1 Problem
Cassette auto-detection does not work any more after Service Pack 5 has been installed on the
PRID station. This happens only if the PC has a Pentium III processor inside.
2 Reason
The current Java runtime version (classes.jar) has problems when run together with Service
Pack 5 and Pentium III processors.
3 Workaround
There are two workarounds responding to that problem.
Perform only one of the following actions:
Copying the file “classes.jar”
3.1
Change the Target line in the Properties Window
3.2
3.1 Copying the file “classes.jar”
Note: The file classes.jar can be downloaded from the mailbox.
Copy the file “classes.jar” to the directory “c:\prid”:
<copy A:\classes.jar C:\prid\classes.jar>
Restart the program.
printed in Germany
FIELDSERVICE-BULLETIN
3.2 Change the Target line in the Properties Window
DD+DIS269.99E
On your desktop, you can
find the PRID icon (see
arrow).
click the right mouse
•
button once to open a
pop-up menu.
click on properties (see
•
arrow).