Agilent Technologies E1401-90493 User Manual

Calibration Guide

Agilent Technologies ESA Spectrum Analyzers
This manual provides documentation for the following instruments:
Agilent ESA-E Series E4402B (9 kHz - 3.0 GHz) E4404B (9 kHz - 6.7 GHz) E4405B (9 kHz - 13.2 GHz) E4407B (9 kHz - 26.5 GHz)
and
Manufacturing Part Number: E4401-90493
Supersedes E4401-90475
Printed in USA
March 2005
© Copyright 2000-2005 Agilent Technologies, Inc.
Notice
The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
The following safety symbols are used throughout this manual. Familiarize yourself with the symbols and their meaning before operating this instrument.
WAR NING Warning denotes a hazard. It calls attention to a procedure
which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.
CAUTION Caution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, could result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
WAR NING This is a Safety Class 1 Product (provided with a protective
earthing ground incorporated in the power cord. The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
WAR NING If this product is not used as specified, the protection provided
by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.
CAUTION Always use the three-prong ac power cord supplied with this product.
Failure to ensure adequate earth grounding by not using this cord may cause product damage.
CAUTION This instrument has autoranging line voltage input, be sure the supply
voltage is within the specified range.
2
Where to Find the Latest Information
Documentation is updated periodically. For the latest information about Agilent ESA Spectrum Analyzers, including firmware upgrades and application information, please visit the following Internet URL:
http://www.agilent.com/find/esa.
3
4
Contents
1. Calibrating
Calibration Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2. Performance Verification Tests
Tests included in this section: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
1. 10 MHz Reference Output Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
2. 10 MHz High-Stability Frequency Reference Output Accuracy: Agilent E4401B, E4402B,
E4404B, E4405B, and E4407B (Option 1D5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4401B, E4402B,
E4403B, and E4411B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B,
E4407B, and E4408B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5. Frequency Span Accuracy: Agilent E4401B and E4411B . . . . . . . . . . . . . . . . . . . . . . . . 55
6. Frequency Span Accuracy:
Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B . . . . . . . . . . . . . . . . . 59
7. Noise Sidebands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407B . . . . 70
9. System-Related Sidebands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
10. Residual FM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
11. Sweep Time Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
12. Display Scale Fidelity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
13. Input Attenuation Switching Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
14. Reference Level Accuracy: Agilent E4401B and E4411B . . . . . . . . . . . . . . . . . . . . . . 107
15. Reference Level Accuracy: Agilent E4402B,
E4403B, E4404B, E4405B, E4407B, and E4408B. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
16. Resolution Bandwidth Switching Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411B . . 129
18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B,
5
Contents
E4405B, E4407B, and E4408B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .136
19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411B . . . . . . . . . . . .143
20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B,
E4407B, and E4408B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .150
21. Resolution Bandwidth Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .157
22. Frequency Response: Agilent E4401B and E4411B . . . . . . . . . . . . . . . . . . . . . . . . . . .162
23. Frequency Response, Agilent E4402B and E4403B . . . . . . . . . . . . . . . . . . . . . . . . . . .172
24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B . . . . . . . . . . .187
25. Frequency Response (Preamp On): Agilent E4401B . . . . . . . . . . . . . . . . . . . . . . . . . .213
26. Frequency Response (Preamp On): Agilent E4402B . . . . . . . . . . . . . . . . . . . . . . . . . .221
27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B . . . . . . . .231
28. Other Input-Related Spurious Responses: Agilent E4401B and E4411B . . . . . . . . . . .244
29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B,
E4407B, and E4408B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .249
30. Spurious Responses: Agilent E4401B and E4411B . . . . . . . . . . . . . . . . . . . . . . . . . . .255
31. Spurious Responses: Agilent E4402B and E4403B . . . . . . . . . . . . . . . . . . . . . . . . . . .263
32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B . . . . . . . . . . . .273
33. Gain Compression: Agilent E4401B, E4402B, E4403B, and E4411B . . . . . . . . . . . . .290
34. Gain Compression: Agilent E4404B, E4405B, E4407B, and E4408B . . . . . . . . . . . . .294
35. Displayed Average Noise Level: Agilent E4401B and E4411B . . . . . . . . . . . . . . . . . .299
36. Displayed Average Noise Level: Agilent E4402B and E4403B . . . . . . . . . . . . . . . . . .309
37. Displayed Average Noise Level: Agilent E4404B and E4405B . . . . . . . . . . . . . . . . . .316
38. Displayed Average Noise Level: Agilent E4407B and E4408B . . . . . . . . . . . . . . . . . .323
39. Residual Responses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .330
40. Fast Time Domain Amplitude Accuracy: Agilent E4401B, E4402B, E4404B, E4405B,
and E4407B (Option AYX) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .333
41. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4401B and
E4411B (Option 1DN or 1DQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .335
6
Contents
42. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4402B,
E4403B, E4404B, E4405B, E4407B, and E4408B (Option 1DN) . . . . . . . . . . . . . . . . . . . 340
43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)
343
44. Tracking Generator Level Flatness:
Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B (Option 1DN) . . . . 350
45. Tracking Generator Harmonic Spurious Outputs: Agilent E4401B and E4411B
(Option 1DN or 1DQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 356
46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,
E4405B, E4407B, and E4408B (Option 1DN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 360
47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Op-
tion 1DN or 1DQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 364
48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B,
E4404B, E4405B, E4407B, and E4408B (Option 1DN) . . . . . . . . . . . . . . . . . . . . . . . . . . 370
49. Tracking Generator LO Feedthrough: Agilent E4402B, E4403B, E4404B,
E4405B, E4407B, and E4408B(Option 1DN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 376
50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B,
E4405B, and E4407B (Option 1D6) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 379
51. Gate Mode Additional Amplitude Error: Agilent E4401B, E4402B, E4404B, E4405B,
and E4407B (Option 1D6) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 384
52. First LO OUTPUT Power Accuracy (Option AYZ only) . . . . . . . . . . . . . . . . . . . . . . 387
53. IF INPUT Accuracy (Option AYZ only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 390
54. Comms Frequency Response (Option BAC or BAH) . . . . . . . . . . . . . . . . . . . . . . . . . 394
58. GSM Phase and Frequency Error (Options BAH and B7E) . . . . . . . . . . . . . . . . . . . . . 406
59. Comms Absolute Power Accuracy (Options BAC or BAH) . . . . . . . . . . . . . . . . . . . . 409
3. Performance Verification Test Records
Agilent E4401B Performance Verification Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . 428
Agilent E4402B Performance Verification Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . 450
7
Contents
Agilent E4403B Performance Verification Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . .479
Agilent E4404B Performance Verification Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . .495
Agilent E4405B Performance Verification Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . .525
Agilent E4407B Performance Verification Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . .557
Agilent E4408B Performance Verification Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . .596
Agilent E4411B Performance Verification Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . .617
4. If You Have a Problem
What You’ll Find in This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .636
Before You Call Agilent Technologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .637
How to Return Your Analyzer for Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .640
8

1Calibrating

9
Calibrating
Calibration verifies that the analyzer performance is within all specifications. It is time consuming and requires extensive test equipment. Calibration consists of all the performance tests. For a complete listing of the performance tests, see the performance verification tests table for your specific analyzer.
Allow the analyzer to warm up in accordance with the temperature stability specifications before performing the tests in this chapter.
None of these test procedures involve removing the cover of the analyzer.
10 Chapter 1
Calibrating

Calibration Cycle

Calibration Cycle
The performance tests in chapter 2 should be used to check the analyzer against its specifications once every year. Specifications are listed in the Specifications Guide.

Performance Verification Test Tables

The tables on the following pages list the performance tests in Chapter
2 , “Performance Verification Tests,” required for each model number.
Perform all the tests marked with a dot in the “Std” (standard) column. If any options are installed in the analyzer, also perform all tests marked with a dot in the appropriate option column.
Chapter 1 11
Calibrating Calibration Cycle
Table 1-1 Agilent E4401B Performance Verification Tests
Calibration for Instrument Option:
Performance Test Name
Stda1DN
1.
10 MHz Reference Output Accuracy
2. 10 MHz High-Stability Frequency Reference Output Accuracy
3. Frequency Readout and Marker Frequency Count Accuracy
5. Frequency Span Accuracy
7. Noise Sidebands
9. System-Related Sidebands
10. Residual FM
11. Sweep Time Accuracy
12. Display Scale Fidelity
13. Input Attenuation Switching Uncertainty
14. Reference Level Accuracy
16. Resolution Bandwidth Switching Uncertainty
17. Absolute Amplitude Accuracy (Reference Settings)
19. Overall Absolute Amplitude Accuracy
21. Resolution Bandwidth Accuracy
22. Frequency Response
25. Frequency Response (Preamp On)
28. Other Input-Related Spurious Responses
30. Spurious Responses
33. Gain Compression
35. Displayed Average Noise Level
39. Residual Responses
40. Fast Time Domain Amplitude Accuracy
41. Tracking Generator Absolute Amplitude and Vernier Accuracy
43. Tracking Generator Level Flatness
45. Tracking Generator Harmonic Spurious Outputs
47. Tracking Generator Non-Harmonic Spurious Outputs
50. Gate Delay Accuracy and Gate Length Accuracy
51. Gate Mode Additional Amplitude Error
a. Perform these tests for all E4401B analyzers. b. Perform this test only on analyzers not equipped with Option 1D5.
b
••
••
••
1DQ
1DS
1D5
1D6
AYX
12 Chapter 1
Calibrating
Calibration Cycle
Table 1-2 Agilent E4402B Performance Verification Tests
Calibration for Instrument Option:
Performance Test Name
1.
10 MHz Reference Output Accuracy
2. 10 MHz High-Stability Frequency Reference Output Accuracy
3. Frequency Readout and Marker Frequency Count Accuracy
6. Frequency Span Accuracy
7. Noise Sidebands
8. Noise Sidebands - Wide Offsets
9. System-Related Sidebands
10. Residual FM
11. Sweep Time Accuracy
12. Display Scale Fidelity
13. Input Attenuation Switching Uncertainty
15. Reference Level Accuracy
16. Resolution Bandwidth Switching Uncertainty
18. Absolute Amplitude Accuracy (Reference Settings)
20. Overall Absolute Amplitude Accuracy
21. Resolution Bandwidth Accuracy
23. Frequency Response
26. Frequency Response (Preamp On)
29. Other Input-Related Spurious Responses
31. Spurious Responses
33. Gain Compression
36. Displayed Average Noise Level
39. Residual Responses
40. Fast Time Domain Amplitude Accuracy
42. Tracking Generator Absolute Amplitude and Vernier Accuracy
44. Tracking Generator Level Flatness
46. Tracking Generator Harmonic Spurious Outputs
48. Tracking Generator Non-Harmonic Spurious Outputs
49. Tracking Generator L.O. Feedthrough
50. Gate Delay Accuracy and Gate Length Accuracy
51. Gate Mode Additional Amplitude Error
54. Comms Frequency Response
55.
(This test has been deleted.)
b
a
Std
1D5
1DS
1DN
1D6
AYX
BAC
B7E
BAH
Chapter 1 13
Calibrating Calibration Cycle
Table 1-2 Agilent E4402B Performance Verification Tests
Calibration for Instrument Option:
Performance Test Name
1D5
Stda1DN
56.
(This test has been deleted.)
57.
(This test has been deleted.)
58.
GSM Phase and Frequency Error
59. Comms Absolute Power Accuracy (Options BAC or BAH)
a. Perform these tests for all E4402B analyzers. b. Perform this test only on analyzers not equipped with Option 1D5. c. Perform this test only on instruments having Option BAH, GSM Measurement Personality.
c
1DS
1D6
AYX
BAC
BAH
••
B7E
14 Chapter 1
Calibrating
Calibration Cycle
Table 1-3 Agilent E4403B Performance Verification Tests
Calibration for
Performance Test Name
1. 10 MHz Reference Output Accuracy
3. Frequency Readout and Marker Frequency Count Accuracy
6. Frequency Span Accuracy
7. Noise Sidebands
9. System-Related Sidebands
10. Residual FM
11. Sweep Time Accuracy
12. Display Scale Fidelity
13. Input Attenuation Switching Uncertainty
15. Reference Level Accuracy
16. Resolution Bandwidth Switching Uncertainty
18. Absolute Amplitude Accuracy (Reference Settings)
20. Overall Absolute Amplitude Accuracy
21. Resolution Bandwidth Accuracy
23. Frequency Response
29. Other Input-Related Spurious Responses
31. Spurious Responses
33. Gain Compression
36. Displayed Average Noise Level
39. Residual Responses
42. Tracking Generator Absolute Amplitude and Vernier Accuracy
44. Tracking Generator Level Flatness
46. Tracking Generator Harmonic Spurious Outputs
48. Tracking Generator Non-Harmonic Spurious Outputs
49. Tracking Generator L.O. Feedthrough
Instrument Option:
a
Std
1DN
a. Perform these tests on all E4403B analyzers.
Chapter 1 15
Calibrating Calibration Cycle
Table 1-4 Agilent E4404B Performance Verification Tests
Perform anc e Test Name
1.
10 MHz Reference Output Accuracy
b
2. 10 MHz High-Stability Frequency Reference Output
Calibration for Instrument Option:
1D5
Stda1DN
1DS
1D6
AYX
BAC
Accuracy
4. Frequency Readout and Marker Frequency Count
Accuracy
6. Frequency Span Accuracy
7. Noise Sidebands
8. Noise Sidebands - Wide Offsets
9. System-Related Sidebands
10. Residual FM
11. Sweep Time Accuracy
12. Display Scale Fidelity
13. Input Attenuation Switching Uncertainty
15. Reference Level Accuracy
16. Resolution Bandwidth Switching Uncertainty
18. Absolute Amplitude Accuracy (Reference Settings)
20. Overall Absolute Amplitude Accuracy
21. Resolution Bandwidth Accuracy
24. Frequency Response
27. Frequency Response (Preamp On)
29. Other Input-Related Spurious Responses
32. Spurious Responses
34. Gain Compression
37. Displayed Average Noise Level
39. Residual Responses
40. Fast Time Domain Amplitude Accuracy
42. Tracking Generator Absolute Amplitude and Vernier
Accuracy
44. Tracking Generator Level Flatness
46. Tracking Generator Harmonic Spurious Outputs
48. Tracking Generator Non-Harmonic Spurious Outputs
49. Tracking Generator L.O. Feedthrough
50. Gate Delay Accuracy and Gate Length Accuracy
51. Gate Mode Additional Amplitude Error
54. Comms Frequency Response
55.
(This test has been deleted.)
B7E
BAH
16 Chapter 1
Calibrating
Table 1-4 Agilent E4404B Performance Verification Tests
Calibration Cycle
Perform anc e Test Name
56.
(This test has been deleted.)
57.
(This test has been deleted.)
58.
GSM Phase and Frequency Error
c
Calibration for Instrument Option:
1D5
Stda1DN
1DS
1D6
AYX
BAC
59. Comms Absolute Power Accuracy (Options BAC or BAH)
a. Perform these tests on all Agilent E4404B analyzers. b. Perform this test only on analyzers not equipped with Option 1D5. c. Perform this test only on instruments having Option BAH, GSM Measurement Personality.
B7E
BAH
Chapter 1 17
Calibrating Calibration Cycle
Table 1-5 Agilent E4405B Performance Verification Tests
Calibration for Instrument Option:
Performance Test Name
1.
10 MHz Reference Output Accuracy
2. 10 MHz High-Stability Frequency Reference Output Accuracy
4. Frequency Readout and Marker Frequency Count Accuracy
6. Frequency Span Accuracy
7. Noise Sidebands
8. Noise Sidebands - Wide Offsets
9. System-Related Sidebands
10. Residual FM
11. Sweep Time Accuracy
12. Display Scale Fidelity
13. Input Attenuation Switching Uncertainty
15. Reference Level Accuracy
16. Resolution Bandwidth Switching Uncertainty
18. Absolute Amplitude Accuracy (Reference Settings)
20. Overall Absolute Amplitude Accuracy
21. Resolution Bandwidth Accuracy
24. Frequency Response
27. Frequency Response (Preamp On)
29. Other Input-Related Spurious Responses
32. Spurious Responses
34. Gain Compression
37. Displayed Average Noise Level
39. Residual Responses
40. Fast Time Domain Amplitude Accuracy
42. Tracking Generator Absolute Amplitude and Vernier Accuracy
44. Tracking Generator Level Flatness
46. Tracking Generator Harmonic Spurious Outputs
48. Tracking Generator Non-Harmonic Spurious Outputs
49. Tracking Generator L.O. Feedthrough
50. Gate Delay Accuracy and Gate Length Accuracy
51. Gate Mode Additional Amplitude Error
54. Comms Frequency Response
55.
(This test has been deleted.)
b
a
Std
1D5
1DS
1DN
1D6
AYX
BAC
B7E
BAH
18 Chapter 1
Calibrating
Calibration Cycle
Table 1-5 Agilent E4405B Performance Verification Tests
Calibration for Instrument Option:
Performance Test Name
1D5
Stda1DN
56.
(This test has been deleted.)
57.
(This test has been deleted.)
58.
GSM Phase and Frequency Error
59. Comms Absolute Power Accuracy (Options BAC or BAH)
a. Perform these tests on all E4405B analyzers. b. Perform this test only if the analyzer is not equipped with Option 1D5. c. Perform this test only on instruments having Option BAH, GSM Measurement Personality.
c
1DS
1D6
AYX
BAC
B7E
BAH
Chapter 1 19
Calibrating Calibration Cycle
Table 1-6 Agilent E4407B Performance Verification Tests
Calibration for Instrument Option:
Performance Test Name
1.
10 MHz Reference Output Accuracy
b
2. 10 MHz High-Stability Frequency Reference Output Accuracy
4. Frequency Readout and Marker Frequency Count Accuracy
6. Frequency Span Accuracy
7. Noise Sidebands
8. Noise Sidebands - Wide Offsets
9. System-Related Sidebands
10. Residual FM
11. Sweep Time Accuracy
a
Std
1DS
1DN
1D5
1D6
AYX
BAC
AYZ
BAH
B7E
12. Display Scale Fidelity
13. Input Attenuation Switching Uncertainty
15. Reference Level Accuracy
16. Resolution Bandwidth Switching Uncertainty
18. Absolute Amplitude Accuracy (Reference Settings)
20. Overall Absolute Amplitude Accuracy
21. Resolution Bandwidth Accuracy
24. Frequency Response
27. Frequency Response (Preamp On)
29. Other Input-Related Spurious Responses
32. Spurious Responses
34. Gain Compression
38. Displayed Average Noise Level
39. Residual Responses
40. Fast Time Domain Amplitude Accuracy
42. Tracking Generator Absolute Amplitude and Vernier
Accuracy
44. Tracking Generator Level Flatness
46. Tracking Generator Harmonic Spurious Outputs
48. Tracking Generator Non-Harmonic Spurious Outputs
49. Tracking Generator L.O. Feedthrough
50. Gate Delay Accuracy and Gate Length Accuracy
51. Gate Mode Additional Amplitude Error
20 Chapter 1
Calibrating
Table 1-6 Agilent E4407B Performance Verification Tests
Calibration for Instrument Option:
Performance Test Name
52. First LO OUTPUT Power Accuracy
53. IF Input Accuracy
a
Std
1DN
1D5
1DS
1D6
AY X
Calibration Cycle
BAC
AYZ
BAH
B7E
54. Comms Frequency Response
55.
(This test has been deleted.)
56.
(This test has been deleted.)
57.
(This test has been deleted.)
58.
GSM - Phase and Frequency Error
c
59. Comms Absolute Power Accuracy (Options BAC or BAH)
a. Perform these tests on all E4407B analyzers. b. Perform this test only on analyzers not equipped with Option 1D5. c. Perform this test only on instruments having Option BAH, GSM Measurement Personality.
••
Chapter 1 21
Calibrating Calibration Cycle
Table 1-7 Agilent E4408B Performance Verification Tests
Calibration for
Performance Test Name
1. 10 MHz Reference Output Accuracy
4. Frequency Readout and Marker Frequency Count Accuracy
6. Frequency Span Accuracy
7. Noise Sidebands
9. System-Related Sidebands
10. Residual FM
11. Sweep Time Accuracy
12. Display Scale Fidelity
13. Input Attenuation Switching Uncertainty
15. Reference Level Accuracy
16. Resolution Bandwidth Switching Uncertainty
18. Absolute Amplitude Accuracy (Reference Settings)
20. Overall Absolute Amplitude Accuracy
21. Resolution Bandwidth Accuracy
24. Frequency Response
29. Other Input-Related Spurious Responses
32. Spurious Responses
34. Gain Compression
38. Displayed Average Noise Level
39. Residual Responses
42. Tracking Generator Absolute Amplitude and Vernier Accuracy
44. Tracking Generator Level Flatness
46. Tracking Generator Harmonic Spurious Outputs
48. Tracking Generator Non-Harmonic Spurious Outputs
49. Tracking Generator L.O. Feedthrough
Instrument Option:
a
Std
1DN
a. Perform these tests on all E4408B analyzers.
22 Chapter 1
Calibrating
Table 1-8 Agilent E4411B Performance Verification Tests
Performance Test Name
Calibration Cycle
Calibration for
Instrument Option:
Stda1DN
1. 10 MHz Reference Output Accuracy
3. Frequency Readout and Marker Frequency Count Accuracy
5. Frequency Span Accuracy
7. Noise Sidebands
9. System-Related Sidebands
10. Residual FM
11. Sweep Time Accuracy
12. Display Scale Fidelity
13. Input Attenuation Switching Uncertainty
14. Reference Level Accuracy
16. Resolution Bandwidth Switching Uncertainty
17. Absolute Amplitude Accuracy (Reference Settings)
19. Overall Absolute Amplitude Accuracy
21. Resolution Bandwidth Accuracy
22. Frequency Response
28. Other Input-Related Spurious Responses
30. Spurious Responses
33. Gain Compression
35. Displayed Average Noise Level
39. Residual Responses
41. Tracking Generator Absolute Amplitude and Vernier Accuracy
43. Tracking Generator Level Flatness
45. Tracking Generator Harmonic Spurious Outputs
47. Tracking Generator Non-Harmonic Spurious Outputs
1DQ
a. Perform these tests on all E4411B analyzers.
Chapter 1 23
Calibrating Before You Start

Before You Start

This brief procedure should be performed before starting the performance verification tests:
1. Switch the analyzer on and let it warm up for five minutes.
2. If the analyzer is an Agilent E4402B, E4403B, E4404B, E4405B, E4407B, or E4408B, connect a cable from AMPTD REF OUT to the
50 Input.
3. Press alignments to finish.
4. Read the remainder of this section before you start any of the tests, and make a copy of the Performance Verification Test Record described below in "Recording the test results."
System, Alignments, Align Now, All, and wait for the auto

Recording the test results

Performance verification test records, for each analyzer, are provided in the chapter following the tests.
Each test result is identified as a TR Entry in the performance tests and on the performance verification test record. We recommend that you make a copy of the performance verification test record, record the test results on the copy, and keep the copy for your calibration test record. This record could prove valuable in tracking gradual changes in test results over long periods of time.

Performing Self-Alignment

Perform a complete self-alignment at least once per day, or if the analyzer fails a verification test. To perform a self-alignment, press
System, Alignment, Align Now, All. The instrument must be up to
operating temperature in order for this test to be valid. If the analyzer continuously fails one or more specifications, complete any remaining tests and record all test results on a copy of the test record. Then refer to Chapter 4 , “If You Have a Problem,” for instructions on how to solve the problem.
24 Chapter 1
Calibrating

Periodic verification of operation

The analyzer requires periodic verification of operation. Under most conditions of use, you should test the analyzer at least once a year with the complete set of performance verification tests.

Test equipment you will need

The following tables list the recommended test equipment for the performance tests. The tables also list recommended equipment for the analyzer adjustment procedures which are located in the Agilent ESA Spectrum Analyzers Service Guide. Any equipment that meets the critical specifications given in the table can be substituted for the recommended model.
Table 1-9 Recommended Test Equipment
Before You Start
Equipment
Digital Multimeter Input Resistance
DVM Test Leads For use with HP/Agilent 3458A Digital
Universal Counter Frequency Range: 10 MHz ±
Frequency Standard Frequency: 10 MHz
Critical Specifications for Equipment
Substitution
10 megohms
Accuracy: ±10 mV on 100 V range
Multimeter
Time Interval Range: 25 ms to 100 ms Single Trigger Operation Range: 2.5 Vdc to –2.5 Vdc External Reference Input Gate Time: Frequency Resolution: 0.01 Hz
Timebase Accuracy (Aging): <1
10 seconds
9
×10
/day
100 Hz
Recommended
HP/Agilent Model
3458A P,A,T
34118B T
53132A P,A,T
5071A P,A
Use
a
Chapter 1 25
Calibrating Before You Start
Table 1-9 Recommended Test Equipment
Equipment
Critical Specifications for Equipment
Substitution
Oscilloscope Bandwidth: dc to 100 MHz
Vertical Scale Factor of 0.5 V to 5 V/Div Two channels Minimum Timebase Setting: <100 ns Digitizing display with pulse width and time interval measurement functions Delta –T measurement accuracy in 200 ns / div: <450 ps
Power Meter Compatible with HP/Agilent 8480
series power sensors. dB relative mode. Resolution: 0.01 dB Reference Accuracy: ±1.2%
RF Power Sensor (2 required)
Frequency Range: 100 kHz to 3 GHz Maximum SWR:
1.60 (100 kHz to 300 kHz)
1.20 (300 kHz to 1 MHz)
1.1 (1 MHz to 2.0 GHz)
1.18 (2.0 GHz to 3.0 GHz) Amplitude range: –25 dBm to 10 dBm
Recommended
HP/Agilent Model
54820A T
E4419B P,A,T
8482A P,A,T
Use
a
Microwave Power Sensor
Power Sensor, Low Power
Synthesized Signal Generator
Frequency Range: 50 MHz to 26.5 GHz Maximum SWR:
1.15 (50 MHz to 100 MHz)
1.10 (100 MHz to 2 GHz)
1.15 (2 GHz to 12.4 GHz)
1.20 (12.4 GHz to 18 GHz)
1.25 (18 GHz to 26.5 GHz) Amplitude range: –25 dBm to 0 dBm
Frequency Range: 50 MHz to 3.0 GHz Amplitude Range: –20 dBm to –70 dBm Maximum SWR:
1.4 (10 MHz to 30 MHz)
1.15 (30 MHz to 3.0 GHz)
Frequency Range: 100 kHz to 1500 MHz Amplitude Range: –35 to 16 dBm SSB Noise: <–120 dBc/Hz at 20 kHz offset
8485A P,A,T
8481D P,A,T
8663A P,A
26 Chapter 1
Calibrating
Table 1-9 Recommended Test Equipment
Before You Start
Equipment
Wide Offset Phase Noise Signal Generator
Signal Generator (Option BAH)
Spectrum Analyzer, Microwave
Critical Specifications for Equipment
Substitution
Frequency Range: 1 GHz ±1 MHz Amplitude Range: 0 dBM
±5 dB
Phase Noise:
<−131 dBc/Hz typical @ 100 kHz offset <−145 dBc/Hz typical @ 1 MHz offset <−147 dBc/Hz typical @ 5 MHz offset <−149 dBc/Hz typical @ 10 MHz offset
Frequency Range: 900 MHz to 1800 MHz Amplitude Range: –30 to 0 dBm Phase Error: <0.5
°
Frequency Error: <2.5 Hz
Frequency Range: 100 kHz to 7 GHz Relative Amplitude Accuracy: 100 kHz to 3.0 GHz: <±1.8 dB Frequency Accuracy: <±10 kHz at 7 GHz
Recommended
HP/Agilent Model
8665B P
E4433B
Option UN8, 1E5
8563E P,T
Use
P
a
Synthesized Sweeper (2 required)
Frequency Range: E4407B or E4408B: 10 MHz to
26.5 GHz All others: 10 MHz to 13.2 GHz Frequency Accuracy (CW): ±0.02% Leveling Modes: Internal and External Modulation Modes: AM Power Level Range: –40 to 16 dBm
Function Generator Frequency Range: 0.1 Hz to 15 MHz
Frequency Accuracy: ±0.02% Waveform: Triangle, Square, Sine
Attenuator/Switch Driver
Compatible with HP/Agilent 8494G and 8496G Programmable step attenuators
83630/40/50B
83620/30/B
40/50B
33120A or
3325B
11713A P
P, A ,T
P, A ,T
Chapter 1 27
Calibrating Before You Start
Table 1-9 Recommended Test Equipment
Equipment
Attenuator, 1 dB Step
Attenuator, 10 dB Step
Attenuator, 10 dB Fixed
Attenuator, 6 dB Fixed (2 required)
Critical Specifications for Equipment
Substitution
Attenuation Range: 0 to 11 dB Frequency Range: 50 MHz ±1 MHz Connectors: Type-N female Calibrated at 50 MHz with accuracy of 1 to 11 dB attenuation: ±0.010 dB.
Attenuation Range: 0 to 110 dB Frequency Range: 50 MHz ±1 MHz Connectors: Type-N female Calibrated at 50 MHz with accuracy of:
0 to 40 dB attenuation: ±0.020 dB 50 to 100 dB attenuation: ±0.065 dB 110 dB attenuation: ±0.075 dB
Nominal attenuation: 10 dB Frequency Range: dc to 3 GHz Connectors: Type-N(m) and Type-N(f)
Nominal attenuation: 6 dB Frequency Range: 50 MHz
±1 MHz
VSWR: <1.1: 1 at 50 MHz
Recommended
HP/Agilent Model
8494A/G P
8496A/G P
8491A
Option 010
8491A
Option 006
Use
P
P
a
Attenuator, 20 dB Fixed
Attenuator Interconnect Kit
Nominal attenuation: 20 dB Frequency Range: 100 kHz to 3 GHz VSWR: <1.2: 1 at
3 GHz
Mechanically and electrically connects HP/Agilent 8494A/G and HP/Agilent 8496A/G
a. P = Performance Test, A = Adjustment, T = Troubleshooting
Table 1-10 Recommended Accessories
Equipment
Directional Bridge Frequency Range: 5 MHz to 3 GHz
Critical Specifications for Accessory
Substitution
Directivity: >40 dB Coupling factor: 16 dB nominal Insertion Loss: 2 dB maximum
8491A
Option 020
11716 Series
Recommended
HP/Agilent Model
86205A P
P
Use
a
28 Chapter 1
Calibrating
Table 1-10 Recommended Accessories
Before You Start
Equipment
Power Splitter (for E4401B/ 02B/03B/11B)
Power Splitter (for E4404B/ 05B/07B/08B)
Directional Coupler (for E4404B/05B/ 07B/08B)
Termination, 50
(2 required for Option 1DN)
Critical Specifications for Accessory
Substitution
Frequency Range: 9 kHz to 13.2 GHz Insertion Loss: 6 dB nominal Output Tracking: <0.25 dB Equivalent Output SWR: <1.22:1
Frequency Range: 9 kHz to 26.5 GHz Insertion Loss: 6 dB nominal Output Tracking: <0.25 dB Equivalent Output SWR: <1.22:1
Frequency Range: 2 GHz to 15 GHz Directivity: >16 dB Max.VSWR: 1.35:1 Transmission Arm Loss: <1.5 dB (nominal) Coupled Arm Loss: ~ 10 dB (nominal)
Impedance: 50
nominal
Connector: Type-N (m)
Recommended
HP/Agilent Model
11667A P,A
11667B
87300B
909A P,T
Use
a
Termination, 50
Termination, 75
Impedance: 50 (nominal) Connector:
(Option 1DQ, 1DP)
50 MHz Low Pass Filter
300 MHz Low Pass Filter
1 GHz Low Pass Filter
BNC (m)
Impedance: 75
(nominal) (2 required
for Option 1DQ) (1 required for Option 1DP)
Cutoff Frequency: 50 MHz Rejection at 65 MHz: Rejection at 75 MHz:
>40 MHz >60 dB
Cutoff Frequency: 300 MHz Rejection at >435 MHz:
>45 dB
Cutoff Frequency: 1 GHz Rejection at >2 GHz:
>60 dB
11593A P,A
909E
P, T
Option 201
0955-0306 P
0955-0455 P
0955-0487 P
Chapter 1 29
Calibrating Before You Start
Table 1-10 Recommended Accessories
Equipment
1.8 GHz Low Pass Filter (for
Critical Specifications for Accessory
Substitution
Cutoff Frequency: 1.8 GHz Rejection at >3 GHz:
>45 dB
E4404/5/7/8B)
4.4 GHz Low Pass Filter (for
Cutoff Frequency: 4.4 GHz Rejection at >5.5 GHz: >42 dB
E4404/5/7/8B)
a. P = Performance Test, A = Adjustment, T = Troubleshooting
Recommended
HP/Agilent Model
0955-0491
(2 required)
9135-0005 or
360D
(2 required)
Use
P
P
a
30 Chapter 1
Calibrating
Table 1-11 Recommended Adapters
Before You Start
Recommended
Critical Specifications for Adapter Substitution
HP/Agilent
Use
Model
BNC (m) to BNC (m) 1250-0216 P,T
BNC tee (f,m,f) 1250-0781 A,T
Type-N (f) to APC 3.5 (f) 1250-1745 P,A,T
Type-N (f) to BNC (m) 1250-1477 P,T
Type-N (f) to BNC (m), 75
(2 required for Option 1DQ)
1250-1534 P,A,T
(1 required for Option 1DP)
Type-N (m) to BNC (f) (4 required) 1250-1476 P,A,T
Type-N (m) to BNC (m) (2 required) 1250-1473 P,T
Type-N (m) to BNC (m), 75
(Option 1DP) 1250-1533 P,A,T
Type-N (m) to Type-N (m) 1250-1472 P,T
Type-N (m) to Type-N (m) 1250-1475 P,A,T
Type-N (f) to Type-N (f), 75
Type-N (f), 75
, to Type-N (m), 50 Ω (Option 1DP) 1250-0597 P,A,T
(Option 1DP) 1250-1529 P,A,T
a
Type-N (m) to SMA (m) 1250-1636 P
50 to 75
Minimum Loss
11852B P,A,T Frequency Range: dc to 1.5 GHz Insertion Loss: 5.7 dB, nominal (Option 1DP)
Type N(f) to Type N(f) 1250-0777
Type N(f) to BNC(f), 75 ohm (Option 1DP only) 1250-1535
Type N (m) to APC 3.5 (f) (3 required) 1250-1744
APC 3.5 (f) to APC 3.5 (f) 1250-1749
Dual Banana to BNC (f) 1251-2277 P,A,T
Type N (m) to BNC (f) (2 required) 1250-0780
a. P = Performance Test, A = Adjustment, T = Troubleshooting
Chapter 1 31
Calibrating Before You Start
Table 1-12 Recommended Cables
Recommended
Critical Specifications for Cable Substitution
HP/Agilent
Use
Model
Frequency Range: dc to 1 GHz Length:
122 cm (48 in)
10503A P,A,T
Connectors: BNC (m) both ends (4 required)
Type-N, 62 cm (24 in) 11500C P,T
Type-N, 152 cm (60 in) (2 required) 11500D P,A,T
Frequency Range: dc to 310 MHz
10502A P,T Length: 23 cm (9 in) Connectors: BNC (m) both ends
BNC, 75
Cable, Test Length:
, 30 cm (12 in) (Option 1DP) 5062-6452 P,A,T
85680-60093 T
91 cm (36 in)
Connectors: SMB (f) to BNC (m) (2 required)
APC 3.5 Cable
8120-4921 P,A,T Frequency: 9 kHz to 26.5 GHz Connectors: APC 3.5 (m) Length: >92 cm (36 in) (2 required)
a
a. P = Performance Test, A = Adjustment, T = Troubleshooting
32 Chapter 1

2 Performance Verification Tests

These tests verify the electrical performance of the analyzer. Allow the analyzer to warm up in accordance with the temperature stability specifications before performing the tests.
33
Performance Verification Tests

Tests included in this section:

Tests included in this section:
1. 10 MHz Reference Output Accuracy
2. 10 MHz High-Stability Frequency Reference Output Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option 1D5)
3. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4401B, E4402B, E4403B, and E4411B
4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B, E4407B, and E4408B
5. Frequency Span Accuracy: Agilent E4401B and E4411B
6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B
7. Noise Sidebands
8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407B
9. System-Related Sidebands
10. Residual FM
11. Sweep Time Accuracy
12. Display Scale Fidelity
13. Input Attenuation Switching Uncertainty
14. Reference Level Accuracy: Agilent E4401B and E4411B
15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B.
16. Resolution Bandwidth Switching Uncertainty
17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411B
18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B
19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411B
20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B
21. Resolution Bandwidth Accuracy
22. Frequency Response: Agilent E4401B and E4411B
34 Chapter 2
Performance Verification Tests
Tests included in this section:
23. Frequency Response, Agilent E4402B and E4403B
24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B
25. Frequency Response (Preamp On): Agilent E4401B
26. Frequency Response (Preamp On): Agilent E4402B
27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B
28. Other Input-Related Spurious Responses: Agilent E4401B and E4411B
29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B
30. Spurious Responses: Agilent E4401B and E4411B
31. Spurious Responses: Agilent E4402B and E4403B
32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B
33. Gain Compression: Agilent E4401B, E4402B, E4403B, and E4411B
34. Gain Compression: Agilent E4404B, E4405B, E4407B, and E4408B
35. Displayed Average Noise Level: Agilent E4401B and E4411B
36. Displayed Average Noise Level: Agilent E4402B and E4403B
37. Displayed Average Noise Level: Agilent E4404B and E4405B
38. Displayed Average Noise Level: Agilent E4407B and E4408B
39. Residual Responses
40. Fast Time Domain Amplitude Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option AYX)
41. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4401B and E4411B (Option 1DN or 1DQ)
42. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B (Option 1DN)
43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)
44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B (Option 1DN)
45. Tracking Generator Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option 1DN or 1DQ)
Chapter 2 35
Performance Verification Tests
Tests included in this section:
46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B (Option 1DN)
47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option 1DN or 1DQ)
48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B (Option 1DN)
49. Tracking Generator LO Feedthrough: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B(Option 1DN)
50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option 1D6)
51. Gate Mode Additional Amplitude Error: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option 1D6)
52. First LO OUTPUT Power Accuracy (Option AYZ only)
53. IF INPUT Accuracy (Option AYZ only)
54. Comms Frequency Response (Option BAC or BAH)
55. This Test Has Been Deleted (Modulation Accuracy
Rho)
56. This Test Has Been Deleted (CDMA Modulation Accuracy EVM)
57. This Test Has Been Deleted (CDMA Code Domain Power)
58. GSM Phase and Frequency Error (Options BAH and B7E)
59. Comms Absolute Power Accuracy (Options BAC or BAH)
36 Chapter 2
Performance Verification Tests
Tests included in this section:

Calibration

To perform calibration:
1. Run all performance verification tests listed in Column 1 of Table 2-1 that are applicable to the analyzer being calibrated and the installed options. Refer to the Performance Verification test tables in Chapter 1 for a list of the tests that are applicable to the analyzer being calibrated.
2. If any of the performance verification tests fail, perform the appropriate calibration adjustments listed in Column 2 of Table 2-1 which corresponds to the failure.
3. Repeat all of the performance verification tests listed in Column 1 of
Table 2-1 if any calibration adjustments were made in step 2. This
will confirm that there is no interaction between adjustments which could negatively impact analyzer performance.
Table 2-1 lists the performance verification tests and adjustments
needed for calibration. The performance tests are located in the following pages of this chapter. Adjustment information is located in the service guide.
Chapter 2 37
Performance Verification Tests
Tests included in this section:
Table 2-1 Calibration Requirements
Tes t
1. 10 MHz Reference Output
2. 10 MHz High-Stability Frequency
3. Frequency Readout and Marker
4. Frequency Readout and Marker
5. Frequency Span Accuracy None
6. Frequency Span Accuracy None
7. Noise Sidebands None
8. Noise Sidebands - Wide Offsets None
9. System-Related Sidebands IF Amplitude
10. Residual FM None
11. Sweep Time Accuracy None
Performance Verification Tests Calibration Adjustments
#
10 MHz Frequency Reference
Accuracy
Reference Output Accuracy
Frequency Count Accuracy
Frequency Count Accuracy
Adjustment
10 MHz Frequency Reference Adjustment
None
None
12. Display Scale Fidelity IF Amplitude
13. Input Attenuation Switching Uncertainty
14. Reference Level Accuracy IF Amplitude
15. Reference Level Accuracy IF Amplitude
16. Resolution Bandwidth Switching Uncertainty
17. Absolute Amplitude Accuracy (Reference Settings)
18. Absolute Amplitude Accuracy (Reference Settings)
19. Overall Absolute Amplitude Accuracy
20. Overall Absolute Amplitude Accuracy
21. Resolution Bandwidth Accuracy IF Amplitude
22. Frequency Response Frequency Response
50 MHz Amplitude Reference
IF Amplitude
None
None
Frequency Response Adjustment
None
23. Frequency Response Frequency Response
38 Chapter 2
Table 2-1 Calibration Requirements
Performance Verification Tests
Tests included in this section:
Tes t
24. Frequency Response Frequency Response
25. Frequency Response (Preamp On) Frequency Response
26. Frequency Response (Preamp On) Frequency Response
27. Frequency Response (Preamp On) Frequency Response
28. Other Input-Related Spurious
29. Other Input-Related Spurious
30. Spurious Responses None
31. Spurious Responses None
32. Spurious Responses None
33. Gain Compression None
34. Gain Compression None
35. Displayed Average Noise Level Frequency Response
Performance Verification Tests Calibration Adjustments
#
None
Responses
None
Responses
36. Displayed Average Noise Level Frequency Response
37. Displayed Average Noise Level Frequency Response
38. Displayed Average Noise Level Frequency Response
39. Residual Responses None
40. Fast Time Domain Amplitude Accuracy
41. Tracking Generator Absolute Amplitude and Vernier Accuracy
42. Tracking Generator Absolute Amplitude and Vernier Accuracy
43. Tracking Generator Level Flatness Tracking Generator ALC and
44. Tracking Generator Level Flatness Tracking Generator ALC and
45. Tracking Generator Harmonic Spurious Outputs
46. Tracking Generator Harmonic Spurious Outputs
None
Tracking Generator ALC and Tracking Generator Frequency Slope
Tracking Generator ALC and Tracking Generator Frequency Slope
Tracking Generator Frequency Slope
Tracking Generator Frequency Slope
None
None
Chapter 2 39
Performance Verification Tests
Tests included in this section:
Table 2-1 Calibration Requirements
Tes t
47. Tracking Generator Non-Harmonic
48. Tracking Generator Non-Harmonic
49. Tracking Generator L.O.
50. Gate Delay Accuracy and Gate
51. Gate Mode Additional Amplitude
52. First LO OUTPUT Power Accuracy LO Power
53. IF INPUT Accuracy IF INPUT Correction
54. Comms Frequency Response Frequency Response Error Correction
55. Deleted None
56. Deleted None
Performance Verification Tests Calibration Adjustments
#
None
Spurious Outputs
None
Spurious Outputs
LO Power
Feedthrough
None
Length Accuracy
None
Error
57. Deleted None
58. GSM Phase and Frequency Error None
59. Comms Absolute Power Accuracy IF Amplitude Adjustment
Manual use of Agilent 8494G and Agilent 8496G Attenuators with the Agilent 11713A
When using the programmable versions of the 1 dB and 10 dB step attenuator (Agilent 8494G and Agilent 8496G), the Agilent 11713A Attenuator/ Switch Driver must be used to control the attenuators. The Agilent 8494G 1 dB step attenuator should be connected as Attenuator X and the Agilent 8496G 10 dB step attenuator should be connected as Attenuator Y.
Use Table 2-2 to determine which of the Attenuator X and Attenuator Y settings are use to set the step attenuators to the desired value. In the columns labeled Attenuator X and Attenuator Y, a “1” indicates that section is on (the LED in the button will be lit), while a “0” indicates that section is off. For example, if the 1 dB step attenuator should be set to 2 dB and the 10 dB step attenuator should be set to 60 dB, sections 2, 6, and 7 should be on (lit) and all other sections should be off.
40 Chapter 2
Performance Verification Tests
Tests included in this section:
Table 2-2 Agilent 11713A Settings for Agilent 8494G and Agilent 8496G
1 dB Step
Atten
(dB)
0 0000 0 0000
1 1000 10 1000
2 0100 20 0100
3 1100 30 1100
4 0010 40 0010
5 1010 50 1010
6 0110 60 0110
7 1110 70 1110
8 0011 80 0011
9 1011 90 1011
10 0111 100 0111
11 1111 110 1111
Attenuator X 10 dB
Step
1234 5678
Atten
(dB)
Attenuator Y
Chapter 2 41
Performance Verification Tests

1. 10 MHz Reference Output Accuracy

1. 10 MHz Reference Output Accuracy
The setability is measured by changing the settings of the digital-to-analog converter (DAC), which controls the frequency of the timebase. The difference in frequency for each DAC step is calculated and compared to the specification.
The related adjustment for this performance verification test is the “10 MHz Reference Frequency Adjustment.”

Equipment Required

Universal counter (Instructions are for Agilent 53132A. For Agilent 5316B, refer to its user documentation.) Frequency standard Cable, BNC, 122-cm (48-in) (2 required)
Figure 2-1 10 MHz Reference Test Setup

Procedure

1. Connect the equipment as shown in Figure 2-1. The frequency standard provides the reference for the universal counter.
2. Check that the analyzer is not in external reference mode. If Ext Ref appears on the screen, the analyzer is in external reference mode. If the analyzer is in external reference mode, disconnect the external reference.
3. Ensure that the analyzer has been on and in internal frequency mode
for at least five minutes before proceeding.
4. Set the universal counter controls as follows:
a. Press
b. Press any one of the arrow keys until TIME is displayed.
c. Press
42 Chapter 2
Gate & ExtArm.
Gate & ExtArm again. Using the arrow keys, set the time to
10 s.
Performance Verification Tests
1. 10 MHz Reference Output Accuracy
d. Press
e. On Channel 1, press
f. On Channel 1, press
g. On Channel 1, press
Enter.
50 /1 M until the LED is lit.
x10 Attenuator until the LEd is extinguished.
AC/DC until the LED next to DC is
extinguished.
h. On Channel 1, press
i. On Channel 1, press
100 kHz Filter until the LED is extinguished.
Trigger/Sensitivity until Auto Trig is
displayed.
j. Use the arrow keys to toggle to off.
k. Press
Freq & Ratio.
5. Wait for the universal counter reading to stabilize. Record the universal counter reading in Table 2-3 as Counter Reading 1 with
0.1 Hz resolution.
6. Set the analyzer by pressing the following keys:
System, Alignments, Timebase, Fine
7. Record the number in the active function block of the analyzer in
Table 2-3 as Timebase Fine.
8. Press the
(up arrow) key on the analyzer.
9. Wait for the frequency counter reading to stabilize. Record the frequency counter reading in Table 2-3 as Counter Reading 2 with
0.1 Hz resolution.
10.Press the
(down arrow) key on the analyzer 2 times.
11.Wait for the frequency counter reading to stabilize. Record the frequency counter reading in Table 2-3 as Counter Reading 3 with
0.1 Hz resolution.
12.Press initial values. Press the
Preset on the analyzer to return the DAC settings to their
Factory Preset softkey, if it is displayed.
13.Subtract Counter Reading 1 from Counter Reading 2 and record the difference in Table 2-3 as the Positive Frequency Change.
Positive Frequency Change Counter Reading 2 Counter Reading 1=
14.Subtract Counter Reading 3 from Counter Reading 1 and record the difference in Table 2-3 as the Positive Frequency Change.
Negative Frequency Change Counter Reading 1 Counter Reading 3=
15.Of the Positive Frequency Change and negative frequency change values recorded in Table 2-3, record the largest value in Table 2-3 as
Chapter 2 43
Performance Verification Tests
1. 10 MHz Reference Output Accuracy
the maximum frequency change.
16.Divide the maximum frequency change by two and record the result as the settability.
Table 2-3 10 MHz Reference Accuracy Worksheet
Description Measurement
Counter Reading 1 ___________ Hz
Timebase Fine DAC ___________ Hz
Counter Reading 2 ___________ Hz
Counter Reading 3 ___________ Hz
Positive Frequency Change ___________ Hz
Negative Frequency Change ___________ Hz
Maximum Frequency Change ___________ Hz
Settability ___________ Hz
44 Chapter 2
Performance Verification Tests
2. 10 MHz High-Stability Frequency Reference Output Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option 1D5)
2. 10 MHz High-Stability Frequency Reference Output Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option 1D5)
This test applies only to analyzers equipped with the high-stability frequency reference, Option 1D5. If your analyzer does not have Option 1D5, perform the “10 MHz Reference Output Accuracy” test instead.
This test measures the warmup characteristics of the 10 MHz reference oscillator. The ability of the 10 MHz oscillator to meet its warmup characteristics gives a high level of confidence that it will also meet its yearly aging specification.
The universal counter is connected to the 10 MHz REF OUT. After the analyzer has been allowed to cool for at least 60 minutes, the analyzer is powered on. A frequency measurement is made five minutes after power is applied and the frequency is recorded. Another frequency measurement is made 10 minutes later (15 minutes after power is applied) and the frequency is recorded. A final frequency measurement is made 60 minutes after power is applied. The difference between each of the first two frequency measurements and the last frequency measurement is calculated and recorded.
The related adjustment for this performance verification test is the “10 MHz Reference Frequency Adjustment.”

Equipment Required

Universal counter (Instructions are for Agilent 53132A. For Agilent
5316B, refer to its user documentation.)
Frequency standard Cable, BNC, 122-cm (48-in) (2 required)
Figure 2-2 Frequency Reference Test Setup
Chapter 2 45
Performance Verification Tests
2. 10 MHz High-Stability Frequency Reference Output Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option 1D5)

Procedure

The analyzer must have been allowed to sit with the power off for at least 60 minutes before performing this procedure. This adequately simulates a cold start.
1. Allow the analyzer to sit with the power off for at least 60 minutes
before proceeding.
2. Switch the power to the analyzer on. Record the Power On Time below.
Power On Time___________
3. Connect the equipment as shown in Figure 2-2. The frequency standard provides the reference for the universal counter. Disconnect any cable to the 10 MHz REF INPUT of the analyzer.
4. Check that the analyzer is not in external reference mode. Ext Ref will appear on the display if the analyzer is in external reference mode. If the analyzer is in external reference mode, disconnect the external reference.
5. Set the universal counter controls as follows:
a. Press
Gate & ExtArm.
b. Press any one of the arrow keys until TIME is displayed.
c. Press
Gate & ExtArm again. Using the arrow keys, set the TIME to
10s.
d. Press
Enter
e. On Channel 1, press 50 /1 M to light the LED next to 50 Ω.
f. On Channel 1, press
g. On Channel 1, press
AC/DC to extinguish the LED next to DC.
x10 Attenuator to extinguish the LED next to
x10 Attenuator.
h. On Channel 1, press
100 kHz Filter to extinguish the LED next to
100 kHz Filter.
i. On Channel 1, press
Trigger/Sensitivity until Auto Trig is
displayed.
j. Use the arrows keys to toggle to off.
k. Press
Freq & Ratio
6. Continue with the next step 5 minutes after the Power On Time noted in step 2.
7. Wait for the universal counter reading to make at least two readings. Record the universal counter reading in Table 2-3 as Counter Reading 1 with 0.001 Hz resolution.
46 Chapter 2
Performance Verification Tests
2. 10 MHz High-Stability Frequency Reference Output Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option 1D5)
8. Continue with next step 15 minutes after the Power On Time noted in step 2.
9. Wait for the universal counter reading to make at least two readings. Record the universal counter reading in Table 2-3 as Counter Reading 2 with 0.001 Hz resolution.
10.Continue with next step 60 minutes after the Power On Time noted in step 2.
11.Wait for the universal counter reading to make at least two readings. Record the universal counter reading in Table 2-3 as Counter Reading 3 with 0.001 Hz resolution.
12.Calculate the 5 Minute Warm-up Error (in ppm) by subtracting Counter Reading 3 from Counter Reading 1 and dividing the result by 10.
5 Minute Warm-up Error Counter Reading 1 Counter Reading 3()10=
NOTE Dividing the frequency by 10 is equivalent to dividing the difference
first by 10 MHz (to normalize the difference to the reference frequency) and then multiplying by 1 x 106 to convert the result to parts-per-million (ppm).
13.Record the 5 Minute Warm-up Error (in ppm) in the performance verification test record as Test Record entry 1.
14.Calculate the 15 Minute Warm-up Error (in ppm) by subtracting Counter Reading 3 from Counter Reading 2 and dividing the result by 10.
15 Minute Warm-up Error Counter Reading 2 Counter Reading 3()10=
15.Record the 15 Minute Warm-up Error in the performance verification test record as Test Record entry 2.
Table 2-4 10 MHz Reference Accuracy Worksheet
Description Measurement
Counter Reading 1 ___________ Hz
Counter Reading 2 ___________ Hz
Counter Reading 3 ___________ Hz
Chapter 2 47
Performance Verification Tests

3. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4401B, E4402B, E4403B, and E4411B

3. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4401B, E4402B, E4403B, and E4411B
The frequency readout accuracy of the analyzer is tested with an input signal of known frequency. By using the same frequency standard for the analyzer and the synthesized sweeper, the frequency reference error is eliminated.
There are no related adjustment procedures for this performance test.

Equipment Required

Synthesized sweeper Adapter, Type N (f) to APC 3.5 (m) Cable, Type N, 183 cm (72 in) Cable, BNC, 122 cm (48 in)
Additional Equipment for 75 Ω Input
Pad, minimum loss Adapter, Type N (f) to BNC (m), 75

Procedure

Figure 2-3 Frequency Readout and Marker Frequency Accuracy Test
Setup
This performance test consists of two parts:
“Part 1: Frequency Readout Accuracy”
48 Chapter 2
Performance Verification Tests
3. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4401B, E4402B, E4403B, and E4411B
“Part 2: Marker Count Accuracy”
Perform “Part 1: Frequency Readout Accuracy” before “Part 2: Marker
Count Accuracy”.
Part 1: Frequency Readout Accuracy
1. Connect the equipment as shown in Figure 2-3. Remember to connect the 10 MHz REF OUT of the analyzer to the 10 MHz REF INPUT of the synthesized sweeper.
CAUTION Use only 75 cables, connectors, or adapters on instruments with 75
inputs, or the input connector will be damaged.
2. Perform the following steps to set up the equipment:
a. Press
INSTRUMENT PRESET on the synthesized sweeper, then set
the controls as follows:
CW, 1.490 GHz (Agilent E4401B and E4411B) CW, 1.5 GHz (Agilent E4402B and E4403B) POWER LEVEL, –10 dBm
b. Press Preset on the analyzer. Press the Factory Preset softkey, if it
is displayed. Set the analyzer by pressing the following keys:
FREQUENCY, 1.5 GHz SPAN, 20
MHz
3. Press Peak Search (or Search) on the analyzer to measure the frequency readout accuracy.
4. Record the marker frequency reading in the performance verification test record.
5. Repeat step 3 and step 4 above for the remaining spans listed in
Table 2-5.
Table 2-5 Frequency Readout Accuracy
Analyzer Span
(MHz)
Test Record Entry,
Actual
Marker Frequency
20 1)
10 2)
1 3)
Part 2: Marker Count Accuracy
Perform "Part 1: Frequency Readout Accuracy" before performing this procedure.
1. Press
Chapter 2 49
Preset on the analyzer. Press the Factory Preset softkey, if it is
Performance Verification Tests
3. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4401B, E4402B, E4403B, and E4411B
displayed. Set the analyzer to measure the marker count accuracy by pressing the following keys:
FREQUENCY, 1.490 GHz (Agilent E4401B and E4411B) FREQUENCY, 1.5 GHz (Agilent E4402B and E4403B) SPAN, 10 MHz BW/Avg Freq Count, Marker Count (On) Resolution (Man), 1 Hz
, Res BW, 100 kHz (Man)
2. Press Peak Search (or Search), then wait for a count be taken (it may take several seconds).
3. Record the counter (Cntr1) frequency reading as Test Record entry 4 of the performance verification test record.
4. On the analyzer, press
5. Press
Peak Search (or Search), then wait for a count be taken (it may
SPAN, 1 MHz.
take several seconds).
6. Record the counter (Cntr1) frequency reading as Test Record entry 5 of the performance verification test record.
50 Chapter 2
Performance Verification Tests

4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B, E4407B, and E4408B

4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B, E4407B, and E4408B
The frequency readout accuracy of the analyzer is tested with an input signal of known frequency. By using the same frequency standard for the analyzer and the synthesized sweeper, the frequency reference error is eliminated.
There are no related adjustment procedures for this performance test.

Equipment Required

Synthesized sweeper Adapter, Type N (m) to APC 3.5 (f) Adapter, APC 3.5 (f) to APC 3.5 (f) Cable, APC 3.5, 91 cm (36 in) Cable, BNC, 122 cm (48 in)

Additional Equipment for Option BAB

Adapter, APC 3.5 (f) to APC 3.5 (f)
Figure 2-4 Frequency Readout and Marker Count Accuracy Test Setup
Chapter 2 51
Performance Verification Tests
4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B, E4407B, and E4408B

Procedure

This performance verification test consists of two parts:
“Part 1: Frequency Readout Accuracy” “Part 2: Marker Count Accuracy”
Perform “Part 1: Frequency Readout Accuracy” before “Part 2: Marker Count Accuracy.”
Part 1: Frequency Readout Accuracy
1. Connect the equipment as shown in Figure 2-4. Remember to connect the 10 MHz REF OUT of the analyzer to the 10 MHz REF INPUT of the synthesized sweeper.
Option BAB only: Use the APC 3.5 adapter to connect the cable to the analyzer input.
2. Perform the following steps to set up the equipment:
a. Press
INSTRUMENT PRESET on the synthesized sweeper, then set
the controls as follows:
CW, 1.5 GHz POWER LEVEL
, –10 dBm
b. Press Preset on the analyzer. Press the Factory Preset softkey, if it
is displayed. Set the analyzer by pressing the following keys:
FREQUENCY, 1.5 GHz
20 MHz
SPAN,
3. Press Peak Search (or Search) on the analyzer to measure the frequency readout accuracy.
4. Record the marker (Mkr1) frequency reading in the performance verification test record as indicated in Table 2-6.
5. Change to the next analyzer span setting listed in
Table 2-6.
6. Repeat step 3 through step 5 for each analyzer frequency and span setting and synthesized sweeper CW frequency setting listed in
Table 2-6 for the analyzer being tested.
Table 2-6 Frequency Readout Accuracy
Synthesized
Sweeper CW
Frequency
(MHz)
1500 20 1.5 1)
52 Chapter 2
Analyzer
Span
(MHz)
Analyzer
Center
Frequency
(GHz)
Test
Record
Entry
Frequency
(GHz)
4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B,
Table 2-6 Frequency Readout Accuracy
Performance Verification Tests
E4407B, and E4408B
Synthesized
Sweeper CW
Frequency
(MHz)
1500 10 1.5 2)
1500 1 1.5 3)
4000 20 4.0 4)
4000 10 4.0 5)
4000 1 4.0 6)
9000 20 9.0 7)
9000 10 9.0 8)
9000 1 9.0 9)
16000 20 16.0 10)
Analyzer
Span
(MHz)
Stop here for Agilent E4404B.
Stop here for Agilent E4405B.
Analyzer
Center
Frequency
(GHz)
Frequency
Test
Record
Entry
(GHz)
16000 10 16.0 11)
16000 1 16.0 12)
21000 20 21.0 13)
21000 10 21.0 14)
21000 1 21.0 15)
Part 2: Marker Count Accuracy
Perform "Part 1: Frequency Readout Accuracy" before performing this procedure.
1. Press
Preset on the analyzer. Press the Factory Preset softkey, if it is
displayed. Set the analyzer to measure the marker count accuracy by pressing the following keys:
FREQUENCY, 1.5 GHz
, 20 MHz
SPAN BW/Avg Freq Count, Marker Count (On) Resolution (Man)
2. Press
, Res BW, 100 kHz (Man)
Peak Search (or Search), then wait for a count be taken (it may
take several seconds).
3. Record the counter (Cntr1) frequency reading in the performance
Chapter 2 53
Performance Verification Tests
4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B,
E4407B, and E4408B
verification test record as indicated in Table 2-7.
4. Repeat step 2 and step 3 for each analyzer center frequency and span setting and synthesized sweeper CW frequency setting listed in
Table 2-7 for the analyzer being tested.
Performance verification test “Frequency Readout Accuracy and Marker Count Accuracy” is now complete.
Table 2-7 Marker Count Accuracy
Synthesized Sweeper CW Frequency
MHz GHz MHz Test Record
1500 1.5 20 16)
1500 1.5 1 17)
4000 4.0 20 18)
4000 4.0 1 19)
9000 9.0 20 20)
9000 9.0 1 21)
16000 16.0 20 22)
16000 16.0 1 23)
21000 21.0 20 24)
Analyzer Center Frequency
Stop here for Agilent E4404B.
Stop here for Agilent E4405B.
Analyzer Span
Counter
Frequency
(Cntr1)
Entry
21000 21.0 1 25)
54 Chapter 2
Performance Verification Tests
5. Frequency Span Accuracy: Agilent E4401B and E4411B
5. Frequency Span Accuracy: Agilent E4401B and E4411B
For testing each frequency span, two synthesized sources are used to provide two precisely-spaced signals. The analyzer marker functions are used to measure this frequency difference.
There are no related adjustment procedures for this performance test.

Equipment Required

Synthesized sweeper Synthesized signal generator Power splitter Adapter, Type-N (m) to Type-N (m) Adapter, Type-N (f) to APC 3.5 (f) Cable, Type-N, 152-cm (60-in) (2 required) Cable, BNC, 122-cm (48-in)
Additional Equipment for 75 Ω Input
Pad, minimum loss Adapter, Type-N (f), to BNC (m), 75

Procedure

Full Span Frequency Span Readout Accuracy
1. Connect the equipment as shown in Figure 2-5. Note that the power splitter is used as a combiner. The synthesized signal generator provides the frequency reference for the synthesized sweeper.
2. Press displayed.
3. Press follows:
4. On the synthesized signal generator, set the controls as follows:
Preset on the analyzer. Press the Factory Preset softkey, if it is
PRESET on the synthesized sweeper and set the controls as
CW, 1350 MHz POWER LEVEL
FREQUENCY, 150 MHz AMPLITUDE
, –5 dBm
, 0 dBm
Chapter 2 55
Performance Verification Tests
5. Frequency Span Accuracy: Agilent E4401B and E4411B
Figure 2-5 Frequency Span Readout Accuracy Test Setup
5. Adjust the analyzer center frequency, if necessary, to place the lower frequency on the second vertical graticule line (one division from the left-most graticule line).
6. On the analyzer, press
Single. Wait for the completion of a new
sweep, then press the following keys:
Peak Search (or Search) Marker, Delta Peak Search Next Peak
(or Search)
The two markers should be on the signals near the second and tenth vertical graticule lines (the first graticule line is the left-most).
7. If necessary, continue pressing
Next Peak until the active marker is
on the right-most signal (1350 MHz).
8. Record the marker delta (
Mkr1) frequency reading as Test Record
entry 1 of the performance verification test record.
100 kHz and 100 MHz Frequency Span Readout Accuracy
1. Set the analyzer by pressing the following keys:
FREQUENCY, Start Freq, 10 MHz Stop Freq Sweep, Sweep
, 110 MHz
(Cont)
2. On the synthesized sweeper set the controls as follows:
56 Chapter 2
Performance Verification Tests
5. Frequency Span Accuracy: Agilent E4401B and E4411B
CW, 100 MHz POWER LEVEL
, –5 dBm
3. Set the synthesized signal generator controls as follows:
FREQUENCY, 20 MHz AMPLITUDE
, 0dBm
4. Adjust the analyzer center frequency to center the two signals on the display.
5. On the analyzer, press
Single. Wait for the completion of a new
sweep, then press the following keys:
Peak Search (or Search) Marker, Delta Peak Search Next Peak
(or Search)
The two markers should be on the signals near the second and tenth vertical graticule lines (the first graticule line is the left-most).
6. If necessary, continue pressing on the right-most signal (100 MHz). Record the marker delta (
Next Peak until the active marker is
Mkr1) frequency reading in the performance test record as Test Record entry 2.
7. Press
Marker, More, Marker All Off on the analyzer.
8. Change to the next equipment settings listed in Table 2-8.
9. On the analyzer, press Single. Wait for the completion of a newsweep, then press the following keys:
Peak Search (or Search) Marker, Delta Peak Search Next Peak
(or Search)
10.If necessary, continue pressing Next Peak until the marker delta is on the right-most signal. Record the marker delta ( reading in the performance test record.
11.Repeat step 7 through step 10 for the remaining analyzer span settings listed in Table 2-8.
Table 2-8 Frequency Span Readout Accuracy
Analyzer
Start
Frequency
(MHz)
0 1500 150 1350 1)1500
Chapter 2 57
Analyzer
Stop
Frequency
(MHz)
Synthesized
Signal
Generator
Frequency
(MHz)
Synthesized
Sweeper
Frequency
(MHz)
Mkr1) frequency
Test
Record
Entry
Analyzer
Span
(MHz)
Performance Verification Tests
5. Frequency Span Accuracy: Agilent E4401B and E4411B
Table 2-8 Frequency Span Readout Accuracy
Analyzer
Start
Frequency
(MHz)
10 110 20 100 2)100
10 10.1 10.01 10.09 3)0.1
800 900 810 890 4)100
800 800.1 800.01 800.09 5)0.1
1400 1500 1410 1490 6)100
1499 1499.1 1499.01 1499.09 7)0.1
Analyzer
Stop
Frequency
(MHz)
Synthesized
Signal
Generator
Frequency
(MHz)
Synthesized
Sweeper
Frequency
(MHz)
Test
Record
Entry
Analyzer
Span
(MHz)
58 Chapter 2
Performance Verification Tests

6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B

6. Frequency Span Accuracy:
Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B
For testing each frequency span, two synthesized sources are used to provide two precisely-spaced signals. The analyzer marker functions are used to measure this frequency difference.
There are no related adjustment procedures for this performance test.

Equipment Required

Synthesized sweeper Synthesized signal generator Power splitter Adapter, Type-N (m) to Type-N (m) Adapter, Type-N (f) to APC 3.5 (f) Cable, Type-N, 152-cm (60-in) (2 required) Cable, BNC, 122-cm (48-in)

Additional Equipment for Option BAB

Adapter, Type-N (m), to APC 3.5 (f)

Procedure

Full Span Frequency Span Readout Accuracy
1. Connect the equipment as shown in Figure 2-6. Note that the power splitter is used as a combiner. The synthesized signal generator provides the frequency reference for the synthesized sweeper.
2. Press displayed. Set the analyzer by pressing the following keys:
3. Press PRESET on the synthesized sweeper and set the controls as follows:
4. On the synthesized signal generator, set the controls as follows:
Preset on the analyzer. Press the Factory Preset softkey, if it is
FREQUENCY, Stop Freq, 3 GHz
CW, 2700 MHz POWER LEVEL
FREQUENCY, 300 MHz AMPLITUDE
, –5 dBm
, 0 dBm
Chapter 2 59
Performance Verification Tests
6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B
Figure 2-6 Frequency Span Readout Accuracy Test Setup
5. Adjust the analyzer center frequency, if necessary, to place the lower frequency on the second vertical graticule line (one division from the left-most graticule line).
6. On the analyzer, press
Single. Wait for the completion of a new
sweep, then press the following keys:
Peak Search (or Search) Marker, Delta Peak Search Next Peak
NOTE The two markers should be on the signals near the second and tenth
(or Search)
vertical graticule lines (the first graticule line is the left-most).
7. If necessary, continue pressing
Next Peak until the active marker is
on the right-most signal (2700 MHz).
8. Record the marker delta (
Mkr1) frequency reading as Test Record
entry 1 of the performance verification test record.
100 kHz and 100 MHz Frequency Span Readout Accuracy
1. Set the analyzer by pressing the following keys:
FREQUENCY, Start Freq, 10 MHz Stop Freq Sweep, Sweep
, 110 MHz
(Cont)
2. On the synthesized sweeper set the controls as follows:
CW, 100 MHz
60 Chapter 2
Performance Verification Tests
6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B
POWER LEVEL, –5 dBm
3. Set the synthesized signal generator controls as follows:
FREQUENCY, 20 MHz AMPLITUDE
, 0dBm
4. Adjust the analyzer center frequency to center the two signals on the display.
5. On the analyzer, press
Single. Wait for the completion of a new
sweep, then press the following keys:
Peak Search (or Search) Marker, Delta Peak Search Next Peak
(or Search)
The two markers should be on the signals near the second and tenth vertical graticule lines (the first graticule line is the left-most).
6. If necessary, continue pressing
Next Peak until the active marker is
on the right-most signal (100 MHz). Record the marker delta ( Mkr1) frequency reading in the performance test record as Test Record entry 2.
7. Press
Marker, More, Marker Off on the analyzer.
8. Change to the next equipment settings listed in Table 2-9.
9. On the analyzer, press
Single. Wait for the completion of a new
sweep, then press the following keys:
Peak Search (or Search) Marker, Delta Peak Search Next Peak
(or Search)
10.If necessary, continue pressing Next Peak until the marker delta is on the right-most signal. Record the marker delta (
Mkr1) frequency
reading in the performance test record.
11.Repeat step 7 through step 10 for the remaining analyzer span settings listed in Table 2-9.
Chapter 2 61
Performance Verification Tests
6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B
Table 2-9 Frequency Span Readout Accuracy
Analyzer
Start
Frequency
(MHz)
0 3000 300 2700 1) 3000
10 110 20 100 2) 100
10 10.1 10.01 10.09 3) 0.1
800 900 810 890 4) 100
800 800.1 800.01 800.09 5) 0.1
1400 1500 1410 1490 6) 100
1499 1499.1 1499.01 1499.09 7) 0.1
Analyzer
Stop
Frequency
(MHz)
Synthesized
Signal
Generator
Frequency
(MHz)
Synthesized
Sweeper
Frequency
(MHz)
Test
Record
Entry
Analyzer
Span
(MHz)
62 Chapter 2
Performance Verification Tests

7. Noise Sidebands

7. Noise Sidebands
A 1 GHz CW signal is applied to the input of the analyzer. The marker functions are used to measure the amplitude of the carrier and the noise level at 10 kHz, 20 kHz, and 30 kHz above and below the carrier. The 100 kHz offset from the carrier is also tested on Agilent E4401B and E4411B analyzers. For all other analyzers, the 100 kHz offset from the carrier is tested in performance test 8. Noise Sidebands - Wide
Offsets: Agilent E4402B, E4404B, E4405B and E4407B.
There are no related adjustment procedures for this performance test.

Equipment Required

Synthesized signal generator Cable, Type-N, 152-cm (60-in)
Additional Equipment for 75 Ω Input
Pad, minimum loss Adapter, Type-N (f), to BNC (m), 75

Additional Equipment for Option BAB

Adapter, Type-N (f), to APC 3.5 (f)
Figure 2-7 Noise Sidebands Test Setup
CAUTION Use only 75 cables, connectors, or adapters on instruments with 75
connectors, or the connectors will be damaged.
Chapter 2 63
Performance Verification Tests
7. Noise Sidebands

Procedure

This performance test consists of five parts:
Part 1: Noise Sideband Suppression at 10 kHz (measured in 1 kHz RBW) Part 2: Noise Sideband Suppression at 10 kHz (measured in 100 Hz RBW) Part 3: Noise Sideband Suppression at 20 kHz Part 4: Noise Sideband Suppression at 30 kHz Part 5: Noise Sideband Suppression at 100 kHz (Agilent E4401B and E4411B only)
• For E4401B or E4411B, do Parts 1, 3, 4 and 5.
• For E4402B, E4404B, E4405B or E4407B with Option 1DR and serial prefixes greater than or equal to US4510, SG4510, or MY4510, do Parts 2, 3 and 4.
• For all other ESA series analyzers do Parts 1, 3 and 4.
A worksheet is provided at the end of this procedure for calculating the noise sideband suppression.
Part 1: Noise Sideband Suppression at 10 kHz (measured in 1 kHz RBW)
1. Perform the following steps to set up the equipment:
Set the synthesized signal generator controls as follows:
FREQUENCY, 1000 MHz AMPLITUDE AMPLITUDE, 6 dBm (75 Ω Input only) AM OFF FM
OFF
, 0dBm (50 Input only)
2. Connect the equipment as shown in Figure 2-7.
3. Press
Preset on the analyzer. Press the Factory Preset softkey, if it is
displayed. Set the analyzer by pressing the following keys:
FREQUENCY, 1 GHz
, 10 MHz
SPAN AMPLITUDE
, Attenuation 10 dB (Man)
4. Press the following analyzer keys to measure the carrier amplitude:
Peak Search (or Search) FREQUENCY, Signal Track (On) SPAN, 50 kHz BW/Avg Video BW FREQUENCY, Signal Track (Off)
, 1kHz
, 30 Hz (Man)
64 Chapter 2
Performance Verification Tests
7. Noise Sidebands
Det/Demod, Detector, Sample Sweep, Sweep Time, 5 sec Single
Wait for the completion of a sweep, then press Peak Search (or
Search).
5. Press the following analyzer keys to measure the noise sideband level at 10 kHz:
Marker, Delta
, Function, Marker Noise (or Noise)
More AMPLITUDE, –10 dBm FREQUENCY, Center Freq, SPAN, Zero Span Single
CF Step, 10 kHz
Record the marker amplitude noise reading in Table 2-10 as the Upper Sideband Noise Level at 10 kHz.
6. Press the following analyzer keys to measure the noise sideband level at
10 kHz:
FREQUENCY, Center Freq, , Single
Record the marker amplitude noise reading in Table 2-10 as the Lower Sideband Noise Level at
7. Press
FREQUENCY, Center Freq,
10 kHz.
8. Continue with Part 3: Noise Sideband Suppression at 20 kHz.
Part 2: Noise Sideband Suppression at 10 kHz (measured in 100 Hz RBW)
1. Perform the following steps to set up the equipment:
Set the synthesized signal generator controls as follows:
FREQUENCY, 1000 MHz AMPLITUDE
, OFF
AM FM
, OFF
, 0dBm
2. Connect the equipment as shown in Figure 2-7.
3. Press
Preset on the analyzer. Press the Factory Preset softkey, if it is
displayed. Set the analyzer by pressing the following keys:
FREQUENCY, 1 GHz SPAN
, 10 MHz
AMPLITUDE
Chapter 2 65
, Attenuation 10 dB (Man)
Performance Verification Tests
7. Noise Sidebands
4. Press the following analyzer keys to measure the carrier amplitude:
Peak Search (or Search) FREQUENCY, Signal Track (On) SPAN, 50 kHz BW/Avg Video BW FREQUENCY, Signal Track (Off) Det/Demod, Detector, Sample Sweep, Sweep Time, 5 sec Single
, 100 Hz
, 30 Hz (Man)
Wait for the completion of a sweep, then press Peak Search (or
Search).
5. Press the following analyzer keys to measure the noise sideband level at 10 kHz:
Marker, Delta
, Function, Marker Noise (or Noise)
More AMPLITUDE, –10 dBm FREQUENCY, Center Freq, SPAN, Zero Span Single
CF Step, 10 kHz
Record the marker amplitude noise reading in Table 2-10 as the Upper Sideband Noise Level at 10 kHz.
6. Press the following analyzer keys to measure the noise sideband level at
10 kHz:
FREQUENCY, Center Freq, , Single
Record the marker amplitude noise reading in Table 2-10 as the Lower Sideband Noise Level at
7. Press
Preset on the analyzer. Press the Factory Preset softkey, if it is
10 kHz.
displayed. Set the analyzer by pressing the following keys:
FREQUENCY, 1 GHz
, 10 MHz
SPAN AMPLITUDE
, Attenuation 10 dB (Man)
8. Press the following analyzer keys to measure the carrier amplitude:
Peak Search (or Search) FREQUENCY, Signal Track (On) SPAN, 50 kHz BW/Avg Video BW FREQUENCY, Signal Track (Off) Det/Demod, Detector, Sample
, 1 kHz
, 30 Hz (Man)
66 Chapter 2
Performance Verification Tests
7. Noise Sidebands
Sweep, Sweep Time, 5 sec Single
Wait for the completion of a sweep, then press Peak Search (or
Search).
9. Press the following analyzer keys to measure the noise sideband level at 10 kHz:
Marker, Delta More
, Function, Marker Noise (or Noise)
AMPLITUDE, –10 dBm
10.Continue with Part 3: Noise Sideband Suppression at 20 kHz
Part 3: Noise Sideband Suppression at 20 kHz
1. Press the following analyzer keys to measure the noise sideband level at 20 kHz:
FREQUENCY, CF Step, 20 kHz Center Freq, Single
Record the marker amplitude noise reading in Table 2-10 as the Upper Sideband Noise Level at 20 kHz.
2. Press the following analyzer keys to measure the noise sideband level at
20 kHz:
FREQUENCY, Center Freq, , Single
Record the marker amplitude noise reading in Table 2-10 as the Lower Sideband Noise Level at
3. Press
↑.
20 kHz.
Part 4: Noise Sideband Suppression at 30 kHz
1. Press the following analyzer keys to measure the noise sideband level at 30 kHz:
FREQUENCY, CF Step, 30 kHz Center Freq, Single
Record the marker amplitude noise reading in Table 2-10 as the Upper Sideband Noise Level at 30 kHz.
2. Press the following analyzer keys to measure the noise sideband level at
Chapter 2 67
30 kHz:
Performance Verification Tests
7. Noise Sidebands
FREQUENCY, , Single
Record the marker amplitude noise reading in Table 2-10 as the Lower Sideband Noise Level at
30 kHz.
3. Press
↑.
4. If the analyzer is an Agilent E4401B or E4411B, continue with Part
5: Noise Sideband Suppression at 100 kHz (Agilent E4401B and E4411B only). Otherwise, continue with the next step.
5. In Table 2-10, record the more positive values (either Upper Noise Sideband Level, or Lower Noise Sideband Level) at the three offset frequencies and record them in the table as the Worst-Case offsets. Record these Worst-Case offset values in the performance verification test record in locations 1, 2 and 3 respectively.
6. This completes the Noise Sidebands test for the Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B analyzers.
Part 5: Noise Sideband Suppression at 100 kHz (Agilent E4401B and E4411B only)
1. Press the following analyzer keys to measure the noise sideband level at 100 kHz:
CF Step, 98 kHz Center Freq, Single
2. Press Peak Search (or Search) and record the marker amplitude noise reading in Table 2-10 as the Upper Sideband Noise Level at 100 kHz.
NOTE A spur may exist at 100 kHz offset from the carrier. Measuring at ±98
kHz offset from the carrier will yield a noise sideband level worse than the reading at ±100 kHz offset.
3. Press the following analyzer keys to measure the noise sideband level at
100 kHz:
FREQUENCY, , Single
4. Press Peak Search (or Search) and record the marker amplitude noise reading in Table 2-10 as the Lower Sideband Noise Level at
100
kHz.
5. In Table 2-10, record the more positive values (either Upper Noise Sideband Level, or Lower Noise Sideband Level) at the four offset frequencies and record them in the table as the Worst-Case offsets. Record these Worst-Case offset values in the performance
68 Chapter 2
verification test record in locations 1, 2, 3 and 4 respectively.
6. This completes the Noise Sidebands test for the Agilent E4401B or E4411B analyzer.
Table 2-10 Noise Sidebands Worksheet
Noise Sideband Amplitude
Performance Verification Tests
7. Noise Sidebands
Offset (kHz)
10 kHz 1)
20 kHz 2)
30 kHz 3)
100 kHz
a
Upper (dBc/Hz)
Lower (dBc/Hz)
Worst (dBc/Hz)
Test Record Entry
4)
a. This offset frequency is tested on Agilent E4402B,
E4404B, E4405B, and E4407B analyzers as part of perfor­mance test 8. Noise Sidebands - Wide Offsets: Agilent
E4402B, E4404B, E4405B and E4407B.
Chapter 2 69
Performance Verification Tests

8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407B

8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407B
A 1 GHz CW signal is applied to the input of the analyzer. The marker functions are used to measure the amplitude of the carrier and the noise level at 100 kHz, 1 MHz, 5 MHz, and 10 MHz above and below the carrier.
There are no related adjustment procedures for this performance test.

Equipment Required

Wide offset phase noise signal generator Cable, Type-N, 152-cm (60-in)

Additional Equipment for Option BAB

Adapter, Type-N (f), to APC 3.5 (f)
Figure 2-8 Noise Sidebands Test Setup

Procedure

This performance test consists of four parts:
Part 1: Noise Sideband Suppression at 100 kHz Part 2: Noise Sideband Suppression at 1 MHz Part 3: Noise Sideband Suppression at 5 MHz Part 4: Noise Sideband Suppression at 10 MHz
Perform part 1 before performing parts 2-4 of this procedure.
A worksheet is provided at the end of this procedure for calculating the
70 Chapter 2
Performance Verification Tests
8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407B
noise sideband suppression.
Part 1: Noise Sideband Suppression at 100 kHz
1. Perform the following steps to set up the equipment:
Set the synthesized signal generator controls as follows:
FREQUENCY, 1 GHz AMPLITUDE AM
OFF
OFF
FM
, 0dBm
2. Connect the equipment as shown in Figure 2-8.
3. Press
Preset on the analyzer. Press the Factory Preset softkey, if it is
displayed. Press the following keys on the analyzer:
FREQUENCY, 1 GHz
, 1MHz
SPAN
4. Press the following analyzer keys to measure the carrier amplitude:
Peak Search (or Search) FREQUENCY, Signal Track (On) SPAN, 10 kHz FREQUENCY Det/Demod, Detector, Sample BW/Avg
, Signal Track (Off)
, 1kHz
5. Adjust the signal generator AMPLITUDE to place the signal peak within 0.5 dB of the reference level on the analyzer. Press Single and then press
Peak Search (or Search).
6. Press the following analyzer keys:
Marker, Delta More
, Function, Marker Noise (or Noise)
AMPLITUDE, –40 dBm SPAN, Zero Span
7. Press BW/Avg, Average On 10, Enter to set the analyzer to video average 10 sweeps.
8. Press the following analyzer keys to measure the noise sideband level at 100 kHz:
FREQUENCY, CF Step, 99.8 kHz Center Freq, Single
Wait for the analyzer to average ten sweeps.
9. Press
Peak Search (or Search) and record the marker amplitude noise
reading in Table 2-11 as the Upper Sideband Noise Level at 100 kHz.
Chapter 2 71
Performance Verification Tests
8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407B
NOTE A spur may exist at 100 kHz offset from the carrier. Measuring at ±98
kHz offset from the carrier will yield a noise sideband level worse than the reading at ±100 kHz offset.
Press the following analyzer keys to measure the noise sideband level at
100 kHz:
FREQUENCY, , Single
Wait for the analyzer to average ten sweeps.
10.Press
Peak Search (or Search) and record the marker amplitude noise
reading in Table 2-11 as the Lower Sideband Noise Level at
100 kHz.
11.Press
↑.
Part 2: Noise Sideband Suppression at 1 MHz
1. Press the following analyzer keys to measure the noise sideband level at 1 MHz:
FREQUENCY, CF Step, 980 kHz Center Freq, Single
Wait for the analyzer to average ten sweeps.
2. Press
Peak Search (or Search) and record the marker amplitude noise
reading in Table 2-11 as the Upper Sideband Noise Level at 1 MHz.
NOTE A spur may exist at 1 MHz offset from the carrier. Measuring at ±980
kHz offset from the carrier will yield a noise sideband level worse than the reading at ±1 MHz offset.
Press the following analyzer keys to measure the noise sideband level at
1 MHz:
FREQUENCY, , Single
Wait for the analyzer to average ten sweeps.
3. Press
4. Press
72 Chapter 2
Peak Search (or Search) and record the marker amplitude noise
reading in Table 2-11 as the Lower Sideband Noise Level at
↑.
1 MHz.
Performance Verification Tests
8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407B
Part 3: Noise Sideband Suppression at 5 MHz
1. Press the following analyzer keys to measure the noise sideband level at 5 MHz:
FREQUENCY, CF Step, 5 MHz Center Freq, Single
Wait for the analyzer to average ten sweeps.
2. Press
Peak Search (or Search) and record the marker amplitude noise
reading in Table 2-11 as the Upper Sideband Noise Level at 5 MHz.
3. Press the following analyzer keys to measure the noise sideband level at
5 MHz:
FREQUENCY, , Single
Wait for the analyzer to average ten sweeps.
4. Press reading in Table 2-11 as the Lower Sideband Noise Level at
5. Press
Peak Search (or Search) and record the marker amplitude noise
5 MHz.
↑.
Part 4: Noise Sideband Suppression at 10 MHz
1. Press the following analyzer keys to measure the noise sideband level at 10 MHz:
FREQUENCY, CF Step, 9.99 MHz Center Freq, Single
Wait for the analyzer to average ten sweeps.
2. Press
Peak Search (or Search) and record the marker amplitude noise
reading in Table 2-11 as the Upper Sideband Noise Level at 10 MHz.
NOTE A spur may exist at 10 MHz offset from the carrier. Measuring at ±9.99
MHz offset from the carrier will yield a noise sideband level worse than the reading at ±10 MHz offset.
Press the following analyzer keys to measure the noise sideband level at
10 MHz:
FREQUENCY, , Single
Wait for the analyzer to average ten sweeps.
3. Press
Chapter 2 73
Peak Search (or Search) and record the marker amplitude noise
Performance Verification Tests
8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407B
reading in Table 2-11 as the Lower Sideband Noise Level at 10 MHz.
4. In Table 2-11, record the more positive values (either Upper Noise Sideband Level, or Lower Noise Sideband Level) at the four offset frequencies and record them in the table as the Worst-Case offsets. Record these Worst-Case offset values in the performance verification test record in locations 1, 2, 3 and 4 respectively.
Table 2-11 Noise Sidebands Worksheet
Noise Sideband Amplitude
Offset (kHz)
100 kHz 1)
1 MHz 2)
5 MHz 3)
10 MHz 4)
Upper (dBc/Hz)
Lower (dBc/Hz)
Worst (dBc/Hz)
Test Record Entry
74 Chapter 2
Performance Verification Tests

9. System-Related Sidebands

9. System-Related Sidebands
A 500 MHz CW signal is applied to the input of the analyzer. The marker functions are used to measure the amplitude of the carrier and the amplitude of any system-related sidebands more than 30 kHz away from the carrier. System-related sidebands are any internally generated sidebands related to the line, power supply or local oscillator.
There are no related adjustment procedures for this performance test.

Equipment Required

Synthesized signal generator Cable, Type-N, 152-cm (60-in)
Additional Equipment for 75 Ω Input
Pad, minimum loss Adapter, Type-N (f), to BNC (m), 75

Additional Equipment for Option BAB

Adapter, Type-N (f), to APC 3.5 (f)
Figure 2-9 System-related Sidebands Test Setup
CAUTION Use only 75 cables, connectors, or adapters on instruments with 75
connectors, or the connectors will be damaged.

Procedure

1. Perform the following steps to set up the equipment:
Chapter 2 75
Performance Verification Tests
9. System-Related Sidebands
a. Set the synthesized signal generator controls as follows:
FREQUENCY, 500 MHz
AMPLITUDE, 0dBm (50 Input only)
AMPLITUDE, 6 dBm (75 Input only) AM Off FM Off
b. Connect the equipment as shown in Figure 2-9.
c. Press
Preset on the analyzer. Press the Factory Preset softkey, if it
is displayed. Set the analyzer by pressing the following keys:
FREQUENCY, 500 MHz SPAN, 10 MHz
2. Set the analyzer to measure the system-related sideband above the signal by performing the following steps:
a. Press the following keys:
Peak Search (or Search) FREQUENCY, Signal Track (On) SPAN, 200 kHz BW/Avg, 1kHz Video BW, 30 Hz (Man)
Allow the analyzer to take two complete sweeps. Then press the following keys:
FREQUENCY, Signal Track (Off) CF Step, 130 kHz (Man)
b. Press
Peak Search (or Search), then Marker, Delta.
Single and wait for the completion of the sweep. Press
c. Press the following keys:
FREQUENCY
(step-up key)
3. Measure the system-related sideband above the signal by pressing
Single on the analyzer. Wait for the completion of a new sweep, then
Peak Search (or Search).
press
4. Record the marker delta amplitude as Test Record entry 1 of the performance verification test record.
5. Set the analyzer to measure the system-related sideband below the signal by pressing the following keys:
FREQUENCY
(step-down key) (step-down key)
6. Measure the system-related sideband below the signal by pressing
76 Chapter 2
Performance Verification Tests
9. System-Related Sidebands
Single. Wait for the completion of a new sweep, then press Peak Search (or Search).
Record the marker delta amplitude as Test Record entry 2 of the performance verification test record.
Chapter 2 77
Performance Verification Tests

10. Residual FM

10. Residual FM
This test measures the inherent short-term instability of the analyzer LO system. With the analyzer in zero span, a stable signal is applied to the input and slope detected on the linear portion of the IF bandwidth filter skirt. Any instability in the LO transfers to the IF signal in the mixing process. The test determines the slope of the IF filter in Hz/dB and then measures the signal amplitude variation caused by the residual FM. Multiplying these two values yields the residual FM in Hz.
For instruments having Option 1DR (Narrow Bandwidths), since the 10 Hz resolution bandwidth filter is digitally implemented, its slope is well known. The measured amplitude variation is simply multiplied by the known slope to yield the residual FM in a 10 Hz resolution bandwidth.
There are no related adjustment procedures for this performance test.

Equipment Required

Synthesized signal generator Cable, Type-N, 152-cm (60-in)
Additional Equipment for 75 Ω Input
Pad, minimum loss Adapter, Type-N (f), to BNC (m), 75

Additional Equipment for Option BAB

Adapter, Type-N (f), to APC 3.5 (f)
Figure 2-10 Residual FM Test Setup
78 Chapter 2
Performance Verification Tests
10. Residual FM
CAUTION Use only 75 cables, connectors, or adapters on instruments with 75
connectors, or the connectors will be damaged.

Procedure

This performance test consists of two parts:
“Part 1: Residual FM” “Part 2: Residual FM for Option 1DR and 1D5”
Perform “Part 2: Residual FM for Option 1DR and 1D5” in addition to Part 1, only if your analyzer is equipped with Option 1DR and 1D5.

Part 1: Residual FM

Determining the IF Filter Slope
1. Connect the equipment as shown in Figure 2-10.
2. Set the synthesized signal generator controls as follows:
FREQUENCY, 1000 MHz AMPLITUDE, –10 dBm (50 Input only) AMPLITUDE, –4 dBm (75 Input only) AM OFF FM OFF
3. Press Preset on the analyzer. Press the Factory Preset softkey, if it is displayed. Set the analyzer by pressing the following keys:
FREQUENCY, Center Freq, 1 GHz
, 1MHz
SPAN AMPLITUDE AMPLITUDE, 39.8 dBmV (75 Input only) Scale/Div, 2dB BW/Avg
, Ref Level, –9 dBm (50 Input only)
, Res BW, 1 kHz
4. On the analyzer, press the following keys:
Peak Search (or Search) SPAN, Span Zoom, 5kHz
Wait for the Span 5kHz message to appear, then press:
Peak Search (or Search), Marker , Marker Ref Lvl, Marker, Off
5. On the analyzer, press the following keys:
Single (Wait for the sweep to finish) Peak Search (or Search) Meas Tools, Delta
6. On the analyzer, rotate knob counterclockwise until the marker
Chapter 2 79
Performance Verification Tests
10. Residual FM
delta (Mkr1) amplitude reads −8dB ±0.3 dB.
7. Press delta (
If you have difficulty achieving the
Delta, then rotate the knob counterclockwise until the marker
Mkr1) reads −4dB ±0.3 dB.
±0.3 dB setting, then make the
following analyzer settings:
Sweep, Sweep (Cont) SPAN, 2kHz BW/Avg, Video BW, 30 Hz (Man)
Repeat step 5 through step 7.
8. Divide the marker delta ( delta (
Mkr1) amplitude in dB to obtain the slope of the resolution
bandwidth filter. For example, if the marker delta ( is 275 Hz and the marker delta (
Mkr1) frequency in Hertz by the marker
Mkr1) frequency
Mkr1) amplitude is 3.92 dB, the
slope would be equal to 70.15 Hz/dB. Record the result below:
Slope ____________ Hz/ dB
Measuring the Residual FM
9. On the analyzer, press:
Marker, Off Peak Search
(or Search) Meas Tools Delta
10.Rotate the knob counterclockwise until the marker delta ( amplitude reads
10 dB ±0.3 dB.
Mkr1)
11.On the analyzer, press the following keys:
Marker, Normal Marker Single BW/Avg SPAN Sweep 100 ms
→, Mkr CF
, Video BW (Man), 1kHz
, Zero Span
, Sweep Time
, Single
The displayed trace should be about five divisions below the reference level. If it is not, press
Sweep, Sweep (Cont), FREQUENCY,
and use the knob to place the displayed trace about five divisions below the reference level. Press
12.On the analyzer, press
Pk-Pk Search. Read the marker delta (Mkr1) amplitude, take its
Peak Search (or Search),
Single.
absolute value, and record the result as the Deviation.
Deviation ____________ dB
13.Calculate the Residual FM by multiplying the Slope recorded in
80 Chapter 2
Performance Verification Tests
10. Residual FM
step 8 by the Deviation recorded in step 12.
Record this value as Test Record entry 1 (Residual FM, 1 kHz Res BW) in the performance verification test record.

Part 2: Residual FM for Option 1DR and 1D5

Perform this additional procedure only if Option 1DR and Option 1D5 are present. Perform “Part 1: Residual FM” before performing this procedure.
1. Press
Preset on the analyzer. Press the Factory Preset softkey, if it is
displayed. Set the analyzer by pressing the following keys:
FREQUENCY, Center Freq, 1GHz SPAN
, 1MHz
AMPLITUDE AMPLITUDE, Ref Level, 39.8 dBmV (75 Input only) Scale/Div, 2dB
, Ref Level, –9 dBm (50 Input only)
2. On the analyzer press the following keys:
Peak Search (or Search) SPAN, Span Zoom, 5kHz
Wait for the “Span 5 kHz” message to appear. Press the following keys:
BW/Avg, Res BW, 10 Hz (Man) SPAN, 100 Hz
3. On the analyzer, press the following keys:
Peak Search (or Search) Marker , Mkr Ref Lvl Marker Peak Search (or Search) Meas Tools Delta
, Off
4. On the analyzer, rotate the knob counterclockwise until the marker delta (
Mkr1) amplitude reads −10 dB ±0.3 dB.
5. On the analyzer, press the following keys:
Marker, Normal Marker Single BW/Avg
SPAN,
Sweep Single
→, Mkr CF
, Video BW, 10 Hz (Man)
Zero Span
, Sweep Time, 20 ms
The displayed trace should be about five divisions below the
Chapter 2 81
Performance Verification Tests
10. Residual FM
reference level. If it is not, press Sweep, Sweep (Cont), FREQUENCY, and use the knob to place the displayed trace about five divisions below the reference level. Press
Single.
6. On the analyzer, press
Pk-Pk Search. Read the marker delta (Mkr1) amplitude, take its
Peak Search (or Search),
absolute value, and record the result as the Deviation.
Deviation ____________ dB
7. Calculate the Residual FM by multiplying the deviation recorded in
step 6 by 0.426 Hz/dB. This is the slope of the 10 Hz Res BW filter at
10 dB below the peak of the filter.
Record this value as Test Record entry 2 (Residual FM (10 Hz RBW)) in the performance verification test record.
82 Chapter 2
Performance Verification Tests

11. Sweep Time Accuracy

11. Sweep Time Accuracy
This test uses a function generator to amplitude-modulate a 500 MHz CW signal from another signal generator. The analyzer demodulates this signal in zero span to display the response in the time domain. The marker accuracy.
If the analyzer is equipped with Option AYX, also perform “Fast Time Domain Amplitude Accuracy” in addition to this procedure.
There are no related adjustment procedures for this performance test.

Equipment Required

function on the analyzer is used to read out the sweep time
Function generator Synthesized signal generator Cable, Type-N, 152-cm (60-in) Cable, BNC, 120-cm (48-in)
Additional Equipment for 75 Ω Input
Pad, minimum loss Adapter, Type-N (f), to BNC (m), 75

Additional Equipment for Option BAB

Adapter, Type-N (f), to APC 3.5 (f)
Figure 2-11 Sweep Time Accuracy Test Setup
CAUTION Use only 75 cables, connectors, or adapters on instruments with 75
Chapter 2 83
Performance Verification Tests
11. Sweep Time Accuracy
connectors, or the connectors will be damaged.

Procedure

1. Set the synthesized signal generator to output a 500 MHz, 10 dBm, CW signal. Set the AM and FM controls to off.
Input only: Set output level to 4dBm.
75
2. Set the function generator to output a 2 kHz, 1.14 Vp-p triangle waveform signal.
3. Connect the equipment as shown in Figure 2-11.
4. Press
Preset on the analyzer. Press the Factory Preset softkey, if it is
displayed. Set the analyzer by pressing the following keys:
FREQUENCY, 500 MHz SPAN, 10 MHz SPAN, Span Zoom, 50 kHz
5. Wait for the Span 50 kHz message to appear. Set Signal Track (Off).
6. Set the span to 0 Hz and set the analyzer as follows:
BW/Avg, Res BW,
Sweep, Sweep Time, 5 ms
3MHz
AMPLITUDE, Scale Type (Lin)
Peak Search (or Search) More, Search Parameters (or Search Criteria) Peak Excursion, 3 dB
Adjust the synthesized signal generator amplitude as necessary for a mid-screen display (marker amplitude should read approximately 110 mV).
7. Set the synthesized signal generator modulation source to EXT DC. Set AM ON at 90% modulation.
8. On the analyzer, press
Tri g then Video. Set the video trigger level to
110 mV (mid-screen).
9. On the analyzer, press
Peak Search (or Search), 0, s, Meas Tools, Next Pk Right. This is
press
Single. After the completion of the sweep,
the marked signal.
10.Press
Next Pk Right eight times so the delta marker is on the eighth signal
Marker, Delta, then Peak Search (or Search) and press
peak from the marked signal.
11.Read the marker delta (
84 Chapter 2
Mkr1) time. Calculate the sweeptime
Performance Verification Tests
11. Sweep Time Accuracy
accuracy as follows:
Mkr1 0.8 Sweep Time×()
Sweep Time Accuracy 100
NOTE The sweep time accuracy is defined as a percentage of the indicated
------------------------------------------------------------------------
×=
Sweep Time
sweep time, not of the indicated signal separation. Therefore, it is appropriate to divide the difference between the marker delta (
Mkr1)
reading and the nominal signal separation by the sweep time, rather than dividing by the nominal signal separation.
12.Record the calculated sweeptime accuracy in Table 2-12.
13.If the analyzer is not equipped with Option AYX, fast time domain sweeps, or Option B7D, DSP and Fast ADC, repeat step 9 through
step 12 only for sweeptime settings between 5 ms and 10 s as
indicated in Table 2-12. For each sweeptime setting, set the function generator to the frequency indicated in Table 2-12.
14.If the analyzer is equipped with Option AYX, fast time domain sweeps, or Option B7D, DSP and Fast ADC, repeat step 9 through
step 12 for all sweeptime settings as indicated in Table 2-12. For
each sweeptime setting, set the function generator to the frequency indicated in Table 2-12.
Table 2-12 Sweep Time Accuracy
Analyzer Sweep
Time Setting
5 ms 2.0 kHz 1)
20 ms 500.0 Hz 2)
100 ms 100.0 Hz 3)
1 s 10.0 Hz 4)
10 s 1.0 Hz 5)
The following entries only apply to analyzers equipped with Option AYX or B7D.
1 ms 10.0 kHz 6)
500
µs 20.0 kHz 7)
100 µs 100.0 kHz 8)
Synthesizer Function
Generator Frequency
Marker
Delta
Mkr1)
(
Reading
Sweep
Time
Accuracy
(%)
Tes t
Record
Entry
Chapter 2 85
Performance Verification Tests

12. Display Scale Fidelity

12. Display Scale Fidelity
A 50 MHz CW signal is applied to the input of the analyzer through two calibrated step attenuators. The attenuators are the amplitude reference standard. The source is adjusted for a response at the reference level. The attenuators are then set to achieve a nominal amplitude below the reference level. The analyzer amplitude marker is compared to the actual total attenuation to determine the scale fidelity error.
The test is performed in both log and linear amplitude scales.
The related adjustment for this performance test is “IF Amplitude.”

Equipment Required

Synthesized signal generator 1 dB step attenuator 10 dB step attenuator 6 dB fixed attenuator (2 required) Attenuator switch driver (if programmable step attenuators are used) Cable, Type-N 152-cm (60-in) (2 required) Cable, BNC 122-cm (48-in) Attenuator interconnect kit
Additional Equipment for 75 Ω Input
50 to 75 minimum loss pad Adapter, Type-N (f), to BNC (m), 75

Additional Equipment for Option BAB

Adapter, Type-N (f), to APC 3.5 (f)
86 Chapter 2
Figure 2-12 Scale Fidelity Test Setup
Performance Verification Tests
12. Display Scale Fidelity
CAUTION Use only 75 cables, connectors, and adapters on instruments with
75 connectors, or the connectors will be damaged.

Procedure

Calculate the Actual Attenuation Errors
1. From the calibration data supplied with the 10 dB step attenuator, enter into Column 4 of Table 2-13 through Table 2-17 the actual attenuation for the corresponding nominal attenuation settings. Enter data in Table 2-14, Table 2-16, and Table 2-17 only if the analyzer has Option 1DR.
If the calibration data does not indicate an actual attenuation value for the 0 dB setting, enter 0 dB.
NOTE The Agilent 8496G programmable attenuator has four attenuator
sections consisting of 10 dB, 20 dB, and 40 dB attenuators. If using the Agilent 8496G programmable attenuator, enter the calibration data for the section three, 40 dB step, rather than the section four, 40 dB step.
2. From the calibration data supplied with the 1 dB step attenuator, enter into Column 5 of Table 2-13 through Table 2-17 the actual attenuation for the corresponding nominal attenuation settings. Enter data in Table 2-14, Table 2-16, and Table 2-17 only if the analyzer has Option 1DR.
Chapter 2 87
Performance Verification Tests
12. Display Scale Fidelity
If the calibration data does not indicate an actual attenuation value for the 0 dB setting, enter 0 dB.
NOTE The Agilent 8494G programmable attenuator has four attenuator
sections consisting of 1 dB, 2 dB, 4 dB, and 4 dB attenuators. If using the Agilent 8494G programmable attenuator, enter the calibration data for the section three, 4 dB step, rather than the section four, 4 dB step.
3. For each row in Table 2-13 and Table 2-14, add the 10 dB and the 1 dB Step Attenuator Actual Attenuation values (Columns 4 and 5) and place the results into the Total Actual Attenuation (Column 6).
Total Actual Attenuation 1 dB Step Attenuator Actual Attenuation
=
10 dB Step Attenuator Actual Attenuation+
Example for –36 dB from REF LVL setting:
1 dB Step Attenuator Actual Attenuation 6 dB()5.998 dB=
10dB Step Attenuator Actual Attenuation 30 dB()30.012 dB=
Total Actual Attenuation 5.998dB 30.012 dB 36.010 dB =+=
4. Enter the total actual attenuation (0 dB from the reference level) below:
Total actual attenuation (0 dB from Ref Level) = __________ dB

Part 1: Log Display Scale Fidelity, Analog Bandwidths

Setup for Log Scale Measurement
1. Connect the equipment as indicated in Figure 2-12.
2. Press displayed. Set the analyzer by pressing the following keys:
Preset on the analyzer. Press the Factory Preset softkey, if it is
System, Alignments, Auto Align, Off FREQUENCY, 50 MHz SPAN, 45 kHz BW/Avg, 3 kHz BW/Avg, Video BW, 1 kHz
3. Preset the synthesized signal generator, then press Blue Key, Special,
0.0. Press the following keys:
FREQUENCY, 50 MHz AMPLITUDE,
+9 dBm
4. Set the 1 dB step attenuator to 0 dB.
5. Set the 10 dB step attenuator to 0 dB.
88 Chapter 2
Performance Verification Tests
12. Display Scale Fidelity
6. Press Peak Search (or Search) on the analyzer.
7. Adjust the synthesized signal generator amplitude until the analyzer marker amplitude reads 0 dBm
75
Input: Adjust the synthesized signal generator amplitude until
the analyzer marker reads 48.75 dBmV ±
NOTE Do not adjust the synthesized signal generator amplitude after the
±0.1 dB.
±0.1 dB.
reference is established.
CLFE Total Actual Attenuation Mkr Reading Total Actual Atten 0 dB from Ref Level()+=
8. On the analyzer, press
Marker, Delta.
Measure the Cumulative Log Fidelity
1. Perform step 2 to step 4 for each measurement value in
Table 2-13.
2. Set the 1 dB and 10 dB step attenuators as indicated in Column 2
and Column 3 of Table 2-13 for the various dB from REF LVL settings.
For settings of –64 dB and lower, press the following keys:
BW/Avg, Average (On) 5, Enter
3. Press Peak Search (or Search) on the analyzer and record the marker
delta (
Mkr1) reading in Column 7 of Table 2-13.
4. Calculate the Cumulative Log Fidelity Error (CLFE) as follows, and record the result in the performance verification test record as indicated in Column 8 of Table 2-13:
Table 2-13 Cumulative and Incremental Log Scale Fidelity Worksheet,
Analog Resolution Bandwidths Measured at 3 kHz
Column 1Column 2Column 3Column 4Column 5Column 6Column
7
dB from
REF LVL
(dB)
0 (Ref) 0 0 0 (Ref) 0 (Ref) NA
10 dB
Step
Atten
Nominal
Attenu-
ation
(dB)
1 dB
Step
Atten
Nominal
Attenu-
ation
(dB)
10 dB
Step
Atten
Actual
Attenu-
ation
(dB)
1 dB Step
Atten
Actual
Attenu-
ation
(dB)
Tot a l
Actual
Attenu-
ation
(dB)
Marker
Delta
( Mkr1)
Reading
(dB)
Chapter 2 89
Column 8Column
Test Record Entry –
CLFE
(dB)
9
Tes t Record Entry –
ILFE
(dB)
Performance Verification Tests
12. Display Scale Fidelity
Table 2-13 Cumulative and Incremental Log Scale Fidelity Worksheet,
Analog Resolution Bandwidths Measured at 3 kHz
Column 1Column 2Column 3Column 4Column 5Column 6Column
7
dB from
REF LVL
(dB)
–4 0 4 1) 22)
–8 0 8 2) 23)
–12 10 2 3) 24)
–16 10 6 4) 25)
–20 20 0 5) 26)
–24 20 4 6) 27)
–28 20 8 7) 28)
–32 30 2 8) 29)
–36 30 6 9) 30)
10 dB
Step
Atten
Nominal
Attenu-
ation
(dB)
1 dB Step
Atten
Nominal
Attenu-
ation
(dB)
10 dB
Step
Atten
Actual
Attenu-
ation
(dB)
1 dB Step
Atten
Actual
Attenu-
ation
(dB)
Tot a l
Actual
Attenu-
ation
(dB)
Marker
Delta
( Mkr1)
Reading
(dB)
Column 8Column
Test Record Entry –
Record Entry –
CLFE
(dB)
9
Tes t
ILFE
(dB)
–40 40 0 10) 31)
–44 40 4 11) 32)
–48 40 8 12) 33)
–52 50 2 13) 34)
–56 50 6 14) 35)
–60 60 0 15) 36)
–64 60 4 16) 37)
–68 60 8 17) 38)
–72 70 2 18) 39)
–76 70 6 19) 40)
–80 80 0 20) 41)
–84 80 4 21) NA
Calculate Incremental Log Fidelity
1. Calculate the Incremental Log Fidelity Error (ILFE) for dB from REF LVL settings of –4 dB to
80 dB using the current and previous
Cumulative Log Fidelity Errors (CLFEs):
90 Chapter 2
Performance Verification Tests
12. Display Scale Fidelity
ILFE CLFE(current) CLFE(previous)=
Example Calculation for ILFE at –20 dB from REFLVL setting:
Previous CLFE 16 dB from REF LVL()0.07 dB=
Current CLFE 20 dB from REF LVL()0.02 dB=
ILFE 20– dB()0.02 dB 0.07 dB()0.09 dB==
2. Record the result in the performance verification test record as indicated in Column 9 of Table 2-13.

Part 2: Log Display Scale Fidelity, Digital Bandwidths

This section is for analyzers with Option 1DR (narrow resolution bandwidths) only.
Setup for Log Scale Measurement
1. Set the following parameters on the analyzer:
SPAN, 150 Hz BW/Avg, 10 Hz BW/Avg, 100 Hz (ESA-L Series only) BW/Avg, Video BW, 3 Hz
(ESA-E Series only)
2. Press Peak Search (or Search) on the analyzer.
3. Adjust the amplitude of the synthesized signal generator until the analyzer marker amplitude reads 0 dBm
Input only: Adjust the amplitude of the synthesized signal
75
±0.1 dB.
generator until the analyzer marker amplitude reads
48.75 dBmV
±0.1 dB.
4. Set the 1 dB and the 10 dB step attenuators to 0 dB.
NOTE Do not adjust the synthesized signal generator amplitude after the
reference is established.
5. On the analyzer, press
Marker, Delta.
Measure the Cumulative Log Fidelity
1. Perform step 2 to step 4 for each measurement value in
Table 2-14.
2. Set the 1 dB and 10 dB step attenuators as indicated in Table 2-14
for the various dB from REF LVL settings.
For settings of –84 dB and lower, press the following keys:
Chapter 2 91
Performance Verification Tests
0 dB from R
12. Display Scale Fidelity
BW/Avg, Average (On) 5, Enter
3. Press Peak Search (or Search) on the analyzer and record the marker delta (
Mkr1) reading in Column 7 of Table 2-14.
4. Calculate the Cumulative Log Fidelity Error (CLFE) as follows and record the result in the performance verification test record as indicated in Column 8 of Table 2-14:
CLFE Total Actual Attenuation MkrReading Total Actual Atten
Table 2-14 Cumulative and Incremental Log Scale Fidelity Worksheet,
Option 1DR Narrow Resolution Bandwidths
Column 1Column 2Column 3Column 4Column 5Column 6Column 7Column 8Column
9
dB from
REF
LVL
(dB)
0 (Ref) 0 0 0 (Ref) 0 (Ref) NA
–4 0 4 43) 68)
–8 0 8 44) 69)
–12 10 2 45) 70)
10 dB
Step
Atten
Nominal
Attenu-
ation
(dB)
1 dB Step
Atten
Nominal
Attenu-
ation
(dB)
10 dB
Step
Atten
Actual
Attenu-
ation
(dB)
1 dB Step
Atten
Actual
Attenu-
ation
(dB)
Tot a l
Actual
Attenu-
ation
(dB)
Marker
Delta
( Mkr1)
Reading
(dB)
Test Record Entry –
CLFE
(dB)
Tes t Record Entry –
ILFE
(dB)
(+=
–16 10 6 46) 71)
–20 20 0 47) 72)
–24 20 4 48) 73)
–28 20 8 49) 74)
–32 30 2 50) 75)
–36 30 6 51) 76)
–40 40 0 52) 77)
–44 40 4 53) 78)
–48 40 8 54) 79)
–52 50 2 55) 80)
–56 50 6 56) 81)
–60 60 0 57) 82)
92 Chapter 2
Performance Verification Tests
12. Display Scale Fidelity
Table 2-14 Cumulative and Incremental Log Scale Fidelity Worksheet,
Option 1DR Narrow Resolution Bandwidths
Column 1Column 2Column 3Column 4Column 5Column 6Column 7Column 8Column
9
dB from
REF
LVL
(dB)
–64 60 4 58) 83)
–68 60 8 59) 84)
–72 70 2 60) 85)
–76 70 6 61) 86)
–80 80 0 62) 87)
–84 80 4 63) NA
–88 80 8 64) NA
–92 90 2 65) NA
–96 90 6 66) NA
–98 90 8 67) NA
10 dB
Step
Atten
Nominal
Attenu-
ation
(dB)
1 dB Step
Atten
Nominal
Attenu-
ation
(dB)
10 dB
Step
Atten
Actual
Attenu-
ation
(dB)
1 dB Step
Atten
Actual
Attenu-
ation
(dB)
Tot a l
Actual
Attenu-
ation
(dB)
Marker
Delta
( Mkr1)
Reading
(dB)
Test Record Entry –
CLFE
(dB)
Tes t Record Entry –
ILFE
(dB)
Calculate incremental log fidelity:
Calculate the Incremental Log Fidelity Error (ILFE) for dB from REF LVL settings of –4 dB to
80 dB using the current and previous
Cumulative Log Fidelity Errors (CLFEs):
ILFE CLFE current()CLFE previous()=
Record the result in the performance verification test record as indicated in Column 9 of Table 2-14.

Part 3: Linear Display Scale Fidelity, Analog Bandwidths

Setup for linear scale measurement:
1. Press displayed. Set the analyzer by pressing the following keys:
FREQUENCY, 50 MHz
Preset on the analyzer. Press the Factory Preset softkey, if it is
System, Alignments, Auto Align, Off
Chapter 2 93
Performance Verification Tests
12. Display Scale Fidelity
SPAN, 10 kHz BW/Avg, 3 kHz AMPLITUDE, Scale Type (Lin) Input/Output
(or Input), Input Z Corr (50) (75 Input only)
2. Preset the synthesized signal generator, by pressing
Special, 0, 0. Press the following keys:
FREQUENCY, 50 MHz AMPLITUDE, –3 dBm AMPLITUDE, 4 dBm (75 Input only)
(50 Input only)
Blue Key,
3. Set the 1 dB step attenuator to 0 dB.
4. Set the 10 dB step attenuator to 0 dB.
5. Press
Peak Search (or Search) on the analyzer.
6. Adjust the synthesized signal generator amplitude until the analyzer marker amplitude reads 223.6 mV
NOTE Do not adjust the amplitude of the synthesized signal generator after
±4 mV.
the reference is established.
Calculate ideal marker amplitude:
7. Considering Total Actual Attenuation at the 0 dB from REF LVL setting to be ATref, and the Total Actual Attenuation at any other dB from REF LVL setting to be ATmeas, calculate the Ideal Mkr Reading, in millivolts, as follows, and enter the result in Column 7 of
Table 2-15.
Ideal Mkr Reading mV()1000 0.05 10×
=
ATmeas–ATref+()10
For example, if ATref = 0.012 dB and ATmeas = 7.982, the Ideal Mkr Reading for the
8 dB from Ref Level setting would
be:
Ideal Mkr Reading(mV) 1000 0.05 10×
7.982–0.012+()10
89.3 mV==
Measure Linear Fidelity:
8. Perform step 9 to step 11 for each measurement value in
Table 2-15.
9. Set the 1 dB and 10 dB step attenuators as indicated in Column 2 and Column 3 of Table 2-15 for the dB from REF LVL settings.
10.Press delta (
Peak Search (or Search) on the analyzer and record the marker
Mkr1) amplitude reading as the actual Mkr reading in
Column 8 of
94 Chapter 2
Performance Verification Tests
12. Display Scale Fidelity
Table 2-15.
11.Calculate the Linear Fidelity Error (LFE) as a percentage of reference level (RL), and record the result in the performance verification test record as indicated in Column 9 of Table 2-15.
Actual Mkr Reading Ideal Mkr Reading
LFE % of RL()100
----------------------------------------------------------------------------------------------------
×=
223.6 mV
Example calculation for LFE(% of RL):
Actual Mkr Reading 85.0 mV=
Ideal Mkr Reading 89.3 mV=
85.0 89.3
LFE %of RL()100
---------------------------
×=
223.6
LFE %of RL()1.92% of RL=
Table 2-15 Linear Scale Fidelity Worksheet, Analog Resolution
Bandwidths Measured at 3 kHz
Column 1Column 2Column 3Column 4Column 5Column 6Column 7Column 8Column
9
dB from
REF
LVL
(dB)
0 (Ref) 0 0 0 (Ref) 0 (Ref) 0 (Ref)
–4 0 4 93)
–8 0 8 94)
–12 10 2 95)
–16 10 6 96)
–20 20 0 97)
10 dB
Step
Atten
Nominal
Attenu-
ation
(dB)
1 dB
Step
Atten
Nominal
Attenu-
ation
(dB)
10 dB
Step
Atten
Actual
Attenu-
ation
(dB)
1 dB Step
Atten
Actual
Attenu-
ation
(dB)
Tot a l
Actual
Attenu-
ation
(dB)
Ideal
Mkr
Reading
(mV)
Actual
Mkr
Reading
(mV)
Test Record Entry –
LFE
(% of
RL)

Part 4: Linear Display Scale Fidelity, Digital Bandwidths

This section is for analyzers with Option 1DR (narrow resolution bandwidths) only.
Chapter 2 95
Performance Verification Tests
12. Display Scale Fidelity
Setup for linear scale measurement:
1. Set the following parameters on the analyzer:
SPAN, 100, Hz (ESA-E Series only) SPAN, 300, Hz (ESA-L Series only) BW/Avg, 10 Hz (ESA-E Series only) BW/Avg, 100 Hz (ESA-L Series only) Video BW, 1 Hz
2. Press Peak Search (or Search) on the analyzer.
3. Set the 1 dB and 10 dB step attenuators to 0 dB.
4. Adjust the synthesized signal generator amplitude until the analyzer marker amplitude reads 223.6 mV
NOTE Do not adjust the synthesized signal generator amplitude after the
±4 mV.
reference is established.
Calculate ideal marker amplitude:
5. Considering Total Actual Attenuation at the 0 dB from REF LVL setting to be ATref and the Total Actual Attenuation at any other dB from REF LVL setting to be ATmeas, calculate the Ideal Mkr Reading, in millivolts, as follows and enter the result in Column 7 of
Table 2-16.
Ideal Mkr Reading mV()1000 0.05 10×
=
ATmeas–ATref+()10
Measure linear fidelity:
6. Perform step 7 to step 9 for each measurement value in Table 2-16.
7. Set the 1 dB and 10 dB step attenuators as indicated in Table 2-16 for the dB from REF LVL settings.
8. Press delta (
Peak Search (or Search) on the analyzer and record the marker
Mkr1) amplitude reading as the actual Mkr reading in
Column 8 of Table 2-16.
9. Calculate the Linear Fidelity Error (LFE) as a percentage of reference level (RL), and record the result in the performance verification test record as indicated in Column 9 of Table 2-16.
Actual Mkr Reading Ideal Mkr Reading
LFE %of RL()100
----------------------------------------------------------------------------------------------------
×=
223.6 mV
96 Chapter 2
Performance Verification Tests
12. Display Scale Fidelity
Table 2-16 Linear Scale Fidelity Worksheet, Option 1DR Narrow
Resolution Bandwidths
Column 1Column 2Column 3Column 4Column 5Column 6Column 7Column 8Column
9
dB from
REF
LV L
(dB)
0 (Ref) 0 0 0 (Ref) 0 (Ref) 0 (Ref)
–4 0 4 98)
–8 0 8 99)
–12 10 2 100)
–16 10 6 101)
–20 20 0 102)
10 dB
Step
Atten
Nominal
Attenu-
ation
(dB)
1 dB Step
Atten
Nominal
Attenu-
ation
(dB)
10 dB
Step
Atten
Actual
Attenu-
ation
(dB)
1 dB Step
Atten
Actual
Attenu-
ation
(dB)
Tot a l
Actual
Attenu-
ation
(dB)
Ideal
Mkr
Reading
(mV)
Actual
Mkr
Reading
(mV)
Test Record Entry –
LFE
(% of
RL)
10.If the analyzer has a 75W Input press:
Input/Output (or Input) Input Z Corr (75)
Zero Span Log Fidelity, Digital Bandwidths
This section is for analyzers with Option 1DR (narrow resolution bandwidths) only.
Setup for zero span measurements
11. Set the 1 dB step attenuator to 11 dB and the 10 dB step attenuator to 110 dB.
12. Press
13. Set the 1 dB and 10 dB step attenuators to 0 dB.
14. Set the following parameters on the analyzer:
15. On the analyzer press
Preset on the analyzer. Press the Factory Preset softkey, if it is
displayed. Set the analyzer by pressing the following keys:
System, Alignments Align Now,
FREQUENCY, 50 MHz SPAN, 100 Hz SPAN, 500 Hz (ESA-L Series only) BW/Avg, 10 Hz (ESA-E Series only) BW/Avg, 100 Hz (ESA-L Series only)
All (wait for the alignment sequence to complete)
(ESA-E Series only)
Peak Search (or Search), Marker –>, and Marker –> CF.
Chapter 2 97
Performance Verification Tests
12. Display Scale Fidelity
16. Press SPAN, Zero Span on the analyzer.
17.Adjust the synthesized signal generator amplitude until the analyzer marker amplitude reads 0 dBm
75 Input: Adjust the synthesized signal generator amplitude until the analyzer marker amplitude reads 48.75 dBmV
NOTE Do not adjust the synthesized signal generator amplitude after the reference is
established.
± 0.1 dB.
± 0.1 dB.
18. On the analyzer, press
Marker, Delta.
Measure the Cumulative Log Fidelity
19.On the analyzer, press
Single, BW/Avg, Average, 5, Enter.
20.Perform step 22 to step 24 for each measurement value in
Table 2-17.
21.Set the 1 dB and 10 dB step attenuators as indicated in Column 2 and Column 3 of Table 2-16 for the various dB from REF LVL settings.
22.Press
Single and wait for “VAvg 5” to be displayed to the right of the
graticule area.
23.Record the marker delta (
Mkr1) amplitude reading in Column 7 of
Table 2-16.
24.Calculate the Cumulative Log Fidelity Error (CLFE) as follows, and record the result in the performance verification test record as indicated in Column 8 of Table 2-16.
CLFE = Total Actual Attenuation +
Mkr Reading – Total
Actual Atten (0 dB from Ref Level)
98 Chapter 2
Performance Verification Tests
12. Display Scale Fidelity
Table 2-17 Zero Span Cumulative Log Fidelity Worksheet, Opt 1DR Narrow
Resolution Bandwidths
Column 1 Column 2 Column 3 Column 4 Column 5 Column 6 Column 7 Column 8
dB from
REF LVL
(dB)
10 dB
Step
Atten
Nominal
Attenu-
ation
(dB)
1 dB Step
Atten
Nominal
Attenu-
ation
(dB)
10 dB
Step
Atten
Actual
Attenu-
ation
(dB)
1 dB Step
Atten
Actual
Attenu-
ation
(dB)
Total
Actual
Attenu-
ation
(dB)
DMkr
Reading
(dB)
Tes t
Record
Entry-
CLFE
(dB)
0 (Ref) 0 0 0 (Ref) 0 (Ref)
–4 0 4 103)
–8 0 8 104)
–12 10 2 105)
–16 10 6 106)
–20 20 0 107)
–24 20 4 108)
–28 20 8 109)
–32 30 2 110)
–36 30 6 111)
–40 40 0 112)
–44 40 4 113)
–48 40 8 114)
–52 50 2 115)
–56 50 6 116)
–60 60 0 117)
–64 60 4 118)
–68 60 8 119)
–70 70 0 120)
Post-Test Instrument Restoration
25. Remove the RF cable from the analyzer input connector.
26.On the analyzer, press the following keys:
Preset System
, Alignments, Auto Align, All
Chapter 2 99
Performance Verification Tests

13. Input Attenuation Switching Uncertainty

13. Input Attenuation Switching Uncertainty
A 50 MHz CW signal is applied to the input of the analyzer through two calibrated step attenuators. The attenuators are the amplitude reference standard. The source is adjusted for a response at the reference level. The internal attenuators are then varied between settings and the external attenuators are changed accordingly to maintain the same input level at the mixer. The analyzer marker functions are used to measure the amplitude differences. The actual attenuation values of the step attenuators are used to correct the marker amplitude readings yielding the input attenuation switching error.
The related adjustment for this performance test is “Frequency Response.”

Equipment Required

Synthesized signal generator 1 dB step attenuator 10 dB step attenuator Attenuator switch driver (if programmable step attenuators are used) 6 dB fixed attenuator (2 required) Cable, Type-N 152-cm (60-in) (2 required) Cable, BNC 122-cm (48-in) Attenuator interconnect kit
Additional Equipment for 75 Ω Input
50 to 75 minimum loss pad Adapter, Type-N (f), to BNC (m), 75

Additional Equipment for Option BAB

Adapter, Type-N (f), to APC 3.5 (f)
100 Chapter 2
Loading...