Agilent Technologies N9340A User Manual

Agilent N9340A Handheld Spectrum Analyzer
User’s Guide
Notices
© Agilent Technologies, Inc. 2007 No part of this manual may be reproduced
Manual Part Number
N9340-90001
Edition
Second edition, April 2007
Printed in China Agilent Technologies, Inc.
Qianfeng Hi-Tech Industry Park Chengdu Hi-Tech Industrial Development Zone (West District) Chengdu 611731, P.R .Ch ina
Software Revision
This guide is valid for A.01.00 revisions of the Agilent N9340A Handheld Spectrum Analyzer software.
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not pro­ceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNI NG
A WARNING notice denotes a hazard. It calls attention to an operating pro­cedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR
252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Notices
© Agilent Technologies, Inc. 2007 No part of this manual may be reproduced
Manual Part Number
N9340-90001
Edition
Second edition, April 2007
Printed in China Agilent Technologies, Inc.
Qianfeng Hi-Tech Industry Park Chengdu Hi-Tech Industrial Development Zone (West District) Chengdu 611731, P.R .Ch ina
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR
252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
A CAUTION notice denotes a hazard. It calls attention to
CAU-CAUTION
an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Tab le of Co nt en ts
1 N9340A Overview
Introduction 2
Functionality 2 Optimized Usability 3
Front Panel Overview 4
Top Panel Overview 5
Display Annotations 6
Working with Batteries 7
Installing a Battery 7 Viewing Battery Status 7 Charging a Battery 8
2 Getting Started
Checking Shipment and Order List 10
Power Requirements 11
AC Power Cords 12
Safety Considerations 13
Electrical Requirements 15 Electrostatic Discharge (ESD) Precautions 15 Instrument Markings 16
Powering on the Analyzer for the First Time 17
Power on your N9340A 17
Preparation for Use 18
Setting up your N9340A 18 Testing buttons and screen pixel 18 Checking Instrument Information 19
Making a Basic Measurement 20
Viewing a Signal 21 Figure 1. View a signal (1 GHz, 0 dBm) 21
Contact Agilent Technologies 24
3 Making Measurements
Measuring Multiple Signals 26
Comparing Signals on the Same Screen 26
1
Measuring a Low-Level Signal 31
Improving Frequency Resolution and Accuracy 36
Making Distortion Measurements 37
One-button Power Measurement 43
Making a Stimulus Response Transmission Measurement 47
Measuring Stop Band Attenuation of a Low-pass Filter 49
Making a Reflection Calibration Measurement 51
Measuring Return Loss Using the Reflection Calibration Routine 54
4 System Setting
Visual and Audio Adjustment 56
Quick Display Adjustment 56 Manual Display Adjustment 56 Setting Key Beep 56
System Setting 57
General system settings 57 Ext Input 58
File 60
Saving a file 60 Loading a file 63
Show System 64
Show Error 65
Perform a Time Base Calibration 66
Factory Default Settings 67
5 Key Reference
Amplitude 70
Ref level 70 Attenuation 70 Preamp 71 Scale/DIV 71 Scale Type 71 Ref Offset 72
2
Y Axis Units 72
BW/SWP 73
RBW 73 VBW 74 VBW/RBW 74 Avg Type 74 Sweep Time 76 Sweep 77
Enter 79
ESC/CLR 80
Frequency 81
Center Frequency 81 Start Frequency 81 Stop Frequency 81 CF Step 82
Marker 83
Marker 83 Normal 83 Delta 84 Off 84 Mode 86 Marker Trace 87 All Off 87
Meas 88
Span 89
Span 89 Full 89 Zero 89 Last Span 89
TRACE 90
Trace 90 Clear Write 90 Max Hold 90 Minimum Hold 91 View 91 Blank 91 Detector 91 Average 93
3
Save Trace 94 Save As 94 Recall Trace 94
Limit 95
Limit Line 95 Limit Pattern 95 Set Pattern 95 Limit Type 95 Beep 96 Save Pattern 96 Recall Pattern 96
6 SCPI Command Reference
SCPI Language Basics 98
Basic Knowledge Requirement 99 Command Categories 99 Command Syntax 99 Standard Notations 100 Default unit of parameters 101
Common Commands 102
Clear Status 102 Identification Query 102 Reset 102
CALCulate Subsystem 103
Limit Line Subsection 103
DEMOdulation Subsystem 112
AM Demodulation 112 FM Demodulation 112
DISPlay Subsystem 113
Turn the Entire Display On/Off 113 Trace Y-Axis Scaling 113 Trace Y-Axis Reference Level Offset 114 Display Mode 115 Brightness 116
INITiate Subsystem 117
Continuous or Single Sweep 117
4
Initiate a single sweep 117
INSTrument Subsystem 118
Select One button measurement 118
MEASure Subsystem 119
OBW Subsection 119 ACPR Subsection 119
SENSe Subsystem 123
[:SENSe]:AVERage Subsection 123 FREQuency Subsection 128 SWEep Subsection 132
SYSTem Subsystem 133
Ext Input 133 Date Query 133 Time Query 133 Error Information Query 133
TRACe Subsystem 134
Tracking Generator Subsystem 135
State 135 Output Amplitude 135 Amplitude Offset 136 Normalization 137 Normalization Reference Level 137 Normalization Reference Position 138
TRIGer Subsystem 139
External Trigger Slope 139
5
Trigger Source 139
UNIT Subsystem 141
7Error Messages
Error Message List 144
8 Menu Map
AMPTD 150
BW/SWP 151
FREQ 152
MARKER 153
MEAS 154
SPAN 155
SYS 156
TRACE 157
Limit 158
6
Agilent N9340A Handheld Spectrum Analyzer
1 N9340A Overview
Agilent Technologies
1
1 N9340A Overview

Introduction

An Agilent N9340A is a handheld spectrum analyzer with a frequency range from 100 kHz to 3 GHz, tunable to 9 kHz. The analyzer provides ultimate measurement flexibility in a package that is ruggedized for field environments and convenient for mobile applications.

Functionality

An Agilent N9340A provides your with comprehensive functionality set and measurement convenience, including:
1 Pass/Fail judgement
Sets limit lines on the display for a quick Pass/Fail judgement.
2 Four- trace detection display
Displays as many as four independent traces simultaneously, being detected in any of the following modes: POS, NEG, Sample or Normal.
3 Marker frequency readout
Provides fine resolution down to a low of 1 Hz with a built- in frequency counter.
2 N9340A User’s Guide
N9340A Overview 1
4 Power Suite
Provides one-button measurement functionality on OBW ACPR
(Occupied Bandwidth), channel power, and
(Adjacent Channel Power Ratio).
5 High- sensitivity measurement
Includes a 3 GHz pre- amplifier (Option N9340A- PA3) enabling highly sensitive measurement which helps in the analysis of low-level signals.

Optimized Usability

An Agilent N9340A handheld spectrum analyzer provides enhanced usability:
7.2-inch sunlight viewable screen (
640 480×
pixels) enables you to read the scans easily and clearly in sun light.
Arc- shaped handle and rugged rubber casing ensure a comfortable and firm hold and ruggedness.
One- button measurement function provides an easy, fast access to power measurement.
File saving or loading via USB interface.
Instant screen control for power saving: pressing and holding [Preset] hardkey for over two seconds turns off the screen. Pressing any key on the front panel activates the screen.
N9340A User’s Guide 3
1 N9340A Overview

Front Panel Overview

14
N9340A
HANDHELD SPECTRUM ANALYZER
SYS MODE MEAS TRACE
100 kHz - 3.0 GHz
BW/ SWP
12
13
11
2DEF 3GHI1ABC
6
5MNO4JKL
PQR
VWX7STU9YZ_
8
LIMIT
0SAVE
ESC/CLR
MARKER
PRESET
ENTER
FREQ SPANAMPTD
10
9
8 7 6
2 3
1
4 5
Caption Function
1 Power Switch Toggles the analyzer between on and off
2 SYS Brings up a submenu for system setup
3 Function keys Includes hardkeys: FREQ, SPAN, AMPTD, BW/SWP, TRACE,
4 Preset Returns the analyzer to a known state
5 Enter Confirms a parameter selection or configuration
6 MARKER Activates marker function
7 ESC/CLR Exits and closes the dialog box or clears the character input
8 Save Quick saves the current trace
9 LIMIT Sets limit lines for quick Pass/Fail judgement
10 Arrow keys Increases or decreases a parameter step by step
11 Knob Selects an option item or edits a numerical parameter
12 Softkeys Indicates current menu functions
13 Speaker Active when in demodulation mode
14 Screen Displays measured traces and status
MEAS, and MODE
4 N9340A User’s Guide

Top Panel Overview

9
50 VDC MAX
dBm (2 W) MAX
33
RF INPUT 50 RF OUT 5 0
12-18 VDC
VDC
12-18
80 W MAX
55W MAX
1
2
1. DC Connector
Provides input for the DC power source via an AC-DC adapter. (Make sure that the line-power source outlet has a protective ground contact.)
2. Charging indicator
Lights when the battery is charging.
3. External power indicator
Lights when an external DC power applied to the analyzer via the AD- DC adapter.
4. USB interface (for PC connection)
Provides a connection between the analyzer and a PC for data transfer of further process.
5. USB interface
Provides an connection between the analyzer and an external USB memory disk.
6. LAN Interface (For future expansion)
7. RF OUT Connector (For future expansion)
8. EXT TRIG IN/REF IN (BNC, Female)
Connects to an external TTL signal or a 10 MHz reference signal. The TTL signal is used to trigger the analyzer’s internal sweep.
9. RF IN Connector (Type N, Female)
Accepts an external input with a frequency range from 100 kHz to 3 GHz, tunable to 9 kHz.
Ext. Pow er
Chargi ng
N9340A Overview 1
8
EXT TRIG IN/ EXT REF IN
PC
4
3
7
6
5
PC
N9340A User’s Guide 5
1 N9340A Overview

Display Annotations

1
2
13
14
12
3
4
5
6
8
7
9
10
11
Description Associated Function Key
1Time and Date [SYS] > {Setting} > {General}
2 Reference Level [AMPTD]
3 Amplitude Scale [AMPTD]
4 Detector Mode [TRACE] > {More (1 of 2)} > {Detector}
5 Center Frequency [FREQ] > {Center Freq}
6 Resolution Bandwidth [BW/SWP] > {RBW}
7 Display Status Line Displays analyzer status and error messages.
8Video Bandwidth [BW/SWP] > {VBW}
9Frequency Span [SPAN]
10 Sweep Time [BW/SWP] > {Sweep Time}
11 Key Menu See key label description in the key reference for
12 Key Menu Title Refers to the current activated menu function
13 Marker Frequency [Marker]
14 Marker Amplitude [Marker]
more information.
6 N9340A User’s Guide
N9340A Overview 1

Working with Batteries

Use the original battery supplied with the instrument or a genuine equivalent replacement from Agilent.

Installing a Battery

Insert a battery, ensuring correct orientation with the
CAU-CAUTION
electrodes correctly aligned.
Step Notes
1 Open the battery cover Use a cross-head type screwdriver to
2 Insert the battery
3 Close the battery cover Push the cover closed, then refasten
remove the screw, then pull the cover open.
the screw.

Viewing Battery Status

Determine the battery status by:
Checking the battery icon, which indicates the approximate level of charge. This is in the lower- right corner of the screen.
Press [SYS] > {More (1 of 3)} > {Show System} > {Page down} to check the current battery information.
N9340A User’s Guide 7
1 N9340A Overview

Charging a Battery

You can only charge the battery when it is in the analyzer either in operating or when the analyzer is shut down.
first using the analyzer. This typically takes about four hours.
Never use a modified or damaged AC-DC adapter.
CAU-CAUTION
1 Install the battery in the analyzer.
2 Simply attach the AC-DC adapter and switch on
external power.
3 The charge indicator lights, indicating that the
battery is charging. When the battery is fully charged, the green charging indicator turns off.
During charging and discharging, the battery will monitor and report its voltage, current, and temperature. If any of these monitored conditions exceed their safety limits, the battery will terminate any further charge or discharge until the error condition is corrected.
Fully charge the battery before
8 N9340A User’s Guide
Agilent N9340A Handheld Spectrum Analyzer
2 Getting Started
Conventions
1. A pair of curly brackets { } indicates a softkey, for example
{Start} refers to the Start softkey.
2. A pair of square brackets [ ] indicates a hardkey,
for example, [ENTER] refers to the ENTER hardkey.
3. An analyzer refers to Agilent N9340A handheld spectrum
analyzer in the following chapters.
4. A DUT refers to a device under test.
9
Agilent Technologies
9
2 Getting Started

Checking Shipment and Order List

We recommend you check the shipment and order list according to the following procedure when you receive the shipment.
Inspect the shipping container for damages. Signs of damage may include a dented or torn shipping container or cushioning material that indicate signs of unusual stress or compacting.
Carefully remove the contents from the shipping container, and check your shipment.
For any question or problem, contact Agilent Technologies Customer Contact Center (CCC) for consultant and service. See “Contact Agilent
Technologies" on page 24.
10 N9340A User’s Guide

Power Requirements

The AC power supply must meet the following requirements
Voltage : 100 VAC to 240 VAC
Frequency: 50 Hz to 60 Hz
Power: Maximum 62 W
The analyzer is equipped with a three- wire power cord, in accordance with international safety standards. This cable grounds the analyzer cabinet when connected to an appropriate power line outlet. The cable appropriate to the original shipping location is included with the analyzer.
Various AC power cables are available that are unique to specific geographic areas. You can order additional AC power cables for use in different areas. The table AC Power Cords lists the available AC power cables, the plug configurations, and identifies the geographic area in which each cable is appropriate.
The detachable power cord is the product disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the product. The front switch is only a standby switch and do not disconnect instrument from LINE power.
:
Getting Started 2
N9340A User’s Guide 11
A
2 Getting Started

AC Power Cords

Plug Type Cable Part
Number
8121-1703 BS 1363/A Option 900
250V 10A
8120-0696 AS 3112:2000 Option 901
250V 10A
8120-1692 IEC 83 C4 Option 902
250V 16A
8120-1521 CNS 10917-2
125V 10A
8120-2296 SEV 1011 Option 906
250V 10
8120-4600 SABS 164-1 Option 917
230V 15A
8120-4754 JIS C8303 Option 918
125V 15A
8120-5181 SI 32 Option 919
250V 16A
8120-8377 GB 1002 Option 922
a
Plug Description
/NEMA 5-15P
For use in Country & Region
United Kingdom, Hong Kong, Singapore, Malaysia
Australia, New Zealand
Continental Europe, Korea, Indonesia, Italy, Russia
Option 903
Unite States, Canada, Taiwan, Mexi c o
Switzerland
South Africa, India
Japan
Israel
China
250V 10A
12 N9340A User’s Guide

Safety Considerations

Agilent has designed and tested the N9340A handheld spectrum analyzer for Measurement, Control and Laboratory Use in accordance with Safety Requirements IEC 61010- 1: 2001, UL 61010- 1 (2004), and CSA C22.2 No.61010- 1- 04. The tester is supplied in a safe condition. The N9340A is also designed for use in Installation Category II and Pollution Degree 2 per IEC 61010 and IEC 60664 respectively.
Read the following safety notices carefully before you start to use a N9340A handheld spectrum analyzer to ensure safe operation and to maintain the product in a safe condition.
Getting Started 2
WARNING
Personal injury may result if the analyzer’s cover is removed. There are no operator-serviceable parts inside. Always contact Agilent qualified personnel for service. Disconnect the product from all voltage sources while it is being opened.
WARNING
This product is a Safety Class I analyzer. The main plug should be inserted in a power socket outlet only if provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
WARNING
WARNING
Electrical shock may result when cleaning the analyzer with the power supply connected. Do not attempt to clean internally.
case only.
Use a dry soft cloth to clean the outside
Always use the three-pin AC power cord supplied with this product. Failure to ensure adequate earth grounding by not using this cord may cause personal injury and product damage.
N9340A User’s Guide 13
2 Getting Started
WARNING
WARNING
CAU-CAUTION
CAU-CAUTION
CAU-CAUTION
NOTE
Danger of explosion if the battery is incorrectly replaced. Replace only with the same type recommended.
Do NOT dispose of batteries in a fire. Do NOT place batteries in the trash. Batteries must be
recycled or disposed of properly.
Recharge the battery only in the analyzer. If left unused, a fully charged battery will discharge itself over time.
Never use a damaged or worn-out adapter or battery. Charging the batteries internally, even while the analyzer is powered off, the analyzer may become warm. To avoid overheating, always disconnect the analyzer from the AC adapter before storing the analyzer into the soft carrying case.
We strongly recommend you to connect the automotive adapter to the power output connector for IT equipment, when charging the battery on your automotives.
The VxWorks operating system requires full conformity to USB 1.1 or USB 2.0 standards from a USB disk. Not all the USB disk are built that way. If you have problems connecting a particular USB disk, please reboot the analyzer before inserting the next.
Temperature extremes will affect the ability of the battery to charge. Allow the battery to cool down or warm up as necessary before use or charging.
Storing a battery in extreme hot or cold places will reduce the capacity and lifetime of a battery. Battery storage is recommended at a temperature of less than 45
o
C.
The LAN port is reserved for future expansion. It is
NOTE
inactive at this version.
14 N9340A User’s Guide
Getting Started 2
°°°

Environmental Requirements

A N9340A is designed for use under the following conditions:
Operating temperature: –10 C to +50 C (with battery) 0 C to +40 C (with adapter)
Storage temperature: –40 C to +70 C
Humidity: 95% or less
Altitude: 9200 m
°
°°

Electrical Requirements

This analyzer allows you to use either the original lithium- ion battery pack or the external AC- DC adapter shipped with the analyzer for power supply to the analyzer.

Electrostatic Discharge (ESD) Precautions

This analyzer was constructed in an ESD protected environment. This is because most of the semiconductor devices used in this analyzer are susceptible to damage by static discharge.
Depending on the magnitude of the charge, device substrates can be punctured of destroyed by contact or mere proximity of a static charge. The result can cause degradation of device performance, early failure, or immediate destruction.
These charges are generated in numerous ways, such as simple contact, separation of materials, and normal motions of persons working with static sensitive devices.
When handling or servicing equipment containing static sensitive devices, adequate precautions must be taken to prevent device damage or destruction. Only those who are thoroughly familiar with
N9340A User’s Guide 15
2 Getting Started
industry accepted techniques for handling static sensitive devices should attempt to service circuitry with these devices.

Instrument Markings

The CE mark shows that the product complies with all relevant European Legal Directives (If accompanied by a year, it signifies when the design was proven).
The CSA mark is a registered trademark of the Canadian Standards Association.
N10149
ISM1-A
The C- Tick mark is a registered trademark of the Australian Spectrum Management Agency.
This symbol is an Industrial Scientific and Medical Group 1 Class A product (CISPR 11, Clause 4)
The instruction manual symbol: indicates that the user must refer to specific instructions in the manual.
The standby symbol is used to mark a position of the instrument power switch.
16 N9340A User’s Guide
Getting Started 2

Powering on the Analyzer for the First Time

A N9340A is fitted with a transflective screen, which is viewable under all lighting conditions. In bright lighting conditions, the display is brighter and easier to read when you allow light to fall directly on the screen.
Use original standard adapter for AC-DC conversion.
CAU-CAUTION
The maximum RF input level of an average continuous power is 33 dBm (or + connecting a signal into the analyzer that exceeds the maximum level.

Power on your N9340A

Insert the battery into the analyzer or connect the analyzer to an external line power supply via the AC-DC adapter, then press the power switch on the front panel of your N9340A to power on the analyzer.
Allow the analyzer to warm- up for 30 minutes before making a calibrated measurement. To meet its specifications, the analyzer must meet operating temperature conditions.
50 VDC signal input). Avoid
Install battery
Press Power Switch
N9340A User’s Guide 17
Use stick stand
2 Getting Started

Preparation for Use

Use [SYS] hardkey to check or set the system settings of your analyzer.

Setting up your N9340A

1 Press [SYS] > {Setting} > {General} to set time and
power saving mode:
Press {Time/Date} to set the time and date for
your analyzer.
Press {Power Manager} to select a power- saving
mode from the follow three modes: turning off backlight, turning off screen display, and turning off both backlight and screen display.
2 Press [SYS] > {Setting} > {Language} to select a
language displayed on screen.
3 To manually adjust the display brightness and
contrast.
Press [SYS] > {Brightness} and then rotate the
knob to adjust display brightness.
Press [SYS] > {Contrast} and then rotate the
knob to adjust display contrast.
4 Press [SYS] > {More(2 of 3)} > {Key Beep} to toggle
the buzzer beep function between on and off.

Te st in g buttons and screen pixel

1 Press [SYS] > {More (2 of 3)} > {Diagnostics} > {Front
Panel} to test all the front panel keys except the [PRESET] hardkey and power switch.
2 Press [SYS] > {More (2 of 3)} > {Diagnostics} > {Black
Pixel} or {White Pixel} to test the basic display
ability of the LCD.
18 N9340A User’s Guide
Getting Started 2

Checking Instrument Information

1 Press [SYS] > {More (1 of 3)} > {Show System} to
display the system information.
2 Press [SYS] > {More (2 of 3)} > {Option} to display
the option information.
3 Press [SYS] > {More (1 of 3)} > {Show Error} to
display the error information.
N9340A User’s Guide 19
2 Getting Started

Making a Basic Measurement

This section provides information on basic analyzer operations with the assumption that you understand the front and top layout, and display annotations of your analyzer. If you do not, please refer to “Front Panel Overview" on page 4 and
“Top Panel Overview" on page 5.
For more information on making measurements, please refer to “Making Measurements" on page 25”.
Entering Data
When setting measurement parameters, there are several ways to enter or modify the value of an active function:
1 Using Knob
Increases or decreases the current value.
2 Using Arrow Keys
Increases or decreases the current value by step.
Press [FREQ] > {CF Step} to set the frequency by an auto- coupled step (Step = Span/10, when {CF Step} mode is set to Auto).
3 Using Numeric Keypad
Enters a specific value. Then press a terminator (either a specified unit softkey or [ENTER]) to confirm input.
4 Using Unit Softkeys
Terminates a value input which requires a unit.
5 Using Enter Key
Terminates an entry or confirms a selection.
20 N9340A User’s Guide
Getting Started 2

Viewing a Signal

1 Use a signal generator to generate a
continuous- wave signal (1 GHz, 0 dBm).
2 Press [SYS] > {More (1 0f 3)} > {Preset} and select
Default to toggle the preset setting to
factory- defined status.
3 Press the green [Preset] key to restore the
analyzer to its factory- defined setting.
4 Connect the generator’s RF OUT connector to
analyzer’s RF IN connector on the top panel.
5 Press [FREQ] > {Center Freq} > 1> {GHz} to set the
center frequency to 1 GHz.
6 Press [SPAN] > 5 > {MHz} to set the frequency
span to 5 MHz.
7 Press [MARKER] > {Peak Search} > {Peak} to place a
marker (labeled 1) at the highest peak (1 GHz) on the display.
The values of amplitude and frequency of the Marker appear both in the function block and in the upper- right corner of the screen.
Use the knob, the arrow keys, or the softkeys in the Peak Search menu to move the marker and read out the value of both frequency and amplitude displayed on the screen.
Figure 1 View a signal (1 GHz, 0 dBm)
N9340A User’s Guide 21
2 Getting Started

Some helpful tips

Performing a time-base calibration
A N9340A provides a manual calibration function to calibrate the time base.
Before calibration, allow the analyzer to warm up for about 30 minutes.
Use a BNC cable to connect a 10 MHz reference signal to the EXT TRIG IN connector of your N9340A, then press [SYS] > {More (2 of 3)} > {Calibration} > {Time
Base} to initiate a calibration.
Selecting a preset type
The analyzer has two types of preset setting for you to choose from:
Default
Restores the analyzer to its factory- defined setting.
User
Restores the analyzer to a user- defined setting.
Powering on the instrument always brings up a
NOTE
factory-defined setting no matter what the Preset setting.
Saving a User-defined Preset
If you constantly use system setting that is not the factory default, please refer to the following steps to create a user- defined system setting:
1 Set analyzer parameters by the knob, the arrow
keys or the numeric keypad.
2 Press [SYS] > {More (1 of 3)} > {Preset} > {Save User}
to save the current parameters as the user preset setting.
3 Press [SYS] > {More (1 of 3)} > {Preset} (User) to set
the preset mode to user defined system setting.
22 N9340A User’s Guide
Upgrading Firmware
When updating firmware, make sure there will be a
CAU-CAUTION
constant power supply for at least 10 minutes. Power cut off during the updating process can cause damage to the instrument.
A N9340A provides an easy and fast access for firmware update.
1 Set up a folder named N9340DATA in the root
directory of your USB memory stick.
2 Store the firmware update application into the
folder named N9340DATA.
3 Insert the USB memory stick into the USB
connector on the top panel of your analyzer.
4 Press [SYS] > {More (2 of 3)} > {Update} >
{Firmware} to activate the firmware updating
procedure. Then the instrument will then perform the update automatically.
Adding an option
Pressing [SYS] > {More (2 of 3)} > {Option} > {Add} brings up a dialog box for entering the option license code. Use the numeric keypad to input the option license code and use the [Enter] key as a terminator. If the analyzer recognises the option license code, an message “Option install successfully” appears in the status line, otherwise a message “Invalid option licence” appears in the status line.
Getting Started 2
N9340A User’s Guide 23
2 Getting Started

Contact Agilent Technologies

Agilent has Sales and Service Offices around the world to provide you with complete support.
Go to http://www.agilent.com/find/assist, for help with:
product selection, configuration, and purchases.
technical and application assistance, and
consulting.
rental and leasing options, and refurbished equipment.
repair, calibration, education and training.
If you do not have access to the internet, call the appropriate number shown below. Or contact your local Agilent Technologies Sales and Service Office.
United States (tel) 800 829 4444
Canada (tel) 877 894 4414
Latin America (tel) +1 (305) 269 7500
China (tel) 800 810 0189
Korea (tel) 080 769 0800
Japan (tel) +81 426 56 7832
Taiwan (tel) 0800 047 866
Europe (tel) +31 20 547 2111
Australia (tel) 1 800 629 485
Other Asia Pacific Countries
(fax) 800 829 4433
(fax) 800 746 4866
(fax) 800 820 2816
(fax) 080 769 0900
(fax) +81 426 56 7840
(fax) 0800 286 331
(fax) +61 (3) 9210 5947
(tel) +65 6375 8100 (fax) +65 6755 0042 Email: tm_ap@agilent.com
24 N9340A User’s Guide
Agilent N9340A Handheld Spectrum Analyzer
3 Making Measurements
25
3 Making Measurements

Measuring Multiple Signals

This section provides information on measuring multiple signals.

Comparing Signals on the Same Screen

An N9340A provides an easy function for you to compare frequency and amplitude differences between signals, such as radio or television signal spectra. Using Delta Marker function allows you to compare two signals when both appear on the screen at the same time. In this following example, a 50 MHz input signal is used to measure frequency and amplitude differences between two signals on the same screen. Delta marker is used to demonstrate this comparison.
1 Press [PRESET] to set the analyzer to a factory
default setting.
2 Input a signal (0 dB, 50 MHz) to the RF IN
connector of the analyzer.
3 Set the analyzer start frequency, stop frequency
and reference level to view the 50 MHz signal and its harmonics up to 100 MHz:
Press [FREQ] > {Start Freq} > 40 > {MHz}
Press [FREQ] > {Stop Freq} >110 > {MHz}
Press [AMPTD] > {Ref Level} > 0 > {dBm}
4 Press [MARKER] > {Peak search} to place a marker
on the highest peak on the display (50 MHz). The {Next Left PK} and {Next Right PK} softkeys are available to move the marker from peak to peak.
5 Press [MARKER] > {Delta} to anchor the first
marker (labeled as 1) and activate a delta marker:
The label on the first marker now reads 1R, indicating that it is the reference point.
26 N9340A User’s Guide
Making Measurements 3
6 Move the second marker to another signal peak
using the front panel knob or by using the {Peak Search} softkey:
Press [MARKER] > {Peak Search} > {Next Right PK} or
{Next Left PK}.
To increase the resolution of the marker readings, turn on
NOTE
the frequency count function. For more information, please refer to “Improving Frequency Resolution and
Accuracy" on page 36
Figure 2 Delta pair marker with signals on the same
screen
N9340A User’s Guide 27
3 Making Measurements

Resolving Signals of Equal Amplitude

In this example a decrease in resolution bandwidth is used in combination with a decrease in video bandwidth to resolve two signals of equal amplitude with a frequency separation of 100 kHz.
Notice that the final RBW selection to resolve the signals is the same width as the signal separation while the VBW is slightly narrower than the RBW.
1 Connect two sources to the analyzer input as
shown below.
Figure 3
Setup for obtaining two signals
N9310A
RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby On
Signal generator
N9310A
RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby On
FUNCTIONS
Utility
Preset
AM I/Q
Frequency
Amplitude FM
Local
File
Sweep
LF Out
TriggerPulseM
Enter
MOD
9
7
8
On/Off
6
4
5
3
102
RF
·
On/Off
LF OUT RF O UT 50
REVERSE PWR 4W MAX 30VDC
FUNCTIONS
Utility
Preset
AM I/Q
Frequency
Amplitude FM
Local
File
Sweep
Trigger
PulseM
LF Out
Enter
MOD
7
8
9
On/Off
4
5
6
1
3
2
RF
·
0
On/Off
LF OUT RF OUT 50
REVERSE PWR 4W MAX 30VDC
Directional coupler
Signal generator
2 Set one source to 300 MHz. Set the frequency of
the other source to 300.1 MHz. Set both source amplitudes to –20 dBm.
3 Setup the analyzer to view the signals:
• Press [PRESET].
Press [FREQ] > {Center Freq} > 300.05 > {MHz}.
Press [SPAN] > {Span} > 2 > {MHz}.
Press [BW/SWP] > {RBW} > 30 > {kHz}.
28 N9340A User’s Guide
Making Measurements 3
Use the knob or the arrow keys to further reduce the resolution bandwidth and better resolve the signals. As you decrease the resolution bandwidth, you improve resolution of the individual signals and also increased the sweep timed. For fastest measurement times, use the widest possible resolution bandwidth. Under factory preset conditions, the resolution bandwidth is coupled to the span.
Figure 4 Resolving signals of equal amplitude
N9340A User’s Guide 29
3 Making Measurements
Resolving Small Signals
Hidden by Large Signals
This example uses narrow resolution bandwidths to resolve two input signals with a frequency separation of 50 kHz and an amplitude difference of 60 dB.
1 Connect two sources to the analyzer input
connector as shown in Figure 3 on page 28.
2 Set one source to 300 MHz at –10 dBm. Set the
other source to 300.05 MHz at –70 dBm.
3 Set the analyzer as follows:
Press [PRESET]
Press [FREQ] > {Center Freq} > 300.05 > {MHz}
Press [SPAN] > [500] > kHz.
Press [BW/SWP] > {RBW} > 300 > {Hz}
4 Reduce the resolution bandwidth filter to view
the smaller hidden signal. Place a delta marker on the smaller signal:
Press [MARKER] > {Peak Search}
Press [MARKER] > {Delta}
Press [MARKER] > {Peak Search} > {Next Right PK}
or {Next Left PK}.
Figure 5 Resolving small signal hidden by a large signal
30 N9340A User’s Guide

Measuring a Low-Level Signal

This section provides information on measuring low-level signals and distinguishing them from spectrum noise. There are four main useful techniques as follows to measure a low-level signal.

Reducing Input Attenuation

The ability to measure a low- level signal is limited by internally generated noise in the spectrum analyzer. The input attenuator affects the level of a signal passing through the analyzer. If a signal is very close to the noise floor, reducing input attenuation will bring the signal out of the noise.
1 Preset the analyzer:
2 Input a signal (1 GHz, –80 dBm) to the analyzer
RF input connector.
3 Set the center frequency, span and reference
level:
Press [FREQ] > {Center Freq} > 1 > {GHz}
Press [SPAN] > 5 > {MHz}
Press [AMPTD] > {Ref Level} > –40 > {dBm}
4 Move the desired peak (in this example, 1 GHz)
to the center of the display:
Press [MARKER] > {Peak Search}
Press [MARKER] > {Marker To} > {To Center}
Figure 6 A signal closer to the noise level (Atten:10 dB)
Making Measurements 3
N9340A User’s Guide 31
3 Making Measurements
5 Reduce the span to 1 MHz and if necessary
re- center the peak.
Press [SPAN] > [1] > {MHz}.
6 Set the attenuation to 20 dB. Note that
increasing the attenuation moves the noise floor closer to the signal level.
Press [AMPTD] > {Attenuation} > 20 > {dB}.
Figure 7 A signal closer to the noise level (Atten: 20 dB)
7 Press [AMPTD] > Attenuation > 0 > {dB} to set the
attenuation to 0 dB.
Figure 8 A signal closer to the noise level (Atten: 0 dB)
32 N9340A User’s Guide
Making Measurements 3

Decreasing the Resolution Bandwidth

Resolution bandwidth settings affect the level of internal noise without affecting the level of continuous wave (CW) signals. Decreasing the RBW by a decade reduces the noise f loor by 10 dB.
1 Refer to “Reducing Input Attenuation" on
page 31, and follow steps 1, 2 and 3.
2 Decrease the resolution bandwidth:
Press [BW/SWP] > {RBW}, and toggle RBW
setting to Man (manual), then decrease the resolution bandwidth using the knob, the
arrow keys or the numeric keypad. The low level signal appears more clearly because the noise level is reduced.
Figure 9 Decreasing resolution bandwidth
N9340A User’s Guide 33
3 Making Measurements
Using the Average Detector and
Increased Sweep Time
When the analyzer’s noise masks low- level signals, changing to the average detector and increasing the sweep time smoothes the noise and improves the signal’s visibility. Slower sweeps are required to average more noise variations.
1 Refer to “Reducing Input Attenuation" on
page 31, and follow steps 1, 2 and 3.
2 Press [TRACE] > {More (1 of 2)} > {Detector} >
{Average} to select the average detector.
3 Press [BW/SWP] > {Sweep Time} to set the sweep
time to 500 ms. Note how the noise smoothes out, as there is more time to average the values for each of the displayed data points.
4 Press [BW/SWP] > {Avg Type} to change the
average type.
Figure 10 Using the average detector
34 N9340A User’s Guide
Making Measurements 3

Tra c e Ave ra gin g

Averaging is a digital process in which each trace point is averaged with the previous average for the same trace point. Selecting averaging, when the analyzer is auto coupled, changes the detection mode to sample, smoothing the displayed noise level.
This is a trace processing function and is not the same as
NOTE
using the average detector (as described on page 34).
1 Refer to the first procedure “Reducing Input
Attenuation" on page 31, and follow steps 1, 2
and 3.
2 Press [TRACE] > {Average} (On) to turn average on.
3 Press 50 > [ENTER] to set the average number to
50.
As the averaging routine smoothes the trace, low level signals become more visible.FM
Figure 11 Trace averaging
N9340A User’s Guide 35
3 Making Measurements

Improving Frequency Resolution and Accuracy

Using the frequency counter to improve frequency resolution and accuracy.
Marker count properly functions only on CW signals or
NOTE
discrete spectral components. The marker must be > 40 dB above the displayed noise level.
1 Press [PRESET]. (Factory Preset)
2 Input a signal (1 GHz, –30 dBm) to the
analyzer’s RF IN connector.
3 Set the center frequency to 1 GHz and the span
to 5 MHz:
4 Press [MARKER] > {More (1 of 2)} > {Mode} > {Freq
Count} to turn the frequency counter on.
5 Move the marker by rotating the knob, half- way
down the skirt of the signal response.
6 Press [MARKER] > {More (1 of 2)} > {Mode} >
{Normal} to turn off the marker counter.

Figure 12 Using the frequency counter

36 N9340A User’s Guide
Making Measurements 3

Making Distortion Measurements

This section provides information on measuring and identifying signal distortion.

Identifying Analyzer Generated Distortion

High level input signals may cause analyzer distortion products that could mask the real distortion measured on the input signal. Use trace and the RF attenuator to determine which signals, if any, are internally generated distortion products. In this example, use a signal from a signal generator to determine whether the harmonic distortion products are generated by the analyzer.
1 Input a signal (200 MHz, –10 dBm) to the
analyzer RF IN connector.
2 Set the analyzer center frequency and span:
Press [PRESET]. (Factory Preset)
Press [FREQ] > {Center Freq} > 400 > {MHz}.
Press [SPAN] > 700 > {MHz}.
The signal produces harmonic distortion products (spaced 200 MHz from the original 200 MHz signal)
Figure 13 Harmonic distortion
3 Change the center frequency to the value of the
first harmonic:
N9340A User’s Guide 37
3 Making Measurements
Press [MARKER] > {Peak Search}.
Press [MARKER] > {Marker To} > {To Center}.
4 Change the span to 50 MHz and re- center the
signal:
Press [SPAN] > 50 > {MHz}.
Press [MARKER] > {Peak Search}.
5 Set the attenuation to 0 dB:
Press [AMPTD] > {Attenuation} > 0 > {dB}.
Press [MARKER] > {Marker To} > {To Ref}.
6 To determine whether the harmonic distortion
products are generated by the analyzer, first save the trace data in trace 2 as follows:
Press [TRACE] > {Trace (2)}.
Press [TRACE] > {Clear Write}.
7 Allow trace 2 to update (minimum two sweeps),
then store the data from trace 2 and place a delta marker on the harmonic of trace 2:
Press [TRACE] > {View}.
Press [MARKER] > {Peak Search}.
Press [Marker] > {Delta}.
Please refer to figure below shows the stored data in trace 2 and the measured data in trace 1. The Marker indicator reads the difference in amplitude between the reference and active markers.
38 N9340A User’s Guide
Making Measurements 3
Figure 14 Identifying Analyzer Distortion (O dB atten)
8 Press [AMPTD] > {Attenuation} > 10 > {dB} to
increase the RF attenuation to 10 dB.
Figure 15 Identifying Analyzer Distortion (10 dB atten)
The marker amplitude readout comes from two sources:
Increased input attenuation causes poorer
signal- to- noise ratio. This causes the marker
to be positive.
The reduced contribution of the analyzer
circuits to the harmonic measurement causes
the Marker to be negative. Large marker readout indicates significant measurement errors. Set the input attenuator to minimize the absolute value of marker.
N9340A User’s Guide 39
3 Making Measurements

Third-Order Intermodulation Distortion

Two- tone, third- order intermodulation distortion is a common test in communication systems. When two signals are present in a non-linear system, they may interact and create third- order intermodulation distortion (TOI) products that are located close to the original signals. System components such as amplifiers and mixers generates these distortion products. In this example we test a device for third- order intermodulation using markers. Two sources are used, one set to 300 MHz and the other to 301 MHz.
1 Connect the equipment as shown in figure below.
N9310A
RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby
On
Signal generator
N9310A
RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby On
Signal generator
FUNCTIONS
Frequency
Utility
Preset
AM I / Q
Sweep
Amplitude FM
Local
File
Enter
LF Out
TriggerPul seM
MOD
7
8
9
On/Off
4
5
6
102
3
RF
·
On/Off
LF OUT RF OU T 5 0
Directional
REVERSE PWR 4W MAX 30VDC
coupler
FUNCTIONS
Frequency
Utility
Preset
AM I/Q
Sweep
Amplitude FM
Local
File
Enter
LF Out
TriggerPulseM
MOD
7
8
9
On/Off
4
5
6
3
102
RF
·
On/Off
LF OUT RF OUT 50
REVERSE PWR 4W MAX 30VDC
This combination of signal generators, low pass filters, and directional coupler (used as a combiner) results in a two- tone source with very low intermodulation distortion.
Although the distortion from this setup may be better than the specified performance of the analyzer, it is useful for determining the TOI performance of the source/analyzer combination. After the performance of the source/analyzer combination has been verified, the DUT (Device under test, for example, an amplifier) would be inserted between the directional coupler output
40 N9340A User’s Guide
Making Measurements 3
and the analyzer input.
The coupler should have a high degree of isolation
NOTE
between the two input ports so the sources do not intermodulate.
2 Set one source (signal generator) to 300 MHz
and the other source to 301 MHz, for a frequency separation of 1 MHz. Set the sources equal in amplitude as measured by the analyzer (in this example, they are set to –5 dBm).
3 Set the analyzer center frequency and span:
Press [PRESET]. (With Preset Type of Default)
Press [FREQ] > {Center Freq} > 300.5 > {MHz}.
Press [SPAN] > 5 > {MHz}.
4 Reduce the RBW until the distortion products
are visible:
Press [BW/SWP] > {RBW}, and reduce the RBW
using the knob, the arrow keys or the numeric
keypad.
5 Move the signal to the reference level:
Press [MARKER] > {Peak Search}.
Press [MARKER] > {Marker To} > {To Ref}.
6 Reduce the RBW until the distortion products
are visible:
Press [BW/SWP] > {RBW}, and reduce the RBW
using the knob, the arrow keys or the numeric
keypad.
7 Activate the second marker and place it on the
peak of the distortion product (beside the test signal) using the Next Peak:
Press [MARKER] > {Delta}.
Press [MARKER] > {Peak Search} > {Next Left
(Right) PK}.
8 Measure the other distortion product:
Press [MARKER] > {Normal}.
N9340A User’s Guide 41
3 Making Measurements
Press [MARKER] > {Peak Search} > {Next Left
(Right) Peak}.
9 Measure the difference between this test signal and the second distortion product.
Press [Marker] > {Normal}.
Press [MARKER] > {Peak Search} > {Next Left
(Right) Peak}.
Figure 16 TOI test screen
42 N9340A User’s Guide
Making Measurements 3

One-button Power Measurement

N9340A provides one-button measurement functionality on OBW power and ACPR easy- to- use access for your measurement of interest.

Measuring OBW

Occupied Bandwidth (OBW) integrates the power of the spectrum in the displayed green frame. The measurement defaults to 99% of the occupied bandwidth power. Occupied bandwidth measurement are made in
either single or continuous sweep mode.
Selecting a Method
Press [MEAS] > {OBW} > {Method} to toggle the occupied bandwidth measurement mode between
percentage (%)and dBc.
• Setting Percentage (%)
Press {OBW} > {Method} > {Percent} to edit the percentage of signal power used when determining the occupied bandwidth. Press {%} to set the percentage ranging from 10.00% to 99.99%.
•Setting dBc
Press {OBW} > {Method} > {dBc} to specify the power level used to determine the emission bandwidth as the number of dB down from the highest signal peak, within the occupied bandwidth span. This function is an independent calculation from the OBW calculation. The x dB Bandwidth result is also called the emissions bandwidth (EBW). This sets the green frame (f1 and f2 are the two cross points of the signal and the pair of vertical lines of the displayed green frame) to your specified dB value below the maximum power, and compute the total power in the frame. The emission bandwidth is then calculated ranging from 0.1 dB to 100 dB (the default value is set to 26 dB).
(Occupied Band Width), channel
(Adjacent Channel Power Ratio) as an
N9340A User’s Guide 43
3 Making Measurements

Measuring ACPR

The adjacent channel power ratio (ACPR) measures the power ratio between the main channel power and the adjacent channel power.
Center Frequency
Sets the center frequency of the main channel power.
Key Access: [MEAS] > {ACPR} > {Center Freq}
Main Channel
Specifies the range of integration used in calculating the power in the main channel. Use the knob, the arrow keys, or the numeric keypad to set the bandwidth.
Key Access: [MEAS] > {ACPR} > {Main Channel}
Adjacent Channel
Specifies the range of integration used in calculating the power in the adjacent channel. Use the knob, the arrow keys, or the numeric keypad to set the bandwidth.
Key Access: [MEAS] > {ACPR} > {Adj Channel}
Channel Space
Sets the space value between the center frequency of main channel power and that of the adjacent channel power.
Key Access: [MEAS] > {ACPR} > {Channel Space}
44 N9340A User’s Guide
Making Measurements 3

Measuring Channel Power

Measures the power and power spectral density in the channel bandwidth that you specified. One pair of vertical lines of the displayed green frame indicates the edges of the channel bandwidth. The center frequency, reference level, and channel bandwidth must be set by the user. The power calculation method used to determine the channel power is a traditional method known as the integration bandwidth (IBW) method. A swept spectrum is used as the basis for this calculation. Therefore, it is important to set the resolution bandwidth correctly before making this
measurement using the following formula:
Span
×=
-------------
n
RBW k
Where k is a value between 1.2 and 4.0 and n is the number of trace points.
Center Frequency
Sets the center frequency of the display.
Key Access: [MEAS] > {Channel Power} > {Center Freq}
Integration Bandwidth
Specifies the bandwidth of integration bandwidth, which can range from 100 Hz to 3 GHz to calculate the power in a channel. For example, set the main (center) channel bandwidth. Note that the green frame indicates the integration bandwidth. Change the integration bandwidth using the knob, the arrow keys, or the numeric keypad. Key Access: [MEAS] > {Channel Power} > {Int BW}
N9340A User’s Guide 45
3 Making Measurements
Channel Bandwidth
Sets the analyzer span for the channel power measurement using the knob, the arrow keys or the numeric keypad. Be sure the span is set between 1 and 10 times the integration bandwidth. Key Access: [MEAS] > {Channel Power}
46 N9340A User’s Guide
Making Measurements 3

Making a Stimulus Response Transmission Measurement

The procedure below describes how to use a built-in tracking generator to measure the rejection of a low pass filter, a type of transmission measurement.
1 To measure the rejection of a low pass filter,
connect the equipment as shown below. A 370 MHz low- pass filter is used as a DUT in this example.

Figure 17 Transmission Measurement Test Setup

DUT
SYS MODE MEAS TRACE
N9340A
HANDHELD SPECTRUM ANALYZER
100 kHz - 3.0 GHz
2
DEF3GHI1ABC
5
MNO4JKL6PQR
VWX7STU9YZ_
8
LIMIT
0
SAVE
ESC/CLR
MARKER
PRESET
ENTER
BW/
FREQ SPANAMPTD
SWP
2 Press [Preset] to perform a factory preset.
3 Set the start and stop frequencies and resolution
bandwidth:
Press [FREQ] > {Start Freq} > 100 > {MHz}
Press [FREQ] > {Stop Freq} > 1 > {GHz}
Press [BW/SWP] > {RBW} > 1 > {MHz}
4 Turn on the tracking generator and if necessary,
set the output power to –10 dBm:
Press [Mode] > {Track Generator} > {Amplitude (On)} >
–10 > {dBm}.
Excessive signal input may damage the DUT. Do not
CAU-CAUTION
exceed the maximum power that the device under test can tolerate.
N9340A User’s Guide 47
3 Making Measurements
5 Press [Sweep] > {Sweep Time (Auto)} to put the
sweep time into stimulus response auto coupled mode.
6 Increase measurement sensitivity and smooth the
noise:
Press [BW/SWP]> {RBW} >30 > {kHz}
Press [BW/SWP] > {VBW} > 30 > {kHz}
A decrease in displayed amplitude is caused by tracking error.
7 Connect the cable from the tracking generator
output to the analyzer input. Store the frequency response in trace 4 and normalize:
Press [MEAS] > {Normalize} > {Store Ref} (1 4) >
{Normalize (On)}
8 Reconnect the DUT to the analyzer and change
the normalized reference position:
Press [MEAS] > {Normalize} > {Norm Ref Posn} > 8 >
[ENTER]
9 Measure the rejection of the low- pass filter:
Press [Marker] > {Normal} > 370 > MHz, {Delta} > 130
> {MHz}
The marker readout displays the rejection of the filter at 130 MHz above the cutoff frequency of the low- pass filter.

Figure 18 Measure the Rejection Range

48 N9340A User’s Guide
Making Measurements 3

Measuring Stop Band Attenuation of a Low-pass Filter

When measuring filter characteristics, it is useful to look at the stimulus response over a wide frequency range. Setting the analyzer x- axis (frequency) to display logarithmically provides this function. The following example uses the tracking generator to measure the stop band attenuation of a 370 MHz low pass filter.
1 Connect the DUT as shown in Figure 16. This
example uses a 370 MHz low pass filter
2 Press [Preset] to perform a factory preset.
3 Set the start and stop frequencies:
Press [FREQ] > {Start Freq} > 100 > {MHz}
Press [FREQ] > {Stop Freq} > 1 > {GHz}
Press [AMPTD] > {Scale Type} > {Log}
4 Press [BW/SWP] > {RBW} > {10}> {kHz} to set the
resolution bandwidth to 10 kHz.
Excessive signal input may damage the DUT. Do not
CAU-CAUTION
exceed the maximum power that the device under test can tolerate.
5 Turn on the tracking generator and if necessary,
set the output power to -10 dBm:
Press [Mode] > {Track Generator} > {Amplitude (On)} >
–10 > {dBm}.
6 Press [Sweep] > {Sweep Time (Auto)} to put the
sweep time into stimulus response auto coupled mode.
Adjust the reference level if necessary to place the signal on screen.
7 Connect the cable (but not the DUT) from the
tracking generator output to the analyzer input. Store the frequency response into trace 4 and normalize:
N9340A User’s Guide 49
3 Making Measurements
Press [MEAS] > {Normalize} > {Store Ref} (1 4) > {Normalize (On)}
8 Reconnect the DUT to the analyzer. Note that the
units of the reference level have changed to dB, indicating that this is now a relative measurement.
9 To change the normalized reference position:
Press [MEAS] > {Normalize} > {Norm Ref Posn} > {8} >
[ENTER]
10Place the reference marker at the specified
cutoff frequency: Press [MARKER] > {Normal} > {370} > MHz
11 Set the second marker as a delta frequency of
37 MHz:
Press {Delta} > 37 >MHz
12In this example, the attenuation over this
frequency range is 19.16 dB/octave (one octave above the cutoff frequency).
13Use the front- panel knob to place the marker at
the highest peak in the stop band to determine the minimum stop band attenuation. In this example, the peak occurs at 600 MHz. The attenuation is 51.94 dB.

Figure 19 Minimum Stop Band Attenuation

50 N9340A User’s Guide
Making Measurements 3

Making a Reflection Calibration Measurement

The following procedure makes a reflection measurement using a coupler or directional bridge to measure the return loss of a filter. This example uses a 370 MHz low- pass filter as the DUT. The calibration standard for reflection measurements is usually a short circuit connected at the reference plane (the point at which the device under test (DUT) is connected.) See Figure
19. A short circuit has a reflection coefficient of 1 (0 dB return loss). It reflects all incident power and provides a convenient 0 dB reference.
1 Connect the DUT to the directional bridge or
coupler as shown below. Terminate the unconnected port of the DUT.
Figure 20 Reflection Measurement Short Calibration Test
Setup
Short Circuit
Or
DUT
Coupled Port
If possible, use a coupler or bridge with the correct test
NOTE
port connector for both calibrating and measuring. Any
SYS MODE MEAS TRACE
N9340A
HANDHELD SPECTRUM ANALYZER
100 kHz - 3.0 GHz
2
DEF3GHI1ABC
5
MNO4JKL6PQR
VWX7STU9YZ_
8
LIMIT
0
SAVE
ESC/CLR
MARKER
PRESET
ENTER
BW/
FREQ SPANAMPTD
SWP
adapter between the test port and DUT degrades coupler/bridge directivity and system source match.
Ideally, you should use the same adapter for the calibration and the measurement. Be sure to terminate the second port of a two port device.
N9340A User’s Guide 51
3 Making Measurements
2 Connect the tracking generator output of the
analyzer to the directional bridge or coupler.
3 Connect the analyzer input to the coupled port
of the directional bridge or coupler.
4 Press [Preset] to perform a factory preset.
5 Turn on the tracking generator and if necessary,
set the output power to –10 dBm:
Press [Mode] > {Track Generator} > {Amplitude (On)} >
–10 > {dBm}
Excessive signal input may damage the DUT. Do not
CAU-CAUTION
exceed the maximum power that the device under test can tolerate.
6 Set the start and stop frequencies and resolution
bandwidth:
Press [FREQ] > {Start Freq} > 100 > {MHz}
Press [FREQ] > {Stop Freq} > 1 > {GHz}
Press [BW/Avg] > {RBW} > 1 > MHz
7 Replace the DUT with a short circuit.
8 Normalize the trace:
Press [MEAS] > {Normalize} > {Store Ref (1 4)}> {Normalize (On)}
This activates the trace 1 minus trace 4 function and display the results in trace 1. The normalized trace or flat line represents 0 dB return loss. Normalization occurs each sweep. Replace the short circuit with the DUT.
Since the reference trace is stored in trace 4, changing
NOTE
trace 4 to Clear Write invalidates the normalization.
52 N9340A User’s Guide

Figure 21 Short Circuit Normalized

Making Measurements 3
N9340A User’s Guide 53
3 Making Measurements

Measuring Return Loss Using the Reflection Calibration Routine

This procedure uses the reflection calibration routine in the proceeding procedure “Making a
Reflection Calibration Measurement" on page 51 to
calculate the return loss of the 370 MHz low pass filter.
1 After calibrating the system with the above
procedure, reconnect the filter in place of the short circuit without changing any analyzer settings.
2 Use the marker to read return loss. Position the
marker with the front- panel knob to read the return loss at that frequency.
Rotate the knob to find the highest peak and the readout is the maximum return loss.
Measuring the Return Loss of the Filter
54 N9340A User’s Guide
Agilent N9340A Handheld Spectrum Analyzer
4 System Setting
Agilent Technologies
55
4System Setting

Visual and Audio Adjustment

Quick Display Adjustment

The analyzer provides you a quick adjustment for both brightness and contrast.
Press [SYS] > {Display} to toggle the screen display status between

Manual Display Adjustment

Adjusting Brightness
Press [SYS] > {Brightness} then rotate the knob to adjust screen brightness.
Adjusting Contrast
Press [SYS] > {Contrast} then rotate the knob to adjust screen contrast.

Setting Key Beep

Activates the key beep function as an indicator of key operation.
Key Access: [SYS] > {More (1 of 3)} > {Key Beep}
Normal and Light.
56 N9340A User’s Guide

System Setting

Includes general system settings, displayed language setting, and external input setting.

General system settings

Provides the following system setting options:
Time/Date
Press [SYS] > {Setting} > {General} > {Time/Date} to set the date and time of the analyzer. The analyzer requires you to input the time in a HHMMSS format, and the date in a YYYYMMDD format.
Power Manager
Press [SYS] > {Setting} > {General} > {Power manager} to select a power saving mode from the followings. The three modes provide a choice for setting the idle time limit as 5 minutes, 15 minutes, 30 minutes or 4 hours.
Backlight
The analyzer turns off the backlight of the
LCD screen after a pre- defined idle time.
Press any key to re-activate the backlight after
the backlight power-saving mode has been
triggered.
LCD
The analyzer turns off the LCD display after a
pre- defined idle time. Press any key to
re- activate the LCD display after the LCD
display power- saving mode has been triggered.
Backlight & LCD
The analyzer turns off the both the LCD
display and the backlight after a pre- defined
idle time. Press any key to re-activate the
backlight and LCD display after the backlight
and LCD power-saving mode has been
triggered.
System Setting 4
N9340A User’s Guide 57
4System Setting

Ext Input

The external Ref and Tri g functions are not available at the
NOTE
same time.
Toggles the channel for external input between Ref and Tr ig . Ref refers to a 10 MHz reference signal; Tri g refers to a TTL signal.
Key Access:
External Reference (Ref)
Use the external reference function as follows:
1 Input a 10 MHz signal to the EXT TRIG IN/REF IN
connector.
2 Press [SYS] > {Setting} > {Ext Input} > {Ref}
the external reference signal input.
The analyzer then turns off its internal reference.
External Trigger (Trig)
When an external TTL signal is used for triggering function, the analyzer uses the inner reference as default.
Use the external trigger function as follows:
1 Press [SYS] > {Setting} > {Ext Input} > {Trig} to enable
the external TTL signal input.
2 Press [SPAN] > {Zero Span} to activate the Tr ig ge r
function.
3 Access the associated softkeys to select the
rising edge (Ext Rise) or the falling edge (Ext Fall) as the trigger threshold
The trace will halt in external trigger mode till the trigger
NOTE
threshold is met or the free run function is activated.
[SYS] > {Setting} > {Ext Input}
to enable
58 N9340A User’s Guide
System Setting 4
Quick saving or loading a system setup
Quick saving a setup
Press [SYS] > {Setting} > {Save Setup} to save the current system settings.
Refer to “File Type" on page 60 for more information about file types.
This quick saving utility of system state is equal to the system
NOTE
state saving utility under the [SYS] > {File}.
Quick recalling a setup
• Load default
Press [SYS] > {Setting} > > {Recall Setup} > {Load default} to recall the factory default setup.
•User Setup
Press [SYS] > {Setting} > {Recall Setup} > {User Setup} to recall a customized setup that has been saved previously.
N9340A User’s Guide 59
4System Setting

File

Pressing [SYS] > {File} accesses to the menu that allows you to manage the file saving and loading.
A USB memory stick of FAT32 or FAT16 format and with only
NOTE
one memory zone is primarily required when using USB memory stick for file saving and loading.
File Type
Pressing [SYS] > {File} > {File Setup} > {File type} allows you to select a file type from one of the following:
• Trace (*.DAT) A trace file records trace data and controls.
• Screen (*.JPG) A screen file records graphic information of the current screen.
State (*.STA) A state file records the current controls and settings of the analyzer.
• Pattern (*.PTN) A pattern file records the limit line settings.
Spectrum Mask (*.MSK) A mask file records the spectrum mask settings. You can use the N9340A PC software to edit a mask file and transfer the mask to your N9340A.
Setup (*.SET) A setup file records the system setting information, such as language, date/time, and power saving mode.
Saving Path
Pressing [SYS] > {File} > {File Setup} > {Save Path} allows you to select a path from one of the following:
Local memory
External USB memory stick
60 N9340A User’s Guide
System Setting 4

Saving a file

After you select a file type, you need to edit a file name by
NOTE
pressing the numeric and alphabetic hardkeys on the right side of the analyzer’s front panel.
Refer to the following three steps to save a file:
1 Press [SYS] > {File} > {File setup} to select the save
path and file type.
2 Edit a file name.
A file name can be consist of letters and digits. A single key stroke on the numerical keypad inputs a digital number; and consecutive key stroke selects and inputs a letter.
3 Press {Save} to save the file.
When a file saving completes, the message “File Saved” displays on the bottom line of the screen.
If you toggle the saving path to USB, files are automatically
NOTE
saved into a folder named as N9340DATA in the root directory of the USB memory stick. If the USB memory stick connection can not be updated, remove the USB memory stick, then cycle the power and re-insert the USB memory stick again.
Quick saving a trace or a screen
Pressing [ESC/CLR] > [Save] allows you to quickly save a trace or a screen to either the local memory or an external USB device, depending upon the setup of the save path. For more information about file setup please refer to “File Type" on page 60 and
“Saving Path" on page 60.
Three steps for saving a trace or a screen:
1 Press [SYS] > {File} > {File Setup} > {Shortcut Type} to
indicate the shortcut type to either a trace or a screen copy.
2 Press [SYS] > {File} > {File Setup} > {Save Path} to either
local memory or an external USB disk.
N9340A User’s Guide 61
4System Setting
3 Press [ESC], [Save] to save a trace or a screen copy to your
pre-defined memory. The trace or screen will be automatically assigned with a default file name. A trace will be assigned with a file name, such as HYTRACE, HYTRACE_1, and consecutive decimal numbers appended. A screen copy will be assigned with a file name, as HYSCREEN, HYSCREEN_1 and consecutive decimal numbers appended to the alphabetic file name.
To quick save or load a setup, please refer to “Quick saving
or loading a system setup" on page 59.
Viewing file list
Refer to the following two steps to view file list:
1 Select the directory you wish to view.
Press [SYS] > {File} > {View} to toggle the displayed file list from local memory or an external USB memory stick.
2 Select the file type you wish to view.
Press [SYS] > {File} > {File setup} > {File type} to select a file type.
62 N9340A User’s Guide

Deleting a file

A deleted file can NEVER be recovered. Carefully
CAU-CAUTION
BE
decide whether you need to delete the file(s).
1 Press [SYS] > {File} > {View} to select a directory.
2 Press [SYS] > {File} > {File setup} > {File type} to
select a file type.
3 Rotate the knob to highlight a file.
4 Press {Delete} to delete a selected file.

Loading a file

A screen file (*.JPG) can not be loaded onto the
NOTE
analyzer.
1 Press [SYS] > {File} > {View} to select a directory.
2 Press [SYS] > {File} > {File setup} > {File type} to
select a file type.
3 Rotate the knob to highlight a file.
4 Press {Load Now} to recall a pre- saved file.
System Setting 4
N9340A User’s Guide 63
4System Setting

Show System

Displays the following hardware, software and battery information of your analyzer:
Machine Model Power Source
MCU Firmware Version Battery Info
DSP Firmware Version Name
FPGA Firmware Version Serial NO.
RF Firmware Version Capacity
RF Module S/N Temperature
KeyBoard Module S/N Charge Cycles
Elapsed Time Voltage
Whole Times Current
This Run Time Charge Status
Temperature Remain Time
Source Voltage Host ID
Key Access: [SYS] > {More (2 of 3)} > {Show system}
64 N9340A User’s Guide

Show Error

Accesses a list of the 30 most recent error messages reported by the analyzer. The most recent error will appear at the bottom of the list. If the error list is longer than 30 entries, the analyzer reports an error message [–350, Query overflow]. When in remote control, and the error list is greater than 30 entries, the error display is halted and a message at the bottom of the list informs the user that error exceed 30.
Key Access: [SYS] > {More (2 of 3)} > {Show Error}
For more information, refer to “Error Messages" on
page 143.
System Setting 4
N9340A User’s Guide 65
4System Setting

Perform a Time Base Calibration

As soon as the calibration function triggers, it interrupts the current measurement is interrupted and a gauge displays on the LCD. The gauge simply indicates calibration action rather than calibration course, as the calibration time is unpredictable. When the calibration is finished, the LCD displays a calibration, and the interrupted measurement restarts.
Time base calibration takes during a short time only when the
NOTE
inner temperature is stable. When the inner temperature is increasing, calibration takes a long-time course or will fail. If the input reference signal is abnormal, the calibration cycle will take a long and unpredictable time to exit, and the LCD displays an error message.
Key Access: [SYS] > {More (2 of 3)} > {Calibration}
66 N9340A User’s Guide

Factory Default Settings

Parameter Default Setting
Center Frequency 1.5 GHz Start Frequency 0.0 Hz Stop Frequency 3.0 GHz Span 3.0 GHz Reference Level 0.0 dBm Attenuation Auto (20 dB) Scale/DIV 10 dB/DIV Scale Type Log RBW Auto (1 MHz) VBW Auto (1 MHz) Average Type Log Power Sweep time Auto Sweep Mode Normal Trace 1 Clear write Trac e 2 B l a n k Trac e 3 B l a n k Trac e 4 B l a n k Trac e 1 Detect ion Pos Peak Trace 2 Detection Pos Peak Trace 3 Detection Pos Peak Trace 4 Detection Pos Peak Trace Average All Off Marker All Off File Type Trace Save Path Local Mode Spectrum Analyzer External Input type Trigger in (TTL signal input)
System Setting 4
For more information about saving a user-defined preset, refer to “Selecting a preset type” and
“Saving a User-defined Preset" on page 22.
N9340A User’s Guide 67
4System Setting
68 N9340A User’s Guide
Agilent N9340A Handheld Spectrum Analyzer
5 Key Reference
Agilent Technologies
69
5Key Reference

Amplitude

Activates the reference level function and access the associated softkeys to set functions that affect the way data on the vertical axis is displayed or corrected.

Ref level

Activates the reference level function.
The reference level is the amplitude power or voltage represented by the top graticule on the display. Changing the value of the reference level changes the absolute amplitude level (in the selected amplitude units) of the top graticule line. To change the reference level, use the arrow keys, the knob, or the numeric keypad. Pressing any digit (0 through 9) on the numeric keypad brings up the terminator menu.
Key access: [AMPTD]

Attenuation

Toggles the input attenuation between Auto or Man.
The input attenuation ranges from 1dB to 51 dB and allows you to set the attenuation level in 1 dB increments when Attenuation is set to Man.
The input attenuator, normally coupled to the reference level control, reduces the power level of the analyzer input signal at the input mixer. The attenuator is recoupled when Attenuation is set to Auto.
To prevent damage to the input mixer, do not exceed a
CAU-CAUTION
power level of +33 dBm at the input. A granted power level for stable sensitive measurement is less than 20 dBm.
Key access: [AMPTD]
70 N9340A User’s Guide

Preamp

Toggles the internal preamp between On and Off.
Preamp results in a correction being applied to compensate for the gain of the preamp at 20 dBm so that amplitude readings show the value at the input connector.
When the preamp is on, a PA indication appears on the left
NOTE
side of the display.
The preamp is set to on in frequency bands from 1 MHz to 3 GHz, otherwise the correction is not applied even though the PA indication still appears.
Key access: [AMPTD]

Scale/DIV

Sets the logarithmic units per vertical graticule division on the display.
The Scale/Div function is only available when the Scale Type
NOTE
key is set to Log.
Key access: [AMPTD]

Scale Type

Key Reference 5
Toggles the scale type of vertical graticule divisions between logarithmic and linear.
When Scale Type is set to Log, a Logarithmic units is activated ranging from 1 to 10 dB per division.
When Scale Type is set to Lin, the vertical divisions are linearly scaled and the default amplitude units are volts. The top of the display is set to the reference- level value and the bottom graticule becomes zero volts. (Each division of the tragically is one- tenth of the reference level in volts.)
Key access: [AMPTD] > {Scale Type}
N9340A User’s Guide 71
5Key Reference

Ref Offset

Adds an offset value which ranges from -327.6 dB to +327.6 dB to the displayed reference level.
Reference-level offsets are only entered by using the numeric
NOTE
keypad. Entering an offset does not affect the trace or the attenuation value.
Reference- level offsets are used when gain or loss occurs between a device under test and the analyzer input. Thus, the signal level measured by the analyzer may be referred to as the level at the input of an external amplitude conversion device. When an amplitude offset is entered, its value appears in the lower left corner of the display.
Key access: [AMPTD]

Y Axis Units

Sets the amplitude units.
Amplitude units are maintained for both logarithmic and linear
NOTE
modes depending on the Scale Type setting.
Key access: [AMPTD] > {Y Axis Units}
72 N9340A User’s Guide

BW/SWP

Activates the bandwidth function and accesses the associated softkeys to control resolution bandwidth, video bandwidth and sweep time.
RBW
Changes the 3- dB resolution bandwidth on the analyzer from 30 Hz to 1 MHz in a 1, 3, 10 sequence using the knob, step keys or the numeric keypad.
If an unavailable bandwidth is entered using the numeric
NOTE
keypad, the closest available bandwidth in the 1, 3, 10 sequence is used.
As the resolution bandwidth is decreased, the sweep time is modified to maintain amplitude calibration. Resolution bandwidth is also related to span. As span is decreased, resolution bandwidth decreases. As the resolution bandwidth changes, the video bandwidth, if in auto couple mode, changes to maintain the VBW/RBW ratio.
In non-zero span mode, the value of RBW is determined by
NOTE
whether the setting for RBW is Auto or Man (manual).
Key Reference 5
A “#”mark appears next to RBW on the display when it is not coupled.
Key access: [BW/SWP]
N9340A User’s Guide 73
5Key Reference
VBW
Changes the analyzer post- detector filter from 3 Hz to 1 MHz in a 1, 3, 10 sequence using the knob, the step keys, or the numeric keypad.
If an unavailable bandwidth entered using the numeric
NOTE
keypad, the closest available bandwidth in the 1, 3, 10 sequence is used.
As the video bandwidth is decreased, the sweep time is increased to maintain amplitude calibration.
A “#”mark appears next to RBW on the display when it is not coupled.
Key access: [BW/SWP]

VBW/RBW

Selects the ratio between the video and resolution bandwidths.
If signal responses near the noise level are visually masked by the noise, the ratio can be set to less than 1 to lower the noise. The knob and step keys change the ratio in a 1, 3, 10 sequence.
When preset is set to default, or VBW/RBW is set to Auto,
NOTE
the ratio is set to 1.
Key access: [BW/SWP]

Avg Type

Accesses the functions to automatically or manually choose from one of the following averaging type:
74 N9340A User’s Guide
Key Reference 5
Log Pwr
Averages the data as appropriate for the logarithmic scaled y axis.
When average type is set to Log Pwr, “LgAv” appears on the left side of the display.
Key access: [BW/SWP] > {Avg Type}
Pwr
Converts the scale from logarithmic to linear power units, and then averages the power trace data.
With average type set to Pwr, “PAvg” appears on the left side of the display.
Key access: [BW/SWP] > {Avg Type}
Voltage
Averages the data as appropriate for the linearly scaled y- axis scale.
With average type set to Voltage, “VAvg” appears on the left side of the display.
Key access: [BW/SWP] > {Avg Type}
There are actually four types of averaging processes within an N9340A. All, except VBW filtering, are affected by this setting. They are listed as follows:
• Trace averaging
Averages signal amplitudes on a trace- to- trace basis. (Press Log Pwr or Pwr)
•Average detector
Averages signal amplitudes during the time or frequency interval represents the result as a single measurement point.
•Noise marker
Averages signal amplitudes across measurement points to reduce variations for noisy signals.
• VBW filtering
Performs a form of averaging of video signal.
N9340A User’s Guide 75
5Key Reference

Sweep Time

Selects the length of time the analyzer takes to tune across the displayed frequency span (or, in zero span, the time the analyzer takes to sweep the full screen) using the knob, the arrow keys, or numeric keypad.
Reducing the sweep time increases the sweep rate.
NOTE
Key access: [BW/SWP]
In non-zero spans:
When the sweep time is auto- coupled, the analyzer selects the optimum (shortest) sweep time ranging from 10 ms to 1000 s for the current settings. This selection process is influenced by three factors:
Frequency Span
The selected Resolution Bandwidth
The selected Video Bandwidth
Note that changing one of the three factors above changes sweep time. The sweep time is in direct ratio to frequency span and is in reverse ratio to the selected resolution bandwidth and video bandwidth when auto- coupled.
With sweep time set to a value less than the minimum
CAU-CAUTION
auto-coupled value, an annotation “Data out of range” appears in the active function block and the sweep time is automatically converted to the minimum auto-coupled value.
In zero span:
The auto- coupled status is invalid in zero span. You can only set the sweep time manually. The sweep time ranges from 6 s to 200 s in zero span.
76 N9340A User’s Guide
µ
Key Reference 5
In FFT Mode
In FFT mode, the sweep time is auto- coupled as default. And the submenu of sweep time is invalid in FFT mode.

Sweep

Activates the sweep mode function and accesses associated softkeys to set functions that affect the way a trace sweeps across the display.
Sweep
Toggles the analyzer between the continuous sweep mode and the single sweep mode.
Sweep (Single)
Puts the analyzer in a single sweep mode. In this mode, pressing [BW/SWP] > {Single Sweep} enables a sweep.
Key access: [BW/SWP] > {Sweep}
Sweep (Cont)
Performs one sweep after another as soon as it is triggered. Pressing [Preset] (When Preset Type is set to Default) or turning the power on, sets the sweep mode to continuous sweep as default.
Key access: [BW/SWP] > {Sweep}
Sweep Mode
An N9340A provides two sweep modes to meet different sweep time requirements:
•Normal
Activates the sweep mode as default.
Key access: [BW/SWP] > {Sweep} > {Sweep Mode}
N9340A User’s Guide 77
5Key Reference
•Fast
Activates the fast sweep mode.
Fast sweep mode provides a fast measurement function that decreases the sweep time. However, this mode brings a decrease of amplitude accuracy.
Fast sweep is valid only when span is no less than 50 MHz.
NOTE
Measurement data is not ganruanteed in fast sweep mode.
Key access: [BW/SWP] > {Sweep} > {Sweep Mode}
Single Sweep
When the analyzer is in continuous sweep mode and not in a measurement mode ([MEAS] > Measure Off), pressing [BW/SWP] > {Single Sweep} convert the continuous sweep to single sweep and executes a sweep after the trigger condition is met. If the analyzer is already in single sweep, pressing [BW/SWP] > {Single Sweep} executes a new sweep as soon as the trigger condition is met.
If Average is on ([TRACE] > {More (1 of 2)} > {Average} > On), pressing [BW/SWP] > {Single Sweep} halts averaging of the current value, and continue averaging from the current count number as soon as a single sweep function triggers.
Key access: [BW/SWP] > {Sweep}
78 N9340A User’s Guide

Enter

Key Reference 5
Terminates date entry and enters into the analyzer a numerical value keyed from numeric keypad on the front panel.
Terminate filename entries.
N9340A User’s Guide 79
5Key Reference

ESC/CLR

Provides mainly two types of functions:
Clear
Clears any numeric entry and cancels the active function.
Clears any title entry and cause the title to revert to the previous name.
Clears input or output overloads
Clears error messages from the status line along
the bottom of the display.
Association with other functions
Provides a permitted setting for triggering other functions as follows:
Activates LIMIT function for pass/fail judgements.
Activates SAVE function for current trace saving.
When pressing [ESC/CLR], a current parameter
CAU-CAUTION
annotation in the lower left corner disappears which is an indicator of available settings of LIMT and SAVE.
80 N9340A User’s Guide

Frequency

Activates the center frequency function, and accesses the menu of frequency functions. The center frequency, or start and stop frequency values appear below the graticule on the display.
When changing both the center frequency and the span,
NOTE
change the frequency first since the span can be limited by the frequency value.

Center Frequency

Activates the center frequency function which allows you to set the horizontal center of the display to a specified frequency.
Key access: [FREQ]

Start Frequency

Sets the frequency at the left side of the graticule. The left and right sides of the graticule correspond to the start and stop frequencies. When these frequencies are activated, their values are displayed below the graticule in place of center frequency and span.
Key access: [FREQ]
Key Reference 5

Stop Frequency

Sets the frequency at the right side of the graticule.The left and right sides of the graticule correspond to the start and stop frequencies. When these frequencies are activated, their values are displayed below the graticule in place of center frequency and span.
Key access: [FREQ]
N9340A User’s Guide 81
5Key Reference

CF Step

Changes the step size for the center frequency function. Once a step size has been selected and the center frequency function is activated, the arrow keys change the center frequency by the step- size value. The step size function is useful for finding harmonics and sidebands beyond the current frequency span of the analyzer. When auto-coupled, the center size is set to one division (10 percent of the span).
Key access: [FREQ]
82 N9340A User’s Guide

Marker

Marker

Normal

Key Reference 5
Accesses the marker control softkeys to select the type and number of markers. Markers are diamond- shaped characters that identify points of traces. Up to six pairs of markers may appear on the display simultaneously; only one pair can be controlled at a time. The marker that is controlled is called the “active” marker. Pressing [MARKER] activates the Normal menu key.
Selects one of the six possible markers. A marker that has already been selected will become active when it is turned on. If a marker has been turned on and assigned to a specific trace, it will become active on that trace if that marker is selected.
Key access: [MARKER]
Activates a single frequency marker at the center of the active trace if a marker is not already displayed. If a marker is displayed before the Normal function is enabled, the marker is enabled at the position of the selected marker. The marker number is indicated above the marker. The knob and/or arrow keys move the marker left or right. If a value is entered from the numeric keypad, the marker is moved to the trace point nearest to that value. Annotation in the active function block and in the upper right corner indicates the frequency and amplitude of the marker (time and amplitude, if in zero span.) Pressing Normal turns off the Delta function and moves the active marker to the delta marker position.
Key access: [MARKER]
N9340A User’s Guide 83
5Key Reference

Delta

Activates a second marker at the position of the first marker. (If no marker is present, two markers appear at the center of the display). The amplitude and frequency (or time) of the first marker is fixed. The marker number is indicated above the delta marker, and the same number is indicated with an R (for example, 1R) above the reference marker. Use the data controls to position the delta marker. Annotation in the active function block and in the upper- right corner of the display indicates the frequency (or time) and amplitude differences between the two markers. The markers will be turned off if the scale type is changed between log and linear.
Press Delta again moves the reference marker to the active
NOTE
marker position, so you are allowed to make delta measurements from differing reference points without having to turn off the makers and begin again.
The delta marker function permits signal-to-noise
NOTE
measurements provided the signal is a single spectral component (sinusoid). Place a normal marker on the signal, press Delta, place the delta marker in the noise, and activate Marker Noise. The indicated amplitude indicates signal-to-noise/Hz.
Key access: [MARKER]
Off
Turns off the active function markers.
Key access: [MARKER]
84 N9340A User’s Guide
Key Reference 5

Peak Search

Place a marker on the highest peak based on the setting of the Search Criteria as follows:
•Peak
Place a marker on the highest peak.
Key access: [MARKER]
•Next LF Peak
Moves the marker to the next peak to the left of the current marker. The signal peak must exceed the peak threshold value by the peak excursion value. If there is no peak to the right, the marker will not move and the No Peak Found error message will appear on the display.
Key access: [MARKER]
•Next RT Peak
Moves the marker to the next peak to the right of the current marker. The signal peak must exceed the peak threshold value by the peak excursion value. If there is no peak to the right, the marker will not move and the No Peak Found error message will appear on the display.
Key access: [MARKER]
•Auto Search
In non- zero span, pressing Auto Search turns all six markers on and puts them on the six highest peak in the current frequency span.
In zero span, the function is not available.
Key access: [MARKER]
N9340A User’s Guide 85
5Key Reference

Marker To

Accesses the following marker function menu keys:
•To Center
Sets the center frequency of the analyzer to the marker frequency. In Delta mode, pressing [Marker] > {Marker To} > {To Center} sets the center frequency to the marker delta value. The function is not available in zero span.
When the frequency scale is in log mode, the center
NOTE
frequency is not at the center of the display.
Key access: [MARKER] > {To Center}
•To Ref
Changes the reference level to the active marker value, moving the marked point to the reference level (top line of the graticule). In Delta mode, pressing [Marker] > {Marker To} > {To Ref} sets the reference level to the amplitude difference between the markers.
Key access: [MARKER] > {Marker To}

Mode

Accesses the following menu keys that allows you to change the active marker readout mode.
•Normal Activates the default readout mode.
Freq Count
Activates a frequency counter mode.
Frequency counter is properly effective only on CW signals or
NOTE
discrete peaks. For a valid reading, the marker must be greater than 25 dB above the noise. Freq Count function is not supported under Marker Delta condition.
86 N9340A User’s Guide
•Noise
Activates a noise readout mode for evaluating power density.
Noise mode operates only on noise peaks and not supported
NOTE
under Marker Delta condition.
Key access: [MARKER] > {More (1 of 2)} > {Mode}

Marker Trace

Activates a marker on the trace if there are no markers turned on. If a marker is currently active, press Marker Trace until 1, 2, 3, or 4 is underlined. The active marker will be moved to the selected trace.
Selecting automatically selected trace. The selection order is to look for the lowest numbered trace in the following order of trace modes:
Clear write Max hold
Min. hold View
If there is no trace in any of these modes, it selects trace 1 as default.
Key access: [MARKER] > {Marker Trace}
Auto will move the marker to the
Key Reference 5

All Off

Turns off all of the markers, including markers used for signal track and demodulation. Marker annotation is also removed.
Key access: [MARKER] > {All Off}
N9340A User’s Guide 87
5Key Reference

Meas

In spectrum analysis mode, pressing [Meas] brings up the menu for a suite of power measurements such as adjacent channel power, occupied bandwidth, and channel power.
Refer to “One-button Power Measurement" on page 43 for more information about these measurements.
88 N9340A User’s Guide

Span

Key Reference 5
Activates the span function and accesses the submenu of span functions. Pressing [SPAN] allows you to change the frequency range symmetrically about the center frequency. The frequency- span readout describes the total displayed frequency range. To determine frequency span per horizontal graticule division (when the frequency scale type is set to linear), divide the frequency span by 10.

Span

Allows you to enter a frequency span range value.
Key access: [SPAN]

Full

Changes the span to full span showing the full frequency range of the analyzer.
Key access: [SPAN]

Zero

Changes the frequency span to zero. In this mode, the current center frequency is displayed in the time domain (the x axis is displayed in units of time), like a conventional oscilloscope.
Key access: [SPAN]

Last Span

Changes the span to the previous span setting.
Key access: [SPAN]
N9340A User’s Guide 89
5Key Reference

TRACE

Accesses the trace keys to store and manipulate trace information. Each trace comprises a series of data points in which amplitude information is stored. The analyzer updates the information for any active trace with each sweep.
If you have selected Channel Power, OBW, or ACPR in the MEAS menu, refer to “One- button Power
Measurement" on page 43.

Tr ac e

Selects the trace for current use.
Key access: [TRACE]

Clear Write

Erases any data previously stored in the selected trace and continuously displays any signals during the sweep of the analyzer. This function is activated for trace 1 at power on and factory preset.
Key access: [TRACE]

Max Hold

Maintains the maximum level for each trace point of the selected trace, and updates each trace point if a new maximum level is detected in successive sweeps.
Changing the vertical scale (Amplitude, Scale Type, Log or
NOTE
Line), or pressing PRESET, or turning averaging on (TRACE, Average (On)) or switching widow in Zone Span restarts the
held trace.
Key access: [TRACE]
90 N9340A User’s Guide
Loading...