Agilent Technologies 664xA, 665xA, 667xA, 668xA, 669xA User Manual

OPERATING GUIDE
GPIB DC POWER SUPPLIES
Agilent Technologies Models
664xA, 665xA, 667xA, 668xA, and 669xA
sA
Agilent Part No. 5964-8267 Printed in Malaysia
Microfiche Part No. 5964-8268 January, 2005

Certification

Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members.
Warranty
This Agilent Technologies hardware product is warranted against defects in material and workmanship for a period of one year from date of delivery. Agilent Technologies software and firmware products, which are designated by Agilent Technologies for use with a hardware product and when properly installed on that hardware product, are warranted not to fail to execute their programming instructions due to defects in material and workmanship for a period of 90 days from date of delivery. During the warranty period Agilent Technologies will, at its option, either repair or replace products which prove to be defective. Agilent Technologies does not warrant that the operation of the software, firmware, or hardware shall be uninterrupted or error free.
For warranty service, with the exception of warranty options, this product must be returned to a service facility designated by Agilent Technologies. Customer shall prepay shipping charges by (and shall pay all duty and taxes) for products returned to Agilent Technologies for warranty service. Except for products returned to Customer from another country, Agilent Technologies shall pay for return of products to Customer.
Warranty services outside the country of initial purchase are included in Agilent Technologies' product price, only if Customer pays Agilent Technologies international prices (defined as destination local currency price, or U.S. or Geneva Export price).
If Agilent Technologies is unable, within a reasonable time to repair or replace any product to condition as warranted, the Customer shall be entitled to a refund of the purchase price upon return of the product to Agilent Technologies.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Customer, Customer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation and maintenance. NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. Agilent Technologies SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Exclusive Remedies
THE REMEDIES PROVIDED HEREIN ARE THE CUSTOMER'S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
Assistance
The above statements apply only to the standard product warranty. Warranty options, extended support contracts, product maintenance agreements and customer assistance agreements are also available. Contact your nearest Agilent Technologies Sales and Service offices for further information on Agilent Technologies' full line of Support Programs.
2

Safety Summary

The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer's failure to comply with these requirements.
GENERAL
Any LEDs used in this product are Class 1 LEDs as per IEC 825-l. This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 du Canada.
ENVIRONMENTAL CONDITIONS
With the exceptions noted, all instruments are intended for indoor use in an installation category II environment. They are designed to operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters. Refer to the specifications tables for the ac mains voltage requirements and ambient operating temperature range. Exceptions: Agilent Technologies Models 6680A, 6681A, 6682A, 6683A, 6684A, 6690A, 6691A and 6692A are intended for use in an installation category III
1
Category II - Local level for connection to household outlets for 120 V, 230 V, etc.
2
Category III - Distribution level or cases where the reliability and availability of the equipment are subject to special requirements.
2
environment.
BEFORE APPLYING POWER
Verify that the product is set to match the available line voltage and the correct fuse is installed.
GROUND THE INSTRUMENT.
This product is a Safety Class 1 instrument (provided with a protective earth terminal). To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The instrument must be connected to the ac mains through a three-conductor power cable, with the third wire firmly connected to an electrical ground (safety ground) at the power outlet. For instruments designed to be hard-wired to the ac power lines (supply mains), connect the protective earth terminal to a protective conductor before any other connection is made. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury. If the instrument is to be energized via an external autotransformer for voltage reduction, be certain that the autotransformer common terminal is connected to the neutral (earthed pole) of the ac power lines (supply mains).
FUSES
Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short circuited fuseholders. To do so could cause a shock or fire hazard.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes.
KEEP AWAY FROM LIVE CIRCUITS
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified service personnel. Do not replace components with power cable connected. Under certain conditions, dangerous voltages may exist even with the power cable removed. To avoid injuries, always disconnect power, discharge circuits and remove external voltage sources before touching components.
DO NOT SERVICE OR ADJUST ALONE
Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
DO NOT EXCEED INPUT RATINGS
This instrument may be equipped with a line filter to reduce electromagnetic interference and must be connected to a properly grounded receptacle to minimize electric shock hazard. Operation at line voltages or frequencies in excess of those stated on the data plate may cause leakage currents in excess of 5.0 mA peak.
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT
Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. Return the instrument to an Agilent Technologies Sales and Service Offices for service and repair to ensure that safety features are maintained.
Instruments which appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
1
, pollution degree 2
3
Safety Symbol Definitions
Symbol Description Symbol Description
Direct current
Alternating current
Both direct and alternating current
Three-phase alternating current
Earth (ground) terminal
Protective earth (ground) terminal
Frame or chassis terminal
Terminal for Neutral conductor on permanently installed equipment
Terminal is at earth potential(Used for measurement and control circuits designed to be operated with one terminal at earth potential.)
Terminal for Line conductor on permanently installed equipment Caution, risk of electric shock
Caution, hot surface
Caution (refer to accompanying documents)
In position of a bi -stable push control
Out position of a bi -stable push control
On (supply)
Off (supply)
Standby (supply). Units with this symbol are not completely disconnected from ac mains when this switch is off. To completely disconnect the unit from ac mains, either disconnect the power cord or have a qualified electrician install an external switch.
The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.
attention to an operating procedure, or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed beyond a CAUTION sign until the indicated conditions are fully understood and met.
The CAUTION sign denotes a hazard. It calls

Printing History

The edition and current revision of this manual are indicated below. Reprints of this manual containing minor corrections and updates may have the same printing date. Revised editions are identified by a new printing date. A revised edition incorporates all new or corrected material since the previous printing date. Changes to the manual occurring between revisions are covered by change sheets shipped with the m anual. In some cases, the manual change applies only to specific instruments. Instructions provided on the change sheet will indicate if a particular change applies only to certain instruments.
 Copyright 2001, 2002, 2005 Agilent Technologies Inc. Edition 1 - July, 2001 Update1 - March, 2002 Update2 - January, 2005
This document contains proprietary information protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated into another language without the prior consent of Agilent Technologies. Information contained in this document is subject to change without notice.
Herstellerbescheinigung
Diese Information steht im Zusammenhang mit den Anforderungen der Maschinenläminformationsverordnung vom 18 Januar 1991. * Schalldruckpegel Lp <70 dB(A) * Am Arbeitsplatz * Normaler Betrieb * Nach EN 27779 (Typprufung).
Manufacturer's Declaration
This statement is provided to comply with the requirements of the German Sound Emission Directive, from 18 January
1991. * Sound Pressure Lp <70 dB(A) *At Operator Position * Normal Operation * According to EN 27779 (Type Test).
4
DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturer’s Name and Address
Responsible Party Agilent Technologies, Inc. Agilent Technologies (Malaysia) Sdn. Bhd 550 Clark Drive, Suite 101 Budd Lake, New Jersey 07828 USA
Declares under sole responsibility that the product as originally delivered
EMC Information ISM Group 1 Class A Emissions
Safety Information and Conforms to the following safety standards.
This DoC applies to above-listed products placed on the EU market after:
January 1, 2004 Date Bill Darcy/ Regulations Manager
For further information, please contact your local Agilent Technologies sales office, agent or distributor, or
Agilent Technologies Deutschland GmbH, Herrenberger Straβe 130, D71034 Böblingen, Germany
Revision: B.00.00 Issue Date: Created on 11/24/2003 3:23
Alternate Manufacturing Site
Product Names
Model Numbers
Product Options
Complies with the essential requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CE Marking accordingly.
As detailed in Electromagnetic Compatibility (EMC), Certificate of Conformance Number
Assessed by: Celestica Ltd, Appointed Competent Body
Malaysia Manufacturing Bayan Lepas Free Industrial Zone, PH III 11900 Penang, Malaysia
a) Single Output 2,000 Watt System dc Power Supplies b) Single Output 2,000 Watt Manually Controlled dc Power Supplies c) Single Output 5,000 Watt System dc Power Supplies d) Single Output 6,500 Watt System dc Power Supplies a) 6671A, 6672A 6673A, 6674A, 6675A b) 6571A, 6572A 6573A, 6574A, 6575A c) 6680A, 6681A, 6682A, 6683A, 6684A d) 6690A, 6691A, 6692A e) E4356A This declaration covers all options and customized products based on the above products.
CC/TCF/02/020 based on Technical Construction File (TCF) HPNJ2, dated June 4, 2002
Westfields House, West Avenue Kidsgrove, Stoke-on-Trent Straffordshire, ST7 1TL United Kingdom
IEC 61010-1:2001 / EN 61010-1:2001 Canada: CSA C22.2 No. 1010.1:1992 UL 61010B-1: 2003
Document No. 6x7y668xA.11.24.doc
PM
5
DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturer’s Name and Address
Responsible Party Agilent Technologies, Inc. Agilent Technologies (Malaysia) Sdn. Bhd 550 Clark Drive, Suite 101 Budd Lake, New Jersey 07828 USA
Declares under sole responsibility that the product as originally delivered
EMC Information ISM Group 1 Class A Emissions
Safety Information and Conforms to the following safety standards.
This DoC applies to above-listed products placed on the EU market after:
January 1, 2004 Date Bill Darcy/ Regulations Manager
For further information, please contact your local Agilent Technologies sales office, agent or distributor, or
Agilent Technologies Deutschland GmbH, Herrenberger Straβe 130, D71034 Böblingen, Germany
Revision: B.00.00 Issue Date: Created on 11/24/2003 3:26
Alternate Manufacturing Site
Product Names
Model Numbers
Product Options
Complies with the essential requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CE Marking accordingly.
As detailed in Electromagnetic Compatibility (EMC), Certificate of Conformance Number
Assessed by: Celestica Ltd, Appointed Competent Body
Malaysia Manufacturing Bayan Lepas Free Industrial Zone, PH III 11900 Penang, Malaysia
a) Single Output 500 Watt System dc Power Supplies b) Single Output 500 Watt Manually Controlled dc Power Supplies c) Single Output 500 Watt System Solar Array Simulator
a) 6651A, 6652A 6653A, 6654A, 6655A b) 6551A, 6552A 6553A, 6554A, 6555A c) E4350B, E4351B
This declaration covers all options and customized products based on the above products.
CC/TCF/00/074 based on Technical Construction File (TCF) HPNJ1, dated Oct. 27, 1997
Westfields House, West Avenue Kidsgrove, Stoke-on-Trent Straffordshire, ST7 1TL United Kingdom
IEC 61010-1:2001 / EN 61010-1:2001 Canada: CSA C22.2 No. 1010.1:1992 UL 61010B-1: 2003
PM
6x4yA6x5yAE435xA.b.11.24doc.doc
Document No.
6
DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturer’s Name and Address
Responsible Party Agilent Technologies, Inc. Agilent Technologies (Malaysia) Sdn. Bhd 550 Clark Drive, Suite 101 Budd Lake, New Jersey 07828 USA
Declares under sole responsibility that the product as originally delivered
EMC Information ISM Group 1 Class A Emissions
Safety Information and Conforms to the following safety standards.
This DoC applies to above-listed products placed on the EU market after:
January 1, 2004 Date Bill Darcy/ Regulations Manager
For further information, please contact your local Agilent Technologies sales office, agent or distributor, or
Agilent Technologies Deutschland GmbH, Herrenberger Straβe 130, D71034 Böblingen, Germany
Revision: B.00.00 Issue Date: Created on 11/24/2003 3:26
Alternate Manufacturing Site
Malaysia Manufacturing Bayan Lepas Free Industrial Zone, PH III 11900 Penang, Malaysia
Product Names
Model Numbers
Product Options
Complies with the essential requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CE Marking accordingly.
As detailed in Electromagnetic Compatibility (EMC), Certificate of Conformance Number
Assessed by: Celestica Ltd, Appointed Competent Body
a) Single Output 200 Watt System dc Power Supplies b) Single Output 200 Watt Manually Controlled dc Power Supplies
a) 6641A, 6642A 6643A, 6644A, 6645A b) 6541A, 6552A 6543A, 6544A, 6545A
This declaration covers all options and customized products based on the above products.
CC/TCF/00/074 based on Technical Construction File (TCF) HPNJ1, dated Oct. 27, 1997
Westfields House, West Avenue Kidsgrove, Stoke-on-Trent Straffordshire, ST7 1TL United Kingdom
IEC 61010-1:2001 / EN 61010-1:2001 Canada: CSA C22.2 No. 1010.1:1992 UL 61010B-1: 2003
PM
6x4yA6x5yAE435xA.a.11.24doc.doc
Document No.
7

Table of Contents

Certification 2 Safety Summary 3 Printing History 4 Declarations 5 Table of Contents 8
GENERAL INFORMATION 11
Introduction 11 Safety Considerations 12 Instrument Identification 12 Options 12 Accessories 13 Description 13
Front Panel Programming 14 Remote Programming 14 Analog Programming 14
Output Characteristic 15
General 15 Downprogramming 15
Specifications and Supplemental Characteristics 15
INSTALLATION 43
Inspection 43
Damage 43 Packaging Material 43 Items Supplied 43
Location and Temperature 44
Bench Operation 44 Rack Mounting 44 Temperature Performance 44
Input Power Source 45
Installing the Series 664xA and 665xA Power Cord 45 Installing the Series 667xA Power Cord 45 Installing the Series 668xA Power Cord 47 Installing the Series 669xA Power Cord 48
TURN-ON CHECKOUT 51
Introduction 51 Preliminary Checkout (All Models) 51 Power-On Checkout (All Models) 52 Using the Keypad (All Models) 52
Shifted Keys 52 Backspace Key 52
Output Checkout (All Models) 52
Checking the Voltage Function 53 Checking the Current Function 54
Checking the Save and Recall Functions (All Models) 55 Determining the GPIB Address (All Models) 55 In Case of Trouble 55
Line Fuse 55 Series 664xA and 665xA Supplies 56 Series 667xA Supplies 56 Series 668xA Supplies 57 Series 669xA Supplies 57
Error Messages (All Models) 57
Selftest Errors 57 Power-On Error Messages 57
8
Checksum Errors. 58 Runtime Error Messages 58
USER CONNECTIONS 59
Rear Panel Connections (All Models) 59 Load Wire Selection (All Models) 59 Analog Connector (All Models) 60 Digital Connector (All Models) 60 Connecting Series 664xA and 665xA Power Supplies to the Load 61
Output Isolation 61 Load Considerations 61 Local Voltage Sensing 62 Remote Voltage Sensing 63 Connecting One Supply to the Load 64 Connecting Supplies in Auto-Parallel 65 Connecting Supplies in Series 67 External Voltage Control 67
Connecting Series 667xA Power Supplies to the Load 68
Output Isolation 68 Load Considerations 69 Local Voltage Sensing 70 Remote Voltage Sensing 70 Connecting One Power Supply to a Single Load 72 Connecting One Power Supply To Multiple Loads 72 Connecting Supplies in Auto-Parallel 73 Connecting Supplies in Series 74 External Voltage Control 75
Connecting Series 668xA and 669xA Power Supplies to the Load 76
Output Isolation 76 Load Considerations 76 Local Voltage Sensing 77 Remote Voltage Sensing 77 Connecting One Power Supply to a Single Load 79 Connecting One Power Supply to Multiple Loads 79 Connecting Supplies in Auto-Parallel 80 Connecting Supplies in Series 81 External Voltage Control 82
Controller Connections 83
Stand-Alone Connections 83 Linked Connections 83
FRONT PANEL OPERATION 85
Introduction 85 Getting Acquainted 85 Programming the Output 88
Establishing Initial Conditions 88 Programming Voltage 89 Programming Overvoltage Protection 89 Programming Current 90 Programming Overcurrent Protection 91
CV Mode vs. CC Mode 91 Unregulated Operation 92 Saving and Recalling Operating States 92 Turn-On Conditions 92 Setting the GPIB Address 93
Types of Power Supply GPIB Addresses 93 Changing the Power Supply GPIB Address 93
CALIBRATION 95
9
Introduction 95 Equipment Required 95 General Procedure 95
Parameters Calibrated 95 Test Setup 96
Front Panel Calibration 96
Entering the Calibration Values 96 Saving the Calibration Constants 96 Disabling the Calibration Mode 96 Changing the Calibration Password 96 Recovering From Calibration Problems 99 Calibration Error Messages 99
Calibration Over the GPIB 100
Calibration Language Dictionary 100 CAL:CURR 100 CAL:CURR:LEV 100 CAL:CURR:MON (Series 668xA/669xA only) 101 CAL:PASS 101 CAL:SAVE 101 CAL:STAT 101 CAL:VOLT 102 CAL:VOLT:LEV 102 CAL:VOLT:PROT 102 Agilent BASIC Calibration Program 102
OPERATION VERIFICATION 105
Introduction 105 Test Equipment Required 105
List of Equipment 105 Current Monitoring Resistor 105
Performing the Tests 107
General Measurement Techniques 107 Programming the Power Supply 107 Order of Tests 107 Turn-on Checkout 107 Voltage Programming and Readback Accuracy 107 Current Programming and Readback Accuracy 108
LINE VOLTAGE CONVERSION 119
Series 664xA and 665xA Power Supplies 119 Series 667xA Power Supplies 120 Series 668xA/669xA Power Supplies 121
DIGITAL PORT FUNCTIONS 123
Digital Connector 123 Fault/Inhibit Operation 123 Changing the Port Configuration 126 Digital I/O Operation 126 Relay Link Operation 127
CURRENT LOOP COMPENSATION (SERIES 668XA ONLY) 129
Introduction 129 Function of Loop Compensation 129 Setting The Loop Compensation Switch 132
USING AGILENT 668XA/669XA SERIES POWER SUPPLIES IN AUTOPARALLEL 133
Introduction 133
OUTPUT BUS BAR OPTIONS 135
Option 601 Installation 135 Option 602 Installation 136
INDEX 137
10
1

General Information

Introduction

Two guides are shipped with your power supply - an Operating Guide (this document) and a Programming Guide. You will find information on the following tasks in these guides:
Quick Document Orientation
Topic Location
Calibrating the power supply Appendix A - this guide Compatibility programming language Appendix B - Programming Guide Configuring the digital port Appendix D - this guide Line voltage: Connecting ac power source Chapter 2 - this guide Converting the ac source voltage Appendix B - this guide Source current, frequency, and power ratings Chapter 1- this guide Operator replaceable parts Chapter 1- this guide Operator troubleshooting Chapter 3 - this guide Output impedance characteristics Chapter 1- this guide Power supply accessories Chapter 1- this guide Power supply operating characteristics Chapter 1- this guide Power supply options Chapter 1- this guide Power supply performance specifications Chapter 1- this guide Programming discrete fault inhibit (DFI) operation Chapter 4 - Programming Guide from the analog port Chapter 4 - this guide from the front panel Chapter 5 - this guide over the GPIB Chapter 2 - Programming Guide remote inhibit (RI) operation Chapter 4 - Programming Guide status registers Chapter 4 - Programming Guide Quick operating checkout (without load) Chapter 3 - this guide Rack mounting Chapter 2 - this guide SCPI programming language Chapter 3 - Programming Guide Wiring analog programming port Chapter 4 - this guide discrete fault indicator (DFI) operation Appendix D - this guide digital port Appendix D - this guide GPIB controller Chapter 4 - this guide load or loads Chapter 4 - this guide local sensing Chapter 4 - this guide remote inhibit (RI) operation Chapter 4 - this guide remote sensing Chapter 4 - this guide
1
See the Table of Contents for complete list of topics.
1
General Information 11

Safety Considerations

This power supply is a Safety Class 1 instrument, which means it has a protective earth terminal. That terminal must be connected to earth ground through a power source equipped with a 3-wire ground receptacle. Refer to the Safety Summary page at the beginning of this guide for general safety information. Before installation or operation, check the power supply and review this guide for safety warnings and instructions. Safety warnings for specific procedures are located at appropriate places in the guide. warning

Instrument Identification

The power supply is identified by a unique serial number that provides the following information:
US = The letter indicates the country of manufacture, where US = USA; MY = Malaysia. 33430177 = The first four digits indicate the year and week of manufacture or last significant design change. Add 1960 to
the first two digits to determine the year. For example, 32=1992, 33=1993, etc. The third and forth digits specify the week of the year (43 = the 43rd week). The last four digits is the number assigned to each unit.

Options

Option Description Used with Agilent Series
664xA 665xA 667xA 668xA 669xA
100 120 220 240 200 230 208 400 601 602 831 832 834 842 843
844
861
862
Input power 87-106 Vac, with power cord x x Input power 104-127 Vac, with power cord x x Input power 191-233 Vac, with power cord x x Input power 209-250 Vac, with power cord x x Input power 174-220 Vac, without power cord x Input power 191-250 Vac, without power cord x Input power 180-235 Vac, 3-phase, without power cord x x Input power 360-440 Vac, 3-phase, without power cord x x Output connector kit required for bench applications x x Bus bar spacers for paralleling power supplies x x Power cord, 12 AWG, UL listed, CSA certified, without plug x Power cord, 4mm2, harmonized, without plug x Power cord, 10 AWG, UL listed, CSA certified, without plug x Power cord, 4mm2, harmonized, with IEC309 32A/220V plug x Power cord, 12 AWG, UL listed, CSA certified, with JIS
C8303 25A/250V plug Power cord, 10 AWG, UL listed, CSA certified, with NEMA
L6-30P-30A/250V locking plug Power cord, 10 AWG, 4-wire, 300V, 25A, 90°C, UL listed,
CSA certified, without plug Power cord, 8 AWG, 4-wire, 300 V, 35A, 90°C, UL listed,
CSA certified, without plug Power cord, 2.5mm2, 4-wire, 450V, 20A, 70°C, harmonized,
without plug
x
x
x
x
x x
General Information 12
List of Options (continued)
Option Description Used with Agilent Series
664xA 665xA 667xA 668xA 669xA
908
909
Rack mount kit (Agilent 5062-3974) x
Rack mount kit (Agilent 5062-3977)
Support rails (E3663AC) are required.
Rack mount kit (Agilent 5062-3977 & 5062-3974)
Support rails (E3663AC) are required.
Rack mount kit with handles (Agilent 5062-3975) x
Rack mount kit with handles (Agilent 5062-3983)
Support rails (E3663AC) are required.
Rack mount kit with handles (Agilent 5062-3983 &
5062-3974) Support rails (E3663AC) are required.
x x
x x
x x
x x

Accessories

Description Agilent No.
Fuse replacement kit for Series 668xA 16 AM for 360-440 Vac, 3-phase line 5060-3512 30 AM for 180-235 Vac, 3-phase line 5060-3513 Fuse replacement kit for Series 669xA 16 AM for 360-440 Vac, 3-phase line 2110-1077 30 AM for 180-235 Vac, 3-phase line 2110-1078 GPIB cable (all models)
0.5 meters (1.6 ft) 10833D
1.0 meter (3.3 ft) 10833A
2.0 meters (6.6 ft) 10833B
4.0 meters ( 13 .2 ft) 10833C Serial link cable (all models)
2.0 meters (6.6 ft) 5080-2148 Slide mount kit heavy duty, for Series 667xA/668xA/669xA 1494-0058 standard, for Series 665xA 1494-0059 standard, for Series 664xA 1494-0060

Description

These units form a family of unipolar, GPIB programmable power supplies organized as follows:
Family Power Agilent Models
Series 664xA 200 W 6641A, 6642A, 6643A, 6644A, 6645A Series 665xA 500 W 6651A, 6652A, 6653A, 6654A, 6655A Series 667xA 2000 W 6671A, 6672A, 6673A, 6674A, 6675A Series 668xA 5000 W 6680A, 6681A, 6682A, 6683A, 6684A Series 669xA 6670 W 6690A, 6691A, 6692A
Each power supply is programmable locally from the front panel or remotely via a rear-panel analog control port.
General Information 13
Operational features include:
Constant voltage (CV) or constant current (CC) output over the rated output range.
Built-in overvoltage (OV), overcurrent (OC), and overtemperature (OT) protection.
Automatic turn-on selftest.
Pushbutton nonvolatile storage and recall of up to 5 operating states (4 in Series 668xA/669xA supplies).
Local or remote sensing of output voltage.
Auto-parallel operation for increased total current.
Series operation for increased total voltage.
Analog input for remote programming of voltage and current.
Voltage output for external monitoring of output current.
User calibration from the front panel.

Front Panel Programming

The front panel has both rotary (RPG) and keypad controls for setting the output voltage and current. The panel display provides digital readouts of the output voltage and current. Other front panel controls permit:
Enabling or disabling the output.
Setting the overvoltage protection (OVP) trip voltage.
Enabling or disabling the overcurrent protection (OCP) feature.
Saving and recalling operating states.
Setting the GPIB address.
Reading GPIB error message codes.
Calibrating the power supply, including changing the calibration protection password.

Remote Programming

The power supply may be remotely programmed via the GPIB bus and/or from an analog input port. GPIB programming is with SCPI (Standard Commands for Programmable Instruments) commands that make the power supply programs compatible with those of other GPIB instruments. (A software-controlled Compatibility Mode also permits programming in the command set of the Agilent 6030xA Autoranging Series.) In addition to control functions, SCPI programming permits writing to the front panel LCD and complete calibration functions. Power supply status registers permit remote monitoring of the following conditions:
Overvoltage, overcurrent, overtemperature, and unregulated states.
Operating mode (constant voltage or constant current).
State of the RI (remote inhibit) input signal.
Power-on status (PON).
Status of the output queue (QYE).
Pending triggers (WTG).
GPIB interface programming errors (CME, DDE, and EXE).
Calibration state (enabled or disabled).
The status registers can be programmed to generate an output fault signal (FLT) upon the occurrence of one or more selected status events.

Analog Programming

The power supply has an analog port for remote programming. The output voltage and/or current of the power supply may be controlled by individual dc programming voltages applied to this port. The port also provides a monitor output that supplies a dc voltage proportional to the output current.
General Information 14

Output Characteristic

General

The power supply can operate in either CV (constant voltage) or CC (constant current) over its voltage and current ratings (see Table 1-l). The operating locus is shown by the Output Characteristic Curve in Table 1-2. The operating point is determined by the voltage setting (V Point 1 is defined by the load line cutting the operating locus in the constant-voltage region. This region defines the CV mode. Point 2 is defined by the load line cutting the operating locus in the constant-current region. This region defines the CC mode.
), the current setting (Is), and the load impedance. Two operating points are shown.
s

Downprogramming

The power supply can sink current for more rapid down programming in the CV mode. For Series 664xA and 665xA supplies, this capability is defined by the second quadrant area (-I sink about 20% of their maximum rated positive output current. For Series 667xA, 668xA, and 669xA power supplies, this is an uncharacterized current-sinking area that provides a limited downprogramming capability.
) of the Output Characteristic Curve. These supplies can
s

Specifications and Supplemental Characteristics

Tables 1-1 through 1-4 list the specifications and supplemental characteristics for the Series 664xA, 665xA, 667xA, 668xA, and 669xA power supplies. The organization is as follows:
Series Specifications Characteristics
6641A-6645A Table l-la Table l-lb 6651A-6655A Table 1-2a Table 1-2b 6671A-6675A Table 1-3a Table 1-3b 6680A-6684A Table 1-4a Table 1-4b 6690A-6692A Table 1-5a Table 1-5b
Specifications are performance parameters warranted over the specified temperature range.
Supplemental Characteristics are not warranted but are descriptions of performance determined either by design or type
testing.
General Information 15
Table 1-1a. Performance Specifications for Series 664xA
1
Parameter Agilent Model Number
6641A 6642A 6643A 6644A 6645A
Output Ratings Voltage: Current:@ 40°°°°C Current:@ 50°°°°C Current:@ 55°°°°C
0 - 8 V 0 - 20 V 0 - 35 V 0 - 60 V 0 - 120 V 0 - 20 A 0 - 10 A 0 - 6 A 0 - 3.5 A 0 - 1.5 A 0 - 18 A 0 - 9 A 0 - 5.4 A 0 - 3.2 A 0 - 1.4 A 0 - 17 A 0 - 8.5 A 0 - 5.1 A 0 - 3.0 A 0 -1.4 A
Programming Accuracy (@ 25 ± 5 °C) Voltage: 2 0.06% +
Current: 0.l5 % +
5 mV
26 mA 13 mA
10 mV 15 mV
6.7 mA
26 mV 51 mV
4.1 mA
1.7 mA
Ripple & Noise (from 20 Hz to 20 MHz with outputs ungrounded, or with either output terminal grounded) Constant Voltage: rms Constant Voltage: p-p Constant Current: rms
300 µV 300 µV 400 µV 500 µV 700 µV
3 mV 3 mV 4 mV 5 mV 7 mV
10 mA
5 mA 3 mA 1.5 mA 1 mA
Readback Accuracy (from front panel or over GPIB with respect to actual output @ 25 ±:5 °C) Voltage: 2 0.07% + +Current 0.15% +
-Current 0.35% +
6 mV 15 mV 25 mV 40 mV 80 mV 18 mA 9.1 mA 5 mA 3 mA 1.3 mA 40 mA 20 mA 12 mA 6.8 mA 2.9 mA
Load Regulation (change in output voltage or current for any load change within ratings) Voltage Current:
1 mV 2 mV 3 mV 4 mV 5 mV
1 mA 0.5 mA 0.25 mA 0.25 mA 0.25 mA
Line Regulation (change in output voltage or current for any line change within ratings Voltage: Current:
0.5 mV 0.5 mV 1 mV 1 mV 2 mV 1 mA 0.5 mA 0.25 mA 0.25 mA 0.25 mA
Transient Response Time (for the output voltage to recover to its previous level (within 0.1% of the rated voltage or 20 mV, whichever is greater) following any step change in load current up to 50% of the rated current.
< 100 µs
AC Input Ratings (selectable via internal switching - see Appendix C)
Nominal line voltage:
Frequency:
Output Terminal Isolation
Notes:
2
3
1
For Supplemental Characteristics, see Table 1-1b.
Specification may degrade slightly when unit is subjected to an RF field 3V/meter. For 230Vac operation, unit is internally set to 240Vac
100, 120, 220, 240 Vac (-13%, +6 %)
230 Vac
±240 Vdc (maximum, from chassis ground)
3
(-10%, +10%)
50/60 Hz
General Information 16
Table 1-1b. Supplemental Characteristics for Series 664xA
1
Parameter Agilent Model Number
6641A 6642A 6643A 6644A 6645A
Output Programming Range (maximum programmable values)
Voltage: Current: Overvoltage Protection (OVP):
8.190 V 20.475 V 35.831 V 61.425 V 122.85 V
20.475 A 10.237 A 6.142 A 3.583 A 1.535 A
8.8 V 22.0 V 38.5 V 66.0 V 132.0 V
Average Resolution Voltage: Current: Overvoltage Protection (OVP):
2 mV 5 mV 10 mV 15 mV 30 mV 6 mA 3 mA 2 mA 1.2 mA 0.5 mA
13 mV 30 mV 54 mV 93 mV 190 mV
Accuracy Overvoltage Protection (OVP): Analog Programming (VP):*
160 mV 400 mV 700 mV 1.2 V 2.4 V
0.36% + 6 mV 15 mV 27 mV 45 mV 90 mV
Analog Programming (IP):*
7.6% + 18 mA 9.2 mA
1.5% + 5.5 mA 3.2 mA 1.4 mA
Current Monitor (+IM):*
7.7% + 65 mA 32 mA
1.6 % + 8.1 mA 7.1 mA 1.8 mA *Referenced to supply output Drift Temperature Stability (following a 30-minute warmup, change in output over 8 hours under constant line, load, and ambient temperature)
Voltage: 0.02% + Current: 0.02% + Temperature Coefficients (change per °C)
0.4 mV 1 mV 2 mV 3 mV 6 mV 16 mA 6 mA 3 mA 2 mA 1 mA
Voltage: +Current: Voltage Readback: +Current Readback:
--Current Readback: Overvoltage Protection (OVP):
Analog Programming (VP):
Analog Programming (IP):
Current Monitor (+IM):
Maximum Input Power:
60 ppm + 0.1 mV 0.2 mV 0.3 mV 0.5 mV 1.1 mV 95 ppm + 0.82 mA 0.41 mA 0.18 mA 0.12 mA 0.04 mA 60 ppm + 0.2 mV 0.5 mV 0.75 mV 1.3 mV 2.6 mV 95 ppm + 1.2 mA 0.62 mA 0.33 mA 0.20 mA 0.08 mA
110 ppm + 1.2 mA 0.62 mA 0.33 mA 0.20 mA 0.08 mA
200 ppm + 1.6 mV 3.3 mV 5 mV 13 mV 24 mV
60 ppm + 0.1 mV 0.25 mV 0.4 mV 0.7 mV 1.25 mV
90 ppm + 0.56 mA 0.28 mA 0.17 mA 0.1 mA 0.04 mA
75 ppm + 0.61 mA 0.3 mA 0.06 mA 0.06 mA 0.02 mA
480 VA; 400 W, 60 W with no load
Notes: 1For Performance Specifications, see Table 1-la.
:
General Information 17
Table 1-lb. Supplemental Characteristics for Series 664xA (continued)
1
Parameter Agilent Model Number
6641A 6642A 6643A 6644A 6645A
Maximum AC Line Current Ratings 100 Vac nominal: 120 Vac nominal: 220 Vac nominal: 230 Vac nominal: 240 Vac nominal:
Maximum Reverse Bias Current:
With AC input power applied and the dc output reverse biased by an external dc source, the supply will continuously
4.4 A rms
3.8 A rms
2.2 A rms
2.1 A rms
2.0 A rms
withstand without damage a current equal to its output current rating (see Table 1- 1a).
Remote Sensing Capability Voltage Drop Per Lead:
Load Regulation:
Load Voltage:
Up to 1/2 of rated output voltage.
Add 3 mV to spec (see Table l-la) for each l-volt change in the + output lead due to load current changes.
Subtract voltage drop in load leads from specified output
voltage rating. Command Processing Time (Average time for output voltage to change after receipt of digital data when the supply is connected directly to the GPIB Bus):
Downprogrammer Current Capability (± 15%):
5.8 A 2.5 A 1.5 A 0.9 A 0.75 A
20 ms
Output Voltage Programming Response Time Rise/Fall Time (time for output to change from 90 % to 10% or from 10% to 90% of its total excursion):
<15 ms
Settling Time (time for output change to settle within 1 LSB (0.025% x rated voltage) of its final value):
Monotonicity:
Auto-Parallel Configuration:
<60 ms
Output is monotonic over entire rated voltage, current, and
temperature range.
Up to 3 identical models
Analog Programming (IP & VP) Input Signal:* Input Impedance: *Signal source must be isolated.
Current Monitor Output (+IM):
Savable States Nonvolatile Memory Locations: Nonvolatile Memory Write Cycles: Prestored State (factory default):
10 k, nominal
0 to -5 V represents zero to full-scale current output
5 ( 0 through 4)
0 to -5 V
40,000, typical
Location 0
Notes: lFor Performance Specifications, see Table l-la.
General Information 18
Table 1-1b. Supplemental Characteristics for Series 664xA (continued)
Parameter All Models
Digital Port Characteristics
GPIB Interface Capabilities
Serial Link Capabilities
Recommended Calibration Interval:
Safety Compliance Complies with: Designed to comply with:
RFI Suppression (complies with):
Dimensions Width:
Height (including removable feet): Depth (including safety cover):
Note 1: For Performance Specifications, see Table l-la.:
Weight Net: Shipping:
Output Characteristic Curve:
14.2 kg (31.4 lb)
16.3 kg (36 lb)
1
(see Table 1-6)
(see Table 1-6)
(see Table 1-6)
1 year
CSA 22.2 No.231,IEC 348
UL 1244
CISPR-ll, Group 1, Class B
425.5 mm (16.75 in)
88.1 mm (3.5 in)
439 mm (17.3 in)
Notes: lFor Performance Specifications, see Table l-la.
General Information 19
Table 1-1b. Supplemental Characteristics for Series 664xA (continued)
Parameter All Models
Output Impedance Curves (Typical):
1
Notes: lFor Performance Specifications, see Table l-la.
General Information 20
Table 1-2a. Performance Specifications for Series 665xA
1
Parameter Agilent Model Number
6651A 6652A 6653A 6654A 6655A
Output Ratings Voltage: Current:@ 40°°°°C Current:@ 50°°°°C Current:@ 55°°°°C
0 - 8 V 0 - 20 V 0- 35 V 0 - 60 V 0 - 120 V 0 - 50 A 0 - 25 A 0 - 15 A 0 - 9 A 0 - 4 A 0 - 45 A 0 - 22.5 A 0 - 13.5 A 0 - 8.1 A 0 - 3.6 A
0 - 42.5 A 0 - 21.3 A 0 - 12.8 A 0 - 7.7 A 0 -3.4 A
Programming Accuracy (@ 25 ± 5 °C) Voltage: 2 0.06% +
Current: 0.l5 % +
5 mV
60 mA 25 mA
10 mV 15 mV
13 mA
26 mV 51 mV
8 mA
4 mA
Ripple & Noise (from 20 Hz to 20 MHz with outputs ungrounded, or with either output terminal grounded) Constant Voltage: rms Constant Voltage: p-p Constant Current: rms
300 µV 300 µV 400 µV 500 µV 700 µV
3 mV 3 mV 4 mV 5 mV 7 mV
25 mA
10 mA 5 mA 3 mA 2 mA
Readback Accuracy (from front panel or over GPIB with respect to actual output @ 25 ± 5 °C) Voltage: 2 0.07% + +Current 0.15% +
-Current 0.35% +
6 mV 15 mV 25 mV 40 mV 80 mV
67 mA 26 mA 15 mA 7 mA 3 mA 100 mA 44 mA 24 mA 15 mA 7 mA
Load Regulation (change in output voltage or current for any load change within ratings) Voltage: Current:
1 mV 2 mV 3 mV 4 mV 5 mV 2 mA 1 mA 0.5 mA 0.5 mA 0.5 mA
Line Regulation (change in output voltage or current for any line change within ratings Voltage: Current:
0.5 mV 0.5 mV 1 mV 1 mV 2 mV 2 mA 1 mA 0.75 mA 0.5 mA 0.5 mA
Transient Response Time (for the output voltage to recover to its previous level (within 0.1% of the rated voltage or 20 mV, whichever is greater) following any step change in load current up to 50% of the rated current.
< 100 µs
AC Input Ratings (selectable via internal switching - see Appendix C)
Nominal line voltage:
Frequency:
Output Terminal Isolation
Notes:
2
3
1
For Supplemental Characteristics, see Table 1-2b.
Specification may degrade slightly when unit is subjected to an RF field 3V/meter. For 230Vac operation, unit is internally set to 240Vac
100, 120, 220, 240 Vac (-13%, +6 %)
230 Vac
±240 Vdc (maximum, from chassis ground)
3
(-10%, +10%)
50/60 Hz
General Information 21
Table 1-2b. Supplemental Characteristics for Series 665xA
1
Parameter Agilent Model Number
6651A 6652A 6653A 6654A 6655A
Output Programming Range (maximum programmable values)
Voltage: Current: Overvoltage Protection (OVP):
8.190 V 20.475 V 35.831 V 61.425 V 122.85 V
51.188 A 25.594 A 15.356 A 9.214 A 4.095 A
8.8 V 22.0 V 38.5 V 66.0 V 132.0 V
Average Resolution Voltage: Current: Overvoltage Protection (OVP):
2 mV 5 mV 10 mV 15 mV 30 mV 15 mA 7 mA 4 mA 2.5 mA 1 mA 13 mV 30 mV 54 mV 93 mV 190 mV
Accuracy Overvoltage Protection (OVP):* Analog Programming (VP):*
160 mV 400 mV 700 mV 1.2 V 2.4 V
0.36% + 6 mV 15 mV 27 mV 45 mV 90 mV
Analog Programming (IP):*
7% + 75 mA 31 mA 16 mA 8 mA 5 mA
Current Monitor (+IM):*
7% + 730 mA 400 mA 120 mA 80 mA 75 mA *Referenced to supply output Drift Temperature Stability (following a 30-minute warmup, change in output over eight hours under constant line, load, and ambient temperature)
Voltage: 0.02% + Current: 0.02% + Temperature Coefficients (change per °C)
0.4 mV 1 mV 2 mV 3 mV 6 mV 40 mA 15 mA 8 mA 5 mA 2.5 mA
Voltage: +Current: Voltage Readback: +Current Readback:
-Current Readback: Overvoltage Protection (OVP):
Analog Programming (VP):
Analog Programming (IP):
Current Monitor (+IM):
Maximum Input Power
60 ppm + 0.1 mV 0.2 mV 0.3 mV 0.5 mV 1.1 mV 90 ppm + 1.4 mA 0.7 mA 0.3 mA 0.2 mA 0.2 mA 60 ppm + 0.2 mV 0.5 mV 0.75 mV 1.3 mV 2.6 mV 90 ppm + 1.7 mA 0.9 mA 0.5 mA 0.3 mA 0.2 mA
105 ppm + 1.7 mA 0.9 mA 0.5 mA 0.3 mA 0.2 mA
200 ppm + 1.6 mV 3.3 mV 5 mV 13 mV 24 mV
60 ppm + 0.1 mV 0.25 mV 0.4 mV 0.7 mV 1.25 mV
90 ppm + 1.4 mA 0.7 mA 0.3 mA 0.2 mA 0.15 mA
80 ppm + 1.4 mA 0.7 mA 0.3 mA 0.2 mA 0.15 mA
1380 VA; 1100 W, 120 W with no load
Notes: 1For Performance Specifications, see Table 1-2a.
General Information 22
Table 1-2b. Supplemental Characteristics for Series 665xA (continued)
1
Parameter Agilent Model Number
6651A 6652A 6653A 6654A 6655A
Maximum AC Line Current Ratings 100 Vac nominal: 120 Vac nominal: 220 Vac nominal: 230 Vac nominal: 240 Vac nominal:
Maximum Reverse Bias Current:
12 A rms (15 AM fuse) 10 A rms (12 AM fuse)
5.7 A rms (7 AM fuse)
5.5 A rms (7 AM fuse)
5.3 A rms (7 AM fuse) With AC input power applied and the dc output reverse biased by an external dc source, the supply will continuously withstand without damage a current equal to its output current rating (see Table 1- 2a) .
Remote Sensing Capability Voltage Drop Per Lead:
Load Regulation:
Load Voltage:
Up to 1/2 of rated output voltage.
Add 3 mV to spec (see Table l-2a) for each l-volt change in the + output lead due to load current changes.
Subtract voltage drop in load leads from specified output voltage rating.
Command Processing Time (Average time for output voltage to change after receipt of digital data when the supply is connected directly to the GPIB Bus):
Downprogrammer Current Capability (± 15%):
11.6 A 5 A 3 A 1.8 A 1.5 A
20 ms
Output Voltage Programming Response Time Rise/Fall Time (time for output to change from 90 % to 10% or from 10% to 90% of its total excursion):
<15 ms
Settling Time (time for output change to settle within 1 LSB (0.025% x rated voltage) of its final value):
Monotonicity:
Auto-Parallel Configuration:
Output is monotonic over entire rated voltage, current, and temperature range.
Up to 3 identical models
<60 ms
Analog Programming (IP & VP) Input Signal:* Input Impedance: *Signal source must be isolated.
Current Monitor Output (+IM):
Savable States Nonvolatile Memory Locations: Nonvolatile Memory Write Cycles: Prestored State (factory default):
10 k, nominal
0 to -5 V represents zero to full-scale current output.
5 ( 0 through 4)
0 to -5 V
40,000, typical
Location 0
Notes: lFor Performance Specifications, see Table l-2a.
General Information 23
Table 1-2b. Supplemental Characteristics for Series 665xA (continued)
Parameter All Models
Digital Port Characteristics
GPIB Interface Capabilities
Serial Link Capabilities
Recommended Calibration Interval:
Safety Compliance Complies with: Designed to comply with:
RFI Suppression (complies with):
Dimensions Width:
Height (including removable feet): Depth (including safety cover):
Weight Net: Shipping:
Output Characteristic Curve:
25 kg (54 lb) 28 kg (61 lb)
1
(see Table 1-6)
(see Table 1-6)
(see Table 1-6)
1 year
CSA 22.2 No.231,IEC 348
UL 1244
CISPR-ll, Group 1, Class B
425.5 mm (16.75 in)
132.6 mm (5.22 in)
497.8 mm (19.6 in)
Notes: lFor Performance Specifications, see Table l-2a.
General Information 24
Table 1-2b. Supplemental Characteristics for Series 665xA (continued)
Parameter All Models
Output Impedance Curves (Typical):
1
Notes: lFor Performance Specifications, see Table l-2a.
General Information 25
Table 1-3a. Performance Specifications for Series 667xA
1
Parameter Agilent Model Number
6671A 6672A 6673A 6674A 6675A
Output Ratings Voltage: Current:@ 0 to 55°°°°C
Programming Accuracy (@ calibration temperature* ± 5 °C)
0 - 8 V 0 - 20 V 0- 35 V 0 - 60 V 0 - 120 V
0 - 220 A 0 - 100 A 0 - 60 A 0 - 35 A 0 - 18 A
Voltage: 0.04% +
Current: 0 . l % +
8 mV
125 mA 60 mA
20 mV 35 mV
40 mA
60 mV 120 mV
25 mA
12 mA
Ripple & Noise (from 20 Hz to 20 MHz with outputs ungrounded, or with ei ther output terminal grounded) Constant Voltage: rms Constant Voltage: p-p Constant Current: rms
650 µV 750 µV 800 µV
7 mV 9 mV 9 mV 11 mV 16 mV
200 mA
100 mA 40 mA 25 mA 12 mA
1.25 mV 1.9 mV
Readback Accuracy (from front panel or over GPIB with respect to actual output @ calibration temp1 ± 5 °C) Voltage: 0.05% + ±±±±Current: 0.1% +
12 mV 30 mV 50 mV 90 mV 180 mV
150 mA 100 mA 60 mA 35 mA 18 mA
Load Regulation (change in output voltage or current for any load change within ratings) Voltage: 0.002% + Current: 0.005% +
300 µV 650 µV
10 mA 7 mA 4 mA 2 mA 1 mA
1.2 mV 2 mV 4 mV
Line Regulation (change in output voltage or current for any line change within ratings Voltage: 0.002% + Current: 0.005% +
300 µV 650 µV
10 mA 7 mA 4 mA 2 mA 1 mA
1.2 mV 2 mV 4 mV
bisep>
Transient Response Time (for the output voltage to recover to its previous level (within 0.1% of the rated voltage or 100 mV, whichever is greater) following any step change in load current up to 50% of the rated current.
< 900 µs
AC Input Ratings (selectable via internal switching - see Appendix C) Nominal line voltage:
220, 230, 240 Vac (191-253 Vac range)
200 Vac (174-220 Vac range)*
*below 185 Vac, derate output voltage linearly to:
7.8 V 18.0 V 31.5 V 56.5 V 108 V Frequency:
Output Terminal Isolation
±240 Vdc (maximum, from chassis ground)
50/60 Hz
Notes: 1For Supplemental Characteristics, see Table 1-3b.
General Information 26
Table 1-3b. Supplemental Characteristics for Series 667xA
1
Parameter Agilent Model Number
6671A 6672A 6673A 6674A 6675A
Output Programming Range (maximum programmable values)
Voltage: Current: Overvoltage Protection (OVP):
8.190 V 20.475 V 35.831 V 61.425 V 122.85 V
225.23 A 102.37 A 61.43 A 35.83 A 18.43 A
10.0 V 24.0 V 42.0 V 72.0 V 144.0 V
Typical Resolution Voltage: Current: Overvoltage Protection (OVP):
2 mV 5 mV 10 mV 15 mV 30 mV 55 mA 25 mA 15 mA 8.75 mA 4.5 mA 15 mV 35 mV 65 mV 100 mV 215 mV
Accuracy ( @ calibration temp ±5 °C)* Overvoltage Protection (OVP):* Analog Programming (VP): Analog Programming (IP): Current Monitor (+IM):
*Calibration temp = 25° C
200 mV 500 mV 900 mV 1.15 V 3.0 V
± 0.3%
± 7%
±7%
Drift Temperature Stability (following a 30-minute warmup, change in output over eight hours under constant line, load, and ambient temperature)
Voltage: 0.02% + Current: 0.02% + Temperature Coefficients (change per °C after 30-minute warmup)
0.24 mV 0.6 mV 1 mV 1.8 mV 3.6 mV 69 mA 35 mA 20 mA 10 mA 6 mA
Voltage: Current: Voltage Readback: ±±±±Current Readback: Overvoltage Protection (OVP):
Analog Programming (VP):
Analog Programming (±±±±IP):
Current Monitor (+IM):
Maximum Input VA and Power
Maximum AC Line Current Ratings
50 ppm + 0.04 mV 0.2 mV 0.7 mV 1.2 mV 2.4 mV 75 ppm + 25 mA 12 mA 7 mA 4 mA 2 mA 60 ppm + 0.1 mV 0.3 mV 1 mV 1.2 mV 3 mV 85 ppm + 30 mA 15 mA 9 mA 5 mA 2.5 mA
200 ppm + 1.8 mV 5 mV 8 mV 13 mV 25 mV
60 ppm + 0.1 mV 0.3 mV 0.5 mV 0.7 mV 1.5 mV
275 ppm + 26 mA 14 mA 9 mA 5 mA 3 mA
50 ppm + 3 mA 2 mA 1 mA 0.6 mA 0.3 mA
3800 VA; 2600 W, 100 W with no load
200 Vac
nominal:
230 Vac
19 A rms (25 AM fuse)
19 A rms (25 AM fuse)
nominal:
Maximum Reverse Bias Current:
With AC input power applied and the dc output reverse biased by an external dc source, the supply will continuously withstand without damage a current equal to its output current rating (see Table 1-3a).
Notes: 1For Performance Specifications, see Table 1-3a.
General Information 27
Table 1-3b. Supplemental Characteristics for Series 667xA (continued)
1
Parameter Agilent Model Number
6671A 6672A 6673A 6674A 6675A
Remote Sensing Capability Voltage Drop Per Lead: Load Voltage:
Up to 1/2 of rated output voltage. Subtract voltage drop in load leads from specified output voltage rating.
Load Regulation: Degradation due to load lead drop in--output: mV (regulation) = Vdrop(R
sense-
)/k
Degradation due to load lead drop in + output:
mV (regulation) = V where R
sense
- and R
drop(Rsense
sense
+)/k + 2V
drop(Vrating
+ are resistances of respective sense leads and k is the following model-dependent
)/(V
+ 10 V)
rating
value: 6671A=1; 6672A=1.82; 6673A=4.99; 6674A=10; 6675A=16.2
Command Processing Time (Average time for output voltage to change after receipt of digital data when the supply is connected directly to the GPIB Bus):
20 ms
Output Voltage Programming Response Time** Rise/Fall Time (time for output to change from 90 % to 10% or from 10% to 90% of its total excursion):***
30 ms 60 ms 130 ms 130 ms 195 ms
Full-load programming speed up/down time (time for output to settle within 4 LSBs of the final value):***
85 ms 190 ms 380 ms 380 ms 600 ms
No-load downprogrammiug discharge time (time for output to fall to 0.5 V when programmed from full voltage to
zero volts):
130 ms 250 ms 350 ms 600 ms 600 ms
** All values exclude command processing time. *** With full resistive load = V Monotonicity:
Auto-Parallel Configuration:
RATED/IRATED.
Output is monotonic over entire rated voltage, current, and temperature range.
Up to 3 identical models
Analog Programming (IP & VP) Input Signal:* VP Input Signal:** (0 to ) VP Input Impedance: IP to -IP Differential Input Signal: (0 to )
*Signal source must be isolated.
-4.72 V -4.24 V -4.25 V -4.24 V -3.97 V
60 k, nominal
+7.79 V +6.81 V +6.81 V =7.01 V +6.34 V
** Referenced to output signal common.
Current Monitor Output (+IM): Output Signal:* (-0.25 to ) Output Impedance:
+9.05 V +7.70 V +7.70 V +7.93 v +7.15 V
490
* Corresponds to 0% to 100% output current. Savable States Nonvolatile Memory Locations: Nonvolatile Memory Write Cycles: Prestored State (factory default):
5 ( 0 through 4)
40,000, typical
Location 0
Notes: lFor Performance Specifications, see Table l-3a.
General Information 28
Table 1-3b. Supplemental Characteristics for Series 667xA (continued)
Parameter All Models
Digital Port Characteristics
GPIB Interface Capabilities
Serial Link Capabilities
Recommended Calibration Interval:
Safety Compliance Complies with: Designed to comply with:
RFI Suppression (complies with):
Dimensions Width:
Height (including removable feet): Depth (including safety cover):
Weight Net: Shipping:
Output Characteristic Curve:
1
(see Table 1-6)
(see Table 1-6)
(see Table 1-6)
1 year
CSA 22.2 No.231,IEC 348
UL 1244
CISPR-ll, Group 1, Class B
425.5 mm (16.75 in)
145.1 mm (5.71 in) 640 mm (25.2 in)
27.7 kg (61 lb)
31.4 kg (69 lb)
Notes: lFor Performance Specifications, see Table l-3a.
General Information 29
Table 1-3b. Supplemental Characteristics for Series 667xA (continued)
Parameter All Models
Output Impedance Curves (Typical):
1
Notes: lFor Performance Specifications, see Table l-3a.
General Information 30