A&D PD4000243 User Manual

Multi Function Weighing Indicator
WM : PD4000243
This is a hazard alert mark.
This mark informs you about the operation of the product.
Note This manual is subject to change without notice at any time to improve the product. No part
of this manual may be photocopied, reproduced, or translated into another language without the prior written consent of the A&D Company.
Contents
1. Compliance........................................................................................ 4
1.1.1. Compliance with FCC rules.......................................................4
1.1.2. Compliance with European Directive.........................................4
2. Outline and Features.........................................................................5
2.1. Precaution.....................................................................................6
2.2. Front Panel....................................................................................7
2.2.1. Keys..........................................................................................7
2.2.2. Symbols ....................................................................................8
2.3. Rear Panel...................................................................................10
3. Installation........................................................................................11
3.1. Mounting Indicator.......................................................................11
3.2. Connecting Loadcell Cable..........................................................12
3.2.1. Verifying Loadcell Output and Input Sensitivity.......................13
3.3. Wiring Power Cord ......................................................................14
3.4. Installing Options.........................................................................15
4. Basic Operation...............................................................................16
4.1. Key Operation Examples............................................................. 16
4.1.1. Standby Mode.........................................................................16
4.1.2. Cursor Operation.....................................................................16
4.1.3. Inputting Characters................................................................16
4.1.4. The Way of Calling a Code .....................................................17
4.1.5. The Way of Entering a Correction Mode.................................17
4.1.6. The Way of Entering Menu......................................................18
4.2. Status Chart (Mode map) ............................................................19
5. Calibration........................................................................................20
5.1. Actual Load Calibration (using Mass)..........................................21
5.2. Digital Span (Calibration without Mass)....................................... 22
5.3. Gravity Acceleration Correction...................................................23
5.3.1. Gravity Acceleration Reference...............................................23
5.4. Calibration Error ..........................................................................24
6. Applications .....................................................................................25
6.1. Hopper Scale with Material Code................................................25
6.1.1. Definition of Material Code......................................................25
6.1.2. Recalling Material Code..........................................................25
6.1.3. Editing Principle Parameters of Material Code........................26
6.1.4. Referring next Material Code...................................................26
6.1.5. Editing Full Parameters of Material Code................................27
6.2. Simple Hopper Scale with Recipe Code......................................30
6.2.1. Definition of Recipe Code (Formal Function)...........................30
6.2.2. Using a Recipe Code ..............................................................31
6.2.3. Construction of Recipe Code...................................................31
6.2.4. Recalling a Recipe Code.........................................................31
6.2.5. Arranging Material Code in Recipe Code................................32
6.2.6. Editing Full Parameters of Recipe Code .................................32
©A&D Co. ltd., AD-4402 Instruction Manual International Version 1875-1A-IE 200105
6.3. System Design of Hopper Scale..................................................34
6.3.1. Operation and I/O Design........................................................34
6.3.2. Design Example......................................................................34
7. Weighing Mode................................................................................ 35
7.1.1. Contents of Batch Weighing Mode..........................................35
7.2. Batch Weighing Mode .................................................................36
7.2.1. Selection of Batch Weighing....................................................37
7.3. Sequential Weighing Mode..........................................................38
7.3.1. Normal Batching of Sequential Weighing................................38
7.3.2. Loss-in Weight of Sequential Mode.........................................41
7.3.3. Compensation Sequence........................................................43
7.3.4. Entrance Sequence.................................................................45
7.3.5. Discharge Sequence...............................................................47
7.3.6. Plain Recipe Sequence...........................................................49
7.3.7. Automatic Selection of Supplying Mat.....................................51
7.3.8. Nozzle Control Sequence (vacuum cleaner)........................... 52
7.3.9. Mixture Sequence...................................................................53
7.3.10. Safety Check Function............................................................55
7.3.11. Pause and Emergency Stop....................................................55
7.3.12. Restart Sequences from Pause...............................................56
7.3.13. Automatic Free Fall Compensation.........................................57
7.3.14. Real Time Free Fall Compensation.........................................58
7.4. Customer Programmed Control
(Comparison Output)
...............................59
7.4.1. Normal Batching of Comparison Function...............................60
7.4.2. Loss-in Weight of Comparison Function..................................62
7.5. Other Functions...........................................................................64
7.5.1. Re-Zero Operation..................................................................64
7.5.2. Zero Tracking Function ...........................................................64
7.5.3. Tare.........................................................................................65
7.5.4. Preset Tare (Fixed Tare Function) ..........................................65
7.5.5. Customizing Function Key (Key Design).................................65
7.5.6. Customizing Sub Display ........................................................66
7.5.7. Graphic Display.......................................................................67
7.5.8. Accumulation Operation..........................................................68
7.5.9. Undoing Accumulation Operation............................................ 68
7.5.10. Clearing (Deleting) Accumulation Data ...................................68
7.5.11. Error Message and Alarm .......................................................69
7.5.12. Graphic Status Indicator..........................................................71
7.5.13. Memory Backup......................................................................71
8. Interface...........................................................................................72
8.1. Control I/O Function ....................................................................72
8.1.1. Interface circuit........................................................................72
8.1.2. Timing Chart............................................................................73
8.2. Built-in RS-485 Interface .............................................................74
8.2.1. Connection..............................................................................74
8.2.2. Settings of Parameters............................................................75
8.2.3. Timing Chart............................................................................76
8.2.4. General Data Format...............................................................77
8.2.5. A&D Data Format.................................................................... 78
8.2.6. Address...................................................................................78
8.2.7. Command List.........................................................................79
Page 2 AD-4402
8.3. Built-in Current Loop Output........................................................ 82
8.3.1. Connection..............................................................................82
8.3.2. Communication Modes............................................................82
8.3.3. Data Format............................................................................83
8.4. BCD Output of Option, OP-01.....................................................83
8.5. Relay Output of Option, OP-02....................................................87
8.6. RS-422/485 Interface of Option, OP-03.......................................88
8.7. RS-232C Interface of Option, OP-04...........................................91
8.8. Parallel I/O of Option, OP-05.......................................................92
8.9. Analog Output of Option, OP-07.................................................. 93
9. Maintenance....................................................................................94
9.1.1. Basic Operation.......................................................................94
9.2. Monitor Mode...............................................................................94
9.2.1. Monitoring Control I/O Function...............................................94
9.2.2. Monitoring Built-in RS-485 Interface........................................94
9.2.3. Monitoring Built-in Current Loop Output..................................95
9.2.4. Monitoring A/D Converter........................................................ 95
9.2.5. Monitoring BCD Output of OP-01............................................95
9.2.6. Monitoring Relay Output of OP-02 ..........................................95
9.2.7. Monitoring RS-422/485 Interface of OP-03.............................96
9.2.8. Monitoring RS-232C Interface o f OP-04..................................96
9.2.9. Monitoring Parallel I/O of OP-05..............................................96
9.2.10. Monitoring Analog Output of OP-07 ........................................96
9.3. Test Mode....................................................................................97
9.3.1. Testing Control I/O Function ...................................................97
9.3.2. Testing Built-in RS-485 Interface.............................................97
9.3.3. Testing Built-in Current Loop Output....................................... 97
9.3.4. Testing A/D Converter............................................................. 98
9.3.5. Testing BCD Output of OP-01.................................................98
9.3.6. Testing Relay Output of OP-02 ...............................................98
9.3.7. Testing RS-422/485 Interface of OP-03..................................98
9.3.8. Testing RS-232C Interface of OP-04.......................................99
9.3.9. Testing Parallel I/O of OP-05...................................................99
9.3.10. Testing Analog Output of OP-07 .............................................99
9.4. Initializing Parameters...............................................................100
9.5. Remote Operation.....................................................................102
10. Function List ..................................................................................103
10.1.1. Operation Keys......................................................................103
10.1.2. Outline of the Function List....................................................104
10.2. Referring Parameters................................................................105
10.3. Parameter Settings....................................................................105
10.4. Parameter List...........................................................................106
11. Specifications.................................................................................132
11.1. Dimensions................................................................................135
11.2. Accessories...............................................................................135
12. References....................................................................................136
12.1. Abbreviation...............................................................................136
12.2. ASCII Code for AD-4402...........................................................137
12.3. Index..........................................................................................138
AD-4402 Page 3
1.
1. Compliance
1.1.
1.1.1.
1.1.1. Compliance with FCC rules
1.1.1.1.1.1.
1.1.2.
1.1.2. Compliance with
1.1.2.1.1.2.
Note: The displayed value may be adversely affected under extreme electromagnetic
Compliance with FCC rules
Compliance with FCC rulesCompliance with FCC rules
Please note that this equipment generates, uses and can radiate radio frequency energy. This eq uipment has be en tested and h as been fou nd to comply with the limits of a Class a computing device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide reasonable protection against interference when this equipment is operated in a commercial environment. If this unit is operated in a residential area it may cause some interference and under these circumstances the user would be required to take, at his own expense, whatever measures are necessary to eliminate the interference. (FCC = Federal Communications Commission in the U.S.A.)
This appliance complies with the statutory EMC (Electromagnetic Compatibility) directive 89/336/EEC and t he Low Voltag e Directive 73 /23/EEC for safety of electrical equipment designed for cer tai n vol tag es.
influences.
Compliance
ComplianceCompliance
Compliance with European Directive
Compliance with Compliance with
European Directive
European DirectiveEuropean Directive
Page 4 AD-4402
2.
2. Outline and Feature
2.2.
The AD-4402 is the multi-function weighing indicator for batch weighing and filling weighing. This indicator has control I/O for weighing sequence and options.
Large display This indicator has blue vacuum fluorescent display (VFD). The character height of the main display is 18 mm. Current weighing data, material names, comparison references and accumulation data are displayed in the same time. Material names and recipe
Operation guidance Message that assists current operation are di splayed in the fron t panel, any one could operate the indicator without instruction manuals.
Full weighing sequences The AD-4402 can combine plural mater i al s and the pl ain mi xture function is equipped. Filling nozzle and agitation sequence is equipped. Using the forecast control function, the flow control can be performed that is equivalent to A/D conversion of 1000 times per second.
Outline and Featuressss
Outline and FeatureOutline and Feature
RS-485 interface 32 indicators can be co nnected to a pr ogramma ble controll er or a per sonal compu ter. These protocols are according to public formats.
Options There are built-in options of AC 250 V direct drive relay, serial interface, parallel interface, analog output and etc. There are built-in options of CC-Link, DeviceNet, PROFIBUS. There are three expansion slots to install options.
Check mode during operation The monitor mode can confirm system situation during operation. The test mode can test Input / Output interface. Even if there is not monitor instrument, interface can be confirmed.
Recipes and raw material data stored in the indicator The recipe is described a combination of material codes and weights. The material code is described th e weighi ng sequence param eters for a r aw material .
Water-resistant panel The classification code of the front panel is equivalent to IP-65 of IEC 529 using accessory rubber packing. The "IP-65" code is explained as follows: IP: International Protection. 6: Against ingress of solid foreign objects.
Dust-tight. No ingress of dust.
5: Against ingress of water with harmful effects.
Protected against water jets (no powerful jets). Water projected in jets against the enclosure from any direction shall have no harmful effects.
AD-4402 Page 5
2.1.
2.1. Precaution
2.1.2.1.
Befor any use, confirm the following articles for the safty operation.
Grounding the indicator Ground the indicator certainly. The earth terminal is the rear panel. Separate this earth l ine from other g round line like a motor, inverter or a power source. Unless the indicator is grounded, it may cause to receive an electric shock, be happen operation error or catch fire
Use adaptable power cord Confirm the AC voltage and current of the power cord. If the voltage range of cord is lower than power line voltage, it may cause of a leak or catching fire. Use compression terminals to connect the power cord to the terminal of the rear panel.
Fuse The fuse is installed to prevent the indicator from catching fire. The indicator is equipped many safety circuits. Therefore, the fuse is not broken in normal operation. If the fuse i s broken , do no t repl ace the fuse and co ntac t your local or A&D dealer. This trouble may cause of an electric discharge of thunder.
Precaution
PrecautionPrecaution
Splashing water The indicator is not water-resistant. When the indicator is mou nted to panel with accessory rubber, the front panel is equivalent to IP-65.
Flammability gas Do not install the indicator in any flammable gas.
Radiation of the indicator Space out instruments to radiate heat sufficiently.
Removing the cover Remove the power cord terminals i n the side o f the power source before r emoving the cover to avoid receiving an electric shock. Do not touch the internal circuit within 10 seconds after turning off the indicator because of receiving an electric shock.
Page 6 AD-4402
2.2.
r
r
r
2.2. Front Panel
2.2.2.2.
Main display Weighing data is displayed.
Front Panel
Front PanelFront Panel
Graphic status indicato
Status indicator
Sub-display Materials, Accumulation data, parameters and operation guidance are displayed.
2.2.1.
2.2.1. Keys
2.2.1.2.2.1.
Keys
KeysKeys
Unit indicato
Standby indicato
Pressing this key, the key works as the F1 key. Pressing the SHIFT key and this key, the key works as t he F3 key.
Pressing this key, the key works as the F2 key. Pressing the SHIFT key and this key, the key works as t he F4 key.
The key to select a function of the key.
The key to open the material code or recipe code. Pressing the SHIFT key and this key, the key works as the m ater ial code edit key. Pressing the ENTER key and this key, the key works as the recipe code edit key.
AD-4402 Page 7
The key to move the cursor or scroll the function number. Press and hold the SHIFT key and press the key is used to decrease the code number. The key to select alphabetical keys, upper keys, lower keys or numerical keys.
Alphanumeric keys. The escape key. Pressing and holdi ng the key ab ove three seconds
in normal weighing mode, the display is turned off (standby mode). The ESC key is used to undo the last key and to return to the last mode. The ENTER key for parameter settings. The key to be turned on the display. Pressing the
The key to select net or gross The tare key. The key is displayed the net value that subtracts tare
weight from a current weighing.
The zero key to zero current weighing display.
key and this key, the key works as the menu key.
2.2.2.
2.2.2. Symbols
2.2.2.2.2.2.
Main display Gross or net is displayed.
Sub display Code numbers, operation guidance, graph, comparison parameter
Unit indicator The indicator is displayed that the weighing unit is selected in the
Status indicator The current weighing status is displayed.
Graphic status indicator
STABLE Lighting the sign, the current weighing display is stable.
Symbols
SymbolsSymbols
and other are displayed selectiv el y.
calibration mode. Refer to section "5. Calibra tion".
The classification number is displayed, when occurred an error or informed an alarm.
The current weighing situation is displayed with symbols.
GROSS Displaying the gross data in the main display, the sign is lighted.
Page 8 AD-4402
NET Displaying the net data in the main display, the sign is lighted.
TARE ENT Tare entered.
Storing the net value, the sign is lighted.
HOLD Fixing the main display, the sign is lighted.
CZ Center of zero.
When the gross weigh t is in t h e ce nt er of t he zero p oin t, th e s ign is lighted.
ZR.ERR Zero error.
Error message for zero ing the gross data of the main display.
SQ.ERR The sequence error sign.
A message for weighing sequence error.
ALARM 1 An error sign for over load or emergency stop mode.
ALARM 2 A fatal error sign. Example: The wire form loadcell is broken.
Standby indicator In the standby mod e, al l inter faces ar e turn ed off an d int ernal circ uit
works only.
FULL
Z. BAND
When the gross data exceeds the full limit, the sign is lighted.
The zero band sign. When the gross data is within the ran ge of the ze ro band (arou nd the zero point), the sign is lighted.
F.FLOW
M.FLOW
D.FLOW
FINISH
The full flow g a t e sign.
The medium flow gate sign.
The dribble flow gate sign.
The finish sign.
AD-4402 Page 9
2.3.
,
A
A
A
2.3. Rear Panel
2.3.2.3.
Loadcell terminal. Eight pieces of 350 loadcell can connect in parallel.
Rear Panel
Rear PanelRear Panel
Built-in RS-485 terminal. The possibility: to read weighing data, write parameters, connect 32 units of the indicator with the multi-drop connection
Main power switch
Power cord terminal
C85V
~ 250V
Option slot to connect maximum three options. Example: BCD output, Relay output, Analog output and field bus.
Control I/O to connect to external control units. 11 input terminals, 11 output terminals,
n input comm on terminal n output common term inal
Built-in current loop output Use to connect A&D external monitor
Page 10 AD-4402
Caution
Remove the power cord before installing the indicator and other. Build in the option before installing the indicator.
The indicator can mount on the panel using the slide rail. If the accessory packing rubber is used, the front panel is equivalent to IP-65 of IEC 529.
3.
3. Installation
3.3.
3.1.
3.1. Mounting Indicator
3.1.3.1.
Installation
InstallationInstallation
Mounting Indicator
Mounting IndicatorMounting Indicator
AD-4402 Page 11
Caution
Share the loadcell cable from noise-generating device and these power lines beacuse loadcell signal is sensitive.
We recommend you to use the 6 wire shielded cable to prevent loss of weighing precision. The loadcell cable l eng th i s shor ter than 5 m, y ou may be use a 4 w ir e shi elded c abl e with terminals 1 & 2 shorted (EXC+ & SEN+ shorted) and terminals 3 & 4 shorted (EXC- & SEN- shorted ).
3.2.
3.2. Connecting Loadcell Cable
3.2.3.2.
Connecting Loadcell Cable
Connecting Loadcell CableConnecting Loadcell Cable
Adaptable Compression Terminal Parts
Use the adaptable compression terminal parts to the cables
Loadcell Output Adjustment for Zero Calibration (Zero Point)
When a message "CERR2" is displayed, zero point of zero ca libration is too large. When a message "CERR3" is displayed, zero point of zero calibration is too small. Use the resist er more than 50 kΩ with low (good) temperature coefficient, when adding a resister to adjust the loadcell output to indicator terminals.
In Case of Too Lage Output In Case of Too Small Output
Page 12 AD-4402
3.2.1.
3.2.1. Verifying
3.2.1.3.2.1.
Verifying Loadcell Output and Input Sensitivity
VerifyingVerifying
The input sensitivity of the indicator is 0.3µV/division or more. Adapt to the following inequality, w hen you desig n a weighing instrume nt using the indicator a nd loadc ell(s).
Caution
A change in input voltage sensitivity is equiva lent to a one division change of the display. Select as large an input voltage sensitivity voltage as possible so that the weighing interval becomes stable. Consider the leverage if a lever is used.
Loadcell Output and Input Sensitivity
Loadcell Output and Input Sensitivity Loadcell Output and Input Sensitivity
Weighing instrument using one loadcell.
Weighing instrument using multi-loadcell
Verification Example
Verification Example
Verification ExampleVerification Example
Design: Loadcell N=1
Rated capacity A=750 [kg] Rated output B=3 [mV/V]
Excitation voltage E=5000 [mV] Weighing interval D=0.05 [kg] Weighing capacity 300 [kg]
3.0∗∗≤
3.0
DBE
A
DBE
NA
A:Rated capacity of loadcell [kg] B:Rated output [mV/V] D:Weighing interval [kg] E:Excitation voltage [mV] N:Number of loadcells
05.035000≥=∗
750
regard the in strument as a good design.
3.01
. Therefore,
AD-4402 Page 13
Caution
Gorund the indicator with terminal E to avoid receiving an electric shock and an error due to discharge a static electricity. Share the ground wire from electrical device that generats noise . Do not use unstable power source. Share the power cord form the moter system (as noise-generating device) to avoid operation error.
The power source can use AC 85V to AC 250V with 50 Hz or 60 Hz.
3.3.
3.3. Wiring Power Cord
3.3.3.3.
Wiring Power Cord
Wiring Power CordWiring Power Cord
Adaptable Compression Terminal Parts
Page 14 AD-4402
Caution
Remove the power cord before operation to install the option. Do not touch an inside parts within ten seconds after removing the power cord because you may receive an electric shock. Do not forget to tighten the screw. If the screw is not tightened, it may cause short circuit or an error due to noise.
Three option boards can install in the slots. Initialize the RAM data in accordance with section 9.4. Initializing Parameters.
3.4.
3.4. Installing Options
3.4.3.4.
Installing Options
Installing OptionsInstalling Options
AD-4402 Page 15
4.
4. Basic Operation
4.4.
4.1.
4.1. Key Operation Examples
4.1.4.1.
This section is described the way of key operation.
4.1.1.
4.1.1. Standby Mode
4.1.1.4.1.1.
OFF Press and hold the OFF key above three seconds in the
ON The ON key is used to turn on the indicator.
4.1.2.
4.1.2. Cursor
4.1.2.4.1.2.
Basic Operation
Basic OperationBasic Operation
Key Operation Examples
Key Operation ExamplesKey Operation Examples
Standby Mode
Standby ModeStandby Mode
weighing mode. Then the indicator enters the standby mode and displays standby indicator. In the standby mode, All interface is turned off and internal circuit works only.
Cursor Operation
CursorCursor
Operation
Operation Operation
There is the cursor on a segment (an item) that is turned on and off.
The key is used to move the cursor forward.
SHIFT + Press and hold the SHIFT key and press the key is used to
move the cursor backward.
ENTER The ENTER key is used to enter the selected item. ESC The ESC k ey is used to re turn to t he las t mode and t o und o the
last key operation.
4.1.3.
4.1.3. Inputting
4.1.3.4.1.3.
The character can be input in a current segment (an item) in adaptable mode.
A/a The A/a key is used to change numerical key, upper keys, lower Alphanumerical The alphanumerical keys and the ENTER key is used to enter ENTER The ENTER key is used to specify the alphanumerical data.
ESC The ESC key is used to undo the last key operation and to
Inputting Characters
InputtingInputting
Characters
Characters Characters
keys and alphabetical key. the parameters and to select a code number directly.
return to the last mode.
Page 16 AD-4402
4.1.4.
4.1.4. The Way of Calling a Code
4.1.4.4.1.4.
The Way of Calling a Code
The Way of Calling a CodeThe Way of Calling a Code
In Case of a Material Code:
Step 1 Suppose that is set the function parameter [5qf- 8] to [0] . Step 2 Press the CODE RECALL key in weighing mode. Step 3 Set the number of a material code with the following keys:
The key is used to increase the code number.
SHIFT + Press and hold the SHIFT key and press the key is used to
decrease the code number.
Numerical The numerical keys and the ENTER key is used to select a
code number directly and to enter the parameters.
ENTER The ENTER key is used to specify the number. ESC The ESC key is used to undo the last key and to return to the
last mode.
In Case of a Recipe Code:
Step 1 Suppose that is set the function parameter [5qf- 8] to [1] or [2]. Step 2 Press the CODE RECALL key in weighing mode. Step 3 Set the number of a recipe code with the following keys:
, SHIFT + , Numerical, ENTER, ESC keys
4.1.5.
4.1.5. The Way of Entering a Correction Mode
4.1.5.4.1.5.
The Way of Entering a Correction Mode
The Way of Entering a Correction ModeThe Way of Entering a Correction Mode
In Case of a Material Code:
Step 1 Press and hold the SHIFT key and press the CODE RECALL key in weighing mode. Step 2 Select a number of a material code using the following keys:
, SHIFT + , Numerical, ENTER, ESC keys Step 3 Edit some items of a material code using numerical keys and the ENTER key. Step 4 Press the ESC key to return to weighing mode.
In Case of a Recipe Code:
Step 1 Press and hold the ENTER key and press the CODE RECALL key in weighing mode. Step 2 Select a number of a recipe code using the following keys:
, SHIFT + , Alphanumerical, A/a, ENTER, ESC keys Step 3 Edit some items of a recipe code using alphanumeric keys and the ENTER key. Step 4 Press the ESC key to return to weighing mode.
AD-4402 Page 17
4.1.6.
4.1.6. The Way of Entering Menu
4.1.6.4.1.6.
Step 1 Press and hold the ENTER key and press the key in weighing mode.
Then the first layer of menu is display.
Step 2 Use the following keys in the menu :
Step 3 Press the ESC key to return to weighing mode several times.
The Way of Entering Menu
The Way of Entering MenuThe Way of Entering Menu
, SHIFT, Alphanumerical, A/a , ENTER, ESC keys
Page 18 AD-4402
4.2.
4.2. Status Chart (Mode map
4.2.4.2.
Status Chart (Mode map))))
Status Chart (Mode mapStatus Chart (Mode map
AD-4402 Page 19
5.
5. Calibration
5.5.
The indicator, which is connected loadcell unit, can weigh the "weight" value on the loadcell pan and display its "mass" value. The calibration function is used to adjust the weighing value (displaying value) so that the weighing system can weigh correctly. There are two way of the calibration. The "actual load calibration" uses a rated mass and zero output from the loadcell. The "digital span" inputs arbitrary values (calculated by hand). T hese methods are selected in the calibration procedure. There is a compensation function of the "gravity acceleration correction". This function is used, when a calibrated weighing system is moved to other place. These calibration para meters are stored in the indicator without any power supply.
Calibration
CalibrationCalibration
Common Calibration Items
Unit The "g", "kg" and "t" or "lb" can be selected. Decimal point The decimal point can be selected form "not used" to "four
decimal places". Minimum division The minimum division of the weighing display. Weighing capacity The maximum display of the weighing display.
Items for the "Actual Load Calibration"
Common items Unit, decimal point, minimum di vi sion and w eighing capacity Zero point adjustment A zero point output is used from the loadcell unit. Span adjustment Rated mass is place on the weighing pan and is weighed.
The sensitivity is adjusted. This sensitivity is the same as "
sensitivity " of digital span.
Items for "Digital Span"
Common items Unit, decimal point, minimum di vi sion and w eighing capacity Zero point output The numerical data is input as zero point output of loadcell
unit. Rated capacity The rated capacity of the loadcell is input.
The sensitivity of the loadcell is input.
Caution
When the CAL switch on the A/D board is "DISABLE", any calibration can not perform. Do not perform any calibration during a weighing sequnce operation. Entering calibration mode during a weighing sequnce operation, the weighing sequnce operation is terminated. Calibrate the weighing system, when a weighing sequnce operation does not work The accuracy of the "Digital Span (Calibration without Mass)" is 1/1000. Do not use a "loadcell summing box", the "digital span" is performed. It is necessary that the loadcell sensitivity is exactly known, if the "digital span" is used.
Page 20 AD-4402
5.1.
5.1. Actual Load Calibration (using Mass)
5.1.5.1.
ESC key If you want to return to the weighing mode during the
ENTER key When the key is pressed, the procedure stores a current
Step 1 Press and hold the ENTER key and press the key to display the
menu in a weighing mode.
Step 2 Press the key twice to select the menu CAL.
Press the ENTER key to enter the calibration mode. Step 3 Press the ENTER key to enter the menu CAL. Step 4 Select a unit using the num erical keys and press the ENTER key
to store it. Step 5 Select a decimal point using the numerical keys and press the
ENTER key to store it.
Actual Load Calibration (using Mass)
Actual Load Calibration (using Mass)Actual Load Calibration (using Mass)
calibration mode, press the ESC key anytime. And it has effect until the last displayed parameter. Example: zero adjustment only, etc.
parameter and proceeds to next step.
Step 6 Select a minimum division using the numer ical keys and press the
ENTER key to store it. Step 7 Select a weighing capcity usi ng the n um eri cal key s and press the
ENTER key to store it. Step 8 Perform the zero point adjustment.
Place nothing on the weighing pan and press the ENTER key to
store it after the STABLE indicator is displayed.
Whether the STABLE indicator is displayed or not, if you want to
store it, wait for ten seconds and press the ENTER key. Step 9 Specify a total mass value to place on the weig hing pan using the
numerical keys and press the ENTER key to store it. Step10 Place the specifyed mass on the weighing pan and press the
ENTER key to store it after the STABLE indicator is displayed.
Whether the STABLE indicator is displayed or not, if you want to
store it, wait for ten seconds and press the ENTER key. Step11 Press the ESC key to return the weighing mode.
AD-4402 Page 21
5.2.
5.2. Digital Span (Calibration without Mass)
5.2.5.2.
ESC key If you want to return to the weighing mode during the
ENTER key When the key is pressed, the procedure stores a current
Step 1 Press and hold the ENTER key and press the key to display the
menu in a weighing mode. Step 2 Press the key twice to select the menu CAL.
Press the ENTER key to enter the calibration mode. Step 3 Press the ENTER key to enter the menu CAL. Step 4 Select a unit using the num erical keys and press the ENTER key
to store it. Step 5 Select a decimal point using the numerical keys and press the
ENTER key to store it.
Digital Span (Calibration without Mass)
Digital Span (Calibration without Mass)Digital Span (Calibration without Mass)
calibration mode, press the ESC key anytime. And it has effect until the last displayed parameter. Example: zero adjustment only, etc.
parameter and proceeds to next step.
Step 6 Select a minimum division using the numer ical keys and press the
ENTER key to store it. Step 7 Select a weighing capcity usi ng the n um eri cal key s and press the
ENTER key to store it. Step 8 Press the F1 key to proceed to the digital span procedure. Step 9 Input the zero point value usi ng the nu mer i cal keys and press the
ENTER key to store it. Step10 Input the rated capacity of a lo adcell using the nu merical key s and
press the ENTER key to store it. Step11 Input the sensitivity of the loadcell in the unit of mV/V using the
numerical keys and press the ENTER key to store it. Step12 Press the ESC key to return the weighing mode.
Advise The digital span can be used for trimming of the actual l oad calibration using mass.
Page 22 AD-4402
5.3.
5.3. Gravity Acceleration Correction
5.3.5.3.
Gravity Acceleration Correction
Gravity Acceleration CorrectionGravity Acceleration Correction
The function compensates the weighing error due to the difference of gravity acceleration.
G1 The place where the weighing system is calibrated. G2 The place where the weighing system is used.
ESC key If you want to return to the weighing mode during the
calibration mode, press the ESC key anytime.
ENTER key When the key is pressed, the procedure stores a current
parameter and proceeds to next step.
Step 1 Press and hold the ENTER key and press the key to display the menu in a
weighing mode. Step 2 Press the key twice to select the menu CAL. Press the ENTER key to enter the
calibration mode. Step 3 Select the menu G with the key. Press the ENTER key to enter it. Step 4 Input the gravity acceleration at G1 using the num erical keys and press the
ENTER key to store it. Step 4 Input the gravity acceleration at G2 using the num erical keys and press the
ENTER key to store it. Step 5 Press the ESC key to return the weighing mode.
5.3.1.
5.3.1. Gravity Acceleration Reference
5.3.1.5.3.1.
Amsterdam 9.813 m/s2Manila 9.784 m/s Athens 9.800 m/s2Melbourne 9.800 m/s Auckland NZ 9.799 m/s2Mexico City 9.779 m/s Bangkok 9.783 m/s2Milan 9.806 m/s Birmingham 9.813 m/s2New York 9.802 m/s Brussels 9.811 m/s2Oslo 9.819 m/s Buenos Aires 9.797 m/s2Ottawa 9.806 m/s Calcutta 9.788 m/s2Paris 9.809 m/s Chicago 9.803 m/s2Rio de Janeiro 9.788 m/s Copenhagen 9.815 m/s2Rome 9.803 m/s Cyprus 9.797 m/s2San Francisco 9.800 m/s Djakarta 9.781 m/s2Singapore 9.781 m/s Frankfurt 9.810 m/s2Stockholm 9.818 m/s Glasgow 9.816 m/s2Sydney 9.797 m/s Havana 9.788 m/s2Tainan 9.788 m/s Helsinki 9.819 m/s2Taipei 9.790 m/s Kuwait 9.793 m/s2Tokyo 9.798 m/s Lisbon 9.801 m/s2Vancouver, BC 9.809 m/s London (Greenwich) 9.812 m/s2Washington DC 9.801 m/s Los Angeles 9.796 m/s2Wellington NZ 9.803 m/s Madrid 9.800 m/s2Zurich 9.807 m/s
Gravity Acceleration Reference
Gravity Acceleration ReferenceGravity Acceleration Reference
2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2
AD-4402 Page 23
5.4.
5.4. Calibration Error
5.4.5.4.
Error Code Treatment and Situation
CERR1
CERR2
CERR3
CERR4
CERR5
CERR6
CERR7
CERR8
CERR9
CERR10
CERR11
Calibration Error
Calibration ErrorCalibration Error
Resolution (Weighing ca paci ty / minimum division) is exceeds th e l i mita ti on. Increase minimum division or decrease weighing capacity. The initial load (no load output) is larger than 2mV/V. Confirm the loadcell cable. Negative loadcell output value. Check wiring. Confirm the loadcell cable. Mass value exceeds the we ighing capacity. Use a mass within the weighing capacity. (Decrease mass value) Mass value is too light for the calibration. Increase mass value. The loadcell output to be equivalent to minimum division is too small. Use more rough minimum division. The polarity of loadcell output is inversed. Confirm the loadcell cable. The mass value of the weighing capacity exceeds 3.2 mV/V. Confirm the mass and weighing capacity. Gravity acceleration is out of range. Correct the value within the range of 9.770 ~ 9.835 m/s Zero output of loadcell unit is out of range. Trim the zero output within 0.0 ~ 2.0 mV/V. The loadcell output to be equivalent to minimum division is out of range. Trim the output within 0.0 ~ 3.2 mV/V.
2
.
Page 24 AD-4402
6.
6. Applications
6.6.
6.1.
6.1. Hopper Scale with Material Code
6.1.6.1.
In the section, ap plicatio ns are explained according to the right hopper scale that performs batch weighing using a material code.
An application is explained mixture of materials using a recipe code.
The foundation of hopper scale design is explained.
6.1.1.
6.1.1. Definition of Material Code
6.1.1.6.1.1.
Applications
ApplicationsApplications
Hopper Scale with Material Code
Hopper Scale with Material CodeHopper Scale with Material Code
Definition of Material Code
Definition of Material CodeDefinition of Material Code
The material code is necessary to store the details before use. And the code is called with code number in a weighing. The material code is used in the procedure that performs batch weighing or loss-in weigh. As the result of t he procedure , a constant w eig ht of the materi al, is cal led "full " or "full filling", can be got. The material code consists of some index number (name) and some comparison values to get a constant weight of the material. The AD-4402 can store a hundred kinds of material codes.
6.1.2.
6.1.2. Recalling
6.1.2.6.1.2.
The following steps are the explanation to recall the material code stored in the indicator. Suppose that the recipe code is not used. (The menu setting] - [Sequence] - [Basic] - [Recipe mode]
Caution
The material code can be recalled during the last weighing. But the code effects after the batch finish (after finishing the last weighing).
Step 1 Press the CODE RECALL key.
Recalling MMMMaterial
RecallingRecalling
aterial CCCCode
aterial aterial
ode
odeode
is set to [5q f- 8] [0] )
[Function] - [Function
Then the material code blinks. Step 2 Enter the material code using numerical keys
The details of the material code are displayed in the sub-display. Step 3 Press the ENTER key to decide the code.
AD-4402 Page 25
6.1.3.
6.1.3. Editing Principle Parameters of Material Code
6.1.3.6.1.3.
You can edit the parameters of target weight, free fall and etc. displayed on the sub­display during a weighing. And items of sub-display can be selected at the menu [Function] - [Function setting] - [General] - [Sub-display].
Editing Principle Parameters of Material Code
Editing Principle Parameters of Material CodeEditing Principle Parameters of Material Code
Caution
If the flash memory is selected for memory backup (0tHf-11 sequential weighing is stopped.
Step 1 Press and hold the SHIFT key and press the CODE RECALL key.
Then the material code blinks. Step 2 Enter the material code using numerical keys
The details of the material code are displayed in the sub-display. Step 3 Press the ENTER key to decide the code. Step 4 Select a parameter using the key on the sub-display. Step 5 Enter the parameter using numerical keys and press the ENTER key to store it. Step 6 If you continue the change, proceed step 4 and 5. Step 7 If you want to finish the ch ange, press the ESC key to return to weighing mode
0tHf-11), a current
0tHf-110tHf-11
several times.
6.1.4.
6.1.4. Referr
6.1.4.6.1.4.
You can refer to next material code in the sequential mode that uses plural material codes. Suppose that the recipe code is not used. (The menu setting] - [Sequence] - [Basic] - [Recipe mode]
Step 1 Press the CODE RECALL key.
Then principle parameters of the next material code are displayed in the sub-
display. Step 2 Press the ESC key to return to current mode.
Referring
ReferrReferr
ing next Material Code
inging
next Material Code
next Material Code next Material Code
is set to [5q f- 8] [0])
[Function] - [Function
Page 26 AD-4402
6.1.5.
6.1.5. Editing Full Parameters of Material Code
6.1.5.6.1.5.
Editing Full Parameters of Material Code
Editing Full Parameters of Material CodeEditing Full Parameters of Material Code
A material code consists of the following parameters.
Name
Material Code Material name Material Hopper No. Final Free Fall Preliminary Optional Preliminary Over Under Zero Band Full Tare Supplementary Flow Open Timer Supplementary Flow Close Timer Automatic Free Fall Range Initial Dribble Flow Initial Medium Flow Total Weight Total Counts
Display Name Display
Symbol
Code Mat Name grain Mat Hopper Hopper 1 Final Final 10.00 kg Free Fall FFall 0.01 kg Preliminary Plm 1.00 kg OP.Preliminary OPPlm 2.00 kg Over Over 0.10 kg Undr Undr 0.10 kg Zero Band 0Band 0.02 kg Full Full 0.05 kg Tare Tare 5.00 kg SF open timer SFOT 0.00 s SF close SFCT 0.00 s AFFC range AFFC 0.00 kg Initial DF IDF 0.00 kg Initial MF IMF 0.00 kg Tot Tot 10.00 kg Tot# Tot# 1
Code
Display Example
11
Memory
Backed up RAM ( factory setting) or flash memory
Backed up RAM
These parameters are stored in backup memory without power supply. Refer to the backup method [0tHf-11] of the function list.
Caution
If the flash memory is selected for memory backup [0tHf-11 sequential weighing is stopped.
0tHf-11], a current
0tHf-110tHf-11
Edit Material Code
Step 1 Press and hold the ENTER key and press the key.
Then menu MatEdit blinks. Step 2 Press the ENTER key to enter the material code edit. Then menu Edit blinks. Step 3 Press the ENTER key to enter menu edit. Step 4 Select the material code using numerical keys and press the ENTER. Step 5 Enter the material name using alpanumerical keys and press the ENTER key. Step 5 Edit other parameters using numerical keys, ENTER key and key. Step 6 If you want to finish the ch ange, press the ESC key to return to weighing mode
several times.
AD-4402 Page 27
Search Material Code
Use this menu to search blank material code.
Step 1 Press and hold the ENTER key and press the key.
Then menu MatEdit blinks. Step 2 Press the key to select menu Search. And press the ENTER key. Step 3 Then the message is displayed. Step 4 Press the ENTER key to prceed next step.
Then the result is displayed. Step 5 Press the ESC key to return to weighing mode several times.
Delete Material Code
The parameter of the material code can be reset in the following menu.
Total value Setpoints Total of a material code All material code All total
Example of Deleting Total Value
Step 1 Press and hold the ENTER key and press the key.
Then menu MatEdit blinks. Step 2 Press the key to select menu Delete. And press the ENTER key. Step 3 Select menu Total using the key. And press the ENTER key. Step 4 Enter the material code using numerical keys and press the ENTER key. Step 5 Press the ESC key to return to weighing mode several times.
Copy Material Code
The parameters of material code are copied. This copy includes a total weight value and times of accumulation.
Step 1 Press and hold the ENTER key and press the
Then menu Step 2 Press the key to select menu Copy. And press the ENTER key. Step 4 Specify a original code number using numerical keys and press the ENTER key. Step 5 Specify a duplicated code num ber using numerical keys and press the ENTER key.
MatEdit blinks.
key.
Step 6 Press the ESC key to return to weighing mode several times.
Page 28 AD-4402
Loading...
+ 114 hidden pages