A&D AD-8527 User Manual

AD-8527
r
r
y
y
y
y
C
A
Quick USB Adapter
Instruction Manual
Quick USB Adap ter
AD-8527
© 2009 A&D Company, Limited. All rights reserved.
No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of A&D Company, Limited.
The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice.
1WMPD4001958A
When operating this instrument, be sure to observe the following:
Do not open the case to repair the AD-8527. Only
qualified personnel can do that. Attempting repairs yourself may cause damage to the AD-8527. Damage caused by attempting to do the repair yourself will void the warranty
With the cap attached over the connectors, the
AD-8527 is protected against water splashing (equivalent to IP65). Please note that it will not endure immersion in water or high pressure running water.
The AD-8527 is supplied power and is turned on only
when it is connected to a weighing instrument or the USB of a personal computer.
CANCEL (+)
Using the AD-8527
Caution
.
Precautions on Usage
Caution
1. Introduction
This manual describes how the AD-8527 works and how to get the most out of it in terms of performance. Please read this manual completely before using the AD-8527.
2. Features
When connected to the RS-232C interface of a weighing instrument and to a personal computer using USB simultaneously, the AD-8527 can transmit the weighing data that the weighing instrument outputs in real time to the personal computer. This enables a personal computer without COM ports to receive data.
The AD-8527 functions similar to a USB keyboard. Thus, no software installation such as a driver is not necessary. Once connected as described above, the weighing data can be transmitted to the personal computer. Note that Command transmission from the personal computer to the weighing instrument is not available.
The AD-8527 can transmit the data to any application such as Excel, Word and Notepad for Windows.
As power is supplied to the AD-8527 from the weighing instrument or USB, no special power source is required.
The AD-8527 has no data storage function. To store data without using a personal computer, use the AD-1688 Weighing Data Logger.
Note: The AD-8527 is supplied power and is turned on only
when it is connected to a weighing instrument or the USB of a personal computer.
3. Unpacking
When unpacking, confirm that everything is contained.
AD-8527 Quick USB Adapter USB extension cable (0.5 m) RS-232C cable with a D-Sub 9-pin connector (1 m) : AX-KO3571-100 RS-232C cable with a D-Sub 25-pin connector (1 m) : AX-KO3572-100 RS-232C cable with a 7-pin DIN connector (1 m) : AX-KO3573-100 Instruction manual (this document)
ap
CANCEL
Quick USB Adapter
AD-8527
(+)
D-8527
USB extension cable
RS-232C cable with a D-Sub 9-pin connector
RS-232C cable with a D-Sub 25-pin connector
RS-232C cable with a 7-pin DIN connector
Quick USB Adapter
AD-8527
CANCEL
(+)
When the AD-8527 is in use, the cap can be attached on the bottom.
Each side has an opening. Use the openings as follows.
CANCEL
Quick USB Adapter
AD-8527
B
(+)
A: Attach a strap to carry. B: Connect the cap and the main unit
so that the cap will not be lost.
A
4. Connection and Key Operations
Quick USB Adapter
CANCEL
USB cable
RS-232C cable
AD-8527
RS-232C
(+)
Personal computer
Weighing instrument
USB connecto
RS-232C connecto
Displa
SET ke
CANCEL ke
AD-8527
(+) ke
When pressed and held
(To keep pressing
for 2 seconds or more)
Displays the function setting menus.
SET
Key
When pressed
(To press and
release immediately)
Displays the date and time.
Starts or stops data transmission
(+) None
at a set interval. (Only in the interval mode)
CANCEL None None
5. Displaying the Date and Time
At the standby display, press the SET key to display the year, date and time.
1. Standby display
2. The display indicates the year and date for 2 seconds each,
3. Press any key to return to the standby display.
OK
Press
and then indicates the time.
DATE DATE
OK
6. Setting the Date and Time
Note: The date and time were set to the Japanese
standard time at the factory when shipped. Before use, set the date and time to your local time.
Change the date and time as follows.
1. Refer to “5. Displaying the Date and Time” to display the date and time. When the year, date or time is displayed, press and hold the SET key for 2 seconds or more. The display flashes and the AD-8527 enters the setting mode.
Using the following keys to set the date and time.
(+) key: Increase the value by one CANCEL key: Decrease the value by one
When the above keys are held pressed, the value is increased or decreased continuously
SET key: Confirm the setting and go to the next item.
2. After setting the year, month, day, hour and minute, press the SET key to return to the standby display.
DATE DATE DATE
OK
7. Transmitting the Weighing Data
Transmitting data from a weighing instrument to a personal computer.
1. Connect the AD-8527 to the USB port of a personal computer and connect the AD-8527 to a weighing instrument using the RS-232C cable.
Note: Use the cable appropriate to the RS-232C
connector of the weighing instrument.
Power is supplied to the AD-8527
and the standby display appears.
If either USB or RS-232C is not
connected, “- - - -“ appears.
2. Press the PRINT key of the weighing instrument to transmit the data to the AD-8527.
Then, the AD-8527 transmits the data to
the personal computer as keyboard input signals. When transmitted, a buzzer sounds and “out” appears for about 2 seconds.
Note: If data is not input into Excel correctly, refer to
“10.2. If the Data Fails to be Input into Excel Correctly”. When transmitting GLP output and statistical results, set the weighing instrument “Data output pause (pU5e)” to “Pause (1.6 seconds)”.
3. Repeat step 2 as necessary.
4. Disconnect the weighing instrument from the AD-8527.
5. Disconnect the personal computer from the AD-8527.
Continued to the reverse side
OK
Standby display
OK
-1- -2- -3-
ID
Numb
ht
From the front
8. Function Settings
In the function setting mode, settings of the AD-8527 performance and communication can be changed. Each function has parameters. By changing the parameter, specify the performance of the function. The set parameters are maintained in memory even if the AD-8527 is disconnected from the weighing instrument or personal computer.
8.1. Setting Procedure
1. At the standby display, press and
hold the SET key for 2 seconds or more.
2. The function display appears. (Refer to “8.2. Function List”.) Press the (+) key to select a function, and press the SET key to
display the parameters.
When the CANCEL key is pressed,
or after 10 seconds of non operation, the AD-8527 returns to the standby display.
3. The parameter of the selected function appears. The current setting has the lower right. The flashing parameter can be changed.
4. Using the following keys to change the parameter.
(+) key: Display the parameters one by one CANCEL key: Go to the next function without changes. SET key: Confirm the setting and go to the next
After changes, a buzzer sounds, the parameter and the appear for 1 second, and then the next function appears.
5. Press the CANCEL key to return to the standby display.
mark at the
function.
mark
OK
Press and hold
Press
Current setting
OK
8.2. Function List
Baud rate
Function
Parameter 600, 1200, 2400*, 4800, 9600, 19200 bps
Description
Data bit, parity
Parameter 7bit-Even*, 7bit-Odd, 8bit-None
Number of data on a line
Parameter 1*,2, 3, 4, 5, 6, 7
Buzzer
Parameter
Sounds the buzzer*, Does not sound
Data transmission mode
Parameter
Standard mode (All data)
*, Interval mode
Data to transmit
Interval time
Parameter 1*,2,5,10,15,20,30,60,120,300,600,900,1200,1800,3600
Parameter Weight values only*, All data (with a header and unit)
When the interval mode is selected
SEC SEC
Date and time data
DATE
SEC
Parameter
Adds the date and time*, Does not add
DATE
SEC
DATE
SEC
Date format
DATE
(Note1)
Parameter YMD*, MDY, DMY
DATE DATE DATE
(Note 1) Set this function according to the setting of the personal
computer used.
*: Factory setting
About the number of data on a line, interval mode, data to transmit, date and time data, refer to the following sections
.
8.3. Data to Transmit
Adding the date and time
The date and time when the AD-8527 receives the data can be transmitted to a personal computer. The date, time and weight are separated by tabs.
e.g.: 6/10/2009<Tab>10:41:24<Tab>+00123.45<Enter>
Example below is when the data is transmitted to Excel
Selecting data to transmit
Two ways of transmitting data are available; weight values only or all data. Transmitting weight values only is convenient when performing statistical calculations using Excel. When performing GLP output, transmitting all data is required. For example, the received data is “ST,+00123.45 g”. When transmitting weight values only is selected, “+00123.45” will be transmitted. When transmitting all data is selected, “ST,+00123.45 g” will be transmitted.
Number of data on a line
This function is useful for storing data with different contents or items at the same time, for example, ID and number of samples in a counting scale. When the number is set, <CR>s of up to the setting are converted into commas. Once the number of <CR>s is reached the setting, a new line starts. For example, the ID, number of samples and weight are sent individually. Using this function, these three data items can be placed on one line as shown in (B).
Example data received to transmit to Excel ID,000123<CR><LF> QT,+00000056 PC<CR><LF> ST,+0003.102 g<CR><LF>
(A) When the number of data on a line is 1, the ID, number
of samples and weight are input on separate lines.
(B) When the number of data on a line is 3, the ID, number
of samples and weight are input on one line.
samples
er of Weig
9. Interval Mode
In the interval mode, the weighing data is transmitted at a certain interval.
Available intervals are:
1, 2, 5, 10, 15, 20, 30 seconds / 1, 2, 5, 10, 15, 30 minutes / 1 hour
1. Connect the AD-8527 to a personal computer using USB.
2. Connect the AD-8527 to a weighing instrument using the RS-232C cable.
3. Select the interval mode (int) for the data transmission mode (out) of the function setting and set the interval time (120 seconds in the example).
Press the CANCEL key to return to the
standby display. In the interval mode, the
mark turns
on at the lower right.
4. Set the data output mode of the weighing instrument to the stream mode.
For details on the setting procedure, refer to the instruction
manual of the weighing instrument.
5. Press and hold the (+) key for 2 seconds or more. The AD-8527 starts to transmit the data at a set interval. “out” is displayed and the mark flashes while the data is transmitted.
6. To stop transmitting the data, press and hold the (+) key. The
mark stops flashing and turns on.
SEC
OK
OK
OK
10. Errors
10.1. Error Display
Error Description Remedy
The supplied voltage is low.
The baud rate and parity of the received data is different from the set values.
Confirm that the USB is connected correctly.
Set the same values for both the AD-8527 and the weighing instrument.
Internal clock error such as low backup battery
Ask for repair.
10.2. If the Data Fails to be Input into Excel Correctly
If data is not input into Excel correctly when using Windows XP, do as follows.
1. Open “Control Panel”, and then “Regional and Language Options”.
2. Select the Languages tab and click on the Details button.
3. Select the Advanced tab and check the Turn off advanced text servises box.
11. Specifications
Devices Balances and scales with an to be connected
Applicable OS Windows 2000 / XP / Vista Clock error ±1 minute / month Dedicated power supply Not required
Operating temperature 5°C to 40°C Operating humidity 85%RH or less, no condensation Dimensions 55x103x16.5 mm Mass Approximately 60 g