Agilent Technologies Optimizing Technical Note

Optimizing the Agilent high-throughput analysis system for high performance and precision
Abstract
The Agilent 220 micro plate sampler—integrated into the Agilent 1100
Series LC system—is the ideal tool to analyze large numbers of struc-
turally distinct compounds. It offers the full potential of well plate
technology:
response times. This is especially useful for drug discovery, combi-
natorial chemistry, medicinal chemistry and natural product analysis.
• The flexibility of having a sampler mode for high-speed sample
analysis and a fraction collector mode for sample isolation and
purification.
• An automated injector program for sample preparation such as
derivatization, dilution and mixing.
• The ability to store and inject over 4,600 samples (using 384-well
plates) for unattended high-throughput.
Technical Note
Equipment
All experiments were carried out on the Agilent 1100 Series HPLC system consisting of
• Agilent 1100 Series vacuum degasser,
• Agilent 1100 Series binary pump,
• Agilent 1100 Series ther­mostatted column compart ment,
• Agilent 1100 Series diode array detector, and
• Agilent 220 micro plate sampler.
The system was controlled using the Agilent ChemStation (version A.07.01) and the micro plate sam­pling software (version A.03.01).
Injection principle
In contrast to the Agilent 1100 Series autosamplers the Agilent 220 MPS works with a fixed-size sample loop injection (figure 1). While the valve is in loading posi­tion the amount of drawn sample is injected into the fixed size sam­ple loop capillary and the surplus of sample is flushed through the loop into waste. To achieve
optimal performance overfilling of the sample loop is recommended. The overfill volume depending on loop size is shown in table 1.
The reason why overfilling is nec­essary is the hydrodynamic behav­ior of fluids as they pass through tubing. A process called laminar flow takes place under conditions in which molecules close to the tubing walls are slowed by fricton­al forces. The result is a bullet­shaped profile in which the mole­cules in the center of the stream travel roughly twice the velocity of those at the tubing wall3. The surplus of sample volume also ensures that the dead volume of the injector port capillary of approximately 4.2 µl is completely flushed with sample. After switch­ing the valve into the injection and run position the amount of sample in the loop is applied to the system.
To apply sample amounts smaller than the loop volume the Centered
Loop Fill or the Partial Loop Fill & Inject commands can be used.
For the Centered Loop Fill tech­nique a sandwich of 1-µl air gaps and sample is injected into the sample loop. The volume drawn before the air gaps and the sample volume is calculated by the soft­ware to position the sample vol­ume in the middle of the sample loop. The precision for this tech­nique is lower than for complete loop fill because only the
Introduction
The Agilent 220 micro plate sam­pler (MPS)1is an essential part of Agilent’s system for combinatorial chemistry and high throughput HPLC analysis. It combines high sample capacity, high speed and sampling/fractionation capabili­ties in one system. In combination with the Agilent 1100 Series HPLC system using mass selective detection it is a complete system that fulfills the requirements of combinatorial chemistry and high­throughput analysis, offering robustness, ruggedness, sensitivi­ty, selectivity and speed. The inte­grated system plus the single soft­ware platform simplifies system setup, operation and management of large amounts of data. Details of the Agilent 220 MPS and how it works in an Agilent system for combinatorial chem­istry or high throughput screening is described in another Agilent technical note2. In this note we explain how to further optimize the hardware and software setup of the Agilent 220 MPS to achieve higher performance and preci­sion.
Loop Volume [µl] Sample Volume [µl]
535 10 40 20 55 50 95
Figure 1 Injection principle of the Agilent 220 micro plate sampler
6
5
4
3
2
1
6
5
4
3
2
1
Loading position Injection and run position
to column
to column
from pump
from pump
waste
waste
injection port
4.2 µl dead volume!
sample loop (fixed size)
sample loop (fixed size)
Table 1 Sample loop sizes and recommended sample volumes
0
1
2
3
4
5
0 20406080100
Sample Volu me [µl]
RSD Area [%]
5 µl Sample Loop
20 µl Sample Loop
mechanical precision of the dilu­tor syringe determines the inject­ed sample volume. For Partial Loop Fill & Inject the sample vol­ume plus a relatively large rinse volume is drawn from the sample. The rinse volume is used to rinse the injection port and the sample loop before the actual sample vol­ume is pushed into the sample loop. The precision is higher than for Centered Loop Fill but also more sample volume is required.
Parameters to optimize
Parameters influencing the perfor­mance of the sampler in sampling mode are:
1. Drawn sample volume (complete loop fill)
2. Draw speed (complete loop fill)
3. Size of air gap (complete loop fill)
4. Dilutor syringe size and sample loop size (centered loop fill)
1. Drawn sample volume
(complete loop fill)
Although the size of the sample loop determines the injected sam­ple volume the amount of drawn sample volume also influences the precision of the measurement3. Figure 2 shows the precision of peak areas for ten measurements for different drawn sample vol­umes
For the drawn sample amount of 5 µl the precision of peak area is very high because 5 µl is just enough sample to fill the injection port capillary (dead volume
4.2 µl). Increasing the sample vol-
ume also increases the precision of peak area. When the recom­mended sample volume is used the precision is usually optimized. More sample volume does not increase the precision any further.
If not enough sample volume for sufficient overfill of the sample loop is available, a lower precision of peak area results.
2. Draw speed (complete loop fill)
The draw speed for the sample as well as the inject speed influences the precision of the measurement as shown in figure 3.
Figure 3 Precision of peak area for different draw and inject speeds (5-µl sample loop, complete loop fill)
0.0
0.2
0.4
0.6
0.8
1.0
1.2
0.5 ml/min 1 ml/min 2 ml/min 3 ml/min
Draw Speed
RSD Area [%]
Figure 2 Precision of peak area for different sample volumes (5 and 20-µl sample loop, complete loop fill)
The best precision is achieved with the lowest draw speed of 0.5 ml/min, which is still higher than the default draw speed of the Agi­lent ChemStation (0.2 ml/min). Table 2 shows the necessary com­bined times for the Create Air Gap, Draw Sample and Inject steps for different sample volumes and different draw speeds.
The times required for the Create Air Gap, Draw Sample and Inject steps are in the range of a few sec­onds, which shouldn't make much difference for normal chromato­graphic runs. For high throughput analysis with run times in the range of 60–120 seconds a time of
13.56 seconds (55-µl sample vol-
ume, draw speed 0.5 ml/min) is about 10–20 % of the whole run time. Therefore it might be neces­sary to use higher draw speeds. For a draw speed of 3 ml/min the
Create Air Gap, Draw Sample
and Inject time of 2.26 seconds is only 2–4 % of the whole run time. The disadvantage of higher draw speed is lower precision of the analysis.
3. Size of air gap (complete loop fill)
It is strongly recommended to insert a Create Air Gap command before the Draw Sample com­mand in the CC-Mode method. The air gap prevents mixing of the sample with the solvent in the tube connected to the dilutor syringe. The optimal size of the air gap can be seen in figure 4.
An air gap between the sample and the solvent in the tube con­nected to the dilutor syringe is necessary to avoid mixing of sam­ple and solvent. The size of the air
gap should be between 1 and 4 µl, with a recommended size of 2–3 µl. The size of the air gap must not be larger than the dead volume of the injection port capillary of about 4.2 µl. For an air gap size of 5 µl, for example, some of the air gap is moved into the sample loop during injection. The effect is evi­dent for an air gap size of 10 µl. The loop is half filled with air from the air gap. Therefore less sample volume is applied to the system.
Figure 4 Precision of peak area for different air gap sizes
Draw Speed 0.5 ml/min 1 ml/min 3 ml/min
Sample Volume 35 µl 8.76 s 4.38 s 1.46 s 55 µl 13.56 s 6.78 s 2.26 s
Table 2 Times for Create Air Gap, Draw Sample and Inject steps for different sample volumes and draw speeds
0
1
2
3
4
5
0 µl 3 µl 5 µl 10 µl
Air Gap Size
RSD Area [%]
Dilutor syringe size and sam­ple loop size (centered loop fill)
To inject sample volumes smaller than the sample loop size the Cen- tered Loop Fill command is used. Figure 5 shows how a sandwich of air gaps and sample is injected into the sample loop. The sizes of the air gaps are calculated by the software by placing the sample in the middle of the sample loop (fig­ure 5).
For Centered Loop Fill the preci­sion of the drawn sample amount depends on the mechanical preci­sion of the dilutor syringe. There­fore, the precision of peak area for Centered Loop Fill is lower than for complete loop fill
4,5
. The solvent volumes for different sam­ple volumes are shown in table 3 (100-µl sample loop). While the air gap volumes are fixed (1 µl) the upper solvent volume is calculat­ed by the software by placing the sample volume in the middle of the sample loop. The area preci­sions for different sample vol­umes for two different dilutor syringe sizes are shown in figure 6.
The precision of peak area increases by applying more sam­ple volume to the sample loop. It is recommended to fill the sample loop to at least 50 % of the loop size. Figure 6 shows that the pre­cision of peak area is about 1 % at 50 µl sample volume. When using the 500-µl syringe the values are slightly higher than for the 100-µl syringe
4,5
.
Figure 5 Centered Loop Fill
Needle
Upper Solvent Volume
Upper Air Gap
Sample drawn from vial
Lower Air Gap
Solvent already in Sample Loop
Injection Port
Tubing
Sample Loop
Figure 6 Precision of area for different sample volumes for two different dilutor syringe sizes
0
2
4
6
8
10
12
14
16
5 µl 10 µl 20 µl 50 µl 80 µl
Sample Volume
RSD Area [%]
100 µl Syringe 500 µl Syringe
Sample Upper Solvent Upper Air Gap Sample Lower Air Gap
Volume [µl] Volume [µl] Volume [µl] Volume [µl] Volume [µl]
551 15 1 10 48 1 10 1 20 43 1 20 1 50 28 1 50 1 80 13 1 80 1
Table 3 Solvent volumes, upper and lower air gap sizes for different sample volumes (100-µl sample loop)
Partial Loop Fill
The CC-Mode software (version A.03.01) offers a new Partial Loop Fill feature. In addition to the sample volume a rinse volume is drawn from the sample. The rinse volume is used to flush the sample loop before the actual sample vol­ume is placed in it. The advantage over the Centered Loop Fill is better precision (figure 7), however, the sample volume required is much higher. There­fore, the decision which partial loop fill mode is used depends on the application. If high repro­ducibility is required and large sample volume is available the Partial Loop Fill should be select­ed. If only small sample volumes are available the Centered Loop Fill with the disadvantage of lower reproducibility has to be used.
Carry over
The Agilent 220 MPS and its soft­ware offer different options to minimize carry over, such as the wash vial and the rinse options6. The effects of these options on carry over are shown in table 4.
Important commands to remove carry over are the rinse functions
Rinse Needle Outside, Rinse Needle Inside and Rinse Injection Port. For the sample used in this
experiment the effect of the wash vial on the carry over was mini­mal, however, it becomes very important for samples with higher viscosity.
Figure 7 Area precision for Centered and Partial Loop Fill
0
1
2
3
5 µl 10 µl 30 µl
Sample Volu me
RSD Area [%]
Centered Loop Fill
Partial Loop Fill
Carry over [%]
Sample concentration No rinse or
Wash Vial Wash Vial
for 1 s,
Rinse Needle
[mg/l] wash function for 1 s
Outside
and
Rinse Injection Port
with 100 µl, 2 ml/min
100 3.2 2.9 0.1 200 1.9 1.9 0.1 300 2.2 1.6 0.1 500 1.9 1.3 0.1
1000 1.7 1.4 0.1
Table 4 Carry over results
Results
The Agilent 220 micro plate sam­pler is a precise instrument for high throughput analysis of large numbers of samples. To optimize
the performance for sample injec­tion the parameters and settings listed in table 5 have to be taken into consideration.
Parameters/Settings
Drawn sample volume If enough sample volume is available the sample loop should be (complete loop fill) overfilled with the recommended sample volume. The re-
commended sample volumes for different sample loop sizes are shown in table 1.
Draw speed The draw speed for
Create Air Gap, Draw Sample
and
Inject
(complete loop fill) should be set to 0.5 ml/min or less. Higher draw speeds lead to
decreasing precision. The time saved is minimal.
Size of air gap Create an air gap before sample is drawn. (complete loop fill) The size of the air gap should be between 1–4 µl, 2–3 µl are
recommended. The size of the air gap must not exceed 4 µl.
Dilutor syringe size and When using
Centered Loop Fill
a sample loop filled to at least sample loop size 50 % should be used. (centered loop fill) The volume of the dilutor syringe should be the same
or higher than the sample loop, but still as small as possible.
Partial Loop Fill The partial loop filling technique is used when high reproducibili
ty is required and large sample volume is available.
Carry over To avoid carry over use the commands
Rinse Needle Inside
,
Rinse Needle Outside
and
Rinse Injection Port
.
Use the
Wash Vial
command, especially for samples with high
viscosity.
Conclusion
This technical note described the optimization of the Agilent 220 MPS hardware and software for highest precision. The optimal set­tings for drawn sample volume, draw speed, size of the air gap and dilutor syringe size and sam­ple loop size were evaluated.
Furthermore, it was demonstrated how carry over could be mini­mized. This information should help optimize the Agilent 220 MPS for high performance and preci­sion.
Table 5 Summary of results
Copyright © 2000 Agilent Technologies All Rights Reserved. Reproduction, adaptation or translation without prior written permission is prohibited, except as allowed under the copyright laws.
Printed 07/2000 Publication Number 5980-0494E
References
1. “Agilent 220 micro plate sampler with Agilent 1100 Series: A solu­tion for high throughput HPLC analyses and purification” Agilent brochure, 2000, publication num- ber 5968-9101E
2. “Agilent 220 micro plate sampler with Agilent 1100 Series system for flexible, high throughput HPLC analyses” Agilent technical note, 1999, publication number 5968­5322E
3. John W. Dolan “Injection Loop Adsorption” LC/GC International, 1996, 4, 530-533
4. John W. Dolan “Autosampler Pre­cision” LC/GC International, 1998, 10, 910-914
5. John W. Dolan “Maintaining Autosampler Performance” LC/GC International, 1997, 7, 418-422
6. Jin Y. Huang, Travis Culley, John W. Dolan “Late Elution and Carry­over Peaks - A Case Study” LC/GC
International 1999, 4, 208-211
Udo Huber is an application chemist based at Agilent Technologies, Waldbronn, Germany
For more information on our products and ser­vices, visit our website at: http://www.agilent.com/chem
Loading...