18.2. Index .................................................................................................................................. 87
2 BM series
1. Introduction
This manual describes how the BM series balances work and how to get the most out of them in
terms of performance. Read this manual thoroughly before using the balance and keep it at hand
for future reference.
About This Manual
This manual consists of the following five parts:
Basic operation
Adapting to the environment
Selecting functions
Interface and communication
Maintenance
..............................
..........................
..................................
...........
..........
Describes precautions, the balance's construction and basic
operation.
Describes response (and stability) adjustment to adapt to the
environment where there is vibration or drafts, the way to
maintain weighing precision in a variation of ambient
temperature, calibration and calibration test.
Describes functions of the balance.
Describes the serial interface used for communicating with a
computer that requests weighing data and controls the
balance, and for use with a printer
Describes maintenance, error codes, troubleshooting,
specifications and options.
BM series 3
1.1. Features
A built-in DC static eliminator can eliminate static electricity from the weighing sample before the
measurement, reducing weighing error.
Each electrode unit of the eliminator is designed to be removed, cleaned and replaced.
Automatic self calibration, using the internal mass, adapting to changes to temperature changes.
Response adjustment adapting to drafts and/or vibration automatically.
Memory function to store weighing data and calibration data.
When only weighing data is stored, a maximum of 200 data can be stored.
Interval mode to store the weighing data periodically.
Good laboratory practice (GLP) data can be output using the RS-232C serial interface.
A built-in clock and calendar that can add the time and date to the output data.
Underhook, for measuring density and weighing magnetic materials.
Multiple weighing units with most of the common units used around the world.
Gram, Milligram, Counting mode, Percent mode, Ounce (Avoir), Troy Ounce, Metric carat, Momme,
Pennyweight, Grain (UK), Tael (preset in the factory), Tola (India), Messghal, Density mode
BM-22 is equipped with a smart range function to weigh with the precision range (a higher
resolution) when zeroing a tare weight within the weighing capacity.
The test tube holder is included in the accessories of BM-20 and BM-22.
Density mode for calculating the density of a solid.
BM series balances are equipped with an RS-232C serial interface and a USB interface to
communicate with a computer. When printing data using the RS-232C interface, by using the USB
interface, the data can be transmitted to a computer at the same time. Communications between
the BM series balance using the RS-232C interface and a Windows computer using the Windows
communication tools software (WinCT) make building a system very easy.
Connecting the USB interface of the balance, by cable to a Windows computer, allows
transmission of the weighing data to Excel or Word.
When multiple balances have the BM-08 Ethernet interface installed in place of the USB interface
and connected to a LAN, data can be acquired from each of them using the WinCT-Plus software.
The weighing data can be stored by connecting the accessory data logger (AD-1688) directly,
when not using a computer.
4 BM series
1.2. Compliance
1.2.1. Compliance With FCC Rules
Please note that this equipment generates, uses and can radiate radio frequency energy. This
equipment has been tested and has been found to comply with the limits of a Class A computing
device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide
reasonable protection against interference when equipment is operated in a commercial
environment. If this unit is operated in a residential area, it may cause some interference and
under these circumstances the user would be required to take, at his own expense, whatever
measures are necessary to eliminate the interference.
(FCC = Federal Communications Commission in the U.S.A.)
1.2.2. Compliance With EMC Directives
This device features radio interference suppression and safety regulation in compliance
with the following Council Directives
Council directive 89/336/EEC EN61326 EMC directive
Council directive 73/23/EEC EN60950 Safety of Information Technology Equipment
The CE mark is an official mandatory European marking.
Please note that any electronic product must comply with local laws and regulations when sold or
used anywhere outside Europe.
BM series 5
2. Unpacking The Balance
Unpack the balance carefully. Keep the packing material to be used for transporting the balance
in the future. See the illustrations to confirm that everything is contained.
Parts For Each Products
BM-22, BM-20
Small weighing Large weighing
pan for filter pan for filter
Weighing pan Pan support
Test tube holder
Small fine range ring
Fine range ring
Tweezers
Dust plate AX-MX-36
φ15, 0.8 φ12, 0.3 φ8, 0.05
Large Medium Small
Round aluminum analytical pans (
each 10 pieces
BM-252
Weighing pan Pan support
Fine range ring
Breeze break ring
Dust plate
)
BM-500, BM-300,
BM-200
Weighing pan
Pan support
Breeze break ring
Dust plate
Common Parts
AC adapter AC adapter labels
Note
Please confirm that the AC adapter type is
correct for your local voltage and receptacle type.
USB cable
6 BM series
Position of placing
AC adapter labels
Display cover
Windows communication
software (WinCT)
Separation plate
If unnecessary, remove it.
D-sub 9pins
Data logger
AD-1688
The weighing data can be
stored when connecting
the data logger directly.
Tweezers
AD-1689
Tweezers for handling a weight.
Use it to avoid temperature changes
due to having your hand in the
weighing chamber, thus to perform
accurate calibration.
J
The DC Static eliminator
Electrode units of the
DC static eliminator
Separation plate
If unnecessary, remove it.
Leveling foot
oint handle
When jointing the right (left) slide
door handle, the right (left) glass
door can be opened or closed
with the left (right) joint handle.
Bubble spirit level
Glass door handle
The handle for the glass door.
The handle can be joined to the joint handle
USB interface
AC adaptor jack
External key jack
Cap the terminal with accessory
rubber cap when it is not used.
Serial interface RS-232C
D-sub 9 pins
Anti theft device
Grounding terminal
LED of the static eliminator
Glass door handle
Neutralization
area
Example of BM-20
Weighing
chamber
Display
Keys
Display cover
BM series 7
3. Installing The Balance And Precautions
3.1. Installing The Balance
Step
1 Refer to "3.2. Precautions Before Use (Environmental Condition And Preparations)"
concerning the place to install the balance. Place the balance on a firm weighing table.
Step
2 Assemble the weighing pan and other parts in the weighing chamber, in accordance with the
product type and use.
BM-500, BM-300, BM-200 BM-252
Weighing pan
Pan support
Breeze break ring
Dust plate
BM-22, BM-20
Test tube holder
Large weighing pan for filter
Small weighing pan for filter
Weighing pan
Pan support
Small fine range ring
Fine range ring
Dust plate
Use the fine range ring to avoid errors caused by drafts
when weighing with a minimum display of 0.01 mg.
Fine range ring
Choose the weighing pan to
adapt to the filter size.
8 BM series
Use the holder for
micro tube or test
tube of φ12 mm
or less.
Step
k
A
k
3 If the separation plate is not necessary, remove
the screw and take it off. Refer to next page
"The Separation Plate".
Step
4 Adjust the level of the balance using the leveling feet.
Step
5 Confirm that the AC adaptor type is correct for your
local voltage and power receptacle type.
Step
6 Connect the specified AC adapter to the balance.
Warm up the balance for at least one hour with
nothing on the weighing pan.
Step
7 Confirm correct weighing. If necessary, calibrate
the balance. Refer to "8. Calibration".
BM-200, BM-300,
BM-500, BM-252
Leveling foot
Bubble spirit level
Note Please confirm that the AC adapter
type is correct for your local voltage
and receptacle type.
Example of outlet
Separation plate
Screw
Hoo
BM-252
BM-22, BM-20
C adaptor jac
Grounding terminal
BM series 9
The Separation Plate
k
3
3
k
Note: Take care that the separation plate is not broken when operating it.
Removing the Separation Plate
Step
1 Support the separation plate and
remove the screw from the plate.
Step
2 Lift the plate and remove hooks.
Step
3 Rotate the plate on the axis of the back edge.
Rotate the plate on the axis of the front edge.
Step
4 Remove the plate from the chamber.
Screw
Hoo
Step
1
Attaching the Separation Plate
Use the arrows in reverse to attach the plate.
Step
5 Insert the plate into the chamber. (Reversed
Step
6 Rotate the plate on the axis of the front edge.
Level the plate. (Reversed
Step
3)
Step
4)
Step
Step
Step
Step
1
2
Step
4
Step
7 Insert hooks and hook the plate. (Reversed
Step
8 Install and fix the screw. (Reversed
Step
1)
Step
2)
Hoo
Separation plate
3.2. Precautions Before Use (Environmental Condition And Preparations)
To ensure that you get the most from your balance, please try to follow these conditions as closely as
possible. Consider these conditions for BM-20 and BM-22 that are sensitive instrument specially.
The best operating temperature is about 20°C / 68°F at about 50% Relative Humidity.
The weighing room should be free of dust.
The weighing table should be solid and free from vibration, drafts (such as frequently opening
doors or windows) and as level as possible. We recommend to use the anti-vibration table
(AD-1670) and remote controller (AD-8922A) for BM-20 and BM-22.
Do not install the balance where it will be subject to vibration. Corners of rooms are best.
Do not install the balance near a heater, air conditioner, or in a breeze.
Do not install the balance in direct sunlight and excessive temperature changes.
Do not use the balance near other equipment which produces magnetic fields.
Adjust the level of the balance using the leveling feet.
Please warm-up the balance for at least one hour. Plug-in the AC adapter as usual.
Calibrate the balance before using and after moving it to another location.
Ensure a stable power source when using the AC adapter.
Do not place or use the balance where there is flammable or corrosive gas present.
10 BM series
3.3. Cautions During Use (For Precision Weighing)
Note the following items to get accurate weighing data.
Discharge static electricity from the weighing material. When
weighing sample (plastics, insulator, etc.) could have a static
charge, the weight value is influenced. Ground the balance, and
Eliminate the static electricity using the built-in static eliminator.
Try to keep the ambient humidity at or above 45%RH in the room.
Use a metal shield case.
Wipe a charged material (plastic sample etc.) with a damp cloth.
This balance uses a strong magnet as part of the balance assembly,
so please use caution when weighing magnetic materials. If there is
a problem, use the underhook (on the bottom of the balance) to
suspend the material away from the influence of the magnet.
Eliminate temperature differences between the weighed sample and
the environment. When a sample is warmer (cooler) than the ambient
temperature, the sample will appear lighter (heavier) than the true
mass. This error is due to the rising (falling) draft next the sample. If
you touch the sample, the same type error will occur. Do not touch the
sample directly with your hand. Use tweezers or other tools.
Charged material
Metal case
Grounding
Magnet material
Draft
40°C 20°C
Weighing pan
Do not drop things upon the weighing pan, or place a weight
beyond the range of the balance on the weighing pan.
Make each weighing gently and quickly to avoid errors due to
changes in the environmental conditions.
We recommend that the fine range ring and separation plate are
used to avoid weighing error caused by drafts when BM-252 displays
a weighing value in unit of 0.01 mg (minimum display: 0.01 mg).
We recommend that the small fine range ring, fine range ring
and separation plate are used to avoid weighing error caused by
drafts when BM-22 and BM-20 display a weighing value in unit of
0.01 mg and 0.001 mg (minimum display: 0.01 mg and 0.001 mg).
Take into consideration the affect of air buoyancy on a sample
when more accuracy is required.
Do not use a sharp instrument (such as a pencil or ball point pen) to
press the keys, use your finger only.
Press the RE-ZERO key before each weighing to prevent possible
errors.
Shock
Fine range ring
BM-252
Small fine range ring
Fine range ring
BM-20
BM-22
Avoid foreign matter (dust, liquid or metal fragments) that could get
inside the balance.
Operate your balance gently. Shorten the operation time as much
as possible (Opening and closing door, putting and removing
sample). Use a pair of tweezers to avoid temperature changes due
to heat from inserting your hand into the weighing chamber.
BM series 11
3.4. Cautions After Use (Management Of The Balance)
Avoid mechanical shock to your balance.
Do not disassemble the balance. Contact your local A&D dealer if your balance needs service
or repair.
Do not use solvents to clean the balance. For best cleaning, wipe with a dry lint free cloth or a
lint free cloth that is moistened with warm water and a mild detergent.
Avoid foreign matter (dust, liquid or metal fragments) that could get inside the balance.
3.5. Cautions For Power Supply
Do not remove the AC adapter while the internal mass is in motion, for example, right after the
AC adapter is connected, or during calibration using the internal mass.
If the AC adapter is removed under the conditions described above, the internal mass will be
left unsecured, that may cause mechanical damage when the balance is moved.
Before removing the AC adapter, confirm that zero is displayed in the weighing mode, then
press the ON:OFF key.
When the AC adapter is connected, the balance is in the standby mode if the standby indicator
is on. This is a normal state and does not harm the balance. For accurate weighing, we
recommend that you always plug in your balance so it can warm up.
Connect the AC adapter for BM-22 and BM-20 normally.
12 BM series
4. Display Symbols And Key Operation
r
A
P
r
y
r
Display
Processing indicator
Weighing in liquid of density mode
Stabilization indicator
Standby indicator
of power suppl
Weighing data or stored data
The current data numbe
Weighing in air of density mode
rior notice indicator of
automatic self calibration
The interval memory function is used to store periodical weighing data. Refer to "12. Data Memory".
Blinking indicators
The amount of stored data with data memory function
Humidity (%)
Response indicator for approx. 30
seconds when weighing starts.
Right and left glass door indicato
Air pressure (hPa)
Standby indicator of
interval memory function
Indicator of static eliminato
Units.
Refer to "5. WeighingUnits".
ctive indicator of
interval memory function
Key Operation
Press and release the key immediately or "Click the key"
Key When pressed and released When pressed and held
The key to turn the display ON and OFF. The standby indicator is displayed when the
display is turned off. The weighing mode is enabled when the display is turned on. This key
is available anytime. Pressing the key during operation will interrupt the operation and turn
the display OFF.
The key to perform calibration using the
internal mass.
The key to switch the preset weighing units
stored in the function table.
Refer to "5. Weighing Units".
In the weighing mode, the key to turn the
minimum weighing value ON and OFF.
In the counting or percent mode, the key
to enter the sampling mode.
The key to outputs the weighing data to a
printer or personal computer (or store it in
memory) depending on the function table
settings. (Factory setting = output)
The key to display other items of the
calibration menu.
The key to perform automatic response
adjustment.
The key to enter the function table mode or
repeatability test mode.
Refer to "10. Function Table" or "6.7. Repeatability Test".
No function at factory setting.
By changing the function table:
"Title block" and "End block" for GLP
report are output.
The data memory menu is displayed.
The key to set the display to zero.
The key to turn the static eliminator ON and OFF.
The key to monitor the date and time, and the environmental sensors (temperature, humidity, air
pressure), for several seconds. Refer to "6.6. Monitoring Clock and Environmental Sensors".
BM series 13
5. Weighing Units
5.1. Units
The units and weighing modes can be selected and stored in the function table as described in
"5.2.Storing The Active Units".
The sequence of displaying them can be arranged to fit the frequency of use.
They are stored and are maintained in non-volatile memory, even if the AC adapter is removed.
If the law in your area permits, you may use all of the units. You can disable the units that you
don't regularly use. And you are able to turn them back on.
If a weighing mode (or unit of mass) has been turned off, that mode or unit will be missing in the
sequence. Tael has four varieties, one of which can be selected and installed at the factory.
Press the MODE key to select a unit or mode for weighing.
For details about the units and modes, see the table below:
Name (unit, mode) Abbreviation Display unit Conversion factor
Gram g
Milligram mg 0.001 g
Counting mode PC -
Percent mode % -
Ounce (Avoir) oz 28.349523125 g
Troy Ounce ozt 31.1034768 g
Metric Carat ct 0.2 g
Momme mom 3.75 g
Pennyweight dwt 1.55517384 g
Grain (UK) GN 0.06479891 g
Tael (HK general, Singapore) 37.7994 g
Tael (HK jewelry) 37.429 g
tl
Tael (Taiwan) 37.5 g
Tael (China)
Tola (India) t 11.6638038 g
Messghal m 4.6875 g
1 g
31.25 g
Density mode DS Refer to "14. Density Measurement"
Density mode
To use the density mode, it must be stored in the function table as described on page 17.
For details about this mode, refer to "14. Density Measurement".
To select this mode, press the MODE key until the processing indictor blinks with the unit
"g" displayed.
14 BM series
Capacity And Minimum Display For BM-252, BM-500, BM-300, BM-200
The tables below indicate the weighing capacity and the minimum display for each balance.
Unit
Gram 520 320 220 0.0001
Milligram 520000 320000 220000 0.1
Ounce (Avoir) 18.34 11.29 7.76 0.00001
Troy Ounce 16.72 10.29 7.07 0.00001
Metric Carat 2600 1600 1100 0.001
Momme 138.7 85.3 58.7 0.0001
Pennyweight 334.4 205.8 141.5 0.0001
Grain (UK) 8024 4938 3395 0.002
Tael (HK general, Singapore) 13.76 8.47 5.82 0.00001
Tael (HK jewelry) 13.89 8.55 5.88 0.00001
Tael (Taiwan) 13.87 8.53 5.87 0.00001
Tael (China) 16.64 10.24 7.04 0.00001
Tola (India) 44.58 27.44 18.86 0.00001
Messghal 110.9 68.3 46.9 0.0001
BM-500 BM-300 BM-200
CapacityMinimum display
Unit
Capacity Minimum display
Gram 250 0.00001
Milligram 250000 0.01
Ounce (Avoir) 8.82 0.000001
Troy Ounce 8.03 0.000001
Metric Carat 1250 0.0001
Momme 66.67 0.00001
Pennyweight 160.7 0.00001
Grain (UK) 3858 0.0002
Tael (HK general, Singapore) 6.61 0.000001
Tael (HK jewelry) 6.68 0.000001
Tael (Taiwan) 6.67 0.000001
Tael (China) 8.00 0.000001
Tola (India) 21.43 0.000001
BM-252
Messghal 53.3 0.00001
BM series 15
Capacity And Minimum Display For BM-22, BM-20
BM-22 is equipped the precision range and standard range of the smart range function.
BM-20
Unit
Gram 22 0.000001
Milligram 22000 0.001
Ounce (Avoir) 0.776 0.0000001
Troy Ounce 0.707 0.0000001
Metric Carat 110 0.00001
Momme 5.87 0.000001
Pennyweight 14.15 0.000001
Grain (UK) 339.5 0.00002
Tael (HK general, Singapore) 0.582 0.0000001
Tael (HK jewelry) 0.588 0.0000001
Tael (Taiwan) 0.587 0.0000001
Tael (China) 0.704 0.0000001
Tola (India) 1.886 0.0000001
Messghal 4.69 0.000001
Precision range
Capacity Minimum display
BM-22
Unit
Gram 5.1 0.000001 22 0.00001
Milligram 5100 0.001 22000 0.01
Ounce (Avoir) 0.180 0.0000001 0.776 0.000001
Troy Ounce 0.164 0.0000001 0.707 0.000001
Metric Carat 25.5 0.00001 110 0.0001
Momme 1.36 0.000001 5.87 0.00001
Pennyweight 3.28 0.000001 14.15 0.00001
Grain (UK) 78.71 0.00002 339.5 0.0002
Tael (HK general, Singapore) 0.135 0.0000001 0.582 0.000001
Tael (HK jewelry) 0.136 0.0000001 0.588 0.000001
Tael (Taiwan) 0.136 0.0000001 0.587 0.000001
Tael (China) 0.163 0.0000001 0.704 0.000001
Precision rangeStandard range
Capacity Minimum displayCapacity Minimum display
Tola (India) 0.437 0.0000001 1.886 0.000001
Messghal 1.09 0.000001 4.69 0.00001
16 BM series
5.2. Storing Active Units
t
s
y
s
The units and modes can be selected and stored in the function table.
The sequence of displaying them can be arranged to fit the frequency of use.
The units stored are maintained in non-volatile memory, even if the AC adapter is removed.
1 Press and hold the RANGE key untilba5fnc of
the function table is displayed in the weighing mode,
then release the key.
2 Press the RANGE key several times to displayUnit.
3 Press the PRINT key to enter the unit selection mode.
4 Specify a unit or mode in the order to be displayed
using the following keys.
RANGE key
RE-ZERO key
Examples
Gram g
Milligram mg
.........
To display the units sequentially.
......
To specify a unit or mode.
The indicatorappears when
the displayed unit or mode is specified.
Unit Display
The indicator is
displayed at the
units specified.
Press and hold
Press several time
Selection ke
Example
Specify the uni
Counting mode PC
Percent mode %
Density mode DS
5 Press the PRINT key to store the units or modes.
The balance displays end and then displays the next
menu item of the function table.
6 Press the CAL key to exit the function table.
Then the balance returns to the weighing mode with the
selected unit.
Store them
BM series 17
5.3. Selecting Unit And Weighing Mode
t
t
Preset the sequence of displaying units and
weighing modes to fit the frequency of use.
Refer to "5.2 . Storing The Active Units".
Pressing the MODE key in the weighing mode,
the unit and weighing mode can be displayed in
order.
BM-22 and BM-20 are preset units in the
factory. (Milligram), (Gram) is displayed
in order.
For BM-252, BM-500, BM-300 and BM-200
are preset units in the factory.
(Gram), (Milligram) is displayed in order.
Density mode
To use the density mode, it must be stored in
the function table as described on page 17.
Refer to "14. Density Measurement" for
details about the mode.
To select this mode, press the MODE key
until the processing indictor blinks with
the unit "g" displayed.
The unit sequence of the factory settings.
Turning on the display.
BM-252
BM-500
BM-20 BM-300
BM-22 BM-200
Counting mode
Percent mode
Ounce (Avoir)
Troy Ounce
Metric Cara
Momme
Pennyweigh
Grain (UK)
Tael
Tola (India)
Messghal
Density mode
18 BM series
5.3.1. Unit Setting Example
s
s
e
The example below sets the units in the order with g (gram) as the first unit followed by PC
(counting mode).
1 Press and hold the RANGE key untilba5fnc of the
function table is displayed in the weighing mode, then release
the key.
2 Press the RANGE key several times to displayUnit.
3 Press the PRINT key to enter the unit selection mode.
4 Press the RE-ZERO key to specify the unit of g.
The stabilization indicatorappears when the unit is
specified.
5 Press the RANGE key several times to display Unit.
6 Press the RE-ZERO key to specify the unit of pcs.
The stabilization indicatorappears when the unit is
specified.
7 Press the PRINT key to store the units.
The balance displays end and then displays the next
menu item of the function table.
8 Press the CAL key to exit the function table. Then the
balance returns to the weighing mode with g, the unit
selected first.
9 Press the MODE key to switch between g and PC.
Specify counting mod
Press and hold
Press several time
Specify gram
Press several time
Store them
BM series 19
6. Weighing
S
r
y
Precautions for the weighing operation
Press the RE-ZERO key each time, before placing a sample on the weighing pan, to prevent
possible errors.
Place a sample in the center of the weighing pan gently.
Temperature changes during measurement may cause weighing error.
Shorten the operation time as much as possible. (
Use a pair of tweezers to avoid a temperature change due to having your hand in the weighing
chamber.
Material with an electrostatic charge or that is magnetic may cause a weighing error.
Do not press keys with a sharp instrument (such as a pencil or ball point pen).
Do not drop things on the pan, or place a weight on the pan that is beyond the weighing range.
Calibrate your balance periodically to maintain weighing accuracy. Refer to section "8. Calibration".
Keep the area clean and dry.
Consider section "3. Precautions" for the weighing operation.
For precision weighing, keep the AC adapter connected to the balance.
6.1. Basic Operation (Gram Mode)
Opening and closing door, putting and removing sample)
Refer to section "4. Display symbols and Key operation" before operation.
Note When turning on the balance with a tare weight placed on
the pan, the balance automatically displays the zero display.
1 Turn on the balance using the ON:OFF key.
2 Select a preset unit using the MODE key, if necessary.
3 Place the tare (container) on the weighing pan, if necessary.
Press the RE-ZERO key to cancel the tare weight.
Then zero is displayed.
Tare : A vessel placed on the pan, but not
to be included in the weighing data.
Example: Container.
4 Place a sample on the pan or in the container.
Close the door.
5 Wait for the stabilization indicator to be
displayed, then read the value.
6 Remove the sample and container from the pan.
Stabilization
indicator
Zero displa
Close door
tabilization
indicato
Press the key
to select a unit,
if necessary.
Ta r e
(Container)
Weighing pan
Sample
Glass door is
not closed
Information
Remove them
20 BM series
6.2. Smart Range For BM-22
r
The smart range for BM-22 consists of the standard range and precision range (high resolution).
Smart range function
These ranges can be switched by the weighing value automatically.
When placing the tare weight within the standard range, if pressing the RE-ZERO key, the
weighing sample can be weighed within the precision range.
When pressing RANGE key, the current range is fixed to the standard range.
Example
1 Press the RE-ZERO key to use the precision range.
The balance displays zero.
2 Place the container (tare) on the weighing pan.
When mass of the container (tare) is within the
standard range, the current range is automatically
switched to the standard range.
3 Press the RE-ZERO key to use the precision range.
The balance displays zero.
4 Place the weighing sample in the container (tare).
When the sample is within the precision range,
the weighing value can be read in high resolution.
Precision range
Container (Tare)
Standard range
Precision range
Weighing matte
Precision range
Precision Range And Standard Range For BM-22
Unit Precision range Standard range
Milligram 0.000 to 5100.009 5100.01 to 22000.84
Gram 0.000000 to 5.100009 5.10001 to 22.00084
Refer to page 14 for other units.
BM series 21
6.3. Counting Mode (PC)
This is the mode to determine the number of objects in a sample based on the standard sample
unit mass. The unit mass means an average mass of the samples. The smaller the variation in
the samples, the more accurate the count will be. The balance is equipped with the Automatic
Counting Accuracy Improvement (ACAI) function to improve the counting accuracy.
Notes
Use samples with a unit mass of 1 mg or more for counting.
If the sample unit mass variable is too large, it may cause a counting error.
To improve the counting performance, use the ACAI function frequently or divide the samples into
several groups and count each group.
Selecting the counting mode
1 Press the MODE key to select the unit (counting mode).
Storing a sample unit mass (Weighing input mode)
2 Press the RANGE key to enter the sample unit mass storing mode.
3 Select the number of samples using the RANGE key.
It may be set to 10, 25, 50 or 100.
Advise
4 Place a tare (container) on the weighing pan, if necessary.
A greater number of samples will yield more accurate
counting result.
Press the RE-ZERO key to cancel the weight (tare).
The number specified in step 3 appears.
Example
: 250 is displayed if 25 is selected in step 3.
Weighing pan
Tare
(Container)
5 Place the number of samples specified on the pan.
In this example, 25 pieces.
6 Wait for the stabilization indicator to be displayed.
Press the PRINT key to calculate and store the unit mass.
Then the balance displays 25 and is set to count samples
with this unit mass. To improve the accuracy of the unit mass,
proceed to step 8.
Notes
If the balance judges that the mass of the samples is too light to
acquire accurate weighing, it displays an error requiring the
addition of more samples to the specified number.
Example: 50 - appears, requiring 25 more samples. Add
25 samples and press the PRINT key. When the unit mass is
stored correctly, the balance proceeds to the counting mode.
If the balance judges that the mass of the samples is too light (
0.0001g
The sample unit mass is stored in non-volatile memory, and is
maintained even if the AC adapter is removed.
) and can not be stored as the unit mass, it displays lo.
under
25 samples
To
Step
8 when using
the ACAI function.
Place samples
Counting Operation
7 Place the samples to be counted on the pan.
Read the result and remove them from the pan.
22 BM series
Counting result
Counting mode using the ACAI function
A
t
The ACAI is a function that improves the accuracy of the unit mass automatically by increasing
the number of samples as the counting process.
ACAI: Automatic Counting Accuracy Improvement
Proceed to step 8 after storing a unit mass at last page.
8 If a few more samples are added, the processing
indicator turns on. To prevent an error, add three or
more. The processing indicator does not turn on if
overloaded. Try to add the same number of samples
as displayed.
9 The balance re-calculates the unit mass while the
processing indicator is blinking. Do not touch the
balance or samples on the pan until the processing
indicator turns off.
10 Counting accuracy is improved when the processing
indicator turns off.
Each time the above operation is performed, a more
accurate unit mass will be obtained. There is no
definite upper limit to the ACAI range for the number
of samples exceeding 100. Try to add the same
number of samples as displayed.
11 Remove all the samples used in ACAI and proceed
with the counting operation using the improved unit
mass.
dd a few more
samples.
The mark turns on
at proper range.
The mark turns on
and off during
calculation.
The mark turns off
after the unit mass
is improved.
From
at last page
step
Repea
6
BM series 23
6.4. Percent Mode (%)
The percent mode displays the weight value in percentage compared to a 100% reference mass
and is used for target weighing or checking the sample variance.
Selecting the percent mode
1 Press the MODE key to select the unit % (Percent mode).
If the percent mode can not be selected, refer to "5. Weighing Units".
Storing the 100% reference mass
2 Press the RANGE key to enter the 100% reference mass
storing mode.
Weighing pan
3 Place a tare (container) on the weighing pan, if necessary.
Press the RE-ZERO key to cancel the weight (tare).
The balance displays 1000 % .
4 Place the sample to be set as the 100% reference mass on
the pan or in the container.
5 Press the PRINT key to store the reference mass.
The balance displays 100.00 % .
Note
The decimal point position can be changed by the 100% mass.
BM-252, BM-500
BM-300, BM-200
100% mass
0.0100 g to 0.0999 g 1 % 0.00100 g to 0.0099 g1 %
0.1000 g to 0.9999 g 0.1 % 0.01000 g to 0.0999 g0.1 %
1.0000 g to 0.01 % 0.10000 g to 0.01 %
If the balance judges that the mass of the sample is too light
(
under 0.01g
) to be used as a reference, it displays lo .
Minimum
display
BM-20, BM-22
100% mass
Minimum
display
Tare
(Container)
100% mass
A 100% reference mass can be stored in the non-volatile
memory and is maintained even if the AC adapter is removed.
6 Remove the sample.
Reading the percentage
7 Place a sample to be compared to the reference mass on
the pan. The displayed percentage is based on the 100%
reference mass.
24 BM series
Place sample
Percentage
6.5. Built-in DC Static Eliminator (Neutralization Device)
y
y
t
The BM series is equipped with four DC static eliminator electrode units to neutralize static
electricity from a charged weighing sample.
When neutralizing static electricity before weighing, stability can be improved, reducing error.
Neutralizing ion : The discharge electrodes of the built-in DC static eliminator continuously
generate bipolar ions by corona discharge. Applying these ions to the weighing
sample neutralizes the static charge.
Static electricity : In general, when the ambient humidity is less than 45%RH, powders, paper,
plastic, nonconductors, etc., easily become charged with static electricity. The
influence of the static electricity may cause a weighing error of several
milligrams. The static eliminator effectively neutralizes the electrical charge.
Operation
1 Place the sample to be weighed on the center
(of the metallic circle) of the separation plate to
neutralize the static charge.
2 Press the ION key to start the neutralization.
The mark and the LED flash.
The neutralization will stop after the preset time
of "Neutralizing time (ion)" in the function table
(the factory setting is 3 seconds).
LED
Static eliminator
Sample
Separation plate
The minimum display of BM-20, BM-22 and BM-252 is
switched to 0.1 mg while performing the neutralization.
When pressing the ION key during the neutralization,
the neutralization stops.
Note
Keep a space between electrodes and the weighing sample.
Placing the sample to close to an electrode may cause the
sample to become charged.
Remove any obstacle between electrodes and the weighing sample.
In the neutralization
Weighing displa
Maintenance Of The Electrode Unit
In general, when using the eliminator for a long time, dust and
stains may stick to the electrodes. Clean them periodically to
maintain performance.
When pins of the electrode are rubbed down and the
neutralization function does not recover after cleaning them,
replace the electrodes with four new units. The standard life
time is approximately 10000 hours.
Replacing Electrode Unit
1 Turn the electrode unit 45 degrees counterclockwise. Remove it.
Ke
Example: Display of BM-20,
BM-22, BM-252
Step
1
Step
2
Cleaning
Electrode
2 Replace all four units with new ones at the same time.
Note
Do not remove and clean the electrodes while the mark and the LED are flashing.
BM series 25
Electrode uni
Remote Control Of The DC Static Eliminator
AX-T
When specifying "on" to [AD8922 control] in the function
settings (8922 ionfnc on), the balance applies the function
of the ION key to the MODE key on the AD-8922.
When specifying "2" to [External control input ] in the function
settings (e-5 ionfnc 2) and assembling external key plug
(AX-T-314A-S) and foot switch (AX-SW128), the balance
applies the function of the ION key to the foot switch
Applies the ION key
to the MODE key on
the AD-8922A.
Example:
External key plug
(AX-T-314A-S)
Ter mi na l 1
Line 1
GND
Applies the ION key to
terminal 1of the foot
switch (AX-SW-123).
Description Of The External Key Plug (AX-T-314A-S) (Option)
External key plug
When specifying [External control input (e-5)] in [Static eliminator (ionfnc)], the following
functions can be applied to the external key plug (AX-T-314A-S) terminals.
Static eliminator External control input Terminal 1 Terminal 2
-314A-S
Ter mi na l 2
Ter mi na l 1
Terminal GND
Line 1
Line 2
GND
ionfnc e-50 PRINT key RE-ZERO keyionfnc e-51 PRINT key ION keyionfnc e-52 ION key RE-ZERO key
factory settings
26 BM series
6.6. Monitoring Clock and Environmental Sensors
y
A
ature
e
e
s
e
s
s
s
e
In the weighing mode, the built-in clock and environmental sensors (temperature, humidity, air
pressure) can be monitored and temperature data can be output with the function settings.
Specifications Of The Environmental Sensors
Sensor Resolution Measurement rangeApplicable range
Temperature ±1.5 °C 5 °C to 40 °C
Humidity ±10 % 0 % to 100 %
Air pressure ±10 hPa 300 hPa to 1100 hPa
5 °C to 40 °C
Caution
The environmental sensors are built into the case. Therefore, these values are not in accordance
with the environmental condition.
Refer to "17. Specifications" for operating condition of the product.
Operation
Press the SELECT key in the weighing
mode. The current time, date and sensor
values (temperature, humidity, air pressure)
are displayed for several seconds in order.
The balance returns to the weighing mode
automatically.
If pressing the SELECT key while
displaying current data, the display
proceeds to next data.
If pressing the CAL key while
displaying current data, the balance
returns to weighing mode.
Environmental sensor
Weighing mod
Tim
5 second
Dat
5 second
10 second
Weighing mod
Humidit
ir pressure
Temper
Note
If adjusting the built-in clock, refer to "10.7.Clock And Calendar Function".
If appending time and date to data output, refer to the [Time/Date output (5-td)] of
"10.2.Details Of The Function Table".
If appending temperature data to the data output, refer to [Temperature output (5-tp)] of
"10.2.Details Of The Function Table".
BM series 27
Loading...
+ 67 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.