Agilent Technologies Optimizing 1100 Highest Performance Technical Note

Introduction
This document will help you to efficiently prepare an Agilent 1100 Series HPLC system for best performance. It describes how to achieve
Optimizing the Agilent 1100 Series System for Highest Performance
• highest sensitivity,
• highest precision,
• low flow rates,
• lowest delay volume,
• lowest carry-over, and
• best results in normal phase chromatography.
We first describe a short standard check-out procedure, which is re­commended before starting to optimize the instrument. This ensures that the instrument has no defects which may cause ghost peaks, increased noise or unsta­ble retention times.
In the second section, we then explain how to optimize the per­formance of an Agilent 1100 Series HPLC system.
1
Agilent Technologies
Innovating the HP Way
Standard Check-out Procedure
What is needed for the stan­dard check-out procedure
• Column: 125 × 4 mm Hypersil ODS, 5 µm (Agilent part number 7982618-
564) Use a new column or a col­umn of known history.
• Solvents: bi-distilled water, acetonitrile
• Isocratic standard sample (Agilent part number 01080-68704)
Checking the equipment for leaks and contamination
1. Fill one solvent bottle with bi­distilled water (channel A) and the other one with acetonitrile (channel B).
2. Open the purge valve and purge both channels with 5 ml/min for at least 10 min. Check outlet tubing of the purge valve. If no air bubbles are observed purg­ing can be stopped.
3. Adjust compressibility and stroke under INSTRUMENT, MORE PUMP. Use the Help key to find appropriate compress­ibility values for the different solvents. With the quaternary pump you only need one com­pressibility value, which should be set to 100. For the binary pump set A to 46 and B to 115. The stroke should be set to auto for both A and B.
4. After the purge process, and after installing the column, set the flow rate to 1 ml/min and close the purge valve.
5. Watch for leaks. Leave the ther­mostatted column compartment open and check each connec­tion for leaks.
6. Watch pressure in the ONLINE SIGNAL screen. The ripple should be very regular, no spik­ing or pressure drops should occur. If they do occur and you can detect no leaks, we recom­mended contacting the Agilent service organization for possi­ble repair.
7. To check for contamination in your system, select the follow­ing settings on your instrument. This is especially important if sensitivity in gradient runs is an issue.
No peaks should be observed if these parameters are set. If, how­ever, the first run does show peaks, run the gradient a second and a third time. Should peaks continue to occur, clean the instrument using different sol­vents or/and passivate it with 65 % nitric acid.
1.
Parameter Setting
Flow rate 1 ml/min
Detection wavelength 210 nm
Stop time 30
Column compartment temperature 36 ºC
Gradient start with 0 % B and
go to 100 % B in 25 min.
Blank run, no injection no vial number in
SAMPLE INFO
screen
Checking for baseline noise
Before running the system the detector lamp should have been switched on for at least half an hour. Select the following settings on your instrument:
Using the Agilent ChemStation choose the PERFORMANCE AND NOISE report and set six 1-minute ranges in the SYSTEM SUITABILITY screen which is located under the REPORT menu. The instrument then calculates the noise automatically. The noise measured over one minute at six different parts in the chro­matogram should be close to the following specifications:
• variable wavelength detector (VWD): ± 0.75 x 10
-5
AU at 254 nm
• diode array detector (DAD): ± 1 x 10
-5
AU at 254 and 750 nm
These specifications are based on a cell with a pathlength of 10 mm, a response time of 2 s, a flow of 1 ml/min methanol and a slit width of 4 nm for the diode array detector.
Figure 1 shows how to evalutate the signal noise based on the ASTM method. The Agilent ChemStation software evaluates ASTM noise automatically.
If the noise drastically differs from the specifications, we recommend contacting the Agilent service organization for possible repair.
Figure 1 Signal noise based on the ASTM method
Parameter Setting
Flow rate 1 ml/min
Detection wavelength 254 nm
Stop time 30
Column compartment temperature 36 ºC
Isocratic: 100 % water
Blank run, no injection no vial number in
SAMPLE INFO
screen
Interval 1 min
123456
Time [min]
Drift
Peak to peak
Loading...
+ 5 hidden pages