in any form or by any means (including
electronic storage and retrieval or
translation into a foreign language)
without prior agreement and written
consent from Agilent Technologies, Inc. as
governed by United States and
international copyright laws.
Manual Part Number
N9340-90001
Edition
Second edition, April 2007
Printed in China
Agilent Technologies, Inc.
Qianfeng Hi-Tech Industry Park
Chengdu Hi-Tech Industrial Development
Zone (West District) Chengdu 611731,
P.R .Ch ina
Software Revision
This guide is valid for A.01.00 revisions of
the Agilent N9340A Handheld Spectrum
Analyzer software.
CAUTION
A CAUTION notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not proceed beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
WARNI NG
A WARNING notice denotes a hazard.
It calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
Warranty
The material contained in this document
is provided “as is,” and is subject to
being changed, without notice, in future
editions. Further, to the maximum extent
permitted by applicable law, Agilent
disclaims all warranties, either express
or implied, with regard to this manual
and any information contained herein,
including but not limited to the implied
warranties of merchantability and fitness
for a particular purpose. Agilent shall not
be liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or performance
of this document or of any information
contained herein. Should Agilent and the
user have a separate written agreement
with warranty terms covering the
material in this document that conflict
with these terms, the warranty terms in
the separate agreement shall control.
Technology Licenses
The hardware and/or software described
in this document are furnished under a
license and may be used or copied only in
accordance with the terms of such
license.
Restricted Rights Legend
If software is for use in the performance of
a U.S. Government prime contract or
subcontract, Software is delivered and
licensed as “Commercial computer
software” as defined in DFAR
252.227-7014 (June 1995), or as a
“commercial item” as defined in FAR
2.101(a) or as “Restricted computer
software” as defined in FAR 52.227-19
(June 1987) or any equivalent agency
regulation or contract clause. Use,
duplication or disclosure of Software is
subject to Agilent Technologies’ standard
commercial license terms, and non-DOD
Departments and Agencies of the U.S.
Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S.
Government users will receive no greater
than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
in any form or by any means (including
electronic storage and retrieval or
translation into a foreign language)
without prior agreement and written
consent from Agilent Technologies, Inc. as
governed by United States and
international copyright laws.
Manual Part Number
N9340-90001
Edition
Second edition, April 2007
Printed in China
Agilent Technologies, Inc.
Qianfeng Hi-Tech Industry Park
Chengdu Hi-Tech Industrial Development
Zone (West District) Chengdu 611731,
P.R .Ch ina
Warranty
The material contained in this document
is provided “as is,” and is subject to
being changed, without notice, in future
editions. Further, to the maximum extent
permitted by applicable law, Agilent
disclaims all warranties, either express
or implied, with regard to this manual
and any information contained herein,
including but not limited to the implied
warranties of merchantability and fitness
for a particular purpose. Agilent shall not
be liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or performance
of this document or of any information
contained herein. Should Agilent and the
user have a separate written agreement
with warranty terms covering the
material in this document that conflict
with these terms, the warranty terms in
the separate agreement shall control.
Technology Licenses
The hardware and/or software described
in this document are furnished under a
license and may be used or copied only in
accordance with the terms of such
license.
Restricted Rights Legend
If software is for use in the performance of
a U.S. Government prime contract or
subcontract, Software is delivered and
licensed as “Commercial computer
software” as defined in DFAR
252.227-7014 (June 1995), or as a
“commercial item” as defined in FAR
2.101(a) or as “Restricted computer
software” as defined in FAR 52.227-19
(June 1987) or any equivalent agency
regulation or contract clause. Use,
duplication or disclosure of Software is
subject to Agilent Technologies’ standard
commercial license terms, and non-DOD
Departments and Agencies of the U.S.
Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S.
Government users will receive no greater
than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
A CAUTION notice denotes a hazard. It calls attention to
CAU-CAUTION
an operating procedure, practice, or the like that, if not
correctly performed or adhered to, could result in damage
to the product or loss of important data. Do not proceed
beyond a CAUTION notice until the indicated conditions
are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to
an operating procedure, practice, or the like that, if not
correctly performed or adhered to, could result in
personal injury or death. Do not proceed beyond a
WARNING notice until the indicated conditions are fully
understood and met.
Tab le of Co nt en ts
1N9340A Overview
Introduction 2
Functionality 2
Optimized Usability 3
Front Panel Overview 4
Top Panel Overview 5
Display Annotations 6
Working with Batteries 7
Installing a Battery 7
Viewing Battery Status 7
Charging a Battery 8
[:SENSe]:AVERage Subsection 123
FREQuency Subsection 128
SWEep Subsection 132
SYSTem Subsystem 133
Ext Input 133
Date Query 133
Time Query 133
Error Information Query 133
TRACe Subsystem 134
Tracking Generator Subsystem 135
State 135
Output Amplitude 135
Amplitude Offset 136
Normalization 137
Normalization Reference Level 137
Normalization Reference Position 138
TRIGer Subsystem 139
External Trigger Slope 139
5
Trigger Source 139
UNIT Subsystem 141
7Error Messages
Error Message List 144
8Menu Map
AMPTD 150
BW/SWP 151
FREQ 152
MARKER 153
MEAS 154
SPAN 155
SYS 156
TRACE 157
Limit 158
6
Agilent N9340A
Handheld Spectrum Analyzer
1
N9340A Overview
Agilent Technologies
1
1N9340A Overview
Introduction
An Agilent N9340A is a handheld spectrum
analyzer with a frequency range from 100 kHz to 3
GHz, tunable to 9 kHz. The analyzer provides
ultimate measurement flexibility in a package that
is ruggedized for field environments and
convenient for mobile applications.
Functionality
An Agilent N9340A provides your with
comprehensive functionality set and measurement
convenience, including:
1 Pass/Fail judgement
Sets limit lines on the display for a quick
Pass/Fail judgement.
2 Four- trace detection display
Displays as many as four independent traces
simultaneously, being detected in any of the
following modes: POS, NEG, Sample or Normal.
3 Marker frequency readout
Provides fine resolution down to a low of 1 Hz
with a built- in frequency counter.
2N9340A User’s Guide
N9340A Overview1
4 Power Suite
Provides one-button measurement functionality
on OBW
ACPR
(Occupied Bandwidth), channel power, and
(Adjacent Channel Power Ratio).
5 High- sensitivity measurement
Includes a 3 GHz pre- amplifier (Option
N9340A- PA3) enabling highly sensitive
measurement which helps in the analysis of
low-level signals.
Optimized Usability
An Agilent N9340A handheld spectrum analyzer
provides enhanced usability:
• 7.2-inch sunlight viewable screen (
640 480×
pixels)enables you to read the scans easily and
clearly in sun light.
• Arc- shaped handle and rugged rubber casing
ensure a comfortable and firm hold and
ruggedness.
• One- button measurement function provides an
easy, fast access to power measurement.
• File saving or loading via USB
interface.
• Instant screen control for power saving: pressing
and holding [Preset] hardkey for over two
seconds turns off the screen. Pressing any key
on the front panel activates the screen.
N9340A User’s Guide3
1N9340A Overview
Front Panel Overview
14
N9340A
HANDHELD SPECTRUM ANALYZER
SYS MODE MEAS TRACE
100 kHz - 3.0 GHz
BW/
SWP
12
13
11
2DEF 3GHI1ABC
6
5MNO4JKL
PQR
VWX7STU9YZ_
8
LIMIT
0SAVE
ESC/CLR
MARKER
PRESET
ENTER
FREQ SPANAMPTD
10
9
8
7
6
23
1
45
CaptionFunction
1 Power SwitchToggles the analyzer between on and off
2 SYSBrings up a submenu for system setup
3 Function keysIncludes hardkeys: FREQ, SPAN, AMPTD, BW/SWP, TRACE,
4 Preset Returns the analyzer to a known state
5 EnterConfirms a parameter selection or configuration
6 MARKERActivates marker function
7 ESC/CLRExits and closes the dialog box or clears the character input
8 Save Quick saves the current trace
9 LIMIT Sets limit lines for quick Pass/Fail judgement
10 Arrow keysIncreases or decreases a parameter step by step
11 Knob Selects an option item or edits a numerical parameter
12 SoftkeysIndicates current menu functions
13 SpeakerActive when in demodulation mode
14 ScreenDisplays measured traces and status
MEAS, and MODE
4N9340A User’s Guide
Top Panel Overview
9
50VDC MAX
dBm (2W) MAX
33
RF INPUT 50RF OUT 5 0
12-18 VDC
VDC
12-18
80W MAX
55W MAX
1
2
1. DC Connector
Provides input for the DC power source via an
AC-DC adapter. (Make sure that the line-power
source outlet has a protective ground contact.)
2. Charging indicator
Lights when the battery is charging.
3. External power indicator
Lights when an external DC power applied to
the analyzer via the AD- DC adapter.
4. USB interface (for PC connection)
Provides a connection between the analyzer and
a PC for data transfer of further process.
5. USB interface
Provides an connection between the analyzer
and an external USB memory disk.
6. LAN Interface (For future expansion)
7. RF OUT Connector (For future expansion)
8. EXT TRIG IN/REF IN (BNC, Female)
Connects to an external TTL signal or a 10 MHz
reference signal. TheTTL signal is used to
trigger the analyzer’s internal sweep.
9. RF IN Connector (Type N, Female)
Accepts an external input with a frequency
range from 100 kHz to 3 GHz, tunable to 9 kHz.
7 Display Status LineDisplays analyzer status and error messages.
8Video Bandwidth[BW/SWP] > {VBW}
9Frequency Span[SPAN]
10 Sweep Time [BW/SWP] > {Sweep Time}
11 Key MenuSee key label description in the key reference for
12 Key Menu TitleRefers to the current activated menu function
13 Marker Frequency[Marker]
14 Marker Amplitude[Marker]
more information.
6N9340A User’s Guide
N9340A Overview1
Working with Batteries
Use the original battery supplied with the
instrument or a genuine equivalent replacement
from Agilent.
Installing a Battery
Insert a battery, ensuring correct orientation with the
CAU-CAUTION
electrodes correctly aligned.
StepNotes
1 Open the battery cover Use a cross-head type screwdriver to
2 Insert the battery
3 Close the battery cover Push the cover closed, then refasten
remove the screw, then pull the cover
open.
the screw.
Viewing Battery Status
Determine the battery status by:
• Checking the battery icon, which indicates the
approximate level of charge. This is in the
lower- right corner of the screen.
• Press[SYS] > {More (1 of 3)} > {Show System} > {Page
down} to check the current battery information.
N9340A User’s Guide7
1N9340A Overview
Charging a Battery
You can only charge the battery when it is in the
analyzer either in operating or when the analyzer
is shut down.
first using the analyzer. This typically takes
about four hours.
Never use a modified or damagedAC-DC adapter.
CAU-CAUTION
1 Install the battery in the analyzer.
2 Simply attach the AC-DC adapter and switch on
external power.
3 The charge indicator lights, indicating that the
battery is charging. When the battery is fully
charged, the green charging indicator turns off.
During charging and discharging, the battery will
monitor and report its voltage, current, and
temperature. If any of these monitored conditions
exceed their safety limits, the battery will
terminate any further charge or discharge until the
error condition is corrected.
Fully charge the battery before
8N9340A User’s Guide
Agilent N9340A
Handheld Spectrum Analyzer
2
Getting Started
Conventions
1. A pair of curly brackets { } indicates a softkey, for example
{Start} refers to the Start softkey.
2. A pair of square brackets [ ] indicates a hardkey,
for example, [ENTER] refers to the ENTER hardkey.
3. An analyzer refers to Agilent N9340A handheld spectrum
analyzer in the following chapters.
4. A DUT refers to a device under test.
9
Agilent Technologies
9
2Getting Started
Checking Shipment and Order List
We recommend you check the shipment and order
list according to the following procedure when you
receive the shipment.
• Inspect the shipping container for damages.
Signs of damage may include a dented or torn
shipping container or cushioning material that
indicate signs of unusual stress or compacting.
• Carefully remove the contents from the shipping
container, and check your shipment.
• For any question or problem, contact Agilent
Technologies Customer Contact Center (CCC) for
consultant and service. See “Contact Agilent
Technologies" on page 24.
10N9340A User’s Guide
Power Requirements
The AC power supply must meet the following
requirements
Voltage :100 VAC to 240 VAC
Frequency:50 Hz to 60 Hz
Power:Maximum 62 W
The analyzer is equipped with a three- wire power
cord, in accordance with international safety
standards. This cable grounds the analyzer cabinet
when connected to an appropriate power line
outlet. The cable appropriate to the original
shipping location is included with the analyzer.
Various AC power cables are available that are
unique to specific geographic areas. You can order
additional AC power cables for use in different
areas. The table AC Power Cords lists the available
AC power cables, the plug configurations, and
identifies the geographic area in which each cable
is appropriate.
The detachable power cord is the product
disconnecting device. It disconnects the mains
circuits from the mains supply before other parts
of the product. The front switch is only a standby
switch and do not disconnect instrument from
LINE power.
:
Getting Started2
N9340A User’s Guide11
A
2Getting Started
AC Power Cords
Plug TypeCable Part
Number
8121-1703BS 1363/AOption 900
250V 10A
8120-0696AS 3112:2000 Option 901
250V 10A
8120-1692IEC 83 C4Option 902
250V 16A
8120-1521CNS 10917-2
125V 10A
8120-2296SEV 1011Option 906
250V 10
8120-4600SABS 164-1 Option 917
230V 15A
8120-4754JIS C8303Option 918
125V 15A
8120-5181SI 32Option 919
250V 16A
8120-8377GB 1002Option 922
a
Plug
Description
/NEMA 5-15P
For use in
Country & Region
United Kingdom, Hong
Kong, Singapore, Malaysia
Australia, New Zealand
Continental Europe, Korea,
Indonesia, Italy, Russia
Option 903
Unite States, Canada,
Taiwan, Mexi c o
Switzerland
South Africa, India
Japan
Israel
China
250V 10A
12N9340A User’s Guide
Safety Considerations
Agilent has designed and tested the N9340A
handheld spectrum analyzer for Measurement,
Control and Laboratory Use in accordance with
Safety Requirements IEC 61010- 1: 2001,
UL 61010- 1 (2004), and CSA C22.2 No.61010- 1- 04.
The tester is supplied in a safe condition. The
N9340A is also designed for use in Installation
Category II and Pollution Degree 2 per IEC 61010
and IEC 60664 respectively.
Read the following safety notices carefully before
you start to use a N9340A handheld spectrum
analyzer to ensure safe operation and to maintain
the product in a safe condition.
Getting Started2
WARNING
Personal injury may result if the analyzer’s cover is
removed. There are no operator-serviceable parts inside.
Always contact Agilent qualified personnel for service.
Disconnect the product from all voltage sources while it
is being opened.
WARNING
This product is a Safety Class I analyzer. The main plug
should be inserted in a power socket outlet only if
provided with a protective earth contact. Any
interruption of the protective conductor inside or outside
of the product is likely to make the product dangerous.
Intentional interruption is prohibited.
WARNING
WARNING
Electrical shock may result when cleaning the analyzer
with the power supply connected. Do not attempt to
clean internally.
case only.
Use a dry soft cloth to clean the outside
Always use the three-pin AC power cord supplied with
this product. Failure to ensure adequate earth grounding
by not using this cord may cause personal injury and
product damage.
N9340A User’s Guide13
2Getting Started
WARNING
WARNING
CAU-CAUTION
CAU-CAUTION
CAU-CAUTION
NOTE
Danger of explosion if the battery is incorrectly replaced.
Replace only with the same type recommended.
Do NOT dispose of batteries in a fire.
Do NOT place batteries in the trash. Batteries must be
recycled or disposed of properly.
Recharge the battery only in the analyzer. If left unused, a
fully charged battery will discharge itself over time.
Never use a damaged or worn-out adapter or battery.
Charging the batteries internally, even while the analyzer
is powered off, the analyzer may become warm. To avoid
overheating, always disconnect the analyzer from the AC
adapter before storing the analyzer into the soft carrying
case.
We strongly recommend you to connect the automotive
adapter to the power output connector for IT equipment,
when charging the battery on your automotives.
The VxWorks operating system requires full conformity to
USB 1.1 or USB 2.0 standards from a USB disk. Not all the
USB disk are built that way. If you have problems
connecting a particular USB disk, please reboot the
analyzer before inserting the next.
Temperature extremes will affect the ability of the battery
to charge. Allow the battery to cool down or warm up as
necessary before use or charging.
Storing a battery in extreme hot or cold places will reduce the
capacity and lifetime of a battery. Battery storage is
recommended at a temperature of less than 45
o
C.
The LAN port is reserved for future expansion. It is
NOTE
inactive at this version.
14N9340A User’s Guide
Getting Started2
°°°
Environmental Requirements
A N9340A is designed for use under the following
conditions:
• Operating temperature:
–10 C to +50 C (with battery)
0 C to +40 C (with adapter)
• Storage temperature: –40 C to +70 C
• Humidity: 95% or less
• Altitude: 9200 m
°
°°
Electrical Requirements
This analyzer allows you to use either the original
lithium- ion battery pack or the external AC- DC
adapter shipped with the analyzer for power
supply to the analyzer.
Electrostatic Discharge (ESD) Precautions
This analyzer was constructed in an ESD protected
environment. This is because most of the
semiconductor devices used in this analyzer are
susceptible to damage by static discharge.
Depending on the magnitude of the charge, device
substrates can be punctured of destroyed by
contact or mere proximity of a static charge. The
result can cause degradation of device
performance, early failure, or immediate
destruction.
These charges are generated in numerous ways,
such as simple contact, separation of materials,
and normal motions of persons working with static
sensitive devices.
When handling or servicing equipment containing
static sensitive devices, adequate precautions must
be taken to prevent device damage or destruction.
Only those who are thoroughly familiar with
N9340A User’s Guide15
2Getting Started
industry accepted techniques for handling static
sensitive devices should attempt to service circuitry
with these devices.
Instrument Markings
The CE mark shows that the product
complies with all relevant European
Legal Directives
(If accompanied by a year, it signifies
when the design was proven).
The CSA mark is a registered trademark
of the Canadian Standards Association.
N10149
ISM1-A
The C- Tick mark is a registered
trademark of the Australian Spectrum
Management Agency.
This symbol is an Industrial Scientific
and Medical Group 1 Class A product
(CISPR 11, Clause 4)
The instruction manual symbol:
indicates that the user must refer to
specific instructions in the manual.
The standby symbol is used to mark a
position of the instrument power switch.
16N9340A User’s Guide
Getting Started2
Powering on the Analyzer for the First Time
A N9340A is fitted with a transflective screen,
which is viewable under all lighting conditions.
In bright lighting conditions, the display is brighter
and easier to read when you allow light to fall
directly on the screen.
Use original standard adapter for AC-DC conversion.
CAU-CAUTION
The maximum RF input level of an average continuous
power is 33 dBm (or +
connecting a signal into the analyzer that exceeds the
maximum level.
Power on your N9340A
Insert the battery into the analyzer or connect the
analyzer to an external line power supply via the
AC-DC adapter, then press the power switch on
the front panel of your N9340A to power on the
analyzer.
Allow the analyzer to warm- up for 30 minutes
before making a calibrated measurement. To meet
its specifications, the analyzer must meet operating
temperature conditions.
50 VDC signal input). Avoid
Install
battery
Press Power
Switch
N9340A User’s Guide17
Use
stick stand
2Getting Started
Preparation for Use
Use [SYS] hardkey to check or set the system
settings of your analyzer.
Setting up your N9340A
1 Press [SYS] > {Setting} > {General} to set time and
power saving mode:
• Press {Time/Date} to set the time and date for
your analyzer.
• Press {Power Manager} to select a power- saving
mode from the follow three modes: turning off
backlight, turning off screen display, and
turning off both backlight and screen display.
2 Press [SYS] > {Setting} > {Language} to select a
language displayed on screen.
3 To manually adjust the display brightness and
contrast.
• Press [SYS] > {Brightness} and then rotate the
knob to adjust display brightness.
• Press [SYS] > {Contrast} and then rotate the
knob to adjust display contrast.
4 Press [SYS] > {More(2 of 3)} > {Key Beep} to toggle
1 Press [SYS] > {More (1 of 3)} > {Show System} to
display the system information.
2 Press [SYS] > {More (2 of 3)} > {Option} to display
the option information.
3 Press [SYS] > {More (1 of 3)} > {Show Error} to
display the error information.
N9340A User’s Guide19
2Getting Started
Making a Basic Measurement
This section provides information on basic analyzer
operations with the assumption that you
understand the front and top layout, and display
annotations of your analyzer. If you do not, please
refer to “Front Panel Overview" on page 4 and
“Top Panel Overview" on page 5.
For more information on making measurements,
please refer to“Making Measurements" on
page 25”.
Entering Data
When setting measurement parameters, there are
several ways to enter or modify the value of an
active function:
1 Using Knob
Increases or decreases the current value.
2 Using Arrow Keys
Increases or decreases the current value by step.
Press [FREQ] > {CF Step} to set the frequency by an
auto- coupled step (Step = Span/10, when
{CF Step} mode is set to Auto).
3 Using Numeric Keypad
Enters a specific value. Then press a terminator
(either a specified unit softkey or [ENTER]) to
confirm input.
5 Press [FREQ] > {Center Freq} > 1> {GHz} to set the
center frequency to 1 GHz.
6 Press [SPAN] > 5 > {MHz} to set the frequency
span to 5 MHz.
7 Press [MARKER] > {Peak Search} > {Peak} to place a
marker (labeled 1) at the highest peak (1 GHz)
on the display.
The values of amplitude and frequency of the
Marker appear both in the function block and in
the upper- right corner of the screen.
Use the knob, the arrow keys, or the softkeys in
the Peak Search menu to move the marker and
read out the value of both frequency and
amplitude displayed on the screen.
Figure 1 View a signal (1 GHz, 0 dBm)
N9340A User’s Guide21
2Getting Started
Some helpful tips
Performing a time-base calibration
A N9340A provides a manual calibration function
to calibrate the time base.
Before calibration, allow the analyzer to warm up
for about 30 minutes.
Use a BNC cable to connect a 10 MHz reference
signal to the EXT TRIG IN connector of your N9340A,
then press [SYS] > {More (2 of 3)} > {Calibration} > {Time
Base} to initiate a calibration.
Selecting a preset type
The analyzer has two types of preset setting for
you to choose from:
• Default
Restores the analyzer to its factory- defined
setting.
• User
Restores the analyzer to a user- defined setting.
Powering on the instrument always brings up a
NOTE
factory-defined setting no matter what the Preset setting.
Saving a User-defined Preset
If you constantly use system setting that is not the
factory default, please refer to the following steps
to create a user- defined system setting:
to save the current parameters as the user
preset setting.
3 Press [SYS] > {More (1 of 3)} >{Preset} (User) to set
the preset mode to user defined system setting.
22N9340A User’s Guide
Upgrading Firmware
When updating firmware, make sure there will be a
CAU-CAUTION
constant power supply for at least 10 minutes. Power cut
off during the updating process can cause damage to the
instrument.
A N9340A provides an easy and fast access for
firmware update.
1 Set up a folder named N9340DATA in the root
directory of your USB memory stick.
2 Store the firmware update application into the
folder named N9340DATA.
3 Insert the USB memory stick into the USB
connector on the top panel of your analyzer.
4 Press [SYS] > {More (2 of 3)} > {Update} >
{Firmware} to activate the firmware updating
procedure. Then the instrument will then
perform the update automatically.
Adding an option
Pressing [SYS] > {More (2 of 3)} > {Option} > {Add}
brings up a dialog box for entering the option
license code. Use the numeric keypad to input the
option license code and use the [Enter] key as a
terminator. If the analyzer recognises the option
license code, an message “Option install
successfully” appears in the status line, otherwise a
message “Invalid option licence” appears in the
status line.
Getting Started2
N9340A User’s Guide23
2Getting Started
Contact Agilent Technologies
Agilent has Sales and Service Offices around the
world to provide you with complete support.
Go to http://www.agilent.com/find/assist, for help
with:
• product selection, configuration, and purchases.
• technical and application assistance, and
consulting.
• rental and leasing options, and refurbished
equipment.
• repair, calibration, education and training.
If you do not have access to the internet, call the
appropriate number shown below. Or contact your
local Agilent Technologies Sales and Service Office.
This section provides information on measuring
multiple signals.
Comparing Signals on the Same Screen
An N9340A provides an easy function for you to
compare frequency and amplitude differences
between signals, such as radio or television signal
spectra. Using Delta Marker function allows you to
compare two signals when both appear on the
screen at the same time.
In this following example, a 50 MHz input signal is
used to measure frequency and amplitude
differences between two signals on the same
screen. Delta marker is used to demonstrate this
comparison.
1 Press [PRESET] to set the analyzer to a factory
default setting.
2 Input a signal (0 dB, 50 MHz) to the RF IN
connector of the analyzer.
3 Set the analyzer start frequency, stop frequency
and reference level to view the 50 MHz signal
and its harmonics up to 100 MHz:
• Press [FREQ] > {Start Freq} > 40 > {MHz}
• Press [FREQ] > {Stop Freq} >110 > {MHz}
• Press [AMPTD] > {Ref Level} > 0 > {dBm}
4 Press [MARKER] > {Peak search} to place a marker
on the highest peak on the display (50 MHz).
The {Next Left PK} and {Next Right PK} softkeys are
available to move the marker from peak to peak.
5 Press [MARKER] > {Delta} to anchor the first
marker (labeled as 1) and activate a delta
marker:
The label on the first marker now reads 1R,
indicating that it is the reference point.
26N9340A User’s Guide
Making Measurements3
6 Move the second marker to another signal peak
using the front panel knob or by using the {Peak Search} softkey:
• Press [MARKER] > {Peak Search} > {Next Right PK} or
{Next Left PK}.
To increase the resolution of the marker readings, turn on
NOTE
the frequency count function. For more information, please
refer to “Improving Frequency Resolution and
Accuracy" on page 36
Figure 2 Delta pair marker with signals on the same
screen
N9340A User’s Guide27
3Making Measurements
Resolving Signals of Equal Amplitude
In this example a decrease in resolution bandwidth
is used in combination with a decrease in video
bandwidth to resolve two signals of equal
amplitude with a frequency separation of 100 kHz.
Notice that the final RBW selection to resolve the
signals is the same width as the signal separation
while the VBW is slightly narrower than the RBW.
1 Connect two sources to the analyzer input as
shown below.
Figure 3
Setup for obtaining two signals
N9310A
RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby
On
Signal generator
N9310A
RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby
On
FUNCTIONS
Utility
Preset
AM I/Q
Frequency
Amplitude FM
Local
File
Sweep
LF Out
TriggerPulseM
Enter
MOD
9
7
8
On/Off
6
4
5
3
102
RF
·
On/Off
LF OUT RF O UT 50
REVERSE PWR
4W MAX 30VDC
FUNCTIONS
Utility
Preset
AM I/Q
Frequency
Amplitude FM
Local
File
Sweep
Trigger
PulseM
LF Out
Enter
MOD
7
8
9
On/Off
4
5
6
1
3
2
RF
·
0
On/Off
LF OUT RF OUT 50
REVERSE PWR
4W MAX 30VDC
Directional
coupler
Signal generator
2 Set one source to 300 MHz. Set the frequency of
the other source to 300.1 MHz. Set both source
amplitudes to –20 dBm.
3 Setup the analyzer to view the signals:
• Press [PRESET].
• Press [FREQ] > {Center Freq} > 300.05 > {MHz}.
• Press [SPAN] > {Span} > 2 > {MHz}.
• Press [BW/SWP] > {RBW} > 30 > {kHz}.
28N9340A User’s Guide
Making Measurements3
Use the knob or the arrow keys to further reduce
the resolution bandwidth and better resolve the
signals.
As you decrease the resolution bandwidth, you
improve resolution of the individual signals and
also increased the sweep timed. For fastest
measurement times, use the widest possible
resolution bandwidth.
Under factory preset conditions, the resolution
bandwidth is coupled to the span.
Figure 4 Resolving signals of equal amplitude
N9340A User’s Guide29
3Making Measurements
Resolving Small Signals
Hidden by Large Signals
This example uses narrow resolution bandwidths to
resolve two input signals with a frequency
separation of 50 kHz and an amplitude difference
of 60 dB.
1 Connect two sources to the analyzer input
connector as shown in Figure 3 on page 28.
2 Set one source to 300 MHz at –10 dBm. Set the
other source to 300.05 MHz at –70 dBm.
3 Set the analyzer as follows:
• Press [PRESET]
• Press [FREQ] > {Center Freq} > 300.05 > {MHz}
• Press [SPAN] > [500] > kHz.
• Press [BW/SWP] > {RBW} > 300 > {Hz}
4 Reduce the resolution bandwidth filter to view
the smaller hidden signal. Place a delta marker
on the smaller signal:
• Press [MARKER] > {Peak Search}
• Press [MARKER] > {Delta}
• Press [MARKER] > {Peak Search} > {Next Right PK}
or {Next Left PK}.
Figure 5 Resolving small signal hidden by a large signal
30N9340A User’s Guide
Measuring a Low-Level Signal
This section provides information on measuring
low-level signals and distinguishing them from
spectrum noise. There are four main useful
techniques as follows to measure a low-level signal.
Reducing Input Attenuation
The ability to measure a low- level signal is limited
by internally generated noise in the spectrum
analyzer.
The input attenuator affects the level of a signal
passing through the analyzer. If a signal is very
close to the noise floor, reducing input attenuation
will bring the signal out of the noise.
1 Preset the analyzer:
2 Input a signal (1 GHz, –80 dBm) to theanalyzer
RF input connector.
3 Set the center frequency, span and reference
level:
• Press [FREQ] > {Center Freq} > 1 > {GHz}
• Press [SPAN] > 5 > {MHz}
• Press [AMPTD] > {Ref Level} > –40 > {dBm}
4 Move the desired peak (in this example, 1 GHz)
to the center of the display:
• Press [MARKER] > {Peak Search}
• Press [MARKER] > {Marker To} > {To Center}
Figure 6 A signal closer to the noise level (Atten:10 dB)
Making Measurements3
N9340A User’s Guide31
3Making Measurements
5 Reduce the span to 1 MHz and if necessary
re- center the peak.
• Press [SPAN] > [1] > {MHz}.
6 Set the attenuation to 20 dB. Note that
increasing the attenuation moves the noise floor
closer to the signal level.
• Press [AMPTD] > {Attenuation} > 20 > {dB}.
Figure 7 A signal closer to the noise level (Atten: 20 dB)
7 Press [AMPTD] > Attenuation > 0 > {dB} to set the
attenuation to 0 dB.
Figure 8 A signal closer to the noise level (Atten: 0 dB)
32N9340A User’s Guide
Making Measurements3
Decreasing the Resolution Bandwidth
Resolution bandwidth settings affect the level of
internal noise without affecting the level of
continuous wave (CW) signals. Decreasing the RBW
by a decade reduces the noise f loor by 10 dB.
1 Refer to “Reducing Input Attenuation" on
page 31, and follow steps 1, 2 and 3.
2 Decrease the resolution bandwidth:
• Press [BW/SWP] > {RBW}, and toggle RBW
setting to Man (manual), then decrease the
resolution bandwidth using the knob, the
arrow keys or the numeric keypad.
The low level signal appears more clearly because
the noise level is reduced.
Figure 9 Decreasing resolution bandwidth
N9340A User’s Guide33
3Making Measurements
Using the Average Detector and
Increased Sweep Time
When the analyzer’s noise masks low- level signals,
changing to the average detector and increasing the
sweep time smoothes the noise and improves the
signal’s visibility. Slower sweeps are required to
average more noise variations.
1 Refer to “Reducing Input Attenuation" on
page 31,and follow steps 1, 2 and 3.
2 Press [TRACE] > {More (1 of 2)} > {Detector} >
{Average} to select the average detector.
3 Press [BW/SWP] > {Sweep Time} to set the sweep
time to 500 ms.
Note how the noise smoothes out, as there is
more time to average the values for each of the
displayed data points.
4 Press [BW/SWP] > {Avg Type} to change the
average type.
Figure 10 Using the average detector
34N9340A User’s Guide
Making Measurements3
Tra c e Ave ra gin g
Averaging is a digital process in which each trace
point is averaged with the previous average for the
same trace point.
Selecting averaging, when the analyzer is auto
coupled, changes the detection mode to sample,
smoothing the displayed noise level.
This is a trace processing function and is not the same as
NOTE
using the average detector (as described on page 34).
1 Refer to the first procedure “Reducing Input
Attenuation" on page 31, and follow steps 1, 2
and 3.
2 Press[TRACE] > {Average} (On) to turn average on.
3 Press50 > [ENTER] to set the average number to
50.
As the averaging routine smoothes the trace, low
level signals become more visible.FM
Figure 11 Trace averaging
N9340A User’s Guide35
3Making Measurements
Improving Frequency Resolution and Accuracy
Using the frequency counter to improve frequency
resolution and accuracy.
Marker count properly functions only on CW signals or
NOTE
discrete spectral components. The marker must be > 40
dB above the displayed noise level.
This section provides information on measuring
and identifying signal distortion.
Identifying Analyzer Generated Distortion
High level input signals may cause analyzer
distortion products that could mask the real
distortion measured on the input signal. Use trace
and the RF attenuator to determine which signals,
if any, are internally generated distortion products.
In this example, use a signal from a signal
generator to determine whether the harmonic
distortion products are generated by the analyzer.
1 Input a signal (200 MHz, –10 dBm) to the
analyzer RF IN connector.
2 Set the analyzer center frequency and span:
• Press [PRESET]. (Factory Preset)
• Press [FREQ] > {Center Freq} > 400 > {MHz}.
• Press [SPAN] > 700 > {MHz}.
The signal produces harmonic distortion products
(spaced 200 MHz from the original 200 MHz signal)
Figure 13 Harmonic distortion
3 Change the center frequency to the value of the
first harmonic:
N9340A User’s Guide37
3Making Measurements
• Press [MARKER] > {Peak Search}.
• Press [MARKER] > {Marker To} > {To Center}.
4 Change the span to 50 MHz and re- center the
signal:
• Press [SPAN] > 50 > {MHz}.
• Press [MARKER] > {Peak Search}.
5 Set the attenuation to 0 dB:
• Press [AMPTD] > {Attenuation} > 0 > {dB}.
• Press [MARKER] > {Marker To} > {To Ref}.
6 To determine whether the harmonic distortion
products are generated by the analyzer, first save
the trace data in trace 2 as follows:
• Press [TRACE] > {Trace (2)}.
• Press [TRACE] > {Clear Write}.
7 Allow trace 2 to update (minimum two sweeps),
then store the data from trace 2 and place a
delta marker on the harmonic of trace 2:
• Press [TRACE] > {View}.
• Press [MARKER] > {Peak Search}.
• Press [Marker] > {Delta}.
Please refer to figure below shows the stored data
in trace 2 and the measured data in trace 1. The
Marker indicator reads the difference in amplitude
between the reference and active markers.
38N9340A User’s Guide
Making Measurements3
Figure 14 Identifying Analyzer Distortion (O dB atten)
8 Press [AMPTD] > {Attenuation} > 10 > {dB} to
increase the RF attenuation to 10 dB.
Figure 15 Identifying Analyzer Distortion (10 dB atten)
The marker amplitude readout comes from two
sources:
• Increased input attenuation causes poorer
signal- to- noise ratio. This causes the marker
to be positive.
• The reduced contribution of the analyzer
circuits to the harmonic measurement causes
the Marker to be negative.
Large marker readout indicates significant
measurement errors. Set the input attenuator to
minimize the absolute value of marker.
N9340A User’s Guide39
3Making Measurements
Third-Order Intermodulation Distortion
Two- tone, third- order intermodulation distortion is
a common test in communication systems. When
two signals are present in a non-linear system,
they may interact and create third- order
intermodulation distortion (TOI) products that are
located close to the original signals. System
components such as amplifiers and mixers
generates these distortion products.
In this example we test a device for third- order
intermodulation using markers. Two sources are
used, one set to 300 MHz and the other to
301 MHz.
1 Connect the equipment as shown in figure below.
N9310A
RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby
On
Signal generator
N9310A
RF Signal Generator 9 kHz - 3.0 GHz
Remote
Standby
On
Signal generator
FUNCTIONS
Frequency
Utility
Preset
AM I / Q
Sweep
Amplitude FM
Local
File
Enter
LF Out
TriggerPul seM
MOD
7
8
9
On/Off
4
5
6
102
3
RF
·
On/Off
LF OUT RF OU T 5 0
Directional
REVERSE PWR
4W MAX 30VDC
coupler
FUNCTIONS
Frequency
Utility
Preset
AM I/Q
Sweep
Amplitude FM
Local
File
Enter
LF Out
TriggerPulseM
MOD
7
8
9
On/Off
4
5
6
3
102
RF
·
On/Off
LF OUT RF OUT 50
REVERSE PWR
4W MAX 30VDC
This combination of signal generators, low pass
filters, and directional coupler (used as a
combiner) results in a two- tone source with very
low intermodulation distortion.
Although the distortion from this setup may be
better than the specified performance of the
analyzer, it is useful for determining the TOI
performance of the source/analyzer combination.
After the performance of the source/analyzer
combination has been verified, the DUT (Device
under test, for example, an amplifier) would be
inserted between the directional coupler output
40N9340A User’s Guide
Making Measurements3
and the analyzer input.
The coupler should have a high degree of isolation
NOTE
between the two input ports so the sources do not
intermodulate.
2 Set one source (signal generator) to 300 MHz
and the other source to 301 MHz, for a
frequency separation of 1 MHz. Set the sources
equal in amplitude as measured by the analyzer
(in this example, they are set to –5 dBm).
3 Set the analyzer center frequency and span:
• Press [PRESET]. (With Preset Type of Default)
• Press [FREQ] > {Center Freq} > 300.5 > {MHz}.
• Press [SPAN] > 5 > {MHz}.
4 Reduce the RBW until the distortion products
are visible:
• Press [BW/SWP] > {RBW}, and reduce the RBW
using the knob, the arrow keys or the numeric
keypad.
5 Move the signal to the reference level:
• Press [MARKER] > {Peak Search}.
• Press [MARKER] > {Marker To} > {To Ref}.
6 Reduce the RBW until the distortion products
are visible:
• Press [BW/SWP] > {RBW}, and reduce the RBW
using the knob, the arrow keys or the numeric
keypad.
7 Activate the second marker and place it on the
peak of the distortion product (beside the test
signal) using the Next Peak:
• Press [MARKER] > {Delta}.
• Press [MARKER] > {Peak Search} > {Next Left
(Right) PK}.
8 Measure the other distortion product:
• Press [MARKER] > {Normal}.
N9340A User’s Guide41
3Making Measurements
• Press [MARKER] > {Peak Search} > {Next Left
(Right) Peak}.
9 Measure the difference between this test signal
and the second distortion product.
• Press [Marker] > {Normal}.
• Press [MARKER] > {Peak Search} > {Next Left
(Right) Peak}.
Figure 16 TOI test screen
42N9340A User’s Guide
Making Measurements3
One-button Power Measurement
N9340A provides one-button measurement
functionality on OBW
power and ACPR
easy- to- use access for your measurement of
interest.
Measuring OBW
Occupied Bandwidth (OBW) integrates the power
of the spectrum in the displayed green frame. The
measurement defaults to 99% of the occupied
bandwidth power.
Occupied bandwidth measurement are made in
either single or continuous sweep mode.
Selecting a Method
Press [MEAS] > {OBW} > {Method} to toggle the
occupied bandwidth measurement mode between
percentage (%)and dBc.
• Setting Percentage (%)
Press {OBW} > {Method} > {Percent} to edit the
percentage of signal power used when determining
the occupied bandwidth. Press {%} to set the
percentage ranging from 10.00% to 99.99%.
•Setting dBc
Press {OBW} > {Method} > {dBc} to specify the power
level used to determine the emission bandwidth as
the number of dB down from the highest signal
peak, within the occupied bandwidth span. This
function is an independent calculation from the
OBW calculation. The x dB Bandwidth result is
also called the emissions bandwidth (EBW). This
sets the green frame
(f1 and f2 are the two cross points of the signal
and the pair of vertical lines of the displayed green
frame) to your specified dB value below the
maximum power, and compute the total power in
the frame. The emission bandwidth is then
calculated ranging from 0.1 dB to 100 dB (the
default value is set to 26 dB).
(Occupied Band Width), channel
(Adjacent Channel Power Ratio) as an
N9340A User’s Guide43
3Making Measurements
Measuring ACPR
The adjacent channel power ratio (ACPR) measures
the power ratio between the main channel power
and the adjacent channel power.
Center Frequency
Sets the center frequency of the main channel
power.
Key Access: [MEAS] > {ACPR} > {Center Freq}
Main Channel
Specifies the range of integration used in
calculating the power in the main channel. Use the
knob, the arrow keys, or the numeric keypad to set
the bandwidth.
Key Access: [MEAS] > {ACPR} > {Main Channel}
Adjacent Channel
Specifies the range of integration used in
calculating the power in the adjacent channel. Use
the knob, the arrow keys, or the numeric keypad to
set the bandwidth.
Key Access: [MEAS] > {ACPR} > {Adj Channel}
Channel Space
Sets the space value between the center frequency
of main channel power and that of the adjacent
channel power.
Key Access: [MEAS] > {ACPR} > {Channel Space}
44N9340A User’s Guide
Making Measurements3
Measuring Channel Power
Measures the power and power spectral density in
the channel bandwidth that you specified. One pair
of vertical lines of the displayed green frame
indicates the edges of the channel bandwidth. The
center frequency, reference level, and channel
bandwidth must be set by the user.
The power calculation method used to determine
the channel power is a traditional method known
as the integration bandwidth (IBW) method. A
swept spectrum is used as the basis for this
calculation. Therefore, it is important to set the
resolution bandwidth correctly before making this
measurement using the following formula:
Span
×=
-------------
n
RBWk
Where k is a value between 1.2 and 4.0 and n is
the number of trace points.
Specifies the bandwidth of integration bandwidth,
which can range from 100 Hz to 3 GHz to calculate
the power in a channel. For example, set the main
(center) channel bandwidth. Note that the green
frame indicates the integration bandwidth. Change
the integration bandwidth using the knob, the
arrow keys, or the numeric keypad.
Key Access: [MEAS] > {Channel Power} > {Int BW}
N9340A User’s Guide45
3Making Measurements
Channel Bandwidth
Sets the analyzer span for the channel power
measurement using the knob, the arrow keys or the
numeric keypad. Be sure the span is set between 1
and 10 times the integration bandwidth.
Key Access: [MEAS] > {Channel Power}
46N9340A User’s Guide
Making Measurements3
Making a Stimulus Response Transmission
Measurement
The procedure below describes how to use a
built-in tracking generator to measure the rejection
of a low pass filter, a type of transmission
measurement.
1 To measure the rejection of a low pass filter,
connect the equipment as shown below.
A 370 MHz low- pass filter is used as a DUT in
this example.
Figure 17 Transmission Measurement Test Setup
DUT
SYS MODE MEAS TRACE
N9340A
HANDHELD SPECTRUM ANALYZER
100 kHz - 3.0 GHz
2
DEF3GHI1ABC
5
MNO4JKL6PQR
VWX7STU9YZ_
8
LIMIT
0
SAVE
ESC/CLR
MARKER
PRESET
ENTER
BW/
FREQ SPANAMPTD
SWP
2 Press[Preset] to perform a factory preset.
3 Set the start and stop frequencies and resolution
bandwidth:
• Press [FREQ] > {Start Freq} > 100 > {MHz}
• Press [FREQ] > {Stop Freq} > 1 > {GHz}
• Press [BW/SWP] > {RBW} > 1 > {MHz}
4 Turn on the tracking generator and if necessary,
The marker readout displays the rejection of the
filter at 130 MHz above the cutoff frequency of
the low- pass filter.
Figure 18 Measure the Rejection Range
48N9340A User’s Guide
Making Measurements3
Measuring Stop Band Attenuation of a
Low-pass Filter
When measuring filter characteristics, it is useful
to look at the stimulus response over a wide
frequency range. Setting the analyzer x- axis
(frequency) to display logarithmically provides this
function. The following example uses the tracking
generator to measure the stop band attenuation of
a 370 MHz low pass filter.
1 Connect the DUT as shown in Figure 16. This
example uses a 370 MHz low pass filter
2 Press[Preset] to perform a factory preset.
3 Set the start and stop frequencies:
• Press [FREQ] > {Start Freq} > 100 > {MHz}
• Press [FREQ] > {Stop Freq} > 1 > {GHz}
• Press [AMPTD] > {Scale Type} > {Log}
4 Press[BW/SWP] > {RBW} > {10}> {kHz}to set the
resolution bandwidth to 10 kHz.
Excessive signal input may damage the DUT. Do not
CAU-CAUTION
exceed the maximum power that the device under test can
tolerate.
5 Turn on the tracking generator and if necessary,
frequency range is 19.16 dB/octave (one octave
above the cutoff frequency).
13Use the front- panel knob to place the marker at
the highest peak in the stop band to determine
the minimum stop band attenuation. In this
example, the peak occurs at 600 MHz. The
attenuation is 51.94 dB.
Figure 19 Minimum Stop Band Attenuation
50N9340A User’s Guide
Making Measurements3
Making a Reflection Calibration Measurement
The following procedure makes a reflection
measurement using a coupler or directional bridge
to measure the return loss of a filter. This example
uses a 370 MHz low- pass filter as the DUT.
The calibration standard for reflection
measurements is usually a short circuit connected
at the reference plane (the point at which the
device under test (DUT) is connected.) See Figure
19. A short circuit has a reflection coefficient of 1
(0 dB return loss). It reflects all incident power
and provides a convenient 0 dB reference.
1 Connect the DUT to the directional bridge or
coupler as shown below. Terminate the
unconnected port of the DUT.
Figure 20 Reflection Measurement Short Calibration Test
Setup
Short
Circuit
Or
DUT
Coupled
Port
If possible, use a coupler or bridge with the correct test
NOTE
port connector for both calibrating and measuring. Any
SYS MODE MEAS TRACE
N9340A
HANDHELD SPECTRUM ANALYZER
100 kHz - 3.0 GHz
2
DEF3GHI1ABC
5
MNO4JKL6PQR
VWX7STU9YZ_
8
LIMIT
0
SAVE
ESC/CLR
MARKER
PRESET
ENTER
BW/
FREQ SPANAMPTD
SWP
adapter between the test port and DUT degrades
coupler/bridge directivity and system source match.
Ideally, you should use the same adapter for the
calibration and the measurement. Be sure to terminate the
second port of a two port device.
N9340A User’s Guide51
3Making Measurements
2 Connect the tracking generator output of the
analyzer to the directional bridge or coupler.
3 Connect the analyzer input to the coupled port
of the directional bridge or coupler.
4 Press[Preset] to perform a factory preset.
5 Turn on the tracking generator and if necessary,
This activates the trace 1 minus trace 4 function
and display the results in trace 1.
The normalized trace or flat line represents 0
dB return loss. Normalization occurs each
sweep. Replace the short circuit with the DUT.
Since the reference trace is stored in trace 4, changing
NOTE
trace 4 to Clear Write invalidates the normalization.
52N9340A User’s Guide
Figure 21 Short Circuit Normalized
Making Measurements3
N9340A User’s Guide53
3Making Measurements
Measuring Return Loss Using the Reflection
Calibration Routine
This procedure uses the reflection calibration
routine in the proceeding procedure “Making a
Reflection Calibration Measurement" on page 51 to
calculate the return loss of the 370 MHz low pass
filter.
1 After calibrating the system with the above
procedure, reconnect the filter in place of the
short circuit without changing any analyzer
settings.
2 Use the marker to read return loss. Position the
marker with the front- panel knob to read the
return loss at that frequency.
Rotate the knob to find the highest peak and
the readout is the maximum return loss.
Measuring the Return Loss of the Filter
54N9340A User’s Guide
Agilent N9340A
Handheld Spectrum Analyzer
4
System Setting
Agilent Technologies
55
4System Setting
Visual and Audio Adjustment
Quick Display Adjustment
The analyzer provides you a quick adjustment for
both brightness and contrast.
Press [SYS] > {Display} to toggle the screen display
status between
Manual Display Adjustment
Adjusting Brightness
Press [SYS] > {Brightness} then rotate the knob to
adjust screen brightness.
Adjusting Contrast
Press [SYS] > {Contrast} then rotate the knob to
adjust screen contrast.
Setting Key Beep
Activates the key beep function as an indicator of
key operation.
Key Access: [SYS] > {More (1 of 3)} > {Key Beep}
Normaland Light.
56N9340A User’s Guide
System Setting
Includes general system settings, displayed
language setting, and external input setting.
General system settings
Provides the following system setting options:
Time/Date
Press [SYS] > {Setting} > {General} > {Time/Date} to set
the date and time of the analyzer.
The analyzer requires you to input the time in a
HHMMSS format, and the date in a YYYYMMDD
format.
Power Manager
Press [SYS] > {Setting} > {General} > {Power manager} to
select a power saving mode from the followings.
The three modes provide a choice for setting the
idle time limit as 5 minutes, 15 minutes, 30
minutes or 4 hours.
• Backlight
The analyzer turns off the backlight of the
LCD screen after a pre- defined idle time.
Press any key to re-activate the backlight after
the backlight power-saving mode has been
triggered.
• LCD
The analyzer turns off the LCD display after a
pre- defined idle time. Press any key to
re- activate the LCD display after the LCD
display power- saving mode has been triggered.
• Backlight & LCD
The analyzer turns off the both the LCD
display and the backlight after a pre- defined
idle time. Press any key to re-activate the
backlight and LCD display after the backlight
and LCD power-saving mode has been
triggered.
System Setting4
N9340A User’s Guide57
4System Setting
Ext Input
The external Ref and Tri g functions are not available at the
NOTE
same time.
Toggles the channel for external input between Ref
and Tr ig . Ref refers to a 10 MHz reference signal; Tri g
refers to a TTL signal.
Key Access:
External Reference (Ref)
Use the external reference function as follows:
1 Input a 10 MHz signal to the EXT TRIG IN/REF IN
connector.
2 Press [SYS] > {Setting} > {Ext Input} > {Ref}
the external reference signal input.
The analyzer then turns off its internal reference.
External Trigger (Trig)
When an external TTL signal is used for triggering
function, the analyzer uses the inner reference as
default.
Press [SYS] > {Setting} > {Recall Setup} > {User Setup}
to recall a customized setup that has been saved
previously.
N9340A User’s Guide59
4System Setting
File
Pressing [SYS] > {File} accesses to the menu that
allows you to manage the file saving and loading.
A USB memory stick of FAT32 or FAT16 format and with only
NOTE
one memory zone is primarily required when using USB
memory stick for file saving and loading.
File Type
Pressing [SYS] > {File} > {File Setup} > {File type} allows
you to select a file type from one of the following:
• Trace (*.DAT)
A trace file records trace data and controls.
• Screen (*.JPG)
A screen file records graphic information of the
current screen.
• State (*.STA)
A state file records the current controls and
settings of the analyzer.
• Pattern (*.PTN)
A pattern file records the limit line settings.
• Spectrum Mask (*.MSK)
A mask file records the spectrum mask settings.
You can use the N9340A PC software to edit a
mask file and transfer the mask to your N9340A.
• Setup (*.SET)
A setup file records the system setting
information, such as language, date/time, and
power saving mode.
Saving Path
Pressing [SYS] > {File} > {File Setup} > {Save Path} allows
you to select a path from one of the following:
• Local memory
• External USB memory stick
60N9340A User’s Guide
System Setting4
Saving a file
After you select a file type, you need to edit a file name by
NOTE
pressing the numeric and alphabetic hardkeys on the right
side of the analyzer’s front panel.
Refer to the following three steps to save a file:
1 Press [SYS] > {File} > {File setup} to select the save
path and file type.
2 Edit a file name.
A file name can be consist of letters and digits.
A single key stroke on the numerical keypad
inputs a digital number; and consecutive key
stroke selects and inputs a letter.
3 Press {Save} to save the file.
When a file saving completes, the message “File
Saved” displays on the bottom line of the screen.
If you toggle the saving path to USB, files are automatically
NOTE
saved into a folder named as N9340DATA in the root directory
of the USB memory stick.
If the USB memory stick connection can not be updated,
remove the USB memory stick, then cycle the power and
re-insert the USB memory stick again.
Quick saving a trace or a screen
Pressing [ESC/CLR] > [Save] allows you to quickly
save a trace or a screen to either the local memory
or an external USB device, depending upon the
setup of the save path. For more information about
file setup please refer to “File Type" on page 60 and
indicate the shortcut type to either a trace or a
screen copy.
2 Press [SYS] > {File} > {File Setup} > {Save Path} to either
local memory or an external USB disk.
N9340A User’s Guide61
4System Setting
3 Press [ESC], [Save] to save a trace or a screen copy to your
pre-defined memory. The trace or screen will be
automatically assigned with a default file name. A trace
will be assigned with a file name, such as HYTRACE, HYTRACE_1, and consecutive decimal numbers appended.
A screen copy will be assigned with a file name, as
HYSCREEN, HYSCREEN_1 and consecutive decimal
numbers appended to the alphabetic file name.
To quick save or load a setup, please refer to “Quick saving
or loading a system setup" on page 59.
Viewing file list
Refer to the following two steps to view file list:
1 Select the directory you wish to view.
Press [SYS] > {File} > {View} to toggle the displayed
file list from local memory or an external USB
memory stick.
2 Select the file type you wish to view.
Press [SYS] > {File} > {File setup} > {File type} to
select a file type.
62N9340A User’s Guide
Deleting a file
A deleted file can NEVER be recovered. Carefully
CAU-CAUTION
BE
decide whether you need to delete the file(s).
1 Press[SYS] > {File} > {View} to select a directory.
Accesses a list of the 30 most recent error
messages reported by the analyzer. The most recent
error will appear at the bottom of the list. If the
error list is longer than 30 entries, the analyzer
reports an error message [–350, Query overflow]. When in remote control, and the error
list is greater than 30 entries, the error display is
halted and a message at the bottom of the list
informs the user that error exceed 30.
For more information, refer to “Error Messages" on
page 143.
System Setting4
N9340A User’s Guide65
4System Setting
Perform a Time Base Calibration
As soon as the calibration function triggers, it
interrupts the current measurement is interrupted
and a gauge displays on the LCD. The gauge simply
indicates calibration action rather than calibration
course, as the calibration time is unpredictable.
When the calibration is finished, the LCD displays
a calibration, and the interrupted measurement
restarts.
Time base calibration takes during a short time only when the
NOTE
inner temperature is stable. When the inner temperature is
increasing, calibration takes a long-time course or will fail. If
the input reference signal is abnormal, the calibration cycle
will take a long and unpredictable time to exit, and the LCD
displays an error message.
Key Access: [SYS] > {More (2 of 3)} > {Calibration}
66N9340A User’s Guide
Factory Default Settings
ParameterDefault Setting
Center Frequency1.5 GHz
Start Frequency 0.0 Hz
Stop Frequency3.0 GHz
Span3.0 GHz
Reference Level0.0 dBm
Attenuation Auto (20 dB)
Scale/DIV10 dB/DIV
Scale TypeLog
RBWAuto (1 MHz)
VBWAuto (1 MHz)
Average TypeLog Power
Sweep timeAuto
Sweep ModeNormal
Trace 1Clear write
Trac e 2B l a n k
Trac e 3B l a n k
Trac e 4B l a n k
Trac e 1 Detect ion Pos Peak
Trace 2 Detection Pos Peak
Trace 3 Detection Pos Peak
Trace 4 Detection Pos Peak
Trace AverageAll Off
MarkerAll Off
File TypeTrace
Save PathLocal
ModeSpectrum Analyzer
External Input typeTrigger in (TTL signal input)
System Setting4
For more information about saving a user-defined
preset, refer to “Selecting a preset type” and
“Saving a User-defined Preset" on page 22.
N9340A User’s Guide67
4System Setting
68N9340A User’s Guide
Agilent N9340A
Handheld Spectrum Analyzer
5
Key Reference
Agilent Technologies
69
5Key Reference
Amplitude
Activates the reference level function and access
the associated softkeys to set functions that affect
the way data on the vertical axis is displayed or
corrected.
Ref level
Activates the reference level function.
The reference level is the amplitude power or
voltage represented by the top graticule on the
display. Changing the value of the reference level
changes the absolute amplitude level (in the
selected amplitude units) of the top graticule line.
To change the reference level, use the arrow keys,
the knob, or the numeric keypad. Pressing any
digit (0 through 9) on the numeric keypad brings
up the terminator menu.
Key access: [AMPTD]
Attenuation
Toggles the input attenuation between Auto or Man.
The input attenuation ranges from 1dB to 51 dB
and allows you to set the attenuation level in 1 dB
increments when Attenuation is set to Man.
The input attenuator, normally coupled to the
reference level control, reduces the power level of
the analyzer input signal at the input mixer. The
attenuator is recoupled when Attenuation is set to
Auto.
To prevent damage to the input mixer, do not exceed a
CAU-CAUTION
power level of +33 dBm at the input. A granted power
level for stable sensitive measurement is less than 20
dBm.
Key access: [AMPTD]
70N9340A User’s Guide
Preamp
Toggles the internal preamp between On and Off.
Preamp results in a correction being applied to
compensate for the gain of the preamp at 20 dBm
so that amplitude readings show the value at the
input connector.
When the preamp is on, a PA indication appears on the left
NOTE
side of the display.
The preamp is set to on in frequency bands from
1 MHz to 3 GHz, otherwise the correction is not
applied even though the PA indication still appears.
Key access: [AMPTD]
Scale/DIV
Sets the logarithmic units per vertical graticule
division on the display.
The Scale/Div function is only available when the Scale Type
NOTE
key is set to Log.
Key access: [AMPTD]
Scale Type
Key Reference5
Toggles the scale type of vertical graticule divisions
between logarithmic and linear.
When Scale Type is set to Log, a Logarithmic units
is activated ranging from 1 to 10 dB per division.
When Scale Type is set to Lin, the vertical divisions
are linearly scaled and the default amplitude units
are volts. The top of the display is set to the
reference- level value and the bottom graticule
becomes zero volts. (Each division of the tragically
is one- tenth of the reference level in volts.)
Key access: [AMPTD] > {Scale Type}
N9340A User’s Guide71
5Key Reference
Ref Offset
Adds an offset value which ranges from -327.6 dB
to +327.6 dB to the displayed reference level.
Reference-level offsets are only entered by using the numeric
NOTE
keypad. Entering an offset does not affect the trace or the
attenuation value.
Reference- level offsets are used when gain or loss
occurs between a device under test and the
analyzer input. Thus, the signal level measured by
the analyzer may be referred to as the level at the
input of an external amplitude conversion device.
When an amplitude offset is entered, its value
appears in the lower left corner of the display.
Key access: [AMPTD]
Y Axis Units
Sets the amplitude units.
Amplitude units are maintained for both logarithmic and linear
NOTE
modes depending on the Scale Type setting.
Key access: [AMPTD] > {Y Axis Units}
72N9340A User’s Guide
BW/SWP
Activates the bandwidth function and accesses the
associated softkeys to control resolution bandwidth,
video bandwidth and sweep time.
RBW
Changes the 3- dB resolution bandwidth on the
analyzer from 30 Hz to 1 MHz in a 1, 3, 10
sequence using the knob, step keys or the numeric
keypad.
If an unavailable bandwidth is entered using the numeric
NOTE
keypad, the closest available bandwidth in the 1, 3, 10
sequence is used.
As the resolution bandwidth is decreased, the
sweep time is modified to maintain amplitude
calibration. Resolution bandwidth is also related to
span. As span is decreased, resolution bandwidth
decreases. As the resolution bandwidth changes,
the video bandwidth, if in auto couple mode,
changes to maintain the VBW/RBW ratio.
In non-zero span mode, the value of RBW is determined by
NOTE
whether the setting for RBW is Auto or Man (manual).
Key Reference5
A “#”mark appears next to RBW on the display
when it is not coupled.
Key access: [BW/SWP]
N9340A User’s Guide73
5Key Reference
VBW
Changes the analyzer post- detector filter from
3 Hz to 1 MHz in a 1, 3, 10 sequence using the
knob, the step keys, or the numeric keypad.
If an unavailable bandwidth entered using the numeric
NOTE
keypad, the closest available bandwidth in the 1, 3, 10
sequence is used.
As the video bandwidth is decreased, the sweep
time is increased to maintain amplitude calibration.
A “#”mark appears next to RBW on the display
when it is not coupled.
Key access: [BW/SWP]
VBW/RBW
Selects the ratio between the video and resolution
bandwidths.
If signal responses near the noise level are visually
masked by the noise, the ratio can be set to less
than 1 to lower the noise. The knob and step keys
change the ratio in a 1, 3, 10 sequence.
When preset is set to default, or VBW/RBW is set to Auto,
NOTE
the ratio is set to 1.
Key access: [BW/SWP]
Avg Type
Accesses the functions to automatically or
manually choose from one of the following
averaging type:
74N9340A User’s Guide
Key Reference5
Log Pwr
Averages the data as appropriate for the
logarithmic scaled y axis.
When average type is set to Log Pwr, “LgAv”
appears on the left side of the display.
Key access: [BW/SWP] > {Avg Type}
Pwr
Converts the scale from logarithmic to linear power
units, and then averages the power trace data.
With average type set to Pwr, “PAvg” appears on
the left side of the display.
Key access: [BW/SWP] > {Avg Type}
Voltage
Averages the data as appropriate for the linearly
scaled y- axis scale.
With average type set to Voltage, “VAvg” appears on
the left side of the display.
Key access: [BW/SWP] > {Avg Type}
There are actually four types of averaging
processes within an N9340A. All, except VBW
filtering, are affected by this setting. They are
listed as follows:
• Trace averaging
Averages signal amplitudes on a trace- to- trace
basis. (Press Log Pwr or Pwr)
•Average detector
Averages signal amplitudes during the time or
frequency interval represents the result as a single
measurement point.
•Noise marker
Averages signal amplitudes across measurement
points to reduce variations for noisy signals.
• VBW filtering
Performs a form of averaging of video signal.
N9340A User’s Guide75
5Key Reference
Sweep Time
Selects the length of time the analyzer takes to
tune across the displayed frequency span (or, in
zero span, the time the analyzer takes to sweep the
full screen) using the knob, the arrow keys, or
numeric keypad.
Reducing the sweep time increases the sweep rate.
NOTE
Key access: [BW/SWP]
In non-zero spans:
When the sweep time is auto- coupled, the analyzer
selects the optimum (shortest) sweep time ranging
from 10 ms to 1000 s for the current settings. This
selection process is influenced by three factors:
• Frequency Span
• The selected Resolution Bandwidth
• The selected Video Bandwidth
Note that changing one of the three factors above
changes sweep time. The sweep time is in direct
ratio to frequency span and is in reverse ratio to
the selected resolution bandwidth and video
bandwidth when auto- coupled.
With sweep time set to a value less than the minimum
CAU-CAUTION
auto-coupled value, an annotation “Data out of range”
appears in the active function block and the sweep time is
automatically converted to the minimum auto-coupled
value.
In zero span:
The auto- coupled status is invalid in zero span.
You can only set the sweep time manually. The
sweep time ranges from 6 s to 200 s in zero span.
76N9340A User’s Guide
µ
Key Reference5
In FFT Mode
In FFT mode, the sweep time is auto- coupled as
default. And the submenu of sweep time is invalid
in FFT mode.
Sweep
Activates the sweep mode function and accesses
associated softkeys to set functions that affect the
way a trace sweeps across the display.
Sweep
Toggles the analyzer between the continuous sweep
mode and the single sweep mode.
• Sweep (Single)
Puts the analyzer in a single sweep mode. In this
mode, pressing [BW/SWP] > {Single Sweep} enables
a sweep.
Key access: [BW/SWP] > {Sweep}
• Sweep (Cont)
Performs one sweep after another as soon as it
is triggered. Pressing [Preset] (When Preset Type
is set to Default) or turning the power on, sets
the sweep mode to continuous sweep as default.
Key access: [BW/SWP] > {Sweep}
Sweep Mode
An N9340A provides two sweep modes to meet
different sweep time requirements:
•Normal
Activates the sweep mode as default.
Key access: [BW/SWP] > {Sweep} > {Sweep Mode}
N9340A User’s Guide77
5Key Reference
•Fast
Activates the fast sweep mode.
Fast sweep mode provides a fast measurement
function that decreases the sweep time. However,
this mode brings a decrease of amplitude accuracy.
Fast sweep is valid only when span is no less than 50 MHz.
NOTE
Measurement data is not ganruanteed in fast sweep mode.
Key access: [BW/SWP] > {Sweep} > {Sweep Mode}
Single Sweep
When the analyzer is in continuous sweep mode
and not in a measurement mode ([MEAS] > Measure Off), pressing [BW/SWP] > {Single Sweep} convert the
continuous sweep to single sweep and executes a
sweep after the trigger condition is met. If the
analyzer is already in single sweep, pressing
[BW/SWP] > {Single Sweep} executes a new sweep as
soon as the trigger condition is met.
If Average is on ([TRACE] > {More (1 of 2)} > {Average} > On), pressing [BW/SWP] > {Single Sweep} halts
averaging of the current value, and continue
averaging from the current count number as soon
as a single sweep function triggers.
Key access: [BW/SWP] > {Sweep}
78N9340A User’s Guide
Enter
Key Reference5
• Terminates date entry and enters into the
analyzer a numerical value keyed from numeric
keypad on the front panel.
• Terminate filename entries.
N9340A User’s Guide79
5Key Reference
ESC/CLR
Provides mainly two types of functions:
Clear
• Clears any numeric entry and cancels the active
function.
• Clears any title entry and cause the title to
revert to the previous name.
• Clears input or output overloads
• Clears error messages from the status line along
the bottom of the display.
Association with other functions
Provides a permitted setting for triggering other
functions as follows:
• Activates LIMIT function for pass/fail judgements.
• Activates SAVE function for current trace saving.
When pressing [ESC/CLR], a current parameter
CAU-CAUTION
annotation in the lower left corner disappears which is an
indicator of available settings of LIMT and SAVE.
80N9340A User’s Guide
Frequency
Activates the center frequency function, and
accesses the menu of frequency functions. The
center frequency, or start and stop frequency
values appear below the graticule on the display.
When changing both the center frequency and the span,
NOTE
change the frequency first since the span can be limited by
the frequency value.
Center Frequency
Activates the center frequency function which
allows you to set the horizontal center of the
display to a specified frequency.
Key access: [FREQ]
Start Frequency
Sets the frequency at the left side of the graticule.
The left and right sides of the graticule correspond
to the start and stop frequencies. When these
frequencies are activated, their values are
displayed below the graticule in place of center
frequency and span.
Key access: [FREQ]
Key Reference5
Stop Frequency
Sets the frequency at the right side of the
graticule.The left and right sides of the graticule
correspond to the start and stop frequencies. When
these frequencies are activated, their values are
displayed below the graticule in place of center
frequency and span.
Key access: [FREQ]
N9340A User’s Guide81
5Key Reference
CF Step
Changes the step size for the center frequency
function. Once a step size has been selected and
the center frequency function is activated, the
arrow keys change the center frequency by the
step- size value. The step size function is useful for
finding harmonics and sidebands beyond the
current frequency span of the analyzer. When
auto-coupled, the center size is set to one division
(10 percent of the span).
Key access: [FREQ]
82N9340A User’s Guide
Marker
Marker
Normal
Key Reference5
Accesses the marker control softkeys to select the
type and number of markers. Markers are
diamond- shaped characters that identify points of
traces. Up to six pairs of markers may appear on
the display simultaneously; only one pair can be
controlled at a time. The marker that is controlled
is called the “active” marker. Pressing [MARKER]
activates the Normal menu key.
Selects one of the six possible markers. A marker
that has already been selected will become active
when it is turned on. If a marker has been turned
on and assigned to a specific trace, it will become
active on that trace if that marker is selected.
Key access: [MARKER]
Activates a single frequency marker at the center
of the active trace if a marker is not already
displayed. If a marker is displayed before the
Normal function is enabled, the marker is enabled
at the position of the selected marker. The marker
number is indicated above the marker. The knob
and/or arrow keys move the marker left or right. If
a value is entered from the numeric keypad, the
marker is moved to the trace point nearest to that
value. Annotation in the active function block and
in the upper right corner indicates the frequency
and amplitude of the marker (time and amplitude,
if in zero span.) Pressing Normal turns off the Delta
function and moves the active marker to the delta
marker position.
Key access: [MARKER]
N9340A User’s Guide83
5Key Reference
Delta
Activates a second marker at the position of the
first marker. (If no marker is present, two markers
appear at the center of the display). The amplitude
and frequency (or time) of the first marker is
fixed. The marker number is indicated above the
delta marker, and the same number is indicated
with an R (for example, 1R) above the reference
marker. Use the data controls to position the delta
marker. Annotation in the active function block
and in the upper- right corner of the display
indicates the frequency (or time) and amplitude
differences between the two markers. The markers
will be turned off if the scale type is changed
between log and linear.
Press Delta again moves the reference marker to the active
NOTE
marker position, so you are allowed to make delta
measurements from differing reference points without having
to turn off the makers and begin again.
The delta marker function permits signal-to-noise
NOTE
measurements provided the signal is a single spectral
component (sinusoid). Place a normal marker on the signal,
press Delta, place the delta marker in the noise, and activate
Marker Noise. The indicated amplitude indicates
signal-to-noise/Hz.
Key access: [MARKER]
Off
Turns off the active function markers.
Key access: [MARKER]
84N9340A User’s Guide
Key Reference5
Peak Search
Place a marker on the highest peak based on the
setting of the Search Criteria as follows:
•Peak
Place a marker on the highest peak.
Key access: [MARKER]
•Next LF Peak
Moves the marker to the next peak to the left of
the current marker. The signal peak must exceed
the peak threshold value by the peak excursion
value. If there is no peak to the right, the marker
will not move and the No Peak Found error message
will appear on the display.
Key access: [MARKER]
•Next RT Peak
Moves the marker to the next peak to the right of
the current marker. The signal peak must exceed
the peak threshold value by the peak excursion
value. If there is no peak to the right, the marker
will not move and the No Peak Found error message
will appear on the display.
Key access: [MARKER]
•Auto Search
In non- zero span, pressing Auto Search turns all six
markers on and puts them on the six highest peak
in the current frequency span.
In zero span, the function is not available.
Key access: [MARKER]
N9340A User’s Guide85
5Key Reference
Marker To
Accesses the following marker function menu keys:
•To Center
Sets the center frequency of the analyzer to the
marker frequency. In Delta mode, pressing [Marker] > {Marker To} > {To Center} sets the center frequency
to the marker delta value. The function is not
available in zero span.
When the frequency scale is in log mode, the center
NOTE
frequency is not at the center of the display.
Key access: [MARKER] > {To Center}
•To Ref
Changes the reference level to the active marker
value, moving the marked point to the reference
level (top line of the graticule). In Delta mode,
pressing [Marker] > {Marker To} > {To Ref} sets the
reference level to the amplitude difference between
the markers.
Key access: [MARKER] > {Marker To}
Mode
Accesses the following menu keys that allows you
to change the active marker readout mode.
•Normal
Activates the default readout mode.
• Freq Count
Activates a frequency counter mode.
Frequency counter is properly effective only on CW signals or
NOTE
discrete peaks. For a valid reading, the marker must be greater
than 25 dB above the noise.
Freq Count function is not supported under Marker Delta
condition.
86N9340A User’s Guide
•Noise
Activates a noise readout mode for evaluating
power density.
Noise mode operates only on noise peaks and not supported
NOTE
under Marker Delta condition.
Key access: [MARKER] > {More (1 of 2)} > {Mode}
Marker Trace
Activates a marker on the trace if there are no
markers turned on. If a marker is currently active,
press Marker Trace until 1, 2, 3, or 4 is
underlined. The active marker will be moved to the
selected trace.
Selecting
automatically selected trace. The selection order is
to look for the lowest numbered trace in the
following order of trace modes:
Clear writeMax hold
Min. holdView
If there is no trace in any of these modes, it
selects trace 1 as default.
Key access: [MARKER] > {Marker Trace}
Auto will move the marker to the
Key Reference5
All Off
Turns off all of the markers, including markers
used for signal track and demodulation. Marker
annotation is also removed.
Key access: [MARKER] > {All Off}
N9340A User’s Guide87
5Key Reference
Meas
In spectrum analysis mode, pressing [Meas] brings
up the menu for a suite of power measurements
such as adjacent channel power, occupied
bandwidth, and channel power.
Refer to “One-button Power Measurement" on
page 43for more information about these
measurements.
88N9340A User’s Guide
Span
Key Reference5
Activates the span function and accesses the
submenu of span functions. Pressing [SPAN] allows
you to change the frequency range symmetrically
about the center frequency. The frequency- span
readout describes the total displayed frequency
range. To determine frequency span per horizontal
graticule division (when the frequency scale type is
set to linear), divide the frequency span by 10.
Span
Allows you to enter a frequency span range value.
Key access: [SPAN]
Full
Changes the span to full span showing the full
frequency range of the analyzer.
Key access: [SPAN]
Zero
Changes the frequency span to zero. In this mode,
the current center frequency is displayed in the
time domain (the x axis is displayed in units of
time), like a conventional oscilloscope.
Key access: [SPAN]
Last Span
Changes the span to the previous span setting.
Key access: [SPAN]
N9340A User’s Guide89
5Key Reference
TRACE
Accesses the trace keys to store and manipulate
trace information. Each trace comprises a series of
data points in which amplitude information is
stored. The analyzer updates the information for
any active trace with each sweep.
If you have selected Channel Power, OBW, or ACPR
in the MEAS menu, refer to “One- button Power
Measurement" on page 43.
Tr ac e
Selects the trace for current use.
Key access: [TRACE]
Clear Write
Erases any data previously stored in the selected
trace and continuously displays any signals during
the sweep of the analyzer. This function is
activated for trace 1 at power on and factory
preset.
Key access: [TRACE]
Max Hold
Maintains the maximum level for each trace point
of the selected trace, and updates each trace point
if a new maximum level is detected in successive
sweeps.
Changing the vertical scale (Amplitude, Scale Type, Log or
NOTE
Line), or pressing PRESET, or turning averaging on (TRACE,
Average (On)) or switching widow in Zone Span restarts the
held trace.
Key access: [TRACE]
90N9340A User’s Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.