Agilent Technologies Optimizing Technical Note

Optimizing the Agilent high-throughput analysis system for high performance and precision
Abstract
The Agilent 220 micro plate sampler—integrated into the Agilent 1100
Series LC system—is the ideal tool to analyze large numbers of struc-
turally distinct compounds. It offers the full potential of well plate
technology:
response times. This is especially useful for drug discovery, combi-
natorial chemistry, medicinal chemistry and natural product analysis.
• The flexibility of having a sampler mode for high-speed sample
analysis and a fraction collector mode for sample isolation and
purification.
• An automated injector program for sample preparation such as
derivatization, dilution and mixing.
• The ability to store and inject over 4,600 samples (using 384-well
plates) for unattended high-throughput.
Technical Note
Equipment
All experiments were carried out on the Agilent 1100 Series HPLC system consisting of
• Agilent 1100 Series vacuum degasser,
• Agilent 1100 Series binary pump,
• Agilent 1100 Series ther­mostatted column compart ment,
• Agilent 1100 Series diode array detector, and
• Agilent 220 micro plate sampler.
The system was controlled using the Agilent ChemStation (version A.07.01) and the micro plate sam­pling software (version A.03.01).
Injection principle
In contrast to the Agilent 1100 Series autosamplers the Agilent 220 MPS works with a fixed-size sample loop injection (figure 1). While the valve is in loading posi­tion the amount of drawn sample is injected into the fixed size sam­ple loop capillary and the surplus of sample is flushed through the loop into waste. To achieve
optimal performance overfilling of the sample loop is recommended. The overfill volume depending on loop size is shown in table 1.
The reason why overfilling is nec­essary is the hydrodynamic behav­ior of fluids as they pass through tubing. A process called laminar flow takes place under conditions in which molecules close to the tubing walls are slowed by fricton­al forces. The result is a bullet­shaped profile in which the mole­cules in the center of the stream travel roughly twice the velocity of those at the tubing wall3. The surplus of sample volume also ensures that the dead volume of the injector port capillary of approximately 4.2 µl is completely flushed with sample. After switch­ing the valve into the injection and run position the amount of sample in the loop is applied to the system.
To apply sample amounts smaller than the loop volume the Centered
Loop Fill or the Partial Loop Fill & Inject commands can be used.
For the Centered Loop Fill tech­nique a sandwich of 1-µl air gaps and sample is injected into the sample loop. The volume drawn before the air gaps and the sample volume is calculated by the soft­ware to position the sample vol­ume in the middle of the sample loop. The precision for this tech­nique is lower than for complete loop fill because only the
Introduction
The Agilent 220 micro plate sam­pler (MPS)1is an essential part of Agilent’s system for combinatorial chemistry and high throughput HPLC analysis. It combines high sample capacity, high speed and sampling/fractionation capabili­ties in one system. In combination with the Agilent 1100 Series HPLC system using mass selective detection it is a complete system that fulfills the requirements of combinatorial chemistry and high­throughput analysis, offering robustness, ruggedness, sensitivi­ty, selectivity and speed. The inte­grated system plus the single soft­ware platform simplifies system setup, operation and management of large amounts of data. Details of the Agilent 220 MPS and how it works in an Agilent system for combinatorial chem­istry or high throughput screening is described in another Agilent technical note2. In this note we explain how to further optimize the hardware and software setup of the Agilent 220 MPS to achieve higher performance and preci­sion.
Loop Volume [µl] Sample Volume [µl]
535 10 40 20 55 50 95
Figure 1 Injection principle of the Agilent 220 micro plate sampler
6
5
4
3
2
1
6
5
4
3
2
1
Loading position Injection and run position
to column
to column
from pump
from pump
waste
waste
injection port
4.2 µl dead volume!
sample loop (fixed size)
sample loop (fixed size)
Table 1 Sample loop sizes and recommended sample volumes
0
1
2
3
4
5
0 20406080100
Sample Volu me [µl]
RSD Area [%]
5 µl Sample Loop
20 µl Sample Loop
mechanical precision of the dilu­tor syringe determines the inject­ed sample volume. For Partial Loop Fill & Inject the sample vol­ume plus a relatively large rinse volume is drawn from the sample. The rinse volume is used to rinse the injection port and the sample loop before the actual sample vol­ume is pushed into the sample loop. The precision is higher than for Centered Loop Fill but also more sample volume is required.
Parameters to optimize
Parameters influencing the perfor­mance of the sampler in sampling mode are:
1. Drawn sample volume (complete loop fill)
2. Draw speed (complete loop fill)
3. Size of air gap (complete loop fill)
4. Dilutor syringe size and sample loop size (centered loop fill)
1. Drawn sample volume
(complete loop fill)
Although the size of the sample loop determines the injected sam­ple volume the amount of drawn sample volume also influences the precision of the measurement3. Figure 2 shows the precision of peak areas for ten measurements for different drawn sample vol­umes
For the drawn sample amount of 5 µl the precision of peak area is very high because 5 µl is just enough sample to fill the injection port capillary (dead volume
4.2 µl). Increasing the sample vol-
ume also increases the precision of peak area. When the recom­mended sample volume is used the precision is usually optimized. More sample volume does not increase the precision any further.
If not enough sample volume for sufficient overfill of the sample loop is available, a lower precision of peak area results.
2. Draw speed (complete loop fill)
The draw speed for the sample as well as the inject speed influences the precision of the measurement as shown in figure 3.
Figure 3 Precision of peak area for different draw and inject speeds (5-µl sample loop, complete loop fill)
0.0
0.2
0.4
0.6
0.8
1.0
1.2
0.5 ml/min 1 ml/min 2 ml/min 3 ml/min
Draw Speed
RSD Area [%]
Figure 2 Precision of peak area for different sample volumes (5 and 20-µl sample loop, complete loop fill)
Loading...
+ 5 hidden pages