Agilent Technologies E1345A User Manual

Agilent E1345A
16-Channel Low-Offset Relay
Multiplexer
Data Sheet
1-Slot, B-size, register based
General purpose, low-offset relay multiplexer
16-channel 3-wire or 8-channel 4-wire multiplexer
Shunt/series signal conditioning elements
Channel scanning with Agilent DMMs
Agilent E1345A
Description
The Agilent E1345A General-Purpose Reed Relay Multiplexer is a B-size, 1-slot, register-based VXI module that switches 16 channels of high, low, and guard each. The multiplexer has low thermal offset performance. This module consists of a B-size component card (labeled E1345-66201) and a screw terminal block that plugs onto the component card. The E1345A is functionally similar to the E1347A.
The E1345A multiplexer is arranged into two banks of eight switches, each having its own common high, low, and guard screw terminals accessible on the terminal block. Tree­switched common high, low, and guard screw terminals and analog bus connections provide access to all 16 channels. The module may be programmed as either a single 1x16 three-wire multiplexer, two independent 1x8 three-wire multiplexers (for four-wire ), or a 1x8 six-wire multiplexer.
Refer to the Agilent Technologies Website for instrument driver availability and downloading instructions, as well as for recent product updates, if applicable.
Configuration
One analog bus cable is shipped with each module, making it easy to connect common outputs together for slot-adjacent modules. If you are using a B-size mainframe, Agilent E1300B or E1301B, use the analog bus cable shipped with the E1326B DMM to connect it to the multiplexer(s).
C-size Adapter
For installing the E1345A in a C-size mainframe, the E1403C active adapter is recommended.
PDFINFO H 5 6 0 1 - 0 1
Product Specifications
Input
DC:
Maximum voltage (any terminal to any other terminal or chassis): 120 Vdc
AC rms:
Maximum voltage (any terminal to any other terminal or chassis): 120 V rms Maximum current (per channel common, non­inductive): 50 mA Maximum power per channel: 1 VA
General Characteristics
Relays: Reed relays
Break-before-make Relays open on power down Relays open on power up
Minimum relay life:
No load: 10E8 operations
Rated load: 10E7 operations Screw terminal wire size: 16 to 26 AWG (1.5, 1.2, 0.9, 0.75, 0.5 mm) Scanning rate: 600 channels/s typ.
General Specifications
DC
Maximum thermal offset per channel, differential Hi-Lo: 4 µV Closed channel resistance: 100 ± 10% Insulation resistance (between any two points): 10E9 Insulation resistance (Hi to Lo, power off): n/a
AC
Minimum bandwidth (-3 dB, 50 Ω source/load): 10 MHz (protection resistors shorted) Crosstalk (channel-to­channel):
100 kHz: –70 dB 10 MHz: –20 dB
Both: n/a Closed channel capacitance:
<150 pF Hi-Lo, <150 pF Lo-Guard, <2000 pF Guard-Chassis
VXI Characteristics
VXI device type: Register based, A16, slave only Size: B Slots: 1 Connectors: P1 Shared memory: None VXI busses: None C-size compatibility: Requires E1403C
Instrument Drivers
See the Agilent Technologies Website (http://www.agilent.com/find/ inst_drivers) for driver availability and downloading.
Command module firmware: Downloadable Command module firmware rev: A.01 I-SCPI Win 3.1: Yes I-SCPI Series 700: Yes C-SCPI LynxOS: Yes C-SCPI Series 700: Yes Panel Drivers: Yes
plug&play
VXI Framework: Yes
plug&play
Framework: Yes
plug&play
VXI Framework: No
Win
Win 95/NT
HP-UX
2
PDFINFO H 5 6 0 1 - 0 2
Loading...