Workaround for “No Auto-Detection of Cassettes”

1 Problem

Cassette auto-detection does not work any more after Service Pack 5 has been installed on the PRID station. This happens only if the PC has a Pentium III processor inside.

2 Reason

The current Java runtime version (classes.jar) has problems when run together with Service Pack 5 and Pentium III processors.

3 Workaround

There are two workarounds responding to that problem. Perform only one of the following actions:
Copying the file “classes.jar”
3.1
Change the Target line in the Properties Window
3.2

3.1 Copying the file “classes.jar”

Note: The file classes.jar can be downloaded from the mailbox.
Copy the file “classes.jar” to the directory “c:\prid”: <copy A:\classes.jar C:\prid\classes.jar>
Restart the program.
printed in Germany
FIELDSERVICE-BULLETIN
HOME

3.2 Change the Target line in the Properties Window

DD+DIS269.99E
On your desktop, you can find the PRID icon (see arrow).
click the right mouse
button once to open a pop-up menu.
click on properties (see
arrow).