A&D BM-500 User Manual

BM Series Additional Information
Overview (Supported software versions: P-1.20 and later)
This supplementary manual describes functions that have been added to the BM series.
Temperature measurement during weighing:
The balance has an embedded temperature sensor.
The temperature of the setup environment of the balance can be displayed and output.
Measurement precision is ±1.5 °C (between 5 and 40 °C).
External control of the static charge eliminator (Ion function):
By connecting an AD-8922A remote controller or using the external input on the rear of
the balance, the static charge eliminator can be switched on and off without needing to
press a button on the balance.
In the weighing mode, press the SELECT key.
The display shows each of the following items for 5
seconds in the following order: time, date, and
temperature. After all items have been displayed,
the display returns to the weighing mode.
To proceed to the next item while an item is
displayed, press the SELECT key.
*The unit for display is °C.
Outputting the Temperature
To add the temperature to the weighing data output, select the “
Displaying the Temperature
change the “
Temperature output setting (added after the setting “ar-d)
5-tp” (add temperature) setting.
Weighing mode
Press SELECT .
Time display
Wait 5 seconds or press SELECT
Date display
Wait 5 seconds or press SELECT
Temperature display
Wait 5 seconds or press SELECT
Weighing mode
dout” (data output) settings and
Not used
Used
Do not output temperature
Output temperature
New
dout
Data output
ar-d
Zero after output
5-tpP
Add temperature
off
on
off
on
Factory setting
Data format for temperature output
Temperature dout 5-tpP on *The unit for output is “C”.
+ 0 2 3 . 4 C
1WMPD4002332
1
CRL
1
F
External Control of the Ion Function (Static Charge Eliminator)
To assign the Ion function to the MODE key of the AD-8922A remote controller, select the
ionfnc” (Static elimination) settings and change the “8922” (AD-8922A control) setting.
To assign the Ion function when using both the AX-T-314A-S external key plug and AX-SW128
foot switch (both sold separately), select the “ the “e-5” (external input control) setting.
ionfnc” (Static elimination) settings and change
Ion function control settings (added after the setting “ion”)
0 3 seconds
ion
Neutralization time
1 10 seconds
2 Manual operation (max. 10 min.)
ionfnc
Static
eliminator
8922
AD-8922A control
e-5S
External input control
(See the figure below
for terminals)
off Do not control the Ion function with the
AD-8922A
on Assign the Ion function to the MODE key of
the AD-8922A.
0 1. PRINT key, 2. RE-ZERO key
1 1. PRINT key, 2. ION key 2 1. ION key, 2. RE-ZERO key
Factory setting
New
*External key plug
AX-T-314A-S
(Sold separately)
Terminal 1
Terminal 2
GND
2