These specifications apply to the Agilent Technologies 8591E, 8593E, 8594E, 8595E, and 8596E spectrum analyzers.
Specifications
All specifications apply over O°C to +55°C. The analyzer will meet its specifications after 2 hours of storage at a constant temperature, within the operating temperature range, 30 minutes after the analyzer is tumed on, and after CAL FREQ and CAL AMPTD (and for the 8593E, 8595E, and 8596E CAL YTF) have been run.
Frequency Specifications
Frequency Range
8591E
50 9 kHz to 1.8 GHz 75 1 MHz to 1.8 GHz
8593E 9 kHz to 22 GHz
Option 026/027 9 kHz to 26.5 GHz
Band LO harmonic = N 0 1 9 kHz to 2.9 GHz 1 1 2.75 GHz to 6.5 GHz 2 2 6.0 GHz to 12.8 GHz 3 3 12.4 GHz to 19.4 GHz 4 4 19.1 GHz to 22.0 GHz 4 4 (Opt. 026/027) 19.1 GHz to 26.5 GHz
8594E
dc coupled 9 kHz to 2.9 GHz ac coupled 100 kHz to 2.9 GHz
8595E
dc coupled 9 kHz to 6.5 GHz ac coupled 100 kHz to 6.5 GHz
8596E
dc coupled 9 kHz to 12.8 GHz ac coupled 100 kHz to 12.8 GHz
Band LO harmonic = N 0 1 9 kHz to 2.9 GHz (dc coupled) 0 1 100 kHz to 2.9 GHz (ac coupled) 1 1 2.75 GHz to 6.5 GHz 2 2 6.0 GHz to 12.8 GHz
Frequency Reference (Opt. 004)
Aging ±2 x 10
-6
/year ±1 x 10-7/year
Temperature Stability ±5 x 10
-6
±1 x 10
-8
Initial Achievable Accuracy ±0.5 x 10
-6
±2.2 x 10
-8
Agilent 8590 E-Series Portable Spectrum Analyzers
Data Sheet
2
Frequency Readout
Accuracy
(Start, Stop, Center, Marker) ±(frequency readout x frequency
reference error
1
+span accuracy +1%
of span +20% of RBW+100 Hz x N*)
Marker Count Accuracy
Frequency Span
10 MHz x N* ±(marker frequency x frequency
reference error
1
+ counter resolution
+100 Hz x N*)
Frequency Span ±(marker frequency x frequency
>10 MHz x N* reference error
1
+ counter
resolution +1 kHz x N*)
Counter Resolution
Frequency Span 10 MHz x N* Selectable from 10 Hz to 100 kHz Frequency Span >10 MHz x N* Selectable from 100 Hz to 100 kHz
Frequency Span
Range 0 Hz (zero span), and
Opt. 130 Std. Min. Min. Max (KHz) (KHz) (GHz)
8591E 1 10 1.8 8593E 1 x N* 10 x N* 19.25 8594E 1 10 2.9 8595E 1 10 6.5 8596E 1 x N* 10 x N* 12.8
Resolution Four digits or 20 Hz x N*
whichever is greater
Accuracy
Span 10 MHz x N* ±2% of span Span >10 MHz x N* ±3% of span
Frequency Sweep Time
Range
Span = 0 Hz, >1 kHz 20 ms to 100 s Span = 0 Hz (Opt. 101) 20 µs to 100 s
Accuracy
20 ms to 100 s ±3% 20 µs to <20 ms (Opt. 101) ±2%
Sweep Trigger Free run, single, line, video,
external
Resolution Bandwidth 1 kHz to 3 MHz (3 dB) in 1-3-10
sequence. 9 kHz and 120 kHz (6 dB) EMI bandwidths.
Option 130 Adds 30, 100, and 300 Hz (3 dB)
bandwidths and 200 Hz (6 dB) EMI bandwidth.
Accuracy ±20% Selectivity (Characteristic)
–60 dB/–3 dB
3 kHz to 10 kHz 15:1 100 kHz to 3 MHz 15.1 1 kHz, 30 kHz 16:1
–40 dB/–3 dB
30 Hz to 300 Hz 10:1
Video Bandwidth Range 30 Hz to 1 MHz in 1,3 sequence
1 Hz to 1 MHz (Opt 130)
Stability
Noise Sidebands (1 kHz RBW, 30 Hz VBW and sample detector) >10 kHz offset from CW signal –90 dBc/Hz + 20 Log N* >20 kHz offset from CW signal –100 dBc/Hz + 20 Log N* >30 kHz offset from CW signal <–105 dBc/Hz + 20 Log N*
Residual FM
8591E
1 kHz RBW, 1 kHz VBW 250 Hz pk-pk in 100 ms 30 Hz RBW, 30 Hz VBW 30 Hz pk-pk in 300 ms
8593E, 94E, 95E, 96E
1 kHz RBW, 1 kHz VBW (250 x N*) Hz pk-pk in
100 ms
30 Hz RBW, 30 Hz VBW (30 x N*) Hz pk-pk in 300 ms
System-Related Sidebands
>30 kHz offset from CW signal –65 dBc + 20 Log N*
Comb Generator Frequency
8593E, 96E I00 MHz fundamental
frequency
Accuracy ±0.007%
* N = LO harmonic. N = 1 for 91E, 94E, 95E
1. Frequency reference error = (aging rate x period of time since adjustment + initial achievable accuracy + temperature stability).
3
Amplitude Specifications
Amplitude specifications do not apply for Analog+ mode and negative peak detector mode except as noted in “Amplitude Characteristics.”
Amplitude Range
Displayed average noise level to +30 dBm
8591E (Opt. 001) Displayed average noise level to
+72 dBmV
Maximum Safe Input Level (input attenuator 10 dB)
Average Continuous Power +30 dBm (1 W)
8591E (Opt. 001) +72 dBmV (0.2 W)
Peak Pulse Power
8591E +30 dBm (1 W) 8591E (Opt. 001) +72 dBmV (0.2 W) 8593E, 94E, 95E, 96E +50 dBm (100 W) for < 10 µs pulse
width and <1 % duty cycle, input attenuation 30 dB.
dc
8591E 25 Vdc 8591E (Opt. 001) 100 Vdc 8593E 0 Vdc 8594E, 95E, 96E 0 V (dc coupled)
50 V (ac coupled)
Gain Compression
>10 MHz 0.5 dB (total power at input mixer2 = –10 dBm)
Displayed Average Noise Level
(Input terminated, 0 dB attenuation, 1 Hz/30 Hz VBW, sample­detector) 8591E 30 Hz RBW 1 kHz RBW
400 kHz to 1 MHz –130 dBm –115 dBm 1 MHz to 1.5 GHz –130 dBm –115 dBm
1.5 GHz to 1.8 GHz –128 dBm –113 dBm
8591E (Opt. 001)
1 MHz to 1.5 GHz –78 dBmV –63 dBmV
1.5 GHz to 1.8 GHz –76 dBmV –61 dBmV
8593E
400 kHz to 2.9 GHz –127 dBm –112 dBm
2.75 GHz to 6.5 GHz –129 dBm –114 dBm
6.0 GHz to 12.8 GHz –117 dBm –102 dBm
12.4 GHz to 19.4 GHz –113 dBm –98 dBm
19.1 GHz to 22 GHz –107 dBm –92 dBm
8593E (Opt. 026/027)
19.1 GHz to 26.5 GHz –102 dBm –87 dBm
8594E
400 kHz to<5 MHz –122 dBm –107 dBm 5 MHz to 2.9 GHz –127 dBm –112 dBm
8595E
400 kHz to 2.9 GHz –125 dBm –110 dBm
2.75 GHz to 6.5 GHz –127 dBm –112 dBm
8596E
400 kHz to 2.9 GHz –125 dBm –110 dBm
2.75 GHz to 6.5 GHz –127 dBm –112 dBm
6.0 GHz to 12.8 GHz –115 dBm –100 dBm
2. Mixer Power Level (dBm) = Input Power (dBm) Input Atten. (dB)
Nominal Dynamic Range
30
Sensitivity 30 Hz RBW
40
Third Order Intermod Second Order Distortion
50
60
70
(30 Hz Res BW)
80
Dynamic Range (dB)
90
100
–70 –60 –50 –40 –30 –20 –10 0 +10
Agilent 8593E (Option 130)
19.1 TO 22 GHz
12.4 TO 19.4 GHz
6.0 TO 12.8 GHz
400 kHz TO 6.5 GHz
10 MHz to 2.9 GHz
>2.75 GHz
Mixer Level (dBm)