THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED
TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY
INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN
AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT
WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is
delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995),
or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in
FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or
disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD
Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined
in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as
defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any
technical data.
Assistance
Product maintenance agreements and other customer assistance agreements are available for Agilent
products.
For any assistance, contact Agilent Technologies. Refer to page 5-7.
ii85057B
Printing Copies of This Document
To print copies of this document, download the PDF file from the Agilent Web site:
•Go to http://www.agilent.com.
•Enter the document’s part number (located on the title page) in the Search box.
The Agilent 85057B 2.4 mm verification kit provides a set of standards with known characteristics, traceable
to a reference (golden) standard in Agilent Technologies calibration lab. This set of standards is used to
verify your measurement calibration and also to verify that your PNA system is operating within its
specifications. The frequency range covered by the 85057B is from 45 MHz to 50 GHz.
Kit Contents
The 85057B verification kit includes the following items:
•20 dB attenuator
•40 dB attenuator
•25 mismatch airline
•50 airline
•storage box
•User’s and Service Guide (this manual)
•USB drive that contains factory-measured verification data for use with the PNA system verification
process
•data sheet for each device that contains factory-measured verification data
•anti-virus scan report
Refer to Chapter 6, “Replaceable Parts,” for a complete list of replaceable part numbers.
IMPORTANTBackup copies of the data sheets, anti-virus scan reports, and the data contained on the USB
drive should be made immediately upon receipt of the verification kit.
NOTEA file containing the verification data for your kit is maintained for one year from the time of
measurement. If you lose this data, contact Agilent. See “Contacting Agilent” on page 5-7.
Calibration Definitions
The 85057B verification kit is intended to be used with the 85056A 2.4 mm calibration kits. Prior to
performing a calibration with your PNA, the calibration kit must be selected and the calibration definitions
for the devices in the kit installed in the PNA. Refer to your PNA Help system for instructions on selecting
the calibration kit and performing a calibration.
Installation of the Calibration Definitions
The calibration definitions for the kit should be permanently installed in the internal memory or hard disk of
the PNA. They may already be resident within the analyzer, or you can download them from Agilent’s
Calibration Kit Definitions Web page at www.na.tm.agilent.com/pna/caldefs/stddefs.html. In addition, the
calibration definitions may be entered manually from the front panel. Refer to your PNA Help system for
1-285057B
General Information
Incoming Inspection
instructions.
Equipment Required but Not Supplied
Certain items are required or recommended for successful operation of the verification kit but are not
included in the kit. Refer to Table on page 6-3 for ordering information.
Incoming Inspection
Verify that the shipment is complete by referring to the contents list in the shipping container.
Check for damage. The foam-lined storage case provides protection during shipping. Verify that this case
and its contents are not damaged.
If the case or any device appears damaged, or if the shipment is incomplete, refer to “Contacting Agilent” on
page 5-7. Agilent will arrange for repair or replacement of incomplete or damaged shipments without
waiting for a settlement from the transportation company. Refer to “Returning a Kit or Device to Agilent
Technologies” on page 5-6 for instructions.
Recording the Device Serial Numbers
In addition to the kit serial number, the devices in this kit are individually serialized (serial numbers are
labeled into the body of each device). Record these serial numbers in Ta bl e 1 - 1. Recording the serial
numbers will prevent confusing the devices in this kit with similar devices from other kits.
Table 1-1 Serial Number Record for the 85057B
DeviceSerial Number
Verific ation k it
20 dB attenuator
40 dB attenuator
50airline
25mismatch airline
___________________________
___________________________
___________________________
___________________________
___________________________
85057B1-3
General Information
Clarifying the Terminology of a Connector Interface
Clarifying the Terminology of a Connector Interface
In this document and in the prompts of the PNA calibration wizard, the sex of connectors and adapters is
referred to in terms of the center conductor. For example, a connector or device designated as 1.85 mm –f–
has a 1.85 mm female center conductor.
A connector gage is referred to in terms of the connector that it measures. For instance, a male connector
gage has a female connector on the gage so that it can measure male devices.
Preventive Maintenance
The best techniques for maintaining the integrity of the devices in this kit include:
•routine visual inspection
•cleaning
•proper gaging
•proper connection techniques
All of these are described in Chapter 3. Failure to detect and remove dirt or metallic particles on a mating
plane surface can degrade repeatability and accuracy and can damage any connector mated to it. Improper
connections, resulting from pin depth values being out of specification (see Table 2-2 on page 2-4), or from
bad connection techniques, can also damage these devices.
1-485057B
2Specifications
2-1
Specifications
Environmental Requirements
Environmental Requirements
Table 2-1Environmental Requirements
ParameterRequired Values/Ranges
Te m p e r a t u r e
Operating
Storage
Error-corrected range
Altitude
Operating< 4,500 meters (ª15,000 feet)
Storage< 15,000 meters (ª50,000 feet)
Relative humidityAlways non–condensing
Operating0 to 80% (26 C maximum dry bulb)
Storage0 to 90%
a. The temperature range over which the calibration standards maintain conformance to their specifications.
b. The allowable network analyzer ambient temperature drift during measurement calibration and during measurements
a
b
when the network analyzer error correction is turned on. Also, the range over which the network analyzer maintains its
specified performance while correction is turned on.
+20 C to +26 C (+68 F to +79 F)
–40 C to +75 C
1 C (1.8 F) of measurement calibration temperature
(–40 F to +167 F)
Temperature—What To Watch Out For
Due to the small dimensions of the devices, electrical characteristics will change with temperature.
Therefore, the operating temperature is a critical factor in their performance, and must be stable before use.
IMPORTANTAvoid unnecessary handling of the devices during use because your fingers are a heat
source.
2-285057B
Specifications
Mechanical Characteristics
Mechanical Characteristics
Mechanical characteristics such as center conductor protrusion and pin depth are not warranted
performance specifications. They are, however, important supplemental characteristics related to electrical
performance. Agilent Technologies verifies the mechanical characteristics of the devices in this kit with
special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any
improper pin depth when the kit leaves the factory.
“Gaging Connectors” on page 3-7 explains how to use gages to determine if the kit devices have maintained
their mechanical integrity. Refer to Table 2-2, “Connector Pin Depths,” for allowable recession.
Pin Depth
Pin depth is the distance the center conductor mating plane differs from being flush with the outer
conductor mating plane. See Figure 2-1. The pin depth of a connector can be in one of two conditions:
•Protrusion is the condition in which the center conductor extends beyond the outer conductor mating
plane. This condition will indicate a positive value on the connector gage.
•Recession is the condition in which the center conductor is set back from the outer conductor mating
plane. This condition will indicate a negative value on the connector gage.
Figure 2-1Connector Pin Depth
Supplemental Characteristics
The following tables list the dimensions of the 50 airline and the 25 mismatch airline. These are
supplemental mechanical characteristics, and from these characteristics you can calculate expected
85057B2-3
Specifications
Mechanical Characteristics
electrical performance.
Table 2-2 Connector Pin Depths
Connectors
millimetersinches
Attenuators
a
Airlines
a. The relationship between the length of the inner conductor and the length of the
outer conductor determines the airline center conductor recession. Refer to
“Gaging the Airline” on page 3-10.
0.000 to
0.0000 to
Allowable Recession
–0.0250.0000 to –0.001
–0.0130.0000 to –0.0005
Using these mechanical dimensions, you can calculate the expected electrical performance with the
equations in the following publications:
•Nelson, Robert E., and Marlene R. Coryell, “Electrical Parameters of Precision, Coaxial, Air-Dielectric
Transmission Lines”, U.S. National Bureau of Standards Monograph No. 96.
•Somlo, P.I., “The Computation of Coaxial Line Step Capacitances”, IEEE Transactions on Microwave
Theory and Techniques, Volume MTT-15, No. 1, January, 1967.
The measurement method in these publications provides a general idea of the expected device
characteristic impedance. Variations in connector interfaces can have a large effect on your actual electrical
measurements.
2-485057B
Specifications
Mechanical Characteristics
Airline Characteristics
The dimensions of the airline outer conductor are shown in Figure 2-2. There are two similar outer
conductors in each kit. They are specifically matched to each center conductor.
The dimensions of the 50 airline and the 25 mismatch airline are shown in Figure 2-3 and Figure 2-4.
CAUTIONThe center and outer conductors of the airlines in this kit have been mechanically measured
and matched. Do not use the center or outer conductors provided in this kit with a center or
outer conductor from any other airline. Damage to the airline or attaching connector may
result.
Figure 2-2Airline Outer Conductor
Dimensionmillimetersinches
D Diameter2.400 ±0.00250.0945 ±0.0001
L Length
Figure 2-3 50Airline Center Conductor
Dimensionmillimetersinches
d 1.0423 ±0.0030.04104 ±0.00012
49.991 ±0.025
1.968 ±0.001
85057B2-5
Specifications
Mechanical Characteristics
Dimensionmillimetersinches
L +0.0025/-0.013
Figure 2-4 25 Mismatch Airline Center Conductor
Dimensionmillimetersinches
+0.0001/
–0.0005
d1.0423 ±0.0080.04104 ±0.0003
d
L
1
l
1
l
2
1.58 ±0.0050.0622 ±0.0002
37.46 ±0.0191.4748 ±0.0007
6.22 ±0.0500.2449 ±0.002
+0.0025/
–0.013+0.0001/–0.0005
2-685057B
Specifications
Electrical Specifications
Electrical Specifications
At the factory, each verification device is electrically characterized on a PNA measurement system. These
factory measurements are traceable to the National Institute of Standards and Technology (NIST) through
mechanical and electrical paths (for more information on traceability, contact Agilent Technologies. Refer to
“Contacting Agilent” on page 5-7.
The factory-measured data for each device is supplied in print and on the USB drive with your kit.
85057B2-7
Specifications
Electrical Specifications
2-885057B
3Use, Maintenance, and Care of the Devices
3-1
Use, Maintenance, and Care of the Devices
Electrostatic Discharge
Electrostatic Discharge
Protection against ESD (electrostatic discharge) is essential while connecting, inspecting, or cleaning
connectors attached to a static-sensitive circuit (such as those found in test sets).
Static electricity can build up on your body and can easily damage sensitive internal circuit elements when
discharged. Static discharges too small to be felt can cause permanent damage. Devices such as calibration
components and devices under test (DUTs), can also carry an electrostatic charge. To prevent damage to the
test set, components, and devices:
•always wear a grounded wrist strap having a 1 M resistor in series with it when handling components
and devices or when making connections to the test set.
•always use a grounded, conductive table mat while making connections.
•always wear a heel strap when working in an area with a conductive floor. If you are uncertain about the
conductivity of your floor, wear a heel strap.
•always ground yourself before you clean, inspect, or make a connection to a static-sensitive device or
test port. You can, for example, grasp the grounded outer shell of the test port or cable connector briefly.
•always ground the center conductor of a test cable before making a connection to the analyzer test port
or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center conductor to the
outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
Figure 3-1 shows a typical ESD protection setup using a grounded mat and wrist strap. For parts numbers of
ESD protection supplies, refer to Ta bl e 6 -3
Figure 3-1ESD Protection Setup
3-285057B
Use, Maintenance, and Care of the Devices
Visual Inspection
Visual Inspection
Visual inspection and, if necessary, cleaning should be done every time a connection is made. Metal
particles from the connector threads may fall into the connector when it is disconnected.
CAUTIONDevices with damaged connectors should be immediately discarded or clearly marked and
set aside for repair. A damaged device will in turn damage any good connector to which it is
attached. Determine the cause of the damage before connecting a new, undamaged
connector in the same configuration.
In some cases, magnification is necessary to see damage to a connector; a magnifying device with a
magnification of 10is recommended. However, not all defects that are visible only under magnification
will affect the electrical performance of the connector. Use the following guidelines when evaluating the
integrity of a connector.
Look for Obvious Defects and Damage First
Examine the connectors first for obvious defects or damage: badly worn plating on the connector interface,
deformed threads or bent, broken, or misaligned center conductors. Connector nuts should move smoothly
and be free of burrs, loose metal particles, and rough spots.
What Causes Connector Wear?
Connector wear is caused by connecting and disconnecting the devices. The more use a connector gets, the
faster it wears and degrades. The wear is greatly accelerated when connectors are not kept clean, or are not
connected properly.
Connector wear eventually degrades performance of the device. Calibration devices should have a long life if
their use is on the order of a few times per week. Replace devices with worn connectors.
The test port connectors on the PNA test set may have many connections each day, and are, therefore, more
subject to wear. It is recommended that an adapter be used as a test port saver to minimize the wear on the
test set’s test port connectors.
Inspect the Mating Plane Surfaces
Flat contact between the connectors at all points on their mating plane surfaces is required for a good
connection. See Figure 2-1 on page 2-3. Look especially for deep scratches or dents, and for dirt and metal
particles on the connector mating plane surfaces. Also look for signs of damage due to excessive or uneven
wear or misalignment.
Light burnishing of the mating plane surfaces is normal, and is evident as light scratches or shallow circular
marks distributed more or less uniformly over the mating plane surface. Other small defects and cosmetic
imperfections are also normal. None of these affect electrical or mechanical performance. If a connector
shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and
inspect it again.
85057B3-3
Use, Maintenance, and Care of the Devices
Visual Inspection
Inspect Female Connectors
Inspect the contact fingers in the female center conductor carefully. These can be bent or broken, and
damage to them is not always easy to see. A connector with damaged contact fingers will not make good
electrical contact and must be replaced.
NOTEThis is particularly important when mating nonprecision to precision devices.
The female connectors in this calibration kit are metrology-grade, precision slotless connectors (PSC).
Precision slotless female connectors are used to improve accuracy. With PSCs on test ports and standards,
the accuracy achieved when measuring at 50 dB return loss levels is comparable to using conventional
slotted connectors measuring devices having only 30 dB return loss. This represents an accuracy
improvement of approximately 10 times.
Conventional female center conductors are slotted and, when mated, are flared by the male pin. Because
physical dimensions determine connector impedance, this change in physical dimension affects electrical
performance, making it very difficult to perform precision measurements with conventional slotted female
connectors.
The precision slotless connector was developed to eliminate this problem. The PSC has a center conductor
with a solid cylindrical shell, the outside diameter of which does not change when mated. Instead, this
center conductor has an internal contact that flexes to accept the male pin.
3-485057B
Use, Maintenance, and Care of the Devices
Cleaning Connectors
Cleaning Connectors
1. Use Compressed Air or Nitrogen
Clean connectors are essential for ensuring the integrity of RF and microwave coaxial connections.
WARNINGAlways use protective eyewear when using compressed air or nitrogen.
Use compressed air (or nitrogen) to loosen particles on the connector mating plane surfaces.
You can use any source of clean, dry, low-pressure compressed air or nitrogen that has an effective
oil-vapor filter and liquid condensation trap placed just before the outlet hose.
Ground the hose nozzle to prevent electrostatic discharge, and set the air pressure to less than 414 kPa
(60 psi) to control the velocity of the air stream. High-velocity streams of compressed air can cause
electrostatic effects when directed into a connector. These electrostatic effects can damage the device.
Refer to “Electrostatic Discharge” on page 3-2 for additional information.
2. Clean the Connector Threads
WARNINGKeep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly closed
container. It is extremely flammable. In case of fire, use alcohol foam, dry chemical, or
carbon dioxide; water may be ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact with eyes, skin, and
clothing. It causes skin irritation, may cause eye damage, and is harmful if swallowed or
inhaled. It may be harmful if absorbed through the skin. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal, state, and local
environmental regulations.
Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn
contaminants on a connector that cannot be removed with compressed air or nitrogen. Refer to Ta bl e 6 -3
on page 6-5 for part numbers for isopropyl alcohol and cleaning swabs.
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the connector threads.
c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean, low-pressure
compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
3. Clean the Mating Plane Surfaces
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the center and outer conductor mating plane surfaces. Refer to Figure 2-1 on page 2-3. When
cleaning a female connector, avoid snagging the swab on the center conductor contact fingers by
using short strokes.
85057B3-5
Use, Maintenance, and Care of the Devices
Cleaning Connectors
c. Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean, low-pressure
compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
4. Inspect the Connector
Inspect the connector to make sure that no particles or residue remain. Refer to “Visual Inspection” on
page 3-3.
3-685057B
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging Connectors
The gages available from Agilent Technologies are intended for preventive maintenance and troubleshooting
purposes only. They are effective in detecting excessive center conductor protrusion or recession, and
conductor damage on DUTs, test accessories, and the calibration kit devices.Do not use the gages for
precise pin depth measurements.
Connector Gage Accuracy
The connector gages are only capable of performing coarse measurements. They do not provide the degree
of accuracy necessary to precisely measure the pin depth of the kit devices. This is partially due to the
repeatability uncertainties that are associated with the measurement. Only the factory—through special
gaging processes and electrical testing— can accurately verify the mechanical characteristics of the
devices.
With proper technique, the gages are useful in detecting gross pin depth errors on device connectors. To
achieve maximum accuracy, random errors must be reduced by taking the average of at least three
measurements having different gage orientations on the connector. Even then, the resultant average can be
in error by as much as ±0.0001 inch due to systematic (biasing) errors usually resulting from worn gages and
gage masters. As the gages undergo more use, the systematic errors can become more significant in the
accuracy of the measurement.
NOTEWhen measuring pin depth, the measured value (resultant average of three or more
measurements) contains measurement uncertainty and is not necessarily the true value.
Always compare the measured value with the observed pin depth limits (which account for
measurement uncertainties) in
Table 2-2 on page 2-4 to evaluate the condition of device connectors.
When to Gage Connectors
Gage a connector at the following times:
•Prior to using a device for the first time, record the pin depth measurement so that it can be compared
with future readings. This serves as a good troubleshooting tool when you suspect damage may have
occurred to the device.
•If either visual inspection or electrical performance suggests that the connector interface may be out of
typical range (due to wear or damage, for example).
• If a verification device is used by someone else or on another system or piece of equipment.
•Initially, after every 100 connections, and after that, as often as experience indicates.
85057B3-7
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging Procedures
Gaging 2.4 mm Connectors
NOTEAlways hold a connector gage by the gage barrel, below the dial indicator. This gives the best
stability, and improves measurement accuracy. (Cradling the gage in your hand or holding it
by the dial applies stress to the gage plunger mechanism through the dial indicator housing.)
1. Select the proper gage for your connector. The gages are intended for performing 2.4 mm pin depth
measurements. Refer to Table 6-3 on page 6-5 for gage part numbers.
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual Inspection” and
“Cleaning Connectors” earlier in this chapter.
3. Zero the connector gage (refer to Figure 3-2):
a. While holding the gage by the barrel, and without turning the gage or the device, connect the gage to
the gage master by interconnecting the male and female connectors. Connect the nut finger tight. Do
not overtighten.
b. Using an open-end wrench to keep the device body from rotating, use the torque wrench included in
the kit to tighten the connecting nut to the specified torque. Refer to “Final Connection Using a
Torque Wrench” on page 3-14 for additional information.
c. As you watch the gage pointer, gently tap the barrel of the gage to settle the reading.
The gage pointer should line up exactly with the zero mark on the gage. If not, adjust the zero set
knob until the gage pointer lines up exactly with the zero mark.
d. Remove the gage master.
4. Gage the device connector (refer to Figure 3-2):
a. While holding the gage by the barrel, and without turning the gage or the device, connect the gage to
the device by interconnecting the male and female connectors. Connect the nut finger-tight. Do not
overtighten.
b. Using an open-end wrench to keep the device body from rotating, use the torque wrench included in
the kit to tighten the connecting nut to the specified torque. Refer to “Final Connection Using a
Torque Wrench” on page 3-14 for additional information.
c. Gently tap the barrel of the gage with your finger to settle the gage reading.
d. Read the gage indicator dial. Read only the black signs,not the red signs.
For maximum accuracy, measure the connector a minimum of three times and take an average of the
readings. After each measurement, rotate the gage a quarter-turn to reduce measurement variations
that result from the gage or the connector face not being exactly perpendicular to the center axis.
e. Compare the average reading with the specifications listed in Table 2-2 on page 2-4.
f. Without turning the gage or the device, remove the device from the gage.
3-885057B
Figure 3-2Gaging 2.4 mm Connectors
Use, Maintenance, and Care of the Devices
Gaging Connectors
85057B3-9
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging the Airline
The airlines in this kit are measured and matched for length at the factory using special fixtures and gages.
Because the gages supplied in the calibration kits do not have the accuracy of the factory gages, use the
following procedure for very general results only. Perform this procedure whenever you suspect that a
center conductor has been switched with another airline or that a device has been damaged.
NOTEAlways hold a connector gage by the gage barrel, below the dial indicator. This gives the best
stability, and improves measurement accuracy.
1. Select a male connector gage and male short for this procedure. Refer to Table 6-3 on page 6-5 for part
numbers.
2. Inspect and clean the mating surfaces and connector threads of the short, airline, and gage. Refer to
“Visual Inspection” and “Cleaning Connectors” earlier in this chapter.
3. Zero the connector gage (refer to Figure 3-3):
a. While holding the gage by the barrel and without turning the gage or the short, connect the gage to
the short by interconnecting the male and female connectors. Connect the nut finger tight. Do not
overtighten.
b. Using an open-end wrench to keep the device body from rotating, use the torque wrench
recommended for use with this kit to tighten the connecting nut to the specified torque. Refer to
“Final Connection Using a Torque Wrench” on page 3-14 for additional information.
c. As you watch the gage pointer, gently tap the barrel of the gage to settle the reading.
d. The gage pointer should line up exactly with the zero mark on the gage. If not, loosen the dial lock
screw on the gage and rotate the gage dial so that the pointer is aligned with the zero mark. Tighten
the dial lock screw.
e. Without turning the short or the gage, remove the short from the gage. Refer to “How to Separate a
Connection” on page 3-18.
Figure 3-3Zeroing the Connector Gage Using the Short
4. Assemble the airline and center conductor (refer to Figure 3-4):
3-1085057B
Use, Maintenance, and Care of the Devices
Gaging Connectors
CAUTIONWear gloves while performing the following steps. You will be touching the exposed center
conductor of the airline. Do not transfer oil or dirt from your fingers to the center conductor.
See “Handling and Storage” on page 3-18.
a. Remove the center conductor from its plastic case. Make sure you select the correct center
conductor for the airline you are connecting. Refer to Figure 2-3 and Figure 2-4 on page 2-6 for
illustrations of both center conductors.
b. Remove the protective end cap from the female end (the end without the connecting nut) of the outer
conductor.
c. Leave the protective end cap on the male end of the airline to prevent the center conductor from
falling out of the outer conductor.
d. Insert the center conductor into the outer conductor so that the female end of the center conductor is
toward the female end of the outer conductor (the end without the connector nut). Refer to Figure
3-4.
Figure 3-4 Assembling the Airline and Center Conductor
5. Attach the short:
a. Without turning the airline or the short, connect the airline to the short by interconnecting the male
and female connectors. Connect the nut finger tight. Do not overtighten.
b. Using an open-end wrench to keep the device body from rotating, use the torque wrench
recommended for use with this kit to tighten the connecting nut to the specified torque. Refer to
“Final Connection Using a Torque Wrench” on page 3-14 for additional information.
c. Remove the protective end cap from the male end of the airline.
85057B3-11
Use, Maintenance, and Care of the Devices
Gaging Connectors
NOTEDo not allow either the center or outer conductor of the airline to come in contact with a
metal or harder surface. The soft gold plating can be displaced, changing the pin depth and
thus the performance of the airline.
d. Carefully press the male end of the center conductor lightly against a firm flat object to seat the
center conductor into the short.
6. Gage the airline (refer to Figure 3-5):
a. While holding the gage by the barrel, and without turning the gage or the airline, connect the gage to
the airline by interconnecting the male and female connectors. Connect the nut finger tight. Do not
overtighten.
b. Using an open-end wrench to keep the device body from rotating, use the torque wrench
recommended for use with this kit to tighten the connecting nut to the specified torque. Refer to
“Final Connection Using a Torque Wrench” on page 3-14 for additional information.
c. Gently tap the barrel of the gage with your finger to settle the gage reading.
Figure 3-5Gaging the Airline
d. Wait approximately 5 minutes for the temperature to stabilize. Do not touch the airline, short, or gage
during this time as your body temperature will affect the temperature of the devices.
The gage reading should be within the pin depth specifications listed in Table 2-2 on page 2-4.
Remember, the gage is intended for coarse measurements only and has an accuracy of 0.0001 in.
3-1285057B
Use, Maintenance, and Care of the Devices
Gaging Connectors
7. Disconnect the short and gage from the airline:
NOTEIf the airline center conductor does not disengage from the device center conductor, gently
pull the center conductors apart and then push the airline center conductor back inside the
outer conductor of the airline.
a. Without turning the airline or the gage, remove the gage from the airline. Refer to “How to Separate a
Connection” on page 3-18.
b. Replace the protective end cap on the airline to prevent the center conductor from sliding out of the
outer conductor.
c. Without turning the airline or the short, remove the short from the airline. Refer to “How to Separate
a Connection” on page 3-18.
d. If you will not be using the airline again immediately, slide the center conductor out of the outer
conductor and store the center conductor in the plastic case provided.
e. Replace the other protective end cap on the outer conductor and store the center and outer
conductors in the foam lined storage case.
85057B3-13
Use, Maintenance, and Care of the Devices
Connections
Connections
Good connections require a skilled operator. The most common cause of measurement error is bad
connections. The following procedures illustrate how to make good connections.
How to Make a Connection
Preliminary Connection
1. Ground yourself and all devices. Wear a grounded wrist strap and work on a grounded, conductive table
mat. Refer to “Electrostatic Discharge” on page 3-2 for ESD precautions.
2. Visually inspect the connectors. Refer to “Visual Inspection” on page 3-3.
3. If necessary, clean the connectors. Refer to “Cleaning Connectors” on page 3-5.
4. Use a connector gage to verify that all center conductors are within the pin depth values listed in Ta b l e
2-2 on page 2-4. Refer to “Gaging Connectors” on page 3-7.
5. Carefully align the connectors. The male connector center pin must slip concentrically into the contact
finger of the female connector.
6. Push the connectors straight together and tighten the connector nut finger tight.
CAUTIONDo not turn the device body. Turn only the connector nut. Damage to the center conductor
can occur if the device body is rotated.
Do not twist or screw the connectors together. As the center conductors mate, there is usually a slight
resistance.
7. The preliminary connection is tight enough when the mating plane surfaces make uniform, light contact.
Do not overtighten this connection.
A connection in which the outer conductors make gentle contact at all points on both mating surfaces is
sufficient. Very light finger pressure is enough to accomplish this.
8. Make sure the connectors are properly supported. Relieve any side pressure on the connection from long
or heavy devices or cables.
Final Connection Using a Torque Wrench
Use a torque wrench to make a final connection. Ta b l e 3 - 1 provides information about the torque wrench
recommended for use with the calibration kit. A torque wrench is not included in the calibration kit. Refer to
Table 6-3 on page 6-5 for part number and ordering information.
Table 3-1 Torque Wrench Information
Co n n e ct or Ty peTor q u e Se t t i ngTor q u e Tol e ra n c e
2.4 mm90 N-cm (8 in-lb)9.0 N-cm (0.8 in-lb)
Using a torque wrench guarantees that the connection is not too tight, preventing possible connector
3-1485057B
Use, Maintenance, and Care of the Devices
Connections
damage. It also guarantees that all connections are equally tight each time.
Prevent the rotation of anything other than the connector nut that you are tightening. It may be possible to
do this by hand if one of the connectors is fixed (as on a test port). However, it is recommended that you use
an open-end wrench to keep the body of the device from turning.
1. Position both wrenches within 90 degrees of each other before applying force. See Figure 3-6. Wrenches
opposing each other (greater than 90 degrees apart) will cause a lifting action which can misalign and
stress the connections of the devices involved. This is especially true when several devices are
connected together.
Figure 3-6Wrench Positions
2. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See Figure 3-7.
Figure 3-7Using the Torque Wrench
3. Apply downward force perpendicular to the wrench handle. This applies torque to the connection
through the wrench.
Do not hold the wrench so tightly that you push the handle straight down along its length rather than
pivoting it, otherwise, you apply an unknown amount of torque.
85057B3-15
Use, Maintenance, and Care of the Devices
Connections
CAUTIONYou don’t have to fully break the handle of the torque wrench to reach the specified torque;
doing so can cause the handle to kick back and loosen the connection. Any give at all in the
handle is sufficient torque.
4. Tighten the connection just to the torque wrench break point. The wrench handle gives way at its
internal pivot point. See Figure 3-7. Do not tighten the connection further.
Connecting the Airline
CAUTIONBefore making any connections to the test set, be sure that bias power to the test set is off,
and take care to avoid electrostatic discharge. Refer to “Electrostatic Discharge” on
page 3-2.
CAUTIONWear gloves while performing the following procedure. You will be touching the exposed
center conductor of the airline. It is important that you do not transfer oil and dirt from your
fingers to this center conductors. Refer to “Handling and Storage” on page 3-18.
Before making the connection, refer to “Preliminary Connection” on page 3-14.
Cables with the appropriate adapters on the ends should be connected to PORT 1 and PORT 2 of the network
analyzer.
1. Remove the center conductor from its plastic case. Make sure you select the correct center conductor
for the airline you are connecting. Refer to Figure 2-3 and Figure 2-4 on page 2-6 for illustrations of both
center conductors.
2. Remove the protective end cap from the female end (the end without the connecting nut) of the airline.
Leave the protective end cap on the male end of the airline to prevent the center conductor from falling
out of the outer conductor.
3. Insert the female end of the center conductor into the outer conductor so that the female end of the
center conductor is toward the female end of the outer conductor (the end without the connector nut).
Figure 3-8Airline Center Conductor Placement
3-1685057B
Use, Maintenance, and Care of the Devices
Connections
NOTETo avoid damaging the airline center conductor, always keep it in direct line with the center
conductor of the device to which it is being connected.
4. Bring the airline—with center conductor installed—toward the cable connector and mate the female
end of the airline center conductor with the center conductor of the cable connector. Refer to Figure 3-9.
Figure 3-9Connecting the Airline
5. Push the airline's female coupling sleeve forward and turn the connecting nut (of the adapter attached to
the cable) to mate the outer conductor of the airline with the adapter. Connect the nut finger tight. Do not
overtighten.
6. Remove the protective end cap from the male end of the airline. Align and insert the male end of the
airline center conductor into the female end of the cable adapter and mate the outer conductors.
Connect the nut finger tight. Do not overtighten.
7. Using an open-end wrench, hold the sliding female coupling sleeve on the female end of the outer
conductor to keep it from rotating. Refer to Figure 3-10 Use the torque wrench recommended for use
with this kit to tighten the adapter connecting nut to the specified torque. Refer to “Final Connection
Using a Torque Wrench” on page 3-14.
8. Using an open-end wrench to keep the cable adapter from rotating, use the torque wrench
recommended for use with this kit to tighten the airline male-end connecting nut to the specified torque.
Figure 3-10 Torquing the Connections
85057B3-17
Use, Maintenance, and Care of the Devices
Handling and Storage
How to Separate a Connection
NOTEDo not turn the device body. Only turn the connector nut. Damage to the center conductor
can occur if the device body is rotated.
1. Use an open-end wrench to prevent the device body from turning.
2. Use another open-end wrench to loosen the connector nut.
3. Complete the disconnection by hand, turning only the connector nut.
4. Pull the connectors straight apart without twisting, rocking, or bending either of the connectors.
NOTEIf disconnecting an airline and the airline center conductor does not disengage from the
device center conductor, gently pull the center conductors apart and then push the airline
center conductor back inside the outer conductor of the airline.
Handling and Storage
•Do install the protective end caps and store the devices in the foam-lined storage case when not in use.
•Do keep connectors and airlines clean.
•Do not store connectors and airlines loose in a box, or in a desk or bench drawer. This is the most
common cause of connector damage during storage.
•Do not touch mating plane surfaces. Natural skin oils and microscopic particles of dirt are easily
transferred to a connector interface and are very difficult to remove.
•Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces
can be damaged if the interface comes in contact with any hard surface.
3-1885057B
4Performance Verification
4-1
Performance Verification
Introduction
Introduction
The performance of your verification kit can only be verified by returning the kit to Agilent Technologies for
recertification. The equipment required to verify the specifications of the devices in the kit has been
specially manufactured and is not commercially available.
How Agilent Verifies the Devices in Your Kit
Agilent verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test system are verified with precision airlines and shorts that
are directly traced to the National Institute of Standards and Technology (NIST). The airline and short
characteristics are developed from mechanical measurements. The mechanical measurements and
material properties are carefully modeled to give very accurate electrical representation. The mechanical
measurements are then traced to NIST through various plug and ring gages and other mechanical
measurements.
2. Each device is electrically tested on this system. For the initial (before sale) testing of the devices,
Agilent includes the test measurement uncertainty as a guardband to guarantee each device meets the
published specification. For recertifications (after sale), no guardband is used and the measured data is
compared directly with the specification to determine the pass or fail status. The measurement
uncertainty for each device is, however, recorded in the calibration report that accompanies recertified
kits.
These two steps establish a traceable link to NIST for Agilent to the extent allowed by the institute’s
calibration facility. The specifications data provided for the devices in the kit is traceable to NIST through
Agilent Technologies.
4-285057B
Performance Verification
Recertification
Recertification
The following will be provided with a recertified kit:
•a new calibration sticker affixed to the case
•a certificate of calibration
•a calibration report for each device in the kit listing measured values, specifications, and uncertainties
NOTEA list of NIST traceable numbers may be purchased upon request to be included in the
calibration report.
Agilent Technologies offers a Standard calibration for the recertification of the kit. For more information,
contact Agilent Technologies. Refer to “Contacting Agilent” on page 5-7.
How Often to Recertify
The suggested initial interval for recertification is 12 months or sooner. The actual need for recertification
depends on the use of the kit. After reviewing the results of the initial recertification, you may establish a
different recertification interval that reflects the usage and wear of the kit.
NOTEThe recertification interval should begin on the date the kit is first used after the
recertification date.
Where to Send a Kit for Recertification
Contact Agilent Technologies for information on where to send your kit for recertification. Refer to
“Contacting Agilent” on page 5-7. Refer to “Returning a Kit or Device to Agilent Technologies” on page 5-6
for details on sending your kit.
85057B4-3
Performance Verification
Recertification
4-485057B
5Troubleshooting
5-1
Troubleshooting
Troubleshooting Process
Troubleshooting Process
If your PNA does not pass performance verification, follow the steps in
Figure 5-1 to determine the cause of the failure and the correct action to take to correct the failure.
5-285057B
Figure 5-1 Troubleshooting Flowchart
Troubleshooting
Troubleshooting Process
85057B5-3
Troubleshooting
Compatible Network Analyzers
Compatible Network Analyzers
The devices in this kit and their data are compatible with the PNA series network analyzers. The USB drive
provided contains the unique factory-measured S-parameter data for each device in this kit. It also contains
the factory measurement uncertainty used in the PNA system verification procedure to calculate the test
limits.
Older models of this verification kit provided data disks for the 8510, 8720 and 8722 analyzers. Since these
analyzers have been discontinued, the data disks are no longer provided in new kits. When old verification
kits that include the data disks are returned to Agilent for recertification, the disks will be reproduced with
new data for each device in the kit. Please specify your VNA model(s) when returning kits for service or
when ordering kit replacement parts.
5-485057B
Troubleshooting
Where to Look for More Information
Where to Look for More Information
This manual contains limited information about PNA series network analyzer system operation. For detailed
information on using a PNA, refer to the PNA Help system. To do so, press the Help key on the front panel of
the PNA.
If you need additional information, see “Contacting Agilent” on page 5-7.
85057B5-5
Troubleshooting
Returning a Kit or Device to Agilent Technologies
Returning a Kit or Device to Agilent Technologies
If your kit or device requires service, contact Agilent Technologies for information on where to send it. See
“Contacting Agilent” on page 5-7 for information. Include a service tag (located at the back of this manual)
on which you provide the following information:
•your company name and address
•a technical contact person within your company, and the person's complete telephone number including
country code and area code
•the model number and serial number of the kit (if returning a complete kit)
•the model number(s) of your network analyzer(s)
•the part number and serial number of each device being returned
•the type of service required
•a detailed description of the problem (if applicable) and how the device was being used when the
problem occurred
5-685057B
Troubleshooting
Contacting Agilent
Contacting Agilent
Assistance with test and measurements needs and information on finding a local Agilent office are available
on the Web at:
www.agilent.com/find/assist
If you do not have access to the Internet, please contact your Agilent field engineer.
NOTE In any correspondence or telephone conversation, refer to the Agilent product by its model
number and full serial number. With this information, the Agilent representative can
determine whether your product is still within its warranty period.
85057B5-7
Troubleshooting
Contacting Agilent
5-885057B
6Replaceable Parts
6-1
Replaceable Parts
Replacing the Verification Data
Replacing the Verification Data
The verification data contains unique performance data that applies to the individual verification devices. No
two devices have the same performance data. It is not a trivial matter to replace lost or damaged data, so it
is important to make one or more backup copies.
If your verification data is lost or damaged, and you have no backup copies, take one of the following
actions:
•If recertification is not required in the near future.
Contact Agilent for replacement verification data - refer to “Contacting Agilent” on page 5-7. Please
specify the information in the table below.
•If recertification will be required in the near future.
Agilent recommends that you have the verification kit recertified early. New verification data will be
generated during the recertification process. Refer to “Recertification” on page 4-3.
Table 6-1 Information to Specify When Ordering Replacement Verification Data
DeviceModel NumberSerial NumberPart Number
Kit--
Device 1--
Device 2--
Device 3--
Device 4--
Needed: (check) PNA USB drive ___; Data sheets ___
Last Recertification: Date _______________; Serviced by: ___________________________________________
6-285057B
Replaceable Parts
Replaceable Parts
Replaceable Parts
Ta b le 6- 2 lists the replacement part numbers for items included in the 85057B verification kit and Figure 6-1
illustrates the attenuators and airlines.
Ta b le 6- 3 lists the replacement part numbers for items not included in the verification kit that are either
required or recommended for successful operation of the kit.
To order a listed part, note the description, the part number, and the quantity desired. Telephone or send your
order to Agilent Technologies. See “Contacting Agilent” on page 5-7.
Table 6-2 Replaceable Parts for the 85057B 2.4 mm Verification Kit
DescriptionQty
Per Kit
Attenuators
20 dB attenuator with data185057BR01
40 dB attenuator with data185057BR02
Airlines
50
airline with data
mismatch airline with data
25
Verification Kit Storage Box
Storage box (without foam pads)15180-7899
Foam pad (for storage box lid)15180-8490
Foam pad (for storage box lower case)185053-80017
User’s and Service Guide
User’s and service guide
a. See “Printing Copies of This Document” on page iii.
a
185057BR03
185057BR04
185057-90015
Agilent
Part Number
85057B6-3
Replaceable Parts
Replaceable Parts
Figure 6-1Component Identification Sheet for the 85057B
6-485057B
Table 6-3 Items Not Included in the Verification Kit
DescriptionQty Agilent
Connector Gages (2.4 mm)
Male connector gage set
Female connector gage set
Short (for gaging airlines)
a
a
a
Wrenches
Replaceable Parts
Replaceable Parts
Part Number
111752-60108
111752-60107
185056-60020
20 mm,
5/16 in, 90N-cm (8 in-lb) torque
90N-cm (8 in-lb) torque18710-1764
a
18710-1765
5/16 in, open-end wrench18720-0015
Spanner wrench108513-20014
ESD Protection Devices
Grounding wrist strap19300-1367
5 ft grounding cord for wrist strap19300-0980
2 x 4 ft conductive table mat and 15 ft ground wire19300-0797
ESD heel strap (for conductive floors)19300-1308
Connector Cleaning Supplies
Isopropyl alcohol----
Cleaning swabs1009301-1243
a. Included in the 85056A 2.4 mm Calibration Kit.
85057B6-5
Replaceable Parts
Replaceable Parts
6-685057B
Index
A
adapters, 1-3
Agilent Technologies
contacting
agreements
customer assistance
warranty, 1-ii
airline
assembling
attaching the short, 3-11
characteristics
connecting
dimensions
center conductor
outer conductor
disconnecting
gaging
mechanical characteristics
mismatch
part numbers
alcohol, isopropyl
as cleaning solvent
precautions for use of
altitude, specifications
assistance
who to contact
attenuator, part numbers
C
calibration
bad
certificate of
constants, See calibration definitions
definitions
entering
permanently stored
kits, intended to be used
report
standard, 4-3
temperature
center conductor
certificate of calibration, 4-3
characteristics