
with selected solvents.
þ Let column equilibrate for at
least 2 hours.
þ For gradient analysis run the
gradient run several times
before starting with the test.
þ Set injection volume to 5 µl.
Is highest performance for gradients in the range from 0 to
5% B and/or from 95 to 100%B
an issue?
þ Select a binary pump for high-
est performance.
For best performance over the
complete gradient range, especially from 0 to 5 % and 95 to 100 %,
we recommend using the binary
pump.
2.