Standard Check-out
Procedure
What is needed for the standard check-out procedure
• Column:
125 × 4 mm Hypersil ODS, 5 µm
(Agilent part number 7982618-
564) Use a new column or a column of known history.
• Solvents:
bi-distilled water, acetonitrile
• Isocratic standard sample
(Agilent part number
01080-68704)
Checking the equipment for
leaks and contamination
1. Fill one solvent bottle with bidistilled water (channel A) and
the other one with acetonitrile
(channel B).
2. Open the purge valve and purge
both channels with 5 ml/min for
at least 10 min. Check outlet
tubing of the purge valve. If no
air bubbles are observed purging can be stopped.
3. Adjust compressibility and
stroke under INSTRUMENT,
MORE PUMP. Use the Help key
to find appropriate compressibility values for the different
solvents. With the quaternary
pump you only need one compressibility value, which should
be set to 100. For the binary
pump set A to 46 and B to 115.
The stroke should be set to
auto for both A and B.
4. After the purge process, and
after installing the column, set
the flow rate to 1 ml/min and
close the purge valve.
5. Watch for leaks. Leave the thermostatted column compartment
open and check each connection for leaks.
6. Watch pressure in the ONLINE
SIGNAL screen. The ripple
should be very regular, no spiking or pressure drops should
occur. If they do occur and you
can detect no leaks, we recommended contacting the Agilent
service organization for possible repair.
7. To check for contamination in
your system, select the following settings on your instrument.
This is especially important if
sensitivity in gradient runs is an
issue.
No peaks should be observed if
these parameters are set. If, however, the first run does show
peaks, run the gradient a second
and a third time. Should peaks
continue to occur, clean the
instrument using different solvents or/and passivate it with 65 %
nitric acid.
1.
Parameter Setting
Flow rate 1 ml/min
Detection wavelength 210 nm
Stop time 30
Column compartment
temperature 36 ºC
Gradient start with 0 % B and
go to 100 % B in
25 min.
Blank run, no injection no vial number in
SAMPLE INFO
screen