Agilent Technologies N9340A User Manual

Agilent N9340A Handheld Spectrum Analyzer
User’s Guide
Notices
© Agilent Technologies, Inc. 2007 No part of this manual may be reproduced
Manual Part Number
N9340-90001
Edition
Second edition, April 2007
Printed in China Agilent Technologies, Inc.
Qianfeng Hi-Tech Industry Park Chengdu Hi-Tech Industrial Development Zone (West District) Chengdu 611731, P.R .Ch ina
Software Revision
This guide is valid for A.01.00 revisions of the Agilent N9340A Handheld Spectrum Analyzer software.
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not pro­ceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNI NG
A WARNING notice denotes a hazard. It calls attention to an operating pro­cedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR
252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Notices
© Agilent Technologies, Inc. 2007 No part of this manual may be reproduced
Manual Part Number
N9340-90001
Edition
Second edition, April 2007
Printed in China Agilent Technologies, Inc.
Qianfeng Hi-Tech Industry Park Chengdu Hi-Tech Industrial Development Zone (West District) Chengdu 611731, P.R .Ch ina
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR
252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
A CAUTION notice denotes a hazard. It calls attention to
CAU-CAUTION
an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Tab le of Co nt en ts
1 N9340A Overview
Introduction 2
Functionality 2 Optimized Usability 3
Front Panel Overview 4
Top Panel Overview 5
Display Annotations 6
Working with Batteries 7
Installing a Battery 7 Viewing Battery Status 7 Charging a Battery 8
2 Getting Started
Checking Shipment and Order List 10
Power Requirements 11
AC Power Cords 12
Safety Considerations 13
Electrical Requirements 15 Electrostatic Discharge (ESD) Precautions 15 Instrument Markings 16
Powering on the Analyzer for the First Time 17
Power on your N9340A 17
Preparation for Use 18
Setting up your N9340A 18 Testing buttons and screen pixel 18 Checking Instrument Information 19
Making a Basic Measurement 20
Viewing a Signal 21 Figure 1. View a signal (1 GHz, 0 dBm) 21
Contact Agilent Technologies 24
3 Making Measurements
Measuring Multiple Signals 26
Comparing Signals on the Same Screen 26
1
Measuring a Low-Level Signal 31
Improving Frequency Resolution and Accuracy 36
Making Distortion Measurements 37
One-button Power Measurement 43
Making a Stimulus Response Transmission Measurement 47
Measuring Stop Band Attenuation of a Low-pass Filter 49
Making a Reflection Calibration Measurement 51
Measuring Return Loss Using the Reflection Calibration Routine 54
4 System Setting
Visual and Audio Adjustment 56
Quick Display Adjustment 56 Manual Display Adjustment 56 Setting Key Beep 56
System Setting 57
General system settings 57 Ext Input 58
File 60
Saving a file 60 Loading a file 63
Show System 64
Show Error 65
Perform a Time Base Calibration 66
Factory Default Settings 67
5 Key Reference
Amplitude 70
Ref level 70 Attenuation 70 Preamp 71 Scale/DIV 71 Scale Type 71 Ref Offset 72
2
Y Axis Units 72
BW/SWP 73
RBW 73 VBW 74 VBW/RBW 74 Avg Type 74 Sweep Time 76 Sweep 77
Enter 79
ESC/CLR 80
Frequency 81
Center Frequency 81 Start Frequency 81 Stop Frequency 81 CF Step 82
Marker 83
Marker 83 Normal 83 Delta 84 Off 84 Mode 86 Marker Trace 87 All Off 87
Meas 88
Span 89
Span 89 Full 89 Zero 89 Last Span 89
TRACE 90
Trace 90 Clear Write 90 Max Hold 90 Minimum Hold 91 View 91 Blank 91 Detector 91 Average 93
3
Save Trace 94 Save As 94 Recall Trace 94
Limit 95
Limit Line 95 Limit Pattern 95 Set Pattern 95 Limit Type 95 Beep 96 Save Pattern 96 Recall Pattern 96
6 SCPI Command Reference
SCPI Language Basics 98
Basic Knowledge Requirement 99 Command Categories 99 Command Syntax 99 Standard Notations 100 Default unit of parameters 101
Common Commands 102
Clear Status 102 Identification Query 102 Reset 102
CALCulate Subsystem 103
Limit Line Subsection 103
DEMOdulation Subsystem 112
AM Demodulation 112 FM Demodulation 112
DISPlay Subsystem 113
Turn the Entire Display On/Off 113 Trace Y-Axis Scaling 113 Trace Y-Axis Reference Level Offset 114 Display Mode 115 Brightness 116
INITiate Subsystem 117
Continuous or Single Sweep 117
4
Initiate a single sweep 117
INSTrument Subsystem 118
Select One button measurement 118
MEASure Subsystem 119
OBW Subsection 119 ACPR Subsection 119
SENSe Subsystem 123
[:SENSe]:AVERage Subsection 123 FREQuency Subsection 128 SWEep Subsection 132
SYSTem Subsystem 133
Ext Input 133 Date Query 133 Time Query 133 Error Information Query 133
TRACe Subsystem 134
Tracking Generator Subsystem 135
State 135 Output Amplitude 135 Amplitude Offset 136 Normalization 137 Normalization Reference Level 137 Normalization Reference Position 138
TRIGer Subsystem 139
External Trigger Slope 139
5
Trigger Source 139
UNIT Subsystem 141
7Error Messages
Error Message List 144
8 Menu Map
AMPTD 150
BW/SWP 151
FREQ 152
MARKER 153
MEAS 154
SPAN 155
SYS 156
TRACE 157
Limit 158
6
Agilent N9340A Handheld Spectrum Analyzer
1 N9340A Overview
Agilent Technologies
1
1 N9340A Overview

Introduction

An Agilent N9340A is a handheld spectrum analyzer with a frequency range from 100 kHz to 3 GHz, tunable to 9 kHz. The analyzer provides ultimate measurement flexibility in a package that is ruggedized for field environments and convenient for mobile applications.

Functionality

An Agilent N9340A provides your with comprehensive functionality set and measurement convenience, including:
1 Pass/Fail judgement
Sets limit lines on the display for a quick Pass/Fail judgement.
2 Four- trace detection display
Displays as many as four independent traces simultaneously, being detected in any of the following modes: POS, NEG, Sample or Normal.
3 Marker frequency readout
Provides fine resolution down to a low of 1 Hz with a built- in frequency counter.
2 N9340A User’s Guide
N9340A Overview 1
4 Power Suite
Provides one-button measurement functionality on OBW ACPR
(Occupied Bandwidth), channel power, and
(Adjacent Channel Power Ratio).
5 High- sensitivity measurement
Includes a 3 GHz pre- amplifier (Option N9340A- PA3) enabling highly sensitive measurement which helps in the analysis of low-level signals.

Optimized Usability

An Agilent N9340A handheld spectrum analyzer provides enhanced usability:
7.2-inch sunlight viewable screen (
640 480×
pixels) enables you to read the scans easily and clearly in sun light.
Arc- shaped handle and rugged rubber casing ensure a comfortable and firm hold and ruggedness.
One- button measurement function provides an easy, fast access to power measurement.
File saving or loading via USB interface.
Instant screen control for power saving: pressing and holding [Preset] hardkey for over two seconds turns off the screen. Pressing any key on the front panel activates the screen.
N9340A User’s Guide 3
1 N9340A Overview

Front Panel Overview

14
N9340A
HANDHELD SPECTRUM ANALYZER
SYS MODE MEAS TRACE
100 kHz - 3.0 GHz
BW/ SWP
12
13
11
2DEF 3GHI1ABC
6
5MNO4JKL
PQR
VWX7STU9YZ_
8
LIMIT
0SAVE
ESC/CLR
MARKER
PRESET
ENTER
FREQ SPANAMPTD
10
9
8 7 6
2 3
1
4 5
Caption Function
1 Power Switch Toggles the analyzer between on and off
2 SYS Brings up a submenu for system setup
3 Function keys Includes hardkeys: FREQ, SPAN, AMPTD, BW/SWP, TRACE,
4 Preset Returns the analyzer to a known state
5 Enter Confirms a parameter selection or configuration
6 MARKER Activates marker function
7 ESC/CLR Exits and closes the dialog box or clears the character input
8 Save Quick saves the current trace
9 LIMIT Sets limit lines for quick Pass/Fail judgement
10 Arrow keys Increases or decreases a parameter step by step
11 Knob Selects an option item or edits a numerical parameter
12 Softkeys Indicates current menu functions
13 Speaker Active when in demodulation mode
14 Screen Displays measured traces and status
MEAS, and MODE
4 N9340A User’s Guide

Top Panel Overview

9
50 VDC MAX
dBm (2 W) MAX
33
RF INPUT 50 RF OUT 5 0
12-18 VDC
VDC
12-18
80 W MAX
55W MAX
1
2
1. DC Connector
Provides input for the DC power source via an AC-DC adapter. (Make sure that the line-power source outlet has a protective ground contact.)
2. Charging indicator
Lights when the battery is charging.
3. External power indicator
Lights when an external DC power applied to the analyzer via the AD- DC adapter.
4. USB interface (for PC connection)
Provides a connection between the analyzer and a PC for data transfer of further process.
5. USB interface
Provides an connection between the analyzer and an external USB memory disk.
6. LAN Interface (For future expansion)
7. RF OUT Connector (For future expansion)
8. EXT TRIG IN/REF IN (BNC, Female)
Connects to an external TTL signal or a 10 MHz reference signal. The TTL signal is used to trigger the analyzer’s internal sweep.
9. RF IN Connector (Type N, Female)
Accepts an external input with a frequency range from 100 kHz to 3 GHz, tunable to 9 kHz.
Ext. Pow er
Chargi ng
N9340A Overview 1
8
EXT TRIG IN/ EXT REF IN
PC
4
3
7
6
5
PC
N9340A User’s Guide 5
1 N9340A Overview

Display Annotations

1
2
13
14
12
3
4
5
6
8
7
9
10
11
Description Associated Function Key
1Time and Date [SYS] > {Setting} > {General}
2 Reference Level [AMPTD]
3 Amplitude Scale [AMPTD]
4 Detector Mode [TRACE] > {More (1 of 2)} > {Detector}
5 Center Frequency [FREQ] > {Center Freq}
6 Resolution Bandwidth [BW/SWP] > {RBW}
7 Display Status Line Displays analyzer status and error messages.
8Video Bandwidth [BW/SWP] > {VBW}
9Frequency Span [SPAN]
10 Sweep Time [BW/SWP] > {Sweep Time}
11 Key Menu See key label description in the key reference for
12 Key Menu Title Refers to the current activated menu function
13 Marker Frequency [Marker]
14 Marker Amplitude [Marker]
more information.
6 N9340A User’s Guide
N9340A Overview 1

Working with Batteries

Use the original battery supplied with the instrument or a genuine equivalent replacement from Agilent.

Installing a Battery

Insert a battery, ensuring correct orientation with the
CAU-CAUTION
electrodes correctly aligned.
Step Notes
1 Open the battery cover Use a cross-head type screwdriver to
2 Insert the battery
3 Close the battery cover Push the cover closed, then refasten
remove the screw, then pull the cover open.
the screw.

Viewing Battery Status

Determine the battery status by:
Checking the battery icon, which indicates the approximate level of charge. This is in the lower- right corner of the screen.
Press [SYS] > {More (1 of 3)} > {Show System} > {Page down} to check the current battery information.
N9340A User’s Guide 7
1 N9340A Overview

Charging a Battery

You can only charge the battery when it is in the analyzer either in operating or when the analyzer is shut down.
first using the analyzer. This typically takes about four hours.
Never use a modified or damaged AC-DC adapter.
CAU-CAUTION
1 Install the battery in the analyzer.
2 Simply attach the AC-DC adapter and switch on
external power.
3 The charge indicator lights, indicating that the
battery is charging. When the battery is fully charged, the green charging indicator turns off.
During charging and discharging, the battery will monitor and report its voltage, current, and temperature. If any of these monitored conditions exceed their safety limits, the battery will terminate any further charge or discharge until the error condition is corrected.
Fully charge the battery before
8 N9340A User’s Guide
Agilent N9340A Handheld Spectrum Analyzer
2 Getting Started
Conventions
1. A pair of curly brackets { } indicates a softkey, for example
{Start} refers to the Start softkey.
2. A pair of square brackets [ ] indicates a hardkey,
for example, [ENTER] refers to the ENTER hardkey.
3. An analyzer refers to Agilent N9340A handheld spectrum
analyzer in the following chapters.
4. A DUT refers to a device under test.
9
Agilent Technologies
9
2 Getting Started

Checking Shipment and Order List

We recommend you check the shipment and order list according to the following procedure when you receive the shipment.
Inspect the shipping container for damages. Signs of damage may include a dented or torn shipping container or cushioning material that indicate signs of unusual stress or compacting.
Carefully remove the contents from the shipping container, and check your shipment.
For any question or problem, contact Agilent Technologies Customer Contact Center (CCC) for consultant and service. See “Contact Agilent
Technologies" on page 24.
10 N9340A User’s Guide

Power Requirements

The AC power supply must meet the following requirements
Voltage : 100 VAC to 240 VAC
Frequency: 50 Hz to 60 Hz
Power: Maximum 62 W
The analyzer is equipped with a three- wire power cord, in accordance with international safety standards. This cable grounds the analyzer cabinet when connected to an appropriate power line outlet. The cable appropriate to the original shipping location is included with the analyzer.
Various AC power cables are available that are unique to specific geographic areas. You can order additional AC power cables for use in different areas. The table AC Power Cords lists the available AC power cables, the plug configurations, and identifies the geographic area in which each cable is appropriate.
The detachable power cord is the product disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the product. The front switch is only a standby switch and do not disconnect instrument from LINE power.
:
Getting Started 2
N9340A User’s Guide 11
A
2 Getting Started

AC Power Cords

Plug Type Cable Part
Number
8121-1703 BS 1363/A Option 900
250V 10A
8120-0696 AS 3112:2000 Option 901
250V 10A
8120-1692 IEC 83 C4 Option 902
250V 16A
8120-1521 CNS 10917-2
125V 10A
8120-2296 SEV 1011 Option 906
250V 10
8120-4600 SABS 164-1 Option 917
230V 15A
8120-4754 JIS C8303 Option 918
125V 15A
8120-5181 SI 32 Option 919
250V 16A
8120-8377 GB 1002 Option 922
a
Plug Description
/NEMA 5-15P
For use in Country & Region
United Kingdom, Hong Kong, Singapore, Malaysia
Australia, New Zealand
Continental Europe, Korea, Indonesia, Italy, Russia
Option 903
Unite States, Canada, Taiwan, Mexi c o
Switzerland
South Africa, India
Japan
Israel
China
250V 10A
12 N9340A User’s Guide

Safety Considerations

Agilent has designed and tested the N9340A handheld spectrum analyzer for Measurement, Control and Laboratory Use in accordance with Safety Requirements IEC 61010- 1: 2001, UL 61010- 1 (2004), and CSA C22.2 No.61010- 1- 04. The tester is supplied in a safe condition. The N9340A is also designed for use in Installation Category II and Pollution Degree 2 per IEC 61010 and IEC 60664 respectively.
Read the following safety notices carefully before you start to use a N9340A handheld spectrum analyzer to ensure safe operation and to maintain the product in a safe condition.
Getting Started 2
WARNING
Personal injury may result if the analyzer’s cover is removed. There are no operator-serviceable parts inside. Always contact Agilent qualified personnel for service. Disconnect the product from all voltage sources while it is being opened.
WARNING
This product is a Safety Class I analyzer. The main plug should be inserted in a power socket outlet only if provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
WARNING
WARNING
Electrical shock may result when cleaning the analyzer with the power supply connected. Do not attempt to clean internally.
case only.
Use a dry soft cloth to clean the outside
Always use the three-pin AC power cord supplied with this product. Failure to ensure adequate earth grounding by not using this cord may cause personal injury and product damage.
N9340A User’s Guide 13
2 Getting Started
WARNING
WARNING
CAU-CAUTION
CAU-CAUTION
CAU-CAUTION
NOTE
Danger of explosion if the battery is incorrectly replaced. Replace only with the same type recommended.
Do NOT dispose of batteries in a fire. Do NOT place batteries in the trash. Batteries must be
recycled or disposed of properly.
Recharge the battery only in the analyzer. If left unused, a fully charged battery will discharge itself over time.
Never use a damaged or worn-out adapter or battery. Charging the batteries internally, even while the analyzer is powered off, the analyzer may become warm. To avoid overheating, always disconnect the analyzer from the AC adapter before storing the analyzer into the soft carrying case.
We strongly recommend you to connect the automotive adapter to the power output connector for IT equipment, when charging the battery on your automotives.
The VxWorks operating system requires full conformity to USB 1.1 or USB 2.0 standards from a USB disk. Not all the USB disk are built that way. If you have problems connecting a particular USB disk, please reboot the analyzer before inserting the next.
Temperature extremes will affect the ability of the battery to charge. Allow the battery to cool down or warm up as necessary before use or charging.
Storing a battery in extreme hot or cold places will reduce the capacity and lifetime of a battery. Battery storage is recommended at a temperature of less than 45
o
C.
The LAN port is reserved for future expansion. It is
NOTE
inactive at this version.
14 N9340A User’s Guide
Getting Started 2
°°°

Environmental Requirements

A N9340A is designed for use under the following conditions:
Operating temperature: –10 C to +50 C (with battery) 0 C to +40 C (with adapter)
Storage temperature: –40 C to +70 C
Humidity: 95% or less
Altitude: 9200 m
°
°°

Electrical Requirements

This analyzer allows you to use either the original lithium- ion battery pack or the external AC- DC adapter shipped with the analyzer for power supply to the analyzer.

Electrostatic Discharge (ESD) Precautions

This analyzer was constructed in an ESD protected environment. This is because most of the semiconductor devices used in this analyzer are susceptible to damage by static discharge.
Depending on the magnitude of the charge, device substrates can be punctured of destroyed by contact or mere proximity of a static charge. The result can cause degradation of device performance, early failure, or immediate destruction.
These charges are generated in numerous ways, such as simple contact, separation of materials, and normal motions of persons working with static sensitive devices.
When handling or servicing equipment containing static sensitive devices, adequate precautions must be taken to prevent device damage or destruction. Only those who are thoroughly familiar with
N9340A User’s Guide 15
2 Getting Started
industry accepted techniques for handling static sensitive devices should attempt to service circuitry with these devices.

Instrument Markings

The CE mark shows that the product complies with all relevant European Legal Directives (If accompanied by a year, it signifies when the design was proven).
The CSA mark is a registered trademark of the Canadian Standards Association.
N10149
ISM1-A
The C- Tick mark is a registered trademark of the Australian Spectrum Management Agency.
This symbol is an Industrial Scientific and Medical Group 1 Class A product (CISPR 11, Clause 4)
The instruction manual symbol: indicates that the user must refer to specific instructions in the manual.
The standby symbol is used to mark a position of the instrument power switch.
16 N9340A User’s Guide
Getting Started 2

Powering on the Analyzer for the First Time

A N9340A is fitted with a transflective screen, which is viewable under all lighting conditions. In bright lighting conditions, the display is brighter and easier to read when you allow light to fall directly on the screen.
Use original standard adapter for AC-DC conversion.
CAU-CAUTION
The maximum RF input level of an average continuous power is 33 dBm (or + connecting a signal into the analyzer that exceeds the maximum level.

Power on your N9340A

Insert the battery into the analyzer or connect the analyzer to an external line power supply via the AC-DC adapter, then press the power switch on the front panel of your N9340A to power on the analyzer.
Allow the analyzer to warm- up for 30 minutes before making a calibrated measurement. To meet its specifications, the analyzer must meet operating temperature conditions.
50 VDC signal input). Avoid
Install battery
Press Power Switch
N9340A User’s Guide 17
Use stick stand
2 Getting Started

Preparation for Use

Use [SYS] hardkey to check or set the system settings of your analyzer.

Setting up your N9340A

1 Press [SYS] > {Setting} > {General} to set time and
power saving mode:
Press {Time/Date} to set the time and date for
your analyzer.
Press {Power Manager} to select a power- saving
mode from the follow three modes: turning off backlight, turning off screen display, and turning off both backlight and screen display.
2 Press [SYS] > {Setting} > {Language} to select a
language displayed on screen.
3 To manually adjust the display brightness and
contrast.
Press [SYS] > {Brightness} and then rotate the
knob to adjust display brightness.
Press [SYS] > {Contrast} and then rotate the
knob to adjust display contrast.
4 Press [SYS] > {More(2 of 3)} > {Key Beep} to toggle
the buzzer beep function between on and off.

Te st in g buttons and screen pixel

1 Press [SYS] > {More (2 of 3)} > {Diagnostics} > {Front
Panel} to test all the front panel keys except the [PRESET] hardkey and power switch.
2 Press [SYS] > {More (2 of 3)} > {Diagnostics} > {Black
Pixel} or {White Pixel} to test the basic display
ability of the LCD.
18 N9340A User’s Guide
Getting Started 2

Checking Instrument Information

1 Press [SYS] > {More (1 of 3)} > {Show System} to
display the system information.
2 Press [SYS] > {More (2 of 3)} > {Option} to display
the option information.
3 Press [SYS] > {More (1 of 3)} > {Show Error} to
display the error information.
N9340A User’s Guide 19
2 Getting Started

Making a Basic Measurement

This section provides information on basic analyzer operations with the assumption that you understand the front and top layout, and display annotations of your analyzer. If you do not, please refer to “Front Panel Overview" on page 4 and
“Top Panel Overview" on page 5.
For more information on making measurements, please refer to “Making Measurements" on page 25”.
Entering Data
When setting measurement parameters, there are several ways to enter or modify the value of an active function:
1 Using Knob
Increases or decreases the current value.
2 Using Arrow Keys
Increases or decreases the current value by step.
Press [FREQ] > {CF Step} to set the frequency by an auto- coupled step (Step = Span/10, when {CF Step} mode is set to Auto).
3 Using Numeric Keypad
Enters a specific value. Then press a terminator (either a specified unit softkey or [ENTER]) to confirm input.
4 Using Unit Softkeys
Terminates a value input which requires a unit.
5 Using Enter Key
Terminates an entry or confirms a selection.
20 N9340A User’s Guide
Loading...
+ 142 hidden pages