This manual applies directly to Agilent 85036B calibration kits with serial
number prefix 3514A, and Agilent 85036E calibration kits with serial number
prefix 3142A. The calibration devices in this kit are individually serialized.
Record the device serial numbers in the table provided in this manual. (See
“Recording the Device Serial Numbers” in Chapter 1.)
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS
SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS.
FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW,
AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN,
INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT
SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE
OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD
AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH
WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Assistan ce
Product maintenance agreements and other customer assistance agreements are availa ble
for Agilent products.
For any assistance, contact Agilent Technologies. Refer to page 5-3.
The Agilent 85036B and 85036E type-N calibration kits are used to calibrate Agilent
network analyzers up to 3 GHz for measurements of components with 75Ω type-N
connectors.
Kit Contents
The 85036B calibration kit contains the following:
• one male and one female 75Ω type-N open termination
• one male and one female 75Ω type-N short termination
• one male and one female 75Ω t ype-N load
• three 75Ω type-N to 75Ω type-N adapters
Refer to T able 6-1 and Figure 6-1 for a complete list of kit contents and their associated
part n u mbers .
The 85036E calibration kit contains t he following:
• one male combination open/short termination
• one male 75Ω type-N load
Refer to T able 6-2 and Figure 6-2 for a complete list of kit contents and their associated
part n u mbers .
Broadband Loads
The broadband loads a re inst rument-grade, 75Ω terminat ions that have been optimized for
performance up to 3 GHz. The rugged internal structure provides for highly repeatable
connections. A distributed resistive element on sapphire provides excellent stability and
return loss.
Opens and Shorts
The opens and shorts are built from parts that are machined to the current state-of the-art
precision machining.
The short’s inner conductors have a one-piece construction, common with the shorting
plane. This construction provides for extremely repeatabl e connection s .
The female open has a separate-piece inner conductor that is made from a
low-dielectric-constant plastic to minimize compensation values.
Both the opens and shorts are constructed so that the pin depth can be controlled very
tightly, thereby minimizing phase errors. Some of the opens and shorts have offsets. The
lengths of these offsets are designed so that the difference in phase of their reflection
coefficients is approximately 180 degrees at all frequencies.
1-285036B/E
General Information
Calibration Kit Overview
Adapters
Like the other devices in the kit, the adapters are built to very tight tolerances to provide
good broadband performance. The adapters utilize a dual-beaded connector structure to
ensure stable, repea table connections. The b eads are des i gned to minimize return loss and
are separated far enough so that interaction between the beads is minimized.
The adapters are designed so that their nominal electrical lengths are the same, which
allows them to be used in calibration procedures for non-insertable devices.
Calibration Definitions
The calibration kit must be selected and the calibration definitions for the devices in the
kit installed in the network analyzer prior to performing a calibration. Refer to your
network analyzer user’s guide for instructions on selecting the calibration kit and
performing a calibration.
The calibration definitions can be:
• resident within the analyzer
• entered from the front panel
Installation of the Calibration Definitions
The calibration definitions for the kit may be permanently installed in the internal
memory or hard disk of the network analyzer.
If the calibration definitions for the kit are not permanently installed in the network
analyzer, they must be manually entered. Refer to your network analyzer user’s guide for
instructions.
Equipment Required but Not Supplied
Gages, torque and open-end wrenches, ESD protective devices, and various connector
cleaning supplies are not included in the calibration kit but are required to ensure
successful operation of the calibration kit. Refer to Table 6-3 on page 6-5 for ordering
information
85036B/E1-3
General Information
Incoming Inspection
Incoming Inspection
Verify that the shipment is complete by referring to Figure 1-1 o n p age 1-5.
Check for damage. The foam-lined storage case provides protection during shipping.
If the case or any device appears damaged, or if the shipment is incomplete, contact
Agilent. See Table 5-1 on page 5-3. Agilent will arrange for repair or replacement of
incomplete or damaged shipments without waiting for a settlement from the
transportation company.
When you send the kit or device to Agilent, in clude a service tag (found near the end of this
manual) with the following information:
• your company name and address
• the name of a technical contact person within your company, and the person's complete
phone number
• the model number and serial number of the kit
• the part number and serial number of the device
• the type of service required
•a detailed description of the problem
1-485036B/E
Figure 1-1 Calibration Kit Contents
General Information
Incoming Inspection
85036B/E1-5
General Information
Recording the Device Serial Numbers
Recording the Device Serial Numbers
In addition to the kit serial number, the devices in this kit are individually serialized
(serial numbers are labeled onto the body of each device). Record these serial numbers in
Table 1-1 for the 850 36B and Table 1-2 for the 85036E. Recording the serial numbers will
prevent confusing the devices in this kit with similar devices in other kits.
Table 1-1Serial Number Reco rd for 85036B
DeviceSerial Number
Calibration kit
Male broadband load
Female broadband load
Male open
Female open body
Male short
Female short
Type-N male to male adapter
Type-N- female to female adapter
Type-N- male to female adapter
This manual applies to any 85036B or 85036E calibration kit whose serial prefix is listed
on the title page. If your calibration kit has a different serial number prefix than the one
listed on the title page, refer to “Calibration Kit History” below for information on how this
manual applies.
Calibration Kit History
This section describes calibration kits with serial number prefixes different than those
listed on the title page.
Agilent 8 5036B kits with serial prefix 3102A and below
• 75 ohm load specification: return loss ≥ 45 db, dc to 2 GHz
• all other kit components unspecified
85036B calibration kits with serial number prefix 3102A and below should meet the
specifications published in this manual. If you wish to have your kit recertified to current
specifications, contact your nearest Agilent service center. For contact information, see
Table 5-1 on page 5-3.
Agilent 8 5036B kits with serial prefix 3103A and below
These calibration kits had a center conductor extender with a smaller body. Therefore, a
different procedure was used for connecting the female open. If you’re using one of the
smaller extenders, replace steps 1 through 3 in “Connecting and Disconnecting the
Two-Piece Female Open (85036B) ” on page 3-9 of this manual with the steps below.
1. Mate the center conductor extender to the male pin of the test port.
2. Connect the open body to the test port by carefully guiding it over the extender.
3. To disconnect the open, carefully remove the open body first.
The part numbers in this manual are the recommended replacement parts for these kits
(including the center conductor extender).
85036B/E1-7
General Information
Clarifying the Sex of a Connector
Clarifying the Sex of a Connector
In this manual, the s ex of c a li bra tion de vices a nd ada pt ers ar e r ef erred to in terms of their
connector interface. For example, a male open has a male connector.
However, during a measurement calibration, the network analyzer softkey menus label a
type-N calibration device with reference to the sex of the analyzer’s test port
connector—not the calibration device connector. For example, the label SHORT(F) on the
analyzer’s display refers to the short that is to be connected to the female test port. This
will be a male short from the calibration kit.
Conversely, connector gages are referred to in terms of the connector that it measures. For
instance, a male connector ga ge has a female connector on the gage so that it can measure
male dev ices.
Preventive Maintenance
The best techniques for maintaining the integrity of the devices in this kit include:
• routine visual inspection
• cleaning
• proper gaging
• proper connection techniques
All of the above are described in Chapter 3 , “Use, Maintenance, and Care of the Devices.”
Failure to detect and remove dirt or metallic particles on a mating plane surface can
degrade repeatability and accuracy and can damage any connector mated to it. Improper
connections, resulting from pin depth values being out of the limits (see Table 2-3 on page
2-3), or from bad connections, can also damage these devices.
1-885036B/E
2Specifications
2-1
Specifications
Environmental Requirements
Environmental Requirements
Table 2-1Environmental Requirements
Parameter Limits
Operating temperature
Error-corrected temperature range
Storage temperature−40 °C to +75 °C (−40 °F to +167 °F)
Altitude
Operation0 to 80% (26 °C maximum dry bu lb)
Storage0 to 95%
a. The temperatu re rang e over which the calibration standards maintain confor ma nce to their
specification s .
b. The allowable network analy z er ambient temp erature drift during measurement calibration
and during measurem ents when t he ne twor k analyzer er ror co rre cti on is t urne d on. Al so, the
range over which the network an aly z er m aintains its spe cified performance while correction
is turned on.
a
b
+18 °C to +28 °C (+64 °F to +82 °F)±1 °C of measurement calibration temperature
Temperature—What to W atch Out For
Changes in temperature can affect electrical characteristics. Therefore, the operating
temperature is a critical factor in performance. During a measurement calibration, the
temperature of the calibration devices must be stable and within the range specified in
Table 2-1.
IMPORTANTAvoid unnecessary handling of the devices during calibration because your
fingers are a heat source.
2-285036B/E
Specifications
Mechanical Specifications
Mechanical Specifications
The mechanical specifications in T able 2-2 apply to the devices in the 85036B and the
85036E 75 Ω type-N calibration kits.
Table 2-2 Mechanical Specifications
DeviceSpecification
Type-N Male OpenInside diameter of outer conductor: 6.9 8 5 to 7.010
Type-N Male ShortInside diameter of outer conductor: 7 ±0.015 mm
Distance from reference plane to shorting plane: 5.29 ±0.013 mm
Supplemental Mechanical Characteristics
Supplemental characteristics are values which are typically met by a majority of the
calibration kit devices tested at Agilent. These supplemental characteristics are intended
to provide information useful in calibration kit applications by giving typical, but
non-warranted performance parameters. Table 2-3 lists the typical characteristics of the
devices in this kit.
Table 2-3Mechanical Characteristics
DeviceCharacteristic
Type-N Male ShortDiameter of male pin: 0.897 ±0.013 mm
All Type-N Male ConnectorsPin depth: 0.207 to 0.210 inch
All Type-N Female ConnectorsPin depth: 0.204 to 0.207 inch
85036B/E2-3
Loading...
+ 34 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.