Operating and Service Manual
Agilent Technologies
85037A/B
Precision Detectors
Serial Numbers
This manual applies directly to Agilent 85037A/B detectors with serial
number 00101 and above.
Part Number: 85037-90013
Printed in USA
November 2001
Supersedes June 1993
© Copyright 1992–1993, 2001 Agilent Technologies, Inc.
Notice.
The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material,
including but not limited to, the implied warranties of merchantability and fitness for a
particular purpose. Agilent shall not be liable for errors contained herein or for incidental
or consequential damages in connection with the furnishing, performatnce, or use of this
material.
Hewlett-Packard to Agilent Technologies Transition
This documentation supports a product that previously shipped under the
Hewlett-Packard company brand name. The brand name has now been changed to Agilent
Technologies. The two products are functionally identical, only our name has changed. The
document still includes references to Hewlett-Packard products, some of which have been
transitioned to Agilent Technologies.
ii
Certification
Agilen tT ec hnologies certies that this pro duct met its published
sp ecications at the time of shipmen t from the factory . Agilen t
further certies that its calibration measuremen ts are traceable to the
United States National Institute of Standards and T ec hnology ,t o
the exten t allo w ed b y the Institute's calibration facilit y , and to the
calibration facilities of other In ternational Standards Organization
mem b ers.
W arranty
This Agilen t instrumen t pro duct is w arran ted against defects in
material and w orkmanship for a p erio d o f one y ear from date of
shipmen t. During the w arran t y p erio d, Agilen t will, at its option,
either repair or replace pro ducts whic h pro v e t o b e defectiv e.
F or w arran t y service or repair, this pro duct m ust b e returned to a
service facilit y designated b y Agilen t. Buy er shall prepa y shipping
c harges to Agilen t and Agilen t shall pa y shipping c harges to return
the pro duct to Buy er. Ho w ev er, Buy er shall pa y all shipping c harges,
duties, and taxes for pro ducts returned to Agilen
t from another
coun try .
Agilen tT ec hnologies w arran ts that its soft w are and rm w are
designated b y Agilen t for use with an instrumen t will execute
its programming instructions when prop erly installed on that
instrumen t. Agilen t d o e s not w arran t that the op eration of the
instrumen t, or soft w are, or rm w are will b e unin terrupted or
error-free.
Limit a tion of W arranty
The foregoing w arran t y shall not apply to defects resulting from
improp er or inadequate main tenance b y Buy er, Buy er-supplied
soft w are or in terfacing, unauthorized mo dication or misuse,
op eration outside of the en vironmen tal sp ecications for the
pro duct, or improp er site preparation or main tenance.
NO OTHER W ARRANTY IS EXPRESSED OR IMPLIED.
HEWLETT-P A CKARD SPECIFICALL Y DISCLAIMS THE
IMPLIED W ARRANTIES OF MER CHANT ABILITY AND
FITNESS F OR A P AR TICULAR PURPOSE.
Ex clusive Remedies
THE REMEDIES PR O VIDED HEREIN ARE BUYER'S SOLE
AND EX CLUSIVE REMEDIES. HEWLETT-P A CKARD SHALL
NOT BE LIABLE F OR ANY DIRECT, INDIRECT, SPECIAL,
INCIDENT AL, OR CONSEQUENTIAL D AMA GES, WHETHER
BASED ON CONTRA CT, TOR T, OR ANY OTHER LEGAL
THEOR Y.
iii
Assistance
Pr o duct maintenanc e agr e ements and other customer assistanc e
agr e ements ar e available for A gilent pr o ducts.
F or any assistanc e, c ontact your ne ar est A gilent T e chnolo gies Sales
and Servic e Oc e.
Safety Notes
Caution
W arning
The follo wing safet y notes are used throughout this man ual.
F amiliarize y ourself with eac h o f the notes and its meaning b efore
op erating this instrumen t.
Caution
denotes a hazard. It calls atten tion to a pro cedure that, if
not correctly p erformed or adhered to, could result in damage to or
destruction of the instrumen t. Do not pro ceed b e y ond a
c aution
sign
un til the indicated conditions are fully understo o d and met.
Warning
denotes a hazard. It calls attention to a procedure which, if
not correctly performed or adhered to, could result in injury or loss
of life. Do not proceed bey ond a
warning
note until the indicated
conditions are fully understood and met.
Instruction
Man ual
The
instruction man ual
sym b ol. The pro duct is mark ed with
this sym b o l when it is necessary for the user to refer to the
instructions in the man ual.
L
iv
Contents
1. General Information
In tro duction . . . . . . . . . . . . . . . . . . 1-1
Pro duct Description . . . . . . . . . . . . . 1-1
Instrumen ts Co v ered b y This Man ual . . . . . . 1-1
Accessories . . . . . . . . . . . . . . . . . . 1-2
Equipmen t Required but Not Supplied . . . . . . 1-2
Recommended T est Equipmen t. . . . . . . . . 1-2
Reection or T ransmission Measuremen ts . . . . 1-2
A C Detection . . . . . . . . . . . . . . . . 1-2
Sp ecications and Supplemen tal Characteristics . . 1-2
2. Installation
Safet y Considerations . . . . . . . . . . . . . . 2-1
Initial Insp ection . . . . . . . . . . . . . . . . 2-1
Preparation for Use . . . . . . . . . . . . . . . 2-2
Electrostatic Disc harge (ESD) . . . . . . . . . 2-2
Static-Safe W orkstation . . . . . . . . . . . 2-2
Static-Safe Practices . . . . . . . . . . . . 2-2
P o w er Requiremen ts . . . . . . . . . . . . . 2-3
Cable Lead Iden tication . . . . . . . . . . . 2-3
Mating Connectors . . . . . . . . . . . . . . 2-3
Connecting an 85037A/B . . . . . . . . . . . 2-3
Mating a Precision 3.5 mm Connector to an SMA
Connector . . . . . . . . . . . . . . . 2-4
Electrical P erformance . . . . . . . . . . 2-4
Op erating En vironmen t. . . . . . . . . . . . . 2-6
Storage and Shipmen t. . . . . . . . . . . . . . 2-6
En vironmen t. . . . . . . . . . . . . . . . . 2-6
P ac k aging . . . . . . . . . . . . . . . . . . 2-6
Returning a Detector for Service . . . . . . . . . 2-7
3. Op eration
Op erating Theory . . . . . . . . . . . . . . . 3-1
A C Detection . . . . . . . . . . . . . . . . 3-1
DC Detection . . . . . . . . . . . . . . . . 3-1
F eatures . . . . . . . . . . . . . . . . . . . . 3-2
Connector T orque V alues . . . . . . . . . . . . 3-2
Op erator's Chec k. . . . . . . . . . . . . . . . 3-3
Op erating Mo des and Sp ecications . . . . . . . 3-3
Dynamic Accuracy . . . . . . . . . . . . . . 3-3
F requency Resp onse . . . . . . . . . . . . . 3-3
Measuremen t System Conguration . . . . . . . . 3-4
Contents-1
DC Detection Measuremen ts . . . . . . . . . . . 3-5
Enabling the DC Detection Mo de . . . . . . . 3-5
Making Accurate DC Measuremen ts . . . . . . 3-5
Op eration . . . . . . . . . . . . . . . . . 3-5
Zeroing an 85037A/B . . . . . . . . . . . . 3-5
A C Detection Measuremen ts . . . . . . . . . . . 3-6
Making Accurate A C Measuremen ts . . . . . . 3-6
Characterizing the Detector (P erforming a P o w er
Calibration) . . . . . . . . . . . . . . . . 3-7
4. P erformance T ests
In tro duction . . . . . . . . . . . . . . . . . . 4-1
Equipmen t Required . . . . . . . . . . . . . 4-1
Return Loss at a Nominal 0 10 dBm . . . . . . . 4-2
Description . . . . . . . . . . . . . . . . . 4-2
Pro cedure . . . . . . . . . . . . . . . . . . 4-3
F requency Resp onse at a Nominal 0 10 dBm . . . . 4-5
Description . . . . . . . . . . . . . . . . . 4-5
Pro cedure . . . . . . . . . . . . . . . . . . 4-6
Conguring the System . . . . . . . . . . . . 4-6
Measuring the Source . . . . . . . . . . . . . 4-7
Measuring the Detector . . . . . . . . . . . . 4-7
Computing the Maxim um Error . . . . . . . . 4-8
Dynamic P o w er Accuracy . . . . . . . . . . . . 4-10
Description . . . . . . . . . . . . . . . . . 4-10
Pro cedure . . . . . . . . . . . . . . . . . . 4-10
First Alternate Dynamic P o w er Accuracy . . . . . 4-11
Description . . . . . . . . . . . . . . . . . 4-11
Second Alternate Dynamic P o w er Accuracy . . . . 4-12
Description . . . . . . . . . . . . . . . . . 4-12
Pro cedure . . . . . . . . . . . . . . . . . . 4-12
Contents-2
5. Adjustmen ts
In tro duction . . . . . . . . . . . . . . . . . . 5-1
Coarse Zero Adjustmen t. . . . . . . . . . . . . 5-2
Description . . . . . . . . . . . . . . . . . 5-2
Pro cedure . . . . . . . . . . . . . . . . . . 5-2
F eedthrough Nulling Adjustmen t. . . . . . . . . 5-3
Description . . . . . . . . . . . . . . . . . 5-3
Pro cedure . . . . . . . . . . . . . . . . . . 5-3
6. Replaceable P arts
In tro duction . . . . . . . . . . . . . . . . . . 6-1
Ordering P arts . . . . . . . . . . . . . . . . . 6-1
7. Service
Error Messages . . . . . . . . . . . . . . . . . 7-1
Repair . . . . . . . . . . . . . . . . . . . . 7-1
Replacing the Detector . . . . . . . . . . . . 7-2
Remo ving the Co v ers . . . . . . . . . . . . . 7-2
Replacing the Cable Assem bly (W1) . . . . . . 7-3
A. Caring for Connectors
Visual Insp ection . . . . . . . . . . . . . . . . A-1
Cleaning . . . . . . . . . . . . . . . . . . A-1
Mec hanical Insp ection . . . . . . . . . . . . . A-2
Index
Contents-3
Figures
2-1. Example of a Static-Safe W orkstation . . . . . . 2-2
2-2. Using Precision 7 m m Connectors . . . . . . . 2-3
2-3. SMA and 3.5 mm Connector Cross-Sections and
SWR P erformance . . . . . . . . . . . . . 2-5
3-1. Detector F eatures . . . . . . . . . . . . . . 3-2
3-2. T ypical Measuremen t Setup . . . . . . . . . . 3-4
4-1. Return Loss Setup . . . . . . . . . . . . . . 4-2
4-2. F requency Resp onse T est Setup . . . . . . . . 4-5
4-3. F requency Resp onse Graph . . . . . . . . . . 4-9
4-4. Dynamic P o w er Accuracy T est Setup . . . . . . 4-11
4-5. First Alternate Dynamic P o w er Accuracy T est Setup 4-11
4-6. Second Alternate Dynamic P o w er Accuracy T est
Setup . . . . . . . . . . . . . . . . . . 4-12
5-1. Coarse Zero/F eedthrough Nulling Adjustmen t . . 5-2
5-2. Coarse Zero Adjustmen t. . . . . . . . . . . . 5-3
5-3. F eedthrough Nulling Adjustmen t. . . . . . . . 5-4
6-1. Detectors and Cable Mark er Kit . . . . . . . . 6-2
6-2. Mo dule Exc hange Program . . . . . . . . . . 6-4
7-1. Remo ving the Detector Co v ers . . . . . . . . . 7-2
7-2. Cable Connections . . . . . . . . . . . . . . 7-4
T ables
Contents-4
1-1. Agilen t 85037A Standard 1 . . . . . . . . . . . 1-3
1-2. Agilen t 85037B Sp ecications 1 . . . . . . . . . 1-4
1-3. Supplemen tal Characteristics . . . . . . . . . 1-5
1-4. Recommended T est Equipmen t. . . . . . . . . 1-6
4-1. Return Loss Equipmen tT able . . . . . . . . . 4-3
4-2. F requency Resp onse Equipmen tT able . . . . . . 4-6
4-3. W orksheet . . . . . . . . . . . . . . . . . . 4-9
4-4. Additional Equipmen t . . . . . . . . . . . . 4-12
4-5. T est Record for 85037A . . . . . . . . . . . . 4-13
4-6. T est Record for 85037A Option 001 . . . . . . . 4-17
4-7. T est Record for 85037B . . . . . . . . . . . . 4-21
6-1. Miscellaneous P arts . . . . . . . . . . . . . . 6-2
6-2. Replaceable P arts and Accessories . . . . . . . 6-3
6-3. Con tacting Agilen t . . . . . . . . . . . . . . 6-5
General Information
1
Introduction
Product Description
This man ual con tains information on op erating, testing, and servicing
the Agilen t 85037A and 85037B precision detectors. Figure 6-1 sho ws
the detectors and the supplied cable mark er kit.
The Agilen t 85037A/B precision detectors are sp ecically designed
for use with an Agilen t 8757D scalar net w ork analyzer and are
not compatible with the Agilen t 8757A/C/E, 8756, or 8755 scalar
net w ork analyzers. These dual 0 dio de detectors ma y b e used in
either A C o r D C detection mo des. F or impro v ed p o w er measuremen t
accuracy v ersus frequency , eac h Agilen t 85037 series precision
detector includes detector sp ecic frequency resp onse data, stored in
an in ternal EEPR OM, whic h i s automatically read b y the 8757D.
When used in conjunction with the 8757D's in ternal p o w er calibrator
(Option 002), these detectors pro vide the maxim um absolute p o w er
measuremen t accuracy .
Agilen t Detector Connector T yp e
85037A T yp e-N (m)
85037A Option 001 Precision 7 m m
85037B Precision 3.5 mm (m)
Instruments Co v ered b y
This Manual
Eac h detector has a unique serial n um b er. The con ten ts of this
man ual apply directly to detectors with serial n um b ers listed on the
title page.
General Information 1-1
Accessories
Equipment Required
but Not Supplied
T able 6-2 lists accessories a v ailable for use with these detectors.
Recommended T est
Equipment
Reflection or
Transmission
Measurements
A C Detection
T able 1-4 lists the equipmen t required to test the detectors. Y ou
ma y substitute an y equipmen t that meets the indicated critical
sp ecications.
Reection and transmission measuremen ts require the follo wing
equipmen t:
One or more detectors.
An Agilen t 8757D scalar net w ork analyzer.
One
of the follo wing:
A directional bridge.
A directional coupler.
As w eep oscillator or syn thesized sw eep er.
F or ratio measuremen ts use:
Ap o w er splitter or dual directional coupler.
A C detection requires the equipmen t listed ab o v e,
plus
either a signal
source capable of 27.778 kHz squarew a v e amplitude mo dulation, or
an external mo dulator.
Specifications and
Supplemental
Characteristics
1-2 General Information
T able 1-1 and T able 1-2 list detector sp ecications whic h are the
p erformance standards or limits against whic hy ou can test the
device.
T able 1-3 lists supplemen tal (t ypical, non-w arran ted) detector
c haracteristics.
T able 1-1. Agilent 85037A Standard
Connector
Standard T yp e-N (m)
Option 001 Precision 7 m m
F requency Range
0.01 to 18 GHz
Return Loss
0.01 to 0.04 GHz 10 dB
0.04 to 18 GHz 20 dB
F requency Resp onse
0.01 to 0.04 GHz
0.04 to 18 GHz
2
6
0.35 dB
6
0.18 dB
Dynamic Range
A Cm o d e +20 to 0 55 dBm
DC mo de +20 to 0 50 dBm
DC Absolute P o w er Accuracy
3, 5
P o w er Corrected Default
(dBm) ( 6 dB) ( 6 dB)
+20 0.25 0.40
+10 0.11 0.40
0
30 0.11 0.40
0
40 0.40 0.80
0
50 0.85 1.30
A C Dynamic P o w er Accuracy
3, 4
1
P o w er Corrected Default
(dBm) ( 6 dB) ( 6 dB)
+20 0.25 0.40
+10 0.11 0.40
0
30 0.11 0.40
0
40 0.40 0.80
0
50 0.85 1.30
0
55 0.85 1.30
T emp erature Co ecien t o f Linearit y
1
The 85037A/B sp ecications only apply when used with the 8757D scalar net w ork analyzer.
2
0
10 dBm, 25 6 5 C
3
The corrected sp ecications apply after a calibration via the 8757D Option 002 in
The default sp ecications apply when the calibrator is not used. P
0.01 dB/ C temp. c hange after calibration.
ternal calibrator.
o w er calibrator uncertain t yi s
included in the 85037A/B corrected sp ecications.
4
Dynamic accuracy refers to measuremen t accuracy as p o w er v aries (in dB) from a 0 dBm reference.
25 6 5 C, 50 MHz, calibration and measuremen t a t the same temp erature.
5
DC mo de, 25 6 5
C, calibration and measuremen t a t the same temp erature.
General Information 1-3
T able 1-2. Agilent 85037B Specifications
1
Connector
F requency Range
3.5 mm (m)
0.01 to 26.5 GHz
Return Loss
0.01 to 0.04 GHz 10 dB
0.04 to 18 GHz 20 dB
18 to 26.5 GHz 18 dB
F requency Resp onse
0.01 to 0.04 GHz
0.04 to 18 GHz
18 to 26.5 GHz
2
6
0.35 dB
6
0.18 dB
6
0.22 dB
Dynamic Range
A Cm o d e +20 to 0 55 dBm
DC mo de +20 to 0 50 dBm
DC Absolute P o w er Accuracy
P o w er Corrected
3, 5
3
(dBm) ( 6 dB) ( 6 dB)
+20 0.25 0.40
+10 0.11 0.40
0
30 0.11 0.40
0
40 0.40 0.80
0
50 0.85 1.30
A C Dynamic P o w er Accuracy
3, 4
Default
P o w er Corrected Default
(dBm) ( 6 dB) ( 6 dB)
+20 0.25 0.40
+10 0.11 0.40
0
30 0.11 0.40
0
40 0.40 0.80
0
50 0.85 1.30
0
55 0.85 1.30
T emp erature Co ecien t o f Linearit y
1
The 85037A/B sp ecications only apply when used with the 8757D scalar net w ork analyzer.
2
0
10 dBm, 25 6 5 C
3
The corrected sp ecications apply after a calibration via the 8757D Option 002 in
The default sp ecications apply when the calibrator is not used. P
0.01 dB/ C temp. c hange after calibration.
ternal calibrator.
o w er calibrator uncertain t yi s
included in the 85037A/B corrected sp ecications.
4
Dynamic accuracy refers to measuremen t accuracy as p o w er v aries (in dB) from a 0 dBm reference.
25 6 5
C, 50 MHz, calibration and measuremen t a t the same temp erature.
5
DC mo de, 25 6 5
C, calibration and measuremen t a t the same temp erature.
1-4 General Information
T able 1-3. Supplemental Characteristics
Cable Length
W eigh t
Net 0.24 kg (0.5 lb)
Shipping 1 k g (2.2 lb)
RF Connector Mec hanical T olerances
T yp e-N male
Recession of the male cen ter conductor 0.207 to 0.210 in
Precision 7 m m
Recession of the cen ter conductor
Collet resilience After y ou depress the collet, it m ust spring bac k out.
Precision 3.5 mm
Recession of the male cen ter conductor 0 t o 0.003 in
1
Because a t yp e-N gage calibration blo c k zeros the gage at a 0.207 inc h oset, the gage displa
0.000 to 0.003 inc hes.
2
With the cen ter conductor collet remo v ed.
(85037A)
(85037A Option 001)
2
(85037B)
1.22m (48 in)
1
0 t o 0.003 in
ys a 0.207 to 0.210 inc h oset as
General Information 1-5
T able 1-4. Recommended T est Equipment
Critical
Item
Sp ecications
T yp e-N Connector Gage Kit Required for 85037A
Agilen t M o del or
P art Num b e r
85054-80011
No substitute
7 m m Connector Gage Kit Required for 85037A Option 001
85050-80012
No substitute
3.5 mm Connector Gage Kit Required for 85037B
11752-60105
No substitute
Collet Extractor T oo l Required for 85037A Option 001
5060-0370
No substitute
Syn thesized Sw eep er Compatible with 8757D 83620A/30A
Scalar Net w ork Analyzer No substitute 8757D
Directional Bridge Compatible with 8757D
85027A/B/C
A C/DC detection
P o w er Meter 436A, 437B or 438A
Calibrated P o w er Sensor T yp e-N
Precision 7 m m
Precision 3.5 mm
10 dB A tten uator T yp e-N
Precision 7 m m
Precision 3.5 mm
8481A
8481A Option 001
8485A
8491B Option 010
8492A Option 010
8493C Option 010
Calibrated Short T yp e-N 11511A
Shielded Op en T yp e-N 85032-60001
Calibrated Short/Shielded Op en Precision 3.5 mm 85037-60001
Precision 7 m m 85021-60001
Adapter T yp e-N (f ) t o BNC (m)
T yp e-N (m) to T yp e-N (m)
3.5 mm (f ) t o 3.5 mm (f )
T yp e-N (m) to APC-7
1250-1534
1250-1475
1250-1749
11525A
(Option 001 only)
T orque W renc h Precision 7 m m 1 2 in-lb
8710-1766
(85037A Option 001)
20 mm 8 in-lb (85037B) 8710-1764
1-6 General Information
Installation
2
Safety
Considerations
Initial Inspection
There are no hazardous v oltages in this detector.
1. Chec k the shipping con tainer and pac k aging material for damage.
2. Chec k that the shipmen t i s complete.
3. Chec k connector, cable, and detector b o dy for mec hanical damage.
4. Chec k the detector electrically:
Either mak e a measuremen t o r test to sp ecications. (See Chapter
3, \Op eration" or Chapter 4, \P erformance T ests.")
If an y o f the follo wing conditions exist, notify y our nearest Agilen t
oce:
incomplete shipmen t
mec hanical damage or defect
failed electrical test
If y ou nd damage or signs of stress to the shipping con
cushioning material, k eep them for the carrier's insp ection. Agilen t
do es not w ait for a claim settlemen t b efore arranging for repair or
replacemen t.
tainer or the
Installation 2-1
Preparation for Use
Caution
Electrostatic Discharge
(ESD)
Do not sub ject the detector to mec hanical sho c k.
ESD can damage the highly sensitiv e micro circuits in this device;
c harges as lo w a s 100 V can destro y a detector.
ESD damage o ccurs most often as y ou connect or disconnect a
device. Use this detector at a static-safe w orkstation and w ear a
grounding strap.
Never
touc h the input connector cen ter con tacts or
the cable con tact pins.
Static-Safe
W orkstation
Figure 2-1 illustrates a static-safe station using t w ot yp es of ESD
protection that y ou can use either together or separately .
A conductiv e table mat and wrist-strap com bination.
A conductiv e o o r mat and heel-strap com bination.
2-2 Installation
Figure 2-1. Example of a Static-Safe W orkstation
Static-Safe
Practices
Before cleaning, insp ecting, or making a connection to a
static-sensitiv e device or test p ort, ground y ourself as far as
p ossible from the test p ort.
Disc harge static electricit y from a device b efore connecting it:
T ouc h the device briey (through a resistor of at least 2 M) to
either the outer shell of the test p ort, or another exp osed ground.
This disc harges static electricit y and protects test equipmen
circuitry .
t
P o w er Requirements
The scalar net w ork analyzer supplies p o w er for the detector.
Cable Lead
Identification
Mating Connectors
Caution
Connecting an
85037A/B
When y ou use more than one detector, use the co ded cable clips
(from the cable mark er kit) to iden tify leads; place matc hing clips on
eac h cable, one at eac h end.
T able 1-3 lists connector mec hanical tolerances.
Conne ctor Car e
(08510-90064) pro vides information on the prop er
Micr owave
main tenance, insp ection, and gaging of connectors. Use the
appropriate torque wrenc h. (See T able 1-4 .)
When tigh tening a connector, do not apply more than the
recommended torque v alue. If y ou torque the connectors with more
pressure than is recommended it can deform the mating surfaces.
Connect a detector to the net w ork analyzer as follo ws:
1. With the cable plug k ey do wn w ard, insert the DC connector in to
the analyzer mating connector.
outer
2. T o secure the DC connector in the analyzer, turn the
shell
clo c kwise.
3. F or a standard detector, connect the RF input b
male connector
outer
shell clo c kwise.
y turning the
4. F or a precision (Option 001) 7 m m connector, refer to Figure 2-2 .
Figure 2-2. Using Precision 7 m m Connectors
Installation 2-3
Mating a Precision 3.5 mm Connector to an SMA Connector
It is p ossible to mate a precision 3.5 mm connector to an SMA
connector, but this is not ideal b ecause the t w o connectors ha v e
sligh tly dieren t dimensions and mec hanical c haracteristics. Mating
a precision 3.5 mm connector to an SMA connector also aects the
electrical p erformance. (See \Electrical P erformance".) Use the
follo wing pro cedure to safely mate 3.5 mm and SMA connectors.
1. Insp ect the SMA connector.
Nev er mate a precision 3.5 mm connector to an SMA connector
in whic h the solid plastic dielectric protrudes in fron t o f the outer
conductor mating plane.
2. Gage b oth connectors. The SMA connector m ust meet the
precision 3.5 mm connector setbac k s p ecications. If not, it will
damage the 3.5 mm connector.
In some SMA connectors, the male con tact pins are not held
securely and are easily pulled out of sp ecication (esp ecially if
the female connector con tact ngers are tigh t). Also, some SMA
male pins are not true pins, but are the cut-o ends of the cen
ter
conductor in semi-rigid coaxial cable. In this case, misalignmen
and burrs are lik ely to o ccur.
t
3. Carefully align the connectors.
4. Push the t w o connectors together with the male con tact pin
precisely concen tric with the female.
5. Do not t wist either connector or device.
6. T urn only the outer male connector n ut.
7. Use a 6 0 N-cm (5 in-lb) torque wrenc
h for the nal connection.
If y ou m ust mak e more than a few connections, use a 3.5 mm-to-3.5
mm adapter to protect the 3.5 mm connector.
Electrical P erformance.
The electrical p erformance of the junction of
t w o precision 3.5 mm connectors is sup erior to the junction of either
t w o SMA connectors, or an SMA connector mated to a precision 3.5
mm connector. (See Figure 2-3 .)
When y ou mate an SMA connector with a precision 3.5 mm
connector, the connection has a t ypical mismatc h (SWR) of
1.10 at 2 GHz. This mismatc h i s less than that of t w o SMA
connectors, but is m uc h higher than that of t w o precision 3.5 mm
connectors.
2-4 Installation
Figure 2-3. SMA and 3.5 mm Connector Cross-Sections and SWR P
erformance
Installation 2-5
Operating
Environment
T emp erature
Humidit y
: Up to 95%. Protect the detector from temp erature
: 0 C t o +55 C.
extremes, whic h can cause condensation.
Storage and
Shipment
Environment
P ackaging
Altitude
: Up to 7,620m (25,000 ft).
Store or ship the detectors in en vironmen ts within the follo wing
limits:
T emp erature : 0
Humidit y
: Up to 95%. Protect the detector from temp erature
25 C t o +75 C.
extremes whic h can cause condensation.
Altitude
: Up to 7,620m (25,000 ft).
Con tainers and materials iden tical to those used in factory-pac k aging
are a v ailable. Con tact y our lo cal Agilen t oce for information. If y ou
pac k age the detector using commercially a v ailable material, follo w
these instructions:
1. W rap the detector in hea vy pap er.
2. Use a strong shipping con tainer that has a double-w all carton of
at least 350-p ound test material.
3. Pro vide a rm cushion that prev en ts mo v emen t inside the
con tainer.
4. Use a 5 t o 7 cm (3 to 4 inc h) la y er of sho c k-absorbing material
around all sides of the detector.
5. Seal the shipping con tainer securely .
6. Mark the shipping con tainer
FRAGILE
.
2-6 Installation
Returning a
Detector for Service
If y ou ship the detector to a Agilen t oce or service cen ter ll out a
blue service tag (pro vided at the bac k o f this man ual), and include
the follo wing information:
1. Compan y name and address.
Do not
cannot b e returned to a P .O. b o x.
2. A tec hnical con tact p erson with a complete phone n um b er.
3. The complete mo del and serial n um b e r o f the detector.
4. The t yp e o f service required (calibration, repair).
5. An y other information, suc h a s failure condition or cause, that
could exp edite service.
When y ou mak e a n inquiry , either b y mail or b y telephone, refer to
the detector b y b oth mo del n um b e r and full serial n um b er.
use an address with a P .O. b o xn um b e r b ecause pro ducts
Installation 2-7
Operation
3
Operating Theory
A C Detection
DC Detection
Cautions
The 85037A/B can detect either unmo dulated RF signals (DC mo de)
or square w a v e amplitude mo dulated RF signals (A C m o de). In
either detection mo de, the detector pro vides a 27.778 kHz square
w a v e signal to the analyzer to in terpret and displa y .
In A C detection, an RF or micro w a v e signal is amplitude mo dulated
with a 27.778 kHz square w a v e. The detector demo dulates (en v elop e
detects) and amplies this signal to pro duce a 27.778 kHz square
w a v e whose p eak-to-p eak v oltage corresp onds to the magnitude of
the original RF signal.
DC mo de requires no mo dulation. The detector dio de con
v erts
the RF signal in to DC v oltage whic h i s then c hopp ed at a
27.778 kHz rate and amplied. The amplied signal is a signal
lik e that pro duced b yA C detection.
Electrostatic disc harge (ESD) can damage the highly sensitiv e
micro circuits in this device; c harges as lo w a s 100 V can destro yy our
detector.
ESD damage o ccurs most often as y ou connect or disconnect a
device. Use this detector at a static-safe w orkstation and w ear a
grounding strap.
Never
touc h the input connector cen ter con tact, or
the cable con tact pins.
Do
not
exceed the recommended torque sp ecication to tigh ten a
connector b ecause greater surface torque can deform the mating
surfaces.
Do
not
apply more than +23 dBm RF CW p o w er or more
than 6 10 Vdc to the detector. Higher p o w er/v oltage can electrically
damage the detector.
Before y ou connect a cable to the detector, alw a ys disc harge the
cable's cen ter conductor static electricit y t o instrumen t-ground.
Do not sub ject the detector to mec hanical sho c k.
Operation 3-1
Features
Connector T orque
V alues
Figure 3-1. Detector Features
Tigh ten the 85037A t yp e-N connector nger-tigh t only .
Tigh ten the 85037A Option 001 with a torque wrenc
h, part n um be r
8710-1766 set at 13.8 cm-kg (12 in-lb).
Tigh ten the 85037B with a torque wrenc h part n um b e r 8710-1764
set at 9.2 cm-kg (8 in-lb).
3-2 Operation
Operator's Check
See \Characterizing the Detector" in this c hapter for a pro cedure
that allo ws y ou to quic kly c hec k the detector. This pro cedure can b e
used as a daily c hec k.
Operating Modes
and Specifications
Dynamic Accuracy
Frequency Response
The Agilen t 85037 series precision detectors ha v e built-in corrections
for b oth frequency range and dynamic accuracy . These corrections
are used to enhance the measuremen t capabilit y o f the detector. The
follo wing text explains ho w the corrections are made. Additional
information regarding detector op eration is a v ailable in the
Sc alar Network A nalyzer Op er ating Manual
.
8757D
The dynamic accuracy and absolute p o w er accuracy of the detector is
measured at the factory . Appro ximately 150 correction constan ts are
then calculated and stored in the EEPR OM of eac h detector. This
pro vides a default correction and is sucien t t o guaran tee the default
dynamic accuracy sp ecications. T o obtain maxim um p erformance,
the detector m ust b e connected to the 50 MHz calibrator output
of the 8757D Option 002 and then c haracterized using the built-in
correction routine in the 8757D. Once this is done, the detector will
meet the enhanced (corrected) sp ecication for dynamic accuracy for
b oth A C and DC measuremen ts. This c haracterization will remain
v alid un til the detector is unplugged. F or optimal p erformance
ho w ev er, the detector m ust b e rec haracterized if its temp erature
c hanges b y more than 6 5
C.
The 85037 uses built-in correction factors to pro vide the b est p ossible
frequency resp onse. The n um b e r o f correction factors will v ary
with eac h detector, but is t ypically 25. These correction factors are
determined at the factory and are stored within the EEPR OM of the
detector. If the 8757D is used with the system in terface, then the
analyzer will automatically determine the prop er correction factor to
use; no user in terv en tion is required. If the 8757 system in terface
is not b eing used, then the frequency related correction factors
will b e determined from the start/stop lab els. These lab els can b e
en tered in to the 8757 b y pressing
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STOP LABEL
) and then en tering the start (or stop) frequency .
4
SYSTEM
NNNNNNNNNNNNNNNNNNNN
5
LABELS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
START LABEL
(or
Operation 3-3
Ho w ev er, if the frequency the detector sees is dieren t from what is
b eing sw ept (for example, when using a mixer or m ultiplier), then the
user m ust man ually en ter the correct detector frequencies as follo ws:
Measurement
System
Configuration
1. Press
2.
3.
4.
5.
6.
7.
Select
Select
Select
Select
Select
Select
4
5
.
CAL
NNNNNNNNNNNNNN
MORE
.
NNNNNNNNNNNNNNNNNNNNNNNNNN
DET FREQ
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DET A,B,C or R
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Start Freq
and use the k eypad to en ter a v alue.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Stop Freq
and use the k eypad to en ter a v alue.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DET FREQ ON
.
.
Figure 3-2 sho ws a t ypical measuremen t setup. A C detection mo de,
the turn-on state of the 8757D analyzer, do es not require an
y s p ecial
setup to initiate a measuremen t. DC detection mo de, ho w ev er,
requires that y ou press sp ecic analyzer k eys in the correct order.
3-4 Operation
Figure 3-2. Typical Measurement Setup
DC Detection
Measurements
DC detection oers absolute p o w er measuremen t capabilit y and the
abilit y t o c haracterize oscillators and mo dulation sensitiv e devices.
Enabling the DC
Detection Mode
Remember
Making Accurate DC
Measurements
1. On the analyzer, press
4
PRESET
5
.
2. Connect the detector.
3. Enable DC mo de:
NNNNNNNNNNNNNNNNNNNNNNN
Press
4
SYSTEM
5
and select
MODE DC
.
NNNNNNNNNNNNNNNNNNNNNNN
Selecting the
MODE DC
softk ey also turns o the source square
w a v e m o dulation.
Y ou m ust enable the DC mo de to access the DC-sp ecic softk eys.
T o mak e DC-mo de measuremen ts as sho wn in Figure 3-2 ,y ou
m ust use an 85027 directional bridge capable of b oth A C and DC
detection.
Operation
Before accurate DC measuremen ts can b e made, t w o quic k routines
m ust b e p erformed; a D C zero and a dynamic p o
w er c haracterization.
(This is referred to as a p o w er calibration on the 8757 and on other
annotations.) The DC zero will impro v e accuracy b elo w 0 35 dBm.
The dynamic c haracterization will pro vide the impro v ed (corrected)
po w er accuracy sp ecication on the 8787D Option 002 only
Measuremen ts can b e p erformed without this c
haracterization (suc h
.
as on instrumen ts without Option 002), ho w ev er only the default
dynamic accuracy sp ecications of the detector are guaran
teed. This
c haracterization includes a D C zero so it is not necessary to p erform
the zero b efore p erforming the c
haracterization.
Zeroing an 85037A/B
When making DC{mo de measuremen ts, y ou m ust p erform a zeroing
op eration to comp ensate for the eects of DC drift and temp erature
uctuations. This is not required in A C detection. The zeroing
op eration eliminates small DC v oltages presen t i n the detector that
w ould otherwise cause amplitude measuremen t errors at lo wp o w er
lev els ( 0 35 dBm and b elo w). Zeroing also establishes the displa y ed
noise lev el (the system noise o or) with no RF signal applied. A
DC zero should b e p erformed at least once ev ery half-hour once the
system has stabilized. A D C zero should b e p erformed more often
as the system w arms up, or if the temp erature is c hanging. The
autozero function is pro vided to automatically p erform DC zeros
p erio dically when the system in terface is used.
Operation 3-5
Note
Before y ou zero the detector, remo v e
al l
RF signals from the detector
input. Ev en a small RF signal presen t during zeroing pro duces
measuremen t errors.
The 8757D analyzer has three t yp es of zeroing:
Autozero
NNNNNNNNNNNNNNNNNNNNNNN
AUTOZRO
turns o the source RF signal output and
automatically zero es the detector.
Rep eat
Autozero
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REPT AZ ON/OFF
selected in terv als.
automatically rep eats autozero at
NNNNNNNNNNNNNNNNNNNN
Man ual Zero
MANUAL
is similar to p o w er meter zeroing. First y ou
m ust either ph ysically remo v e the detector from the
RF signal or turn the RF signal o, then y ou select
NNNNNNNNNNNNNNNNNNNN
MANUAL
to p erform the zero.
Note
A C Detection
Measurements
Making Accurate A C
Measurements
See the analyzer do cumen tation for more information on these
softk eys.
A C detection is the preferred metho d for the ma jorit
yo f
measuremen ts as it oers greater sensitivit y . I t also oers
imm unit y t o noise and drift with time and temp erature. A
C
detection amplitude measuremen ts require a m o dulation en v elop e.
The en v elop e i s pro vided through a 27.778 kHz square w a v e
amplitude mo dulation of the RF test signal. T
v ary dep ending on the source. Figure 3-2 illustrates a t
est set connections
ypical
measuremen t setup with an Agilen t 8360 syn thesized sw eep er
supplying the 27.778 kHz mo dulation.
Accurate A C measuremen ts dep end hea vily on the en v elop e o f
the amplitude mo dulated RF signal. In addition, to obtain the
sp ecied dynamic accuracy ,a c haracterization of the detector (a
po w er cal) m ust rst b e p erformed. Without this c haracterization,
only the default dynamic accuracy is guaran teed. P erforming the
c haracterization will calibrate the detector in
b oth
A C and DC
mo des. See \Characterizing the Detector" in this c hapter.
3-6 Operation
The RF signal m ust b e squarew a v e m o dulated at a rate within 20
Hz of 27.778 kHz. The amplitude ratio of the on p ortion to the
o p ortion of the en v elop e m ust b e at least 30 dB and the ratio of
on-time to o-time should b e within 5% of 50/50. Most Agilen t
sources pro vide this capabilit y .I f y our 8757D scalar net w ork analyzer
is equipp ed with the Option 002 p o w er calibrator, y ou can obtain the
b est p o w er measuremen t accuracy b y follo wing this pro cedure.
Characterizing the
Detector (P erforming
aP o w er Calibration)
T o obtain the b est accuracy , and to meet the \corrected" dynamic
accuracy sp ecications, eac h detector m ust b e c haracterized on the
8757D input on whic h i t will b e used. The c haracterization routine
sim ultaneously corrects b oth A C
and
DC measuremen ts. This
routine is also useful as an op erator's c hec k t o ensure the detector
is op erating prop erly . Only the 8757D with an Option 002 p o w er
calibrator ha v e this capabilit y .
Detector c haracterization b egins at +20 dBm and decreases in 1 d B
incremen ts do wn to 0 35 dBm. T o reduce measuremen t time, p o w er
steps are larger than 1 d B a t l o w er p o w er lev els. Correction v alues
will b e generated o v er this range if the detector is op erational. It
will also correct for an y loss in the detector up to ab out 50 dB. F or
example, lets assume a D C detector has a dynamic range of +20 to
0
50 dBm. A 3 0 d B atten uator is then placed on the detector and a
c haracterization is p erformed. Up on completion, the detector will b e
c haracterized from +20 to 0 20 dBm. Applying a 0 dBm signal to
the detector/pad com bination will result in a 0 dBm reading; not a
0
30 dBm reading whic h i s what the detector itself is actually seeing.
A loss of more than ab out 50 dB ma
y result in an error message
b eing displa y ed on the analyzer. This could o ccur if the detector is
defectiv eo r i f y ou simply forgot to connect the detector under test to
the P o w er Cal Output of the 8757D.
Up on completion of this routine, the dynamic range of the detector
will b e displa y ed; v erify that it is correct for the curren t detection
mo de, taking in to accoun ta n y atten uation due to added pads or
other devices, or due to inheren t loss, suc h a s i n a directional bridge
whic ht ypically has 12.5 dB of loss.
T oc haracterize the detector (p erform a p o w er cal), connect the
detector to the desired input of the 8757D. Connect the RF input of
the detector to the calibrator output. Mak e sure the analyzer has
b een on and the detector connected for at least 30 min utes.
1. Press
2.
3.
4.
5.
Select
Select
Select
Select
4
5
.
CAL
NNNNNNNNNNNNNN
MORE
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER CAL
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DET A,B,C or R
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
START CAL
.
.
A complete c haracterization for one detector tak es ab out 60 seconds.
Once c haracterized, do
not
unplug the detector from the analyzer or
y ou will ha v e t o p erform another c haracterization. This is b ecause
the 8757 automatically recalibrates itself when it senses a detector
has b een remo v ed or connected to the detector inputs. Ho
the analyzer
c an
b e turned o and on again without losing this
w ev er,
information. P erform the c haracterization at least once p e r d a y and
more often if the temp erature v aries b y more than 6 5
C.
Operation 3-7
P erformance T ests
4
Introduction
Equipment Required
Note
Use the pro cedures in this c hapter to test the detector's electrical
p erformance to the sp ecications listed in Chapter 1. None of these
tests require access to the detector's in terior. T o completely test eac h
detector, three tests are required:
1. return loss
2. dynamic accuracy (A C and DC uncorrected)
3. frequency resp onse
Preceding eac h test is an equipmen t table that lists whic h equipmen t
y ou will need for that particular test. Y ou ma y substitute an y
equipmen t that meets the indicated critical sp ecications. Refer to
T able 1-4 .
Before y ou p erform a p erformance test, gage the input connector
on the detector and en ter the results in the test record at the
end of \P erformance T ests." F or descriptiv e illustrations dening
connector tolerances, see
08510-90064).
Micr owave Conne ctor Car e
(part n um be r
P erformance T ests 4-1
Return Loss at a
Nominal 0 10 dBm
Description
The return loss of the 85037 can b e measured using the test system
describ ed in this pro cedure. The test setup is calibrated using an
op en/short to minimize frequency resp onse and phasing errors. Then
the detector under test (DUT) is connected to the
bridge, and its return loss is measured on the 8757D.
The three main sources of error in these measuremen ts come from:
1. bridge directivit y
2. source matc h o f the bridge
3. dynamic accuracy of the 8757D
The rst t w ov ary with frequency while dynamic accuracy v aries
with the measured return loss amplitude. Use the corresp onding,
connector-compatible Agilen t directional bridge, Agilen t 85027/A/B
or C.
test p ort
of the
4-2 P erformance T ests
Figure 4-1. Return Loss Setup
T able 4-1. Return Loss Equipment T able
Equipmen t Common to all Agilen t Detectors
Scalar Net w ork Analyzer 8757D
8360 Series Syn thesized Sw eep er 83620A/30A
Additional Equipmen t Required for 85037A
Directional Bridge 85027C
Shielded Op en P art Num b e r 85032-60001
Short Agilen t 11512A
Adapter T yp e-N (m) to 3.5 mm (f ) P art Num b e r 1250-1744
Additional Equipmen t Required for 85037A Option 001
Directional Bridge 85027A
Calibrated Op en/Short P art Num b e r 85021-60001
Adapter T yp e-N (m) to 3.5 mm (f ) P art Num b e r 1250-1744
Additional Equipmen t Required for 85037B
Directional Bridge 85027B
Calibrated Op en/Short P art Num b e r 85037-60001
Procedure
1. Connect the equipmen t a s sho wn in Figure 4-1 , with nothing
connected to the bridge
test p ort
.
On the 8757D:
Press
4
PRESET
Press Channel
5
.
4
5
(Pressing t wice turns c hannel 2 o ) .
2 54 2
Allo w 3 0 min utes for w arm up.
2. Reset the stop frequency on the source, and if necessary and set
the p o w er:
F or the 85037A:
On the 83620A source:
Press
Press
4
STOP 54 18 54 GHz
4
PO WER LEVEL 54 0
5
.
3 54 dBm
5
.
F or the 85037B:
It is
not
necessary to reset the stop frequency on the Agilen
t
83630A source as it stops automatically at 26.5 GHz.
3. Calibrate the test setup.
Press F unction
4
CAL
5
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Select
SHORT/OPEN
.F ollo w the directions (prompts) app earing
on the CR T.
Connect the short to the
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STORE SHORT
.
Remo v e the short.
test p ort
of the bridge.
P erformance T ests 4-3
Connect the op en to the
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STORE OPEN
test p ort
.
Remo v e the op en.
The CR T will displa y:
SHORT/OPEN CAL SAVED IN CH1 MEM
of the bridge.
.
Note
Press F unction
4
DISPLA Y
5
.
NNNNNNNNNNNNNNNNNNNNNNNNNN
Select
MEAS-MEM
.
4. On the 8757D:
Connect the detector under test to the
Press F unction
Press
4
CURSOR
4
SCALE 54 5 54 dB
5
. Use the cursor to nd the highest trace v alue
5
.
test p ort
of the bridge.
(the w orst case measuremen t), in eac h s p ecication range.
W rite eac hv alue in the test record.
If more frequency resolution is needed at the lo
w end of the
frequency range, rep eat the preceeding pro cedure using a stop
frequency of 50 MHz.
This completes the pro cedure for measuring return loss.
4-4 P erformance T ests
Frequency
Response at a
Nominal 0 10 dBm
Description
The frequency resp onse of the 85037A/B detector is sp ecied as
the maxim um p eak-to-p eak deviation from a constan t input signal
of 0 10 dBm, as measured o v er the sp ecied frequency range.
T o simplify the measuremen t pro cedure, frequency resp onse is
measured with a nominal 0 10 dBm signal applied. First, the source
is c haracterized for frequency resp onse using a calibrated p o w er
meter/sensor com bination. Second, the DUT is c haracterized.
Finally , a p oin t-b y-p oin t dierence is computed, plotted, and
compared to the sp ecication windo w. The man ual test describ ed in
this pro cedure has an appro ximate ro ot sum of the squares (RSS)
uncertain t y ranging up to 6 0.25 dB. This implies that a \go o d "
detector, w ell within the limits of its sp ecications,
out of sp ecication. This measuremen t i s only an indication of the
detector's resp onse within these limits. If greater measuremen
accuracy is desired, a test system that minimizes the sources of
measuremen t uncertain t y will b e required.
c ould
measure
t
Figure 4-2. Frequency Response T est Setup
P erformance T ests 4-5
T able 4-2. Frequency Response Equipment T able
Equipmen t Common to All Agilen t Detectors
Scalar Net w ork Analyzer 8757D
8360 Series Syn thesized Sw eep er 83620A/30A
P o w er Meter 436A/437B/438A
Additional Equipmen t Required for Agilen t 85037A
P o w er Splitter 11667A
P o w er Meter 436A/437B/438A
Calibrated Sensor 8481A
10 dB A tten uator 8491B Option 010
Crystal Detector 8474B
Additional Equipmen t Required for Agilen t 85037A Option 001
P o w er Splitter 11667A Opt 002
P o w er Meter 436A/437B/438A
Calibrated Sensor 8481A Option 001
10 dB A tten uator 8492A Option 010
Crystal Detector 8474A
Additional Equipmen t Required for Agilen t 85037B
P o w er Splitter 11667B
P o w er Meter 436A/437B/438A
Calibrated Sensor 8485A
10 dB A tten uator 8493C Option 010
Crystal Detector 8473D
Procedure
Configuring the System
1. Connect the equipmen t a s sho wn in Figure 4-2 , with nothing
connected to the op en end of the p o
w er splitter.
T urn on all equipmen t.
Allo w 3 0 min utes for w arm up.
2. On the p o w er meter:
Press
4
5
mo de.
dBm
Zero and calibrate the p o w er meter. If y ou are unsure of ho w
to do this, refer to the p o w er meter op erating and service
man ual.
4
Range Hold
Set the
indicated for 50 MHz on the p o w er sensor
5
and
4
P o w er Ref
Cal F actor %
5
should remain o.
on the p o w er meter to the v alue
Cal F actor Chart
3. On the 8757D:
Preset the analyzer, turn c hannel 2 o, and select the detector
mo de for DC:
Press
4
PRESET
5
.
.
4-6 P erformance T ests
Press Channel
4
2 54 2
5
(Pressing t wice turns Channel 2 o ) .
Press Instrumen t State
4
SYSTEM
5
.
NNNNNNNNNNNNNNNNNNNNNNN
Select
MODE DC
.
4. On the source:
Press
4 CW 54 50 54
MHz
5
.
Connect the p o w er meter/sensor to the output of the splitter.
Press
Press
NNNNNNNNNNNNNNNNNNNNNNN
4
5
Ext Det
ALC
4
PO WER LEVEL 54 0
NNNNNNNNNNNNNN
MORE
10 54 dBm
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Coupling Factor
5
and, if necessary , adjust the
4 10 54 dB 5
po w er lev el for a p o w er meter reading within 6 0.05 dB of
0
10 dBm.
.
Measuring the Source
Measuring the Detector
Do not
readjust the p o w er lev el for the remainder of this test.
5. On the source:
Press
4 CW 5
and en ter the test frequencies as sho wn on the w ork
sheet.
F or example:
4
CW 54 .01 54 GHz
5
.
6. Using the Cal F actor Chart on the P o w er Sensor:
a. Set the
Cal F actor %
on the p o w er meter to the v alue
indicated for the test frequency as needed. (Use the nearest
frequency v alue.)
b. Note the reading on the p o w er meter.
c. Record this v alue on to the w orksheet T able 4-3 under the
\Source P o w er" column.
7. Rep eat steps 5 and 6 using, at minim
um, the test frequencies
noted on the w ork sheet.
8. Disconnect the p o w er meter/sensor.
9. With the detector still disconnected from the source, zero the
detector:
Press
4
5
on the 8757D.
CAL
Select
4
DC DET ZERO
5
.
NNNNNNNNNNNNNNNNNNNNNNNNNN
Select
AUTOZERO
.
When the zero is complete, the displa y will indicate:
AUTO ZERO COMPLETE
.
P erformance T ests 4-7
10. Connect the detector to the p o w er splitter.
On the 8757D:
Computing the
Maximum Error
Press F unction
11. On the source:
Press
4 CW 5
only the test frequencies used in steps 5 through 7.
12. Note and record on the w orksheet (under the \Measured P o w er"
column) the v alue indicated b y the 8757D cursor displa y .
13. Rep eat this step un til all of the same frequency p oin ts ha v e b een
measured.
14. Using the v alues recorded in steps 6 and 12, subtract the v alue
in step 6 from the v alue in step 12 for eac h o f the frequencies as
sho wn on the w orksheet.
No w use these v alues to plot a p oin
on the graph on the next page. The p eak to p eak v
determine the frequency resp onse of the detector. Record this
result on the test record.
This completes the pro cedure for measuring frequency
resp onse.
4
CURSOR
and en ter the rst test frequency . Remem b e r t o use
5
to turn the cursor on.
t t o p oin tv ariation curv e
ariations
4-8 P erformance T ests
Figure 4-3. Frequency Response Graph
T able 4-3. W orksheet
Recommended T est F requencies
F requency
in GHz
0. 0 1
0. 0 4
0. 1 0
2
6
10
14
16
18
20
22
24
26 .5
Measured P o w er
(dBm)
Source P o w er
(dBm)
Meas P o w er min us
Source P o w er (dB)
P erformance T ests 4-9
Dynamic P o w er
Accuracy
Description
Procedure
The dynamic p o w er accuracy of the 85037A/B is dep enden t u p o n the
ra w (uncorrected) resp onse of the detector, the correction routine
of the 8757D analyzer, and the accuracy of the 8757D Option 002
calibrator. The accuracy of the calibrator is assumed to b e within
sp ecications. This can b e c hec k ed b y follo wing the p erformance test
pro cedure in the 8757D man ual. The correction routine is xed and
cannot v ary , s o i t i s also assumed to b e accurate. The high accuracy
of the 85037A/B is only guaran teed after p erforming the built-in
c haracterization routine (a p o w er cal) in the 8757D. Once this is
done, p erforming an y dynamic accuracy tests using the same 8757D
calibrator is ineectual b ecause the only error that will b e seen will
b e due to measuremen t rep eatabilit y and an y p ossible temp erature
drift. Therefore, to v erify the dynamic accuracy of the 85037A/B,
only the ra w, uncorrected p erformance is measured. Once this has
b een v eried, the corrected dynamic accuracy ,b y design, m ust b e
within sp ecication. Indep enden tv erication of b oth corrected and
uncorrected sp ecications can b e p erformed b
t w o alternativ e setups sho wn in Figure 4-5 or Figure 4-6 . T
y using either of the
esting of
the ra w dynamic accuracy p erformance is the only test required. The
man ual test pro cedure follo ws.
1. Connect the equipmen t a s sho wn in Figure 4-4 . I f the sw
eep er is
on, turn it o.
2. Allo w equipmen tt o w arm up for 30 min utes.
3. Disconnect the detector.
4. Press
4
PRESET
5
.
5. Reconnect the detector to input A, this forces default correction.
6. Press Channel
7. Press
8.
9.
Select
Press
4 0 54
dBm
4
SYSTEM
NNNNNNNNNNNNNN
MORE
4
CAL
5
4 2 54 2 5
5
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SWEEP MODE
NNNNNNNNNNNNNN
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DET OFFSET
START MEAS
(Pressing t wice turns Channel 2 o ) .
NNNNNNNNNNNNNNNNN
CW ON
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEASURE DET OFS
NNNNNNNNNNNNNNNNN
DET A
. Read the oset v alue. Note this n um be r
as the reference v alue.
10. Press
4
CAL
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRIOR MENU
4
19 54 dBm
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
5
START MEAS
.
11. Read the oset v alue sho wn on the 8757D and subtract from
this the reference v alue in step 9. Record the results on the test
record card under Dynamic Accuracy A C M o de.
12. Rep eat steps 10 and 11 at the p o
w er lev els sho wn on the test
record.
4-10 P erformance T ests
First Alternate
Dynamic P o w er
Accuracy
13. V erify the noise o or in A C m o de.
Press:
4
A V G
NNNNNNNNNNNNNNNNNNNN
5
AVG ON
4
CURSOR
5
and record the a v erage noise o or as displa y ed b y the cursor
reading.
14.
Press
4
PRESET 54 SYSTEM
NNNNNNNNNNNNNNNNNNNNNNN
5
MODE DC
.
15. Rep eat steps 6 through 12 for DC mo de. The detector should b e
rezero ed b e t w een the 0 30 and 0 40 dBm steps.
Figure 4-4. Dynamic P o w er Accuracy T est Setup
Description
Figure 4-5. First Alternate Dynamic P o w er Accuracy T est Setup
This setup is for p erforming indep enden tv erication of dynamic
accuracy in 10 dB incremen ts up to 0 dBm in DC mo de.
P erformance T ests 4-11
Second Alternate
Dynamic P o w er
Accuracy
Description
Figure 4-6. Second Alternate Dynamic P o w er Accuracy T est Setup
Procedure
This setup is for p erforming indep enden
accuracy in 10 dB incremen ts up to +20 dBm in DC mo de.
T able 4-4. Additional Equipment
F unction Generator Agilen t 8116A
50 MHz Bandpass Filter 08757-80027
1. Set the 8757D for CW, DC mo de, cursor on.
2. Set the step atten uator to 20 dB.
3. Adjust the 8116 for a 5 0 MHz sinew
for a cursor reading of 0.0 dBm.
4. Adjust the step atten uator in 10 dB steps and compare the 8757D
cursor reading with the calculated applied p o w er determined
b y the calibrated step atten uator. The 20 dB step on the step
atten uator b ecomes the reference. The other steps are relativ et o
this 20 dB v alue.
5. The detector should b e rezero ed b e t
steps.
tv erication of dynamic
a v e and adjust the amplitude
w een the 0 30 and 0 40 dBm
4-12 P erformance T ests
T able 4-5. T est Record for 85037A
T est F acilit y Rep ort Num be r
Date
Customer
T ested b y
Mo del Am bien t temp erature
C
Serial Num be r Relativ eh umidit y %
Options
Sp ecial Notes
P erformance T ests 4-13
T able 4-5. T est Record for 85037A (2 of 3)
Mo del Rep ort Num be r Date
T est Equipmen t Used Mo del Num b e r T race Num b e r Cal Due Date
1.
2.
3.
4.
5.
6.
7.
8.
4-14 P erformance T ests
T able 4-5. T est Record for 85037A (3 of 3)
Serial Num b er: Rep ort Num b er: Date:
T est
Description Sp ecication
Return Loss at a Nominal 0 10 dBm
0.01 to 0.04 GHz 10 dB
0.04 to 18 GHz 20 dB
F requency Resp onse at a Nominal 0 10 dBm
0.01 to 0.04 GHz
0.04 to 18 GHz
6
6
Dynamic P o w er Accuracy
A C M o d e uncorrected
+19 dBm
+10 dBm
0
30 dBm
0
40 dBm
0
50 dBm
6
6
6
6
6
0.35 dB
0.18 dB
0.4 dB
0.4 dB
0.4 dB
0.8 dB
1.3 dB
Measured
Results
Measuremen t
Uncertain t y
6
1.0 dB
6
2.4 dB
6
0.18 dB
6
0.25 dB
6
0.2 dB
6
0.13 dB
6
0.22 dB
6
0.30 dB
6
0.36 dB
1
Noise Flo or
Dynamic P o w er Accuracy
DC Mo de uncorrected
+19 dBm
+10 dBm
0
30 dBm
0
40 dBm
0
50 dBm
1
Using the equipment and procedures do cumen
0
6
6
6
6
6
ted in this man ual.
55 dBm
0.4 dB
0.4 dB
0.4 dB
0.8 dB
1.3 dB
6
0.2 dB
6
0.13 dB
6
0.22 dB
6
0.3 dB
6
0.36 dB
P erformance T ests 4-15
4-16 P erformance T ests
T able 4-6. T est Record for 85037A Option 001
T est F acilit y Rep ort Num be r
Date
Customer
T ested b y
Mo del Am bien t temp erature
C
Serial Num be r Relativ eh umidit y %
Options
Sp ecial Notes
P erformance T ests 4-17
T able 4-6. T est Record for 85037A Option 001 (2 of 3)
Mo del Rep ort Num be r Date
T est Equipmen t Used Mo del Num b e r T race Num b e r Cal Due Date
1.
2.
3.
4.
5.
6.
7.
8.
4-18 P erformance T ests
T able 4-6. T est Record for 85037A Option 001 (3 of 3)
Serial Num b er: Rep ort Num b er: Date:
T est
Description Sp ecication
Return Loss at a Nominal 0 10 dBm
0.01 to 0.04 GHz 10 dB
0.04 to 18 GHz 20 dB
F requency Resp onse at a Nominal
0
10 dBm
0.01 to 0.04 GHz
0.04 to 18 GHz
6
6
Dynamic P o w er Accuracy
A C M o d e uncorrected
+19 dBm
+10 dBm
0
30 dBm
0
40 dBm
6
6
6
6
0.35 dB
0.18 dB
0.4 dB
0.4 dB
0.4 dB
0.8 dB
Measured
Results
Measuremen t
Uncertain t y
6
.7 dB
6
1.4 dB
6
0.18 dB
6
0.25 dB
6
0.2 dB
6
0.13 dB
6
0.22 dB
6
0.3 dB
1
0
50 dBm
Noise Flo or
Dynamic P o w er Accuracy
DC Mo de uncorrected
+19 dBm
+10 dBm
0
30 dBm
0
40 dBm
0
50 dBm
1
Using the equipment and procedures do cumen ted in this man ual.
6
0
6
6
6
6
6
1.3 dB
55 dBm
0.4 dB
0.4 dB
0.4 dB
0.8 dB
1.3 dB
6
0.36 dB
6
0.2 dB
6
0.13 dB
6
0.22 dB
6
0.3 dB
6
0.36 dB
P erformance T ests 4-19
4-20 P erformance T ests
T able 4-7. T est Record for 85037B
T est F acilit y Rep ort Num be r
Date
Customer
T ested b y
Mo del Am bien t temp erature
C
Serial Num be r Relativ eh umidit y %
Options
Sp ecial Notes
P erformance T ests 4-21
T able 4-7. T est Record for 85037B (2 of 3)
Mo del Rep ort Num be r Date
T est Equipmen t Used Mo del Num b e r T race Num b e r Cal Due Date
1.
2.
3.
4.
5.
6.
7.
8.
4-22 P erformance T ests
T able 4-7. T est Record for 85037B (3 of 3)
Serial Num b er: Rep ort Num b er: Date:
T est
Description Sp ecication
Return Loss at a Nominal 0 10 dBm
0.01 to 0.04 GHz 10 dB
0.04 to 18 GHz 20 dB
18 GHz to 26.5 GHz 18 dB
F requency Resp onse at a Nominal 0 10 dBm
0.01 to 0.04 GHz
0.04 to 18 GHz
18 GHz to 26.5 GHz
6
6
6
Dynamic P o w er Accuracy
A C M o d e uncorrected
+19 dBm
+10 dBm
0
30 dBm
6
6
6
0.35 dB
0.18 dB
0.22 dB
0.4 dB
0.4 dB
0.4 dB
Measured
Results
Measuremen t
Uncertain t y
6
0.7 dB
6
1.4 dB
6
2.0 dB
6
0.18 dB
6
0.25 dB
6
0.3 dB
6
0.2 dB
6
0.13 dB
6
0.22 dB
1
0
40 dBm
0
50 dBm
Noise Flo or
6
6
0
Dynamic P o w er Accuracy
DC Mo de uncorrected
+19 dBm
+10 dBm
6
6
0 dBm + 0.4 dB
0
30 dBm
0
40 dBm
0
50 dBm
1
Using the equipment and procedures do cumen ted in this man ual.
6
6
6
0.8 dB
1.3 dB
55 dBm
0.4 dB
0.4 dB
0.4 dB
0.8 dB
1.3 dB
6
0.3 dB
6
0.36 dB
6
0.2 dB
6
0.13 dB
6
0.22 dB
6
0.22 dB
6
0.3 dB
6
0.36 dB
P erformance T ests 4-23
Adjustments
5
Introduction
Tw o adjustmen ts can b e p erformed on the 85037A/B.
1. coarse zero adjustmen t
2. feedthrough n ulling adjustmen t
Normally these t w o adjustmen ts do not need to b e p erformed. They
should only b e p erformed if the zeroing routine fails.
Before attempting an y adjustmen ts:
T urn on all equipmen t and allo w a minim um of 30 min utes w arm up
time.
Note that the adjustmen ts are in teractiv e and
in the order giv en.
must
b e p erformed
Adjustments 5-1
Coarse Zero
Adjustment
Description
Procedure
The coarse zero adjustmen t cen ters the detector's DC preamp input
oset v oltage within the built-in automatic zero routine's range.
This is accomplished b y adjusting for the lo w est o v erall indication of
noise on the 8757D. When the lev el is minimized, it indicates that
the range has b een prop erly cen tered. Remo v e the detector's outer
co v ers and metal housing to allo w access for making the adjustmen ts.
(Refer to \Remo ving the Co v ers" at the b eginning of Chapter 7,
\Service.")
Figure 5-1. Coarse Zero/Feedthrough Nulling Adjustment
Equipmen t: 8757D scalar net w ork analyzer
1. Connect the equipmen t a s sho wn in Figure 5-1 with no input to
the detector.
2. Press
3.
4
PRESET
Press
4
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
5
CHANV LOGGER
.
NNNNNNNNNNNNNN
5
MORE
.
NNNNNNNNNNNNNNNNNNNNNNN
SERVICE
NNNNNNNNNNNNNNNNNNNN
A4 ADC
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MORE
CHANNEL VOLTS
4. Lo cate the single-turn p oten tiometer on b ottom side of the b oard.
While viewing the
DATA READING
for
INPUT A
, adjust the coarse
zero p oten tiometer (the single turn p ot) for the most negativ e
reading (t ypically ab out 0 5 V). See Figure 5-2 .
5-2 Adjustments
Feedthrough Nulling
Adjustment
Figure 5-2. Coarse Zero Adjustment
Description
Procedure
The feedthrough n ulling is adjusted for b est DC accuracy b elo
0
45 dBm. This is accomplished b y adjusting for the
o v erall indication of noise on the 8757D. When the lev
highest
el of noise is
w
maximized, it indicates that the feedthrough is n ulled.
Equipmen t: 8757D scalar net w ork analyzer
P erform the follo wing steps in order.
Connect the equipmen t a s sho wn in Figure 5-1 .
On the 8757D:
1. Press
2. Press Channel
3.
Press
4
PRESET
4
SYSTEM
5
.
4 2 54 2 5
NNNNNNNNNNNNNN
5
(Pressing t wice turns Channel 2 o ) .
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SWEEP MODE
NNNNNNNNNNNNNNNNN
CW ON
to place the 8757D in
CW mo de.
NNNNNNNNNNNNNNNNNNNNNNN
4. Press
5.
6.
Press
Press
4
SYSTEM
5
MODE DC
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
4
5
SMOOTH ON
SPCL
4
SCALE 54 5 54 dB
.
.
5
.
7. If necessary , use the arro ws k eys to mo v eu p o r d o wn to set the
reference line to the cen ter of the screen. (This will b e 4 divisions
up from the b ottom of the CR T screen.)
8.
9.
Press
Press
4
5
REF
4
CURSOR
NNNNNNNNNNNNNNNNNNNNNNNNNN
REF POSN
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
5
CRSR 0 >REF LVL
.
Adjustments 5-3
10.
Press
4
CAL
indicates
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
5
DC DET ZERO
:REMOVE RF FROM DC DETECTORS.
NNNNNNNNNNNNNNNNNNNN
MANUAL
. The displa y
If y ou ha v e connected
the detector to an RF output, disconnect it.
Press
NNNNNNNNNNNNNN
CONT
.
11.
When the routine is complete the displa y indicates: MANUAL
ZER O COMPLETED.
12. Lo cate the F eedthrough Null m ulti-turn p oten tiometer whic h
is on the top side of the PC b oard, (the same side as the cable
wires). Adjust it one half turn (either direction). See Figure 5-3 .
13. Rep eat steps 9 through 12 un til turning the F eedthrough Null
p oten tiometer in either direction will only
de cr e ase
the displa y ed
noise. If the o v erall noise lev el is less (lo w er on the screen) than
the b eginning v alue noted, rev erse the direction of adjustmen t.
Con tin ue adjusting un til the noise lev el displa y ed is
maximize d
.
5-4 Adjustments
Figure 5-3. Feedthrough Nulling Adjustment
Replaceable P arts
6
Introduction
Ordering P arts
This section pro vides replaceable parts and ordering information.
T o order a part listed in the replaceable parts T able 6-2 , quote the
Agilen t part n um b er, indicate the quan tit y required and address the
order to y our nearest Agilen t oce.
Replaceable P arts 6-1
T o request information on a part that is not listed in the replaceable
parts table, include the instrumen t m o del n um b e r and a description
of the part and its function. Address the inquiry to the nearest
Agilen t oce.
Figure 6-1. Detectors and Cable Mark er Kit
T able 6-1. Miscellaneous P arts
Item
# Description and Quan tit y
1 Plastic half-b o d y c o v er (2) 85025-40006
2 Lab el: w arning max input (1) 85037-80003
Lab el: electrostatic sensitiv e (1) 85037-80002
3 Cable assem bly (1) 85025-60003
4 Screw M2.5 x 0.45; 4 mm-LG (2) 0515-0972
5 Lo c kw asher 2.5 mm (2) 2190-0583
6 Cable mark er kit (1) 5061-1044
Agilen t
P art
Num b e r
6-2 Replaceable P arts
T able 6-2. Replaceable P arts and Accessories
P art Description Agilen t M o del or
P art Num b e r
Rebuilt/Exc hange Assem blies
85037A T yp e-N 85037-69006
85037A Option 001 7m m 85037-69007
85037B 3.5 mm 85037-69008
Accessories
Connector Gage Kits T yp e-N 85054-80011
7m m 85050-80012
3.5 mm 11752-60106
Collet Extractor T oo l F or 85037A Option 001 5060-0370
Adapters T yp e-N (f ) t o BNC (m) 1250-1534
T yp e-N (m) to T yp e-N (m) 1250-1475
3.5 mm (f ) t o 3.5 mm (f ) 1250-1749
T yp e-N (m) to APC-7 11525A
(Option 001 only)
APC-3.5 (f ) t o N (m) 1250-1744
Extension Cables 7.6 meter (25 fo ot) 11679A
61 meter (200 fo ot) 11679B
Connector Care Man ual 08510-90064
Replaceable P arts 6-3
6-4 Replaceable P arts
Figure 6-2. Module Exchange Program
By in ternet, phone, or fax, get assistance with all y our test &
measuremen t needs.
T able 6-3. Contacting Agilent
Online Assistance:
www.agilent.com/find/assist
United States
(tel) 1 800 452 4844
Canada
(tel) 1 877 894 4414
(fax) (905) 206 4120
Europ e
(tel) (31 20) 547 2323
(fax) (31 20) 547 2390
Japan
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
Latin America
(tel) (305) 269 7500
(fax) (305) 269 7599
Australia
(tel) 1 800 629 485
(fax) (61 3) 9210 5947
New Zealand
(tel) 0 800 738 378
(fax) 64 4 495 8950
Asia P acic
(tel) (852) 3197 7777
(fax) (852) 2506 9284
Replaceable P arts 6-5
Service
7
Caution
Error Messages
Repair
This pro duct is susceptible to damage from electrostatic disc harge
(ESD). When y ou p erform an y o f the follo wing pro cedures, w ear a
grounded static-strap and w ork at a static-safe w orkstation.
The message
EEPROM read failed -- A (B,C or R)
Indicates that
the EEPR OM calibration constan ts are either corrupted or unable
to b e read from the detector. The problem could b e in either the
detector or the 8757D. T ry another detector to isolate the cause.
If the detector app ears to b e the cause of the problem, then the
detector needs to ha v e the calibration constan ts regenerated at the
factory . Refer to \Returning a Detector for Service" in Chapter 2,
\Installation."
These detectors ha v e only one repairable item:
The cable assem bly .
If the detector fails electrically , order an exc hange detector:
Remember
1. See the Replaceable parts, T able 6-2 , for the part n um b e r that
matc hes the mo del that y ou are replacing:
85037A
85037A Option 001 or
85037B.
2. F ollo w \The Mo dule Exc hange Program" instructions in
Figure 6-2 .
T o receiv e exc hange credit, y ou
must
return the failed assem bly to
Agilen t i n the exc hange assem bly b o x.
Service 7-1
Repair
Replacing the Detector
Remo ving the Co v ers
1. Remo v e the plastic co v ers from the existing detector. (See
\Remo ving the Co v ers.")
2. Install the co v ers on the replacemen t detector.
3. P erform an op erator's c hec k. (See \Op erator's Chec k" in
Chapter 3.)
4. If the replacemen t i s a restored exc hange detector, return the
defectiv e detector using the pac king material supplied.
1. Place the detector with its narro w side on a at surface, with the
RF connector facing a w a y from y ou. (See Figure 7-1 .)
2. Holding the sides of the detector near the cable end, insert a small
at-blade screwdriv er with a blade no greater than 3.5 mm (1/8
in) b e t w een the side lab el and the raised edge. (See Figure 7-1 .)
Mak e sure y ou insert the screwdriv er as far forw ard on the
detector as p ossible.
3. Rotate the screwdriv er appro ximately 90
,u n til the co v er snaps
apart.
4. Rep eat steps 2 and 3, inserting the screwdriv
of the w a yt o w ard the cable end of the detector. (See the p oin
er appro ximately 2/3
t
sho wn in Figure 7-1 .)
5. Separate the plastic shell halv es. If the co v er do es not separate
easily , rep eat steps 2 through 4 o n the other side of the detector.
(See Figure 7-1 .)
6. T o attac h the co v ers to a replacemen t detector, snap the halv es
together.
Figure 7-1. Remo ving the Detector Co v ers
7-2 Service
Repair
Replacing the Cable
Assembly (W1)
Caution
1. Remo v e the plastic co v ers from the detector. (See \Remo ving the
Co v ers.")
2. Remo v e the t w o screws at the cable end of the metal housing.
3. Slide the metal housing a w a y from the RF connector, completely
exp osing the prin ted circuit (PC) assem bly .
4. Unsolder all cable wires from the PC assem bly .
5. With the PC assem bly facing up, carefully secure the detector
frame in a vice. Do not grip the PC assem bly .
Do not o v er-tigh ten the vice; y ou can b end the frame.
6. Using a 7/16 inc h o p en-ended wrenc h, unscrew the cable hex n ut.
7. Remo v e the old cable assem bly .
8. Remo v e the metal housing from the cable.
9. Carefully slide the metal housing on to the new cable
assem bly , making sure that y ou do not co v er the adjustmen t
p oten tiometers.
10. Solder the new cable wires to the PC assem
7-2.)
bly . (Refer to Figure
11. Ensure that all cable wires are securely connected to the
assem bly .
12. Slide the metal housing o v er the PC assem bly .
13. Using the screws remo v ed in step 2, secure the metal housing to
the frame.
14. Snap the co v ers bac k on.
Service 7-3
Repair
Figure 7-2. Cable Connections
7-4 Service
Caring for Connectors
This app endix pro vides a brief in tro duction to the fundamen tals of
prop er connector care, as imp ortan t t o g o o d measuremen ts as prop er
instrumen t calibration and adjustmen t. This app endix is in tended to
pro vide basic information and tell y ou where to nd more: Agilen t's
Micr owave Conne ctor Car e
A
. (See T able 6-2 for ordering information.)
Remember
Visual Inspection
Cleaning
A damaged connector can destro ya n y connector attac hed to it.
Basic connector care comprises three parts:
1. visual insp ection
2. cleaning
3. mec hanical insp ection
Visually insp ect all system connectors often. Examine connectors
for problems suc h a s con tamination or corrosion, esp ecially on the
con tacting surfaces. Lo ok for deformed threads, burrs, scratc
hes,
rounded shoulders, and similar signs of w ear or damage. An y visible
problem can degrade p erformance; clean, reinsp ect, and, if necessary
replace the connector.
T ry compressed air rst.
Use compressed air to lo osen particles on the connector mating
plane surfaces. Clean air cannot damage a connector, or lea v e
particles or residues b ehind.
,
If a solv en t i s necessary , use
only
isoprop yl alcohol.
If there is dirt or con tamination on a connector that y ou cannot
remo v ed with compressed air, try a foam sw ab or lin t-free cleaning
cloth moistened with isoprop yl alcohol.
Use the least amoun t o f alcohol p ossible.
Av oid getting an y plastic parts in the connectors w et with the
alcohol.
Nev er spra y alcohol directly in to a connector.
Chec k the alcohol p erio dically for con tamination.
Caring for Connectors A-1
Mechanical Inspection
Because coaxial connector mec hanical tolerances can b e v ery precise
(on the order of a few h undredths of microinc hes), ev en a p erfectly
clean, un used connector can cause trouble if it is mec hanically out of
sp ecication. Use a connector gage to mec hanically insp ect coaxial
connectors. Gage a connector at the follo wing times:
Before y ou use it for the rst time.
If either visual insp ection or electrical p erformance indicates that
the connector in terface ma y b e out of sp ecication (due to w ear or
damage, for example).
The device is used on another system or piece of equipmen t.
As a matter of routine: initially after ev ery 100 connections, and
after that as often as exp erience suggests.
A-2 Caring for Connectors
Index
A C detection, 1-2 , 3-1
A
adjustmen ts, 5-1
coarse zero adjustmen t, 5-1 {2
feedthrough n ulling adjustmen t, 5-1 {2
autozero, 4-7
bridge
B
directivit y, 4-2
source matc h, 4-2
cable
C
lead iden tication, 2-3
replacing, 7-3
cable assem bly, 7-3
c haracterizing the detector, 3-7
c haracterizing the source, 4-7
cleaning connectors, A-1
coarse zero adjustmen t, 5-1
computing the maxim um error, 4-8
w orksheet, 4-9
connector
care, A-1
cleaning, A-1
mating, 2-3
mating a 3.5 mm to an SMA connector, 2-4
mec hanical insp ection, A-2
torque v alues, 3-2
visual insp ection, A-1
con tacting Agilen t, 6-5
co v er
attac hing, 7-2
remo ving, 7-2
DC detection, 3-1
D
detector
c haracterization, 3-7
connecting, 2-3
po w er calibration, 3-7
replacing, 7-2
zeroing, 3-5
detector rebuild/exc hange, 7-1
dynamic accuracy, 3-3
dynamic accuracy of the 8757D, 4-2
dynamic p o w er accuracy, 4-10
Index-1
electrical p erformance, 4-1
E
electrostatic disc harge (ESD), 2-2
cautions, 3-1
equipmen t, 4-6
exc hange credit, 7-1
feedthrough n ulling adjustmen t, 5-1
F
o or mat
use, 2-2
frequency resp onse, 4-5
correction constan ts, 3-3
correction factors, 3-3
gaging, 2-4
G
heel strap
H
use, 2-2
initial insp ection, 2-1
I
installation, 2-1
mat
M
use, 2-2
measuremen t errors
sources, 4-2
measuremen ts
A C detection, 3-6
DC detection, 3-5
measuremen t system conguration, 3-4
measuring the detector, 4-7
measuring the source, 4-7
mixer or m ultiplier, when using
man ually en tering frequencies, 3-3
mo dule exc hange program, 6-4
Index-2
op erating en vironmen t, 2-6
O
op eration
theory, 3-1
ordering parts, 6-1
pac k aging, 2-6
P
p erformance tests, 4-1
dynamic accuracy, 4-1 {10
frequency resp onse, 4-1 , 4-5
return loss, 4-3
po w er
requiremen ts, 2-3
reection measuremen ts, 1-2
R
remo ving co v ers, 7-2
repairable parts, 7-1
replaceable parts, 6-1
replacing
cable, 7-3
detector, 7-2
return loss, 4-3
test description, 4-2
safet y considerations, 2-1
S
service, 7-1
returning detector for service, 2-7
setup
A C detection mo de, 3-4 {6
DC detection mo de, 3-4 {5
static-safe
practices, 2-2
w orkstation, 2-2
storage and shipmen t, 2-6
en vironmen t, 2-6
table mat
T
use, 2-2
test records, 4-13
transmission measuremen ts, 1-2
W
Z
W1 cable
replacemen t, 7-3
wrist strap
use, 2-2
zeroing, 3-6
autozero, 3-6
man ual zero, 3-6
rep eat zero, 3-6
Index-3