Results
The Agilent 220 micro plate sampler is a precise instrument for
high throughput analysis of large
numbers of samples. To optimize
the performance for sample injection the parameters and settings
listed in table 5 have to be taken
into consideration.
Parameters/Settings
Drawn sample volume If enough sample volume is available the sample loop should be
(complete loop fill) overfilled with the recommended sample volume. The re-
commended sample volumes for different sample loop sizes are
shown in table 1.
Draw speed The draw speed for
Create Air Gap, Draw Sample
and
Inject
(complete loop fill) should be set to 0.5 ml/min or less. Higher draw speeds lead to
decreasing precision. The time saved is minimal.
Size of air gap Create an air gap before sample is drawn.
(complete loop fill) The size of the air gap should be between 1–4 µl, 2–3 µl are
recommended.
The size of the air gap must not exceed 4 µl.
Dilutor syringe size and When using
Centered Loop Fill
a sample loop filled to at least
sample loop size 50 % should be used.
(centered loop fill) The volume of the dilutor syringe should be the same
or higher than the sample loop, but still as small as possible.
Partial Loop Fill The partial loop filling technique is used when high reproducibili
ty is required and large sample volume is available.
Carry over To avoid carry over use the commands
Rinse Needle Inside
,
Rinse Needle Outside
and
Rinse Injection Port
.
Use the
Wash Vial
command, especially for samples with high
viscosity.
Conclusion
This technical note described the
optimization of the Agilent 220
MPS hardware and software for
highest precision. The optimal settings for drawn sample volume,
draw speed, size of the air gap
and dilutor syringe size and sample loop size were evaluated.
Furthermore, it was demonstrated
how carry over could be minimized. This information should
help optimize the Agilent 220 MPS
for high performance and precision.
Table 5
Summary of results